iwaki electromagnetic metering pumpiwaki-pumps.com.au/manuals/eh_w.pdf · 2020. 5. 24. · thank...

64
Read this manual before use of product IWAKI Electromagnetic Metering Pump EH-W Type Instruction Manual

Upload: others

Post on 02-Feb-2021

9 views

Category:

Documents


0 download

TRANSCRIPT

  • Read this manual before use of product

    IWAKIElectromagnetic Metering Pump

    EH-W Type

    Instruction Manual

    T315-2 '03/01

    ( )Country codes

    IWAKI CO.,LTD. 6-6 Kanda-Sudacho 2-chome Chiyoda-ku Tokyo 101-8558 JapanTEL:(81)3 3254 2935 FAX:3 3252 8892(http://www.iwakipumps.jp)

    U.S.A. : IWAKI WALCHEM CorporationAustralia : IWAKI Pumps Australia Pty. Ltd.Singapore : IWAKI Singapore Pte. Ltd.Indonesia : IWAKI Singapore (Indonesia Branch)Malaysia : IWAKIm Sdn. Bhd.Taiwan : IWAKI Pumps Taiwan Co., Ltd.Thailand : IWAKI (Thailand) Co.,Ltd.Hong Kong : IWAKI Pumps Co., Ltd.China : GFTZ IWAKI Engineering & Trading Co., Ltd.China : IWAKI Pumps Co., Ltd. (Beijing office)China : IWAKI Pumps (Shanghai) Co., Ltd.Philippines : IWAKI Chemical Pumps Philippines, Inc.Korea : IWAKI Korea Co.,Ltd.

    TEL : (1)508 429 1440 FAX : 508 429 1386TEL : (61)2 9899 2411 FAX : 2 9899 2421TEL : (65)763 2744 FAX : 763 2372TEL : (62)21 690 6607 FAX : 21 690 6612TEL : (60)3 7803 8807 FAX : 3 7803 4800TEL : (886)2 8227 6900 FAX : 2 8227 6818TEL : (66)2 320 1303 FAX : 2 322 2477TEL : (852)2 607 1168 FAX : 2 607 1000TEL : (86)20 8435 0603 FAX : 20 8435 9181TEL : (86)10 6442 7713 FAX : 10 6442 7712TEL : (86)21 6272 7502 FAX : 21 6272 6929TEL : (63)2 888 0245 FAX : 2 843 3096TEL : (82)2 3474 0523 FAX : 2 3474 0221

    Germany : IWAKI EUROPE GmbHItaly : IWAKI Italia S.R.L.Denmark : IWAKI Pumper A/SSweden : IWAKI Sverige ABFinland : IWAKI Suomi OyNorway : IWAKI Norge ASFrance : IWAKI France S.A.U.K. : IWAKI PUMPS (UK) LTD.Switzerland : IWAKI (Schweiz) AGAustria : IWAKI (Austria) GmbHHolland : IWAKI Holland B.V.Spain : IWAKI Iberica Pumps, S.A.Belgium : IWAKI Belgium n.v.

    TEL : (49)2154 9254 0 FAX : 2154 1028TEL : (39)02 990 3931 FAX : 02 990 42888TEL : (45)48 24 2345 FAX : 48 24 2346TEL : (46)8 511 72900 FAX : 8 511 72922TEL : (358)9 2742714 FAX : 9 2742715TEL : (47)66 81 16 60 FAX : 66 81 16 61TEL : (33)1 69 63 33 70 FAX : 1 64 49 92 73TEL : (44)1743 231363 FAX : 1743 366507TEL : (41)32 3235024 FAX : 32 3226084TEL : (43)2236 33469 FAX : 2236 33469TEL : (31)297 241121 FAX : 297 273902TEL : (34)943 630030 FAX : 943 628799TEL : (32)1430 7007 FAX : 1430 7008

  • Thank you for having selected of Iwaki's EH/W type electromagnetic meteringpump.

    This instruction manual, which is divided into 5 sections, namely "Safety Section,""Outline Section," "Installation Section," "Operation Section" and "MaintenanceSection," deals with the correct handling and operation procedures for the pump.

    You are requested to read this manual prior to using the pump, to ensure safe andlong operation. If external signals are to be input into the pump, please consult us inadvance.

    * Please make sure the end user gets a copy of this manual.

    ContentsIMPORTANT INSTRUCTION ······················································· 1

    Safety Section ········································································ 2~4

    OUTLINE OF PRODUCT ······················································ 5~131. Unpacking and Inspection ························ 62. Operation Principle ··································· 63. Identification Codes ······························· 7,84. Main Features ·········································· 95. Specifications ···································· 10,116. Description of Parts ··························· 12,13

    INSTALLATION ····································································· 14~251. Before Use ············································· 152. Notes on Operation ·························· 15~173. Installation, Piping, and Wiring ········· 18~25

    PUMP OPERATION ····························································· 26~441. Operation ········································· 27~322. Methods of Pump Operation ············ 33~44

    MAINTENANCE ···································································· 45~571. Causes of Trouble and

    Troubleshooting ······································ 462. Maintenance and Inspection ·················· 473. Disassemble and Reassembly ········· 48~514. Accessories ············································ 525. Names of Parts and

    External Dimensions ························ 53~57

    ■ Application of Warranty and Repair Service ·················· 58

    This instruction manual should be kept on hand by the end user for quick reference. It isrecommended that each user, after reading the instruction manual thoroughly, place it in aposition close to the pump system and where it may be easily accessed by any user atany time whenever necessary.

    Concerning any points you wish to know more about, contact Iwaki, an Iwaki branch office,or an Iwaki distributor.

  • - 1 -

    Caution

    Types of Symbols

    WarningNonobservance or misapplication of the contents of the "Warning" section could lead to a serious accident, including death or injury.

    Nonobservance or misapplication of the contents of the "Caution" section could lead to a serious physical injury to the user or serious damage to the product.

    Indicates that "Warning" or "Caution" must be exercised. Inside this triangle, a concrete and practical image provided as a warning or caution message is depicted.

    Indicates a prohibited action or procedure. Inside or near this circle, a concrete and practical image of the activity to be avoided is depicted.

    Indicates an important action or procedure which must be performed or carried out without fail. Failure to follow the instructions herein can lead to malfunction or damage to the pump.

    For the Safe and Correct Handling of the Pump

    • Read the "Safety Instructions" sections carefully to prevent accidents involving your customers orother personnel and to avoid damage or loss of other assets. Always follow the instructions andadvice found in these sections.

    • Observe and abide by the instructions described in this manual. These instructions are veryimportant for protecting pump users from dangerous conditions and situations related with theuse of the pump system.

    • The symbols relate to the following meanings described below:

    Important Instruction

  • - 2 -

    WARNING • Turn off the power supply.

    Working without disconnecting the power supply may cause an electrical shock.

    Before engaging upon any working procedures involving the pump, make sure to

    turn the power supply switch off and to stop the pump and other related devices.

    • Terminate operation!

    When you detect or become aware of a dangerous sign or abnormal condition

    during operation, terminate the operation immediately and start it from the beginning

    again.

    • For specified application only.

    The use of a pump in any application other than those clearly specified may result in

    injury or damage to the pump. Use the pump strictly in accordance with the pump

    specifications and application range.

    • No remodeling!

    Never remodel a pump. Otherwise, a serious accident may result. Iwaki will not be

    responsible for any accident or damage of any kind which is caused by the user

    remodeling the pump without first obtaining permission or instructions from Iwaki.

    • Wear protectors.

    If you touch or come in contact with any type of hazardous chemical liquid, including

    but not limited to chemicals, you may experience a serious injury. Wear protective

    gear (protective mask, gloves, etc.) during the pump operation.

    • Operating site must be free of water and humidity

    The pump is not designed to be water-proof or dust-proof. The use of the pump in

    places where water splashes or humidity is high may result in an electrical shock or

    short circuit.

    CAUTION

    • Qualified operators only!

    The pump operator and pump operation supervisor must not allow any operators

    who have little or no knowledge of the pump to run operate the pump. Pump

    operators must have a sound knowledge of the pump and its operation.

    • Specified power only.

    Do not operate the pump on voltage which is not specified on the nameplate.

    Failure to do so may result in damage or fire. Only the specified power level is to be

    applied.

    Safety Section

    Electrical Shock

    Prohibited

    No Remodeling

    Prohibited

    Wear protective gear

    Prohibited

    Prohibited

  • - 3 -

    CAUTION• Do not run the pump dry.

    Do not run the pump dry (without liquid inside the pump). Heat generated as a

    result of abrasion between elements inside the pump during operation without liquid

    may damage the inside of the pump.

    • Do not wet or dampen!

    If an electric part or wiring gets wet with the liquid spilled over accidentally, a fire or

    electrical shock may be caused. Install the system in a place free from liquid

    spillage or leakage.

    • Ventilate!

    Poisoning may result during an operation which involves toxic or odorous liquid.

    Ventilate the operating site sufficiently.

    • Spill-out accident!

    Protective measures should be taken against any accidental spill-out or leakage of

    the operating liquid as a result of unexpected damage on the pump or the related

    piping.

    • Damaged pump

    Never operate a damaged pump. A damaged pump may cause leakage or

    electrical shock.

    • Do not damage or change power cable!

    Do not scratch, damage, process, or pull the power cable forcibly. An extra load

    onto the cable, such as heating the cable or placing something heavy on the cable,

    may damage the cable and finally cause a fire or an electrical shock.

    • Arrange grounding!

    Do not operate the pump without connecting the grounding wire. Otherwise, an

    electrical shock may result. Make sure the grounding wire is connected with the

    grounding terminal.

    • Install an earth leakage breaker (option)!

    The operation of a pump without using an earth leakage breaker may cause an

    electrical shock. Please purchase an optional leakage breaker and install in the

    system.

    • Handling of power cable

    Use of a defective or damaged power cable may result in a fire or electrical shock.

    Handle the power cable carefully.

    Prohibited

    Electrical Shock

    Caution

    Caution

    Caution

    Grounding

    Safety Section

    Do not wet or dampen

    Prohibited

    Electrical Shock

  • - 4 -

    Safety Section

    CAUTION • Follow the instruction manual

    Replace the consumable parts by following the descriptions in the instruction manual.

    Do not disassemble any part of the pump if the disassembling procedure for the part

    in question is not included in the instruction manual.

    • Limited operating site and storage

    Do not install or store the pump in the following places:

    * Places where a flammable gas or material is used or stored.

    * Places where the ambient temperature is extremely high (40°C or higher) or

    extremely low (0°C or lower).

    • Disposal of used pump

    Disposal of used or damaged pumps must be done in accordance

    with the relevant local laws and regulations. (Consult a licensed

    industrial waste products disposing company.)

    Prohibited

  • - 5 -

    OUTLINE OF PRODUCT

    1. Unpacking and Inspection ................. 6

    2. Operation Principle ............................ 6

    3. Identification Codes ........................... 7

    4. Main Features ................................... 9

    5. Specifications .................................. 10

    6. Description of Parts ......................... 12

  • - 6 -

    After unpacking the pump, check the following

    points to ascertain that the product is exactly as you

    ordered. If you find anything wrong, please get in

    touch with your dealer.

    z Do the model, discharge pressure, voltage, etc.,

    shown on the nameplate represent what you

    ordered?

    x Is the following accessory included in the

    package?

    • Key protector. . . . . . . . . . . . . . . . . . . . . . . . . . . 1

    * Install the key protector in the control unit.

    c Has the pump been damaged in transit? Are the

    bolts and nuts loose?

    1. Unpacking and Inspection

    The Iwaki EH series electromagnetic pump is a

    diaphragm metering pump of the 'linear

    electromagnetic drive system', which means

    electromagnetic force drives the diaphragm directly.

    It consists of a pump unit, a drive unit and a control

    unit. Reciprocating movement is caused by the

    electromagnetic force produced by the pulse current

    sent from the control unit, and the force of a spring.

    The reciprocating movement is transmitted to the

    diaphragm directly connected to a plunger, to

    change the capacity of the pump chamber. The

    changed capacity and the functioning of the valves

    in the pump unit drive the pump.

    2. Operation Principle

    Iwaki MeteringPump

    CAPACITY

    mR/min

    MAX.PRESSURE MPa

    STROKE RATE spm

    VOLTAGE

    MODEL

    MFG.No.

    IWAKI CO.,LTD. TOKYO JAPAN

    CURRENT ATHERMALLY PROTECTED

    POWER CONSUMPTION

    1P41

    6987

    Discharge

    Discharge pressure

    Voltage

    Power consumption

    Electric current

    Model

    Manufacturingnumber

    W

    V

    Key protector

    Control unit Pump unit

    Stroke lengthadjusting dial

    Drive unit(body of pump) Plunger

    Diaphragm

    Spring

  • ■ Pump Body

    EH - B 20 VC - 100 P W 1 - PH MMq w e r t y u i o !0

    q Series code

    w Drive unit type (average of power consumption)

    B: 16W C: 24W

    e Effective diameter of diaphragm

    10: 10mm 15: 15mm 20: 20mm

    30: 30mm 35: 35mm

    r Materials of pump unit (liquid-contacting parts)

    - 7 -

    3. Identification Codes

    Parts

    Code

    PumpHead

    Valve Ball O ring Gasket Diaphragm

    Hastelloy C276

    VC

    VH

    PC

    PH

    FC

    PVC

    GFRPP

    PVDF

    Ceramic

    Hastelloy C276

    Ceramic

    Ceramic

    FKM

    EPDM

    FKM

    EPDM

    FKM

    EPDM

    FKM

    EPDM

    FKM

    EPDM

    FKM

    EPDM

    PCTFE

    PTFEPTFE

    +EPDM

    Valve seat

    Material code PVC: Transparent vinyl chloride

    GFRPP: Glass-fiber reinforced polypropylene

    PVDF: Polyvinylidene fluoride

    FKM: Fluoroelastmer

    EPDM: Ethylene propylene diene methylene

    PTFE: Polytetrafluoroethylene

    PCTFE: Polychlorotrifluoroethylene

    t Line voltage

    100: AC 110/115 V 20E: AC 220-240 V

    y Power cable type

    P: With an attaching plug

    No indication: With press-fit terminals (for 220/240V)

    u Control unit type

    W: Water treatment type

    i Connection tube diameter (inner diameter × outer diameter)2: φ4 × φ63: φ6 × φ85: φ9 × φ126: φ10 × φ12

  • - 8 -

    EHC - 100B P W - PHq

    q

    w

    e

    r

    t

    y

    w e r t y u

    Type of control unit

    Line voltage

    Power supply cord type

    Control function type

    Controller unit model code

    Type of drive unit(power consumption)

    EHC

    W: Water treatment type

    uIndication of specialspecifications

    01~99: Special specificationsNo indication: Standard specifications

    PH: pH controlPH1PH2PH2BOR: ORP controlCD: Conductivity Control

    B: For EH-B pump (16W)C: For EH-C pump (24W)

    100: AC110/115V20E: AC220, 230, 240V

    P: With attaching plugNo indication: With press-fit terminals

    o Control function type

    * PH: pH type (pH control function, A.T.C: positive thermistor 6.8Ω at 25 ºC)PH1: pH type (pH control function, A.T.C: Pt1000)

    * PH2: pH type (pH control function, A.T.C: Pt100)

    PH2B: pH type (pH control function, A.T.C: Pt100 with BNC plug)

    OR: ORP type (oxidation-reduction potential control function)

    CD: Conductivity type (conductivity control function)

    *!0 Special specification code

    01~99: None-standard materials, different diameter fitting, etc.

    The specifications marked with asterisks are special specifications, while the rest are standard specifications.

    ■ Control Unit

    ■ Points to Note

    q Do not detach the control unit unless necessary.

    w The control unit must not be attached to a pump with a different drive unit coad or line voltage code. (Verify with

    the name plate.)

    CautionIf the control unit is attached to a pump with a different drive unit code or line voltage code, it will cause a problem or

    failure of in the electronic circuits of the control unit and/or the drive unit of the pump.

  • - 9 -

    ■ Control Unit

    The pump operation is electronically controlled by the control unit which has a built-in CPU of its own (emphasis is

    laid on the facility of controlling operation and functions).

    PH model ············The pump stroke rate is controlled in proportion to the pH value measured by the electrode, for

    accurate control of pump discharge.

    OR model············The pump stroke rate is controlled in proportion to the ORP value measured by the electrode, for

    accurate control of pump discharge.

    CD model············The pump stroke rate is controlled in proportion to the Conductivity value measured by the senser,

    for accurate control of pump discharge.

    ■ Simple operation

    As the operating keys of the control unit are used to run the pump and to control the flow rate, system control of the

    flow rate can be carried out easily.

    ■ Memory backup function

    Due to the memory backup function, set values remain stored even after the power has been turned off. When power is

    restored, the pump operates with these set values.

    4. Main Features

    Pump unit• Reciprocating movement of the diaphragm

    changes the capacity of the pump chamber,to drive the pump.

    Stroke length control unit• Adjusts or sets the discharge per stroke in the range of 0 to 100%.

    Drive unit• Causes the diaphragm to reciprocate, by means of the

    electromagnetic force and the force of the spring in responseto command signals received from the control unit.

    Control unit• Controls pump operation, flow rate and

    adjustment, and setting of the stroke rate.

  • - 10 -

    5. Specifications

    Note 1: The table shows the results of tests with clear water at the rated voltage and normal temperature.

    Note 2: The maximum discharge is the amount under the maximum discharge pressure (stroke length

    100%; stroke rate 100%). Under a lower pressure, the discharge is larger than shown in the table.

    Note 3: Operating ambient temperature: 0~40°CNote 4: Liquid temperature: 0~60°C (VC, VH : 0~40°C)Note 5: Permissible voltage fluctuation: Within ±10% of ratingNote 6: The connection tube size φ 6 × φ 8 is not applicable to PC and PH types. Note 7: The FC type does not have B15, B30, or C15.

    Note 8: The applicable sizes of connection tubes for the FC type are φ 4 × φ 6 for B10, B20, and C20 andφ 10 × φ 12 for C30 and C35.

    * Please consult us for special specifications such as slurry liquid handling.

    * Please note that for the purpose of improvement, parts which have no effect on specifications and fitting sizes are

    subject to change without prior notice.

    ■ Sound level

    The sound level of the each type of the pump is 60 dB (A scale) or less at a distance of one meter

    Note 6: Permissible voltage fluctuation: Within ±10% of rating

    * Please consult us for special specifications such as in the case of handling slurry.

    * Please note that for the purpose of improvement, parts which have no effect on specifications and fitting

    sizes are subject to change without prior notice.

    50/60Hz

    EH-B10 38 0.04-0.11 0.98

    0 ~ 360

    40 ~ 100(0.4~1.0)

    16EH-B15 70 0.04-0.19 0.69

    EH-B20 105 0.06-0.29 0.39 20 ~ 100(0.2~1.0)

    EH-B30 230 0.13-0.64 0.20

    EH-C15 90 0.05-0.25 0.98

    24EH-C20 130 0.07-0.36 0.69 20 ~ 100

    (0.25~1.25)EH-C30 270 0.15-0.75 0.34EH-C35 450 0.25-1.25 0.20

    FC/SHCapacity[ml/min]

    Capacity[ml/stroke]

    Max.pressure[MPa]

    Stroke frequency[spm]

    Stroke length[%](mm)

    Average powerconsumption[W]

    EH-B10 38 0.04-0.11 0.98 40 ~ 100(0.4~1.0)

    EH-C20 120 0.07-0.33 0.6920 ~ 100

    (0.25~1.25)24EH-C30 260 0.14-0.72 0.34

    EH-C35 410 0.23-1.14 0.20

    0 ~ 360EH-B20 95 0.05-0.26 0.39 20 ~ 100

    (0.2~1.0)

    16

    ■ Pump specifications

    VC/VHPC/PH

    Capacity[ml/min]

    Capacity[ml/stroke]

    Max.pressure[MPa]

    Stroke frequency[spm]

    Stroke length[%](mm)

    Average powerconsumption[W]

  • - 11 -

    ■ Specifications of Control Unit

    Mode Mode

    Change

    PHAUTOMAN

    Code ORAUTOMAN

    Entering by keys

    CDAUTOMAN

    KAUTO• SET• PB• spm• Lo• Hi

    KMAN• spm

    0~10mS/cmCell Constant 0.80~2.20

    Temp.Coefficient:: -9.99~9.99% / ºC

    KAUTO• SET• PB• spm• Lo• Hi

    KMAN• spm

    -1999~+1999mVZERO

    4 keys START/STOP , ▲ , s , FUNCBy recieving stop signal from outside

    4 Digit LCD Measure, Set Value, Error, and othersGreen LED

    No voltage contactBy recieving stop signal

    Synchronized output with pump stroke

    Lower than "Low" or higher than "Hi"

    Open collector Sink type (Isolated)Max.voltage DC35V, Min. Current DC2.5mA

    KAUTO• SET• PB• spm• Lo• Hi

    KMAN• spm

    Prameters

    KeysStop

    DisplayStroke

    Sensor

    StopStop

    SynchronizedstrokeLo/HiWithin

    SET Value

    Form

    Function

    Indication

    Input

    Output*1(Select one)

    0~14pHZERO, SPAN (ATC)

    With SET Value±0.2mS / cm

    With SET Value±20mV

    With SET Value±0.20pH

    DC0~1V (0~10mS/cm)Max. load

    10kΩ or more

    DC0~1V (-1999~1999mV)Max. load

    10kΩ or more

    DC0~1V (0~14pH)Max. load

    10kΩ or moreRecord out

    0~9.99mS/cm-9.99~9.99mS/cm0~360spm-9.99~9.99mS/cm-9.99~9.99mS/cm

    0~360spm

    -1999~+1999mV-1999~+1999mV0~360spm-1999~+1999mV-1999~+1999mV

    0~360spm

    0~14pH-14~14pH0~360spm0~14pH0~14pH

    0~360spm

    Note: *1 Standard output is set "STOP". Select one output from Table

    When select output except STOP, inform with ordering.

  • - 12 -

    6. Description of Parts

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    Unit displayRemains lit while the temperature display is selected. This display is not provided on the "OR" model.

    Display panelSet values and error messages are displayed.

    ALM displayRemains lit if the measured value exceeds the upper or lower limit value.

    Unit displayRemains lit when stroke rate display is selected. Flashes if the stroke rate has reached the upper limit value.

    Unit displayRemains lit when the measured value or parameter is displayed. Displays "mV" for an OR type and "mS/cm" for a CD type.

    Hi displayFlashes if the measured value exceeds the upper limit value. Remains lit when the parameter is displayed.

    Lo displayFlashes if the measured value exceeds the lower limit value. Remains lit when the parameter is displayed.

    Function keyPressed to determine a set value or store a value. Also used in multiple-key operation in combination with another key for other selection purpose.

    PB displayRemains lit when proportional band value is selected for parameter display.

    Operation mode displayRemains lit when automatic operation mode (AUTO) is selected. Flashes during automatic operation.

    Operation mode displayRemains lit when manual operation mode (MAN) is selected. Flashes during manual operation.

    ZERO displayRemains lit when the calibration set value is selected for parameter display.

    ON lampRemains lit while power is supplied. Flashes in synchronization with the pump stroke.

    SPAN displayRemains lit when the calibration set value is selected for parameter display.

    START/STOP keyPressed to start or stop pump operation.

    Shift keyPressed to move to the next rightward digit for value setting. Also used in multiple-key operation in combination with the function key for the selection of an operation mode.

    SET displayRemains lit when the SET value is selected for parameter display. Also remains lit when any other set value is displayed.

    ZERO is not provided on the CD type, which employs CELL instead. SPAN is not provided on the OR type and the CD type. The CD type employs % instead. For detailed operation procedures for these parameters, refer to the "Operation" section."Multiple-key operation" means that the particular key is pressed with another key simultaneously. In this unit, the function key and another key are pressed simultaneously.The operation surface of the control unit is covered with protective film for shipment. Remove this film before using the control unit.

    UP keyCalls up the set value while a parameter is displayed. Increases the value of the flashing digit when a value is to be set. Used also in multiple-key operation in combination with the function key for the selection of the display content.

    Control unit

    Stroke rate adjusting dial

    Drivel unit

    Pump unit

    ■ Names of Parts

    ■ Control unit

  • - 13 -

    ■ Basic display patterns

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    Display pattern Description

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    AUTO mode operation is under way.The value may be 0 spm with the pump operation suspended, depending on the set value and the measured value.

    The pump is in standby condition in MAN mode.Under normal conditions, WAIT is displayed while the pump is in standby condition or the stroke rate is displayed while the pump is in operation.

    The control unit is ready for mode selection.MAN or AUTO remains flashing. This mode selection display can be accessed from the WAIT (standby) display.

    Set value is being changed.The parameter display has been replaced by this display and the set value is ready to be changed. The value of the flashing digit can be changed by pressing the UP key.

    Error displayIf an attempt has been made to enter a value out of the setting range as set value, this display flashes for 5 seconds.

    Error displayIf the parameter PB has been set at 0.00 or -0.00, this display flashes for 5 seconds.

  • - 14 -

    INSTALLATION

    1. Before Use ........................................15

    2. Notes on Operation .......................... 15

    3. Installation, Piping, and Wiring ........ 18

  • - 15 -

    Caution• Turn off the power supply.

    Working without disconnecting the power supply may cause an electrical shock. Before engaging upon

    any working procedures involving the pump, make sure to turn the power supply switch off and to stop

    the pump and other related devices.

    • Terminate operation!

    When you detect or become aware of a dangerous sign or abnormal condition during operation,

    terminate the operation immediately and start it from the beginning again.

    • Specified power only.

    Do not operate the pump on voltage which is not specified on the nameplate. Failure to do so may

    result in damage or fire. Only the specified power level is to be applied.

    • Keep from heat or flame.

    Do not place any dangerous materials or flammable objects near the pump for the prevention of fire or

    accident.

    • Damaged pump

    Never operate a damaged pump. A damaged pump may cause leakage or electrical shock.

    1. Before Use

    “Strictly observe the following points.”

    Operators and maintenance service staff must read the instruction manual thoroughlybefore using the products. Do not operate the pump system unless all of the contents inthe manual are completely understood.

    2. Notes on operation

    • Dropping the pump or subjecting it to strong impacts

    may result in faulty performance. Handle the pump

    with care.

  • - 16 -

    • The pump can be operated outdoors, as the pump

    employs a simple water and dust proof structure.

    When installing the pump, avoid places exposed to

    direct sunlight or direct rain with an ambient

    temperature of above 40ºC, or with a relative humidity

    of above 85%. Though the pump has a simple

    waterproof and dustproof structure, a sheltered location

    is recommended.

    • Select an installation site convenient for future

    maintenance and inspection, and fix the pump on a

    level floor so that it is free of vibrations.

    • Ventilate!

    Poisoning may result during an operation which

    involves toxic or odorous liquid. Ventilate the

    operating site sufficiently.

    • Do not wet or dampen!

    If an electric part or wiring gets wet with the liquid

    spilled over accidentally, a fire or electrical shock may

    be caused. Install the system in a place free from liquid

    spillage or leakage.

    • Install an earth leakage breaker (option)!

    The operation of a pump without using an earth leakage

    breaker may cause an electrical shock. Please purchase

    an optional leakage breaker and install in the system.

    • The control unit can be detached. However, it should

    not be detached unless unavoidable. Never use the

    detached control unit with the pump models other than

    the one delivered.

    Prohibited

    Caution

  • - 17 -

    • Arrange grounding!

    Do not operate the pump without connecting the

    grounding wire. Other wise, an electrical shock may

    result. Make sure the grounding wire is connected with

    the grounding terminal.

    • Limited operating site and storage

    Do not install or store the pump in the following places:

    * Places where a flammable gas or material is used or

    stored.

    * Place where the ambient temperature is extremely high

    (40ºC or higher) or extremely low (0ºC or lower).

    • Cleaning

    Wiping the pump body or the nameplate with a cloth

    soaked in a solvent such as benzene, thinner or

    kerosene may remove or change the color of the

    coating. Use a dry cloth or a cloth soaked in water or

    neutral detergent.

    • If the pump is used for industrial porpose, the electric

    power cable of the pump must be installed in the

    enclosure which satisfies IP54 or in the atmosphere

    corresponding to IP54

    Prohibited

    Prohibited

  • - 18 -

    3. Installation, Piping, and Wiring

    ■ Installation

    z Installation

    Install the pump at a place where the ambient

    temperature does not exceed 40ºC and the relative

    humidity does not exceed 85%. (There should be no

    dew condensation inside the control unit.) The place

    should be convenient for future maintenance and

    inspection. A sheltered location is recommended.

    x Place the pump as close to the suction tank as possible

    so as to realize flooded suction piping (install the pump

    at a lower level than the tank on the suction side).

    c Fix the pump firmly by means of m5 screws on a level

    floor free from splashing liquid.

    If the pump is inclined, the discharge may be decreased

    or may even reach zero.

    v If the pump is used for a liquid that tends to generate air

    bubbles (sodium hypochlorite, hydrazine solution, etc.),

    make sure to install it in a cool, dark place away from

    direct sunlight. When a tank, etc., is used, flooded

    suction piping is required.

    Discharge side

    Suction side

  • - 19 -

    ■ Piping

    z Piping of pump fittings

    For discharge side and suction side piping, use the tube

    provided. The connections should be firm enough to

    prevent liquid leakage or air suction.

    • Since the fitting nuts are made of plastic, take care not

    to tighten them excessively, or damage may occur to

    the fittings

    • If the tube provided is not long enough, use, for

    example, PVC braided tube with a diameter matching

    that of the pump fittings.

    x Piping for air bleeding

    As a safety measure, prior to bleeding air, connect the

    tube to the discharge port of the air vent unit so that the

    bled air can be discharged into the suction tank, a drain

    plate, etc.

    (See 'Air Bleeding' on page 23.)

    [Note]

    The B30, C30, C35 and FC type is not equipped

    with an air vent unit.

    c Installation of check valve

    The pump is provided with a check valve to prevent

    overfeeding. The check valve must be used in any of

    the following cases:

    q The liquid surface on the suction side is higher than

    that on the discharge (fig. A.)

    The pressure at the feeding position is atmospheric

    pressure, and liquid is fed at a position lower than

    the suction side liquid surface.

    tube

    Fitting nut

    Fitting

    Check valve

    Discharged into the suction tank

    Ssuction side

    Discharge side

    Fig. A

  • - 20 -

    w If the pressure on the suction side is higher than that

    on the discharge side (Fig. B).

    e The liquid surface on the discharge side liquid is

    higher than that of the suction side, but the

    difference between the two levels is 5 m or less (Fig.

    C).

    r The load pressure (pipie resistance, discharge pump

    head, etc.) applied to the pump is 0.13 MPa

    {1.3 kgf/cm2} or less.

    • Install the check valve at the edge of the discharge side

    connecting tube. It must be least 1 m away from the

    pump .

    • The connection port of the CA type check valve can be

    mated with not only the tube but R1/2 and R3/8 threads.

    Cut off the unnecessaly section with a pipe cutter.

    (For applicable pump types, see 'Accessories' on page

    52.)

    • Besides the CA type, the CB double-ended type

    (optional item) for connection to braided tube is

    available.

    • A back pressure valve (optional item) can be used in

    place of the check valve. For details, can Iwaki.

    Fig. B

    Fig. C

    R1/2 R3/8 Diameter φ 9

    CA type check valve

    BV type back pressure valve

    BVC type back pressure valve

    CB type check valve

    5m

  • - 21 -

    I. Wiring

    ■ Points to be checked before wiring

    z Make sure that the main power switch is in OFF position (or power supply is disconnected).

    x The pump is turned ON or OFF by operating the key on the control unit. For ON/OFF switching of the power

    supply by the use of relay, use a relay with a contact rating of 15A or more. A relay with a lower rating than that

    may be destroyed (by contact fusion). To meet the application need to stop the pump frequently, use the STOP

    function with the power supply continuously connected. The figure below shows the recommended wiring.

    Use the contact "a" as shown above, especially when the pump ON/OFF switching is interlocked with the power

    ON/OFF switching of another water supply pump.

    c The electrical connection should be carried out by an

    authorized electrician in accordance with local

    regulations. (Use quality wiring materials only and

    observe the technical standards of electrical installation

    and interior wiring regulations.)

    • Connection with 110V power supply

    The power cable has a connection plug at its end.

    Insert the plug into a 110V plug socket.

    • Connection with 220-240V power supply

    The power cable has a press-fit terminal at its end.

    Connect the terminal firmly.

    Caution

    Do not connect the control unit with a power source which

    supplies power higher than that which is rated. Otherwise,

    damage to the electric circuits in the control unit may be

    sustained.

    EH/W4

    6

    Ry Contact "b"

    Contact"a"

    Relay

    Powersupply

    Caution

    Only the operator/service staff who is trained for safety handling procedures should be in charge of the

    related electric arrangement and control of the power source. Failure to observe this instruction may result

    in injury to personnel or damage to property.

  • - 22 -

    v Make sure to establish a ground.

    b Do not obtain a power supply from the same power

    service outlet that also supplies power to high-power

    electric equipment which generates surge voltage which

    results in a malfunction of the electric circuits in the

    control unit. Be careful about the noise generated by an

    inverter or the like as well.

    n The control unit can be easily detached from the pump

    drive unit. However, do not connect the control unit

    with a pump which is specified for a different operation

    voltage or drive unit. In order to prevent damage to the

    electronic circuit and the drive unit, use the control unit

    with the delivered pump only.

    • Surge voltage

    The electronic circuits in the control unit may be

    affected by an excessively strong surge voltage. It is

    recommended that the control unit should be distanced

    from high-power electric equipment of 200V or higher

    that generates surge voltage. The following measures

    may be taken, depending on the circumstances.

    q Use a surge absorbing element (such as a varistor

    with surge resistance of 2000A or more) at the

    power connection of the pump.

    w Install a noise-cutting transformer.

    Surge absorbing element

    Noise-cutting transformer

  • - 23 -

    ■ Wiring

    • Names of parts

    • Connection of pH or ORP electrode

    Set the respective wiring terminals of the pH electrode as shown in the

    figure. Check the terminals with the respective tags attached for a

    correct connection. The electrode which has a model number ending

    with "TV" has 5 wires on it. The electrode which has a model number

    ending with "V" has 2 wires on it. In the case of the TV type, remove

    the resistor installed at the T-T terminals before connecting the wires.

    An OPR electrode has two wires only, which are not to be connected

    to the terminals T and E. No resistor is attached, either.

    GLASSE T T REF

    Green Black Red Grey Grey (Large)

    TC Ref pH

  • - 24 -

    Caution

    All the wires are prepared especially for the control unit. Do not attempt to modify or extend them, for such a attempt

    may affect measurement accuracy.

    Caution

    If a wire needs to be extended, make sure to use a relay box. Connect the wires within the relay box correctly.

    Follow the instruction manual attached to the electrode. In the case of a pH electrode provided with a temperature

    compensating element, keep the resistor installed at the T-T terminals and do not connect any wire to the T terminal.

    (Terminate the ends of wires upon wiring.)

    • Connection of conductivity sensor

    Conductivity sensor ESC-150P1-06SV only is applicable to the

    CD type. Note that no other type of sensor is applicable.

    Wire the sensor as shown in the figure. Be careful with the

    color mark on each wire. Insert the top of each wire into the

    terminal and fasten the connection with the use of a flat-blade

    precision screwdriver. The fastening torque should be 4 kgf·cm

    or lower.

    Caution

    All the wires are prepared especially for the control unit. Do not attempt to modify or extend them, for such a attempt

    may affect measurement accuracy.

    • When stop function is used

    Stop function refers to a method to stop the

    pump operation temporarily by sending an

    external no-voltage contact signal (contact "a").

    The wiring required for the use of this function

    should be done in accordance with the following steps. Connect the wires to the removable connectors "4" through

    "6". After completing the connection, pull out the connectors once and then return them to their original place.

    Before inserting the tops of the wires to the terminals, unsheathe them for a length of 5 mm from the ends. Fasten

    the connectors with a flat-blade precision screwdriver. The fastening torque should be 4 kgf·cm or lower.

    • For open collector

    Polarity should not be mistaken. "4" is for plus (+) and "6" is for minus (-). (Max.

    rating 5V, 1 mA)

    • For relay or other contact means

    A relay or a contact should be designed for electronic circuits and has a min. load of 1

    mA or less.

    1 2 3 4 5

    Green Black White Red Shielding wire

    PUMP

    External contact

    ON

    CLOSEOPEN

    4 5 6

    Wiring for open collector

  • - 25 -

    • When a recording output (recorder) is used

    The recorder will output in the range of DC0 through 1V for the measurement ranges of respective device models.

    The connected load should be 10 kΩ or more. Take the same wiring as in the case of the stop function above.Connect wires to the connectors "5" and "6." Be careful not to mistake the polarity, with "5" for + and "6" for -.

    • When open collector output is used

    In the case of PH or OR model, connect the wires to the + and - terminals on the terminal block on the substrate.

    In the case of the CD model, connect wires to terminals 1 and 2 on the connector CN12. ("1" is for +

    and "2" is for -.)

    Use an M3 press-fit terminal at the end of a wire for the PH or OR model. Unsheathe the end of the

    wire by 5 mm for the CD model.

    The output setup is shown in the figure on the right. The min. suction current is 2.5 mA. Use a

    converter or the like to drive a relay or AC device.

    ■ Wiring Method

    Caution

    Do not carry out wiring with the pump in operation. Be sure to turn the power off before wiring. Otherwise, an electric

    shock or short circuit may be caused which will result in the malfunction of the pump.

    Caution

    The internal circuit remains electrically charged immediately after the power is turned off. Wiring work should be begun

    at about 1 minute following the turn-off.

    z Remove the retaining screw on the terminal box and open the terminal box.

    x Take out the removable connector and the PCB connector.

    c Remove other cable nuts which are not used for wiring. Take out the blind cap and the cable gasket from the

    terminal box. Then, set the cable nut and the cable gasket onto the cable to be connected. Insert the cable inside the

    terminal box.

    v Use a proper connection tool for a reliable connection.

    b Upon completion of the connection work above, return the connectors to their original place.

    Caution

    Check that the PCB connector is inserted firmly. Failure to secure the PCB in a complete connection may cause a

    malfunction of the pump system.

    n Install the terminal box by pressing the flange against the recess in the control unit housing. Fasten the retaining

    screw firmly. Then, adjust the excess cable inside the terminal box and fasten the cable nut.

    Caution

    A cable which is 7.8 mm in outer diameter should be used in wiring. Any cable smaller than this size will result in poor

    sealing and cause a malfunction of the system.

    m Designated ferrite core (Model: ZCAT, manufacturer: TDK), which is a noise reduction part, should be installed

    near the external input signal cable. Install this near the control unit as closely as possible.

  • - 26 -

    PUMP OPERATION

    1. Operation ......................................... 27

    2. Methodsn of Pump Operation .......... 33

  • - 27 -

    ■ Test run (air bleeding to adjustment of discharge)

    When the pump installation, piping and wiring have been comleted and the operating procedure has been fully

    understood, preparations for pump operation should be made bleeding the air and adjusting the discharge.

    * For the operating procedure, refer to 'Methods of Pump Operation' on page 33.

    ■ Air Bleeding

    Operate the pump and feed the liquid to the pump unit for air bleeding.

    Air bleeding means removing air trapped in the suction side tube and the pump head. When the pump is to be used for

    the first time or the liquid tank has been replace, make sure to bleed air by following the steps described below.

    Caution

    Air bleeding requires some care as liquid splashes out of the discharge port of the air vent unit. Do not forget to connect

    the tube to the air discharge port and return the air to the liquid tank, a drain plate, atc.

    1. Operation

    • Starting up and stopping of pump ("MAN" control mode)

    z Turn the power on.

    • When power is applied, the ON lamp (green) lights and

    "MAN" mode is established. (Upon initial power-up)

    x If "AUTO" is displayed, change "AUTO" to "MAN" by

    pressing the key while the key is kept

    depressed.

    indicates blinking of the display. When the

    pump is in operation, press the key to select

    standby mode before changing to "MAN."

    * Every time you press the key, "MAN" or

    "AUTO" flashes alternately.

    STARTS T O P

    F U N C

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    PB

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    *

  • - 28 -

    c Press the key to select "MAN." Then, "WAIT"

    will be displayed.

    v Press the key once.

    • The display panel on the control unit displays "360"

    and the pump starts operating. (Stroke rate initial set

    value: 360 spm)

    * Every time you press the key, the pump starts or

    stops operating.

    STARTS T O P

    STARTS T O P

    F U N C

    Caution• Do not run the pump with the discharge-side valve closed.

    Operating the pump with the discharge-side valve fully closed may cause a liquid leakage or a rupture

    in the piping. Never operate the pump with the discharge-side valve completely closed.

    • Do not operate the pump dry.

    Operating a pump filled with liquid after dry operation may result in liquid leakage. Make sure to

    operate the pump with sufficient liquid inside.

    * If the pump is operated dry for a long time (30 min. or longer), the pump may be heated excessively,

    which will result in liquid leakage due to the deformation of the pump unit (pump head, valve case, etc.)

    and/or loosened pump head assembly.

    • Loosened pump head

    Loosened installation bolts on the pump head may result in liquid leakage.

    * Prior to the initial pump run, fasten the 4 hex socket bolts. (During pump storage or transportation, the

    hex socket bolts may be loosened.)

    * Fastening torque: 2.16 N·m {22 kgf·cm}

    Tighten all the hex socket bolts fully by applying an equal torque in a diagonal order among the bolts.

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    *

  • - 29 -

    ■ Procedure for air bleeding in EH-B10, 15, 20, C15, and 20 typesz Open the air vent valve to put the pump in operation.

    The air vent valve can be opened when the air vent

    adjusting screw is turned counterclockwise (about haif

    a rotation).

    x In this state, leave the pump running for about 10

    minutes to let the air out.

    • Bleed air from the tube on the suction side and the

    pump head, and fill them with water.

    c Close the air discharge port by turning the air vent

    adjusting screw.

    v Finally, make sure that no liquid leaks from any part.

    ■ Procedure for air bleeding in EH-B30, C30, and C35 typesz Return the tube connected to the fitting nut on the discharge side to the liquid tank or a drain plate.

    • If the check valve is attached to the discharge side valve, it should be detached before the above step z.

    Caution

    Some types of liquid will damage skin and machine parts. If your hand or any of the pump parts is exposed to such

    liquid, wipe it away immediately.

    x Run the pump for about 10 minutes to bleed air.

    c After letting the air out of the pump head and fitting the pump head with water, reconnect the discharge side tube to

    the regular piping.

    v Finally, make sure that there is no leakage in any part.

  • ■ Adjustment of DischargeThe set values of the stroke length and stroke rate should be determined with full consideration of the operating

    conditions of the pump and the properties of the liquid to be handled. Adjust the discharge.

    • There are two ways of adjusting the discharge: adjustment of stroke rate and adjustment of stroke length.

    • The former is generally used. Adjustment of stroke length is done for the range that cannot be covered adjustment

    of stroke rate.

    Adjustment of stroke rate (For reference only)

    The discharge should be adjusted properly, referring to the following description.

    q If the liquid tends to generate air bubbles (sodium hypochlorite (NaOC1), hydrazine solution (N2H2H2O), etc.), keep

    the stroke length at around 100% and adjust the stroke rate.

    • When the stroke length is set at a very low level, the discharge may gradually decrease.

    w When the back pressure on the discharge side is high, keep the stroke length at around 100% and adjust the stroke

    rate.

    e If the reaction to a discharge per shot fluctuates violently due to neutralization or titration, shorten the stroke length

    to reduce the discharge and use the stroke rate for further adjustment.

    z Procedure for Discharge Adjustment Viewed from the

    angle of pump performance, the following is the best

    way.

    q Set the stroke length at 100% and roughly adjust the

    discharge by the stroke rate.

    w Measure the discharge.

    e If the discharge is less than the required amount, increase the stroke rate and measure the discharge again.

    r Fine-turn the discharge by adjusting the stroke length.

    t Measure the discharge for the last time to confirm that the required amount is discharged.

    - 30 -

    360spm

    (Storke length fixed at 100%)

    270spm

    (Storke length fixed at 75%)

    180spm

    (Storke length fixed at 50%)

    Dis

    char

    ge

    Adjusted stroke length

    Fixed stroke rate

  • z Adjustment of stroke rate

    The stroke rate is set by operating the control unit keys.

    • Adjust the stroke rate per minute of the plunger in the

    range of 0 to 360 spm by the control unit.

    x Adjustment of stroke length

    The stroke length is adjusted by changing the degree of

    plunger return.

    q Turn the power on and adjust the discharge of the pump

    by rotating the stroke length adjusting dial while the

    pump is in operation.

    w The graph on the left shows the relation between

    discharge and stroke length, which is shown in % of the

    amount indicated on the nameplate as 100.

    • In practical use, set the stroke length in the range of 50

    to 100%.

    Do not rotate the stroke length adjusting dial when the

    pump is out of operation.

    - 31 -

    Dis

    char

    ge

    Adjusted stroke length

    Fixed stroke rate

    Dis

    char

    ge

    Adjusted stroke length

    Fixed stroke rate

  • - 32 -

    ■ Starting Loaded OperationWhen air bleeding and discharge adjustment have been completed, the pump is ready for loaded operation.

    See that the entire setting is complete, then start pump operation by pressuing the key.

    * For the operating and setting procedures, refer to 'Methods of Pump Operation' on page 33.

    ■ Suspension of Operation• Suspection of operation for a long time (over a month)

    Before stopping the pump, run it with clear water for about 30 minutes to wash the liquid-contacting parts and

    inside of the piping.

    • Resumption of operation

    If the pump dose not suck or feed liquid when the power is turned on, clean the valve sets and remove the solids, if

    any. If air enters the pump head, bleed it and readjust the discharge. (See pages 22~26.)

  • - 33 -

    The operating profile and the discharge amount of EH Series Pump (for water treatment process) is

    controlled by the control unit (with the contents of setting input or changed as necessary). Functions

    differ the selected operation mode. Read this section carefully before actual operation.

    Page

    I. Manual Operation ·······························································34

    II. Setting and Changing of Modes ·········································35

    III. Calling Up, Setting andChanging of Parameters···············································36~38

    IV. Calibration and Operationof Electrode ··································································38~41

    V. Setting of Parameter andDescription of Performance ················································42

    VI. Other Display Messages andSpecial Procedure ························································43~44

    2. Methods of Pump Operation

  • - 34 -

    I. Manual Operation

    ■ Start-up and Stop

    • Starting up procedure

    Before applying power, check the specified power supply on the nameplate.

    Caution

    Never use a power source other than the specified one. This will prevent trouble or damage.

    z Power supply

    When power is supplied, all display elements light up

    once. Then, they will go out instantly to make a normal

    display. The contents valid before the previous power-

    off are displayed.

    Note: The contents of the display differ depending

    on the selected mode.

    • For MAN mode, "WAIT" or stroke rate is displayed.

    (The explanation in this section is given on the

    assumption that "WAIT" is displayed.)

    • For AUTO mode, the pH value is displayed on the PH

    model (voltage mV on OR model and conductivity

    mS/cm on the CD model). (Upon the initial power-up,

    "WAIT" is displayed in MAN mode.)

    x Starting and stopping

    The pump starts running when the key is pressed

    once.

    the pump stops running when the key is pressed

    again.

    * Each press of the key selects starting and

    stopping alternately.

    STARTS T O P

    STARTS T O P

    STARTS T O P

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    *

  • - 35 -

    II. Setting and Changing of Modes

    ■ Setting and Changing of Modes

    • MAN' and 'AUTO' modes are available for selection.

    MAN mode:··················· Manual operation (for air elimination or test run)

    AUTO mode: ················· pH control operation with PH model

    ORP control operation with OR model

    Conductivity control operation with CD model

    z Stop the pump through key operation and put it in

    standby condition (with 'WAIT' displayed).

    x While holding the key down, press the

    key. Then, the LCD display starts flashing, indicating

    that the mode can be changed.

    c If the key is pressed while 'AUTO' is flashing,

    'MAN' is displayed in flashing mode.

    * Each time the key is pressed, 'AUTO' or 'MAN'

    is alternately displayed in flashing mode.

    v Select a display from the flashng display and press the

    key to enter the election. (now the mode

    changing process is completed.)

    F U N C

    F U N C

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    PB

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    spm

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    *

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

  • - 36 -

    III. Calling Up, Setting and Changing of Parameters

    ■ Parameters

    Parameters, data which determine the operation and control of pump, store respective set values specified previously.

    The applicable parameteres available in each model and each mode are as shown below.

    ■ Calling up parameter

    Parameters can be called up and displayed in the operation

    panel as necessary.

    z Press the key once to display the first parameter.

    Each time the key is pressed, the following

    parameters are displayed one at a time in sequence.

    When all the parameters have been displayed in such a

    manner, the first parameter is displayed again. (Each

    time the key is pressed, the parameters are

    displayed in sequence downward from the top as

    indicated in the above table for each mode.)

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    Mode Parameter Initialvalue Setting range Step

    Initialvalue Setting range Step

    Initialvalue Setting range Step

    PH model OR model CD model

    360spm

    7.00pH

    4.00pH

    7.00pH

    5.00pH

    360spm

    2.00pH

    12.00pH

    360spm

    0mV

    0mV

    1000mV

    360spm

    -1500mV

    1500mV

    MAN

    AUTO

    SET

    ZERO

    SPAN

    CELL

    %

    SET

    Pb

    spm

    Lo

    Hi

    0~360spm

    0.00~14.00pH

    0.00~14.00pH

    0.00~14.00pH

    -14.00~14.00pH

    0~360spm

    0.00~14.00pH

    0.00~14.00pH

    0~360spm

    0mV

    -1999~1999mV

    -1999~1999mV

    0~360spm

    -1999~1999mV

    -1999~1999mV

    1spm 360spm 0~360spm 1spm

    1.00 0.80~2.20 0.01

    2.00% -9.99~9.99% 0.01%

    5.00mS/cm 0.00~9.99mS/cm 0.01mS/cm

    -2.5mS/cm -9.99~9.99mS/cm 0.01mS/cm

    360spm 0~360spm 1spm

    0.00mS/cm 0.00~9.99mS/cm 0.01mS/cm

    9.99mS/cm 0.00~9.99mS/cm 0.01mS/cm

    0.01pH

    0.01pH

    0.01pH

    0.01pH

    1spm

    0.01pH

    0.01pH

    1spm

    1mV

    1mV

    1spm

    1mV

    1mV

  • - 37 -

    ■ Changing parameter

    z Call up a parameter to be changed. Press the key

    to display a parameter to be changed.

    x Press the key to start the set value display section

    flashing.

    indicates blinking of the display.

    c Press the key again, to move the position of the

    digit to flash. Each press of the said key moves the

    digit rightward one at at a time.

    v Each press of the key, then, increase the value of

    the digit selected.

    (0/1/2/3·········/9/0)

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    PB

    PB

    PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

  • - 38 -

    b With the desired value displayed, press the key

    to enter the value. (The set value display section stops

    flashing.)

    n When setting a parameter with a minus sign, call up the

    parameter and press the key first. Then, press

    the key once to flash a minus sign.

    Warning

    Press the key for the storage of a set value in the

    memory. Each time a set value is changed, check that no

    digit is flashing. Also, make sure to press the key.

    IV.Calibration and Handling of Electrode

    Warning

    Calibration of electrode is an indispensable process in the use of the control unit. When using an electrode for the first

    time, washing it, replacing and recalibrating it, even if the ZERO or SPAN described in the following procedure has been

    set up, press the key to make the display flash and press the key to memorize the value newly.

    z Change the mode, if MAN mode has been selected, into

    the AUTO mode.

    x Prepare two types of pH standard liquid. (pH7 liquid

    and pH4 liquid)

    c Press the key to call up parameter ZERO. Place

    the pH electrode in the pH7 standard liquid for 1 to 2

    minutes. ( lights up and all the digits are

    displayed.)

    v Set the value of the pH7 standard liquid.

    Press the key to make the highest digit flash.

    Pressing the key while the digit is flashing

    increases the value by 1. Pressing the key then

    moves the flash to the next digit.

    For the set value, take the value on the package of the

    standard liquid in accordance with the temperature of

    the liquid. (The value indicated for an electrode

    without a temperature compensating element is

    specified for a temperature of 25 °C.)

    indicates blinking of the display.

    Even when setting the parameter to the same value as the

    current one, carry out steps c and v above.

    ZERO

    F U N C

    F U N C

    F U N C

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    LoPB

    MAN

    ZERO SPAN

    AUTO °C ALMPB

    PB

    PB

    SET

    spm

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    pH

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    Caution

    In the step of calibration, make sure that the pH

    electrode is soaked in each pH standard liquid

    completely. Be careful not to expose the electrode to

    the air or use the same liquid for both ZERO and

    SPAN. Otherwise, incorrect values will be stored in the

    memory which will result in abnormal operation.

    F U N C

  • - 39 -

    b Press the key to enter the value. The display

    stops flashing.

    With the above step completed, the signal from the pH

    electrode and set value are memorized.

    n Press the key to call up parameter SPAN. Place

    the pH electrode in the pH4 standard liquid for 1 to 2

    minutes. ( lights up.)

    m Set the value of the pH4 standard liquid.

    Press the key to make the highest digit flash.

    Pressing the key while the digit flashes increases

    the value by 1. Pressing the key then moves the

    flash to the next digit.

    For the set value, take the value on the package of the

    standard liquid in accordance with the temperature of

    the liquid. (The value indicated for an electrode

    without a temperature compensating element is

    specified for a temperature of 25 °C.)

    indicates blinking of the display.

    Even when setting the parameter to the same value as the

    current one, carry out steps n and m above.

    , Press the key to enter the value. The display

    stops flashing.

    With the above step completed, the signal from the pH

    electrode and the set value are memorized.

    . With all the above steps completed, the calibration of

    an electrode is completed.

    F U N C

    SPAN

    F U N C

    PB

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    PB

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    PB

    PB

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    pH

    Hi

    Lo

  • - 40 -

    z Press the key to call up parameter ZERO.

    x Disconnect the wire once before connecting an ORP

    electrode or after connecting it and short-circuit the

    terminals GLASS and REF on the terminal block.

    c Press the key to make the highest digit flash.

    Wait for about 1 to 2 minutes before pressing the

    key.

    indicates blinking of the display.

    v Press the key to return to normal display

    (measurement status). Connect the ORP electrode and

    measure the voltage of the quinhydrone standard liquid.

    b Wait for 1 or 2 minutes. Then read the indicated

    voltage.

    Confirm that the indicated value is at 260±20 mV.

    z In the case of the CD type unit, setting of the cell

    constant of the conductivity sensor is made instead of

    calibration.

    First, press the key to call up parameter CELL.

    x Check the cell constant specified for the conductivity

    sensor to be used. Press the key to make the

    highest digit flash. Pressing the key while the

    digit flashes increases the value by 1. Pressing the

    key then moves the flash to the next digit.

    indicates a blinking of the display.

    c Press the key to enter the value. The display

    stops flashing.

    Warning

    The cell constant, being a unique value of a sensor, varies

    with the type of sensor. Never forget to set the constant

    prior to the measurement of conductivity. Measuring

    conductivity without setting the constant may cause a

    malfunction or a display error.

    F U N C

    PB

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    mV

    Hi

    Lo

    PB

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    spm

    mV

    Hi

    Lo

    PB

    MAN

    ZERO SPAN

    AUTO °C ALMSET

    mV

    Hi

    Lo

    MAN

    ZERO SPAN

    AUTO °C ALMSET GAIN

    spm

    mV

    Hi

    Lo

    PB

    MAN

    CELL SPAN

    AUTO °C ALMSET

    spm

    mS/cm

    Hi

    Lo%

    PB

    MAN

    CELL SPAN

    AUTO °C ALMSET

    spm

    mS/cm

    Hi

    Lo%

    PB

    MAN

    CELL SPAN

    AUTO °C ALMSET

    spm

    mS/cm

    Hi

    Lo%

    F U N C

  • - 41 -

    v The temperature compensation coefficient is set at

    2%/°C for standard operation.

    ■ Handling instructions for pH/ORP electrodes and conductivity sensor

    • Refer to the "Wiring" section and make sure to carry out calibration with the standard liquid.

    • Electrodes are vulnerable to impact. They may be damaged if dropped or knocked against another part. Handle the

    electrodes carefully.

    • Special considerations and/or measures may be necessary when installing the electrode under such conditions as

    described below.

    z Position where bubbles are generated.

    x Measurement of liquid which contains slurry.

    c Electrolyzed liquid

    v Connection or extension of wire when the electrode is distanced from the pump unit

    • Clean the pH electrode periodically (once a month or more frequently). Use water to wash the electrode as a

    general rule. When dirt adheres to the electrode stubbornly, use 1N (specified) of hydrochloric acid or nitric acid.

    Warning

    Abrasive force at any degree of strength applied on the electrode may scratch the surface, which will affect measuring

    accuracy or lower the speed of response. Use a soft cloth to wipe the surface of the electrode if stains or dirt are difficult

    to remove by washing.

    • Clean the ORP electrode periodically (once a month or more frequently). Use water to wash the electrode as a

    general rule. When dirt adheres to the electrode stubbornly, use 1N (specified) of hydrochloric acid or nitric acid.

    Furthermore, if the stain or dirt is difficult to remove by washing, use a soft cloth to wipe the metallic surface on the

    end of the electrode.

    • Refer to the "Wiring" section and set the cell constant to the value specified on the sensor before use.

    • The cable of the sensor must not be extended or cut short. Changing the cable length may change the cell constant

    to finally result in a measurement error.

    • Rotate the cover on the top of the sensor to remove it. Then, wash the sensor with water. If stain or dirt is difficult

    to remove by washing with water, wipe the electrode with a soft cloth soaked in neutral detergent and wash it again

    with water.

    PB

    MAN

    CELL

    AUTO °C ALMSET

    spm

    mS/cm

    Hi

    Lo%

  • - 42 -

    V. Setting of parameters and operation mechanism

    ■ Setting of parameters and operation mechanism (PH model)

    • Operation mechanism

    • The stroke rate of the pump is represented by the real line in the figure above. The range of proportional control

    is called proportional band (PB). The spm value indicates the max. stroke rate of the pump. In the example

    above, the stroke rate is fixed at 300 spm for the pH values at or lower than 2.50. The PB value is assumed to be

    added to the SET value, which increases the stroke rate in the figure above as the pH measurement decreases.

    For example, the stroke rate for the measured value 4.00 pH stands at 120 spm.

    • The parameters are set before shipment from the plant. Set them to the required values in accordance with the

    application purpose before use.

    • If the measurements exceed the value set for Lo or Hi, the alarm function runs with display and/or alarm output

    (optional function). The values for Lo and Hi can be set as required upon the condition, however, a setting of

    Lo>Hi should be avoided.

    • When the measured value is at or lower than the Lo set value:

    The display ALM is lit and Lo flashes.

    spm360

    270300

    180

    90

    0 2 45

    6 7 8 10 12 14pH

    Parameterq SETw PBe spm Lo Hi

    5.00pH-2.50pH300spm2.00pH

    12.00pH

    e

    w

    q

    2.5

    -2.50

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

  • - 43 -

    VI. Other Display Messages and Special Procedure

    ■ Error messages

    If any value out of the permissible set value range is

    entered by mistake, error indication ( 'E_ST' ) will be

    displayed for 5 seconds and the value valid before the

    setting change step will be resumed. Carry out value

    changing step for a correct value.

    • If 0.00 or -0.00 is specified for the setting of parameter

    Pb, "E_Pb" will flash for 5 seconds. Then, the value

    registered in the previous setting will be displayed.

    Then, carry out the set value changing step for the

    correct value (only in the AUTO mode).

    ■ How to select the content of display

    This control unit can display pH value, voltage (mV), conductivity (mS/cm), stroke rate (spm), and temperature (°C)

    (PH type and CD type only).

    z While holding key down, press the key.

    The currect measured value starts flashing.

    x Press the key next to make the temperature

    display 'ºC' flashing. (No temperature display is

    available with OR type.)

    indicates blinking of the display.

    F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

    AUTOMAN

    ZERO SPAN

    °C ALMSET PB

    spm

    mS/cm

    Hi

    Lo

    STARTS T O P F U N CSTARTS T O P F U N C

  • - 44 -

    c Press the key again to display stroke rate in flash

    mode.

    The above display contents are shown alternately every

    time the key is pressed.

    v With the desired display presented, press the key

    to register it.

    • When stroke rate is display in 'spm'

    The pump is equipped with stop function whith

    temporarily stops the pump operation upon receiving

    contact signal from outside. When this function is

    activated and only if the stroke rate is displayed, the

    display as shown below is presented.

    * Note that you cannot operate the key when the

    stop function is activated by the external contact signal.

    ■ Special procedure

    When the display is not in flash mode, pressing the key for more than one second will turn the parameter display

    back to the basic display (selected display content) in the current mode 'MAN' or 'AUTO'. This procedure is employed

    when the parameter should be reset or when only one parameter should be changed with call-up for all the other

    parameters canceled.

    F U N C

    STARTS T O P

    F U N C

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    Hi

    Lo

    OPENCLOSE

    ON

    PUMP

    Contact

    spm

    pH

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

    *

    STARTS T O P F U N CSTARTS T O P F U N C

    MAN

    ZERO SPAN

    AUTO °C ALMSET PB

    spm

    pH

    Hi

    Lo

  • - 45 -

    MAINTENANCE

    1. Maintenance and Inspection ........... 46

    2. Causes of Trouble and Troubleshooting................................ 47

    3. Disassembly and Reassembly ..........48

    4. Accessories.......................................52

    5. Names of Parts and ExternalDimensions .......................................53

  • - 46 -

    ■ Daily check

    While the pump is in operatior, pay attention to the following points. If anything is out of order, see 'Causes of

    Trouble and Trouble shooting' on page 46 and deal with it appropriately. When any of the parts is due for replacement,

    replace them with a new ones.

    1. Maintenance and Inspection

    Part QuantityReplacement

    TimeRemarks

    Valve set

    Diaphragm

    O ring

    2 sets

    1

    *

    Approx.8,000 hrs.

    * For number, see "Names of Parts and Structure ".

    No. Check Point Detail/Remedy Checking method

    1 Does pump lift liquid

    normally?

    • Is liquid normally fed?

    • Is suction pressure/discharge pressure at

    normal level?

    • Has liquid undergone quality change,

    crystallization, or solidification?

    • By flow meter or visual

    • Check nameplate.

    • Visual

    2 Abnormal noise or

    vibration?

    • Abnormal functioning of pump may cause

    abnormal noise or vibration.

    • Visual and aural

    3 Is there liquid or air

    suction in any joint of

    pump or piping?

    • Tighten leaking joint.

    • Excessive air bubbles in discharged liquid

    mean air is sucked into system.

    • Visual

    s The durability of the expendable parts depends on the pressure, temperature, and properties of the

    liquid handled. The value in the above table were obtained from a continuous test run of the pump

    using clear water at room temperature. Take it as a guideline for replacement.

    s The valve set illustrated above is applicable to VC, VH, PC and PH types only.

    ■ Expendable parts

  • - 47 -

    Cause

    ● Faulty wiring or breakage.● Lowered voltage.

    ● Damage to electronic circuit of controlunit.

    ● Air sucked in from suction piping.● Valve gasket not installed.● Valve set fitting in wrong direction.● Pump is air-locked.● Pump stroke length is too short.

    ● Suction or discharge side valve cloggedwith foreign matter.

    ● Adhesion of valve onto valve seat.

    ● Suction or discharge side valve cloggedwith foreign matter.

    ● Air trapped in pump.● Overfeeding.● Diaphragm out of order or damaged.

    ● Valve CS or fitting insufficently tightened.● Pump head in sufficiently tightened.

    ● Diaphragm is damaged.● O ring and valve gasket not installed.

    ● Power supply cord is routed in a wrongway.

    ● Stained electrode.● Scratch, crack, deterioration on electrode.

    2. Causes of Trouble and Troubleshooting

    Trouble

    Pump does not start.

    No liquid flow.

    Discharge fluctuates.

    Liquid leaks.

    Measured value fluctuatesconsiderably.

    Measuring response is extremelypoor.

    Troubleshooting

    K Correct wiring.K Find cause and raise voltage

    to rated level.K Replace electronic circuit.

    K Rectify piping.K Install it.K Reassemble it.K Bleed air.K Drive pump with 100%

    stroke length. Then resetstroke length.

    K Disassemble, inspect, andclean.

    K Disassemble, inspect, andclean.

    K Disassemble, inspect, andclean.

    K Bleed air.K Install check valve.K Replace it.

    K Tighten it.K Tighten it. Clamping torque:

    2.16 N·m {22 kgf·cm}.K Replace it.K Install them.

    K Change cord routing.

    K Clean electrode.K Replace electrode.

  • ■ Disassembly and Reassembly of Pump UnitBefore disassembly the pump unit, as a safety measure, turn the power off and lower the pressure in the pump to

    atmospheric pressure. Then, remove the piping and disassemble the pump unit with reference to 'Names of Parts and

    Structure of Pump' on page 53.

    ■ Replacement of Valve Sets

    z Removal of the discharge side valve

    q Loosen the fitting nut and detach the tube

    connected with the pump, paying attention to

    the residual liquid in the pump.

    Warning

    Some types of liquid may be harmful to the human

    skin and pump parts. If your hand or parts of the

    pump are exposed to the liquid, wipe it off

    thoroughly.

    w Rotate the locknut counterclockwise while

    lifting it up by hand and detach the air vent

    unit A.

    e Loosen air vent unit B by means of a wrench

    and detach it. Then, take the valve set out of

    the pump head.

    • In the EH-B30, C30, C35 or FC type pump,

    loosen the connecting port by means of a

    wrench and detach it. Then, take the valve

    set out of the pump head.

    x Removal of thesunction side valve set

    q Loosen the fitting nut and detch the tube.

    • Be careful of the residual liquid.

    w Loosen the fitting by means of a wrench and take the valve set out of the fitting.

    • If scratches or wear and tear are found in the removed valve set, replace it with a new one.

    - 48 -

    3. Disassembly and Reassembly

    4

    5

  • - 49 -

    For reassembly, follow the steps of disassembly in

    reverse, paying attention to the following points.

    • Order of arrangement and direction of insertion

    of valve set:

    An errer in the order of arranging a valve set or

    in the direction on its insertion will cause

    defective liquid feeding (leaking or a decrease in

    discharge).

    • Make sure to insert the O rings.

    z Assembly of suction side valve set

    Fit the valve set into the fitting, and screw the

    connecting port into the pump head by hand.

    Next, tighten the connecting port about a 1/4

    rotation by means of a wrench.

    x Assembly of discharge side valve set

    Fit the valve set into the pump head. Insert the

    lock nut into the air vent valve B and tighten the

    locknut.

    ■ Air vent assemblyUse a wrench to loosen the lock nut. Since the air

    vent valve can be rotated 360 degrees, the tube

    connection position can be selected freely to satisfy

    your requirements. The lock nut shall be loosened

    to make sure that the tube is not positioned over the

    pump head. Then, fix the position by fastening the

    lock nut. When fastening the lock nut, hold the air

    vent valve A with your hand and rotate the lock nut

    clockwise when viewed from above the pump unit

    by hand. Then, use a wrench to rotate the lock nut

    about 1/4 turn for further tightening.

    Air vent unit A

    Lock nut

    Pump head

    • Order of Valve set arrangement

    6

    1411131211131217

    1714111312111312

    3

    4

  • - 50 -

    ■ Replacement of diaphragm

    z Set the stroke length adjusting dial at 100%.

    x Remove the pump head from the pump unit by loosening the four hexagon socket bolts with a hexagon bar wrench.

    c Hold the periphery of the diaphragm and rotate it

    counterclockwise to detach it from the plunger

    pin. (The diaphragm is screwed onto the plunger

    pin.)

    • In some cases, a certain number of diaphragm

    spacers are inserted behind the diaphragm and

    the retainer for the purpose of positioning. Be

    careful not to lose them when replacing the

    diaphragm.

    For reassembly, follow the steps of disassembly in

    reverse, paying attention to the following.

    z Firtst, make sure that the stroke length of the

    pump is set at 100%.

    • Start up the pump for setting the stroke length.

    Then, stop the pump and turm the power off.

    x Insert the retainer and diaphragm spacer(s) onto

    the threaded part of a new diaphragm and screw

    the plunger pin into it.

    • Fit the concave side of the retainer onto the

    threaded part of the diaphragm.

    c Attach the pump head(1) to the pump body by

    tightening the hexagon socket bolt.

    (Clampimg torque: 2.16N·m {22kgf·cm})

    Pump head

    Plunger Pin

    Diaphragm spacer

    Retainer

    Diaphragm

  • - 51 -

    ■ Detachment of Control Unit* Do not detach the control unit from the pump drive unit unless unavoidable.

    q While operating the pump, rotate the stroke length adjusting dial to 0% reading, then stop the pump and turn off the

    power swtich.

    w Next, wait for about one minute and pull out the dial. Then unscrew the control unit setscrew with a Philip

    screwdriver. Finally, pull up the control unit out of the pump unit.

  • - 52 -

    4. Accessories

    ■ Specification of check valve, back pressure valve (Optional Items)

    Type

    SetPressure

    MPa{kgf/cm2}

    Diameterof Fitting

    mm

    Materialof Liquid

    ContactingParts

    Applicable Pump

    Material Codeof Liquid

    ContactingParts of Pump

    CA-1V-4 ◊ 6CA-1E-4 ◊ 6CA-2V-9

    CA-2E-9

    CA-2VL-9

    CA-2EL-9

    CA-1VC-4 ◊ 6CA-1VE-4 ◊ 6CA-1VC-6

    CA-1VE-6

    CA-2VC-9

    CA-2VE-9

    CA-2VCL-9

    CA-2VEL-9

    BVC-1VV-4H

    BVC-1VV-9H

    0.17 ± 0.04{1.7 ± 0.4}

    0.17 ± 0.04{1.7 ± 0.4}

    0.2 ± 0.02{2 ± 0.2}

    +0.04-0.03+0.4-0.3

    +0.04-0.03+0.4-0.3

    +0.02-0.01+0.2-0.1

    ø 4 ◊ ø 6

    ø 4 ◊ ø 6

    ø 4 ◊ ø 6

    ø 6 ◊ ø 8

    ø 9 ◊ ø 12

    ø 9 ◊ ø 12

    ø 9 ◊ ø 12

    GFRPP

    PVC

    PVC

    EH-B10·15·20EH-C15·20

    EH-B10·15·20EH-C15·20

    EH-B10·15·20EH-C15·20

    EH-B30EH-C35

    EH-B30EH-C35

    EH-C30

    EH-C30

    EH-C30

    EH-B30EH-C35

    PC

    PH

    PC

    PH

    PC

    PH

    VC

    VH

    VC

    VH

    VC

    VH

    VC

    VH

    VC

    BVC-1TV-4H

    BVC-1TV-10H

    0.2 ± 0.02{2 ± 0.2}

    0.10{1.0 }

    0.10{1.0 }

    0.05{0.5 }

    0.05{0.5 }

    +0.02-0.01+0.2-0.1

    ø 4 ◊ ø 6

    ø 10 ◊ ø 12PVDF

    EH-B10·20EH-C20

    EH-C30

    EH-C35

    FC

  • - 53 -

    ■ Exploded View

    5. Names of Parts and External Dimensions

    Air vent assembly

    Control unit

    Pump proper

    Valve set

    (PW-SW)

    Valve set

    Body of pump

    Terminal box Cable gasketBlind cap

    Cable nut

    Gasket

    Body of controller

    Retaining screw

    Cap

    Stroke lengthadjusting dial

    Key protectorCap

    connector

  • ■ Pump Unit• EH-B10, B15, B20, C15, C20 Types

    • EH-B30, C30, C35 Types

    - 54 -

    No. QuantityName1 1

    1

    1 (2)

    3 (2)

    1(0)

    1(0)

    1

    1

    1

    1(0)

    4

    4

    4

    2

    2

    1

    4

    1

    1

    1

    1

    2

    3

    4

    5

    6

    7

    8

    9

    10

    11

    12

    13

    14

    17

    18

    19

    22

    23

    24

    25

    Pump head

    Bracket

    Connecting port

    Fitting nut

    Air vent unit A

    Lock nut

    Diaphragm

    Insert bolt

    Retainer

    Air vent unit B

    Valve guide

    Valve seat

    Valve ball

    Valve gasket

    O ring

    Diaphragm spacer

    Hex socket bolt(with spring washer and plate washer)

    Diaphragm seal

    Air vent adjusting screw

    Adjusting screw nameplate

    O ring

    The figures in parentheses indicate quantitiesin the cases of the EH-B30, C30, and C35 types.

  • - 55 -

    • E