Intrinsically Safe Pressure Gauge Model AI DESCRIPTION As an in-process gauge or portable calibrator, the hand-held Model AI delivers long-lasting performance with enhanced accu- racy. Accuracy of 0.05 % coupled with an enhanced resolution, 4½-digit display makes this unit an excellent transfer standard for calibrating pressure measuring equipment. The AI is rugged and features high/low pressure capture with a field-adjustable update rate from three-to-ten times per sec- ond. These units are manufactured from stainless steel and use sensor technology which has no moving parts, requires little recalibration and has excellent overpressure tolerance. NEMA 4 construction provides EMI and RFI protection. Units of measure are field-selectable, and the unit is offered with gage, absolute, vacuum, or compound reference. The Model AIW is powered by one or two common 9 V alkaline batteries. The Model AIV is powered by an 11 Vdc to 32 Vdc power supply. The AI Digital Gauges have the following complete Factory Mu- tual Approved Classifications: Class I: Gases and vapor Class II: Combustible dust Class III: Fibers Intrinsically safe systems limit the thermal and electrical energy to the hazardous area to help prevent ignition. Division 1: present or likely to be present in normal operation Division 2: not present in normal operation Group A: Acetylene Group B: Hydrogen Group C: Acetaldehyde, Ethylene, Methyl Ether Group D: Acetone, Gasoline, Methanol, Propane Group E: Metal dust Group F: Carbon dust Group G: Grain dust FEATURES High accuracy: 0.05 % full scale Factory Mutual approved Ranges to 10000 psi - gage, absolute, vacuum, or com- pound Stainless steel construction 4½ digit display with 0.5 in height On/off switch Optional handles Optional carrying case and panel mounting ring 9 V battery power High/low capture and store Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Upload: others

Post on 23-Jan-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Intrinsically Safe Pressure GaugeModel AI

DESCRIPTIONAs an in-process gauge or portable calibrator, the hand-held

Model AI delivers long-lasting performance with enhanced accu-

racy. Accuracy of 0.05 % coupled with an enhanced resolution,

4½-digit display makes this unit an excellent transfer standard

for calibrating pressure measuring equipment.

The AI is rugged and features high/low pressure capture with a

field-adjustable update rate from three-to-ten times per sec-

ond. These units are manufactured from stainless steel and use

sensor technology which has no moving parts, requires little

recalibration and has excellent overpressure tolerance. NEMA 4

construction provides EMI and RFI protection. Units of measure

are field-selectable, and the unit is offered with gage, absolute,

vacuum, or compound reference.

The Model AIW is powered by one or two common 9 V alkaline

batteries. The Model AIV is powered by an 11 Vdc to 32 Vdc

power supply.

The AI Digital Gauges have the following complete Factory Mu-

tual Approved Classifications:

Class I: Gases and vapor

Class II: Combustible dust

Class III: Fibers

Intrinsically safe systems limit the thermal and electrical energy

to the hazardous area to help prevent ignition.

Division 1: present or likely to be present in normal operation

Division 2: not present in normal operation

Group A: Acetylene

Group B: Hydrogen

Group C: Acetaldehyde, Ethylene, Methyl Ether

Group D: Acetone, Gasoline, Methanol, Propane

Group E: Metal dust

Group F: Carbon dust

Group G: Grain dust

FEATURES• Highaccuracy:0.05%fullscale

• FactoryMutualapproved

• Rangesto10000psi-gage,absolute,vacuum,orcom-

pound

• Stainlesssteelconstruction

• 4½digitdisplaywith0.5inheight

• On/offswitch

• Optionalhandles

• Optionalcarryingcaseandpanelmountingring

• 9Vbatterypower

• High/lowcaptureandstore

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 2: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model AI

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Accuracy 0.05%offullscaleBSFL

Pressure range, 0 psi to: 15, 50, 100, 200, 500, 750, 1000, 1500, 2000, 5000, 7500, 10000 psi

High and low capture Standard

Update speed Adjustable from 3 to 10 per sec

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating(standard)

-1 °C to 71 °C [30 °F to 160 °F]

Temperature, operating(optional)

-18 °C to 82 °C [0 °F to 180 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA 4 (also provides temperature range of -18 °C to 82 °C [0 °F to 180 °F])

Power, Model AIW Oneortwo9Valkalinebatteries(in-cluded)

Power, Model AIV 11 Vdc to 32 Vdc @ 100 mA (3 ft cable included)

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4 LCD digits - 12,7 mm [0.5 in] high

Pressure port 9/16-18 UNF-2B - female straight thread with SAE spec J514 o-ring boss

Wetted parts Stainless steel

Case material Stainless steel

Low battery indication Standard

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi)

Maximum safe over pressure (psi)1

Incremental display steps

0 psi to 15 psi = 15

75 0.005

0 psi to 50 psi = 50

250 0.01

0 psi to 100 psi = 100

500 0.02

0 psi to 200 psi = 200

1000 0.05

0 psi to 500 psi = 500

1500 0.1

0 psi to 750 psi = 750

1500 0.2

0 psi to 1000 psi = 1K

2000 0.2

0 psi to 1500 psi = 1.5K

3000 0.5

0 psi to 2000 psi = 2K

4000 0.5

0 psi to 5000 psi = 5K

7500 1

0 psi to 7500 psi = 7.5K

12000 2

0 psi to 10000 psi = 10K

15000 2

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 3: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

SensingandControl

AutomationandControlSolutions

Honeywell

1985 Douglas Drive North

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008688-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model AI Intrinsically Safe Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specificationsmaychangewithoutnotice.Theinformationwesupply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

FormoreinformationaboutSensingandControlproducts,visitwww.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

Notes

1.Thelimit/relaymodelsincludetwoLEDstatusindicatorsonfrontfaceand two form C relays (normally open, common, normally closed) that are rated at a maximum 24 Vdc/Vac at 1 A or 48 Vdc/Vac at 1/2 A.

ORDER gUIDE

W = Battery powered (9 V alkaline)*V = 11 Vdc to 32 Vdc (3 ft. cable)

* = Example

SERIESAI

AI W 200 G

POWER

PRESSURE RANGE (PSI)

REFERENCEG = Gage (std) reads zero at atmosphere *A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt) scaled in In Hg (15 psi or lower)C = Compound (opt) reads both positive & negative

(vacuum) pressure

ELECTRICAL CONNECTION

Z

15 = 0-1550 = 0-50100 = 0-100200 = 0-200*

500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20005K = 0-50007.5K = 0-750010K = 0-10000

Z = No electrical connection (Model AIW - 9 V battery)*C = 3 foot cable (Model AIV)Call factory for availability of other electrical connections and cable lengths.

Each gauge has the following scaling built in for field selection:FIELD SELECTABLE STANDARD UNITS OF MEASURE

Bar = BarmBar = MillibarkPa = KilopascalsMPa = Megapascals

Ft H20 = Feet of waterIn H20 = Inches of waterIn Hg = Inches of mercurymmHg = mm of mercury

Consult factory for other units of measure not listed

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 4: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Digital Pressure Test GaugeModel AK

DESCRIPTIONThe hand-held Model AK is an in-progress gauge or por-

table calibrator and delivers long-lasting, enhanced accuracy

performance. Accuracy of 0.05 % coupled with an enhanced

resolution, 4½-digit display makes this unit an excellent transfer

standard for calibrating pressure measuring equipment.

The AK is rugged and features high/low pressure capture with

a field-adjustable update rate from three to 10 times per sec-

ond. These units are manufactured from stainless steel and use

sensor technology which has no moving parts, requires little

recalibration and has excellent overpressure tolerance. Optional

NEMA 4 rating extends the operating temperature to -18 °C to

82 °C [0 °F to 180 °F]. Units of measure are field-selectable and

the unit is offered with gage, absolute, vacuum, or compound

reference.

* with optional NEMA 4 rating

FEATURES• Highaccuracy:0.05%fullscale

• 0°Fto180°Foperatingtemperature*

• Enhancedresolution

• Stainlesssteelconstruction

• Field-selectableunitsofmeasure

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• On/Offswitch

• CEcompliant(certainmodels,consultfactory)

• NEMA4rating(optional)

• Optionalcarryingcaseandpanelmountingring

• Optionalpressureports

• NISTtraceable(optional)

• 9Vbatterypower

• High/lowcaptureandstore

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 5: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model AK

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Accuracy 0.05%offullscaleBFSL

Pressure ranges 0 psi to 15, 50, 100, 200, 500, 750, 1000, 1500, 2000, 5000, 7500, 10000 psi

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating (standard)

-1 °C to 71 °C [30 °F to 160 °F]

Temperature, operating (optional)

-18°Cto82°C[0°Fto180°F](NEMA4optionrequired)

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA2(NEMA4optional,exceptAKT)

Display 4½LCDdigits-12,7mm[0.5in]high

Power, Model AKW Oneortwo9Valkalinebatteries(in-cluded)

Power, Model AKT 110Vacadapter@60Hz(included)

Power,ModelAKV 11Vdcto32Vdc@100mA(3ftcableincluded)

MECHANICAL SPECIFICATIONS

Characteristic Measure

Physical diameter 93,98mm[3.7in]

Pressure port 9/16-18UNF-2B,femalestraightthreadwithSAEspecJ5140-ringboss

Wetted parts Stainlesssteel

Case material Stainlesssteel

Lowbatteryindication Standard

Highandlowcapture Standard

Update speed Adjustable from 3 to 10 per second

Face membrane Tactile feedback raised memory chips

Calibration data Storedonmemorychip

W = Battery Powered (9 V alkaline)*T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)

ORDER CODE* = Example

SERIESAK*

AK W 200 G

POWER

PRESSURE RANGE (PSI)15 = 0-1550 = 0-50100 = 0-100200 = 0-200*

500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20005K = 0-50007.5K = 0-750010K = 0-10000

REFERENCEG = Gage (std) reads zero at atmosphere *A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt) scaled in In Hg (15 psi or lower)C = Compound (opt) reads both positive & negative

(vacuum) pressure

ELECTRICAL CONNECTIONZ = No electrical connection (Model AKW - 9 V battery)*C = 3 foot cable (Model AKV)T = ac adapter (Model AKT)

Call factory for availability of other electrical connections & cable lengths.

Z

OPTIONAL4 = NEMA 4 (Not available on AKT)

NEMA 4 option also provides temperature range of 0-180°F

4

Each gauge has the following scaling built in for field selection:FIELD SELECTABLE STANDARD UNITS OF MEASURE

Bar = BarmBar = MillibarkPa = KilopascalsMPa = Megapascals

Ft H20 = Feet of waterIn H20 = Inches of waterIn Hg = Inches of mercurymmHg = mm of mercury

Consult factory for other units of measure not listed

ORDER CODES

ADDITIONAL SPECIFICATIONS

Pressurerange(psi) Maximum safe over pressure(psi)1

Incrementaldisplaysteps

0 psi to 15 psi = 15 75 0.005

0 psi to 50 psi = 50 250 0.01

0 psi to 100 psi = 100 500 0.02

0 psi to 200 psi = 200 1000 0.05

0 psi to 500 psi = 500 1500 0.1

0 psi to 750 psi = 750 1500 0.2

0 psi to 1000 psi = 1K 2000 0.2

0 psi to 1500 psi = 1.5K 3000 0.5

0 psi to 2000 psi = 2K 4000 0.5

0 psi to 5000 psi = 5K 7500 1

0 psi to 7500 psi = 7.5K 12000 2

0 psi to 10000 psi = 10K 15000 2

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 6: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

SensingandControl

AutomationandControlSolutions

Honeywell

1985DouglasDriveNorth

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008689-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model AK Digital Pressure Test Gauge

Warranty. Honeywellwarrantsgoodsofitsmanufactureasbeing free of defective materials and faulty workmanship. Honeywell’sstandardproductwarrantyappliesunlessagreedtootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specificwarrantydetails.IfwarrantedgoodsarereturnedtoHoneywellduringtheperiodofcoverage,Honeywellwillrepairor replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literatureandtheHoneywellwebsite,itisuptothecustomertodetermine the suitability of the product in the application.

Specificationsmaychangewithoutnotice.Theinformationwesupply is believed to be accurate and reliable as of this printing. However,weassumenoresponsibilityforitsuse.

FormoreinformationaboutSensingandControlproducts,visitwww.honeywell.com/sensing or call [email protected]

WARNINGPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINGMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

Notes

1.Urefaciblafeuipitnostinvullamzzril2.Urefaciblafeuipitnostinvullamzzrildolestoconsentnisdelipsusci-

lis nismodiamcor sustinit am eugait velestinit ut nibh elisim eugait nim iril ut del duisi.

3.Urefaciblafeuipitnostinvullamzzrildolestoconsentnisdel4.Urefaciblafeuipitnostinvullamzzrildolestoconsentnisdel5.Urefaciblafeuipitnostinvullamzzrildolestoconsentnisdel

6.Urefaciblafeuipitnostinvullamzzrildolestoconsentnisdelipsusci-lis nismodiamcor sustinit am eugait velestinit ut nibh elisim eugait nim iril ut del duisi.

MOUNTING DIMENSIONS AND CHARACTERISTICS

For reference only

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 7: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Model JA

DESCRIPTIONThe Series JA digital pressure test gauge with 0.1 % full scale

accuracy uses transducer technology and a stainless steel

diaphragm for high overpressure protection. The transducer

technology provides enhanced accuracy over the entire pres-

sure range. The JA has no moving parts and thus, may provide

a long life with fewer re-calibrations. The stainless steel NEMA

construction also provides EMI and RFI protection. The JA

provides high resolution with an easy to read digital display.

There are no operator errors due to interpolation of hash marks

or parallax errors.

The Model JAB provides a 4 mA to 20 mA, two-wire output.

The Model JAE provides a 0 Vdc to 5 Vdc output. The Model

JAR has two programmable limits and relays with no analog

output. The Model JAX provides a 0 Vdc to 5 Vdc output with

two programmable limits and relays for process control or alarm

indication. The Model JAW is powered by one or two common

9 V alkaline batteries. The Model JAT is powered by a 110 Vac

adapter. The Model JAV is powered by an 11 Vdc to 32 Vdc

power supply.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• 4mAto20mA(two-wire)and0Vdcto5Vdcanalogout-

puts

• Twoprogrammablelimitsandrelays(optional)

• Highaccuracy-0.1%fullscale

• Highandlowdetection-Microprocessorbased

• Rangesto10000psi,gageorabsolute

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• Zerooffset/tare

• On/offswitchdisablefeature

• Twoprogrammablelimitsandrelays(optional)

• NEMA4rating(optional)

• Optionalcarryingcaseandpanelmountingring

• NISTtraceable(optional)

High Accuracy Digital Pressure Gauge

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 8: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JA

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.1%fullscale(betterthantestgaugeaccuracy)

Pressure range 0 psi to 10, 20, 50, 100, 200, 500, 1000, 2000, 5000 and 10000 psi

Highandlowcapture Standard

Update speed 3 times per sec

Zero and span signal adjustment

Standard: Models JAB, JAE, JAX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA2(optionalNEMA4)

Power, Model JAW Oneortwo9Valkalinebatteries(in-cluded)

Power, Model JAT 110Vacadapter@60Hz(included)

Power, Model JAV 11Vdcto32Vdc@100mA(3ftcableincluded)

Power, Model JAB 11Vdcto32Vdc(dependingonloopresistance)@20mA

Power, Model JAE 11 Vdc to 32 Vdc @ 100 mA

Power, Model JAR 11 Vdc to 32 Vdc @ 100 mA

Power, Model JAX 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 3 ft cable standard: Models JAB, JAE, JAR, JAX, JAV

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4½LCDdigits-12,7mm[0.5in]high

Pressure port 1/4-18NPTmale(200psiandbelow)

1/4-18 UNF 2B female straight thread withSAEspecJ514o-ringboss(500psiandabove)

Wetted parts Stainless steel

Case material Stainless steel

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

Low battery indication Standard: Model JAW

Limits and relays Models JAR and JAX1

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi)

Maximum safe over pressure (psi)2

Incremental display steps

0 psi to 10 psi = 10 50 0.01

0 psi to 20 psi = 20 100 0.01

0 psi to 50 psi = 50 250 0.05

0 psi to 100 psi = 100

500 0.1

0 psi to 200 psi = 200

1000 0.1

0 psi to 500 psi = 500

1500 0.5

0 psi to 1000 psi = 1K

2000 1

0 psi to 2000 psi = 2K

4000 1

0 psi to 5000 psi = 5K

7500 5

0 psi to 10000 psi = 10K

15000 5

2 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 9: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008690-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JA High Accuracy Digital Pressure Gauge

Warranty. Honeywellwarrantsgoodsofitsmanufactureasbeing free of defective materials and faulty workmanship. Honeywell’sstandardproductwarrantyappliesunlessagreedtootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywellduringtheperiodofcoverage,Honeywellwillrepairor replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literatureandtheHoneywellwebsite,itisuptothecustomertodetermine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However,weassumenoresponsibilityforitsuse.

For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

ORDER gUIDE

* = ExampleZG1KWJA ( )

The following order code is used ONLY if you want this option.

4 = NEMA 4 (not available on Model JAT)

G = Gage (std) reads zero at atmosphere*A = Absolute (opt) reads zero at vacuum(15 psi or higher)

ELECTRICAL CONNECTION

PRESSURE RANGE (PSI)

Models with Output SignalB = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = 2 programmable limits & relaysX = 0 Vdc to 5 Vdc with two limits & relays

Models with No Output SignalW = Battery powered (9 V alkaline) *T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)

MODEL DESIGNATION

SERIESJA = High accuracy 0.1% full scale*

0-10 =100-20 =200-50 =500-100=100

REFERENCE

0-200 = 2000-500 = 5000-1000 = 1K*0-2000 = 2000

0-5000 = 5K0-10000 = 10K

This ordercode is usedONLY if you

wantNEMA 4Rating.

Z = No electrical connection (Model JAW - 9 volt battery)*C = 3 foot cableT = ac adapter (Model JAT)B = Bendix (opt) PTIH-10-6P or equal (connector sold separately)Call factory for availability of other electrical connections and cablelengths.

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercuryConsult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

NEMA 4 CONSTRUCTION

NOTES1. The limit/relay models include two LED status indicators on front

faceandtwoformCrelays(normallyopen,common,normallyclosed)thatareratedatamaximum24Vdc/Vacat1Aor48Vdc/Vac at 1/2 A.

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 10: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Differential Digital Pressure GaugeModel JD

FEATURES• Ranges 1 psi to 10000 psi differential

• Field selectable units of measure

• 4 mA to 20 mA (two-wire) or 0 Vdc to

5 Vdc analog outputs (optional)

• Two programmable limits and relays (optional)

• High and low detection with microprocessor memory

• 4½ digit display with 0.5 in height

• Accuracy 0.2 % full scale

• Optional NEMA 4 rating

• NIST traceable

• CE compliant (certain models, consult factory)

• On/off switch with auto-off feature

DESCRIPTIONThe hand-held Model JD wet/wet differential pressure gauge de-

livers long-lasting, enhanced accuracy performance. Accuracy

of 0.2 % coupled with an enhanced resolution, 4½-digit display

makes this unit an excellent transfer standard for calibrating

pressure measuring equipment.

The JD is rugged and features high/low pressure capture with an

update rate of three times per second. These units are manufac-

tured from stainless steel and use sensor technology which has

no moving parts, requires little recalibration, and has excellent

overpressure tolerance.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 11: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2 Honeywell • Sensing and Control

Model JD

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.2 % full scale (better than test gage accuracy)

Pressure range 0 psid to 1, 5, 5, 10, 15, 25, 50, 75, 100, 150, 200, 300, 500, 750, 1000, 1500, 2000, 3000, 5000, 7500, 10000 psid

High and low capture Standard

Update speed 3 times per sec

Zero and span signal adjustment

Standard: Models JDB, JDE, JDX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA 2 (optional NEMA 4)

Power, Model JDW One or two 9 V alkaline batteries (included)

Power, Model JDT 110 Vac adapter @ 60 Hz (included)

Power, Model JDV 11 Vdc to 32 Vdc @ 100 mA (3 ft cable included)

Power, Model JDB 11 Vdc to 32 Vdc (depending on loop resistance) @ 20 mA

Power, Model JDE 11 Vdc to 32 Vdc @ 100 mA

Power, Model JDR 11 Vdc to 32 Vdc @ 100 mA

Power, Model JDX 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 3 ft cable (standard) on Models: JDB, JDE, JDR, JDX, JDV

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4½ LCD digits - 12,7 mm [0.5 in] high

Pressure port 1/8-27 NPT female on rages to 750 psid; 1/4-18 NPT female on ranges 1000 psid and above

Wetted surfaces Stainless steel

O-ring material 1 psid to 750 psid - Viton

1000 psid to 10000 psid - Teflon-coated stainless steel

Housing Stainless steel

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

Low battery indication Standard on Model JDW

Limits and relays Models JDR and JDX

ACCESSORIES ORDER CODES

Order code Accessory

AD12 Pressure adapter (Zinc plated steel) 1/8-27 NPTM by 1/4-18 NPTF*

AD5 Pressure adapter (Zinc plated steel) 1/4-18 NPTM to 1/4-18 NPTM*

AD2 Pressure adapter (Zinc plated steel) 1/4-18 NPTM to 7/16-20 UNF F*

AD3 Pressure adapter (Zinc plated steel) 1/4-18 NPTM to 7/16-20 UNFM*

NISTCERTS NIST certification (in addition to NIST statement)

* Stainless steel available

ADDITIONAL SPECIFICATIONS

Pressure range (psi)

Maximum line pressure (psi)1

Maximum safe over pressure (psi)2

Incremental display steps (psi) - positive only

0 psi to 1 psi 1500 1500 0.001

0 psi to 2 psi 1500 1500 0.001

0 psi to 5 psi 1500 1500 0.005

0 psi to 10 psi 1500 1500 0.01

0 psi to 15 psi 1500 1500 0.01

0 psi to 25 psi 1500 1500 0.02

0 psi to 50 psi 1500 1500 0.05

0 psi to 75 psi 1500 1500 0.05

0 psi to 100 psi 1500 1500 0.1

0 psi to 150 psi 1500 1500 0.1

0 psi to 200 psi 1500 1500 0.1

0 psi to 300 psi 1500 1500 0.2

0 psi to 500 psi 1500 1500 0.5

0 psi to 750 psi 1500 1500 0.5

0 psi to 1000 psi 1500 1500 1

0 psi to 1500 psi 3500 2250 1

0 psi to 2000 psi 4000 3000 1

0 psi to 3000 psi 5000 4500 2

0 psi to 5000 psi 7000 7000 5

0 psi to 7500 psi 9500 9500 5

0 psi to 10000 psi

12000 12000 5

1 Maximum line pressure is the amount of pressure that can be placed on both sides of the diaphragm

2 Maximum safe overpressure is the pressure which the unit can experi-ence occasionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 12: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Honeywell • Sensing and Control 3

Differential Digital Pressure Gauge

For reference only

MOUNTINg DIMENSIONS AND ChARACTERISTICS

DRAIN PORT2X, 1/8-27 NPT

↑↑

→3.0

3.5

3.7

5.1

← 2.25↑

MOUNTING HOLESTYPICAL EACH

SIDE1/4-28 UNF .3125"

PRESSURE PORT2X, 1/8-27 NPT

Request certified drawing before designingmountings or fixtures. Specifications subject to

change without notice.

3.7

3.0

5.1

← →1.82← →

DRAIN PORT2X, 1/8-27 NPT

↑ PRESSURE PORT2X, 1/8-27 NPT

MOUNTING HOLESTYPICAL EACH

SIDE1/4-28 UNF .3125 in

2.5

↓↑

3.7

5.1

2.25

← 2.25 →←→

2X, .40

OPTIONAL 3 FEETCABLE

PRESSURE PORT2X, 1/4-18 NPT

FEMALE

1 TO 25PSID

50 TO750 PSID

1000 TO10000PSID

allpressure

ranges

→← 2.2

→← 2.2

→← 2.2

ORDER gUIDE

D = Differential unit scaled from 0 psi to positive full scalevalue on the positive pressure side.*

MODEL DESIGNATION

* = Example

SERIESJD*

WJD ZD5

Models with output sIgnalB = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = 2 programmable limits & relaysX = 0 Vdc to 5 Vdc with two limits & relaysModels with no output sIgnalW = Battery powered (9 V alkaline)*T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)PRESSURE RANGE (PSID)

REFERENCE

Bi-directional scaling is available as an option.Please consult the factory.

1K = 0-10001.5K = 0-15002K = 0-20003K = 0-30005K = 0-50007.5K = 0-750010K =0-10000

75 = 0-75100= 0-100150= 0-150200= 0-200300= 0-300500= 0-500750 = 0-750

1 = 0-12 = 0-25 = 0-5*10 = 0-1015 = 0-1525 = 0-2550 = 0-50

ELECTRICAL CONNECTIONZ = No electrical connection (Model JDW - 9 V battery)*C = 3 foot cable (Models JDB, JDE, JDR, JDX & JDV)T = ac adapter (Model JDT)B = Bendix (opt) PTIH-10-6P or equal (connector sold separately)Call factory for availability of other electrical connections and cable lengths.

NEMA 4 CONSTRUCTION4 = NEMA 4 (not available on Model JDT)

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercuryConsult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

This ordercode is

used ONLYif you want

NEMA 4rating

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 13: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008714-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JD Differential Digital Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

NOTES1. The limit/relay models include two LED status indicators on front

face and two form C relays (normally open, common, normally closed) that are rated at a maximum 24 Vdc/Vac at 1 A or 48 Vdc/Vac at 1/2 A.

WARNINgPERSONAL INJURY

• DO NOT USE these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

• The information presented in this catalogue is for reference only. DO NOT USE this document as product installation information.

• Complete installation, operation and maintenance information is provided in the instructions supplied with each product.

Failure to comply with these instructions could result in death or serious injury.

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 14: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Differential Pressure GaugeModel JG

DESCRIPTIONThe hand-held Model JG wet/wet differential pressure gauge de-

livers long-lasting, enhanced accuracy performance. Accuracy

of 0.2 % coupled with an enhanced resolution, 4½-digit display

makes this unit an excellent transfer standard for calibrating

pressure measuring equipment.

The JG is rugged and features high/low pressure capture with an

update rate of three times per second. These units are manufac-

tured from stainless steel and use sensor technology which has

no moving parts, requires little recalibration, and has excellent

overpressure tolerance.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• High accuracy: 0.2 % full scale

• -1 °C to 71 °C [30 °F to 160 °F] operating temperature

• Enhanced resolution

• Stainless steel construction

• On/off switch

• CE compliant (certain models, consult factory)

• 4½ digit display with 0.5 in height

• Optional carrying case and panel mounting ring

• Optional NEMA 4 rating

• 9 V battery power

• High/low capture and store

• Four digit display

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 15: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2 Honeywell • Sensing and Control

Model JG

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Accuracy 0.2 % full scale

Maximum line pressure 500 psid

High and low capture Standard

Update speed 3 times per sec

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA 2

Power, Model JGW One or two 9 V alkaline batteries (included)

Power, Model JGT 110 Vac adapter @ 60 Hz (included)

Power, Model JGV 11 Vdc to 32 Vdc @ 100 mA (3 ft cable included)

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4 LCD digits - 12,7 mm [0.5 in] high

Pressure port 1/4-18 NPT male

Wetted parts Stainless steel

Case material Stainless steel

Low battery indication Standard

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

ORDER gUIDE

W = Battery powered (9 V alkaline)*T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)B = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = Two programmable limits & relaysX = 0 Vdc to 5 Vdc with limits & relays

* = Example

SERIESJG

JG W 5 D

POWER

ELECTRICAL CONNECTIONZ = No electrical connection (Model JGW - 9 V battery)*C = 3 foot cable (Model JGV)T = ac adapter (Model JGT)

Call factory for availability of other electrical connections and cable lengths.

Z

OPTIONAL4 = NEMA 4NEMA 4 option also provides temperature range of 0-180°F

Each gauge has the following scaling built in for field selection:OPTIONAL UNITS OF MEASURE

Bar = BarmBar = MillibarkPa = KilopascalsMPa = Megapascals

Ft H20 = Feet of waterIn H20 = Inches of waterIn Hg = Inches of mercurymmHg = mm of mercury

Consult factory for other units of measure not listed

PRESSURE RANGE (PSI)1 = 0-1(opt)5 = 0-5(opt)15 = 0-1530 = 0-3050 = 0-50100 = 0-100

200 = 0-200*300 = 0-300500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20003K = 0-30005K = 0-50007.5K = 0-750010K = 0-10000

REFERENCED = Differential

MOUNTINg DIMENSIONS AND CHARACTERISTICS

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 16: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008713-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JG Differential Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

• DO NOT USE these products as safety or emergency stop devices or in any other application where failure of the product could result in personal injury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

• The information presented in this catalogue is for reference only. DO NOT USE this document as product installation information.

• Complete installation, operation and maintenance information is provided in the instructions supplied with each product.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. The limit/relay models include two LED status indicators on front face

anåçd two form C relays (normally open, common, normally closed) that are rated at a maximum 24 Vdc/Vac at 1 A or 48 Vdc/Vac at 1/2 A.

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Maximum safe over pressure (psi)1

Model Jg

0 psi to 1 psi = 1 (opt) 10 0.001

0 psi to 5 psi = 5 (opt) 25 0.005

0 psi to 15 psi = 15 75 0.01

0 psi to 30 psi = 30 150 0.02

0 psi to 50 psi = 50 250 0.05

0 psi to 100 psi = 100 500 0.1

0 psi to 200 psi = 200 1000 0.1

0 psi to 300 psi = 300 1200 0.2

0 psi to 500 psi = 500 1500 0.5

0 psi to 750 psi = 750 1500 0.5

0 psi to 1000 psi = 1K 2000 1

0 psi to 1500 psi = 1.5K 3000 1

0 psi to 2000 psi = 2K 4000 1

0 psi to 3000 psi = 3K 6000 2

0 psi to 5000 psi = 5K 7500 5

0 psi to 7500 psi = 7.5K 12000 5

0 psi to 10000 psi = 10K 15000 51 Maximum safe overpressure is the pressure which the unit can experience occasionally without the loss of accuracy or permanent damage

ACCESSORIES ORDER CODES

Stainless steel

Zinc-plated steel

Accessory

SS6 AD6 Pressure adapter (Zinc plated steel) to 1/4-18 NPT female*

SS7 AD7 Pressure adapter (Zinc plated steel) to 1/4-18 NPT male*

SS9 AD9 Pressure adapter (Zinc plated steel) to 7/16-20 UNF male with 37° flare*

CC2 Carrying case

MR2 Panel mounting ring

NISTCERTS NIST certification (in addition to NIST statement)

* Call factory for other fittings

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 17: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Model JH

DESCRIPTION

Using high pressure autoclave fittings, the Series JH digital pres-

sure test gauge with 0.2 % full scale accuracy uses transducer

technology and a stainless steel diaphragm for high over pres-

sure protection. The transducer technology provides enhanced

accuracy over the entire pressure range. The JH has no moving

parts and thus, may provide a long life with fewer re-calibrations.

The stainless steel NEMA construction also provides EMI and

RFI protection.

The JH provides high resolution with an easy to read digital

display. There are no operator errors due to interpolation of hash

marks or parallax errors. The units can be scaled to read in

various engineering units. PSI pressures are read x 10 with the

4½-digit display. The Model JHB provides a 4 mA to 20 mA, two-

wire output. The Model JHE provides a 0 Vdc to 5 Vdc output.

The Model JHR has two programmable limits and relays with no

analog output. The Model JHX provides a 0 Vdc to 5 Vdc output

with two programmable limits and relays for process control

or alarm indication. The Model JHW is powered by one or two

common 9 V alkaline batteries. The Model JHT is powered by a

110 Vac adapter. The JHV, JHB, JHE, JHR and JHX models are

powered by an 11 Vdc to 32 Vdc power supply.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

deactivated.

FEATURES• Pressurerangesfrom15000psito60000psigage

• Twoprogrammablelimitsandrelays(optional)

• Accuracy-0.2%fullscale

• Highandlowdetection-Microprocessorbased

• 4mAto20mA(two-wire)and0Vdcto5Vdcanalogout-

puts(optional)

• Customerrecalibration

• 4½digitdisplaywith0.5inheight

• Zerooffset/tare

• On/offswitchdisablefeature

• Twoprogrammablelimitsandrelays(optional)

• NEMA4rating(optional)

• NISTstatement(standard)

• NISTtraceable(optional)

• Optionalcarryingcaseandpanelmountingring

High Pressure Digital Pressure Gauge

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 18: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JH

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.2%fullscale(betterthantestgaugeaccuracy)

Pressure range 0 psi to 15000, 20000, 25000, 30000, 35000, 40000,50000,60000psi

High and low capture Standard

Update speed 3 times per sec

Zero and span signal adjustment

Standard: Models JHB, JHE, JHX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1°Cto71°C[30°Fto160°F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA2(optionalNEMA4)

Power, Model JHW Oneortwo9Valkalinebatteries(in-cluded)

Power, Model JHT 110Vacadapter@60Hz(included)

Power, Model JHV 11Vdcto32Vdc@100mA(3ftcableincluded)

Power, Model JHB 11Vdcto32Vdc(dependingonloopresistance)@20mA

Power, Model JHE 11 Vdc to 32 Vdc @ 100 mA

Power, Model JHR 11 Vdc to 32 Vdc @ 100 mA

Power, Model JHX 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 3 ft cable standard: Models JHB, JHE, JHR, JHX, JHV

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98mm[3.7in]

Display 4½LCDdigits-12,7mm[0.5in]high

Pressure port 9/16-18UNF-2Bfemale(perAutoclaveAEF250-C)

Wetted parts Stainless steel

Case material Stainless steel

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

Low battery indication Standard: Model JHW

Limits and relays Models JHR and JHX1

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Maximum safe over pressure (psi)2

Incremental display steps

0 psi to 15000 psi = 15K 20000 1 x 10

0 psi to 20000 psi = 20K 25000 1 x 10

0 psi to 25000 psi = 25K 31000 1 x 10

0 psi to 30000 psi = 30K 37000 2 x 10

0 psi to 35000 psi = 35K 43000 5 x 10

0 psi to 40000 psi = 40K 50000 5 x 10

0 psi to 50000 psi = 50K 62000 5 x 10

0psito60000psi=60K 75000 5 x 10

2 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 19: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008691-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JH High Pressure Digital Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

For more information about Sensing and Control products, visit www.honeywell.com/sensingorcall+1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. The limit/relay models include two LED status indicators on front

faceandtwoformCrelays(normallyopen,common,normallyclosed)thatareratedatamaximum24Vdc/Vacat1Aor48Vdc/Vac at 1/2 A.

ORDER gUIDE

ORDER CODE* = Example

ZG50KWJH ( )

ELECTRICAL CONNECTION

4 = NEMA 4 (not available on Model JHT)NEMA 4 CONSTRUCTION

REFERENCEG = Gage - Unit reads zero at atmosphere*

PRESSURE RANGE (PSI)15K = 0-1500020K = 0-2000025K = 0-2500030K = 0-30000

35K = 0-3500040K = 0-4000050K = 0-50000*60K = 0-60000

Models with Output SignalB = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = 2 programmable limits & relaysX = 0 Vdc to 5 Vdc with two limits & relays

Models with No Output SignalW = Battery powered (9 V alkaline) *T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft cable)

MODEL DESIGNATION

SERIESJH*

Z = No electrical connection (Model JHW - 9 V battery)*C = 3 foot cableT = ac adapter (Model JHT)B = Bendix (opt) PTIH-10-6P or equal (connector sold separately)Call factory for availability of other electrical connections and cablelengths.

This ordercode is

used ONLYif you want

NEMA 4rating

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercury

Consult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 20: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Digital Pressure GaugeModel JK

DESCRIPTIONThe Series JK digital pressure test gauge with 0.2 % full scale

accuracy uses transducer technology and a stainless steel

diaphragm for high overpressure protection. The transducer

technology provides enhanced accuracy over the entire pres-

sure range. The JK has no moving parts and thus, may provide

a long life with fewer re-calibrations. The stainless steel NEMA

construction also provides EMI and RFI protection. The JK

provides high resolution with an easy to read digital display.

There are no operator errors due to interpolation of hash marks

or parallax errors.

The Model JKW is powered by one or two common 9 V alkaline

batteries. The Model JKT is powered by a 110 Vac adapter. The

Model JKV is powered by an 11 Vdc to 32 Vdc power supply.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• Rangesto10000psi

• 0.2%accuracy

• Fieldselectableengineeringunits

• Available-18°Cto82°C[0°Fto180°F]operatingtem-

perature4

• Stainlesssteelconstruction

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• 9Vbatterypower

• High/lowcaptureandstore

• On/offswitch

• CEcompliant(certainmodels,consultfactory)

• Optionalcarryingcaseandpanelmountingring

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 21: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JK

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Range See additional specifications table

Accuracy 0.2 %1

Update speed 3 per second

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1°Cto71°C[30°Fto160°F](-18°Cto82°C[0°Fto180°F]withoptionalNEMA4rating)

Rating NEMA2,std.(NEMA4optionalexceptmodelJKT)

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Display LCD,41/2-digit,1,3cm[0.5in]high

Power supply, Model JKW 1or29-voltbatteries(included)

Power supply, Model JKT 110voltacadaptor@60Hz(included)

Power supply, Model JKV 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 0,9m[3ft]cable(std.)formodelJKV

Low battery indication Standard on model JKW

Calibration data Stored on memory chip

Highandlowcapture Standard

MECHANICAL SPECIFICATIONS

Characteristic Measure

Pressure port 1-4/18NPTmale(300psiandbelow)9/16-18UNF-2BfemalestraightthreadwithSAEspecJ514o-ringboss(500psiandabove)

Face membrane Tactile feedback raised buttons

Case material Stainless steel

Wetted parts Stainless steel

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

Pressure range (psi)

1 5 15 30 50 100 200 300 500

Max. safe over pres-sure (psi)

10 25 75 150 250 500 1000 1200 1500

Incremen-tal display steps

0.001 0.005 0.01 0.02 0.05 0.1 0.1 0.2 0.5

Pressure range (psi)

750 1000 1500 2000 3000 5000 7500 10000

Max. safe over pres-sure (psi)

1500 2000 3000 4000 6000 7500 12000 15000

Incremen-tal display steps

0.5 1 1 1 2 5 5 5

ADDITIONAL SPECIFICATIONS

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 22: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985DouglasDriveNorth

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008692-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JK Digital Pressure Gauge

Warranty. Honeywellwarrantsgoodsofitsmanufactureasbeing free of defective materials and faulty workmanship. Honeywell’sstandardproductwarrantyappliesunlessagreedtootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywellduringtheperiodofcoverage,Honeywellwillrepairor replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literatureandtheHoneywellwebsite,itisuptothecustomertodetermine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However,weassumenoresponsibilityforitsuse.

For more information about Sensing and Control products, visit www.honeywell.com/sensingorcall+1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. Accuracies stated are for best fit straight line for all errors including

linearity, hysteresis and repeatability through zero.2. Consult factory for optional units of measure.3. Consult factory for optional electrical connectors and cable lengths.4. -18°Cto82°C[0°Fto180°F]operatingtemperatureonlywith

optional NEMA 4 rating.

ORDER gUIDE

W = Battery powered (9 V alkaline)*T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)

* = Example

JK*

JK

POWER

PRESSURE RANGE (PSI)1 = 0-1(opt)5 = 0-5(opt)15 = 0-1530 = 0-3050 = 0-50100 = 0-100

200 = 0-200*300 = 0-300500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20003K = 0-30005K = 0-50007.5K = 0-750010K = 0-10000

REFERENCEG = Gage (std) reads zero at atmosphere *A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt)C = Compound (opt) reads both positive & negative

(vacuum) pressure

ELECTRICAL CONNECTIONZ = No electrical connection (Model JKW - 9 volt battery)*C = 3 foot cable (Model JKV)T = ac adapter (Model JKT)

Call factory for availability of other electrical connections and cable lengths.

Z

OPTIONAL4 = NEMA 4 (Not available on JKT)

NEMA 4 option also provides temperature range of 0-180°F

4

Each gauge has the following scaling built in for field selection:FIELD SELECTABLE STANDARD UNITS OF MEASURE

Bar = BarmBar = MillibarkPa = KilopascalsMPa = Megapascals

Ft H20 = Feet of waterIn H20 = Inches of waterIn Hg = Inches of mercurymmHg = mm of mercury

Consult factory for other units of measure not listed

W 200 G

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 23: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Food/Pharmaceutical Digital Pressure GaugeModel JP

DESCRIPTIONUsing Tri-Clover sanitary fittings, the Series JP digital pres-

sure test gauge with 0.2% full scale accuracy uses transducer

technology and a stainless steel diaphragm for high over pres-

sure protection. The transducer technology provides enhanced

accuracy over the entire pressure range. The JP has no moving

parts and thus, may provide a long life with fewer re-calibrations.

The stainless steel NEMA construction also provides EMI and

RFI protection.

The JP provides high resolution with an easy to read 4½-digit

display. There are no operator errors due to interpolation of hash

marks or parallax errors. The units can be scaled to read in vari-

ous engineering units such as InHg, Ft H2O, etc. The Model JPB

provides a 4 mA to 20 mA, two-wire output. The Model JPE pro-

vides a 0 Vdc to 5 Vdc output. The Model JPR has two program-

mable limits and relays with no analog output. The Model JPX

provides a 0 Vdc to 5 Vdc output with two programmable limits

and relays for process control or alarm indication.

The Model JPW is powered by one or two common 9 V alkaline

batteries. The Model JPT is powered by a 110 Vac adapter. The

Model JPV is powered by an 11 Vdc to 32 Vdc power supply.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• Tri-Cloversanitaryfittings

• Twoprogrammablelimitsandrelays(optional)

• Testgaugeaccuracy-0.2%fullscale

• Highandlowdetection-Microprocessorbased

• NEMA4

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• Zerooffset/tare

• On/offswitchdisablefeature

• Optionalcarryingcaseandpanelmountingring

• Twoprogrammablelimitsandrelays(optional)

• NISTtraceable(optional)

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 24: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JP

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.2%fullscale(betterthantestgaugeaccuracy)

Pressure range 0 psi to 1, 2, 5, 10, 20, 50, 100, 200, 500 psi

High and low capture Standard

Update speed 3 times per sec

Zero and span signal adjustment

Standard: Models JPB, JPE, JPX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA 4

Power, Model JPW One or two 9 V alkaline batteries (included)

Power, Model JPT 110Vacadapter@60Hz(included)

Power, Model JPV 11 Vdc to 32 Vdc @ 100 mA (3ftcableincluded)

Power, Model JPB 11Vdcto32Vdc(dependingonloopresistance)@20mA

Power, Model JPE 11 Vdc to 32 Vdc @ 100 mA

Power, Model JPR 11 Vdc to 32 Vdc @ 100 mA

Power, Model JPX 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 3 ft cable standard: Models JPB, JPE, JPR, JPX, JPV

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4½ LCD digits - 12,7 mm [0.5 in] high

Pressure port Sanitary flanges: 1.5 in, 2.0 in, 2.5 in, 3.0 in

Wetted parts Stainless steel

Case material Stainless steel

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

Low battery indication Standard: Model JPW

Limits and relays Models JPR and JPX*

MOUNTINg DIMENSIONS AND CHARACTERISTICS

Request certified drawing before designing mountings or fixtures.Specifications subject to change without notice.

3 FEET CABLE EXIT *

1.5" TRI-CLOVERSANITARY FLANGE←

← →

2.2

3.7

5.2

1.2← →

→←

2.0" TRI-CLOVERSANITARY FLANGE

2.5" TRI-CLOVERSANITARY FLANGE

3.0" TRI-CLOVERSANITARY FLANGE

*Models JPB, JPE, JPR, JPX, JPVNo cable for battery powered Model JPW

ADDITIONAL SPECIFICATIONS

Maximum safe over pressure (psi)1, 2

Pressure range 0 to: (psi)

1.5 in flange dia.

2.0 in flange dia.

2.5 in flange dia.

3.0 in flange dia.

Incre-mental dis-play steps

1(opt) 10 10 10 10 0.001

2(opt) 15 15 15 15 0.002

5(opt) 25 25 25 25 0.005

10 50 50 50 50 0.01

15 75 75 75 75 0.01

25 125 125 125 125 0.02

50 250 250 250 250 0.05

100 500 500 500 350 0.1

250 600 600 600 350 0.2

400 600 600 600 NA 0.5

500 600 600 NA NA 0.5

Max. flange pressure at 100 °F

600 600 600 350

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage.

2 Maximum pressure: 1 1/2, 2, 2 1/2 inch dia. flange - 600 psi at 100 °F; 3 inch dia. flange - 350 psi at 100 °F temperatures above; 100 °F reduce maximum flange pressure. Please call factory.

For reference only

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 25: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008693-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JP Food/Pharmaceutical Digital Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. The limit/relay models include two LED status indicators on front

faceandtwoformCrelays(normallyopen,common,normallyclosed)thatareratedatamaximum24Vdc/Vacat1Aor48Vdc/Vac at 1/2 A.

ORDER gUIDE

Z* = Example

G250WJP

ELECTRICAL CONNECTION

G = Gage (std) reads zero at atmosphere*A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt) scaled in In Hg (15 psi or lower)C = Compound (opt) reads both positive and negative

(vacuum) pressure

REFERENCE

SERIESJP*MODEL DESIGNATIONModels with output signalB = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = Two programmable limits & relaysX = 0 Vdc 5 Vdc with two limits & relaysModels with no output signalW= Battery powered (9 V alkaline) *T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)

PRESSURE RANGE (PSI)

Z = No electrical connection (Model JPW - 9 V battery)*C = 3 foot cableT = ac adapter (Model JPT)B = Bendix (opt) PTIH-10-6P or equal (connector sold separately)

SANITARY FLANGE SIZE :P = 1.5 inches (up to 500 psi) (does not comply with 3A standards)Q = 2.0 inches (up to 500 psi)R = 2.5 inches (up to 400 psi)S = 3.0 inches (up to 250 psi)

Call factory for availability of other electrical connections and cable lengths.

1 = 0-1 (opt)2 = 0-2 (opt)5 = 0-5 (opt)10 = 0-10

15 = 0-1525 = 0-2550 = 0-50100 = 0-100

250= 0-250*400= 0-400500= 0-500

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercury

Consult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 26: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Remote Pressure Digital Pressure Gauge

Model JR

DESCRIPTION

With a 6-ft microchange cable connection, the Model JR digital

pressure test gauge with 0.2 % full scale accuracy uses trans-

ducer technology and a stainless steel diaphragm for high

over pressure protection. The transducer technology provides

enhanced accuracy over the entire pressure range. The JR has

no moving parts and may provide a long life with fewer re-cal-

ibrations. The JR provides high resolution with an easy to read

4½-digit display. There are no operator errors due to interpola-

tion of hash marks or parallax errors. The units can be scaled to

read in various engineering units such as InHg, Ft H2O, etc.

The Model JRB provides a 4 mA to 20 mA, two-wire output. The

Model JRE provides a 0 Vdc to 5 Vdc output. The Model JRR

has two programmable limits and relays with no analog output.

The Model JRX provides a 0 Vdc to 5 Vdc output with two pro-

grammable limits and relays for process control or alarm indica-

tion. The Model JRW is powered by one or two common

9 V alkaline batteries. The Model JRT is powered by a 110 Vac

adapter. The Model JRV is powered by an 11 Vdc to 32 Vdc

power supply.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• Portablebatterypoweredavailable

• Twoprogrammablelimitsandrelays(optional)

• Testgaugeaccuracy-0.2%fullscale

• Highandlowdetection-Microprocessorbased

• Rangesto60000psi-gage,absolute,vacuum,orcom-

pound

• Optionalpressureportadapters

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• Zerooffset/tare

• On/offswitchdisablefeature

• Optionalcarryingcaseandpanelmountingring

• NEMA4rating(optional)

• NISTtraceable(optional)

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 27: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JR

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.2%fullscale(betterthantestgaugeaccuracy)

Pressurerange0psito 1, 5, 15, 30, 50, 100, 200, 300, 500, 750, 1000, 1500, 2000, 3000, 5000, 7500, 10000psi(seeSeriesJHforrangesabove10000psi)

High and low capture Standard

Update speed 3 times per sec

Zero and span signal adjustment

Standard:ModelsJRB,JRE,JRX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA2(optionalNEMA4)

Power,ModelJRW Oneortwo9Valkalinebatteries(in-cluded)

Power,ModelJRT 110Vacadapter@60Hz(included)

Power,ModelJRV 11Vdcto32Vdc@100mA(3ftcableincluded)

Power,ModelJRB 11Vdcto32Vdc(dependingonloopresistance)@20mA

Power,ModelJRE 11 Vdc to 32 Vdc @ 100 mA

Power,ModelJRR 11 Vdc to 32 Vdc @ 100 mA

Power,ModelJRX 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 3ftfrombackofunittopowersupply:Models JRB, JRE, JRR, JRX, JRV

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4½ LCD digits - 12,7 mm [0.5 in] high

Pressureport 1/4-18NPTmale(1psito300psi);9/16-18UNF-2BfemalestraightthreadwithSAEspecJ514o-ringboss(500psiandabove)

Wetted parts Stainlesssteel

Case material Stainlesssteel

Face membrane Tactile feedback raised buttons

Calibration data Storedonmemorychip

Low battery indication Standard:ModelJRW

Limits and relays Models JRR and JRX

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Maximum safe over pressure (psi)1

Model JR

0psito1psi=1(opt) 10 0.001

0psito5psi=5(opt) 25 0.005

0 psi to 15 psi = 15 75 0.01

0 psi to 30 psi = 30 150 0.02

0 psi to 50 psi = 50 250 0.05

0 psi to 100 psi = 100 500 0.1

0 psi to 200 psi = 200 1000 0.1

0 psi to 300 psi = 300 1200 0.2

0 psi to 500 psi = 500 1500 0.5

0 psi to 750 psi = 750 1500 0.5

0 psi to 1000 psi = 1K 2000 1

0 psi to 1500 psi = 1.5K 3000 1

0 psi to 2000 psi = 2K 4000 1

0 psi to 3000 psi = 3K 6000 2

0 psi to 5000 psi = 5K 7500 5

0 psi to 7500 psi = 7.5K 12000 5

0 psi to 10000 psi = 10K 15000 5

(SeeSeriesJHforrangesabove10000psi)

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 28: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

SensingandControl

AutomationandControlSolutions

Honeywell

1985DouglasDriveNorth

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008694-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JR Remote Pressure Digital Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed tootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specificationsmaychangewithoutnotice.Theinformationwesupply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

FormoreinformationaboutSensingandControlproducts,visitwww.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. The limit/relay models include two LED status indicators on front

faceandtwoformCrelays(normallyopen,common,normallyclosed)thatareratedatamaximum24Vdc/Vacat1Aor48Vdc/Vac at 1/2 A.

ORDER gUIDE

* = ExampleZG200WJR ( )

4 = NEMA 4 (Not available on Model JRT)NEMA 4 CONSTRUCTION

SERIES

MODEL DESIGNATIONJR*

Models with output signalB = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = Two programmable limits & relaysX = 0 Vdc to 5 Vdc with two limits & relaysModels with No Output SignalW= Battery powered (9 volt alkaline) *T = 110 Volt AC adapter (included)V = 11 to 32 VDC (3 ft. cable)

PRESSURE RANGE (PSI)1 = 0-1(opt)5 = 0-5(opt)15 = 0-1530 = 0-3050 = 0-50100= 0-100

200 = 0-200*300 = 0-300500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20003K = 0-30005K = 0-50007.5K = 0-750010K = 0-10000

See Series JH for ranges above 10,000 psi.

G = Gage (std) reads zero at atmosphere*A = Absolute (opt) reads zero at vacuum(15 psi or higher)V = Vacuum (opt) scaled in inches of mercury (15 psi or lower)C = Compound (opt) reads both positive and negative

(vacuum) pressure

REFERENCE

ELECTRICAL CONNECTIONZ = No electrical connection (Model JRW - 9 V battery)*C = 3 foot cableT = ac adapter (Model JRT)B = Bendix (opt) PTIH- 10-6P or equal (connector sold separately

- back connection only)Call Factory for availability of other electrical connections and cable lengths.

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercury

Consult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

This ordercode is

used ONLYif you want

NEMA 4rating

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 29: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Model JS

DESCRIPTIONThe Model JS digital pressure test gauge with 0.2 % full scale

accuracy uses transducer technology and a stainless steel dia-

phragm for high over pressure protection. The transducer tech-

nology provides enhanced accuracy over the entire pressure

range. The JS has no moving parts and thus, may provide a long

life with fewer re-calibrations. The stainless steel NEMA con-

struction also provides EMI and RFI protection. The JS provides

high resolution with an easy to read digital display. There are no

operator errors due to interpolation of hash marks or parallax er-

rors. The units can be scaled to read in various engineering units

such as InHg, Ft H2O, etc.

The Model JSB provides a 4 mA to 20 mA, two-wire output.

The Model JSE provides a 0 Vdc to 5 Vdc output. The Model

JSR has two programmable limits and relays with no analog

output. The Model JSX provides a 0 Vdc to 5 Vdc output with

two programmable limits and relays for process control or alarm

indication.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• 4mAto20mA(two-wire)and0Vdcto5Vdcanalogoutputs

• Twoprogrammablelimitsandrelays(optional)

• Testgaugeaccuracy-0.2%fullscale

• Highandlowdetection-Microprocessorbased

• Rangesto10000psi-gage,absolute,vacuum,orcom-

pound

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• Zerooffset/tare

• On/offswitchdisablefeature

• Optionalcarryingcaseandpanelmountingring

• Optionalpressureportadapters

• NEMA4rating(optional)

• NISTtraceable(optional)

Digital Pressure Gauge With Output

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 30: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JS

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.2%fullscale(betterthantestgaugeaccuracy)

Pressure range 0 psi to 1,5,15,30,50,100,200,300,500,750,1000,1500,2000,3000,5000,7500,10000psi

High and low capture Standard

Update speed 3timespersec

Zero and span signal adjustment

Standard: Models JSB, JSE, JSX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1°Cto71°C[30°Fto160°F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA2(optionalNEMA4)

Power, Model JSB 11Vdcto32Vdc(dependingonloopresistance)@20mA

Power, Model JSE 11Vdcto32Vdc@100mA

Power, Model JSR 11Vdcto32Vdc@100mA

Power, Model JSX 11Vdcto32Vdc@100mA

Electrical connection 3ftcable(standard)

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98mm[3.7in]

Display 4½LCDdigits-12,7mm[0.5in]high

Pressure port 1/4-18NPTmale(1psito300psi);9/16-18UNF-2BfemalestraightthreadwithSAEspecJ514o-ringboss(500psiandabove)

Wetted parts Stainless steel

Case material Stainless steel

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

Limits and relays Models JSR and JSX

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Maximum safe over pressure (psi)1

Model JS

0psito1psi=1(opt) 10 0.001

0psito5psi=5(opt) 25 0.005

0psito15psi=15 75 0.01

0psito30psi=30 150 0.02

0psito50psi=50 250 0.05

0psito100psi=100 500 0.1

0psito200psi=200 1000 0.1

0psito300psi=300 1200 0.2

0psito500psi=500 1500 0.5

0psito750psi=750 1500 0.5

0psito1000psi=1K 2000 1

0psito1500psi=1.5K 3000 1

0psito2000psi=2K 4000 1

0psito3000psi=3K 6000 2

0psito5000psi=5K 7500 5

0psito7500psi=7.5K 12000 5

0psito10000psi=10K 15000 5

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 31: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985DouglasDriveNorth

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008695-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JS Digital Pressure Gauge With Output

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed tootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

For more information about Sensing and Control products, visit www.honeywell.com/sensingorcall+1-815-235-6847Emailinquiriestoinfo.sc@honeywell.com

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. Thelimit/relaymodelsincludetwoLEDstatusindicatorsonfront

faceandtwoformCrelays(normallyopen,common,normallyclosed)thatareratedatamaximum24Vdc/Vacat1Aor48Vdc/Vacat1/2A.

G = Gage (std) reads zero at atmosphere*A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt) scaled in In Hg (15 psi or lower)C = Compound (opt) reads both positive and negative

(vacuum) pressure

OUTPUT SIGNAL

* = Example

SERIESJS*

B = 4 mA to 20 mA (two-wire)*E = 0 Vdc to 4 VdcR = 2 programmable limits & relaysX = 0 Vdc to 5 Vdc with two limits & relays

CG200BJS

REFERENCE

PRESSURE RANGE (PSI)1K = 0-10001.5K = 0-15002K = 0-20003K = 0-30005K = 0-50007.5K = 0-750010K = 0-10000

1 = 0-1 (opt)5 = 0-5 (opt)15 = 0-1530 = 0-3050 = 0-50

100 = 0-100200 = 0-200*300 = 0-300500 = 0-500750 = 0-750

( )

NEMA 4 CONSTRUCTION4 = NEMA 4

ELECTRICAL CONNECTIONC = 3 foot cable*B = Bendix (opt) - PTIH-10-6P or equal (connector sold seperately)

Call factory for availability of other electrical connections and cable lengths

This ordercode is

used ONLYif you want

NEMA 4rating

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercury

Consult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

ORDER gUIDE

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 32: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Digital Pressure Gauge With VCR Fittings

Model JV

DESCRIPTION

Using VCR, the Series JV digital pressure test gauge with 0. 2%

full scale accuracy uses transducer technology and a stainless

steel diaphragm for high over pressure protection. The transduc-

er technology provides enhanced accuracy over the entire pres-

sure range. The JV has no moving parts and thus, may provide a

long life with fewer re-calibrations.

The stainless steel NEMA construction also provides EMI and

RFI protection. The JV provides high resolution with an easy to

read digital display. There are no operator errors due to interpo-

lation of hash marks or parallax errors. The units can be scaled

to read in various engineering units, such as in-Hg, ft-H20, etc.

The Model JVB provides a 4 mA to 20 mA, two-wire output. The

Model JVE provides a 0 Vdc to 5 Vdc output. The Model JVR has

two programmable limits and relays with no analog output. The

Model JVX provides a 0 Vdc to 5 Vdc output with two program-

mable limits and relays for process control or alarm indication.

The Model JVW is powered by one or two common 9 V alkaline

batteries. The Model JVT is powered by a 110 Vac adapter. The

JVV, JVB, JVE, JVR, and JVX models are powered by an 11 Vdc

to 32 Vdc power supply.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

deactivated.

FEATURES• Passivationoroxygen-cleanedoptions

• Twoprogrammablelimitsandrelays(optional)

• Testgaugeaccuracy-0.2%fullscale

• Highandlowdetection-Microprocessorbased

• Rangesto5000psi-gage,absolute,vacuum,orcompound

• 4½digitdisplaywith0.5inheight

• Customerrecalibration

• Zerooffset/tare

• On/offswitchdisablefeature

• Optionalcarryingcaseandpanelmountingrings

• NEMA4rating(optional)

• NISTtraceable(optional)

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 33: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model JV

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Linearity and hysteresis 0.2%fullscale(betterthantestgaugeaccuracy)

Pressurerange0psito 1, 5, 15, 30, 50, 100, 200, 300, 500, 750, 1000, 1500, 2000, 3000, 5000, 7500, 10000 psi

High and low capture Standard

Update speed 3 times per sec

Zero and span signal adjustment

Standard: Models JVB, JVE, JVX

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA2(optionalNEMA4)

Power,ModelJVW Oneortwo9Valkalinebatteries(in-cluded)

Power,ModelJVT 110Vacadapter@60Hz(included)

Power,ModelJVV 11Vdcto32Vdc@100mA(3ftcableincluded)

Power,ModelJVB 11Vdcto32Vdc(dependingonloopresistance)@20mA

Power,ModelJVE 11 Vdc to 32 Vdc @ 100 mA

Power,ModelJVR 11 Vdc to 32 Vdc @ 100 mA

Power,ModelJVX 11 Vdc to 32 Vdc @ 100 mA

Electrical connection 3ftcable(standard)onModels:JVB,JVE, JVR, JVX, JVV

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4½LCDdigits-12,7mm[0.5in]high

Pressureport Size4stainlesssteelVCR(male)(OtherVCRpartsavailable.Consultfactory)

Wetted parts Stainless steel

Case material Stainless steel

Face membrane Tactile feedback raised buttons

Calibration data Stored on memory chip

Low battery indication Standard on Model JVW

Limits and relays Models JVR and JVX

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Maximum safe over pressure (psi)1

Model JV

0psito1psi=1(opt) 10 0.001

0psito5psi=5(opt) 25 0.005

0 psi to 15 psi = 15 75 0.01

0 psi to 30 psi = 30 150 0.02

0 psi to 50 psi = 50 250 0.05

0 psi to 100 psi = 100 500 0.1

0 psi to 200 psi = 200 1000 0.1

0 psi to 300 psi = 300 1200 0.2

0 psi to 500 psi = 500 1500 0.5

0 psi to 750 psi = 750 1500 0.5

0 psi to 1000 psi = 1K 2000 1

0 psi to 1500 psi = 1.5K 3000 1

0 psi to 2000 psi = 2K 4000 1

0 psi to 3000 psi = 3K 6000 2

0 psi to 5000 psi = 5K 7500 5

0 psi to 7500 psi = 7.5K 12000 5

0 psi to 10000 psi = 10K 15000 5

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 34: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

Golden Valley, MN 55422 USA

+1-815-235-6847

www.honeywell.com/sensing

008696-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model JV Digital Pressure Gauge With VCR Fittings

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

For more information about Sensing and Control products, visit www.honeywell.com/sensing or call +1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. The limit/relay models include two LED status indicators on front

faceandtwoformCrelays(normallyopen,common,normallyclosed)thatareratedatamaximum24Vdc/Vacat1Aor48Vdc/Vac at 1/2 A.

ORDER gUIDE

Z* = Example

G100WJV

1 = 0-1 (opt)5 = 0-5 (opt)15 = 0-1530 = 0-3050 = 0-50

100= 0-100*200= 0-200300= 0-300500= 0-500750= 0-7501K = 0-1000

1.5K = 0-15002K = 0-20003K = 0-30005K = 0-50007.5K = 0-750010K = 0-10000

SERIESJV*MODEL DESIGNATION

4 = NEMA 4 (not available on Model JVT)NEMA 4 CONSTRUCTION

( )

Models with output signalB = 4 mA to 20 mA (two-wire)E = 0 Vdc to 5 VdcR = 2 programmable limits & relaysX = 0 Vdc to 5 Vdc with two limits & relays

Models with no output signalW= Battery powered (9 V alkaline)*T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)

PRESSURE RANGE (PSI)

G = Gage (std) reads zero at atmosphere*A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt) scaled in In Hg (15 psi or lower)C = Compound (opt) reads both positive and negative

(vacuum) pressure

REFERENCE

V

ELECTRICAL CONNECTIONZ = No electrical connection (Model JVW - 9 V battery)*C = 3 foot cableT = ac adapter (Model JVT)B = Bendix (opt) (PTIH-10-6P or equal) connector sold

separatelyCall factory for availability of other electrical connections and cable lengths.

V = Size 4 stainless steel VCR (male)*Other VCR ports available. Please call factory.

VCR PRESSURE FITTING

This ordercode is

used ONLYif you want

NEMA 4rating

Bar = Bar Ft H20 = Feet of waterMBar = Millibar In H20 = Inches of waterkPa = Kilopascals In Hg = Inches of mercuryMPa = Megapascals mmHg = mm of mercuryConsult factory for other units of measure not listed

FIELD SELECTABLE STANDARD UNITS OF MEASURE

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 35: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Intrinsically Safe Pressure GaugeModel PI

DESCRIPTIONAs an in-process gauge or portable calibrator, the hand-held

Model PI delivers long-lasting, enhanced accuracy perfor-

mance. Accuracy of 0.2 % coupled with an enhanced resolution,

4½-digit display makes this unit an excellent transfer standard

for calibrating pressure measuring equipment.

The PI is rugged and features high/low pressure capture with a

field-adjustable update rate from three-to-ten times per second.

These units are manufactured from stainless steel and use sen-

sor technology which has no moving parts, requires little recali-

bration and has excellent overpressure tolerance. NEMA 4 rating

provides EMI and RFI protection. Units of measure are field-

selectable and the unit is offered with gage, absolute, vacuum,

or compound reference.

The Model PIW is powered by one or two common 9 V alkaline

batteries.

The PI Digital Gauges have the following complete Factory Mu-

tual Approved Classifications:

Class I: Gases and vapor

Class II: Combustible dust

Class III: Fibers

Intrinsically safe systems limit the thermal and electrical energy

to the hazardous area to help prevent ignition.

Division 1: present or likely to be present in normal operation

Division 2: not present in normal operation

Group A: Acetylene

Group B: Hydrogen

Group C: Acetaldehyde, Ethylene, Methyl Ether

Group D: Acetone, Gasoline, Methanol, Propane

Group E: Metal dust

Group F: Carbon dust

Group G: Grain dust

FEATURES• Highaccuracy:0.2%fullscale

• FactoryMutualapproved

• Rangesto10000psi-gage,absolute,vacuum,orcom-

pound

• Stainlesssteelconstruction

• On/offswitch

• CEcompliant

• 9Vbatterypower

• Optionalpressureports

• Optionalcarryingcaseandpanelmountingring

• High/lowcaptureandstore

• 4½-digitdisplaywith0.5inheight

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 36: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model PI

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Accuracy 0.2%offullscaleBSFL

Pressure range, 0 psi to: 15,50,100,200,500,750,1000,1500,2000,5000,7500, 10000 psi

High and low capture Standard

Update speed 3 per sec

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating (standard)

-1°Cto71°C[30°Fto160°F]

Temperature, operating (optional)

-18°Cto82°C[0°Fto180°F]

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Rating NEMA 4 (also provides temperature range of -18 °C to 82 °C [0°Fto180°F])

Power, Model PIW Oneortwo9Valkalinebatteries(in-cluded)

Power, Model PIV 11 Vdc to 32 Vdc @ 100 mA (3 ft cable included)

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98mm[3.7in]

Display 4LCDdigits-12,7mm[0.5in]high

Pressure port 9/16-18UNF-2B-femalestraightthreadwithSAEspecJ514o-ringboss

Wetted parts Stainlesssteel

Case material Stainlesssteel

Lowbatteryindication Standard

Face membrane Tactile feedback raised buttons

Calibration data Storedonmemorychip

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Maximum safe over pressure (psi)1

Incremental display steps

0psito15psi=15 75 0.01

0psito50psi=50 250 0.05

0psito100psi=100 500 0.1

0psito200psi=200 1000 0.1

0psito500psi=500 1500 0.5

0psito750psi=750 1500 0.5

0psito1000psi=1K 2000 1

0psito1500psi=1.5K 3000 1

0psito2000psi=2K 4000 1

0psito5000psi=5K 7500 5

0psito7500psi=7.5K 12000 5

0psito10000psi=10K 15000 5

1 Maximum safe overpressure is the pressure which the unit can experience oc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 37: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

SensingandControl

AutomationandControlSolutions

Honeywell

1985DouglasDriveNorth

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008685-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model PI Intrinsically Safe Pressure Gauge

Warranty. Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Honeywell’s standard product warranty applies unless agreed to otherwise by Honeywell in writing; please refer to your order acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the customer to determine the suitability of the product in the application.

Specificationsmaychangewithoutnotice.Theinformationwesupply is believed to be accurate and reliable as of this printing. However, we assume no responsibility for its use.

FormoreinformationaboutSensingandControlproducts,visitwww.honeywell.com/sensingorcall+1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

Notes

1.Thelimit/relaymodelsincludetwoLEDstatusindicatorsonfrontfaceand two form C relays (normally open, common, normally closed) that are rated at a maximum 24 Vdc/Vac at 1 A or 48 Vdc/Vac at 1/2 A.

W = Battery powered (9 v alkaline)*

* = Example

SERIESPI

PI W 200 G

POWER

PRESSURE RANGE (psi)

REFERENCEG = Gage (std) reads zero at atmosphere *A = Absolute (opt) reads zero at vacuum (15 psi or higher)V = Vacuum (opt) scaled in In Hg (15 psi or lower)C = Compound (opt) reads both positive & negative

(vacuum) pressure

ELECTRICAL CONNECTION

Z

15 = 0-1550 = 0-50100 = 0-100200 = 0-200*

500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20005K = 0-50007.5K = 0-750010K = 0-10000

Z = No electrical connection (Model PIW - 9 V battery)*Call factory for availability of other electrical connections and cable lengths.

Each gage has the following scaling built in for field selection:FIELD SELECTABLE STANDARD UNITS OF MEASURE

Bar = BarmBar = MillibarkPa = KilopascalsMPa = Megapascals

Ft H20 = Feet of waterIn H20 = Inches of waterIn Hg = Inches of mercurymmHg = mmof mercury

Consult factory for other units of measure not listed

NEMA (STANDARD)4 = NEMA 4

4

ORDER CODE

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 38: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Portable Pressure CalibratorsModels AC and JC

DESCRIPTIONConstructed with stainless steel wetted parts and totally por-

table, the JC and AC Series pressure calibrators can handle

many rugged applications and harsh environments. The circuitry

is designed to withstand dropping and is immune to electromag-

netic interference. The calibrators are fully battery powered (9 V)

so they are easily transportable from site to site.

This versatile pressure calibrator is available in two accuracies:

JC Series with 0.2 % accuracy and AC Series with 0.05 % ac-

curacy. The calibrators have a 4½-digit display for accurate and

easy readings. The high/low feature captures and holds both the

highest and lowest pressure reading. (Press clear or off to reset).

Operating temperatures are from -18 °C to 82 °C [0 °F to 180 °F].

The set includes the digital gauge, carrying case, NIST certifi-

cate, and pressure adapter.

FEATURES• 0.2%and0.05%fullscaleaccuracy

• Fieldselectableengineeringunits

• Highandlowandpeakmemoryfunction

• Wettedstainlesssteelconstruction

• Highover-pressureprotection

• 4½digitdisplaywith0.5inheight

• EMIandRFIshieldingforalltypesofindustrialenvironments

• NISTtraceable(standard)

• NEMA4ratedforharshenvironments

• Optionalpressureports

• Optionalcarryingcaseandpanelmountingring

• NISTtraceable(optional)

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 39: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Models AC and JC

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Accuracy, JC Series 0.2 % of full scale

Accuracy, AC Series 0.05 % of full scale

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating (standard)

-1 °C to 71 °C [30 °F to 160 °F]

Temperature, operating (optional)

-18°Cto82°C[0°Fto180°F](NEMA4option required)AK only

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Display, AC Series 4½ LCD digits, 0.5 in high

Display, JC Series 4 LCD digits, 0.5 in high

MECHANICAL SPECIFICATIONS

Characteristic Measure

Wettedparts Stainless steel

Case material Stainless steel

Highandlowcapture Standard

Calibration data Stored on memory chip

MOUNTINg DIMENSIONS AND CHARACTERISTICS

For reference only

ADDITIONAL SPECIFICATIONS

Incremental display steps

Pressure range (psi) Maximum safe over pressure (psi)1

Model JC

Model AC

0 psi to 1 psi = 1 (opt) 10 0.001 –

0 psi to 5 psi = 5 (opt) 25 0.005 –

0 psi to 15 psi = 15 75 0.01 0.005

0 psi to 30 psi = 30 150 0.02 –

0 psi to 50 psi = 50 250 0.05 0.01

0 psi to 100 psi = 100 500 0.1 0.02

0 psi to 200 psi = 200 1000 0.1 0.05

0 psi to 300 psi = 300 1200 0.2 –

0 psi to 500 psi = 500 1500 0.5 0.1

0 psi to 750 psi = 750 1500 0.5 0.2

0 psi to 1000 psi = 1K 2000 1 0.2

0 psi to 1500 psi = 1.5K 3000 1 0.5

0 psi to 2000 psi = 2K 4000 1 0.5

0 psi to 3000 psi = 3K 6000 2 –

0 psi to 5000 psi = 5K 7500 5 1

0 psi to 7500 psi = 7.5K 12000 5 2

0 psi to 10000 psi = 10K 15000 5 2

1Maximumsafeoverpressureisthepressurewhichtheunitcanexperienceoc-casionally without the loss of accuracy or permanent damage

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 40: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Sensing and Control

Automation and Control Solutions

Honeywell

1985 Douglas Drive North

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008687-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Models AC and JC Portable Pressure Calibrators

Warranty. Honeywellwarrantsgoodsofitsmanufactureasbeingfreeofdefectivematerialsandfaultyworkmanship.Honeywell’sstandardproductwarrantyappliesunlessagreedtootherwisebyHoneywellinwriting;pleaserefertoyourorderacknowledgementorconsultyourlocalsalesofficeforspecific warranty details. If warranted goods are returned to Honeywellduringtheperiodofcoverage,Honeywellwillrepairor replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

Whileweprovideapplicationassistancepersonally,throughourliteratureandtheHoneywellwebsite,itisuptothecustomertodetermine the suitability of the product in the application.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this printing. However,weassumenoresponsibilityforitsuse.

For more information about Sensing and Control products, visit www.honeywell.com/sensing or call [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

W = Battery powered (9 V alkaline)*T = 110 Vac adapter (included)V = 11 Vdc to 32 Vdc (3 ft. cable)

* = Example• = Available on JC only

SERIES• JC* = 0.2% AC = 0.05%

JC W 200 G

POWER

4 = NEMA 4 (not available on Model JCT or ACT)

NEMA 4 CONSTRUCTION

Ft H20 = Feet of water Bar = BarIn H20 = Inches of water mmHg =mm of mercuryIn Hg = Inches of mercury kPa = Kilopascals

Consult factory for other units of measure not listedNote: PSI is standard

OPTIONAL UNITS OF MEASURE(AC is Front Panel Field Selectable, JC is Factory Set)

O P

T I

O N

A L

PRESSURE RANGE (PSI)Standard

1 = 0-1 (opt)•5 = 0-5 (opt)•

15 = 0-1530 = 0-30•50 = 0-50

100 = 0-100

200 = 0-200*300 = 0-300•500 = 0-500750 = 0-7501K = 0-10001.5K = 0-1500

2K = 0-20003K = 0-3000•5K = 0-50007.5K = 0-750010K = 0-10000

REFERENCEG= Gage (std) reads zero at atmosphere *A= Absolute (opt) reads zero at vacuum (15 psi or higher)V= Vacuum (opt) scaled in In Hg (15 psi or lower)C= Compound (opt) reads both positive & negative

(vacuum) pressure

ELECTRICAL CONNECTIONZ = No electrical connection (Model JCW - 9 volt battery)*C = 3 foot cable (Model JCV)T = ac adapter (Model JCT)

Call factory for availability of other electrical connections and cable lengths.

These ordercodes are

used ONLY ifyou want specialunits of measure

and/orNEMA 4 rating.

( )Z

The following order codes are used ONLY if you want these options.

ORDER CODES

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 41: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Digital Force IndicatorModel LC

DESCRIPTIONThe Model LC digital force indicator is a highly portable force

measurement device. With its 9 V battery supply and compact

size the indicator can easily be carried to remote locations to

perform force measurement calibrations. The unit can measure

compression forces up to 50,000 lbs.

The peak capture feature allows the highest force reading taken

to be displayed with the push of a button. The remote location of

the load sensor minimizes the chance of damage to the elec-

tronic body of the gauge. The stainless steel construction of the

sensor and the indicator makes them often resistant to the occa-

sional mishandling than can occur in the field. Optional handles

are available for the load cell and gauge.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• Rangesto50000lb

• Compressionmeasurement

• High/lowforcecapture

• Zerooffset/tare

• NISTtraceable

• Customerrecalibration

• Optionalhandlesandpanelmountingring

• Optionalcarryingcase

• OptionalNEMA4rating

• Selectableengineeringunits

Model LC shown with optional handle.

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 42: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model LC

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Repeatability 0.2 % of full scale

Range 5 lb to 50000 lb

Highandlowcapture Standard

Update speed 3 per sec (selectable to 10 per sec)

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Power 9 V battery, 11 Vdc to 32 Vdc, 100 Vac

MECHANICAL SPECIFICATIONS

Characteristic Measure

Diameter 93,98 mm [3.7 in]

Display 4½digits

Case material Stainlesssteel

Cell materials Stainlesssteel

Zero/span adjustment Standard

Low battery indication Standard

Face membrane Tactile feedback

Calibration data Storedonmemorychip

MOUNTINg DIMENSIONS

ADDITIONAL SPECIFICATIONS

Pressure range (psi) Incremental display steps

0 psi to 1 psi (opt) = 1 0.001

0 psi to 5 psi (opt) = 5 0.005

0 psi to 25 psi = 25 0.005

0 psi to 50 psi = 50 0.01

0 psi to 100 psi = 100 0.02

0 psi to 250 psi = 250 0.05

0 psi to 500 psi = 500 0.1

0 psi to 1000 psi = 1K 0.2

0 psi to 2000 psi = 2K 0.5

0 psi to 3000 psi = 3K 2

0psito4000psi=4L 2

0 psi to 5000 psi = 5K 1

0 psi to 7500 psi = 7.5K 2

0 psi to 10000 psi = 10K 2

0 psi to 15000 psi = 15K 2

0 psi to 20000 psi = 20K 2

0 psi to 30000 psi = 30K Consult factory

0 psi to 50000 psi = 50K Consult factory

ACCESSORIES ORDER CODES

Order code Accessory

CC2 Standardcarryingcase:39/16Dx713/16Wx81/2L

MR2 Panel mounting ring

NISTCERTS NISTcertification(inadditiontoNISTstate-ment)

HD1 Handle

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 43: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

SensingandControl

AutomationandControlSolutions

Honeywell

1985DouglasDriveNorth

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008671-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model LC Digital Force Indicator

Warranty. Honeywellwarrantsgoodsofitsmanufactureasbeing free of defective materials and faulty workmanship. Honeywell’sstandardproductwarrantyappliesunlessagreedtootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specificwarrantydetails.IfwarrantedgoodsarereturnedtoHoneywellduringtheperiodofcoverage,Honeywellwillrepairor replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literatureandtheHoneywellwebsite,itisuptothecustomertodetermine the suitability of the product in the application.

Specificationsmaychangewithoutnotice.Theinformationwesupply is believed to be accurate and reliable as of this printing. However,weassumenoresponsibilityforitsuse.

FormoreinformationaboutSensingandControlproducts,visitwww.honeywell.com/sensingorcall+1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. Accuraciesstatedareforbestfitstraightlineforallerrorsincluding

linearity, hysteresis and repeatability through zero.2. Consult factory for optional units of measure.3. Consult factory for optional electrical connectors and cable lengths.

ORDER gUIDE

SERIESLC=0.2%

POWERW=Battery powered (9 V alkaline)T=110 Vac adapter (included)V=11 Vdc to 32 Vdc (3 ft. cable)

PRESSURE RANGE (lb) 5=0-5 1K=0-1000 10K=0-10,000

10=0-10 2K=0-2000 15K=0-15,00025=0-25 3K=0-3000 20K=0-20,00050=0-50 4K=0-4000 30K=0-30,000

100=0-100 5K=0-5000 50K-0-50,000250=0-250 7.5K=7500500=0-500

REFERENCEK=Compression

ELECTRICAL CONNECTIONZ = No electrical connection C = 3 foot cableT = ac adapterCall factory for availability of other electrical connections and cable lengths.

UNITS OF MEASURE (Field Selectable)Note: lbs is standardn = NewtonKn = Kilonewtong = GramsKg = KilogramsConsult factory for other units of measure not listed

OPTIONAL NEMA 4 CONSTRUCTION4 = NEMA 4

standard

LC W 100 K Z ( )

This ordercode is

used ONLY ifyou want

NEMA 4 rating.

repeatability

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 44: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

Digital Force IndicatorModel LK

DESCRIPTIONThe Model LK Digital Force Indicator is a highly portable force

measurement device. With its 9 V battery supply and compact

size the indicator can easily be carried to remote locations to

perform force measurement calibrations. The unit can measure

forces in tension and compression over ranges up to 10000 lb.

The peak capture feature allows the highest force reading taken

to be displayed with the push of a button.

The remote location of the load sensor minimizes the chance of

damage to the electronic body of the gauge. The stainless steel

construction of the sensor and the indicator also make them re-

sistant to the occasional mishandling than can occur in the field.

Optional handles are available for the load cell and gauge.

Each unit has a membrane face with raised buttons and tactile

feedback for setup and operation. The high, low, and clear but-

tons are easily accessible on this front membrane. Zero adjust-

ment and zero offset/tare functions are standard on each unit.

Calibration and setup parameters are stored on a memory chip

to protect from loss even when power is interrupted. Unauthor-

ized set ups and calibrations are also blocked with internal

security. Various combinations of the front panel buttons can be

de-activated.

FEATURES• Rangesto10000lb

• Tensionandcompressionmeasurement

• Highandlowforcecapture

• Zerooffset/tare

• NISTtraceable

• Optionalcarryingcaseandpanelmountingring

• OptionalNEMA4rating

• Fieldselectableengineeringunits(lb,N,kN,g,kg)

• Customerrecalibration

Model LK shown with optional handle. Visit our Web site for more information on optional handles.

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 45: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

2Honeywell•SensingandControl

Model LK

PERFORMANCE SPECIFICATIONS

Characteristic Measure

Range Seeadditionalspecificationstable

Accuracy 0.2 %1

Update speed 3persecond(selectableto10/second)

ENVIRONMENTAL SPECIFICATIONS

Characteristic Measure

Temperature, operating -1 °C to 71 °C [30 °F to 160 °F]

Rating NEMA2,std.(NEMA4optionalexceptmodelLKT)

ELECTRICAL SPECIFICATIONS

Characteristic Measure

Display LCD,41/2-digit,1,3cm[0.5in]high

Power supply, Model LKW 1or29-voltbatteries(included)

Power supply, Model LKT 110voltacadaptor@60Hz(included)

Power supply, Model LKV 11Vdcto32Vdc@100mA

Electrical connection 0,9m[3ft]cable(std.)formodelLKV

Low battery indication StandardonmodelLKW

Calibration data Storedonmemorychip

Zero and span adjustment Standard

MECHANICAL SPECIFICATIONS

Characteristic Measure

Case material Stainlesssteel

MOUNTINg DIMENSIONS AND CHARACTERISTICS

Rang-es (lb)

T D H C F A B

1 lb #6-32 UNC

25,4mm [1.00 in]

19,5mm [0.75in]

6,4mm[0.25in]

2,8 mm [0.11 in]

12,5mm [0.50in]

9,7 mm [0.38 in]

5to10lb

#6-32 UNC

25,4mm [1.00 in]

11 mm [0.45in]

6,4mm[0.25in]

1,3 mm [0.05in]

7,9 mm [0.31 in]

4,8mm[0.19 in]

25to100 lb

#10-32 UNF

25,4mm [1.00 in]

13 mm [0.52in]

6,4mm[0.25in]

0,8 mm [0.03 in]

12,5mm [0.50in]

6,4mm[0.25in]

250to1000 lb

1/4”-28UNF

25,4mm [1.00 in]

13 mm[0.52in]

9,7 mm [0.38 in]

0,8 mm [0.03 in]

12,5mm [0.50in]

6,4mm[0.25in]

2000 to 3000 lb

3/8”-24UNF

25,4mm [1.00 in]

18 mm [0.72 in]

12,5mm[0.50in]

0,8 mm [0.03 in]

12,5mm [0.50in]

9,7 mm [0.38 in]

4000to5000lb

1/2”-20UNF

32,0 mm [1.25in]

24mm[0.94in]

16 mm[0.63 in]

0,8 mm [0.03 in]

12,5mm [0.50in]

9,7 mm [0.38 in]

7500to10000 lb

3/4”-16UNF

35mm[1.38 in]

28 mm [1.10 in]

22 mm [0.88 in]

0,8 mm [0.03 in]

12,5mm [0.50in]

9,7 mm [0.38 in]

ADDITIONAL SPECIFICATIONS

Range (lb)

0 to 1 0 to 5 0 to 25

0 to 50

0 to 100

0 to 250

0 to 500

Incre-mental display steps

0.001 0.005 0.02 0.01 0.02 0.05 0.1

Range (lb)

0 to 1000

0 to 2000

0 to 3000

0 to 4000

0 to 5000

0 to 7500

0 to 10000

Incre-mental display steps

0.2 0.5 2 2 1 2 2

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com

Page 46: Intrinsically Safe Pressure Gauge - Steven EngineeringDigital Pressure Test Gauge Model AK DESCRIPTION The hand-held Model AK is an in-progress gauge or por-table calibrator and delivers

SensingandControl

AutomationandControlSolutions

Honeywell

1985DouglasDriveNorth

GoldenValley,MN55422USA

+1-815-235-6847

www.honeywell.com/sensing

008670-1-EN IL50 GLO May 2008Copyright © 2008 Honeywell International Inc. All rights reserved.

Model LK Digital Force Indicator

Warranty. Honeywellwarrantsgoodsofitsmanufactureasbeing free of defective materials and faulty workmanship. Honeywell’sstandardproductwarrantyappliesunlessagreedtootherwisebyHoneywellinwriting;pleaserefertoyourorder acknowledgement or consult your local sales office for specific warranty details. If warranted goods are returned to Honeywellduringtheperiodofcoverage,Honeywellwillrepairor replace, at its option, without charge those items it finds defective. The foregoing is buyer’s sole remedy and is in lieu of all warranties, expressed or implied, including those of merchantability and fitness for a particular purpose. In no event shall Honeywell be liable for consequential, special, or indirect damages.

While we provide application assistance personally, through our literatureandtheHoneywellwebsite,itisuptothecustomertodetermine the suitability of the product in the application.

Specificationsmaychangewithoutnotice.Theinformationwesupply is believed to be accurate and reliable as of this printing. However,weassumenoresponsibilityforitsuse.

FormoreinformationaboutSensingandControlproducts,visitwww.honeywell.com/sensingorcall+1-815-235-6847Email inquiries to [email protected]

WARNINgPERSONAL INJURY

•DONOTUSEtheseproductsassafetyoremergencystopdevicesorinanyotherapplicationwherefailureoftheproductcouldresultinpersonalinjury.

Failure to comply with these instructions could result in death or serious injury.

WARNINgMISUSE OF DOCUMENTATION

•Theinformationpresentedinthiscatalogueisforreferenceonly.DONOTUSEthisdocumentasproductinstallationinformation.•Completeinstallation,operationandmaintenanceinformationisprovidedintheinstructionssuppliedwitheachproduct.

Failure to comply with these instructions could result in death or serious injury.

NOTES1. Accuraciesstatedareforbestfitstraightlineforallerrorsincluding

linearity, hysteresis and repeatability through zero.2. Consult factory for optional units of measure.3. Consult factory for optional electrical connectors and cable lengths.

Courtesy of Steven Engineering, Inc.-230 Ryan Way, South San Francisco, CA 94080-6370-Main Office: (650) 588-9200-Outside Local Area: (800) 258-9200-www.stevenengineering.com