Page 1: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

Triumph with • Improved Peak Shape for Bases • Shorter Run Times • Core-Shell Performance from pH 1-12

Kinetex EVO C18 Core-Shell UHPLC/HPLC Columns

Pssst! Waters® XBridge®

and BEH™ user, now you can…

EVOC18

Page 2: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

2

Less Band Broadening

More Band Broadening

The PresentThe PastTraditional fully porous hybrid

• More band broadening

• Lower sensitivities

• Longer run times

Kinetex® Core-Shell EVO C18

• Sharper peaks • Increased sensitivity

• Higher efficiencies • Shorter run times

O

SiCH2

CH2

O

O

O

Si S

O

O

Si

O O

O

O

Si

O

O Si Si

O

O

Si

O

O

O

Si Si

O

Si SiOOCH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

C

C

O

SiCH2

CH2

O

O

O

Si Si

O

O

Si

O O

O

O

Si

O

O Si Si

O

O

Si

O

O

O

Si Si

O

Si SiOOCH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

CH2

C

Kinetex EVO C18 uses a patented organo-silica grafting

process which incorporates uniform stabilizing ethane

cross-linking to provide resistance to high pH attack

while maintaining mechanical strength of the core-

shell particle.

Introducing Kinetex EVO C18

The Kinetex core-shell EVO C18 provides much

greater LC performance than fully porous products,

giving you the freedom to get improved results,

cost savings, and overall higher productivity. The

patented surface technology of the Kinetex EVO C18

also provides you with the ruggedness to take on 100%

aqueous conditions and reversed phase methods from

pH 1-12.

Kinetex EVO is patented by Phenomenex. U.S. Patent Nos. 7,563,367 and 8,658,038 and foreign counterparts.

© 2015 Phenomenex, Inc. All rights reserved.

Phenomenex l WEB: www.phenomenex.com

Page 3: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

3

The Present How Does Kinetex EVO C18 Perform?

Upgrade from XBridge and BEH C18 columns:

XBridge® and BEH™ C18 Kinetex

EVO C18 Page(s)

3.5 µm vs 5 µm 4

5 µm vs 5 µm 4-5

1.7 µm vs 2.6 µm 6-7

1.7 µm vs 1.7 µm 8-9

5 µm OBD™ vs 5 µm AXIA™ 10

Ordering Information 11

EVOC18

With our vast range of clients, our quality control lab

receives all kind of sample to analyse. This column

is often used to analyse a large range of samples.

The low cost combined with the great range of pH

in which this column can be used saves us a lot of

trouble and gives us a great price/utility ratio.

Pierre-Alexandre Paquette Solartech Laboratories, Inc.

This column works better than any column

I have used for method development.

Jenn Jazikoff Dynamic Pharmaceuticals

“”

Phenomenex l WEB: www.phenomenex.com

Page 4: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

4

With the combination of 100 % aqueous stability, pH ruggedness from 1-12, and the core-shell performance

advantage, you can easily replace old hybrid silica columns and gain immediate method improvements without

increasing backpressure.

Kinetex® 5 µm EVO C18 150 x 4.6 mm

Waters® XBridge® 3.5 µm C18 150 x 4.6 mm

Waters XBridge 5 µm C18 150 x 4.6 mm

min0 2

2

3

1

4

4

5

6

6

7

8 10

mAU

22718

0

20

40

60

80

100

120

140

min0 2 4 6 8 10 12 14

mAU

0

20

40

60

80

100

120

140

22719

2

31

56

4

7

min0 2 4 6 8 10 12 14

mAU

0

20

40

60

80

100

120

140

22720

2

31

56

4

7

AP

P ID

227

18

AP

P ID

227

19A

PP

ID 2

2720

Conditions for all columns:

Column: Kinetex 5 µm EVO C18XBridge 5 µm C18XBridge 3.5 µm C18

Dimensions: 150 x 4.6 mm

Mobile Phase: A: 0.1 % Formic acid in Water B: 0.1 % Formic acid in Acetonitrile

Gradient: 5 % to 95 % B over 10 minutes

Flow Rate: 1.5 mL/min

Temperature: Ambient

Detection: UV @ 254 nm

Sample: 1. Uracil2. Pindolol3. Chlorpheniramine4. Nortriptyline5. 3-Methyl-4-Nitrobenzoic acid6. 5-Methyl Salicyl Aldehyde7. Hexaphenone

Comparative separations may not be representative of all applications.

What an increase in sensitivity & peak shape!

EVOC18

Upgrade Your HPLC Methods for

Greater Sensitivity

143 Bar 364 Bar

171 Bar

Phenomenex l WEB: www.phenomenex.com

Page 5: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

5Comparative separations may not be representative of all applications.

Easy improvements in resolution, productivity, and sensitivity are just a column away. By upgrading your methods

from traditional fully porous hybrid materials to Kinetex 5 µm EVO C18 Core-Shell Technology you can expect

excellent performance and pH durability at low HPLC backpressures!

Kinetex 5 µm EVO C18 150 x 4.6 mm

Waters® XBridge® 5 µm C18 150 x 4.6 mm

2

1

3 4

5

min0 2 4 6 8 10 12 14

mAU

0

5

10

15

20

25

30

22768

mAU

0

2

4

6

8

10

12

14

min0 2 4 6 8 10 12 14

1

2

3

4

5

AP

P ID

227

88A

PP

ID 2

3108

Conditions for all columns:

Column: Kinetex 5 µm EVO C18XBridge 5 µm C18

Dimensions: 150 x 4.6 mm

Mobile Phase: A: 20 mM Sodium Phosphate (pH 7.0 or pH 10.2)B: Methanol/Acetonitrile (50:50)

Isocratic: A/B (30:70)

Flow Rate: 1.25 mL/min

Temperature: 30 ˚C

Detection: UV @ 254 nm

Sample: 1. Protriptyline2. Nortriptyline3. Imipramine4. Amitriptyline5. Clomipramine

min0 2

2

1

3

4

4

5

6 8 10 12 14

mAU

0

5

10

15

20

25

30

22779

mAU

0

2

4

6

8

10

12

14

min0 2

1

2

3

4

5

4 6 8 10 12 14

AP

P ID

227

79A

PP

ID 2

3107

EVOC18

Wow! Better peak shape and better performance in less time!

Advance pH 1-12 Methods Ahead with

Better Peak Shape

pH 7.0

pH 10.2

pH 7.0

pH 10.2

Phenomenex l WEB: www.phenomenex.com

Page 6: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

6

Conditions for all columns unless otherwise noted:Columns: Kinetex 2.6 µm EVO C18

ACQUITY BEH 1.7 µm C18Dimensions: 50 x 2.1 mm

Mobile Phase: A: Water with 0.1% Formic acidB: Acetonitrile with 0.1% Formic acid

Gradient: Time (min)05

% B595

Flow Rate: 0.8 mL/min (Kinetex)0.4 mL/min (BEH)

Temperature: 30 ˚CDetection: MS/MS

Sample: Proprietary Drug Formulation

Comparative separations may not be representative of all applications.

2.6 µm

Waters® ACQUITY® BEH™ 1.7 µm C18

Kinetex 2.6 µm EVO C18

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min0.0

5.0e4

1.0e5

1.5e5

2.0e5

2.5e5

3.0e5

3.5e5

4.0e5

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min0.0

2.0e44.0e46.0e48.0e41.0e51.2e51.4e51.6e51.8e52.0e52.2e52.4e52.6e52.8e5

AP

P ID

231

58

Replace Fully Porous Sub-2 µm Columns for

Greater Productivity

0.4 mL/min407 Bar

Inte

nsity

, cp

sIn

tens

ity, c

ps

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 min0.0

5000.0

1.0e4

1.5e4

2.0e4

2.5e4

3.0e4

3.5e4

4.0e4

4.5e4

5.0e4

5.5e4

0.02000.04000.06000.08000.0

1.0e41.2e41.4e41.6e41.8e42.0e42.2e42.4e42.6e42.8e43.0e4

0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 min

0.8 mL/min352 Bar

Critical Separation from Interference

Kinetex

35% Faster Analysis

Utilize Kinetex® 2.6 µm core-shell columns to take full advantage of the flow rates that are available on your UHPLC

system. Since Kinetex 2.6 µm core-shell columns display similar or better performance at a lower pressure than fully

porous sub-2 µm columns, you can increase your flow rate with the Kinetex 2.6 µm column and achieve some very

useful time savings and productivity gains, while holding onto the necessary performance parameters.

BEH

AP

P ID

231

59

Phenomenex l WEB: www.phenomenex.com

Page 7: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

7

min0 1 2 3

AU

0.5 1.5 2.5 3.5

0

0.1

0.2

0.3

0.4

0.5

Replace Fully Porous Sub-2 µm Columns for

Increased Resolution

Waters® ACQUITY® BEH™ 1.7 µm C18 Kinetex 2.6 µm EVO C18

50 x 2.1 mm 401 Bar

50 x 2.1 mm 192 Bar

Flow Rate: 0.4 mL/min

Temperature: 22 ˚C

Detection: UV @ 210 nm

Sample: 1. Green Coffee Extract (Chlorogenic Acid and Antioxidants)

Conditions for all columns unless otherwise noted:

Column: Kinetex 2.6 µm EVO C18ACQUITY BEH 1.7 µm C18

Dimensions: 50 x 2.1 mm

Mobile Phase: A: 20 mM Potassium phosphate (pH 2.5)

B: Acetonitrile

Gradient: Time (min)056

6.510

% B5707055

Comparative separations may not be representative of all applications.

Kinetex 2.6 µm EVO C18

Run a Kinetex 2.6 µm column in the same dimension as your sub-2 µm fully porous column.

Utilize a longer Kinetex column to further increase resolution

100 x 2.1 mm331 Bar

AP

P ID

230

03

Start

What’s Next?

0 1 min

AU

0

0.2 0.4 0.6 0.8 1.2 1.4 1.6 1.8

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0 1 min

AU

0

0.2 0.4 0.6 0.8 1.2 1.4 1.6 1.8

0.1

0.2

0.3

0.4

0.5

0.6

0.7

AP

P ID

230

01

AP

P ID

230

00

Similar or better performance at a lower backpressure on the Kinetex column.

Rs: 1.10 1.21Rs: 2.31 0.95

Rs: 2.37 1.92

Phenomenex l WEB: www.phenomenex.com

Page 8: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

8

Improve Your UHPLC Analyses with

Faster Results and Better Sensitivity

Comparative separations may not be representative of all applications.

After enjoying the greater sensitivity, resolution, and peak shape given by

the Kinetex EVO C18 for this isolated standard, the same performance

benefits can easily be realized for the more complex extracted sample.

Conditions for both columns:Column: Kinetex 1.7 µm EVO C18

ACQUITY BEH 1.7 µm C18Dimensions: 50 x 2.1 mm

Mobile Phase: A: 0.1 % Formic Acid in WaterB: 0.1 % Formic Acid in Methanol

Gradient: Time (min) 05

% B2

95Flow Rate: 0.5 mL/min

Temperature: 30 º CDetection: UV @ 254 nm

Sample: 1. Berberine

Berberine Standard

Goldenseal Extract (Containing Berberine)

Increase in Peak Height Response57 %

1 2 3 min

mAU

0

20

40

60

80

100

1201

AP

P ID

231

11 a

nd 2

3109

Greater Resolution of Impurities

Kinetex 1.7 µm EVO C18

Kinetex 1.7 µm EVO C18

Waters® ACQUITY® BEH™ 1.7 µm C18

min1 2

1

3

23110mAU

0

25

50

75

100

125

150

175

200

AP

P ID

231

10

By upgrading from fully porous hybrid UHPLC columns to the Kinetex® 1.7 µm EVO C18, you can keep the pH

stability but gain much greater performance. With the core-shell advantage at your disposal your new methods will

be faster and more sensitive, leaving room for greater productivity.

Phenomenex l WEB: www.phenomenex.com

Page 9: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

9

1

2

4

53

min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5

AU

-0.010

0.010.020.030.040.050.060.07

More Performance, Excelling Quality

Comparative separations may not be representative of all applications.

Here’s an illustration of just how powerful a tool the Kinetex® 1.7 µm EVO C18 is. In comparison with multiple batches

of a common hybrid sub-2 µm column line, Kinetex EVO C18 columns show consistently greater performance and

better peak shape for basic compounds. This level of quality is what we strive to give you every day with every

batch, every column, and even every phone call.

Conditions for both columns:Column: Kinetex 1.7 µm EVO C18

ACQUITY BEH 1.7 µm C18Dimensions: 150 x 2.1 mm

Mobile Phase: A: 0.1 % Formic Acid in WaterB: 0.1 % Formic Acid in Acetonitrile

Gradient: Time (min) 056

6.510

% B5

959555

Flow Rate: 0.5 mL/minTemperature: 22 º C

Detection: UV @ 254 nmSample: 1. 4-Hydroxybenzoic Acid

2. Labetol3. Propanolol4. Protriptyline5. Ibuprofen

EVOC18

Kinetex 1.7 µm EVO C18

Waters® ACQUITY® BEH™ 1.7 µm C18

min0 0.5 1 1.5 2 2.5 3 3.5 4 4.5

1

2

4

53

AU

-0.010

0.010.020.030.040.050.060.07

AU

-0.010

0.010.020.030.040.050.060.07

min0 0.5 1

1

1.5 2

2

2.5 3

4

53

3.5 4 4.5

AU

-0.010

0.010.020.030.040.050.060.07

min0 0.5 1

1

1.5 2

2

2.5 3

4

53

3.5 4 4.5

AP

P ID

230

86A

PP

ID 2

3088

AP

P ID

230

87A

PP

ID 2

3089

Batch A

Batch A

Batch BAverage Peak CapacityA&B: 168

Wow! Look at my improved peak

shape for bases!

Batch B

Increase in Peak Height Response

Average Peak CapacityA&B: 262

Phenomenex l WEB: www.phenomenex.com

Page 10: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

10

Kinetex® EVO C18 will easily pave the way to enhanced purification and isolation levels with

its incredible pH durability and core-shell performance gains. Get ready for a future of high

yields and chromatographic improvements over traditional fully porous and hybrid materials.

Kinetex 5 µm EVO C18 150 x 21.2 mm AXIA™

Waters® XBridge® 5 µm C18 150 x 19 mm Prep OBD™22769

0 1 2 3 4 min-2500

0

2500

5000

7500

10000

12500

15000

17500

20000

22500

25000

AP

P ID

227

68

AP

P ID

227

69

Conditions for all columns:

Column: Kinetex 5 µm EVO C18 Axia packedXBridge 5 µm C18 Prep OBD

Dimensions: 150 x 21.2 mm (Kinetex)150 x 19 mm (XBridge)

Mobile Phase: A: 0.1 % TFA in WaterB: 0.1 % TFA in Acetonitrile

Gradient: 5-95 % B in 4 min.Hold for 1 min.Re-equilibrate for 2 min.

Flow Rate: 25 mL/min

Temperature: Ambient

Detection: UV @ 254 nm

Sample: Amitriptyline

Sample Load: 2 mg 4 mg 10 mg 20 mg 40 mg 80 mg

mVm

V

22768

0 1 2 3 4 min-2500

0

2500

5000

7500

10000

12500

15000

17500

20000

22500

25000

Amazing Peak Shape Over All Sample Loads

Enhance Purificationand Isolation Levels

Comparative separations may not be representative of all applications.

www.phenomenex.com/ KinetexApps

www.phenomenex.com/ KinetexEVO

Learn more online about how Kinetex EVO C18 can enhance your HPLC methods!

Phenomenex l WEB: www.phenomenex.com

Page 11: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

11

Ordering Information5 μm Minibore Columns (mm)

SecurityGuard™ ULTRA Cartridges‡

Phases 30 x 2.1 50 x 2.1 100 x 2.1 150 x 2.1 3/pk

EVO C18 00A-4633-AN 00B-4633-AN 00D-4633-AN 00F-4633-AN AJ0-9298for 2.1 mm ID

5 μm MidBore™ Columns (mm)SecurityGuard

ULTRA Cartridges‡

Phases 50 x 3.0 100 x 3.0 150 x 3.0 3/pk

EVO C18 00B-4633-Y0 00D-4633-Y0 00F-4633-Y0 AJ0-9297for 3.0 mm ID

5 μm Analytical Columns (mm)SecurityGuard

ULTRA Cartridges‡

Phases 50 x 4.6 100 x 4.6 150 x 4.6 250 x 4.6 3/pk

EVO C18 00B-4633-E0 00D-4633-E0 00F-4633-E0 00G-4633-E0 AJ0-9296for 4.6 mm ID

5 μm Axia™ Packed Preparative Columns (mm)SecurityGuard

PREP Cartridges*Phases 50 x 21.2 100 x 21.2 150 x 21.2 250 x 21.2 15 x 21.2

/eaEVO C18 00B-4633-P0-AX 00D-4633-P0-AX 00F-4633-P0-AX 00G-4633-P0-AX AJ0-9304

for 21.2 mm ID

5 μm Axia Packed Preparative Columns (mm)SecurityGuard

PREP Cartridges**Phases 50 x 30 100 x 30 250 x 30 15 x 30

/eaEVO C18 00B-4633-U0-AX 00D-4633-U0-AX 00G-4633-U0-AX AJ0-9305

for 30 mm ID

2.6 μm Minibore Columns (mm)SecurityGuard™

ULTRA Cartridges‡

Phases 30 x 2.1 50 x 2.1 100 x 2.1 150 x 2.1 3/pk

EVO C18 00A-4725-AN 00B-4725-AN 00D-4725-AN 00F-4725-AN AJ0-9298for 2.1 mm ID

2.6 μm MidBore™ Columns (mm)SecurityGuard

ULTRA Cartridges‡

Phases 50 x 3.0 100 x 3.0 150 x 3.0 3/pk

EVO C18 00B-4725-Y0 00D-4725-Y0 00F-4725-Y0 AJ0-9297for 3.0 mm ID

2.6 μm Analytical Columns (mm)SecurityGuard

ULTRA Cartridges‡

Phases 50 x 4.6 100 x 4.6 150 x 4.6 3/pk

EVO C18 00B-4725-E0 00D-4725-E0 00F-4725-E0 AJ0-9296for 4.6 mm ID

1.7 μm Minibore Columns (mm)SecurityGuard

ULTRA Cartridges‡

Phases 50 x 2.1 100 x 2.1 150 x 2.1 3/pk

EVO C18 00B-4726-AN 00D-4726-AN 00F-4726-AN AJ0-9298for 2.1 mm ID

‡SecurityGuard ULTRA Cartridges require holder, Part No.: AJ0-9000

*PREP SecurityGuard Cartridges require holder, Part No.: AJ0-8223

**PREP SecurityGuard Cartridges require holder, Part No.: AJ0-8277

Phenomenex l WEB: www.phenomenex.com

If you are not completely

satisfied with Kinetex core-

shell columns, send in

your comparative data to

a similar product with the

Kinetex column within 45

days for a FULL REFUND.

Page 12: Kinetex EVO C18 - Phenomenexphx.phenomenex.com/lib/br27210615_W.pdf · Core-Shell Technology 3 The Present How Does Kinetex EVO C18 Perform? Upgrade from XBridge and BEH C18 columns:

Core-Shell Technology

Terms and Conditions Subject to Phenomenex Standard Terms & Conditions, which may be viewed at www.phenomenex.com/TermsAndConditions.

Trademarks Phenomenex and Kinetex are registered trademarks of Phenomenex. Axia, MidBore, and SecurityGuard are trademarks of Phenomenex. Waters, ACQUITY, UPLC, and XBridge are registered trademarks, BEH Technology and OBD are trademarks of Waters Corporation. SelectScience is a registered trademark of SelectScience Ltd.

Disclaimer Phenomenex is not affiliated with Waters Corporation. Comparative separations may not be representative of all applications.

Axia column and packing technology is patented by Phenomenex. U.S. Patent No. 7, 674, 383

Kinetex EVO is patented by Phenomenex. U.S. Patent Nos. 7,563,367 and 8,658,038 and foreign counterparts.

The opinions stated herein are solely those of the speaker and not necessarily those of any company or organization.

© 2015 Phenomenex, Inc. All rights reserved.

www.phenomenex.comPhenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department at [email protected]

Australiat: +61 (0)2-9428-6444 f: +61 (0)2-9428-6445

[email protected]

Austriat: +43 (0)1-319-1301f: +43 (0)1-319-1300

[email protected]

Belgiumt: +32 (0)2 503 4015 (French)t: +32 (0)2 511 8666 (Dutch)f: +31 (0)30-2383749

[email protected]

Canadat: +1 (800) 543-3681f: +1 (310) 328-7768

[email protected]

Chinat: +86 (0)20 2282-6668f: +86 (0)20 2809-8130

[email protected]

Denmarkt: +45 4824 8048f: +45 4810 6265

[email protected]

Finlandt: +358 (0)9 4789 0063f: +45 4810 6265

[email protected]

Francet: +33 (0)1 30 09 21 10 f: +33 (0)1 30 09 21 11

[email protected]

Germanyt: +49 (0)6021-58830-0f: +49 (0)6021-58830-11

[email protected]

Indiat: +91 (0)40-3012 2400f: +91 (0)40-3012 2411

[email protected]

Irelandt: +353 (0)1 247 5405f: +44 1625-501796

[email protected]

Italyt: +39 051 6327511f: +39 051 6327555

[email protected]

Luxembourgt: +31 (0)30-2418700 f: +31 (0)30-2383749

[email protected]

Mexicot: 001-800-844-5226f: 001-310-328-7768

[email protected]

The Netherlandst: +31 (0)30-2418700 f: +31 (0)30-2383749

[email protected]

New Zealandt: +64 (0)9-4780951f: +64 (0)9-4780952

[email protected]

Norwayt: +47 810 02 005f: +45 4810 6265

[email protected]

Puerto Ricot: +1 (800) 541-HPLCf: +1 (310) 328-7768

[email protected]

Spaint: +34 91-413-8613f: +34 91-413-2290

esp [email protected]

Swedent: +46 (0)8 611 6950f: +45 4810 6265

[email protected]

United Kingdomt: +44 (0)1625-501367f: +44 (0)1625-501796

[email protected]

USAt: +1 (310) 212-0555f: +1 (310) 328-7768

[email protected]

All other countries Corporate Office USA

t: +1 (310) 212-0555f: +1 (310) 328-7768

[email protected]

Triumph with • Improved Peak Shape for Bases • Shorter Run Times • Core-Shell Performance from pH 1-12 Kinetex EVO C18 Core-Shell UHPLC/HPLC Columns

BR

2721

0615

_W