CompactGC4.0

Key Performance

Page 2: Key Performance - Gassite

Proven Performance

n The speed of a MicroGC...... CompactGC4.0 delivers results in seconds

nThe robustness of a process GC...... CompactGC4.0 resists aggressive media thanks to

solid detectors and diaphragm valves

nThe versatility of a laboratory GC...... CompactGC4.0 offers 7 different detectors and

exchangeable columns

Performance prioritisedRobust diaphragm valves, high performance separationcolumns and robust, highly sensitive detectors are key to the CompactGC4.0.

CompactGC4.0 is a reliable performer for a wide rangeof applications. It provides high throughput at low costof ownership. CompactGC is a proven analyser concept,appreciated worldwide.

2

Page 3: Key Performance - Gassite

3

Fast Performance

Reliable Performance

CompactGC4.0 is a fast performer. Results are delivered in seconds, achieved with fast injection, narrow bore columns and fast highly sensitive detectors. The typical runtime is 2-120 seconds. Analysis time is reduced even more by using backflush option and pressure programming to accelerate late eluting components. Cycle time is equal to run time: no additional flush or conditioning steps are needed.

CompactGC4.0 is a reliable performer. Proven GC technology from laboratory GCs andprocess instruments are used for high uptime. Injection valves normally account for the majority of the maintenance costs in gas analysis; therefore robust diaphragm valves are used to guarantee millions of injections.

Rigid, maintenance free detectors with a long lifetime ensure reliable performance for many years. Column switching techniques like backflush and heart-cut effectively protect your analysis column. The durable setup also allows analysis of aggressive components, for example high percentage H2S or ammonia.

Seconds1 2

Methane

Ethene

Ethane

C1 and C2 analysis in only 2 seconds.Seconds

30 50 70 90 110 130

pressure program

Water

Air

Methane

Carbondioxide Nitrous oxide

Ethylene

EthanePropylene

Propane

Accelerated analysis of water and C3 by pressure programming of the carrier gas

2 seconds

robust diaphragm

valves

20 40 60 80 100 120 140 160 180Seconds

BenzeneToluene

Shortening runtime to 80 seconds by using a backflush option

20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180Seconds

BenzeneToluene

Ethylbenzene p-Xylene

m-Xylene

o-Xylene

Styrene

44 48 52 56 60

Name

Air

Methane Carbondioxide

EtheneAmmonia

Hydrogen sulfide

SecondsGAS diaphragm valve offers an extended lifetime and internal purge for extreme low leakrate Solid detectors and diaphragm valves allow analysis of high % NH3 and H2S

Page 4: Key Performance - Gassite

4

Performance from ppt to %

TCD (Thermal Conductivity Detector) and FID (Flame Ionisation Detector) cover many applications from ppm to % permanent gases and hydrocarbons.

PED (Plasma Emission Detector) and PDD (Pulsed Discharge Detector) offer high purity analysis at low ppb level.

PID (Photo Ionisation Detector) detects conjugated hydrocarbon and aromatic components.

PFPD (Pulsed Flame Photometric Detector) selectively measures ppb sulphur.

MS (Mass Spectrometer) provides positive identification of your components.

CompactGC4.0 performs over a wide range of concentrations. A choice of high sensitive detectors covers every component and concentration range:

Thermal Desorption Option A Pre-concentrator module is offered for ppb-ppt sensitivity. A precise volume, typically 25ml to 1 l, is sent to an electrically cooled trap to focus the components of interest. The trap is rapidly heated (100 ºC/s) for fast injection on the analytical column.

CGM Detector

capillary columncarrier gas

sample in

purge gas

micro trap0-350 ‘Cmultibed

sampleout

samplepump

massflow

controller

Thermal Desorption

Option

Page 5: Key Performance - Gassite

5

Performance EverywhereCompactGC4.0 is a versatile performer. The 19” industrial standard enclosure allows versatile use on many locations. Beside laboratory measurement and on-line analysis, single or multiple instruments can be stacked in transportable enclosures for on site measurement.

highly trans-

portable

5

Page 6: Key Performance - Gassite

6

Priority on Performance● Up to 4 channelsCompactGC4.0 offers up to 4 analysis channels with independent injection loops, columns and detectors for analysing a wide range of components.

●Inert sample pathFor the analysis of polar component at low concen-tration levels like ppb Sulphur, an inert sample path is mandatory. This is achieved by the use of high quality materials like Sulfinert/Silconert tubing.

Cost Effective PerformanceCompactGC4.0 is a cost effective performer. Columns are replaced by the user in a minute, withresulting low operational cost and high flexibility.Capillary as well as micro-packed columns canbe installed.

The CompactGC concept is also very maintenancefriendly. Finger tight couplings are used for a 100% leak free connection without the need for tools. All parts are highly accessible thanks to the transparent design.

Easy exchangeable columns / low cost of ownership

Page 7: Key Performance - Gassite

7

●Chromatography Data SystemsInstrument control drivers are available for Chromeleon, OpenLab and EZChrom. All GC parameters like flow, pressure, temperature and detector settings are controlled by the chromatography data system.

●Highly repeatableFirst class GC hardware offers low RSD% and high linearity.

12 14 16 18 20 22 24 26 28 30 32 34 36

Hydrogen

Oxygen

Nitrogen

Methane

Seconds

Chromatogram used for RSD% calculation. 1 % conc. level (H2: 5%)

Page 8: Key Performance - Gassite

ApplicationsCatalyst Screening / On-line analysis / PetroSpeed and Flexibility in High Throughput analysis: short run times for fast screening and easy column replacement when new experiments require it.

High Purity Analysis:Analysis of impurities in bulk gases by Pulsed Discharge Detector. Highly sensitive: low ppb detection limit.

10 30 50 70 90 110 130 150 170

Seconds

Methane

Ethane

PropaneButane

Pentane

Hexane

0 20 40 60 80 100 120 140 160 180 200

Seconds

Oxygen

Hydrogen

Nitrogen

Methane

Carbonmonoxide

Seconds

0 20 40 60 80 100 120 140 160 180 200

CarbondioxideMethaneCarbonmonoxide

Hydrogen

Fast analysis of C1-C6 hydrocarbons (FID)

Hydrocarbons (FID)

Analysis of CO and CO2 on single column (TCD)

5 ppm permanent gases in Helium (PDD)10 ppm CH4 , CO2 and N2O (PDD)

Nitrogen

Seconds

10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170 180 190 200 210 220 230

MethanolHexane

Cyclohexane

2,2,

4-tr

imet

hyl

pen

tan

eH

epta

ne

Toluene

Eth

ylb

enze

ne

m/p

-Xyl

ene

o-X

ylen

e

8

Methane

Carbondioxide

Nitrous Oxide (N2O)

Hyd

rog

en

Air

10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160 170Seconds

Page 9: Key Performance - Gassite

ApplicationsNatural Gas AnalyserNatural Gas Analysers are available with various configurations on CompactGC. Optional channels are offered for low ppm H2 or sulphur components. Calorific Value / BTU calculation is also included.

9

Seconds

80

Nitrogen Methane

Carbon dioxide

Ethane

Propane

Seconds

80

Propane

i-Butane

i-Pentane

C6 Hydrocarbons C7 Hydrocarbons

Fast natural gas analysis using capillary columns (double TCD)

n-Butane

n-Pentane

0 100 200 300 400 500 600 700 800

Seconds

n-Propane

i-Butane n-Butaneneo-Pentane i-Pentane n-Pentane

C6+

Carbondioxide

Oxygen

EthaneNitrogen

NGA using micropacked columns, including O2 / N2 separation (single TCD)

Methane

Calorific Value reporting by GAS NGA calculation package for Chromeleon

Page 10: Key Performance - Gassite

Applications

BiogasFast on-line analysis of Biogas. Optional channel available for low ppm Siloxanes (FID).

Seconds

Biogas analysis using Rt-QBond column (TCD)

20 40 60 80

Carbon dioxide

Ethane

Propane

Seconds

Biogas analysis on Molsieve plot column (TCD)

30 60 90

Methane

Nitrogen

Oxygen

Carbon monoxide

Hydrogen

Hydrogen Sulfide

10

Transformer Oil Gas AnalyserCompactGC with Head Space autosampler offers fully unattended analysis of transformer oil.

Minutes

Morgan Schaffer TOGA control sample 100 ppm (TCD)

0 2 4 6 8 10 12 14

Hydrogen

MethaneCarbonmonoxide

Minutes

Morgan Schaffer TOGA control sample 100 ppm (Methaniser/FID)

0 2 4 6 8 10 12 14 16

Methane

Carbonmonoxide

Carbondioxide

Acetylene

Ethane

Ethylene

Oxygen Nitrogen

Page 11: Key Performance - Gassite

ApplicationsFermentation - Biotech - Online measurementCompactGC is resistent against high water percentages, and is therefore very suitable for fermentation and biotech reactor experiments.

Seconds

Headspace analysis of beer fermentation

145 155 165 175 185 195 205 215 225

Aceet-aldehyde

Ethylacetate

DMSEthanol

Monitoring of solvent reduction by bacteriaSeconds

0 10 20 30 40 50

Methanol

Ethanol

Acetone

Pentane

Methylene Chloride

Methylethylketone

Trichloroethylene

Carbon Tetrachloride

Benzene

Air Water

Low SulphurCompactGC with Pulsed Flame Photometric Detector: Selective analysis of low ppb Sulphur components. Analysis without pre-concentration on a small footprint.

Seconds

1 ppm volatile Sulphur components in N2 balance (PFPD)

Hydrogen Sulfide

Methyl mercaptan

Carbonyl Sulfide

0 90

11

Permanent gases on Rt-QBond (TCD)

12 14 16 18 20 22 24 26 28 30 32 34 36

Hydrogen

Oxygen

Nitrogen

Methane

Seconds

Permanent gases on Molsieve plot column (TCD)

30 40 50 60 70 80 90 100 110 120 130

Seconds

C3-C3=

C2-C2=

N2OCO2

Air

Water

Page 12: Key Performance - Gassite

12

ApplicationsAir toxics, VOC, BTX, Low Sulphur with pre-concentrator moduleppb/ppt concentration levels are measured using the optional Thermal Desorption module. Pre-packed traps with various (multibed) adsorbents are available to cover many applications.

Seconds

180 220 260 300 340 380 420 460 500

Benzene

Toluene

m/p-Xylene

Ethylbenzene

o-Xylene

Time (min) 5 10 15

1 Hydrogen sulfide

2 Methanethiol

3 Dimethyl disulfide (DMS)

4 Dimethyl disulfide (DMDS)

1

2

20 pbb

10 pbb

3

4

12

3

4

100 ppb BTEX in N2 (FID). LOD: 0.5 ppb (25 ml sample) Low level ppb Sulphur components (PFPD). LOD: 0.5 ppb (25 ml sample)

NEWThermal

Desorption Option

Page 13: Key Performance - Gassite

13

Testimonials●With the CompactGC we are able to control the

impurity levels in our production process and

hence increase its efficiency. The return on

investment was positive in a short timescale.

Karel Anthonissen, Process Engineer Philips

Innovative Applications N.V. Turnhout Belgium

●CompactGC offers the robustness of a process

GC and the versatility of a laboratory

instrument. Trace levels of polar components are

correctly analysed due to the inert sample path. In

practise, the CompactGC has proven to be a very

fast and reliable analyser.

Mr. Markus Beinlich, laboratory manager RWE

Metering GmbH Germany

● ‘The CompactGC is a very reliable performer thanks

to the use of robust valves and solid detectors. It is

a very flexible microGC, because we can change

the separation columns ourselves in minutes

when new experimental conditions require that.

And we are very satisfied about the support

offered by GAS.

Prof. Emiel Hensen, Schuit Institute of Catalysis,

Eindhoven University of Technology,

The Netherlands

●‘The concept of the GAS CompactGC

features the best of both worlds: it combines

the advantages of a micro GC with the technique

of a traditional lab GC. Being able to replace or

modify individual parts bypasses disadvantages of

modularity and offers increased flexibility and

reduced costs of ownership’.

Manfred van Zuijlen, Researcher on-line GC, Shell

Global Solutions International BV Amsterdam,

The Netherlands

●The technical choices made (valves, surface

treatment ...) in the CGC’s design ensure a high

level of reliability and robustness. The rapid,

repeatable and linear measurements allow

a significant reduction in their uncertainty. These

qualities make this equipment an obvious choice

for a metrology laboratory such as the LNE.

Christophe Sutour, Technical Manager,

LNE- Laboratoire national de métrologie et d’essais,

Paris, France

●The support and care delivered by GAS

are above expectations.

Kevin Mouthaan, Technician Chemical Engineering

and Catalysis, Delft University of Technology,

The Netherlands

13

Page 14: Key Performance - Gassite

14

SpecificationsDimensions: - 19” standard enclosure, 4HE

- 45 x 18 x 55 cm (w*h*d)

- footprint (w*d):

45*55 (landscape position)

18*55 (portrait position)

- weight: 22 kg (2 ch. Instrument)

Power: 100-240V AC, 47-63 Hz,

12A max

Heated zones: up to 10

Electronic Gas Control - connection: 1/8” Swagelock

- max 7 modules;

up to 14 channels

- max inlet pressure 1000kPa

- max outlet pressure

350/700 kPa

- max splitflow: 100 ml/min

- carrier gas He, Nr, Ar, (H2)

- detector gas depending on

detector

Analysis channels: 1-4; extendable

Valve oven: - independently heated, 40-150 °C

- up to 3 valves

- gas sampling or liquid sampling valves

- diaphragm valves with internal purge

Column ovens: - independently heated, 40-250 °C

- up to 3

Detectors - Thermal Conductivity Detector (TCD)

- Pulsed Discharge Detector (PDD/PDHID)

- Plasma Emission Detector (PED)

- Photo Ionisation Detector (PID)

- Pulsed Flame Photometric Detector

(PFPD)

- Mass Spectrometer (MS)

Detector MDA Linearity Remarks

Thermal Conductivity Detector

400pg C2= /ml He

105

Flame Ionisation Detector

1.5 pg C/sec

>107 Flame out detection

Pulsed Discharge Detector

Low pg range

No radioactive source

Plasma Emission Detector

Low pg range

No radioactive source

Photo Ionisation Detector

1*1012 g Benzene

105

Pulsed Flame Pho-tometric Detector

<1 pg S/sec

2.5 orders (5 orders quadratic)

Detects 28 elements

Mass Spectro- meter, external

See MSspec

Page 15: Key Performance - Gassite

15

SpecificationsColumns - capillary 0.32/0.25mm ID and smaller

- wide bore MXT 0.53mm ID

- micropacked 1/16” OD

Column switching - diaphragm valve

- Deans backflush or heartcut

Pre-Concentrator - Trap low temperature 0-50 ºC

- Trap desorption temperature 50-350 ºC

- Trap heating rate 100 ºC/s

- Trap loading using internal sample pump and

Mass Flow Controller

Sampling - inlet temperature: max. 150 °C,

- inlet pressure: max. 21 bar

(depending on used valve)

- connection: 1/8” Swagelock

- injection volume: 0.06 ul to

5ml (loop)

Repeatability - 0.2% RSD or better (n=10, Ethane, TCD detection)

Data communications - LAN for GC parameter control and data transfer

- Instrument Control drivers for Chromeleon, EZChrom,

OpenLab

- optional CGC editor for GC method editing (LAN/RS-232)

- 8 * external events 24V (valve control)

- 4 * SSR output

- 3 * digital input (remote start-stop-prep)

- optional analogue output (0-1V) for detector signals

Environmental conditions - Ambient operating temperature: 5-45 °C

- Ambient operating humidity: 5-95 %,

(non-condensable)

Page 16: Key Performance - Gassite

www.gassite.com

P.O. Box 2148, 4800 CC Breda, The NetherlandsT: +31 (0)76 5411800 F: +31 (0)842 206757 E: [email protected] I: www.gassite.comV

3-20