linear accelerator quality assurance (1)
TRANSCRIPT
TG-40 and TG-142
Samir Laoui
Date: July 29th, 2015
AAPM Task Group 40 Report “Comprehensive QA for Radiation
Oncology” 1994
Introduction
“Every patient with cancer deserves to receive the best possible management to achieve cure, long-term tumor control or palliation”
Inter-Society Council for Radiation Oncology (ISCRO, 1986)
Delivery of treatment in an accurate and consistent manner is by no means easy to achieve, since the radiation therapy process is a complex interweaving of a number of related tasks for designing and delivering radiation treatments
Background
Quality Assurance: “the maintenance of a desired level of quality in a service or product, especially by means of attention to every stage of the process of delivery or production”. online dictionary
In radiation therapy : A set of procedures that ensure a safe dose delivery to the tumor site while sparing normal tissue, and minimal exposure to the personnel
To achieve this goal, institutions are required to establish a set of baseline standards
Processes of Machine QA
Acceptance Testing
Commissioning
Quality Assurance
Clinical Needs and Purchase
Baseline
Type of radiation equipment QA
Quality Assurance of Linear Accelerator
Quality Assurance of Simulator
Quality Assurance of Brachytherapy
Quality Assurance of Diagnostics CT
Quality Assurance of treatment Planning System (TPS)
Quality Assurance of QA tools
AAPM Task Group 40 Report “Comprehensive QA for Radiation Oncology” 1994
Scope: Guidelines for administrators
QA for External Beam Radiation Therapy Equipment Co-60 Units (Table I)
Linear Accelerators
Simulators
CT scanners
QA of measurement equipment
Treatment planning computer system
External beam Treatment planning
Brachytherapy
Treatment plans QA for individual patients
Tolerances: TG-40
Adopted from AAPM Report 13 (AAP, 1984)
Report 13 used the method of quadratic summation to set the tolerance values
The overall dosimetric uncertainty of ± 5% (ICRU, 1976). This can be only be achieved if every individual task is performed with an accuracy better than ±5%
55522
Linac QA TG-40 (1994)/ Daily
Could seriously affect patient positioning/ registration of radiation fields with target volume
Directly affect dose delivery
Directly affect patient safety
RX may continue for 3-5 % output
Action Level set at 5%, RX has to be halted
Linac QA TG-40 (1994)/ Monthly
Smaller impact on patient
Not likely to change on daily basis
Linac QA TG-40 (1994)/ Annual
Baseline verification
TG-40: Other machines QA
QA of Simulators (Table III)
Should be subject to routine mechanical checks as Linacs
Image quality
QA of CT scanners (No table)
Laser alignment should be checked daily
Image quality has to be checked periodically
Electronic density check is annual
QA of measurement equipment (Table IV)
Redundancy in dose calibration equipment is recommended using Sr-90 source
Other equipment should be calibrated
TG-40: QA of treatment planning
Treatment planning computer system QA (Table V)
(Van Dyke et al, 1993, ICRU 42)
Beam data library
Dose calculation models
Operating instructions
Treatment planning system process (Table VI) Prescription Positioning and Immobilization Data Acquisition Contouring
Data Transfer Dose computation Plan Evaluation Computation of Monitor Units
Beam modifiers Plan implementation
Treatment planning QA for individual patient Treatment planning review
Monitor Unit review
Plan implementation
TG-40: Brachytherapy
Higher tolerances compared to external beam’s, 15%
Sealed sources description and calibration Independent source activity strength verification Traceability
Treatment planning and dosimetry Planning (TPS) Localization (CT) Dose calculation algorithms Patient dose calculation (Checked to within +/-15%) Delivery of treatment Documentation
Remote afterloading Calibration Verification of source position
AAPM Task Group 142 Report “Quality Assurance of Medical
Accelerators” 2009
AAPM Task Group 142 Report “Quality Assurance of Medical Accelerators” 2009
Specific to Linacs
Does Include QA of MLC (TG-50) Include QA of Portal imaging (TG-58) Respiratory Motion (TG-76) Kilovoltage localization (TG-104)
Does not Include instructions for performing QA tests Accelerator commissioning (TG-106) QA for Tomotherapy (TG-148) QA for Robotic Radiosurgery (Tg-135) Positioning (TG-147) VMAT/Arc therapy are not covered (TG-119)
AAPM Task Group 142 Report “Quality Assurance of Medical Accelerators” 2009
Scope: Quality assurance for medical accelerators
Test frequencies
Guidelines for tolerance values
Ancillary treatment devices not in TG-40
Asymmetric jaws
Dynamic/Virtual/Universal wedge
MLC
Radiographic Imaging
Respiratory gating
Special procedures/SRS/SBRT/TBI/TSET/IMRT
TG-142: General
Baseline dosimetric values entered into treatment planning systems to characterize and model the treatment machine directly after calculated plans
Values can deviate from baseline values as a result of Linac malfunction
Accidents
Mechanical breakdown
Linac part replacement
Wear and tear
QA procedures are made in place to detect these changes
The report recommends that institution be proactive in detecting changes based on workload (IMRT, MLC)
TG-142: General
A beam output constancy is an important quantity for an accurate and reproducible dose coverage
The report includes monthly tolerance values that are specific to a constant beam shape. Central axis output (BP) as well as off-axis output are defined during commissioning
TG-40 TG-142
TG-142: Guidelines for Tolerance Values
Acceptance testing Commissioning and acceptance test set the baseline for future dosimetric and
mechanical measurements. This will ensure that the machine is operating within tolerance values
Tolerance and Action Levels* Level 1: Inspection is required
Level 2: Maintenance must be scheduled in the near future
Level 3: Immediate action and/or treatment is halted
* TG-142 does not explicitly define the action level values
Linac QA TG-142 (2009)/ Daily
The tables are differentiated
into non-IMRT/Stereotactic
machines and
IMRT/Stereotactic machines
Linac QA TG-142 (2009)/ Monthly
MONTHLY
TG-142 Vs. TG-40
Linac QA TG-142 (2009)/ Annual
Linac QA TG-142 (2009)/ Annual
•
Ancillary RX devices: Dynamic/Virtual Wedge
•
•
Multileaf Collimation QA
Imaging QA
•
Example: Acuity Daily QA
Reference TG-76
Dynamic phantom which simulate human organ motion are recommended to test target localization and respiratory gated treatment accuracy
Tests include: Beam Energy constancy
Beam output constancy
Temporal accuracy of phase/Amplitude gating windows (example: 250 and 1500 ms with 100 ms tolerance)
Respiratory Gating QA
Summary
TG-142 is considered to be flexible compared with TG-40
TG-142 recommend establishing institution-specific baseline values
Tolerances vary depending on the machine functionality (Non-IMRT, IMRT, SRS/SBRT)
There are overlaps of daily, monthly and annual tests that helps cross check the machine performance with independent techniques
Summary
Recommends that the QA equipment be calibrated per TG-106
Recommends an end-to-end check if a new/revised procedure is introduced
During annual QA, absolute machine output should be calibrated as per the TG-51 calibration protocol (Baselines)