1 Specification and Technical Data Sheet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in multiple chamber configurations for your application. Designed for simplicity in operation and serviceability, these models are ideal for use in research laboratories, bio-containment environments, and animal care facilities. Large Steam Sterilizers Vessel design features a stainless steel, fully-jacketed 316L chamber. The vessel is insulated and mounted on a steel frame, which offers adjustable feet on self-centering floor pads. Horizontally Sliding Doors are energy-efficient, safe and operated by simple touch screen activation. Rectangular chamber eliminates wasted space and reduces high utility costs common to cylindrical or elliptical designs. All chambers are polished to a mirror finish of <10 Ra. Surface finish can be just as critical in determining the corrosion resistance of austenitic stainless steel as the grade. Poor quality finishes can lead to disappointing performance of stainless steel. A highly polished surface will give the best performance in any specified environment. Non-proprietary parts are a hallmark of PRIMUS Steam Sterilizers allowing for immediate diagnostic and replacement of worn components. Gravity, vacuum, and liquid cycles come standard on all models. Multiple test cycles are included for process challenge. Low temperature cycles and effluent decontamination are available for specific applications. Configurable controls are adaptable to meet a variety of applications. Water conservation is available with our PRI-Saver® TRVPS system that offers up to 90% water savings. Predictive maintenance functionality included in our PLC based controls allows for increased uptime through the proactive monitoring of critical components. Ease of service is built into the design and delivered by PRIMUS Authorized Service Agents. Features FOR GENERAL PURPOSE APPLICATIONS CONTENTS Page Features 1 Standards 2-3 Control Systems 4-5 Configuration Options 6-7 Drawings & Dimensions 8-17 Configuration Worksheet 18-19 Intelligent design focuses on ease of use, simplified diagnostics, and clear service access for maximum uptime.

Upload: others

Post on 05-Mar-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

1

Specification and Technical Data Sheet

Form 2076 -3.10 1119 LG

PRIMUS® large steam sterilizers come in multiple chamber configurations for your application. Designed for simplicity in operation and serviceability, these models are ideal for use in research laboratories, bio-containment environments, and animal care facilities.

Large Steam Sterilizers

• Vessel design features a stainless steel, fully-jacketed 316L chamber. The vessel is insulated and mounted on a steel frame, which offers adjustable feet on self-centering floor pads.

• Horizontally Sliding Doors are energy-efficient, safe and operated by simple touch screen activation.

• Rectangular chamber eliminates wasted space and reduces high utility costs common to cylindrical or elliptical designs.

• All chambers are polished to a mirror finish of <10 Ra. Surface finish can be just as critical in determining the corrosion resistance of austenitic stainless steel as the grade. Poor quality finishes can lead to disappointing performance of stainless steel. A highly polished surface will give the best performance in any specified environment.

• Non-proprietary parts are a hallmark of PRIMUS Steam Sterilizers allowing for immediate diagnostic and replacement of worn components.

• Gravity, vacuum, and liquid cycles come standard on all models. Multiple test cycles are included for process challenge. Low temperature cycles and effluent decontamination are available for specific applications.

• Configurable controls are adaptable to meet a variety of applications.

• Water conservation is available with our PRI-Saver® TRVPS system that offers up to 90% water savings.

• Predictive maintenance functionality included in our PLC based controls allows for increased uptime through the proactive monitoring of critical components.

• Ease of service is built into the design and delivered by PRIMUS Authorized Service Agents.

Features

FOR GENERAL PURPOSE APPLICATIONS

CONTENTS PageFeatures 1Standards 2-3Control Systems 4-5Configuration Options 6-7Drawings & Dimensions 8-17Configuration Worksheet 18-19

Intelligent design focuses on ease of use, simplified diagnostics, and clear service access for maximum uptime.

Page 2: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

2Form 2076 -3.10 1119 LG

Specifications

closed and sealed. When sterilizer cycle is complete, the seal retracts by vacuum. The door cannot be opened while a cycle is in progress and the door will not unseal while the chamber is under pressure or vacuum.

Door interlocks on double door sterilizers are programmed to prevent inadvertent opening of door(s). An access key is provided to override door interlocks.

Chamber Drain System

Drain system is designed to prevent pollutants from entering into the water-supply system and sterilizer.

Drain Water Quench

The piping system provides automatic condensing of chamber steam and discharge to the floor drain. Cooling water is added to ensure discharge temperature is below 60°C (140°F). A separate temperature switch is included to regulate the volume of water so as not to exceed the required amount necessary to achieve target temperature.

Vacuum System

Chamber pressure is reduced during the conditioning phase and drying phase through the means of either a standard water ejector or a liquid ring vacuum pump. Following the drying phase, the chamber is returned to atmospheric pressure by admitting air through a 0.3 micron bacteria-retentive filter.

Steam Source

Sterilizers are piped, valved, and trapped to receive building- supplied steam delivered at 50 to 80 psig (344.7 to 551.6 kPa) dynamic. If building steam source is not available, an electric carbon-steel steam generator or electric stainless steel steam generator may be provided to supply steam to the sterilizer. Steam piping is constructed of brass and copper or stainless steel and includes a steam strainer and brass pressure regulator.

Steam to Steam

For applications requiring clean or pure steam, a Steam to Steam Generator is available. The quality of the steam produced will match the quality of the feedwater. Pure Steam¨ SCS1 The stainless steel Steam to Steam Generator with sanitary

fittings comes with double tube sheet construction. Generated from WFI quality water, the water source to the Steam to Steam Generator is provided by others.

Clean Steam¨ SCS2 The stainless steel Steam to Steam Generator with

threaded connections comes with single tube construction. Generated from distilled and/or reverse osmosis water, the Clean Steam sterilizers normally include stainless steel piping for all wetted surfaces in the process loop.

Piping

All piping connections terminate within the confines of the sterilizer and are accessible from the front and side of sterilizer.

Solenoid valves with DIN connectors are arranged for easy removability and can be serviced individually.

Jacket Assembly

A Type 316L stainless steel chamber and a Type 304 stainless steel jacket are welded together to form the sterilizer vessel. Type 316L stainless steel end frame(s) is welded to door end. On single door units, back of chamber is fitted with welded, 316L stainless steel formed head.

Sterilizer vessel is ASME rated at 45 psig (3.06 Bar) and insulated. The Vessel includes one 1.5” (38 mm) inch validation port for customer use.

Steam-supply opening inside the chamber is shielded by a Type 316L stainless steel baffle.

The unique design of the chamber jacket allows for even distribution of heat.

Chamber Finish

The brilliant PRI-Mirror chamber finish found in all PRIMUS models sets the highest standard for surface finish, achieving a 10 Ra (0.026 micron) measurement.

“Surface finish can be just as critical in determining the corrosion resistance of austenitic stainless steel as the grade [58]. Poor quality finishes can lead to disappointing performance of stainless steel and a bright polished surface will usually give the best performance in any specified environment. 1 micron = 39.4 microinch Ra.

Coarse polished finishes with surface roughness values greater than 1 micron have been shown to contain deep grooves where chloride ions can accumulate and destroy the passive film, thereby initiating corrosion attack. In contrast, fine polished finishes with surface roughness values less than 0.5 micron will generally have fewer sites where chloride ions can accumulate” (Parrott and Pitts, 2011).1

Chamber door

Door is constructed of a single formed piece of Type 316L stainless steel. Door is insulated to reduce the surface temperature of the stainless steel door cover.

During cycle operation, door is sealed by a steam-activated door seal. Door seal is constructed of an easily replaceable silicone gasket located in a channel groove in the chamber end ring. To ensure safety, a cycle cannot be started until the door is fully

Construction

Each sterilizer meets applicable requirements of the following listings and standards, and carries the appropriate symbols.• ASME Code, Section VIII, Division 1 for unfired pressure vessels.

The pressure vessel is so stamped; ASME Form U-1 is furnished. Shell and door are constructed to withstand working pressure of 45 psig (310.2kPa)

• ASME Code, Section I, Part PMB for power boilers, if optional steam generator is supplied.

• UL/ICE/CSA61010-1 SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL, AND LABORATORY USE -

• UL/ICE/CSA61010-2-040 SAFETY REQUIREMENTS FOR ELECTRICAL EQUIPMENT FOR MEASUREMENT, CONTROL AND LABORATORY USE PART 2-040: PARTICULAR REQUIREMENTS FOR STERILIZERS AND WASHER-DISINFECTORS USED TO TREAT MEDICAL MATERIALS

• UL 508A Standard for Industrial Control Panels

Standards

Page 3: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

3

Specification and Technical Data Sheet

Specifications, Cont.

Cycle Descriptions

Standard and Optional Cycles:

Prevacuum Cycle (standard) for efficient sterilization of porous, heat-and moisture-stable materials at 100°C to 135°C (212°F to 275°F). Prevacuum cycle utilizes a mechanical air-evacuation system.

Gravity Cycle (standard) for sterilization of heat- and moisture-stable goods at 100°C to 138°C (212°F to 280°F). Gravity cycle utilizes gravity air-displacement principle. Pre-positive pulses for enhanced air removal are available in hard goods applications.

Liquid Cycle (standard) for sterilization of liquids and media in vented borosilicate glass or metal containers at 100°C to 138°C (212°F to 280°F). Liquid cycle uses enhanced solution cooling during exhaust (cooling) phase to control exhaust rate.

Waste Cycle (standard) for processing of laboratory waste. Parameters may need to be adjusted based on specific loads. Bags should not be completely sealed.

Lab Low (Isothermal) Cycle (optional) is for processing heat-sensitive and heat-coagulable solutions in vented borosilicate glass or metal containers at 100°C (212°F). Lab Low utilizes steam to enhance temperature control and prevents layering of steam and air within the chamber. Process maintains positive pressure in chamber to inhibit boiling. Temperature uniformity ±5°C.

Jacket Cooling (Optional) improves exhaust time for liquid loads. Thermostatically controlled water is introduced gradually and in combination with air pressure to control load condition. This process dramatically reduces cooling time.

Air Over Cooling (Optional) provides air pressure to chamber during the exhaust phase to maintain pressure until load temperture is reduced to a set point.

Effluent Decontamination Cycle (Optional) for decontamination of biohazardous waste (BSL-3 and BSL-4 environments). Condensate is held in the chamber and decontaminated before discharge to floor drain. Steam is introduced through bottom of sterilizer chamber, and chamber is exhausted out top side of vessel. During purge and vacuum pulses, all purge and exhaust gases are vented through a 0.2 micron bacterial retentive filter. Optionally, filter housing can be steam jacketed to ensure filter integrity.

Leak Test Cycle (standard) for verification of door seal and piping system integrity.

Bowie-Dick Test is available for 121°C (250°F) and 132°C (270°F) prevacuum cycles.

PRI-Egreen

PRI-Egreen is a standard feature that will shut off utilities to the jacket after the unit has sat idle for a specified period of time. Time is programmable and secured via the PLC in order to meet facility sustainability goals.

PRI-Egreen +

Shutdown may be programmed to activate at the end of any designated cycle or time of day. When activated, control system automatically shuts off all utility valves, conserving steam and water usage. Sterilizer utilities can be restarted either by programmed time or manual operation. A different shutdown and restart time can be programmed for each day.

Drain Cooling

Drain quench is a standard feature on all PRIMUS models. Discharge temperature below 140F.

Sustainability

Chamber condensate alarm activates alarm state if excessive condensate is detected in the vessel chamber drain.

Door interlocks (double door units only) allow only one door to be opened at a time and, during processing, prevent the unload side door from being opened until a satisfactory cycle is complete. If a cycle is aborted, the unload side door cannot be opened.

Pressure relief valve limits the amount of pressure buildup so that the rated pressure in the vessel is not exceeded.

Door and gasket safety switch signals when door seal is energized and tight against the door. Software prevents cycle from starting until the limit switch signal is received. If control loses appropriate signal during cycle, an alarm state is activated.

Emergency stop button (located on front of the sterilizer) is included on all sterilizers with PLC control systems.

Safety Features

1Parrott, R., BSc PhD MIMMM CEng, & Pitts, H., MEng PhD. (2011). Chloride stress corrosion cracking in austenitic stainless steel [PDF]. Heallth and Safety Exclusive.

Page 4: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

4Form 2076 -3.10 1119 LG

Specifications

Cutting edge and configurable, PRIMUS’ PRI-Matic® series is a PLC based control platform designed to exceed the needs of any general purpose, research laboratory, containment or vivarium application. These platforms feature industry leading, non-proprietarty components and pharmaceutical grade coding.

Standard displays ranging from 5.7” to 12.1” provide a full color, touch-sensitive screen. Standard are thirty programmable cycles, adjustable to meet specific processing requirements. Seven cycles come pre-programmed for ease of use. All control configurations are performed through the touch screen display.

Cycle values and operating features may be adjusted and verified prior to cycle operation. User Access, Profiles, Simplified Screens, Interactive P&ID (on certain models) and additional options can be configured or toggled on/off easily in the user-friendly menus.

Critical control system components are housed within a NEMA-rated sealed compartment which protects components from moisture and heat generated during sterilization process.

Operator interface control panel, consisting of touch screen and thermal printer, is located on load or non-sterile end of sterilizer. If the sterilizer is equipped with double doors, an additional touch screen is provided on the unload or sterile end.

• Touch-Sensitive Screen provides users with color, touch-sensitive displays featuring high definition TFT LCD displays. Screen sizes range from 5.7” to 12.1” diagonally with a resolution of 800x600 pixels, a selection with the largest HMI’s (Human Machine Interface) in the market. These screens provide clear, sharp, and bright displays, even in environments with low light, by utilizing the 65,536 available colors. They deliver realistic images and the brightest displays (PRI-Matic 100).

• Thermal Printer located below the touch screen, provides an easy-to-read printed record of all pertinent cycle data on 2-1/4” wide paper. Data is automatically printed at the beginning and end of each cycle and at transition points during the cycle. A duplicate print can be obtained of the last cycle run. Additional print options include Ink-On-Paper Impact Printer, Ethernet printing and PRI-SND. PRI-SND system (Secured Network Device) stores cycle data in a .pdf format that can be accessed over a network.

PLC Control Systems - A Variety of Choices

The time-tested and reliable PRIMUS PSS5 Microcomputer Control is an industrial microprocessor providing accuracy and automation for all customer requirements. This simple, versatile control has become the first choice for both sterilizer users and service technicians.

The PRIMUS PSS5 control system provides for automatic operation through all phases of the sterilization cycle.

Cycle parameters can be selected and programmed by the operator. All cycle phases are monitored visually. Cycle completion indicators are provided both audibly and visually.

The control is an embedded Microcomputer system with non-volatile memory storage eliminating the need for battery back up. The computer consists of solid-state devices, isolated from heat and moisture in a NEMA-rated, shielded enclosure. The Microcomputer control is self-diagnostic and provides information to the operator on a 2-line, 16 character LCD located on the control touchpad. An audible tone accompanies all alarms indicating a problem with the cycle. When specified, cycle parameters may be locked out by the supervisor through the use of touchpad keystrokes.

The control touchpad provides information on cycle progress by LED indicators. Specific information on cycle parameters and progress, in addition to alarms, are shown on the LCD display. Cycle selection and programming are entered by control panel pushbuttons, which provide an audible signal when pressed. Controls include a cycle advance button to allow the operator to manually advance the cycle. The control panel indicates the use of cycle advance and which phases were advance. It continually displays this information until the door is opened at the end of the cycle. Cycle progress including time, temperature, pressure, and alarms are reported on a 24-character thermal or impact printer.

The touchpad contains pushbuttons and the LCD display. Cycle progress is displayed by LEDs on the Operator’s Panel. Inputs are made using the numbered cycle buttons (See Part III, User’s Manual, Figure 2.1-1, Touchpad Display). Double door sterilizers have a touchpad near each door.

Touchpad Control Systems

Page 5: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

5

Specification and Technical Data Sheet

Specifications

Thermal printer take-up spool stores an entire roll of paper, providing cycle records which can be saved for future reference. Three paper tape rolls are furnished with each unit.

• Unload side control panel (equipped on double-door sterilizers only) includes a touch-sensitive screen identical to the operating end screen. Preprogrammed cycles can be started from the unload side control panel. Display concurrently shows the same information as the load side screen display.

• Cycle configuration is performed by accessing the system menu on the touch screen display after authentication. In addition to adjustment of cycle values, the following operating parameters can also be changed through the change values menu:

• Time Display and Printout Units in standard AM/PM or 24-hour military (MIL) time.

• Selectable Cycle Name permits user to name each cycle with any combination of letters, numbers, blank spaces, and underscores.

• Print Interval permits time period adjustmentsv between cycle-status printouts generated during the cycles phases.

• User Access settings permit adjustment of access and security of up to 14 users on PRI-Matic 100 and up to 30 users on PRI-Matic 200.

• Security access code is required to enter the administrator mode (changing values), and service mode. Servicing the sterilizer or accessing change values menu causes display to request the entry of an access code. If access code is not properly entered, display returns to the standby screen, denying user access to the sterilizer or programming. Access to the sterilizer can be limited to 30 operators, each with a different access code.

• Alarms pulses red flags on and off during alarm conditions. Informational text is displayed to guide the user in resolving the alarm. A buzzer that sounds during an alarm and at a different rate for five seconds at cycle completion can be configured manually.

• Temperature Display and Printout Units in Celsius (°C) or Fahrenheit (°F). Temperature is set, displayed, and printed to the nearest 0.1°. Recalibration is not required when changing temperature units from °C to °F and vice versa.

PLC Control Systems, cont.• SD Card is provided for downloading cycle information to a

Customer-furnished Excel spreadsheet file. Up to six months of (one hour) cycles can be stored on the included flash card before card has to be downloaded to PC.

• Battery Backed Memory backs up all cycle memory. In the event of a power failure, the cycle is stopped and cycle data is recorded up to that phase. Once power is restored the system goes into an alarm state indicating there was a power failure. At that point the user must acknowledge the alarm and can then either resume or abort the cycle regardless of current phase.

Predictive Maintenance

Predictive maintenance features allow users and service technicians to monitor the life cycles of major autoclave parts. Monitoring and tracking life cycles of major autoclave components enables easier budgeting, minimizes downtime and increases productivity.

Real-Time Trend Graph Data

Automatically tracks vital temperature and pressure information. Shows process values for chamber, jacket, and load probe.

Step Detail Screen

Displays a description of the current cycle phase step, the conditions needed to advance the step, and any timeout conditions that apply. It is ideal for troubleshooting and acts as a training tool for new users.

User Authorization Levels

Four levels of authorization come standard with increasing varying access permissions. Standard levels include default, operator, technician and administrator. Additional levels can be custom configured.

Page 6: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

6Form 2076 -3.10 1119 LG

Specifications, cont.

Standard Features

Resistance Temperature Detectors (RTDs) are standard for sterilizer temperature control. The chamber drain line RTD monitors and controls temperature variations within the sterilizer chamber. A jacket RTD provides temperature control within the jacket.

Software Calibration is provided for all temperature and pressure inputs. Calibration is available in the service mode and is accessible through the touch screen displays, and is performed using external or internal temperature and pressure sources. Control system provides a printed record of calibration data for verification to current readings.

Cycle Data Records are recorded on the printer tape and can be saved to a customer provided SD card. Data can also be retrieved for on-screen review or sent via e-mail if the system is enabled. Network connection required.

Automatic Steam Shutoff to Jacket is provided for Lab Low and liquid cycles. When activated for Lab Low cycles, the jacket control conducts a timed jacket drain, automatically allowing the operation of cycles at lower temperatures. When activated for liquid cycles, steam supply to the jacket is turned off during exhaust phase, allowing load to cool efficiently.

Insulation, one-inch thick, asbestos and chloride-free fiberglass completely encases the exterior of the sterilizer vessel and is sealed in an aluminum external cover.

Air Backup to door seals is optional on all double door sterilizers, with either bioseal or air differential seals.

Stainless Steel Piping to Chamber delivers steam generated from clean steam source to the chamber and its contents. All steam-to-chamber piping components are constructed of 300 series stainless steel.

Stainless Steel Clean Steam Generator automatically produces clean steam using customer-supplied steam and purified water. Generator is integrally connected to the clean steam-to-chamber piping system.

RTD Load Probes and F0 Sterilization automatically sense the load temperature during cycle operation. A single thermal load probe is sealed through the sterilizer vessel and manually placed in the product container within the chamber prior to cycle operation.

In conjunction with the load probe option, individual cycles can be set to start sterilization phase according to chamber drain temperature or according to load temperature. Also, F0 set points are available for each cycle, allowing for sterilization phase termination based on the calculated F0 value.

Compressed Air to Gasket utilizes compressed air to actuate gasket in lieu of steam.

Options

Bioseal is a 1/4” stainless steel plate which is welded to the chamber and a 1/4” thick silicone gasket that extends between the plate and a stainless steel wall frame which is welded to wall imbeds. The bioseal is provided on the unload side of the sterilizer and prevents passage of airborne microorganisms from the space between the vessel body and the structural wall opening. Steam is the primary source of pressure behind the door seal. All sterilizers with bioseals have air back-up to maintain seal pressure when out of cycle or if the steam source is not available.

Air-Differential Seal (double door units only), provided on the unload side of the sterilizer, minimizes airflow between the dirty and clean sides of the barrier.

Back Cabinet Panel is provided on single door, freestanding units where the unit is accessible on all sides.

Vacuum Pump upgrade is offered in lieu of the standard water ejector vacuum system. This pump increases vacuum rate and decreases vacuum cycle times on larger chambers. This option is standard on units over 100 cu. ft.

0.2 Micron Bacterial Retentive Filter provides sterile air during airbreak at end of cycle.

Additional Chamber Penetration: capped chamber penetration port is located at the side of the vessel preventing interference with other piping. The port provides for thermocouple probes or other test instrumentation.

Form C Dry Contacts provide four relays to communicate equipment status. Selectable statuses include: E-stop power, in cycle, cycle complete, cycle compromised, alarm state, and sterilizer power state.

Piping Options:

Brass and copper piping are standard on most sterilizers being the lowest cost option utilizing off the shelf components.

Stainless steel threaded- generally used with clean steam and offers sufficient corrosion resistance.

Stainless steel sanitary- for pure steam application highest level of cleanliness as they limit entrapment areas where bacteria could form or harbor. They are also corrosion resistant when used with purified water.

Air Compressor, Portable, 115 Vac. is intended for pneumatic valves on sterilizers when an air utility is not provided by the facility. It may also be used as a back-up pressure source for the door seal in bioseal applications.

This portable 2.35 gallon compressor tank that delivers 150 LPM @ 345 KPa (5.3CFM @ 100 PSI).

Seismic tie-down kit conforms to current California Code of Regulations.

Accessories

Page 7: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

7

Specification and Technical Data Sheet

Specifications, cont.

Sterilizers are arranged for either freestanding or pit mounting, as specified. Each sterilizer is equipped with adjustable feet on self-centering floor pads.

On freestanding units, stainless steel cabinet side panels enclose the sterilizer body and piping. Front fascia is included if recessed.

Mounting Arrangement

Our highly trained service specialists provide periodic inspections and adjustments to ensure low-cost, peak performance. PRIMUS representatives can provide information regarding annual maintenance agreements.

Preventative Maintenance

All PRIMUS steam sterilizers are proudly designed and manufactured in the USA. Each unit is constructed of solid stainless steel and built in our quality controlled ASME facility using non-proprietary parts. Constructing our sterilizers in the USA ensures the durability and excellence of each unit we produce. Our employees are important to us and by keeping manufacturing local we are able to ensure high labor standards and employee satisfaction. Satisfied employees make for high quality products which we are able to pass on to our clients.

Our dedication to using non-proprietary parts stems from our commitment to meeting the needs of our customers. Non-proprietary parts ship much faster and we are usually able to deliver within two days.

American Made

Page 8: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

8Form 2076 -3.10 1119 LG

TDS Drawings

Page 9: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

9

Specification and Technical Data Sheet

MODEL N O P Q

Chamber Size(W x H x L)

Inches/Millimeters

35 x 57 x 86889 x 1448 x 2184

49 x 57 x 861245 x 1448 x 2184

49 x 86 x 861245 x 2184 x 2184

61 x 86 x 861549 x 2184 x 2184

Chamber Capacity 99 ft3 / 2812 L 139 ft3 / 3936 L 210 ft3 / 5939 L 261 ft3 / 7393 L

Overall Width 52.31 / 1329 66.31 / 1684 73.81 / 1875 85.81 / 2180

Overall Height1 96.00 / 2438 96.00 / 2438 137.00 / 3480 137.00 / 3480

Overall Length (SD)2,3 101.50 / 2578 101.50 / 2578 101.50 / 2578 101.50 / 2578

Wall Opening Width4 105.00 / 2667 133.00 / 3378 138.00 / 3505.2 174.00 / 4420

Wall Opening Height 83.00 / 2108 83.00 / 2108 124.00 / 3149.6 124.00 / 3149.6

Wall Opening Length 101.50 / 2578 105.50 / 2680 107.50 / 2730 107.50 / 2730

Fascia Width1 107.00 / 2718 135.00 / 3429 140.00 / 3556 176.00 / 4470

Fascia Height 84.00 / 2134 84.00 / 2134 125.00 / 3175 125.00 / 3175

1. Fascia extends 1” beyond wall opening on each side overlapping the wall and sealing the opening.2. Allow minimum of 2” clearance at rear of recessed unit only. Cabinet sides models have rear clearance built in.3. Contact PRIMUS for alternative wall opening.4. Standard Left Side Service/equipment access shown. Optional right side access is available.

Volume / Dimensions Chart Single Door

*Refer to General Arrangement (GA) drawings for details and final connection point to utility services (S-Steam, W-Water, D-Drain, E-Electrical, A-Air).

MODEL J K L M

Chamber Size(W x H x L)

Inches/Millimeters

26 x 63 x 48660 x 1600 x 1219

26 x 63 x 76660 x 1600 x 1930

35 x 57 x 48889 x 1448 x 1219

35 x 57 x 60889 x 1448 x 1524

Chamber Capacity 46 ft3 / 1288 L 72 ft3 / 2040 L 55 ft3 / 1569 L 69 ft3 / 1962 L

Overall Width 40.25 / 1022.35 40.25 / 1022.35 52.31 / 1329 52.31 / 1329

Overall Height1 109.00 / 2768.6 109.00 / 2768.6 96.00 / 2438 96.00 / 2438

Overall Length (SD)2,3 64.00 / 1625.6 92.00 / 2336.8 63.50 / 1613 75.50 / 1918

Wall Opening Width4 86.00 / 2184 86.00 / 2184 105.00 / 2667 105.00 / 2667

Wall Opening Height 94.00 / 2387.6 94.00 / 2387.6 83.00 / 2108 83.00 / 2108

Wall Opening Length 62.25 / 1581.15 90.25 / 2292.35 63.50 / 1613 75.50 / 1918

Fascia Width1 88.00 / 2235 88.00 / 2235 107.00 / 2718 107.00 / 2718

Fascia Height 95.00 / 2413 95.00 / 2413 84.00 / 2134 84.00 / 2134

Page 10: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

10Form 2076 -3.10 1119 LG

TDS Drawings

Page 11: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

11

Specification and Technical Data Sheet

1. Fascia extends 1” beyond wall opening on each side overlapping the wall and sealing the opening.2. Allow minimum of 2” clearance at rear of recessed unit only. Cabinet sides models have rear clearance built in.3. Handle projects from the face of the fascia to the outer radius of the handle 2.375” for Models AA & A. Models B, C, & D the handle projects 4.25”.4. Contact PRIMUS for alternative wall opening.5. Standard Left Side Service/equipment access shown. Optional right side access is available.

MODEL N O P Q

Chamber Size(W x H x L)

Inches/Millimeters

35 x 57 x 86889 x 1448 x 2184

49 x 57 x 861245 x 1448 x 2184

49 x 86 x 861245 x 2184 x 2184

61 x 86 x 861549 x 2184 x2184

Chamber Capacity 99 ft3 / 2812 L 139 ft3 / 3936 L 210 ft3 / 5939 L 261 ft3 / 7393 L

Overall Width 52.31 / 1329 66.31 / 1684 73.81 / 1875 85.81 / 2180

Overall Height1 96.00 / 2438 96.00 / 2438 137.00 / 3480 137.00 / 3480

Overall Length (SD)2,3 105.75 / 2686.1 109.38 / 2778 109.75 / 2788 109.75 / 2788

Wall Opening Width4 105.00 / 2667 133.00 / 3378 138.00 / 3505.2 174.00 / 4420

Wall Opening Height 83.00 / 2108 83.00 / 2108 124.00 / 3149.6 124.00 / 3149.6

Wall Opening Length 100.50 / 2553 104.12 / 2645 104.12 / 2645 104.12 / 2645

Fascia Width1 107.00 / 2718 135.00 / 3429 140.00 / 3556 176.00 / 4470

Fascia Height 84.00 / 2134 84.00 / 2134 125.00 / 3175 125.00 / 3175

Volume / Dimensions Chart Double Door

*Refer to General Arrangement (GA) drawings for details and final connection point to utility services (S-Steam, W-Water, D-Drain, E-Electrical, A-Air).

MODEL J K L M

Chamber Size(W x H x L)

Inches/Millimeters

26 x 63 x 48660 x 1600 x 1219

26 x 63 x 76660 x 1600 x 1930

35 x 57 x 48889 x 1448 x 1219

35 x 57 x 60889 x 1448 x 1524

Chamber Capacity 46 ft3 / 1288 L 72 ft3 / 2040 L 55 ft3 / 1569 L 69 ft3 / 1962 L

Overall Width 40.25 / 1022.35 40.25 / 1022.35 52.31 / 1329 52.31 / 1329

Overall Height1 109.00 / 2768.6 109.00 / 2768.6 96.00 / 2438 96.00 / 2438

Overall Length (SD)2,3 70.13 / 1781.18 98.13 / 2492.5 67.75 / 1720.9 79.75 / 2025.7

Wall Opening Width4 86.00 / 2184 86.00 / 2184 105.00 / 2667 105.00 / 2667

Wall Opening Height 94.00 / 2387.6 94.00 / 2387.6 83.00 / 2108 83.00 / 2108

Wall Opening Length 62.44 / 1585.98 90.44 / 2297.18 62.50 / 1587 74.50 / 1892

Fascia Width1 88.00 / 2235 88.00 / 2235 107.00 / 2718 107.00 / 2718

Fascia Height 95.00 / 2413 95.00 / 2413 84.00 / 2134 84.00 / 2134

Page 12: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

12Form 2076 -3.10 1119 LG

MODEL J K L M N

Chamber Size(W x H x L)

Inches/Millimeters

26 x 63 x 48660 x 1600 x 1219

26 x 63 x 76660 x 1600 x 1930

35 x 57 x 48889 x 1448 x 1219

35 x 57 x 60889 x 1448 x 1524

35 x 57 x 86889 x 1448 x 2184

Pit Depth 12.00 / 304.8 12.00 / 304.8 12.00 / 304.8 12.00 / 304.8 12.00 / 304.8

Pit Length (SD) 61.00 / 1569.4 89.00 / 2260.6 61.00 / 1569.4 73.00 / 1879.6 99.00 / 2514.6

Pit Length (DD) 59.50 / 1511.3 87.50 / 2222.5 59.00 / 1698.6 71.00 / 1803.4 97.00 / 2463.8

Pit Width 81.00 / 2057 81.00 / 2057 98.00 / 2489 98.00 / 2489 98.00 / 2489

MODEL O P Q S

Chamber Size(W x H x L)

Inches/Millimeters

49 x 57 x 861245 x 1448 x 2184

49 x 86 x 861245 x 2184 x 2184

61 x 86 x 861549 x 2184 x2184

37 x 86x 86939.8 x 2184.4 x 2184.4

Pit Depth 12.00 / 304.8 12.00 / 304.8 12.00 / 304.8 18.00 / 457

Pit Length (SD) 101.00 / 2565.4 101.00 / 2565.4 101.00 / 2565.4 N/A

Pit Length (DD) 97.88 / 2486.5 98.50 / 2501.9 98.50 / 2501.9 98.50 / 2501.9

Pit Width 130.00 / 3302 138.00 / 3505.2 165.00 / 4191 117 / 2972

Pit Dimentions

Pit Installation Details

Page 13: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

13

Specification and Technical Data Sheet

Steam Source

Electric Boilers¨ EB Carbon steel. Uses house supplied water. Includes feedwater boost pump.¨ EBC Stainless steel construction for clean steam generation. Includes stainless steel feedwater boost pump.

NOTE: Stainless Steel Boilers shall be operated using only deionized water, having a maximum conductance of 1 microsiemen per cm (1µS/cm) [minimum specific resistivity of 1 megohm per cm (1MW/cm)].

Boiler Utilities

MODEL J K L M N

Chamber Size(W x H x L)

Inches/Millimeters

26 x 63 x 48660 x 1600 x 1219

26 x 63 x 76660 x 1600 x 1930

35 x 57 x 48889 x 1148 x 1219

35 x 57 x 60889 x1448 x 1524

35 x 57 x 86889 x 1448 x 2184

Boiler Size kW/Hr 144.0 180.0 158.0 180.0 240.0

Boiler Steam Output lbs/Hr 433.0 542.0 475.0 542.0 723.0

Stand Alone Model EB-J EB-K EB-L EB-M EB-N

208 Vac, 3Ph Amperes 400 N/A N/A N/A N/A

240 Vac, 3Ph Amperes 347 434 379 434 578

480 Vac, 3Ph Amperes 173 217 190 217 289

V.A.C. 110, 60Hz1 Amperes 10 10 10 10 10

MODEL O P Q S

Chamber Size(W x H x L)

Inches/Millimeters

49 x 57 x 861245 x 1448 x 2184

49 X 86 X 861245 x 2184 x 2184

61 X 86 X 86889 X 1448 X 1524

37 x 86 x 86940 x 2184 x 2184

Boiler Size kW/Hr 360.0 450.0 540.0 450.0

Boiler Steam Output lbs/Hr 1054.0 1355.0 1656.0 1355.0

Stand Alone Model EB-O EB-P EB-Q EB-P

208 Vac, 3Ph Amperes N/A N/A N/A N/A

240 Vac, 3Ph Amperes N/A N/A N/A N/A

480 Vac, 3Ph Amperes 434 542 650 542

V.A.C. 110, 60Hz1 Amperes 10 10 10 10

1. Utilities are for controls only2. Electric boilers are available in Carbon Steel or Stainless Steel. Contact PRIMUS for overall dimensions and utility connections. 3. All models are stand-alone.4. GMP Validatable Option Only. Sanitary piping is required. 5. Low Water Cutoff is standard and the “automatic reset” feature is disabled with this option. The boiler will need to be manually reset.6. Water Quality - For best results, feed water supply should be evaluated prior to initial startup to ensure it is of the quality necessary for the application, various external treatment processes (water

softener, water conditioning, etc.) may be used. Contact PRIMUS for further recommendations.

Page 14: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

14Form 2076 -3.10 1119 LG

STEAM (S) WATER (W) AIR (A) DRAIN (D) ELECTRICAL (E)

Building Steam Supply• Pipe Size: See Below• Quality: Condensate free

97% to 100% saturated vapor (sutibly trapped to ensure dry steam and filtered to remove particulates)

• Pressure: 50 to 80 PSIG Dynamic

Note:1. Steam-to-Steam

Generator requires minimum pressure 65 PSIG house steam

WATER (W)Cold Water Supply• Pipe Size: see Below• Temperature: < 70o F• Pressure: 50-70 PSIG

Dynamic

Instrument Air• Connection: See

Below• Quality: Dry and oil

free• Pressure: 60-80 PSI

Dynamic

Building Drain SystemMinimum 2”• Location: Locate

floor sink directly under sterilizer

Note:1. Exhaust discharge is

cooled to < 140oF2. 12” x 12” x 8” floor

sink is recommended by PRIMUS

3. Floor should be sloped to drain

Building PowerSupply - Dedicated Circuit• Volts: 110• Phase: Single• Amps: 10

Note:Additional circuits required for ancillary and optional equipment i.e., vacuum pump, boost pump, boiler, etc.

MODEL NPT LBS/HR(KG/HR) NPT GPM

(LPM) NPT NPT(Discharge Pipe Size)

J 1” 364.0 (165.1) 3/4” 14 (53) 1/4” 1”

K 1” 480.0 (217.72) 1” 25 (95) 1/4” 1 1/4”

L 1” 430.0 (195.0) 1” 25 (95) 1/4” 1 1/4”

M 1” 480.0 (217.72) 1” 25 (95) 1/4” 1 1/4”

N 1” 685.0 (310.7) 1” 25 (95) 1/2” 1 1/4”

O 1” 890.0 (403.7) 1” 25 (95) 1/2” 1 1/4”

P 1 1/2” 1210.0 (548.85) 1 1/2” 30 (114) 1/2” 1 1/2”

Q 1 1/2” 1440.0 (653.2) 2” 30 (114) 1/2” 1 1/2”

S 1 1/2" 1050.0 (476.3) 1 1/2” 25 (95) 1/2” 1 1/2”

T CONTACT PRIMUS

Electrical Connection and Utilities Consumption1

Provide utility services within 6’-0” of final connection to sterilizer. Optimum sterilizer performance requires the specified utilities.

Boiler Utilities

Page 15: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

15

Specification and Technical Data Sheet

Model KBTU’S/HR

SINGLE DOOR:Through one wall, at fascia

JKLMNOPQ

55665

111924

SINGLE DOOR:Through one wall, service area

JKLMNOPQ

11.812999

11.21719

SINGLE DOOR: Free standing, cabinet total

JKLMNOPQ

161616

17.518243643

Model KBTU’S/HR

DOUBLE DOOR:Through one wall, at fascia

JKLMNOPQS

55666

11192419

DOUBLE DOOR: Through one wall, service area

JKLMNOPQS

161615151522333933

DOUBLE DOOR: Through two walls, at each fascia

JKLMNOPQS

55666

11192419

DOUBLE DOOR: Through two walls, service area

JKLMNOPQS

10108

10101014

15.514

HVAC DATA Heat loss, at ambient of 70o F

Page 16: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

16Form 2076 -3.10 1119 LG

Service and Equipment Access

Warranty

When facing the unload/clean or BioSeal side of the unit, service access is from either left or right and top.

• Wiring is laid side-by- side in Panduit® raceway channels. All wiring is clearly labeled or readily visible.

• Piping components are threaded rigid brass and flared copper fittings, positioned with sufficient space for removal and replacement without disassembling the entire piping assembly.

Wiring and piping components are industrial grade, non-proprietary, and are available through authorized service agencies, local supply houses, or direct from PRIMUS.

Sterilizer pressure vessels manufactured by PRIMUS are waranteed against defects in workmanship and materials under normal use and operation for fifteen years where the sterilizer is continually maintained under the PRIMUS or a PRIMUS Authorized Service Agent (ASA) service contract.

Architectural Notes:

1. Allow sufficient space for traps, shut-offs, filters and other utility supply components.2. Utility (service disconnects) shall be provided and installed “By Others”.3. Building or structure modifications to accommodate the sterilizer, as well as, sterilizer shoring, rigging, cribbing and/or crane requirements into the facility shall be provided “By Others”.4. Provide maximum mechanical and service access space, a minimum of 24”, additional space required when boiler specified. See General Arrangement drawing for placement of ancillary equipment and service access.5. Some options affect utility services and overall dimensions.6. Water Quality - refer to page 8.7. The Manufacturer’s Equipment Warranty does not cover failure due to improper utility provisions.8. Drawings not to scale.9. Wall thickness must be provided on single and double door models recessed through 1 wall, with cabinet sides.10. Floor under pit mounted sterilizers must be water tight and sloped to the drain.

Page 17: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

17

Specification and Technical Data Sheet

Model Sterilizer Size

inches/millimeters

J26 x 63 x 48

660 x 1600 x 1219

K 26 x 63 x 76

660 x 1600 x 1930

L35 x 57 x 48

889 x 1448 x 1219

M35 x 57 x 60

889 x 1448 x 1524

N 35 x 57 x 86

889 x 1448 x 2184

Shipping Dimensions, Single Door 2

58 x 89.5 x 59.51473 x 2273 x 1511

58 x 89.5 x 87.51473 x 2273 x 2223

68.63 x 82.5 x 59.751743 x 2089 x 1518

68.63 x 82.5 x 71.751743 x 2089 x 1823

69.5 x 82.5 x 97.51765 x 2089 x 2476

Shipping Dimensions,

Double Door 2

58 x 89.5 x 601473 x 2273 x 1524

58 x 89.5 x 881473 x 2273 x 2235

68.63 x 82.5 x 60.751743 x 2089 x 1543

68.63 x 82.5 x 72.751743 x 2089 x 1848

69.5 x 82.5 x 98.51765 x 2089 x 2502

Overall Width Dimension 79 / 2006.6 79 / 2006.6 105 / 2667 105 / 2667 105 / 2667

Knockdown Width

Dimensions41 / 1029 41 / 1029 48 / 1227 48 / 1227 48 / 1227

Weight CONTACT PRIMUS FOR WEIGHTS

Crated Weight (Additional) VAN DELIVERY ONLY

Crated Dimensions

42 x 92 x 611067 x 2337 x 15497

42 x 92 891067 x 2337 x 2261

81 x 87 x 742057 x 2210 x 1880

81 x 87 x 862057 x 2210 x 2184

82 x 87 x 1122083 x 2210 x 2845

Crated Cube 137 cu. ft. / 4 cu. m. 200cu. ft. / 6 cu. m. 302 cu. ft. / 9 cu. m. 350 cu. ft. / 10 cu. m. 494 cu ft. / 14 cu m

Loading Equipment 146 lbs / 66.kg 152 lbs / 69kg 205 lbs / 93 kg 166 lbs / 75.3 kg 282 lbs / 128 kg

Boiler 785 lbs / 356 kg 790 lbs / 359 kg 785 lbs / 356 kg 790 lbs / 358 kg CONTACT PRIMUS

1. Door beams on all units may be split for shipping purpose.2. Shipping dimensions are measured to the edge of the heat exchanger and include plumbing on, split beams removed, where applicable, front stainless panel and printer removed.3. Contact PRIMUS for shipping information for “T” size.

Model Sterilizer Size

inches/millimeters

O49 x 57 x 86

1245 x 1448 x 2184

P49 x 86 x 86

1245 x 2184 x 2184

Q61 x 86 x 86

1549 x 2184 x 2184

S37 x 86 x 86

940 x 2184 x 2184

Shipping Dimensions, Single

Door 2

84 x 83 x 1012142 x 2099 x 2570

91 x 120 x 1022311 x 3035 x 2578

103 x 120 x 1022616 x 3035 x 2578

72 x 120 x 1031803 x 3035 x 2578

Shipping Dimensions,

Double Door 2

84 x 83 x 1012142 x 2099 x 2596

91 x 120 x 1032311 x 3035 x 2604

103 x 120 x 1032616 x 3035 x 2604

71.5 x 120 x 1031803 x 3035 x 2578

Overall Width Dimension 126 / 3200 134 / 3391 146 / 3696 107 / 2718

Knockdown Width Dimensions 62 / 1585 71 / 1794 83 / 2099 50 / 1278

Weight CONTACT PRIMUS FOR WEIGHTS

Crated Weight (Additional) VAN DELIVERY ONLY FLAT BED DELIVERY

Crated Dimensions 96 x 87 x 1162438 x 2210 x 2946

104 x 125 x 1172642 x 3175 x 2972

116 x 125 x 1172964 x 3175 x 2972

Crated Cube 561 cu ft. / 16 cu m 880 cu ft. / 25 cu m 981 cu ft. / 28 cu m

Loading Equipment 370 lbs / 168 kg 460 lbs / 209 kg 528 lbs / 240 kg

Boiler CONTACT PRIMUS

Shipping Dimensions, Cubage, & Weights

Page 18: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

18Form 2076 -3.10 1119 LG

CONFIGURATION WORKSHEETUse this worksheet to define your model number and compile your selections in the Configuration Options section of this document. Forward this completed worksheet to the PRIMUS Sales department (fax 402-344-4242) or your local PRIMUS representative. Remember, PRIMUS can address your customization needs. Please contact us for more information.

Contact Information

Chamber Size Options

Project name

Project address

Company name

Contact name

Contact email, phone and fax

Quantity

Specification section/drawing number

Item number

Room number

PRIMUS quote number (completed by PRIMUS)

¨ S5 PSS5 control system¨ S11 PRI-Matic 100 Control System (PSS11)¨ S9 PRI-Matic 200 Control System (PSS9)¨ C11 Serial data output¨ C32 Authorized user access¨ C10 Remote mount control panel¨ C25 Pressure values displayed in PSIA¨ R7 Thermal printer¨ R1 Impact printer

Controls and Related Options

¨ HS House Steam¨ EB Electric Steam Generator - Carbon Steel*¨ EBC Electric Steam Generator - Stainless Steel*¨ SCS Steam to Steam Options - Consult with PRIMUS

for details* *See detail section for specific option selection

Steam SourceDoor¨ DA Single Door¨ DD Double Door

Cabinet¨ CD Panels Both Sides¨ CCL Left Side Panel¨ CCR Right Side Panel

Base Configuration (Orientation)Recessed¨ CA One Wall¨ CB Two Walls

Service and Equipment Access¨ C29 Left Side (Std)¨ C30 Right Side

¨ V4 Validation port¨ P4 Liquid-ring Electric Vacuum Pump¨ P25 Drain line strainer/valve¨ EAC Air compressor; unavailable with option P11

¨ NPN Site Specific Installation Requirements - Consult with PRIMUS for maximum utilization of available space*

*See detail section for specific option selection

General Options

MODEL O P Q S

Chamber Size(w x h x l)

inches/millimeters

49 x 57 x 861245 x 1448 x 2184

49 x 86 x 861245 x 2184 x 2184

61 x 86 x 861549 x 2184 x2184

37 x 86x 86939.8 x 2184.4 x 2184.4

Chamber Volume 139 ft3 / 3936 L 210 ft3 / 5939 L 261 ft3 / 7393 L

Select size ¨ ¨ ¨ ¨

MODEL J K L M N

Chamber Size(w x h x l)

inches/millimeters

26 x 63 x 48660 x 1600 x 1219

26 x 63 x 76660 x 1600 x 1930

35 x 57 x 48889 x 1448 x 1219

35 x 57 x 60889 x 1448 x 1524

35 x 57 x 86889 x 1448 x 2184

Chamber Volume 46 ft3 / 1288 L 72 ft3 / 2040 L 55 ft3 / 1569 L 69 ft3 / 1962 L 99 ft3 / 2812 L

Select size ¨ ¨ ¨ ¨ ¨

Page 19: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

19

Specification and Technical Data Sheet

¨ P8 Effluent decontamination configuration

¨ P8.1 Effluent decontamination as a cycle

¨ V6 BioSeal flange¨ V10 BioSeal extension panel,

stainless steel¨ V12 BioSeal enhanced-gasketed¨ P10 Compressed air to door gasket¨ C23 Remote signaling of sterilizer

status

¨ L4 Loading cart¨ L5 Additional cart shelf¨ L6 Transfer carriage¨ L7 Floor Cart/Bulk Truck¨ L8 Additional Floor Cart Shelves

¨ P3 Boost pump for low water pressure¨ E1A Step-down transformer from 220 VAC

to 110 VAC¨ E1B Step-down transformer from 480 VAC

to 110 VAC¨ E1C Step-down transformer from 480/240

to 240/120 VAC, 1 phase, 0.5 KVA¨ E2 Uninterruptible power supply¨ E3 GFI receptacle¨ E4 Lighted Din Connections

¨ EB1 CS Boiler, 208/3ph¨ EB2 CS Boiler, 240/3ph¨ EB3 CS Boiler, 480/3ph¨ EBC1 SS Boiler, 208/3ph¨ EBC2 SS Boiler, 240/3ph¨ EBC3 SS Boiler, 480/3ph¨ SCS1* Pure steam to steam double tube, sanitary¨ SCS2 Clean steam to steam singe tube, threaded¨ EBO Automatic boiler blow-down

¨ P1 Brass & Copper piping and fittings (standard)

¨ P5 Safety Valve Over Rupture Disk¨ P6 Quench Effluent on Demand

to Drain (standard)¨ P7 Condensate Sampling Valve¨ P9 SS Chamber Piping¨ P9.1 SS Jacket Piping¨ P9.2 SS Lower Piping from first

check valve to drain¨ P10 Compressed Air to Gasket¨ P11† Air Over Pressure Chamber

Cooling¨ P13 Precise Temperature Control¨ P14 Lab Low Process - Low

Temperature Cycle¨ P16 Jacket Cooling¨ P20 Steam Quality Sampler¨ P32 Chilled Water Return¨ P33 Automatic Self Cleaning

Drain line)¨ P34 Pneumatic Valves¨ P36 0.2 Micron Filter

¨ V1 Stainless steel frame¨ V2 Jacket type, 316 L SS¨ V3 Jacket insulation cover, SS¨ V6.1 Additional BioSeal flange¨ V8 Seismic restraints¨ C3* Blind flange assembly,

sanitary fittings¨ C4 Jacket pressure display,

analog, service area¨ C4.1 Panel-mounted analog jacket

pressure gauge¨ C4.2 Panel-mounted digital jacket

pressure gauge¨ C4.3 Chamber and jacket pressure

gauges (International 30-0-100)

¨ D1 GMP Validation Documentation¨ D1.1 Chamber Mapping Only¨ D1.2 Tagging Only¨ D1.3 GMP Document Only¨ D2 IQ/OQ Protocol, PSS5¨ D5 IQ/OQ Protocol, PRI-Matic 200¨ D6 IQ/OQ Protocol, PRI-Matic 100¨ D8† Software Source Code

¨ DH Power-operated door via control panel

¨ DG Power-operated door via mushroom button

¨ CA Recessed, 1 wall¨ CB Recessed, 2 wall¨ C31 Control panel opposite

of service side¨ CF Rear panel

¨ C1* Conax adapter, 4 Gland¨ C1.1 1-1/2 Conax adapter, 8 Gland¨ C1.2 2 Conax adapter, 16 Gland¨ C2* PRI-Matic °F control, °F accumulated¨ C33 Display Centigrade¨ C12 Chamber condensate alarm¨ C13*† Sterilize filter as a cycle¨ C15*† PRI-Matic in NEMA rated enclosure

(dust-proof)¨ C16*† PRI-Matic mounted in NEMA 4

enclosure (waterproof)

¨ C16.1*† PRI-Matic mounted in NEMA 4X stainless enclosure (waterproof )

¨ R3* Strip chart recorder¨ C17* “Sterilizer off” signal¨ C18* Remote sterilizer “in cycle” signal¨ C19 Chamber drain temperature, remote

digital display¨ C28* † Ethernet Printing¨ C33 Centigrade reading (displays and prints temperature in

degrees Centigrade)¨ C34 Automatic vacuum leak rate test¨ C36 Ethernet connection for RS232¨ C37 Form C dry contacts

Bio-containment Options Loading Equipment Options Utility-related Options

Sustainability Options

Steam Source Options - Detail

Piping

Vessel Options

Documentation

General Options

Additional Control Options

† PRI-Matic control only

¨ P30 PRI-Pure reverse osmosis water system

*see page x for add’l details

Page 20: Large Steam Sterilizers · 2020. 7. 29. · 1 Speci˜ca T heet Form 2076 -3.10 1119 LG PRIMUS® large steam sterilizers come in . multiple chamber configurations . for your application

20

Specification and Technical Data Sheet

Form 2076 -3.10 1119 LG

Contact Us

PRIMUS Sterilizer Company, LLC8719 South 135th Street, Suite 300Omaha, NE 68138(402) 344-4200 Fax (402) 344-4242 [email protected]

All information contained in this document is for reference only.

The base language of this document is English. Any translations to other languages must be made from the base language.

CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH APPLICABLE LAWS, REGULATIONS, LOCAL AND NATIONAL REGULATIONS.