ise for tpp and high-resolution respirometry · a) electrodes rinsed with water, 99% ethanol, water...
TRANSCRIPT
OROBOROS INSTRUMENTS
high-resolution respirometry
Oxygraph-2k
www.oroboros.at
ISE for TPP+ and High-Resolution Respirometry
Zuzana Sumbalová, Mario Fasching, Erich Gnaiger
OROBOROS INSTRUMENTS, Innsbruck, Austria
Tetraphenylphosphonium (TPP+) accumulates in the mitochondrial matrix as a function of the mitochondrial membrane potential.
The TPP+ electrode is an ion selective electrode (ISE). The signal [V] of the TPP+ electrode is linearly dependent on the logarithm of free [TPP+].
The ISE system (OROBOROS INSTRUMENTS) consists of separate measuring and reference electrodes. The ISE are inserted through the stoppers into two chambers of the Oxygraph-2k and connected to the potentiometric channels of the O2k-MultiSensor system, for simultaneous recording of oxygen and TPP+.
Development of a TPP+ electrode
Tetraphenylphophonium chloride
Figure 2: Calibration test of a TPP+ electrode performed in Oxygraph-2k chamber in MiRO6, 37ºC with stirring speed 750 rpm. The calibration solution (1 mM TPPCl in 100 mM KCl) was injected automatically with TIP2k. A calibration range 0.1 -10 µM TPP+ was covered in 17 steps at 5 min intervals, using data recording intervals of 1 s.
Table 1: Typical performance parameters of OROBOROS TPP+ electrodes. Values are calculated from non-amplified signal.The critical sensitivity is defined at 40 mV/decade [mV/dec]. At this sensitivity, the critical [TPP+] was 0.7 µM. The absolute value of drift was measured in all calibration runs in the range 2-5 µM TPP+.
A long-term study over 3.5 months To evaluate the long-term stability of 10 TPP+ electrodes without change of membranes with rinsing procedure necessary for removal of inhibitors and uncouplers: water, 99% ethanol, water
10 TPP+ electrodes were followed for 3.5 months after mounting a new membrane:
● one calibration run (Fig. 2) per electrode every 7-14 days ● 10 runs each electrode
Ethanol: 5 electrodes rinsed with water, 99% ethanol, waterWater: 5 electrodes rinsed with water only
Table 2: Performance parameters of TPP+ electrodes in the course of the study. Values shown as mean ± SD are calculated from non-amplified signal. Ethanol: electrodes rinsed with water, ethanol, water. Water: electrodes rinsed with water only, *P<0.05 vs. 1st run in the same group, #P<0.05 vs. water group in the same run.Ethanol group:One electrode needed a new membrane after 7th run
Summary Characteristics of the electrodes:1. High sensitivity: > 40 mV/dec for [TPP+] >0.7 µM
2. Low detection limit: < 0.2 µM TPP+
Long-term performance of the electrodes:1. The zero signal, sensitivity, and critical [TPP+] did not change significantly within 3.5 months (Table 2).
2. The response time increased slightly in the course of the study, there is a trend of a faster increase in response times of ethanol-rinsed vs. water-rinsed electrodes (Figure 5).
Conclusions
1. Improved performance parameters of the OROBOROS TPP+ electrode are the basis for application in mitochondrial physiology with high-resolution respirometry.
2. The low detection limit allows for using low [TPP+] to avoid inhibition of mitochondrial respiration.
3. Rinsing with ethanol is possible, the same membrane can be used on a TPP+ electrode for many experiments. The membranes can be changed at intervals to ensure high performance.
Refferences:
1. Buck RP, Lindner E (1994) Recommendation for nomenclature of ion-selective electrodes (IUPAC recommendations 1994). Pure Appl. Chem. 66: 2527-2536.
Acknowledgement:
This work was supported by FEMtech, a program of the Federal Ministry for Transport, Innovation and Technology (BMVIT) to promote equal opportunities in research and technology.
Figure 3: The performance parameters of a TPP+ electrode over 3.5 months. a) Non-amplified raw signal [V] as a function of log[TPP+]. The parameter of a linear fit slope [mV/dec] represents the sensitivity of ISE. b) Sensitivity [mV/dec] of non-amplified signal (calculated for each point) as a function of log[TPP+]. The red arrow shows the detection limit of 0.2 µM, the black arrows show critical [TPP+] of 0.7 µM (see Table 1).
0
10
20
30
40
50
60
70
80
1.00E-07 1.00E-06 1.00E-05
[TPP+] M
1th
2nd
3rd
4th
5th
6th
7th
8th
9th
10th
-0.220
-0.200
-0.180
-0.160
-0.140
-0.120
-0.100
-0.080
1.00E-07 1.00E-06 1.00E-05
V
[TPP+] M
1th
2nd
3rd
4th
5th
6th
7th
8th
9th
10th
0.2 µM
Slope [mV/dec]
Design
Figure 4: Evaluation of the response time of a TPP+ electrode. a) 1st calibration run b) 10th calibration run
Results
Figure 5: The response time as a function of number of runs. a) electrodes rinsed with water, 99% ethanol, water b) electrodes rinsed withwater only. The response time was defined as the time after which the slope of the raw signal achieved the value -0.004 mV/s (IUPAC recommendation [1]), measured at the calibration step from 0.8 to 1 µM TPP+. The number of data points for time derivation was 10.
a) b)
0
10
20
30
40
50
60
70
80
90
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
response time (s)
T84
T89
T92
T96
T97
n of rinsing
Ethanol - rinsed electrodes
0
10
20
30
40
50
60
70
80
90
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
response time (s)
T86
T88
T93
T94
T95
n of rinsing
Water - rinsed electrodes
a)
b)
a)
b)
Figure 1: Set with 2 TPP+ and 2 reference electrodes
Range [h:min:s]: 1:45:002:20:002:02:301:45:001:27:301:10:000:52:300:35:00
pX C
alib
rate
d (Z
)
-0.12
-0.144
-0.168
-0.192
-0.216
Slop
e pX
(Z) [
mpX
/s]
0.008
0.004
0
-0.004
-0.008
2e-7 4e-7 5e-7 6e-7 8e-7 1.0e-6 1.2e-6 1.4e-6 1.6e-61e-7 2.5e-62e-6 3e-6 4e-6 5e-6 6.98e-6 9.96e-601TIP startP001 TIP end TIP startP001 P002 P003 P004 P005 P006 P007 P008 P009 P010 P011 P012 P013 P014 P015 P016 TIP end
pX R
aw S
igna
l (Z)
[V]
Range [h:min:s]: 0:02:001:27:001:26:401:26:201:26:001:25:401:25:201:25:00
pX C
alib
rate
d (Z
)
-0.154
-0.156
-0.158
-0.16
-0.162
-0.164
Slo
pe p
X (Z
) [m
pX/s
]
0.008
0.004
0
-0.004
-0.008
21P006
pX R
aw S
igna
l (Z)
[V]
Range [h:min:s]: 0:02:002:30:202:30:002:29:402:29:202:29:002:28:40
pX C
alib
rate
d (X
)
-0.14
-0.142
-0.144
-0.146
-0.148
Slo
pe p
X (X
) [m
pX/s
]
0.008
0.004
0
-0.004
-0.008
01P006
pX R
aw S
igna
l (X)
[V]
Features of the OROBOROS TPP+ electrodeISE membrane: + simple mechanical mounting
ISE membrane composition:
+ optimized to increase signal stability
and minimize side effects by chemical
substances such as ethanol and ADPHigh quality inner electrode: + signal stability
PVDF electrode housing:+ minimal O2 storage and diffusion
+ minimal carry-over of inhibitors
Integration into the O2k
chamber :
+ minimal oxygen diffusion
Calibration Rinsing Raw signal Sensitivity Critical [TPP+] |Drift| Response
test group @ 0 TPP+ @ 1-3 µM TPP+ @ 40 mV/dec of raw signal time[mV] [mV/dec] [µM] [mV/s] [s]
1st Ethanol (5) -199 ± 10 51 ± 4 0.9 ± 0.5 0.0007 14 ± 3Water (5) -213 ± 18 (3) 52 ± 4 1.0 ± 0.4 0.0006 16 ± 6
5th Ethanol (5) -200 ± 14 56 ± 6 0.6 ± 0.3 0.0004 49 ± 10*#
Water (5) -199 ± 8 54 ± 3 0.7 ± 0.2 0.0002 28 ± 11
10th Ethanol (4) -189 ± 18 54 ± 7 0.9 ± 0.4 0.0006 52 ± 19*Water (4) -198 ± 12 55 ± 5 0.4 ± 0.2* 0.0003 36 ± 14*
[V]
Raw signal @ 0 TPP+ Sensitivity @ 1-3 µM TPP+ Critical [TPP+] @ 40 mV/dec |Drift| of raw signal[mV] [mV/dec] [µM] [mV/s]
Mean ± SD -196 ± 14 53 ± 5 0.7 ± 0.3 0.0007n 96 99 82 99