IROC Houston QA Center end-to-end

QA phantom program key findings over

the past 15 years

David Followill, Ph.D.

AAPM 2017 Summer Meeting

Denver, CO

July 30, 2017

Page 2: IROC Houston QA Center end-to-end QA phantom program key

Extent of Activities

• 1,837 RT facilities

participate in NCTN and

monitored by IROC

Houston

• 329 other sites (primarily

international also audited

• 56 countries monitoring a

total of 2166 sites

• Largest QA audit program

• Shipping >700 phantoms/yr

Page 3: IROC Houston QA Center end-to-end QA phantom program key

Imaging, Planning and Delivery - QA required at each step

Positioning

and

Immobilization

Image

Acquisition

(CT, MR)

Structure

Segmentation

treatment

planning

File

transfer and

management

Plan

validation

Position

verification

treatment

delivery

Black Box

Page 4: IROC Houston QA Center end-to-end QA phantom program key

Thus the need for an end-to-end

QA audit tool to verify the

intended treatment goal.

Deliver the correct dose

to correct location

as planned

Page 5: IROC Houston QA Center end-to-end QA phantom program key

Phantom Family

2 prostate phantoms

19 SRS phantoms

33 lung phantoms

8 Spine

phantoms 10 liver inserts24 H&N

phantoms

Page 6: IROC Houston QA Center end-to-end QA phantom program key

• Anthropomorphic shape with targets and OARs

• Independent “end to end” audit

• Imaging• Planning/dose calculation• Setup• delivery

• Uniform phantoms and dosimeters

• Standardized analysis

• Uniform pass/fail criteria

• Allows inst. to inst. comparison

• Established infrastructure

Benefits of IROC-H PhantomsPhantom Patient

PhantomPatient

Page 7: IROC Houston QA Center end-to-end QA phantom program key

Photon Phantom Results

Phantom H&NLiver

insertLung Prostate Spine

Irradiations 2052 165 1109 566 336

Pass 1755 (86%) 120 (73%) 921 (83%) 484 (86%) 261 (78%)

Fail 297 45 188 82 75

Criteria 7%/4mm 7%/4mm 5%/5mm 7%/4mm 5%/3mm

Comparison between institution’s plan and delivered dose.

Page 8: IROC Houston QA Center end-to-end QA phantom program key

Problems that have been

detected

Page 9: IROC Houston QA Center end-to-end QA phantom program key

0 10 20 30 40 50 60 70 80 90520

540

560

580

600

620

640

660

680

700

Distance along line

Do

se V

alu

e

Calc

Meas

Criteria: 3%/2 mm

Varian 6 MV IMRT H&N

PAxial

9

0 10 20 30 40 50 60 70 80 90520

540

560

580

600

620

640

660

680

700

Distance along line

Do

se V

alu

e

Calc

Meas

Measurement vs. Monte Carlo

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2.0

LATERAL

L

R L

Page 10: IROC Houston QA Center end-to-end QA phantom program key

Lung Phantom TLD results

TLD results

Irradiation Lung Phantom

0.90

0.92

0.94

0.96

0.98

1.00

1.02

1.04

1.06

1.08

1.10

0 50 100 150 200 250

Irradiation

RP

C/I

nst

Average = 0.967 +/- 2.9%

274 irradiations

Algorithms included:

AAA/Superposition/MonteCarlo

Page 11: IROC Houston QA Center end-to-end QA phantom program key

Lung Phantom TLD resultsTLD results

Irradiation Lung Phantom

0.90

0.92

0.94

0.96

0.98

1.00

1.02

1.04

1.06

1.08

1.10

0 50 100 150 200 250

Irradiation

RP

C/I

nst

Average = 0.967 +/- 2.9%

274 irradiations

Algorithms included:

AAA/Superposition/MonteCarloAverage = 0.994 +/- 3.3%

30 irradiations

Algorithms included: MonteCarlo

Page 12: IROC Houston QA Center end-to-end QA phantom program key

Lung Phantom ResultsPercent of pixels passing 5%/3mm gamma criteria

9690

50

86

52

85

94

31

61

0

20

40

60

80

100

Pin

nac

le S

C

Ecl

ipse

AAA

Ecl

ipse

PB

XiO

SC

XiO

Cla

rkso

n

TomoThera

py SC

Multi

Pla

n MC

Multi

Pla

n PB

Corv

us PB

Pe

rce

nt

of

pix

els

pa

ss

ing

±5

%/3

mm

cri

teri

a

Page 13: IROC Houston QA Center end-to-end QA phantom program key

Proton Phantom Audits

• 136 proton phantoms irradiated, analyzed

Brain Liver Lung Prostate Spine TOTAL

Total Irradiations 25 14 38 38 21 136

# Passed 24 7 24 30 17 102

Pass Rate [%] 96% 50% 63% 79% 81% 75%

Page 14: IROC Houston QA Center end-to-end QA phantom program key

Lung Phantom Results

Page 15: IROC Houston QA Center end-to-end QA phantom program key

Lung Phantom Results

Page 16: IROC Houston QA Center end-to-end QA phantom program key

Lung Phantom Results

Page 17: IROC Houston QA Center end-to-end QA phantom program key

Summary

• Introduction of new technologies present challenges that need to be verified prior to treating patients

• End-to-end QA phantoms detect errors, improve dose delivery accuracy and provide confidence

• Improved compliance with protocol prescription specifications and reduced deviation rate