lcms technology connects to your application · atmospheric pressure ionization (api) esi apci 22....

63
LCMS Technology Connects to Your Application June, 2018 (LCMS Product Specialist) Dr.Rittichai Charoensapyanan 1

Upload: others

Post on 21-Mar-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

LCMS Technology

Connects to Your Application

June, 2018 (LCMS Product Specialist)

Dr.Rittichai Charoensapyanan

1

Page 2: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Fundamental of Liquid Chromatography (LC)

Fundamental of Mass Spectrometer (MS)

LCMS Applications

Topics

2

Page 3: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Fundamental of Liquid Chromatography

3

Page 4: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Liquid Chromatography (LC)

Stationary Phase

Mobile phase (continuous)1.9

4.7

6.3

Retention time Identification

Peak area Quantification

• Liquid Chromatography (LC) : Separation technique which liquid is used as mobile phase

• Separation : Between two phases (Stationary phase and Mobile phase)

• Compounds are separated from each other based on their difference in affinity for the

stationary or mobile phase.

4

Page 5: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Degasser : Remove air bubble in solvents

• Pump : - Mix solvents

- Control the flow rate of mobile phase and analytes

• Autosampler : Inject the sample into a running system

• Column : Separate each components

• Column Compartment : Control a column temperature

• Detector : Detect signal from analytes after separation

HPLC System

5

DAD (UV, VIS) Fluorescence Reflective Index Mass Spectrometer

Page 6: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

HPLC System Range

Basic Automated

Highly economic & reliable

620 bar UHPLC compatible

Flow rates up to 10 mL/min

100 Hz detector range

Modular flexibility

Standard

3rd Generation Modules

620 bar UHPLC compatible

Flow rates up to 10 mL/min

Oven temp. 5 – 80 ºC

100 Hz DAD, MWD, VWD, FLD, CAD

Highest flexibility

x2 Dual LC

Two systems in one

620 bar UHPLC compatible

Flow rates up to 10 mL/min

Oven temp. 5 – 80 ºC

Automated Application Switching

Parallel and Tandem LC

Online SPE-LC

Automated method scouting

Turn key Viper kits for ease of use

RSLC

Binary and Quaternary UHPLCs

1000 bar up to 5 mL/min

800 bar up to 8 mL/min

Oven temp. 5 – 110 ºC

200 Hz DAD, MWD, VWD, FLD

Improved sub 2-µm particle column

compatibility

Ultrafast/ultra resolution system

x2 Dual RSLC

x2 Dual UHPLC System

Two systems in one

1000 bar up to 5 mL/min

800 bar up to 8 mL/min

Oven temp. 5 – 110 ºC

200 Hz DAD, MWD, VWD, FLD

Parallel and Tandem LC

Online SPE-LC

Automated method scouting

Offline 2D-UHPLC

Turn key Viper kits for ease of use

RSLCnano

UHPLC system for Nano/Cap/Micro

20 nL/min – 50 µL/min up to 800 bar

Continuous direct flow

New standard in retention time

precision

Snap-in valves

nanoViper fitting system for easy

operation

Basic Standard x2 Dual LC RSLC x2 Dual RSLC RSLCnano

6

Page 7: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

VanquishTM

Max Pressure 1517 bar

The Highest Pressure HPLC

7

Page 8: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

1.9µm

5µm

u opt

u opt

Linear Velocity (mm/s)

H E

T P

m)

5

10

15

0 61 2 3 4 5

3µm

u opt

Incr

easi

ng C

olum

n Ef

ficie

ncy

Increasing Flowrate

Advantage of Small Particle

Higher efficiency, independent of flow rate means…

Faster runs without loss of performance

8

Page 9: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Efficiency is the key!!!

5µm

1.9µm N = 142,000 plates/m

(189% higher)

N = 75,000 plates /m

Higher resolution – narrower peaks

Higher sensitivity – taller peaks

Higher peak capacity (more peaks / unit time) – narrower peaks

( )k

kNRs

+−=

1 1

41

αα

Selectivity Efficiency Retention

Advantage of Small Particle

9

Page 10: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Increase Speed, Maintain Resolution 200x2.1mm

0 2 4 6 8 10 12 14 16 18Time (min)

12µm

8µm

5µm

3µm

1.9µm600µl/min 655 bar

400µl/min 190 bar

250µl/min 102 bar

100µl/min 56bar

150µl/min 68 bar

Spee

d

Speeding up analysis with 1.9 µm Hypersil GOLD

Advantage of Small Particle

10

Page 11: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

The UltiMate™ 3000 HPLC Systems

Pump

Autosampler

Column Compartment

Detector

With ValvesStandard

Quaternary Dual-GradientBinary

VWD MWD/DAD Fluorescence Corona

Standard Thermostatted + FractionationBasic Automated

Isocratic

Coulochem

11

Page 12: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

HPLC Applications

• Built-in column compartment with 2-position, 6-port

switching valve

Switching Valve (2-position, 6-port)

12

Page 13: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

HPLC Applications

13

Load Position

• Online SPE: Extraction and

separation at the same time !!!

Injection Position

Page 14: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

HPLC Applications

• Parallel LC: Analysis of

two distinct methods !!!

14

Page 15: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Fundamental of Mass Spectrometer

15

Page 16: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

“The basis in mass spectrometry (MS) is the production of ions, that aresubsequently separated or filtered according to their mass-to-charge(m/z) ratio, and detected. The resulting mass spectrum is a plot of the(relative) abundance of the produced ions as a function of the m/z ratio.”

Niessen et al., LC-MS: Principles and Applications, 1992, Marcel Dekker, Inc., New York, p. 29.

What is Mass Spectrometer?

• Measure gas-phase ions

• Operate at very low pressure (10-5 to 10-7 torr)

• Mass spectrometer work with IONS

• Determine the mass are separated according to

their mass-to-charge (m/z) ratio

16

Page 17: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Information Rich Data

17

Page 18: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

(1023.566 x 1) - 1

= 1022.6

(512.287 x 2) - 2

= 1022.6

Mass to charge (m/z) = ( molecular weight + charge ) / charge

Mass Spectrum

18

Page 19: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Liquid

ChromatographyIonization Mass Analysis

Mass Spectrometry: Block Diagram

19

Page 20: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Ion source : Converts sample molecules (neutral) into charged molecules or molecular ions.

• Type of ionization techniques

o Matrix Assisted Laser Desorption Ionization (MALDI)

o Atmospheric Pressure Ionization (API)

- Electrospray Ionization (ESI)

- Atmospheric Pressure Chemical Ionization (APCI)

Ionization

Ion Source

20

No one ionization technique is applicable to all classes of

chemical species !

Page 21: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Atmospheric Pressure Ionization (API)

ESI APCI

21

Page 22: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Ions formed by gas phase chemistry

• Good for volatile / thermally stable

• Good for non-polar analytes

• Good for small molecules (steroids)

• Ions formed by solution chemistry

• Good for thermally labile analytes

• Good for polar analytes

• Good for large molecules (protein/peptide)

Atmospheric Pressure Ionization (API)

ESI APCI

22

Page 23: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analysis

Mass Spectrometry: Block Diagram

23

Page 24: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer

• Triple Quadrupole (QqQ)

• Orbitrap

24

Page 25: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer: Triple Quadrupoles (QqQ)

High-capacity transfer tube (HCTT)

Active collision cell (Q2)

Electrodynamic ion funnel (EDIF)

Ion beam guide with neutral blocker

Asymetric RF drive

HyperQuad quadrupole mass filter (Q1)

Dual-mode discrete-dynode detector

HyperQuad quadrupole mass filter (Q3)

25

Page 26: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer: Triple Quadrupoles (QqQ)

• Q1 and Q3 are “Mass filter” where

ions are scanned by varying the

DC/AC & RF voltages across the

quadrupole set

26

Page 27: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer: Triple Quadrupoles (QqQ)

• Q2 is “Collision Cell” where precursor

ions are fragmented and pass through

Q3 for ion sorting again

Precursor Ion

Fragmentation

(Collision gas: Ar)

Product Ions

27

Page 28: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Scan Modes in QqQ

Purpose: Survey scan

of a chromatographic

peak

Purpose: Quantitation

on a specific m/z

range of ions

m/z 200-400

m/z 250

28

Page 29: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Scan Modes in QqQ

Purpose: Targeted quantitation

29

Page 30: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Fixed m/z: 400

m/z 300 m/z 500

m/z 400

m/z 400 m/z 500

Fixed m/z: 500

m/z 300 m/z 400

m/z 400 Fixed m/z: 400 Fixed m/z: 400

m/z 400

m/z 400

Scan Modes in QqQ

30

Page 31: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

SRM

m/z 300 m/z 400

m/z 400

m/z 250

Q1: Precursor Ion Q3: Product IonQ2: Fragmentation

m/z 400

m/z 250

m/z 150m/z 400

m/z 400

m/z 400

m/z 300

m/z 100

m/z 250

Fixed m/z: 250Fixed m/z: 400

31

Fragment

Page 32: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Scan Modes in QqQ

32

Page 33: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

RT: 0.00 - 75.04 SM: 7G

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75Time (min)

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

52.33

47.88

31.3055.14

34.47 50.24

39.421.00 18.87 23.568.09 24.156.50 17.2211.51 65.28 70.26 72.6363.6542.17 44.24 56.03

31.30

39.8538.39 47.883.23 30.99 40.53 59.413.45 64.64 67.2452.44 73.5755.5327.2610.36 21.9019.6614.03

NL: 2.91E8Base Peak F: + c NSI Full ms [ 400.00-1800.00] MS data14

NL: 7.97E7Base Peak m/z= 1030.90-1031.90 F: + c NSI Full ms [ 400.00-1800.00] MS data14

SIM

Full Scan

Full Scan VS SIM

34

Page 34: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

18 19 20 21 22 23 24 25 26 27Time (min)

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

0

20

40

60

80

100

Rel

ativ

e A

bund

ance

0

RT: 23.76

25.37

23.5327.3522.93

26.6220.19 23.43 24.55 25.4824.6721.25 26.2622.41 24.0220.57 22.1719.61

RT: 23.76

24.5923.13 24.17 25.18 25.3422.93 27.0926.31 26.4623.37

21.89 22.5221.6721.1520.5719.65 20.15

RT: 23.77

RT: 23.77

NL: 2.67E4m/z= 271.50-272.50 F: + c SIM ms [236.50-237.50, 271.50-272.50, 306.50-307.50] MS probe20f_sim

NL: 9.94E3m/z= 306.50-307.50 F: + c SIM ms [236.50-237.50, 271.50-272.50, 306.50-307.50] MS probe20f_sim

NL: 6.10E5m/z= 207.50-208.50 F: + c EI SRM ms2 237.000 [207.999-208.001] MS Genesis Probe20F

NL: 1.06E6m/z= 236.50-237.50 F: + c EI SRM ms2 272.000 [236.999-237.001] MS Genesis Probe20F

Superior Selectivity

Free from sample matrix

SRM

SIM

SIM VS SRM

35

Page 35: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer

• Orbitrap

36

Page 36: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer: Orbitrap

37

Page 37: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Orbitrap Mass Analyzer: Principle of Operation

Makarov A. Anal. Chem. 2000, 72, 1156-1162.

(r,z)

(r,φ)

Image Current

38

qmk

z /=ω

Page 38: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Many Ions Generate a Complex “Transient”

Fourier Transform

Frequency DomainImage Current

Mass Spectrum

qmk

z /=ω

39

Page 39: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Q ExactiveTM

40

Page 40: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Analyzer

• Orbitrap

High Resolution Accurate Mass (HRAM)

42

Page 41: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Resolution

43

794792790788786

100908070605040302010

794792790788786

100908070605040302010

Δm= 0.786

• Ability of a mass spectrometer to distinguish between ions of nearly equal m/z ratios (isobars).

m - measured mass

Δm - peak width measured at 50% peak intensity (Full Width Half Maximum)

Low Resolution

R = 786.6 = 1,000

High Resolution

R = 786.6 = 100,000

0.786

0.007Δm= 0.007

Page 42: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Commercial High Resolution MS Technology Race

Time progression (year)

Mas

s re

solu

tion

(FW

HM

)

Bendix Tof0

50000

100000

150000

200000

250000

300000

350000

400000

450000

500000

1955 1965 1975 1985 1995 2005 2015

Orbitrap Tof / QTof

Ion Trap-Orbitrap

Quad Orbitrap

Tribrid Orbitrap

ORBITRAP’s spectacular climb in performance in a decade!

First Q-Tof

Q-Orbitrap*

New Q-Orbitrap

New Tribrid Orbitrap

Entry Q-Orbitrap

LIT-Orbitrap ETD

44

Page 43: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• The precision of which the mass is measuredby the mass spectrometer.

• Typical way of reporting mass error in ppm(relative measure) or mDa (absolute measure)

Mass error = Measured – Exact Mass x 106

Mass Accuracy

45

( )Exact Mass

C = 12.0000

H = 1.0078

N = 14.0031

O = 15.9949

S = 31.9721

Good

786.6003 (-1.19 ppm)

794792790788786

100908070605040302010

786.60124

787.60463

788.60773

789.61068

786.70 (+124 ppm)

error = 786.6003 – 786.60124 x 106

786.60124

( )error = 786.7000 – 786.60124 x 106

786.60124

( )

Not so Good

Page 44: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Typical mass accuracy capability for various MS types

Source: Metabolomics Fiehn’s lab

Mass Accuracy

46

Page 45: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Mass Accuracy

47

• Increases confidence in identification

CxNxOxHx

C12H22O11

Page 46: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Measured Mass Mass Error (Da) Possible Formula Exact Mass

32.0 ± 0.2

O2 31.9898

CH3OH 32.0261

N2H4 32.0374

S 31.9721

32.02 ± 0.02CH3OH 32.0261

N2H4 32.0374

32.0257 ± 0.002 CH3OH 32.0261CH3OH

• Main advantage: the possibility to determine the elemental composition of individual

molecular or fragment ions, a powerful tool for the structural elucidation or confirmation.

C = 12.0000

H = 1.0078

N = 14.0031

O = 15.9949

S = 31.9721

Mass Resolution and Accuracy

48

Page 47: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

100

892.60 892.65 892.70 892.75 892.80 892.85 892.90 892.95 893.00m/z

892.7862

892.8323

892.7952

892.7377

892.8226

892.7903100

892.7378

892.8246

100892.7973

Mass Resolution and Accuracy

49

Page 48: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Isobaric compounds separation

R = 100,000292.02 292.04 292.06 292.08

m/z

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

292.04031C 10 H 15 O5 N P S

292.02656C8 H11 O 3 N 5 Cl S

0

R = 50,000

R = 35,000

R = 15,000

Thiamethoxam

[M+H]+ = 292.02656

Parathion

[M+H]+ = 292.04031

∆m0.0138 Da

Mass Resolution and Accuracy

50

Page 49: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Isobaric compounds separation

Mass Resolution and Accuracy

51

Page 50: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Removing interferences

High resolution is very important for samples with complex matrix (e.g. biological,

food), since they will contain a significant number of background ions

Mass Resolution and Accuracy

52

Page 51: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

LCMS Applications

53

Page 52: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

LCMS Applications

54

• Pharmaceutical analysis

– Bioavailability studies

– Drug metabolism studies, pharmacokinetics

– Characterization of potential drugs

– Drug degradation product analysis

– Screening of drug candidates

– Identifying drug targets

• Etc.

• Biomolecule characterization

– Proteomics

– Oligonucleotides

• Environmental analysis

– Pesticides on foods

– Soil and groundwater contamination

• Forensic and clinical analysis

• Toxicology

Page 53: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Application in Food

Melamine

SRM Transitions

(Q1) 127 -> 68 (Q3)

(Q1) 127 -> 85 (Q3)

Identification and Quantitation of Melamine in Infant Milk

Varelis et al. Thermo AN62732. 2008

Sample Prep

(SPE)

LC-MS/MS

(Targeted SRM)

LC: AccelaTM

Column: BioBasic AX (Ion Exchange)

Column Temperature: 30ºC

Injection Volume: 1 µL

Mobile Phase: A) 85% ACN + 10% IPA + 5%

Ammonium acetate; B) 90% water and 10%ACN

Flow Rate: 400 µL/min Run Time: 5 min

MS: TSQ Quantum UltraTM

Ionization: Positive ESI

Modes: Targeted SRM

Instrument conditions

55

Page 54: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Identification and Quantitation of Melamine in Infant Milk

• Limit of Detection (LOD): <10 ppb (below the FDA requirement at 1 ppm)

Varelis et al. Thermo AN62732. 2008

Application in Food

56

Page 55: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Application in Halal Food

Determination of Meat Authenticity

Orduna et al. Thermo AN64677. 2016

LC & HRAM MS

Conditions

57

Page 56: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Orduna et al. Thermo AN64677. 2016

Application in Halal Food

Determination of Meat Authenticity

58

Page 57: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Application in Pharmaceutical Products

Li et al. Thermo AN64427. 2015

Discovery and Characterization of Natural Components

LC: Ultimate 3000TM

Column: Hypersil Gold C18

Column Temperature: 35ºC

Injection Volume: 1 µL

Mobile Phase: ACN + 0.1% FA (linear gradient)

Flow Rate: 300 µL/min Run Time: 50 min

MS: Q ExactiveTM

Ionization: Positive and Negative ESI

Modes: dd-MS2

Instrument conditions

59

Page 58: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Application in Pharmaceutical Products

Discovery and Characterization of Natural Components

Extracted Ion Chromatogram and Isotopic Pattern

MS and MS2 Spectrum

MS2 Spectrum Interpretation

60 Li et al. Thermo AN64427. 2015

Page 59: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Application in Pharmaceutical Products

Discovery and Characterization of Natural Components

• 21 Cantharidins were detected

in the blister beetle extract.

• 16 new Cantharidins were

discovered and identified with

mass accuracy <1 ppm by

fragment ion search (FISh)

function, using cantharidinimide

as the parent structure.

Li et al. Thermo AN64427. 201561

Page 60: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

• Peptide Labeling (multiple studies with a single experiment)

62

Application in Proteomics

Page 61: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

63

Application in Proteomics

• Protein Identification

• Expression Level of Protein

• Protein Modification

• Protein Localization

• Peptide Mapping

• Pathways

• Etc.

Page 62: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

http://planetorbitrap.com/

LCMS Technology and Applications

64

Page 63: LCMS Technology Connects to Your Application · Atmospheric Pressure Ionization (API) ESI APCI 22. ... Δm - peak width measured at 50% peak intensity (Full Width Half Maximum) Low

Thank You for Your Attention

65