jeol jem 1230 responsible personell: endy (45279377 ... · rise, notify endy, the filament could be...
TRANSCRIPT
Responsible personell: Endy (45279377)
Online booking is compulsory!
After training you will have access to
working alone on the instrument. All
insertion of samples is done by
responsible MIC personell (Endy and
Reidar).
Jeol JEM 1230
1
Table of content:
The control panel page 3
The sample holder page 4
High tension and beam page 5
Simple beam alignment page 8
Focussing your sample page 9
Inserting camera page 11
Taking pictures page 12
How to save your dm3 images page 14
Converting dm3 images to tiff page 15
Turning off the system page 16
2
JEOL JEM 1230
This switch will
change the
magnification.
Bring it to the left
to go down in
magnification and
right to go up.
Control the brightness of you beam.
Leave these buttons on (green light) to have a stepwise increase/decrease of the brightness and focus.
Focusing knobs:
coarse and fine.
The wobbler is
usefull for adjusting
the focus.
This button is used when acquiring a photo
onto the film negative. This is rarely in use.
Beam shift is used
to align the beam
(this is explained
later).
Control board
3
1 2 3 4 5
Date: Name: Specimen:
The sample holder
The pentaholder will fit 5 samples at a time. MIC personnel will help you insert your samples into
the microscope. Remember to fill out the form with the sample positions.
Once inserted, you can change between samples by turning
the middle wheel (arrow). The number which shows up next to
the white circle indicated the sample position on the holder.
4
1 2 3 4 5
3. 3.
Turning on the high tension and the beam
1. Click onto the circle in front of HT, then click onto ON.
2. A ”Ready” will appear behind the HT within 10 seconds.
3. Now turn on the beam by clicking first into the circle in front of
the beam, the click onto ON.
4. The beam will come on within 20 sec. You will notice the beam
current rising from 33 to 44 µA. If the beam current does not
rise, notify Endy, the filament could be broken.
5. Turn on the power (A) for the gatan camera. Turn on the
cooling (B) for better images. Be careful not to «heat» the
camera which is the same switch as for cooling.
1. 1.
A B
2.
5
Make sure that the ”specimen position” window
is open to orientate yourself on the grid.
Use the ”roller mouse” to move the grid around
in order to find your samples.
You can also move around by clicking inside the
speciment position monitor, but be sure to be in
«point» mode.
Roller mouse. Click onto the arrows to increase the speed.
6
Specimen position
Is the beam well aligned?
Beam slightly missaligned Ok alignment
It is very simple to align the beam yourself:
1. Focus your beam (make it small) using the ”brightness” knob.
2. Press ”beam shift” on the control board (it should be green).
3. Use the right mouse button to move the beam into position.
4. Press ”beam shift” again (green light comes off).
Fluorescent
screen Fluorescent
sample
mirror
7
Focussing your sample
Before focussing your sample, make sure the ocular is in focus for
your eyes.
When there is light on your sample, look into the ocular and use the
black knob on the side to focus the black spot you see inside your
field.
8
Focussing your sample
To focus you sample use the two knobs (coarse and fine) and
focus first on the large fluorescent screen.
For fine focus, bring down the fluorescent sample mirror and
use the oculars. Some people prefer to focus using the
”wobbler”. Press ”wobbler” on the control table and use the two
focusing knobs to stop the image from moving. When image is
still, you have focus.
BUT: remember to turn off the wobbler !
To bring down
the
fluorescent
sample mirror,
pull down this
handle. 9
Focus onto
something with
a lot of
contrast.
Inserting the camera
The Gatan camera has a resolusion of 1024x1024 pixels. The camera is
very sensitive, so make sure the current density is under 10 pA/cm2
before inserting the camera!
Open the software:
DigitalMicrograph.
Insert the camera, see other page. 10
To insert camera, make sure
the current density is benieth
10 pA/cm2. Select ”camera
inserted” (you will hear a
sound and your image on the
fluorescent screen is gone).
Select ”start view” and a
window with the live image
shows up.
This window is for the live
image. The resolusion is not so
good, and the refresh time is
dependant upon the beam
brightness. Yellow means too
little light, red means too much
and green means perfect
illumination condition.
This window is for the picture
acquisition. Press ”start
acquire” when you want to
take a picture.
To change the settings for
both live image and
acquisition image, click onto
the hammer sign (not
recomended).
The rectangular should be in this
range before acquiring an image.
Auto exposure just makes life
a whole lot easier. If you don’t
have a homogenous sample,
you might want to remove the
«auto» function and test out
different exposure times.
Taking pictures
11
Preview picture
Full resolution picture
After you press ”start acquire”,
the image will appear within a
few seconds.
The scalebar on the preview image is wrong, but the scalebar
on the full resolution image is correct.
12
To save this image, go to file – save image as, and save your image in the user-folder
(shortcut on desktop). The image must be saved as a gatan file (dm3). This format is
specific for the gatan software. You will export the images once you finish your session.
13
”File – batch convert”, find your folder and start the convertion from dm3 to tif. You will not loose
your raw data, instead you will get extra copies as tif format.
You can save the files temporarily on C:\users save here\
Converting dm3 images to tif
After you’ve acquired all your images, you can convert them to tif in the following way:
14
Turning off the system when you have finished working
2.
1. 1.
2.
1. Make sure the camera is not inserted. Turn off the
beam by selecting OFF in control window.
2. Turn off the HT by selecting OFF in control window.
3. You will be asked if it’s ok to execute ”HT Mode OFF”,
select OK.
4. Turn off thee cooling (B) and the main power (A) for the
gatan camera control box.
3. A B
15
You should now see light on the fluorescent table.
If it’s dark:
• check if the current density is 0.
• If it’s not, then turn up the brightness (on the control board)
• check that you are in the middle of your grid (specimen position window). Use the mouse
to move around as you could be on top of a mesh.
• check if the camera is still inserted (through the software).
If it’s still dark, it could be a problem with the filament.
→ contact Endy
16
Troubleshooting
17