Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 1
Lab Energy Management Big Ten Mechanical & Energy Conference
October 2014
1
The Approach• Promotion and awareness of Campus Energy Transition Plan
Goals • Be Spartan Green, Face It: Green Looks Good on You, radio, newspaper,
athletic events, fairs, promotional events, website, Twitter, Facebook
• Communication of progress• Institutional reporting, rankings, building reports,
• Behavior Change• Commissioning Energy Treasure Hunt • Environmental Steward Program• Green Certification Program
2
• Energy Educators
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 2
Safe Sustainable Labs – Cross Unit Team
Balancing Laboratory Safety and Energy Efficiency
• MSU’s Energy Transition Plan• Why Focus on Laboratories? • The “Safe Sustainable Labs” Concept
3
Energy Transition Plan Goals
% Greenhouse Gas Emission Reduction
% Campus Renewable Energy
FY2015 30 15
FY2020 45 20
FY2025 55 25
FY2030 65 40
4
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 3
Progress Towards Energy Transition Plan Goals
5
585,867
541,565
506,413486,486 477,090
0
100,000
200,000
300,000
400,000
500,000
600,000
700,000
Basis (FY2010) FY2011 FY2012 FY2013 FY2014
Me
tric
to
ns
of
CO
2Michigan State University East Lansing Campus
Total Scope 1 and Scope 2 Greenhouse Gas Emissions
↓ 18.6%
Safe Sustainable Labs: MSU’s Energy Transition Plan
The Ultimate Vision:Campus Powered by 100% Renewable Energy
6
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 4
Safe Sustainable Labs: MSU’s Energy Transition Plan
• Investment in energy conservation measures• More stringent energy-related building standards• Energy monitoring strategies
Goal 1: Improve the Physical Environment
7
Safe Sustainable Labs: MSU’s Energy Transition Plan
Capacity
Reliability
Environment
Health
Cost
Why is IPF installing the OptiNet system?
MSU’s Energy Transition Plan calls for aggressive investment in energy conservation measures and new technologies.
Improvements in building systems like this will allow the University to achieve a 20% reduction in energy consumption by the year 2020.
8
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 5
Why Focus on Laboratories?
A research building typically uses 5 to 10 times more energy than an office building.
Appropriate ventilation control can result in a 30 to 80% reduction in energy use.
9
The Safe Sustainable Labs ConceptAn integrated approach to energy reduction
in intensive research spaces
Emergency Purge
Button
Energy Efficient
Equipment
Centralized
Demand Control
Ventilation
Fume Hoods
Molecular Plant Science
10
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 6
Working together for safety
Balancing Laboratory Safety and Energy Efficiency
• Conventional Laboratory Ventilation
• OptiNet: Centralized Demand Controlled Ventilation
• Response to Elevated VOCs
• OptiNet Monitoring
• Lab Assessment – Derive Number of Air Changes per Hour
• Locations of OptiNet at MSU
11
How Lab HVAC Works: Variable Air Volume System
A Air VolumeL Lab LlAB Lb
12
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 7
Centralized Demand Controlled Ventilation
Integration of airflow systems with sophisticated occupancy and air-quality sensors.
• Real time air quality monitoring
• Sampled air analyzed for contaminants Sensor Suite
• Responsive Ventilation ACH setback based on occupancy ACH increase in response to
contaminants Varies from 4 ACH to 18 ACH
• Emergency Purge button13
How HVAC Works in a Variable Air Volume Labwith Centralized Demand Control Ventilation
Variable volume
control exhaust boxes
Supply VAV boxes for
temperature control
14
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 8
Centralized Demand Controlled Ventilation
TEMP
15
DASHBOARD for DOLLARS
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 9
Green = Good
Red = Bad
Frequency of Hood
Use
Average % Open
• OptiNet NOT Suitable for some Labs• High Solvent Use and Storage• Cryogen Use in Smaller Space• Biosafety Level 3• Volatile Carcinogen Storage• High Acute Toxicity by Inhalation• High Hood Density• Derived Number of Air Changes per Hour
Environmental Health & Safety Lab Assessments for OptiNet
18
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 10
•Food Science*•Biochemistry*•Molecular Plant Sciences•Anthony Hall* •Bio-Engineering (new building)
* Select Labs Only
Locations of OptiNet at MSU
19
Responsibilities of Lab Personnel
Balancing Laboratory Safety and Energy Efficiency
• Emergency Room Purge
• Room Purge or Hood Purge?
• Safe Sustainable Lab Practices
20
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 11
21
When there is a large chemical spill or hazardous release in the room…
•Press Room Purge Button•Exit Laboratory•Close Lab Doors•Call 911
Emergency Purge Ventilation Procedure
22
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 12
Activate the red emergency room purge button when there is a spill or hazardous release in the room.
Emergency Room Purge or Hood Purge?
Use hood purge button located on the fume hood monitor when there is a spill or release in the hood.
23
Maintain clear area around Emergency Room Purge button
Safe Sustainable Lab Practices
24
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 13
Keep the sash closed as much as possible• Avoid raising sash above the arrow • Lower or close hood when not in active use. • Leaving the lab area? Close sash completely.
Safe Sustainable Lab Practices
25
Minimize openings on hoods with horizontal panes
Position the sash all the way down and slide panes to create as small an open area as practical.
Safe Sustainable Lab Practices
26
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 14
Keep doors and windows closed •This allows correct lab pressurization to be maintained. •Open doors or windows can impair air flow through the hood
Work with volatile carcinogens and toxics in hood
Proper use and storage of chemical
Safe Sustainable Lab Practices
27
Promotion & Awareness
Be Spartan Green Campaign• Website/Green Stories/Projects/Events/Community• Content driven by behavior team research & recommendations
• Energy Educators – How does my building work?
28
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 15
Communicating ProgressGoals and Feedback Critical for Campus Communications
• Key Indicators Report• Project progress• Energy Dashboard
29
Behavior Change – Green Certification
• Recognize, promote and assist departments/students with MSU best practices
• Certification criteria based on space type:o Offices and conference roomso Residential facilitieso Kitchenso Science Labso Information Technology Areas/Computer Labs
• Qualitative and quantitative checkpoints
30
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 16
Funding for Lab Equipment Maintenance
31
32
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 17
33
34
12.713.4
14.0 14.2 14.2 14.315.2 15.4
17.4 17.618.6
21.5
23.4
25.426.5
0.0
5.0
10.0
15.0
20.0
25.0
30.0
Elec
tric
An
nu
al C
on
sum
pti
on
(kw
hr/
GSF
)
Big Ten and Friends Utilities BenchmarkingElectrical Consumpton FY 2012-13
Lab Energy Management Michigan State University
Lynda Boomer, Michigan State University
Todd Wilson, Michigan State University 18
&Questions