iNIBP Case Report – Experiences in OR Linear Inflation Technology Non-invasive Blood Pressure Measurement

Upload: others

Post on 19-Oct-2020

17 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

iNIBP Case Report – Experiences in OR

L i n e a r I n f l a t i o n T e c h n o l o g y N o n - i n v a s i v e B l o o d P r e s s u r e M e a s u r e m e n t

7908 SP64-007 ’14.08. CLB.E.

Printed in Japan

Page 2: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

C O N T E N T S

Introduction …………………………………………………………………………… 2

Technological Description of iNIBP

iNIBP advantages ……………………………………………………………………… 2

iNIBP technology adapts to each patient situation and condition ……………… 4

[Case1] Laparoscopic assisted distal gastrectomy …………… 5

[Case 2] Pelvic lymphadenectomy,

para-aortic lymphadenectomy ………………………… 6

[Case 3] Laparoscopic resection of descending colon …………… 7

[Case 4] Radical prostatectomy ………………………………………… 8

[Case 5] Laparoscopic uterine myomectomy ……………………… 9

Discussion ………………………………………………………………………… 10

Page 3: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

2

IntroductioniNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This case report, which describes the perioperative use of iNIBP, is designed to help clinicians and healthcare professionals understand and appreciate the performance characteristics of iNIBP, including the advantages and limitations of the measurement.

Technology Description of iNIBP

Figure 1 Comparison of the iNIBP inflation method to the conventional deflation method

(Upper) The conventional step-deflation method: A cuff is rapidly inflated to a target inflation pressure, and then deflated step-wise while sensing oscillations.

(Lower) iNIBP inflation method: A cuff is slowly inflated to a lower inflation pressure while simultaneously sensing oscillations.

Figure 2 Histogram of measurement time of iNIBP inflation method and deflation method 323 measurement data collected in 64 cases in operating room were used to compare measurement time of iNIBP inflation method and conventional deflation method.

250

200

150

100

50

05 10 15 20 25 30 35 40 45 50 55 60 65 70 Measurement time

(sec)

Fre

que

ncy

(Num

ber

)

Onodera J. et al. Validation of in�ationary non-invasive blood pressure monitoring in adult surgical patients. J Anesth. 2011 Feb;25(1):127-30.

iNIBP13.0±2.3 sec (avg)

iNIBP De�ation method

De�ation method32.7±13.6 sec (avg)

20 sec shorter

SYS

DIA

SYS

DIA

Cuff pressure

oscillation

Cuff pressure

oscillation

iNIBP advantages

Non-invasive blood pressure (NIBP) is commonly used as an important vital sign to evaluate patients' hemodynamic status. In the traditional oscillometric NIBP measurement using the conventional step-deflation method, a cuf f placed on the patients' l imb is rapidly inf lated to a target pressure greater than systolic, and then def lates step-wise while sensing the oscillations of the arterial wall created by pulsatile blood f low through the ar tery. On the other hand, iNIBP inf lation method completes the measurement while inflating a cuff. Therefore, when compared to the conventional method, the iNIBP measurement t ime is shor ter (Figure 2) and target inf lat ion pressure is lower1, maximizing patient comfor t whi le preserving the accuracy of this important parameter.The conventional step-def lat ion method sets the target inf lat ion pressure based on the prev ious systo l ic b lood pressure (SYS). When the patient's blood pressure increases compared to previous measurements, the target inf lation pressure may be set to a level which is insuf f icient for the measurement. In these cases, another inflation-deflation cycle is required, and may be repeated, increasing both the measurement t ime and potentia l for patient discomfor t (Figure 3.a). When the patient's blood pressure has dropped since the previous measurement, the target inflation pressure is set significantly above the previous systolic pressure, which again leads to longer measurement time and excessive pressure applied to the patient (Figure 3.b). With iNIBP technology from Nihon Kohden, the algorithm ef fects a s low inf lat ion of a cuf f whi le s imultaneously detect ing oscillations, then deflates the cuff as soon as SYS is determined. The advantages of iNIBP are best demonstrated when patient's blood pressure is varying, for example during surgery. As iNIBP inflates while sensing blood pressure variation, it can complete the measurement in a shorter amount of time without applying excessive and unnecessary pressure to the patient.

Page 4: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

3

BP

200

100

9:30 10:30 11:30 12:30 13:30

SYS

DIA

SYS elevated to 200 mmHg

SYS dropped to 100 mmHg

10

140mmHg

210mmHg

250mmHg

SYS

20 30 40 50 60

Time (sec)

Pressure too low

Pressure still too low

De�ationmeasurement

completes

De�ationmeasurement

completes

iNIBPDe�ation method

iNIBP can complete the measurement in a shorter amount of time as it in�ates the cuff while sensing blood pressure variation and detecting oscillations, whereas the de�ation method needs to repeat the in�ation-de�ation cycle when the target in�ation pressure, based on previous systol ic pressure measurements are comparatively low. Shorter measurement

time

300

250

200

150

100

50

0

Cuffpressure(mmHg)

iNIBP in�ates the cuff to only slightly above the SYS while detecting oscillations and completes the measurement in a shorter amount of time, in comparison with the de�ation method, which in�ates a cuff based on a pre-determined target pressure that often signi�cantly exceeds a c t u a l S Y S , l e a d i n g t o l o n g e r measurement times with the increased potential for patient discomfort.

10

130mmHg

240mmHg

SYS

20 30 40 50 60

Time (sec)

iNIBPMeasurement completes as soon as SYS is determined

iNIBPDe�ation method

Lower pressure

300

250

200

150

100

50

0

Cuffpressure(mmHg)

Figure 3.a. iNIBP in blood pressure elevation

Figure 3.b. iNIBP in blood pressure drop

iNIBPMeasurement completes as soon as SYS is determined

Shorter measurement time

Figure 3 iNIBP in elevation and drop in blood pressure

Page 5: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

4

iNIBP technology adapts to each patient situation and condition

iNIBP automatically switches its measurement mode to step-deflation in situations where it cannot determine an accurate blood pressure value with inflation method because of factors related to measurement environment and patient's physiology. Also, iNIBP automatically employs the deflation method when signal reliability of the blood pressure reading, based on the blood pressure determined while inflating, is poor (Figure 4). Examples of these situations and conditions are shown below. When these conditions improve, iNIBP will change the mode to the inflation measurement method. Based on the analysis of actual clinical data, about 80% of measurements in the Operating Room (OR) and 70% of measurement in the Emergency Room (ER) were performed with the inflation measurement method2.

In�ationmeasurement

In�ation measurement

De�ationmeasurement

Unable tocalculate

Low reliability

In�ate a cuff

Detect oscillations

Display SYS, DIA and MAP

Calculate SYS, DIA, and MAP

Determine the reliability

In�ate a cuff until reaching a target cuff pressure

ON OFF

Detect oscillations

Display SYS, DIA and MAP

Calculate SYS, DIA, and MAP

Determine the reliability

De�ate step-wise

Measurementcompletes

Measurementcompletes

No NIBPreading provided

Unable tocalculate

Low reliability

Factor related to measurement environment

Cuff and/or tube are not secure and stable because of body motion or contact with the cuff/tube

Cuff is either too tight or too loose

Non-recommended cuff or an older cuff with worn Velcro® is used

Improperly sized cuff for the patient

Cuff placed over thick clothing

Factor related to patient’s physiology

Frequent arrhythmia

Patient has weak pulsation due to shock or has extremely low blood pressure

(SYS lower than 60 mmHg or DIA lower than 30 mmHg)

Patient has pulse waves with low amplitude (small pulse wave)

Slow pulse rate

Figure 4 Measurement flow of linear inflation technology non-invasive blood pressure measurement – iNIBP

Page 6: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

5

[Case 1] Laparoscopic assisted distal gastrectomy

Disease : Stomach cancer

Anesthesia : General anesthesia with epidural block (volatile anesthetic)

Patient : 68 years old male, height 170 cm, weight 81 kg. Cuff on left arm

Anesthesia time : 8:55 – 15:50

Surgical time : 9:37 – 15:25

In this case, a minimally-invasive procedure was performed. Blood pressure increased approximately 30 mmHg between the second ( A ) and third ( B ) NIBP measurement, then decreased 25 mmHg between the third ( B ) and fourth ( C ) measurement. In cases such as this where blood pressure varies significantly, the conventional deflation method is more likely to repeat the inflation-deflation cycle or set the target inflation pressure too high, leading to longer measurement times. This case shows that iNIBP successfully completed all measurements with the inflation measurement method even in a patient with significant blood pressure variation within a short measurement time.

[Feedback from anesthesiologist]

"I'm curious if NIBP will perform in patients with arrhythmia (e.g. AF, bigeminy or trigeminy). The iNIBP measurement time appears to be shorter compared to the conventional method when used in patients with normal sinus rhythm."

BP

140

120

100

80

60

40

20

80

60

40

20

0

Mea

sure

men

t ti

me

(sec

)

140

100

60

20

Pulserate

9:00 10:00 11:00 12:00 13:00 14:00

Anesthesia starts

Anesthesiaends

Surgerystarts

Surgeryends

De�ationmeasurement

In�ation measurement

PR

SYS

DIA

15:00 16:00

Number of measurement

Measurement time (sec)

Average Minimum Maximum

14.7 13.9 16.9

-

13

0

In�ation

De�ation - -

A

B

C

Page 7: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

6

[Case 2] Pelvic lymphadenectomy, para-aortic lymphadenectomy

Disease : Ovary cancer

Anesthesia : General anesthesia with epidural block (volatile anesthetic)

Patient : 47 years old female, height 161 cm, weight 56 kg. Cuff on left arm

Anesthesia time : 10:55 – 17:06

Surgical time : 11:21 – 16:39

BP

140

120

100

80

60

40

20

80

60

40

20

0

140

100

60

20

10:00 11:00 12:00 13:00 14:00 15:00 16:00 17:00

12.2 11.0 14.2

51.5

10

4 44.6 57.3

Number of measurement

Measurement time (sec)

Average Minimum Maximum

In�ation

De�ation

A B

E

D

C

Mea

sure

men

t ti

me

(sec

)

Pulserate

Anesthesia starts

Anesthesiaends

Surgerystarts

Surgeryends

De�ationmeasurement

In�ation measurement

PR

SYS

DIA

In the first half of the surgery (beginning of surgery until approximately 14:30), the cuff frequently detected noise caused by touching the cuff, so the iNIBP algorithm did not determine blood pressure while inflating the cuff and switched its measurement mode to the deflation method ( A B C and D ). As shown in this case, iNIBP automatically changes its measurement mode depending on the situation in order to provide reliable readings. This case also shows that iNIBP can complete the measurement with the inflation method in a shorter period of time in cases when the patient's blood pressure increases to 43 mmHg ( E ) , where the conventional method will require repetitive inflation-deflation cycles.

[Feedback from anesthesiologist]

“The measurement time of iNIBP did not appear to be significantly shorter compared to the conventional method.”

Page 8: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

7

[Case 3] Laparoscopic resection of descending colon

Disease : Descending colon cancer

Anesthesia : General anesthesia with epidural block (volatile anesthetic)

Patient : 72 years female, height 151 cm, weight 74.5 kg. Cuff on left arm

Anesthesia time : 9:05 – 15:49

Surgical time : 10:00 – 15:03

The patient had low amplitude pulse wave and her blood pressure was relatively low (SYS 60-90/DIA 30-50 mmHg) during surgery. When the amplitude of oscillations is lower than a threshold, where it cannot determine an accurate blood pressure value with inflation mode, iNIBP determines NIBP value with the deflation measurement mode instead of the inflation mode. In this case, the amplitude of patient's pulse wave frequently ranged about the threshold, so iNIBP provided half of the readings with the deflation method. However, iNIBP completed the measurement in a shorter period of time with the inflation method even when blood pressure dropped immediately after the anesthesia induction and when blood pressure elevated after the end of surgery and at awakening of the patient.

[Feedback from anesthesiologist]

“It appears that the iNIBP measurement time was half the conventional NIBP measurement time.”

BP

160

140

120

100

80

60

40

20

80

60

40

20

0

160

120

80

40

9:00 10:00 11:00 12:00 13:00 14:00 15:00 16:00

14.2 10.8 29.7

58.6

13

11 49.0 80.5

Mea

sure

men

t ti

me

(sec

)

Pulserate

Anesthesia starts

Anesthesiaends

Surgerystarts

Surgeryends

De�ationmeasurement

In�ation measurement

PR

SYS

DIA

Number of measurement

Measurement time (sec)

Average Minimum Maximum

In�ation

De�ation

Page 9: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

8

[Case 4] Radical prostatectomy

Disease : Prostate cancer

Anesthesia : General anesthesia with epidural block (volatile anesthetic)

Patient : 65 years old male, height 168 cm, weight 68 kg. Cuff on left arm

Anesthesia time : 11:50 – 17:25

Surgical time : 12:14 – 16:52

In this case, all measurements throughout the surgery, from the anesthesia induction through patient awakening, were performed with the inflation measurement mode under the condition of varying blood pressure.

[Feedback from anesthesiologist]

“I was impressed by the speed of the measurement. iNIBP can be convenient because of the shorter measurement time, which may not be essential.”

BP

140

120

100

80

60

40

20

80

60

40

20

0

140

100

60

20

11:00 12:00 13:00 14:00 15:00 16:00 17:00

13.3 12.0 16.2

-

13

0 - -

Mea

sure

men

t ti

me

(sec

)

Pulserate

Number of measurement

Measurement time (sec)

Average Minimum Maximum

In�ation

De�ation

Anesthesia starts

Anesthesiaends

Surgerystarts

Surgeryends

De�ationmeasurement

In�ation measurement

PR

SYS

DIA

Page 10: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

9

[Case 5] Laparoscopic uterine myomectomy

Disease : Uterine myoma

Anesthesia : General anesthesia (volatile anesthetic)

Patient : 37 years old female, height 151 cm, weight 53 kg. Cuff on left arm

Anesthesia time : 8:38 – 12:43

Surgical time : 9:08 – 12:25

BP

140

120

100

80

60

40

20

80

60

40

20

0

140

100

60

20

8:00 9:00 10:00 11:00 12:00 13:00

13.9 12.8 14.6

39.4

7

2 39.2 39.6

A

B

Mea

sure

men

t ti

me

(sec

)

Pulserate

Number of measurement

Measurement time (sec)

Average Minimum Maximum

In�ation

De�ation

Anesthesia starts

Anesthesiaends

Surgerystarts

Surgeryends

De�ationmeasurement

In�ation measurement

PR

SYS

DIA

After the start of the surgery, the deflation method was used for the third ( A ) and fourth ( B ) measurement because of noise caused by patient's body motion. Subsequently, iNIBP completed all measurements with the inflation measurement method throughout the surgery. This case demonstrates that iNIBP automatically switches its measurement mode to the step-deflation method in a situation where it cannot determine an accurate blood pressure value with the inflation method due to factors that can affect measurement.

[Feedback from anesthesiologist]

“iNIBP's advantage of being able to measure while inflating allows for shorter measurement time, which can make the measurement more comfortable for patients.”

Page 11: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

10

Discussion

Nihon Kohden's Linear Inflation Technology, Non-invasive Blood Pressure Measurement,“iNIBP”, determines blood pressure by detecting oscillations during the initial inflation of the cuff. Compared to the conventional method, iNIBP can provide accurate blood pressure readings in less time without applying excessive pressure to the patient, leading to maximized comfort for patients. iNIBP allows clinicians to accurately measure blood pressure immediately and when this critical vital sign is needed the most, especially during dynamic blood pressure changes during surgery. Nihon Kohden is hopeful that you find this report useful to understand the performance of iNIBP. Please contact us with your ideas and suggestions on how this technology might benefit you and your patients.

Reference1 Jun Onodera, Yoshifumi Kotake et al. Validation of inflationary non-invasive blood pressure monitoring in adult surgical

patients. J Anesth, 2011;25:127-130.2 Junichi Sasaki, Shingo Hori. Validation of inflationary non-invasive blood pressure monitoring in emergency room patients.

Poster presentation at the 34th International Symposium on Intensive Care and Emergency Medicine.

The data shown in this Case Report was provided by Department of Anesthesiology, Toho University Ohashi Medical Center.

Page 12: Linear Infl ation Technology Non ... - Nihon Kohden Europe · iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This

iNIBP Case Report – Experiences in OR

iNIBP Case Report – Experiences in OR

L i n e a r I n f l a t i o n T e c h n o l o g y N o n - i n v a s i v e B l o o d P r e s s u r e M e a s u r e m e n t

7908 SP64-007 ’14.08. CLB.E.

Printed in Japan