lab 1 complete 15-08
DESCRIPTION
lab 1TRANSCRIPT
1.0 INTRODUCTION
Many types of devices have been developed to measure fluid velocities and
flowrates based on the Bernoulli equation. The orifice, nozzle meter and the venturi
meter are the three common used types of flow meter. The operation of each is based on
the same physical principles: an increase in velocity causes a decrease in pressure.
In venturi meter, the pressure difference between the upstream side of the cone
and the throat is measured and provides the signal for the rate of flow. Fluid is
accelerated through a converging cone of angle 15-20o. The orifice meter consists of a
flat orifice plate with a circular hole drilled in it. There is a pressure tap upstream from
the orifice plate and another just downstream. There are in general three methods of
placing the taps. The coefficient of the meter depends upon the position of taps.
In this experiment, the water level readings were taken for each controlled flow
rate after the water was pumped into venturi and orifice meter. There are total 10
readings of pressure drop. The pressure drop against each device is used for calculation.
Then , the operation and performance of each device was compared.
The remainder of the report provides detailed analysis of the Theory and
Working equations for this experiment. The procedure can be found at material and
method section. Presentation of the findings and detailed discussion of the result can be
found in Result and Discussion section. Finally, the conclusion and recommendation part
will show summary of this experiment.
1
2.0 THEORY AND WORKING EQUATIONS
Bernoulli's principle states that in fluid flow, an increase in velocity occurs
simultaneously with decrease in pressure. This principle is a simplification of Bernoulli's
equation which states that the sum of all forms of energy in a fluid flowing along an
enclosed path (a streamline) is the same at any two points in that path. The Bernoulli
equation is an approximate relation between pressure, velocity, and elevation, and is
valid in regions of steady, incompressible flow where the net frictional forces are
negligible. The Bernoulli equation is a very handy equation useful in almost any
engineering application. Bernoulli’s equation neglects friction, but friction effects in
these devices become significant and are normally accounted for by introducing
empirical coefficients and retaining the frictionless form of Bernoulli’s equation, rather
than by introducing the friction term into Bernoulli’s equation. Bernoulli's equation is
only valid if one assumes the following: incompressible fluid (fluid velocity less than
one third the speed of sound) and inviscid flow (this just means that the point in question
along the flow is going to be away from where the flow and the object come into
contact). The value of discharge coefficient, Cd differs for each device. The devices used
in this experiment to measure flow rate are venturi meter, orifice meter, rotameter and
bench device. Each of these meters operates on the principle that a decrease in flow area
in pipes causes an increase in velocity that is accompanied by the decrease in pressure.
In this experiment, a few assumptions are made for data analysis:
1) density of water is constant and taken to be 1000kg/m3
2) the flow is steady and incompressible
3) the pipe is horizontal which means no change in elevation
2
Bernoulli's equation reduces to an equation relating the conservation of energy at two
points in the fluid flow: Z1 = Z2
Bernoulli’s equation
or:
with: or V = Q / A
Venturi meter
The venturi meter was invented by an American engineer, Clemans Herschel
(1842-1930) and named by him after Italian Giovanni Venturi (1746-1822) for his
pioneering work on conical flow section, is the most accurate flow meter in this group,
but it is also the most expensive.The angle of the convergent cone is usually 200, the
length of throat is equal to the throat diameter and the angle of divergent cone is 5 0 to 70
to ensure a minimum loss of energy. Venturi meter consists of a short converging
conical tube, leading to a cylindrical portion called “throat” which is followed by a
diverging section.
A fluid passing through smoothly varying constrictions is subject to changes in
velocity and pressure in order to satisfy the conservation of mass-flux (flow rate). The
reduction in pressure in the constriction can be understood by conservation of energy.
The fluid gains kinetic energy as it enters the constriction, and that energy is supplied by
a pressure gradient force from behind. The pressure gradient reduces the pressure in the
constriction, in reaction to the acceleration. Likewise, as the fluid leaves the constriction,
3
it is slowed by a pressure gradient force that raises the pressure back to the ambient
level. Therefore, the flow rate through a venturi meter is given by:
Flow rate, Q = Cd x Ao [1 – (Ao /A) 2]-1/2 x [2(P2 – P3 )/ ρ]1/2
Where Cd = discharge coefficient (0.97)
Ao = throat area = 2.011 x 10-4 m2
A = inlet area = 3.8 x 10-4 m2
ρ = density of water = 1000 kg/m3
P2 = inlet pressure (Pa)
P3 = throat pressure (Pa)
g = 9.81 ms-2
D2 = inlet diameter = 0.022 m
D3 = throat diameter = 0.016 m
Orifice meter.
The orifice meter has the simplest design and it occupies minimal space as it
consists of a plate with a hole in the middle. It is usually placed in a pipe in which fluid
flows. There are considerable variations in design such as sharp-edged, beveled and
rounded. An orifice meter is used to measures the rate of fluid flow. As fluid flows
through the pipe, it has a certain velocity and a certain pressure. When the fluid reaches
the orifice plate, with the hole in the middle, the fluid is forced to converge to go through
the small hole; the point of maximum convergence actually occurs shortly downstream
of the physical orifice, at the so-called "vena contracta". As it does so, the velocity and
the pressure changes. Beyond the vena contracta, the fluid expands and the velocity and
4
pressure change once again. By measuring the difference in fluid pressure between the
normal pipe section and at the vena contracta, the volumetric and mass flow rates can be
obtained from Bernoulli's equation.
Flow rate, Q = Cd x Ao x [2g (H7-H8)] 1/2 x [1-(Ao/A) 2]-1/2
Where Cd = discharge coefficient (0.61)
D4 = orifice diameter (0.20 m)
D = orifice upstream diameter (0.025 m)
Ao = orifice area (3.1 x 10-4 m)
A = orifice upstream area (4.9 x 10-4 m)
(H7-H8)= pressure difference across orifice (m)
Rotameter
A simple, reliable, inexpensive, and easy-to-install flow meter with low pressure
drop and no electrical connections that gives a direct reading of flow rate for a wide
range of liquids and gases is the variable-area flow meter, also called a rotameter. A
rotameter consists of a vertical tapered conical transparent tube made of glass or plastic
with a float inside that is free to move. As fluid flows through the tapered tube, the float
rises within the tube to location where the float weight, drag force, and buoyancy force
balance each other and the net force acting on the float is zero. The flow rate is
determined by simply matching the position of the float against the graduated flow scale
outside the tapered transparent tube.
Pressure drop
5
Pressure drop can be expressed as the loss of pressure head hL and acceleration of
gravity g.
∆P/ρ=ghL where hL is the head loss
Pipe losses, H = K x V2/2g where: K = coefficient of losses
V = velocity of flow
g = 9.81m/s2
3.0 MATERIALS AND METHODS
The apparatus is shown in figure diagram. The Bernoulli’s Theorem apparatus is
level on the hydraulics bench by using adjustable feet. A small amount of wetting agent
was injected into the test section and had been ensured that the section has the tapered
14˚ tapered duct converging in the direction of flow. To reverse the test section, the total
head probe was withdrawn before releasing the couplings.
The manometer tubes were filled with water to discharge all pockets of air from
the system. By adjusting the flow control valve, the level in the manometers had been
raised or lowered as required. The flow control valve once again had been adjusted to
provide the combination of flow rate and system pressure, which give the largest
convenient difference between the highest and lowest manometer levels. The scale
reading of each manometer level was noted. Three sets of readings of volume per time
had been taken to find the flow rate using the volumetric tank. The specific procedures
of the experiment are as follows:
6
1. The probe was inserted to the end of the parallel position of the duct, and the
tapered portion was moved in 1cm at a time. For each position (100mm, 150mm
and 200mm), the distance from the end of the parallel portion had been noted,
and the scale readings of the manometer levels were recorded.
2. This was repeated to give high flow rates at both high and low static pressure for
different combinations of valve openings.
3. The inlet feed was stopped, the apparatus were drained, the probe was fully
withdrawn, the couplings were undo, the test sections were reversed and the
couplings were replaced.
4. The above procedures were repeated.
Equivalent static pressure in the convergent and divergent configurations was set up
using the remote tapping at the flow control valve.
7
4.0 RESULTS AND DISCUSSION
According to the theory, we found that the flow rate that recorded by rotameter
should be higher than the flow rate of venturi and orifice meter. This shows that losses
occurred at venturi and orifice meter. Losses of energy in pipeline are due to shock from
the sudden change of flow rate at the interchanges of section area, this losses also known
as head loss. That is kinetics energy converted into pressure. Losses can also occur due
to the frictional resistance to flow.
However, there is an abnormal in the data obtained from the venturi, where the
flow rate that calculated are larger than that of the rotameter. This deviates from the
theoretical expectation. This is due to the discrepancies during the experiment such as:
Some bubbles still remain in the manometer tube and the water level was resisted
from raised to the actual height.
8
Since the water level of the manometer board fluctuate all the time, it is difficult for
us to measure or observed the actual water level. What we do is took the average
value of the water level fluctuation height.
According to the data tabulated in table 1.1, 2.1 and 3.1 where the discharge of
the venturi is larger than the orifice. This implies that the losses at the orifice always
greater than that at venturi. This is due to the stopping effect of the flow at the orifice
while the contraction at the venturi neck (point 2) are slower, guided by a tapering tube.
The sudden enlargement from vena- contracta (minimum area) formed downstream in
the orifice with an area which is smaller than the actual area of the orifice cause the flow
rate calculated is smaller compared to the venturi meter.
Other reason that cause the deviation is the head loss in the pipe flow such as:
Major losses ( losses occur in straight pipe )
Minor losses ( losses occur in pipe system components)
Percentage of error of venturi is higher than orifice, therefore the accuracy of
venturi is higher than the orifice. This is because:
Orifice meter has relatively high-pressure loss for a given flow rate.
The orifice plate and meter run must be kept clean and retain the original conditions
specified by the standard.
To assume this, periodic inspection should be conducted to reaffirm condition.
Inspection frequency depends on problems of foreign material collection and
9
possible damage. Inspections will confirm the orifice diameter and coefficient of
discharge.
The most important variable in the equation – the one that directly determines flow-
measurement accuracy – is the differential pressure. Therefore, a major effort should
be made to have as a high a differential as possible (considering flowing conditions),
and the best available differential transducers should be specified. Anything that can
be done should be done to improve this most critical measurement factor.
More sensitive to flow disturbances than some meter.
Flow pattern in the meter does not make meter self- cleaning.
Venturi has low permanent pressure loss.
Venturi can be used on slurries and dirty fluids.
Figure 4.1.1 Graph of Head loss versus Velocity for venturi meter and orifice meter (1st
Trial)Figure 4.1: Graph of head loss versus velocity for venturi and orifice (1st Trial)
y = 0.03x - 0.0088R2 = 1
y = 0.0085x - 0.0048R2 = 0.9973
00.0020.0040.0060.008
0.010.012
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6
velocity, m/s
head
loss
, m
venturiorificeLinear (orifice)Linear (venturi)
Figure 4.1.2 Graph of Head loss versus Velocity for venturi meter and orifice meter (2nd
Trial)
10
Figure 4.1: Graph of head loss versus velocity for venturi and orifice (2nd Trial)
y = 0.0187x - 0.0032R2 = 0.9982
y = 0.008x - 0.0042R2 = 0.9949
0
0.002
0.004
0.006
0.008
0 0.2 0.4 0.6 0.8 1 1.2 1.4
velocity, m/s
head
loss
, m
venturiorificeLinear (orifice)Linear (venturi)
Figure 4.1.3 Graph of Head loss versus Velocity for venturi meter and orifice meter (3 rd
Trial)
y = 0.0275x - 0.0072R2 = 1
y = 0.0091x - 0.0055R2 = 0.9958
00.0020.0040.0060.008
0.010.012
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8
velocity, m/s
head
loss
, m
venturiorificeLinear (orifice)Linear (venturi)
5.0 CONCLUSION AND RECOMMENDATIONS
11
In this experiment, flow rate was measured by using the flow apparatus which
consisted of venturi and orifice. Rotameter was acted as a control to the apparatus and
bench device was used to present the actual value of the flow discharge. From the result
achieved, the average percentage error for the venturi meter is between 3.7% to 16.53%
while the orifice reading deviates in the range of 11.79% to 31.42%. On the other hand,
the percentage of error for rotameter is between 9.33% to 15.56%. From the theory,
venturi meter was preferred for the flow rate measurement than rotameter because
venturi meter had lower pipe losses and percentage error compared to the rotameter. So
venturi meter was more accurate than rotameter.
To avoid any error in the experiment result, all precaution when doing this
experiment might be taken. Discharge hose was ensured properly directed to sump tank
of fibre glass before starting up system. Collection tank drain valve also was ensured left
opened to allow flow discharge back into sump tank. In addition to that, ‘Trapped
bubbles’ in the glass tube or plastic transfer tube was ensured been removed. Any
parallax deviation when taking the reading also might be avoided by reading the result
perpendicular to the water meniscus. Besides that, all kind of organism or unwanted
particle was also made sure not been in the apparatus. Moreover, in order to have a
better fluid flow, the flow apparatus might not dissipate in eddies downstream.
REFERENCES
12
Douglas J.F. and Matthews R.D. (1985). Solving Problems In Fluid Mechanics, Vol. 1.
Longman Publication. New York.
Hauser B.A. (1992). Hydraulics For Operations, 3rd Ed. Lewis Publishers. London
Manohar M. and Krishnamachar P. (1990). Fluid Mechanics, Vol. 2. Hydraulic
Machinery & Advanced Hydraulics. Vikas Publishing House Pvt. Ltd. Melbourne.
Massey B.S. (1997). Mechanics Of Fluids, 6th Ed. Van Nostrand Reinhold
(International). Cambridge.
Munson,Young and Okiishi (1998). Fundamentals of fluid mechanics, 3rd Ed. John
Wileys & Sons: New York.
Streeter Victor L., Benjamin Wylie E., and Bedford Keith W. (1995). Fluid Mechanics,
9th Ed. WCB McGraw-Hill. New York.
13
14
APPENDICES
APPENDIX A
RAW DATA
The following are tables related to readings taken at the first trial.
Table A1.1 : Relationship between pressure head and flow rate for both venturi and orifice
MANOMETER
READING ROTAMETER VOLUME TIME
FLOW
RATE,Q
(L/min) (min) USING BERNOULLI'S EQN
(L/min)
1 2 3 4 5 6 7 8 9 10 VENTURI ORIFICE
17.2 14.3 5.2 11 13.7 14.6 15 9 12.7 11 20 20 1 18.29 15.89
18.4 18.5 11.5 16.8 16.4 16.9 17 13.7 15.8 14.8 15 15 1 16.17 18.2
188 18.6 15.5 16.9 17.8 18.1 17.5 17.0 17.6 17 10 10 1 10.75 7.946
14
Table A1.2 : Comparison of percentage error for Venturi Meter and Orifice with Rotameter readings
BENCH VENTURI ORIFICE ROTAMETER
FLOW
RATE
FLOW
RATE % ERROR FLOW RATE % ERROR FLOW RATE % ERROR
(L/min) (L/min) (L/min) (L/min)
22 18.29 8.55 15.89 20.55 20 10
17.5 16.17 7.8 18.2 9 15 16.7
10 10.75 7.5 7.946 20.45 10 5
Table A1.3 : Comparison between venturi and orifice velocity and head loss
VENTURI ORIFICE
HEAD LOSS,m VELOCITY (m/s) HEAD LOSS ,m VELOCITY (m/s)
0.82*10^-2 1.52 0.74*10^-2 0.54
0.64*10^-2 1.34 0.98*10^-2 0.62
0.28*10^-2 0.89
15
The following are tables related to readings taken at the second trial.
Table A2.1: Relationship between pressure head and flow rate for both venturi and orifice
MANOMETER
READING ROTAMETER VOLUME TIME FLOW RATE,Q
(L/min) (min) USING BERNOULLI'S EQN
(L/min)
1 2 3 4 5 6 7 8 9 10 VENTURI ORIFICE
21.3 20.5 9.2 15.1 17.9 19 19.2 13.2 16.9 15 1 20 1 15.29 13.5
22.1 21.7 15.5 18.6 20 29.7 20.8 17.7 19.5 18.5 15 15 1 15.21 8.91
22.4 22.2 `9.2 20.5 21.3 21.7 21.8 20.3 21.1 20.8 10 10 1 10.58 7.95
16
Table A2.2: Comparison of percentage error for Venturi Meter and Orifice with Rotameter readings
Table A2.3: Comparison between venturi and orifice velocity and head loss
BENCH VENTURI ORIFICE ROTAMETER
FLOW
RATE
FLOW
RATE % ERROR
FLOW
RATE % ERROR FLOW RATE % ERROR
(L/min) (L/min) (L/min) (L/min)
23 15.29 23.55 13.5 32.5 20 15
17 15.21 1.4 8.91 40.67 15 13
11 10.58 5.8 7.95 20.57 10 10
VENTURI ORIFICE
HEAD
LOSS, m VELOCITY (m/s) HEAD LOSS,m VELOCITY (m/s)
0.6*10^-2 1.267 0.54*10^-2 0.46
0.57*10^-2 1.261 0.23*10^-2 0.3
0.28*10^-2 0.88 0.19*10^-2 0.27
17
The following are tables related to readings taken at the third trial.
Table A3.1: Relationship between pressure head and flow rate for both venturi and orifice
MANOMETER
READING ROTAMETER VOLUME TIME FLOW RATE,Q
(L/min) (min)
USING BERNOULLI'S
EQN
(L/min)
1 2 3 4 5 6 7 8 9 10 VENTURI ORIFICE
26 25.1 13.9 19.7 22.4 23.6 23.9 17.8 21.6 19.6 1 20 1 20.44 13.39
26.4 26.9 19.5 22.8 24.3 24.9 25.1 21.8 23.7 22.7 15 15 1 13.38 18.20
26.5 26.2 23.4 24.7 25.4 25.7 21.8 20.3 25.3 24.7 10 10 1 10.22 10.58
18
Table A3.2: Comparison of percentage error for Venturi Meter and Orifice wit Rotameter readings
BENCH VENTURI ORIFICE ROTAMETER
FLOW
RATE
FLOW
RATE % ERROR
FLOW
RATE % ERROR FLOW RATE % ERROR
(L/min) (L/min) (L/min) (L/min)
21.5 20.44 2.20 13.39 33.05 20 7.5
17 13.38 10.80 18.20 21.33 15 13
11.5 10.22 17.50 10.58 5.80 10 15
Table A3.3: Comparison between venturi and orifice velocity and head loss
VENTURI ORIFICE
HEAD
LOSS, m VELOCITY HEAD LOSS,m VELOCITY
1*10^-2 1.694 0.54*10^-2 0.46
0.44*10^-2 1.111 0.98*10^-2 0.62
0.36*10^-2 0.974
19
APPENDIX B
Calculation
Calculations for Table A1.1 and A1.2,
For the venturi,
For the flow 20 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (14.3 – 5.2)(9.81) ]0.5
= 3.049 x 10-4 m3/s = 18.29 liter/min.
Percentage error = (18.29-20.00)/20 x 100
= 8.55 %
For the flow 15 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (18.5 – 11.5)(9.81) ]0.5
= 2.694 x 10-4 m3/s = 16.17 liter/min.
Percentage error = (16.17 – 15.00)/15 x 100
= 7.8 %
For the flow 10 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (18.6 – 15.5)(9.81) ]0.5
= 1.793 x 10-4 m3/s = 10.75 liter/min.
Percentage error = (10.75 – 10.00)/10 x 100
= 7.5 %
20
For the orifice,
For the flow 20 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(15.1 – 9.0)]1/2 x [1- (0.633)2]-1/2
= 2.649 x 10-4 m3/s = 15.89 liter/min.
Percentage error = (15.89-20.00)/10 x 100
= 20.55 %
For the flow 15 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(17.0-13.7)]1/2 x [1- (0.633)2]-1/2
= 3.034 x 10-4 m3/s = 18.20 liter/min.
Percentage error = (18.20-15.00)/15 x 100
= 9.0%
For the flow 10 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(18.5 – 16.5)]1/2 x [1- (0.633)2]-1/2
= 1.324 x 10-4 m3/s = 7.946 liter/min.
Percentage error = (7.946-10)/10 x 100
= 20.54%
21
Calculations for Table A2.1 and A2.2
For the venturi,
For the flow 20 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (20.5 – 9.2)(9.81) ]0.5
= 2.548 x 10-4 m3/s = 15.29 liter/min.
Percentage error = (15.29-20.00)/20 x 100
= 23.55 %
For the flow 15 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (21.7 – 15.5)(9.81) ]0.5
= 2.535 x 10-4 m3/s = 15.21 liter/min.
Percentage error = (15.21-15.00)/15 x 100
= 1.4 %
For the flow 10 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (22.2 – 19.2)(9.81) ]0.5
= 1.764 x 10-4 m3/s = 10.58 liter/min.
Percentage error = (10.58 – 10.00)/10 x 100
= 5.8 %
22
For the orifice,
For the flow 20 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(19.2-13.2)]1/2 x [1- (0.633)2]-1/2
= 2.25 x 10-4 m3/s = 13.50 liter/min.
Percentage error = (13.50-20.00)/10 x 100
= 32.5 %
For the flow 15 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(20.8-17.7)]1/2 x [1- (0.633)2]-1/2
= 1.485 x 10-4 m3/s = 8.91 liter/min.
Percentage error = (8.9-15.00)/15 x 100
= 40.67 %
For the flow 10 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(21.8-20.3]1/2 x [1- (0.633)2]-1/2
= 1.324 x 10-4 m3/s = 7.946liter/min.
Percentage error = (7.946-10)/10 x 100
= 20.54 %
23
Calculations for table A3.1 and A3.2,
For the venturi,
For the flow 20 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (25.1-13.9)(9.81) ]0.5
= 3.407 x 10-4 m3/s = 20.44 liter/min.
Percentage error = (20.44-20.00)/20 x 100
= 2.2 %
For the flow 15 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (25.9-19.5)(9.81) ]0.5
= .2.23 x 10-4 m3/s = 13.38 liter/min.
Percentage error = (13.38-15.00)/15 x 100
= 10.8 %
For the flow 10 L/min.,
Q = 0.97 x 2.011 x 10-4 [ 1-(0.53)2 ]-0.5 x [ 2 (26.2-23.4)(9.81) ]0.5
= 1.704 m3/s = 10.22 liter/min.
Percentage error = (10.22 – 10.00)/10 x 100
= 17.5%
For the orifice,
For the flow 20 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(23.9-17.8)]1/2 x [1- (0.633)2]-1/2
= 2.232 x 10-4 m3/s = 13.39 liter/min.
24
Percentage error = (13.39-20.00)/10 x 100
= 33.05
For the flow 15 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(25.1-21.8)]1/2 x [1- (0.633)2]-1/2
= 3.034 x 10-4 m3/s = 18.20 liter/min.
Percentage error = (18.2-15.00)/15 x 100
= 21.33 %
For the flow 10 L/min.,
Q = 0.61 x 3.1 x 10-4 x [2(9.81)(22.1-20.6]1/2 x [1- (0.633)2]-1/2
= 1.764 x 10-4 m3/s = 10.58liter/min.
Percentage error = (10.58-10)/10 x 100
= 5.8 %
25
CALCULATION FOR HEAD LOSSES OF VENTURI METER,
Hventuri = V2 2 / 2g [ A2 / At - 1]2 ------------ (1)
Hpipe losses = K x V2 2 / 2g -------------(2)
Compare equation (1) and (2), we have
K = [ A2 / At - 1]2
= (3.8 x 10-4 - 1 )2 = 0.22
2.011 x 10-4
Calculations for Table A1.3
V@20L/min = Q/At Hventuri = K x V2 2 / 2g
= 3.049 x 10-4 / 2.011 x 10-4 = 0.22 x (1.525)2 / 2 x 9.81
= 1.525m/s = 0.824 x 10-2 m
V@15L/min = Q/At Hventuri = K x V2 2 / 2g
= 2.694 x 10-4 / 2.011 x 10-4 = 0.22 x (1.34)2 / 2 x 9.81
= 1.34 m/s = 0.64 x 10-2 m
V@10L/min = Q/At Hventuri = K x V2 2 / 2g
= 1.793 x 10-4 / 2.011 x 10-4 = 0.22 x (0.89)2 / 2 x 9.81
= 0.89 m/s = 0.28 x 10-2 m
26
Calculations for Table A2.3
V@20L/min = Q/At Hventuri = K x V2 2 / 2g
= 2.548 x 10-4 / 2.011 x 10-4 = 0.22 x (1.267)2 / 2 x 9.81
= 1.267m/s = 0.6 x 10-2 m
V@15L/min = Q/At Hventuri = K x V2 2 / 2g
= 2.535 x 10-4 / 2.011 x 10-4 = 0.22 x (1.261)2 / 2 x 9.81
= 1.261 m/s = 0.57 x 10-2 m
V@10L/min = Q/At Hventuri = K x V2 2 / 2g
= 1.764 x 10-4 / 2.011 x 10-4 = 0.22 x (0.88)2 / 2 x 9.81
= 0.88m/s = 0.28 x 10-2 m
Calculations for Table A3.3
V@20L/min = Q/At Hventuri = K x V2 2 / 2g
= 3.407 x 10-4 / 2.011 x 10-4 = 0.22 x (1.694)2 / 2 x 9.81
= 1.694m/s = 1 x 10-2 m
V@15L/min = Q/At Hventuri = K x V2 2 / 2g
= 2.23 x 10-4 / 2.011 x 10-4 = 0.22 x (2.23)2 / 2 x 9.81
= 1.109 m/s = 0.44 x 10-2 m
V@10L/min = Q/At Hventuri = K x V2 2 / 2g
= 1.704 x 10-4 / 2.011 x 10-4 = 0.22 x (0.974)2 / 2 x 9.81
= 0.974m/s = 0.38 x 10-2 m
27
CALCULATION OF HEAD LOSSES FOR ORIFICE,
For orifice, there are 2 losses, exit loss and entry loss.
For entry loss, assume that the entry is sharp- edged.
Thus, H1 = 0.5 V2 2 / 2g
For exit loss, H2 = V2 2 / 2g [ A2 / At - 1]2
K = [ A2 / At - 1]2
= ( 3.1 x 10-4 / 4.9 x 10-4 – 1 )
= 0.13
Calculations for Table A1.2,
V@20L/min = Q/A2 H = 0.5 V2 2 / 2g
= 2.649 x 10-4 / 3.1 x 10-4 = 0.5 x (0.54)2 / 2 x 9.81
= 0.54 m/s = 0.0074 m
V@15L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 3.034 x 10-4 / 3.1 x 10-4 = 0.5 x (0.62)2 / 2 x 9.81
= 0.62 m/s = 0.98 x 10-2 m
V@10L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 1.324 x 10-4 / 3.1 x 10-4 = 0.5 x (0.27)2 / 2 x 9.81
= 0.27 m/s = 0.19 x 10-2 m
Calculations for Table A2.3
V@20L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 2.25 x 10-4 / 3.1 x 10-4 = 0.5 x (0.46)2 / 2 x 9.81
= 0.46 m/s = 0.0054 m
28
V@15L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 1.485 x 10-4 / 3.1 x 10-4 = 0.5 x (0.30)2 / 2 x 9.81
= 0.30 m/s = 0.23 x 10-2 m
V@10L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 1.324 x 10-4 / 3.1 x 10-4 = 0.5 x (0.27)2 / 2 x 9.81
= 0.27 m/s = 0.19 x 10-2 m
Calculations for Table A2.3
V@20L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 2.23 x 10-4 / 3.1 x 10-4 = 0.5 x (0.46)2 / 2 x 9.81
= 0.46 m/s = 0.0054 m
V@15L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 3.034 x 10-4 / 3.1 x 10-4 = 0.5 x (0.62)2 / 2 x 9.81
= 0.62 m/s = 0.98 x 10-2 m
V@10L/min = Q/A2 H1 = 0.5 V2 2 / 2g
= 1.324 x 10-4 / 3.1 x 10-4 = 0.5 x (0.27)2 / 2 x 9.81
= 0.27 m/s = 0.19 x 10-2 m
29
APPENDIX C
Figure C1 : Installation drawing for Flowmeter Demonstration Apparatus
30
Figure C2 : Graph of Orifice meter discharge coefficient
31
Figure C3 : Venturi meter discharge coefficient
32