ln22 1v2 revb en bs - leica biosystems · 2015-06-30 · and/or gn 2 as well as objects which are...

44
Leica LN22 Liquid nitrogen freezing device Instructions for Use Leica LN22 – Liquid Nitrogen Freezing Device V 1.2, English – 10/2012 Order No.: 14 0373 80101 RevB Always keep this manual near the instrument. Read carefully prior to operating the instrument.

Upload: others

Post on 31-May-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

  • Leica LN22

    Liquid nitrogen freezing

    device

    Instructions for Use

    Leica LN22 – Liquid Nitrogen Freezing Device

    V1.2, English – 10/2012

    Order No.: 14 0373 80101 RevB

    Always keep this manual near the instrument.

    Read carefully prior to operating the instrument.

  • 3Leica LN22

    The information contained in this documentationrepresents the state of the art in science and

    technology.

    We are under no obligation to update the present

    manual periodically and on an ongoing basis ac-

    cording to the latest technical developments, nor

    to provide our customers with additional copies,

    updates etc. of this manual.

    For erroneous statements, drawings, technical

    illustrations etc. contained in this manual we ex-

    clude liability as far as permissible according tothe national law applicable in each individual

    case. In particular, no liability whatsoever is ac-

    cepted for any financial loss or consequential

    damage caused by or related to compliance with

    statements or other information in this

    manual.

    Statements, drawings, illustrations and other in-

    formation as regards contents or technical de-

    tails of the present manual are not to be consid-ered as warranted characteristics of our

    products.

    These are determined only by the contract provi-

    sions agreed between ourselves and our cus-

    tomers.

    Leica reserves the right to change technical

    specifications as well as manufacturing process-

    es without prior notice. Only in this way is it pos-

    sible to continuously improve the technology and

    manufacturing techniques used in our products.

    This document is protected under copyright

    laws. Any copyrights of this document are re-

    tained by Leica Biosystems Nussloch GmbH.

    Any reproduction of text and illustrations (or of

    any parts thereof) by means of print, photocopy,

    microfiche, web cam or other methods – includ-

    ing any electronic systems and media – requires

    express prior permission in writing by Leica

    Biosystems Nussloch GmbH.

    © Leica Biosystems Nussloch GmbH

    NOTE

    Published by:

    Leica Biosystems Nussloch GmbH

    Heidelberger Strasse 17 - 19

    D-69226 Nussloch

    Germany

    Telephone: +49 (0) 62 24 143-0

    Fax: +49 (0) 62 24 143-2 68

    Internet: http://www.LeicaBiosystems.com

  • 4 Instructions for Use V1.2, RevB – 10/2012

    Table of contents

    1. Important information ................................................................................................................ 61.1 Symbols in the text and their meanings ................................................................................... 6

    1.2 Intended use of instrument ........................................................................................................ 6

    2. Safety ............................................................................................................................................ 72.1 Safety instructions ...................................................................................................................... 7

    3. Instrument components and specifications ....................................................................... 153.1 Technical data ............................................................................................................................ 15

    3.2 Overview and instrument components .................................................................................. 16

    3.3 Standard delivery ...................................................................................................................... 17

    4. Startup ........................................................................................................................................ 184.1 Installation site requirements .................................................................................................. 18

    4.2 Unpacking ................................................................................................................................... 18

    4.3 Preparing the microtome for attachment of the LN22 ......................................................... 19

    4.4 Installing the cryochamber ...................................................................................................... 20

    4.5 Connecting to mains ................................................................................................................. 21

    4.5.1 Checking the voltage selector setting ................................................................................... 21

    4.5.2 Changing the voltage selector setting and inserting the fuses ......................................... 22

    4.5.3 Plugging in the connection cables ......................................................................................... 23

    4.5.4 Grounding the Dewar ................................................................................................................ 23

    4.5.5 Connecting the mains cable .................................................................................................... 24

    4.6 Switching on the instrument .................................................................................................... 24

    4.7 Control unit ................................................................................................................................. 25

    4.8 Specimen holders ...................................................................................................................... 27

    4.8.1 Specimen holder for block embeddings ................................................................................ 27

    4.8.2 Universal specimen holder and flat specimen holder ......................................................... 29

    4.9 Knife holders .............................................................................................................................. 30

    4.9.1 Knife holders for tungsten carbide knives ............................................................................. 30

    4.9.2 Glass knife, CryoDiamond knife, trimming knife Histo ......................................................... 30

    4.10 Preparing the pumping system ............................................................................................... 33

    4.11 Refrigerating .............................................................................................................................. 33

    4.12 Refilling the Dewar vessel ....................................................................................................... 34

    4.13 Heating up the chamber ........................................................................................................... 35

  • 5Leica LN22

    Table of contents

    5. Cleaning and maintenance .................................................................................................... 365.1 Cleaning ...................................................................................................................................... 36

    5.2 Maintenance .............................................................................................................................. 39

    5.2.1 General maintenance instructions ......................................................................................... 39

    5.2.2 Exchanging the mains fuses .................................................................................................... 39

    5.2.3 Exchanging the bulb .................................................................................................................. 40

    5.2.4 Removal of the pump valves .................................................................................................... 41

    6. Appendix .................................................................................................................................... 426.1 Ordering information ................................................................................................................. 42

    7. Warranty and service .............................................................................................................. 43

  • 6 Instructions for Use V1.2, RevB – 10/2012

    ENTER

    (5)

    1.1 Symbols in the text and theirmeanings

    Warnings

    appear in a gray box and are marked by

    a warning triangle .

    Notes,i.e. important user information, appear in

    a gray box and are marked by an infor-

    mation symbol .

    Solvents and reagents that are flamma-ble are identified by this symbol.

    This warning symbol identifies surfacesof the instrument that are hot during op-eration.

    Numbers in parentheses refer to itemnumbers in illustrations.

    Function keys that must be pressed on theinput screen are boldfaced in the text andcapitalized.

    1. Important information

    1.2 Intended use of instrument

    The Leica LN22 is a liquid nitrogen freezing de-

    vice to quickly and effortlessly change from sec-

    tioning at room temperature to freezing section-

    ing applications and vice versa.

    Any other use of the instrument isconsidered improper!

  • 7Leica LN22

    Be sure to comply with the safety instructions and warnings provided in this chapter.

    Be sure to read these instructions, even if you are already familiar with the operation and useof other Leica products.

    2. Safety

    2.1 Safety instructions

    These operating instructions contain important

    information for the operational safety and main-

    tenance of the device and are an important com-

    ponent of the device.

    If additional requirements on accidentprevention and environmental protec-tion exist in the country of operation,

    this Instructions for Use must be sup-plemented by appropriate instructionsto ensure compliance with such re-quirements.

    Device type

    All information provided in this manual applies

    only to the instrument type indicated on the cover

    page.

    An identification label with the serial number is

    attached on the underside of the control unit.

    This instrument was built and tested in accord-

    ance with the safety regulations for electrical

    measuring, control, regulating and laboratory

    devices as specified below:

    • EN 61010-1:2001,

    • EN 61326: 1997+A1: 1998 +A2: 2001

    .

    The protective devices on both instru-ment and accessories must neither beremoved nor modified.

    Only service personnel qualified by

    Leica may repair the instrument andaccess the instrument’s internal com-ponents.

  • 8 Instructions for Use V1.2, RevB – 10/2012

    Safety measures in cryopreparation

    Provided that the safety measures outlined in the present chapter are

    observed, using all safety devices of the instrument as described, there

    is no significant accident risk involved in cryopreparation. For this rea-

    son, please make sure at all times when working with the LN22 system

    that the few, but decisive, rules for safety listed in this chapter are fol-

    lowed.

    When handling liquid nitrogen (LN2), please bear in mind the following:

    • LN2 is extremely cold and boils at -196°C. As a consequence, also ni-

    trogen gas (GN2) escapes from boiling LN

    2 at a very low temperature.

    Bear in mind, that LN2 and/or GN

    2 as well as objects which are in con-

    tact with LN2 / GN

    2 (pipe joints, valves, pipes, containers or stoppers)

    can cause severe frost bite on skin and eyes.

    • When evaporating, LN2 expands at a ratio of approximately 1 : 700,

    i.e. 1 liter of liquid nitrogen produces almost 1 m3 of nitrogen gas

    (GN2). Therefore, please make sure that the working area is kept well

    ventilated when larger quantities of liquid nitrogen evaporate (e.g.

    when pouring LN2 from one container into another). LN

    2 residues from

    containers may only be emptied outdoors: Useful for disposing of liq-

    uid nitrogen is a gravel-filled container or pit, where LN2 will evapo-

    rate quickly and safely.

    • Nitrogen gas (GN2) is odorless and has no taste so it cannot be distin-

    guished from normal air when inhaled. GN2 is nontoxic as a sub-

    stance. Still, a higher GN2 content (normal ratio: about 78 % GN

    2 to

    21 % of oxygen) in the air will cause the oxygen level to diminish, thus

    causing sudden fainting and deep unconsciousness without any pre-

    ceding warning symptoms (such as numbness, dizziness).

    In case of doubt about the oxygen level in the air it is advisable to use

    a gas analyzer with an oxygen level indication range of 0 to 25 %.

    There is danger when the oxygen level drops below 18 %.

    2. Safety

  • 9

    If an unconscious person remains in such a low-oxygen environment,

    death may occur.

    In case of emergency immediately call a physician and an ambu-lance. If breathing stops, apply artificial respiration at once.

    • For the reasons just explained, never place LN2 containers in closed

    storage rooms. Due to defects because of improper handling (jolts,

    mechanical damage) or simply by normal wear, the evaporation rate

    of Dewar vessels may rise to several liters per day. In enclosed

    spaces, this will cause a dangerous reduction of the oxygen level in

    the air.

    • Please remember that the LN22 system evaporates 2 to 3 liters of liq-

    uid nitrogen per hour during normal operation. This quantity of liquid

    corresponds to 2 to 3 m3 of GN2/h. Therefore, make sure your work

    area is kept well ventilated at all times and adhere to the precaution-

    ary measures described above.

    • Be extremely careful when bringing objects at ambient temperature

    into contact with LN2. During the first few moments an insulating layer

    of gas will form around the object. This makes any significant transfer

    of warmth impossible and so comparatively little LN2 will evaporate.

    However, as soon as the object has cooled down sufficiently there

    will often occur strong boiling and splashing of LN2 at a time when it

    is no longer expected.

    2. Safety

  • 10 Instructions for Use V1.2, RevB – 10/2012

    • There are clothing items and protective gear

    with certain characteristics which you should

    specifically avoid when working with LN2: Do

    not wear closed safety goggles (a), boots (c),shoes (e) and gloves (g) into which LN

    2

    splashes can enter, but from where they can-

    not escape. LN2 that has splashed into such

    items will evaporate immediately inside the

    item thus causing severe injuries, before the

    item can be removed or taken off. Therefore,

    always use safety goggles with side protec-

    tion but which are open at the top and at the

    bottom (b), so that LN2 splashes can escape

    immediately. Wear boots only in combination

    with long, loose-fitting (not tight!) pants that

    fall over the boots (d) and completely covertheir upper opening. Only wear open slippers

    (f) in the lab, no shoes or pumps that encloseyour feet entirely.

    In combination with open slippers/sandals only

    wear pants without cuffs, so that LN2 splashes

    will flow off freely. When pouring LN2 from one

    container into another or when inserting the

    Dewar pump into the Dewar, never wear safety

    gloves! As an alternative to gloves you can use

    an open flannel cloth (h) to protect your handswhen touching cold components. Protective

    gloves are useful when handling cold and dry

    objects, however they are not for work with LN2.

    a

    c

    b

    d

    fe

    g h

    • If your skin is injured by splashing LN2, imme-

    diately rinse the affected skin area(s) thor-

    oughly with lukewarm (body temperature!)

    water. In case of serious injuries, immediately

    see a dermatologist.

    If LN2 splashes into your eyes, rinse the

    eye immediately and thoroughly withlukewarm water. If available in your lab,use a special eye-rinsing device for thatpurpose. After rinsing, immediately seean ophthalmologist.

    Fig. 1

    2. Safety

  • 11

    • Once every three months check the actual evaporation rate of your

    Dewar vessels and compare the measuring results to the evaporation

    rates specified by the manufacturer. The evaporation rate of an un-

    damaged metal Dewar is far less than 1 liter per day. Defective

    Dewars with evaporation rates higher than that are a safety risk and

    should be taken out of service or repaired.

    • Standard Dewar vessels, such as the metal Dewar of the LN22 sys-tem, are not pressure vessels, therefore make sure they are closedonly with a special Dewar topper (Fig. 2). These stoppers fit looselyonto the Dewar neck and leave a sufficiently large gap to allow for

    sudden evaporation of GN2. Check the stopper frequently to make

    sure it will not be stuck to the Dewar neck by ice deposits (Fig. 3).Never close a Dewar vessel with stoppers that you have crafted

    yourself (Fig. 4). Through exposure to shock or vibration (e.g. trans-port), a gas-tight closed Dewar vessel can release GN

    2 in an explo-

    sion-like fashion and burst in the process. In that case the entire LN2

    content (35 liters = 35m3 GN2) will evaporate at once (that may e.g.

    happen inside a vehicle during transport). When a Dewar moves, thus

    causing a sudden evaporation of LN2, an appropriately designed stop-

    per will rise due to the gas pressure, thus opening a wide enough gap

    for the GN2 to escape (Fig. 5). The small quantities of LN

    2 which may

    escape as well in such situations do not have to be regarded as a sig-

    nificant safety risk.

    GN2

    GN2

    Fig. 4 Fig. 5

    Tightly fitted stopper

    with seal

    Fig. 2

    Loosely fitted stopper

    Fig. 3

    Ice deposits

    2. Safety

  • 12 Instructions for Use V1.2, RevB – 10/2012

    • Never leave LN2 in open containers where it can exchange with the

    room atmosphere. The boiling point of LN2 (-196°C) is lower than that

    of liquid oxygen (-183°C). If a large LN2-surface is exposed to the air,

    oxygen from the air is absorbed by the LN2 while nitrogen is released

    into the air. LN2 with a high content of liquid oxygen has a bluish color.

    In that case, caution is advised, since concentrated liquid oxygen is a

    fire hazard.

    • Make sure that your Dewar vessel is filled with no other substances

    than LN2. If at the filling location also other gases are refilled, install a

    warning sign there that clearly states: “Liquid nitrogen only”. Checkthe color of the cryogen! Bluish color indicates a high oxygen con-tent (see previous paragraph). If LN

    2 is stored during a prolonged

    period of time, liquid oxygen will accumulate in the LN2 vessel, since

    liquid oxygen has a higher boiling point (-183°C) than liquid LN2

    (-196°C).

    • Never fill a Dewar vessel up to the neck with LN2 (Fig. 6). In case of a

    boiling delay, there is danger that the rapidly expanding gas bubbles

    will force the LN2 in the neck upwards causing it to splash (Fig. 7). For

    this reason, always make sure that the LN2 surface level is below the

    neck of the vessel (Fig. 8). Also bear in mind, that in most cases thelength ‘l’ of the outer insulation differs from the length ‘L’ of the actualDewar neck.

    GN2

    LN2

    LN2 level

    Fig. 6 Fig. 7 Fig. 8

    2. Safety

    LN2 level

  • 13Leica LN22

    Warnings – Installation/Connection to mains

    The room where the instrument is installed must either be well ventilated or be large enough toavoid a dangerous decrease of the oxygen level in the air.

    Please read chapter 2 ‘Safety instructions’ carefully!

    Pay careful attention to chapter 3 - ‘Technical data’.

    The voltage selector is preset to 230V in the factory.

    Before connecting the instrument to the mains power, please check whether the factory setting

    is identical with the local mains power in your lab!

    The mains cable inlet is sealed with an adhesive tape that indicates the factory setting the in-strument is delivered with.

    Severe damage may be caused to the instrument when operating it at an incorrect voltage set-ting.

    Do not change the voltage selector setting with the instrument being connected to the mains!

    The Dewar must be externally grounded!

    The instrument may only be connected to a grounded mains power outlet socket. To connectthe instrument only use one of the mains cables that are supplied together with the instrument!

    Warnings – Maintenance

    Only authorized and qualified service personnel may access the internal components of the in-strument for service and repair!

    Before opening the instrument, switch off the mains switch and pull the mains plug!

    Before exchanging the fuses, switch off the mains switch and pull the mains plug!

    When exchanging fuses, only replacement fuses with the same specification as the originalones may be used. Refer to chapter 3 ‘Technical data’ for the correct specifications.

    Before replacing the lamp, switch off the mains switch and pull the mains plug!

    2. Safety

  • 14 Instructions for Use V1.2, RevB – 10/2012

    Warnings – working at the instrument

    During operation, metal surfaces and components inside the cryochamber can reach tempera-tures from as low as -196°C up to as hot as +110°C!

    Touching these extremely cold or hot parts can cause severe injury!

    For that reason, for any type of manipulations inside the cryochamber you may use only the

    special tools which come as part of the standard delivery and the use of which is described indetail in the present manual.

    Transfer of frozen specimens may also only be done by means of the corresponding specialtools.

    Always attach the specimen holder before inserting the knife holder!

    The light has to be on as long as the knife is inside the cryochamber!

    Exercise extreme caution when handling knives! The cutting edges are extremely sharp andcan cause severe injury!

    Never try to catch a falling knife!

    The knife may only be inserted into the knife holder shortly before sectioning and must be re-moved from the knife holder directly afterwards!

    To insert the knife holder into the cryochamber, always use the corresponding manipulator(part of standard delivery)!

    The knife edge is exposed! Risk of injury!

    Cleaning

    Do not use acetone or xylene to clean the lacquered surfaces of the instrument.

    2. Safety

  • 15

    3. Instrument components and specifications

    3.1 Technical data

    General

    Approvals: The device-specific approval symbols are located

    on the underside of the control unit next to the

    identification label.

    Supply voltage: 100-240V/47-63Hz

    Power draw: 250W

    Room temperature: up to -150°C

    Temperature stability: +/-1°C

    Nitrogen consumption

    in standby mode: approx. 1 l/day

    Dimensions and weight Control unit Cryochamber

    Dimensions: inside outside

    Width: 250mm 120mm 253mm

    Depth: 340mm 100mm 172mm

    Height: 120mm 130mm 171.5mm

    Weight: 8.2kg 7.9kg

    Fuses

    Miniature fuses (5.0 x 20mm): 220-240V: 3.15 AT

    100-120V: 6.3 AT

  • 16 Instructions for Use V1.2, RevB – 10/2012

    3.2 Overview and instrument components

    3. Instrument components and specifications

    1

    2

    5

    3

    4

    1 Cryochamber

    2 Control unit

    3 Liquid nitrogen

    pump

    4 Dewar vessel for

    liquid nitrogen on

    roller base

    5 Connection cable

    6 LN2 connection

    hose

    The liquid nitrogen freezing device LN22 quickly

    converts a rotary microtome into a cryosectioning

    system.

    The LN22 attachment consists of a cryochamber,

    which encloses the sectioning area, and which

    is refrigerated by a permanent flow of liquid ni-

    trogen, pumped from a Dewar vessel. The auto-

    matic pumping mechanism ensures a stable low-

    temperature environment around the frozen

    specimen and the knife. Separate temperature

    control systems maintain stable temperatures of

    both knife and specimen.

    The desired temperature setting for knife and

    specimen are selected on the control unit. The

    control unit, which continuously monitors and

    regulates the selected temperature parameters

    and controls the liquid nitrogen pumping system.

    Actual temperatures of knife, specimen and

    cryochamber are displayed on the control unit at

    all times.

    The cryochamber is mounted onto the microtome

    base plate, thus being completely isolated from

    the microtome feed and drive elements.

    Fig. 9

    6

  • 17

    3. Instrument components and specifications

    3.3 Standard delivery

    The Leica LN22 standard delivery includes:

    • Basic instrument (cryochamber)

    • Dewar,

    • LN2 pump,

    • Control unit

    1 Knife holder for tungsten carbide knives ................................ 0378 25655

    1 Tungsten carbide knife 40° ....................................................... 0216 08724

    1 Round specimen holder 12 mm ................................................ 0378 25653

    1 Tool for specimen holder manipulation .................................. 0378 26193

    1 Tool for knife manipulation ........................................................ 0378 26195

    1 Tool set, replacement fuses, different mains cables

    1 Set of power cords:

    - 1 Power cord ,D’ .................................................................... 0411 13558

    - 1 Power cord ‚USA-C-J‘ ....................................................... 0411 13559

    - 1 Power cord ‚UK‘ .................................................................. 0411 27822

    1 Instructions for Use - G/E/F/S ................................................... 0700 37117

    Please compare the actually delivered parts with the parts listand your order.

    Should there be any discrepancies, please contact your localLeica sales office immediately.

  • 18 Instructions for Use V1.2, RevB – 10/2012

    4.1 Installation site requirements

    - stable, vibration-free laboratory bench with horizontal and even stage

    plate; practically vibration-free floor

    - no direct insulation

    - mains power supply within a range of approx. 1.5 meters

    - no draft

    - plain floor surface

    - handwheel easily and comfortably accessible

    - 10cm minimum distance between the instrument and walls and/or

    other objects on all sides

    The room must be well ventilated and large enough to avoid a dan-gerous decrease of the oxygen level in the air.

    Carefully read the chapter on ‘Safety Instructions’!

    4.2 Unpacking

    The shipment consists of several cardboard boxes, which are all con-

    tained in one large box.

    • Remove all packing material carefully.

    If you find any damages, please contact your freight forwarder immedi-

    ately.

    4. Startup

  • 19Leica LN22

    4.3 Preparing the microtome forattachment of the LN22

    • Move the microtome specimen cylinder to the

    rear limit stop.

    • Pivot the magnifier of microscope to the side.

    • Remove knife holder basis and knife holder.

    • Remove the specimen clamp.

    • Unscrew the fixture for specimen clamps.

    • Lock handwheel (1) in 12-o’clock position (1a).• Instead of the fixture for specimen clamps,

    assemble the cryo specimen arm (3) with the4 shorter screws (standard delivery) (2).

    Fig. 11

    • Rotate the handwheel to place the cryo speci-

    men arm (3) at the low end of the cuttingstroke.

    Fig. 12

    Fig. 10

    22

    4. Startup

    1

    1a3

    3

  • 20 Instructions for Use V1.2, RevB – 10/2012

    4.4 Installing the cryochamber

    • Place the cryochamber onto the guide rails (1)on the microtome base plate and carefully in-

    sert the cryo specimen arm into the opening

    (2) in the cryochamber while sliding the cham-ber towards the microtome.

    • Push the cryochamber towards the microtomeuntil it reaches the limit stop and secure it in

    place with the clamping lever (3) on the rightside of the microtome base plate.

    Fig. 13

    1

    2

    3

    Fig. 15

    • Slide cold bridge (4) onto the cryo specimenarm and secure with the screw.

    4

    • Connect the cable (5) at the cryochamber con-nection.

    5

    Tube

    connection

    Fig. 14

    Connection

    for control

    unit

    4. Startup

  • 21

    4.5 Connecting to mains

    4.5.1 Checking the voltage selector setting

    The voltage selector is preset to 230V in the factory.Before connecting the instrument to the mains power, pleasecheck whether the factory setting is identical with the local mainspower in your lab!Severe damage may be caused to the instrument when operating itat an incorrect voltage setting.

    The voltage selector is located behind the cover of the voltage selector

    housing (3) next to the mains switch (2) on the rear of the control unit.The actual voltage setting can be seen in a small indication window (1) inthe cover (3). See the figure on the following page!

    Check that the voltage selector setting that shows in the window(1) is the appropriate one for the mains supply in your lab.

    If the voltage shown in the indication window does not correspond to

    your local mains supply, the voltage selector setting MUST be changed

    to the appropriate values for your lab prior to connecting the instrument

    to the mains supply.

    4. Startup

  • 22 Instructions for Use V1.2, RevB – 10/2012

    Fig. 16

    4.5.2 Changing the voltage selector setting and inserting the fuses

    The voltage selector can be set to 120V and 230V.

    Do not change the voltage selector setting with the instrument con-nected to the mains!

    14

    1

    5

    139

    11

    10 7

    6

    7

    7

    7

    7

    7

    7

    78

    15

    3

    16

    3

    4

    2

    12

    4. Startup

  • 23

    • Place the tip of the screwdriver into the small recess on the left of

    cover (3) and push very carefully, using the screwdriver as a lever, toremove the housing.

    • Pull out the housing (3) together with the fuses (4) and the voltageselector insert (2).

    • Remove the voltage selector insert (2) from the housing (3).• Insert 2 fuses (4) (both of the same specification) which are appropri-

    ate for the voltage in your lab (see instrument name plate for corre-

    sponding fuse value).

    • Put the voltage selector insert (2) back into the housing (3), so thatthe desired voltage shows in the indication window (16) of the hous-ing.

    • Reinsert the housing (3), together with voltage selector and fuses,into the instrument and push lightly until it locks in place.

    • Check whether the correct voltage now shows in window (16).

    4.5.3 Plugging in the connection cables

    • Connect the cryochamber (15) to the control unit (14) with cable (6)and secure both plugs with screws (7).

    • Connect the cryo specimen arm with cable (8) to the cryochamber(15) and secure both plugs with screws (7).

    • Place the Dewar vessel (10) onto the roller base (11).• Hang up the supporting device (12) for the LN

    2 pump (13), either in the

    arched handle of the Dewar vessel or on the wall, using the supplied

    dowels and screws.

    • Place the LN2 pump onto the supporting device and connect the

    pump cable (9) to the control unit; secure the plug with screws (7).

    4.5.4 Grounding the Dewar

    The Dewar must be externally grounded!

    For safety reasons, metallic surfaces which contain electronic components have to be grounded. In

    the absence of appropriate grounding, housing parts can be above the earth potential and thus dan-

    gerous when touched.

    4. Startup

  • 24 Instructions for Use V1.2, RevB – 10/2012

    4.5.5 Connecting the mains cable

    The instrument is delivered with various country-specific mains ca-bles.

    The instrument may only be connected to a grounded mains poweroutlet socket. To connect the instrument, only use one of the mains

    cables that are supplied together with the instrument!

    • Before plugging in the mains cable, check whether the mains switch

    (1, Fig. 16) on the rear side of the control unit (14, Fig. 16) is in ‘O’ (OFF)position.

    • From the set of mains cables, select the one with the plug that fits into

    your wall outlet.

    • Insert the connector plug of the mains cable into socket (5, Fig. 16).

    • Plug the mains plug into the wall outlet.

    The instrument is now ready for use.

    4.6 Switching on the instrument

    Switching on the cryochamber illumination

    The switch (1) for the cryochamber illumination islocated on the left side of the cryochamber.

    The light has to be on as long as the knifeis inside the cryochamber!

    1

    Fig. 17

    4. Startup

    • Switch on the mains switch at the rear of the

    control unit (1, Fig. 16).

    First, the software version will be displayed for a

    few seconds. Then the actual temperature of

    specimen (1, Fig. 18 a), knife (2, Fig. 18 b) andchamber gas (3, Fig. 18c) is displayed.

  • 25Leica LN22

    ‘TEMPERATURE’ control panel

    • To indicate the selected temperature value,

    press the + / - button (4) below the tempera-ture indication briefly (less than four seconds).

    • To alter the selected temperature value, press

    the + / - button (4) for a longer period of time.

    The temperature measuring unit (K, °C, °F) (12) ischanged via the software. See ‘Control panel -

    OPERATION’ on how to actually change the tem-

    perature indication.

    ‘MEMORY’ control panel

    Parameters that are used often can be saved:

    • To save these parameters, press STO (5) and

    one of the 3 storage buttons (6, TA, TB, TC).

    • To activate the individual saved parameters,

    press the corresponding storage button (6, TA,

    TB, TC).

    4.7 Control unit

    Fig. 18

    Fig. 18a

    Fig. 18b

    1 2 3

    412

    5

    6

    Key combination:Knife + specimen

    4. Startup

    +

  • 26 Instructions for Use V1.2, RevB – 10/2012

    The selected temperature parameters are dis-

    played for about 4 seconds. Then the display

    switches back to actual temperature.

    ‘OPERATION’ control panel

    With the LN2-STOP button (7), the LN

    2 pump can

    be stopped at any time, e.g. for refilling the Dewar.

    The STOP-Alarm button (8) is to switch off thebeep signal, which sounds either when the filling

    level of the Dewar falls below 20% of full or when

    the lowest LN2 level in the cryochamber is

    reached.

    Button H (9) activates chamber heating.

    Fig. 18c

    7 8 9 11

    10

    The flow diagram indicates the LN2 filling level of Dewar vessel and

    cryochamber (10). When the lowest filling level inside the Dewar isreached, the pump is automatically switched off.

    • To change the unit of measurement that is displayed (K, °C, °F) switch

    on the control unit, immediately press the START button (11) and keepit pressed down for at least 7 seconds.

    • With the “+” button of the knife temperature indication (2), situated inthe ‘TEMPERATURE’ control panel, you can change between 1 (K),

    2 (°C) and 3 (°F) (Fig.18a, 12).

    • To exit the selection mode press the H-button (9).

    4. Startup

  • 27Leica LN22

    4.8 Specimen holders

    During operation metal surfaces and components inside thecryochamber can reach temperatures from as low as -196°C up to

    as hot as +110°C!Touching these extremely cold or hot parts can cause severe in-jury!For that reason, for any type of manipulations inside the

    cryochamber you may use only the special tools which come aspart of the standard delivery and the use of which is described indetail in the present manual. Transfer of frozen specimens may alsoonly be done by means of the corresponding special tools.Always attach the specimen holder before inserting the knifeholder!

    1 Freezing station

    2 Cryo specimen arm

    3 Clamping screw of the cold bridge

    4 Cold bridge

    5 Specimen holder

    4.8.1 Specimen holder for blockembeddings

    • Put the specimen together with appropriate

    embedding medium (e.g. 2.3 molar saccharose

    solution) into the specimen holder.

    4 3

    1

    5

    2

    Fig. 19

    Fig. 20

    OCT is very rarely suitable for theselow temperatures!

    4. Startup

  • 28 Instructions for Use V1.2, RevB – 10/2012

    Fig. 21

    Fig. 22

    • Lock the handwheel of the microtome in

    12-o’clock position.

    • Place the specimen holder into the freezingstation, as shown in.

    Extra Fine Screw Thread! Do nottighten too much!

    • Once the embedding medium is thoroughly fro-

    zen, use the special forceps with small radius

    (20) to insert the specimen holders into thecryo specimen arm (2) and tighten with theclamping screw (3) of the cold bridge (4).

    2

    4 3

    20

    4. Startup

  • 29Leica LN22

    4.8.2 Universal specimen holder and flat specimen holder

    • Place the specimen into the specimen holder

    (5) and (6) slightly tighten screw (Use specialtool!)

    • Use the special forceps with large radius to

    insert the specimen holder into the cryo speci-

    men arm and tighten loosely with screw (3) ofcold bridge (4).

    Fig. 23

    6

    5

    • Release handwheel lock.

    • Place the specimen cylinder into the lowest possible position to en-

    sure a fast freezing process.

    • Once the desired temperature has been reached, bring the specimen

    cylinder back in the upper limit position.

    • Lock the handwheel.

    • Tighten screw (6) in specimen holder (5).

    • Loosen screw (3, Fig. 22) of cold bridge (4, Fig. 22) from the cryospecimen arm, orient the specimen holder into the desired cutting

    direction.

    • Tighten screw (3) or cold bridge (4) thoroughly.

    Extra Fine Screw Thread! Do not tighten too much!

    4. Startup

  • 30 Instructions for Use V1.2, RevB – 10/2012

    1 Cryo specimen arm

    2 Cold bridge

    3 Clamping screw of the cold bridge

    4 Specimen holder

    5 Knife holder

    6 Knife clamping screws

    7 Knife holder clamping screw

    8 Temperature sensor for cryochamber

    temperature

    9 Knife holder base

    10 Clamping screw for knife holder base

    11 Cryochamber base plate

    12 Eccentric disc for lateral adjustment of

    the knife holder

    13 Freezing station

    14 Bottom plate of LN2 trough

    15 Cable loop

    16 Clearance angle adjustment

    Fig. 24

    4. Startup

    16 14 2 3

    13 12 6

    1

    4

    8

    7510911

    15

    4.9 Knife holders

    4.9.1 Knife holders for tungsten carbide knives

    Inserting the knife

    Exercise extreme caution when handling knives! The cutting edgesare extremely sharp and can cause severe injury!The knife may only be inserted into the knife holder shortly beforesectioning and must be removed from the knife holder directly af-terwards!

    The knife edge is exposed! Risk of injury!

  • 31

    Inserting the knife holder

    Always attach the specimen holder before inserting the knife

    holder to avoid damaging the knife!To insert the knife holder into the cryochamber, always use the ap-propriate manipulator, which comes as part of standard delivery!

    • Place knife holder (2) using manipulator (4) diagonally from aboveonto knife holder base (9, Fig. 24). Always use the manipulator in or-der to avoid injury.

    • Set the desired clearance angle (see Fig. 24).

    Adjusting the clearance angle

    The clearance angle is adjusted with the no. 3 Allen key, through the

    drillings (16, Fig. 24) in the clearance angle adjusting on the left of theknife holder.

    Limit stop = 2° effective clearance angle 1st scale mark: 5°

    2nd scale mark: 10°

    3rd scale mark: 15°

    • Fasten knife holder (5, Fig. 24) in knife holder base (9, Fig. 24) withscrews (6, Fig. 24).

    • Once the selected temperature has been reached, retighten the

    screws (7, Fig. 24), on the knife holder (5, Fig. 24), and on the knifeholder base (10, Fig. 24) as well as the knife clamping screws(6, Fig. 24).

    4. Startup

    • Move the specimen cylinder to the rear limit stop.

    • Insert knife (1) in the center of knife holder (2); tighten clamping screws(3). If the knife is moved to the left or right edge, tighten only the centerand left or right screw (place remaining screws and tighten slightly).

    1

    4

    2

    3

    Fig. 25

  • 32 Instructions for Use V1.2, RevB – 10/2012

    4. Startup

    Fig. 26

    Adjusting the knife holder laterally:

    • Release the clamping screws (10, Fig. 24) of the knife holder base(9, Fig. 24).

    • Displace the knife holder laterally via the eccentric disc (12, Fig. 24).

    Changing the knives inside the cryochamber

    • Prior to changing knives, move the cryo specimen arm (1, Fig. 24) firstto the rear limit stop and next to the lower limit position.

    • Lock the handwheel.

    • Using special forceps (Fig. 26) knives can be exchanged inside the

    cryochamber at very low temperatures.

    4.9.2 Glass knife, CryoDiamond knife, trimming knife Histo

    Fig. 27

    • Insert the cryo-diamond (8) or glass knife in the knife holder (9); tightenclamping screws (10).

    • See ‘Knife holder for tungsten carbide knives’ on how to proceed.

    The trimming knife kit can be used to trim lateral specimen edges parallel

    before sectioning with glass or diamond knives.

    • It is possible to exchanges the knives inside the cryochamber at very

    low temperatures using the standard forceps. Prior to exchanging

    the knives, displace the cryo specimen arm to the rear limit stop.

    • Lock the handwheel.Fig. 28

    10

    10

    9

    8

    Manipulator

  • 33

    4.10 Preparing the pumping system

    Use caution when handling liquid nitrogen!

    Prior to filling the Dewar vessel with LN2, please refer to the safety

    instructions in chapter 2.

    • Slowly lower the pump into the Dewar vessel if the liquid boils over

    and LN2 escapes from the Dewar, lift the pump partly back out until

    the boiling over stops, then continue to lower the pump carefully until

    it is completely immersed into the Dewar.

    The LN2 connecting pipe consists of an outer insulation tube (1) with a

    polyamide filling tube (2) inside.

    • Connect the polyamide tube (2) to the cryochamber (Fig. 29a), andpress the insulation tube (1) completely into the corresponding open-ing in the cryochamber wall.

    • Place the Dewar vessel close.

    • Open the protective lid (3) and connect the other end of the polya-mide tube to the pump (Fig. 29b).

    The polyamide and insulation tube are both delivered overlength and can

    be shortened according to the needs at the installation site. Always

    make sure, however, that both ends of the polyamide tube are covered

    entirely by the outer insulation tube.

    4. Startup

    2

    1

    2

    1

    3

    Fig. 29

    a

    b

    4.11 Refrigerating

    • Before starting with your cryopreparation work, set the desired tem-

    perature values for specimen and knife.

    • Start the pump by pressing the START button (11, Fig. 18c).

    • Move the cryospecimen arm to the lowest position to ensure the

    quickest possible refrigeration time.

    The reservoir in the cryochamber is filled up with LN2. Once the appropriate level is reached, the filling

    level is maintained constant by means of a level sensor which controls the LN2 pump. When the speci-

    men temperature has reached the selected value, the motorized sectioning function of the microtome

  • 34 Instructions for Use V1.2, RevB – 10/2012

    4.12 Refilling the Dewar vessel

    Use caution when handling liquid nitrogen!Please refer to the safety instructions in chapter 2.

    can be started to help reach a stable specimen temperature through the

    regular immersion of the cold bridge of the cryo specimen arm into the

    liquid nitrogen. The LED above the START button flashes until the knife

    temperature has reached the preselected value.

    If you want to change the temperature during a work session, e.g. from

    -70°C to -120°C, it is recommendable to press the START button once

    again as this ensures that the new temperature is reached as fast as

    possible, since otherwise the instrument will approach the new tempera-

    ture value only very slowly.

    The cryochamber of the LN22 liquid nitrogen freezing device is an open

    chamber system, which can be covered with a lid when relative air hu-

    midity is high or during breaks. It is advisable to cover the chamber dur-

    ing the refrigeration process and reopen it before starting to section. An

    appropriate plastic lid for the chamber comes as part of standard deliv-

    ery.

    While not in operation or while collecting the specimens, alwaysmove the cryo specimen arm to the lowest position to maintain theselected specimen temperature at a constant level.

    The 25 liter Dewar vessel contains enough liquid

    nitrogen to work without interruption for more

    than 8 hours. We advise you to leave the pump in

    the Dewar until the nitrogen content is exhausted.

    The liquid nitrogen consumption of the pump while

    not in use is less than 1 liter / day.

    The Dewar can be filled in two different ways:

    Press the LN2 STOP button (7, Fig. 18c) and wait a

    few minutes until the pump tubing has thawed and

    can thus be bent without difficulty:

    Do NOT remove the LN2 connecting tube from the

    pump. This way you avoid formation of conden-

    sation water inside the upper pump valve.

    1. Slightly lift the pump and refill the Dewar from

    a LN2 storage tank without removing the pump

    entirely.

    4. Startup

  • 35Leica LN22

    2. Remove the pump from the Dewar, cover the

    pump immediately with the yellow protectioncap, place it on the corresponding supportingdevice and refill the Dewar. Before the pump

    is re-immersed into the Dewar it has to be thor-

    oughly dry.

    This second method is very time-consuming

    and can take up to several hours. To simplify

    the process the use of a second Dewar ves-

    sel is recommended. Once a Dewar is empty,

    immediately place it into the second (full)

    Dewar.

    The pump must always be held in UP-RIGHT position to prevent melt water andhumid air from entering the pump valve.

    After taking the pump out of the Dewar, immedi-

    ately cover the Dewar and cover the pump with

    the yellow protective cap.

    Humidity will cause icing of the pump valves dur-

    ing the following refrigeration process thus inter-

    fering with the pumping process. Pumping trou-

    ble is easy to recognize: in spite of the pumping

    motor being activated, no nitrogen will be pumped

    from the Dewar. To eliminate this problem, please

    proceed as described in chapter 5.2.4.

    4.13 Heating up the chamber

    During operation, metal surfaces and components inside thecryochamber can reach temperatures from as low as -196°C up to

    as hot as +110°C! Touching these extremely cold or respectivelyhot parts can cause severe injury!For that reason, for any type of manipulations inside thecryochamber you may use only the special tools which come aspart of the standard delivery and the use of which is described indetail in the present manual.Transfer of frozen specimens may also only be done by means ofthe corresponding special tools.

    • Remove specimens and cryo-tools from the

    cryochamber.

    • Remove sectioning waste since it will fuse

    during heating.

    • Press button (H) on the control unit (9, Fig. 18c).

    After pressing button (H) the temperature of both

    knife and specimen holder will rise quickly to ap-

    proximately +100°C. After about 45 minutes the

    heating process is completed.

    At that point, any residual water inside the cham-

    ber should have been evaporated.

    The control unit automatically switches off the

    heating function once a certain chamber tem-

    perature has been reached. This means that the

    instrument can be left in the H-mode overnight.

    The pump can remain in the Dewar vessel due to

    its low liquid nitrogen consumption while not in

    operation.• To restart the instrument, simply press the

    START button (11, Fig. 18c).

    4. Startup

  • 36 Instructions for Use V1.2, RevB – 10/2012

    5.1 Cleaning

    Do not use acetone or xylene to clean the lacquered surfaces of the

    instrument.

    The cryochamber should be cleaned in regular intervals, especially

    when large quantities of sectioning debris have been produced during a

    work session.

    Since sectioning debris (especially very soft rubber samples, such as

    bitumen) melts during the chamber heating process, we advise you to

    clean the chamber prior to the heating process (however, after all re-

    maining LN2 has evaporated / the chamber has reached ambient tem-

    perature) with a vacuum cleaner or similar tool.

    5. Cleaning and maintenance

    For cleaning, proceed as follows:

    • Remove knife, knife holder, specimen holder

    and cold bridge.

    • Displace the cryo specimen arm to the rear

    limit stop and lower end position.

    • Release the clamping lever (1) on the micro-tome base plate.

    • Carefully pull the chamber forward towards

    yourself until the cryo specimen arm (2) is out-side the chamber.

    • Retighten the clamping lever (1) on the micro-tome base plate.

    1

    2

    • To remove the knife holder system (9), detachboth screws (10).

    • Remove both screws together with the corre-

    sponding washers, using forceps.

    Fig. 30

    Fig. 319

    10

    10

  • 37Leica LN22

    • Vacuum sectioning debris that sticks to cham-

    ber walls and bottom or remove it with ben-

    zine.

    Caution: flammable!

    • All removable parts can be cleaned in an ul-

    trasonic bath.

    • Carefully pull out the cable loop (15) far enoughso that the knife holder base can be placed

    upright on the bottom plate of the

    cryochamber (11).

    • Remove the eccentric disc (12).

    • Detach the 3 screws (17) of the cryochamberbottom plate.

    Fig. 32

    9

    17 17

    111712 15

    Fig. 3311

    9 14

    • Remove cryochamber bottom plate (11) andknife holder base (9).

    • Store the knife holder base on the chamber

    wall.

    • Use the manipulator to remove the bottom

    plate (14) of the LN2 trough.

    21

    Fig. 34

    5. Cleaning and maintenance

  • 38 Instructions for Use V1.2, RevB – 10/2012

    5. Cleaning and maintenance

    • Use one of your hands to hold the cable loop

    on the chamber bottom while inserting the

    knife holder base together with the base plate

    with your other hand. Pay careful attention

    that the cable loop is not crushed between

    bolts (21, Fig. 35) and the chamber bottomplate.

    Fig. 36

    Fig. 37

    9

    17

    17

    111712 15

    • Unscrew knife holder base (9) again, place onedge in the chamber, fasten the chamber bot-

    tom plate (11) with the screws (17).

    • Place eccentric disk (12) in the recess of thebottom plate.

    • Loosely attach the chamber bottom plate (11)to the knife holder base (9) with screws(10, Fig. 31).

    • Reinsert the bottom plate (14) of the LN2 trough

    into the chamber using the manipulator.

    Fig. 35

    10

    • Insert the two screws with the washers in the

    drillings (10) and screw the knife holder base(9) onto the bottom plate.

    Fig. 38

    17 17

  • 39Leica LN22

    5.2 Maintenance

    5.2.1 General maintenance instructions

    Only authorized Leica service engineers may open the instrument

    for maintenance and repair work!Before opening the instrument, switch off the mains switch and pullthe mains plug!

    The instrument is practically maintenance-free. However, to ensure

    smooth and trouble-free operation of the instrument over a long period

    of time, we recommend the following preventive maintenance steps:

    • Have the instrument inspected o n c e a year by an authorized Leica

    Technical Service engineer.

    • Ask for a Service Contract after the warranty period expires.

    For details, please contact your local Leica sales office.

    • Never attempt any repairs on the instrument yourself since this will

    invalidate the warranty.

    5.2.2 Exchanging the mains fuses

    Before exchanging the fuses, switch the instrument off and pull themains plug!When exchanging fuses, only use replacement fuses with the samespecification as the original ones. Refer to chapter 3 ‘Technicaldata’ for the correct specifications.

    5. Cleaning and maintenance

    The mains fuses are located in the voltage selec-

    tor housing (3).• Remove the voltage selector housing (3) as

    described in chapter 5.5.2.

    • Remove the fuses (4) and insert new fuses ofthe same technical specification.

    • Reinsert the housing together with the insert

    (2) and the fuses into the corresponding open-ing in the instrument housing and push lightly

    to lock it in place.

    • Check again whether the correct voltage

    shows in the indication window (16).Fig. 39

    14

    2

    4

    1 16

    3

    16

  • 40 Instructions for Use V1.2, RevB – 10/2012

    5.2.3 Exchanging the bulb

    Before replacing the lamp, switch off the mains switch and pull themains plug!

    The bulb of the chamber illumination is located

    behind a cover on the left side of the cryocham-

    ber.

    • Remove the two screws (1) of the cover (2)with 2.5 screwdriver.

    Fig. 40

    Fig. 41

    • Carefully pull out the cover (2) together withthe cable attached to it on the inside just

    enough to be able to reach the lamp.

    • Remove the lamp (3) and insert a new lamp ofthe same specification.

    • Put the cover (2) back in its place and securewith the screws (1).

    3

    2

    2

    1

    5. Cleaning and maintenance

  • 41

    Fig. 42

    2

    13

    5.2.4 Removal of the pump valves

    The two lower pump valves (1) and (2) can be easily disassembled forthorough drying. Valve (2) is located above valve (1).

    • For disassembly, support the valves (1) and (2) valve block at the pipeend with an Allen key no. 3 at the screw (3) and unscrew the brassvalves with a screwdriver.

    • Dry the valves on a heating panel.

    • Before reinstallation, check whether the ball moves inside the valve

    by shaking it.

    5. Cleaning and maintenance

  • 42 Instructions for Use V1.2, RevB – 10/2012

    6.1 Ordering information

    Knife holders

    Knife holder for triangle glass knife and diamond knife .............................................................. 0378 26169

    Knife holder for tungsten carbide knives ....................................................................................... 0378 25655

    Knives

    Tungsten carbide knife 40°, 30mm .................................................................................................. 0216 08724

    Tungsten carbide knife 50°, 30mm .................................................................................................. 0216 08728

    Trimming knife (usable only together with adapter 0378 26875) ................................................ 0216 25654

    Holder for trimming knife .................................................................................................................. 0378 26875

    Cryo-diamond knife, 4mm, 45°, with truff ....................................................................................... 0366 26427

    Cryo-diamond knife, 4mm, 45°, without truff ................................................................................. 0366 26428

    Cryo-diamond knife, 6mm, 45°, with truff ....................................................................................... 0366 26431

    Cryo-diamond knife, 6mm, 45°, without truff ................................................................................. 0366 26432

    Specimen holders

    EM Universal specimen holder, for RM 21 series and RM 22 series ........................................ 0356 10868

    EM Flat specimen holder .................................................................................................................. 0355 10405

    Special wrench for EM specimen holder 0356 10868 and 0355 10405 ....................................... 0356 10869

    Round specimen holder for round specimen blocks, dia. 12mm ............................................... 0378 25653

    Accessories

    Ionization device ............................................................................................................................ upon request

    Holding pliers for tungsten carbide knives ................................................................................... 0378 26195

    Holding pliers for round specimen holder, dia. 6 and 12mm ....................................................... 0378 26193

    Holding pliers for EM specimen holder .......................................................................................... 0378 26194

    Spare halogen bulb for backlighting system ................................................................................. 0378 32400

    Instructions for Use V1.1 – 06/2006 (German, English, French, Spanish) .................................. 0700 37117

    6. Appendix

  • 43Leica LN22

    7. Warranty and service

    Warranty

    Leica Biosystems Nussloch GmbH guarantees that the contractual

    product delivered has been subjected to a comprehensive quality con-

    trol procedure based on the Leica in-house testing standards, and that

    the product is faultless and complies with all technical specifications

    and/or agreed characteristics warranted.

    The scope of the warranty is based on the content of the concluded

    agreement. The warranty terms of your Leica sales organization or the

    organization from which you have purchased the contractual product

    shall apply exclusively.

    Service information

    If you are in need of technical customer service or spare parts, please

    contact your Leica representative or the Leica dealer where you pur-

    chased the unit.

    Please provide the following information:

    • Model name and serial number of the instrument.

    • Location of the instrument and name of the person to contact.

    • Reason for the service call.

    • Delivery date

    Decommissioning and disposal

    The instrument or parts of the instrument must be disposed of in compli-

    ance with the local laws.

  • 44 Instructions for Use V1.2, RevB – 10/2012

    Notes

    1. Important information 1.1 Symbols in the text and their meanings 1.2 Intended use of instrument 2. Safety 2.1 Safety instructions 3. Instrument components and specifications 15 3.1 Technical data 3.2 Overview and instrument components 3.3 Standard delivery 4. Startup 4.1 Installation site requirements 4.2 Unpacking 4.3 Preparing the microtome for attachment of the LN22 4.4 Installing the cryochamber 4.5 Connecting to mains 4.5.1 Checking the voltage selector setting 4.5.2 Changing the voltage selector setting and inserting the fuses 4.5.3 Plugging in the connection cables 4.5.4 Grounding the Dewar 4.5.5 Connecting the mains cable 4.6 Switching on the instrument 4.7 Control unit 4.8 Specimen holders 4.8.1 Specimen holder for block embeddings 4.8.2 Universal specimen holder and flat specimen holder 4.9 Knife holders 4.9.1 Knife holders for tungsten carbide knives 4.9.2 Glass knife, CryoDiamond knife, trimming knife Histo 30 4.11 Refrigerating 4.12 Refilling the Dewar vessel 4.13 Heating up the chamber