70 February 2011 healthVIE.com Professional Education & Training Load Control Monitoring Using Process Challenge Devices ANSI/AAMI ST79: 2010 update by Martha Young, BS, MS, CSPDT 2. A process challenge device (PCD) is designed to create a challenge to the sterilization process that represents the most difficult item routinely processed. A. True B. False 3. A PCD may include a biological indicator (BI), a BI and a Class 5 integrating indicator, a Class 5 integrating indicator or a Class 6 emulating indicator. A. True B. False 4. The BI PCDs for table-top sterilizers and flash sterilization cycles are user-assembled and should be representative of the load. A. True B. False 5. Using a BI PCD or a CI PCD in each load is optional for routine load release. A. True B. False 6. A BI PCD with a Class 5 integrating indicator should be used to monitor each load containing an implant and the load should be quarantined until the BI result is available. A. True B. False 7. Steam sterilization processes should be routinely monitored with a BI PCD weekly, preferably daily (each day the sterilizer is used). A. True B. False 8. Ethylene oxide sterilization processes should be routinely monitored with a BI PCD in each load. A. True B. False 9. Hydrogen peroxide and ozone sterilization processes should be monitored with a BI PCD at least daily, but preferable in every sterilization cycle. A. True B. False 10. When a BI PCD, a CI PCD, or other CIs, or physical monitors detect a sterilization process failure the load should be quarantined, the sterilizer removed from service, an investigation of the cause initiated, and all loads back to the last negative BI recalled if the reason for the cause of the failure is not immediately identified (usually an operator error). A. True B. False Many thanks to the team at 3M Health Care for working with healthVIE.com to provide the following accredited course. IAHCSMM has awarded 1.5 contact points for completion of this continuing education lesson toward IAHCSMM recertification. The CBSPD has preapproved this inservice for 1.5 contact hours for a period of five (5) years from the date of publication, and to be used only once in a recertification period. This inservice is 3M Health Care Provider approved by the California Board of Registered Nurses, CEP 5770 for 1 contact hour. This form is valid up to five (5) years from the date of publication. Instructions for submitting results are on page 94. healthVIE.com and 3M Health Care will be working collaboratively to provide continuing education courses in monthly editions of healthVIE.com. Objectives After completion of this self-study activity, the learner will be able to: 1. Define load control monitoring and process challenge devices. 2. Develop policies and procedures for using biological indicator process challenge devices for routine release of nonimplant and implant loads. 3. Develop policies and procedures for using biological indicator process challenge devices for routine sterilizer efficacy monitoring. 4. Discuss the value of using a biological indicator process challenge device in every load for load control monitoring. Test Questions 1. The advantage of using a biological indicator process challenge device (BI PCD) in each load for load control is to reduce the risk and cost of healthcare-associated infections (HAIs) by not releasing the loads until the BI results are negative. A. True B. False

Upload: lamnhan

Post on 18-Apr-2018

224 views

Category:

Documents


2 download

TRANSCRIPT

70 February 2011healthVIE.com

Professional Education & Training

Load Control Monitoring Using Process Challenge Devices

ANSI/AAMI ST79: 2010 update by Martha Young, BS, MS, CSPDT

2. A process challenge device (PCD) is designed tocreate a challenge to the sterilization process thatrepresents the most difficult item routinely processed.A. True B. False

3. A PCD may include a biological indicator (BI), a BIand a Class 5 integrating indicator, a Class 5 integrating indicator or a Class 6 emulating indicator.A. True B. False

4. The BI PCDs for table-top sterilizers and flash sterilization cycles are user-assembled and should berepresentative of the load.A. True B. False

5. Using a BI PCD or a CI PCD in each load is optionalfor routine load release.A. True B. False

6. A BI PCD with a Class 5 integrating indicator shouldbe used to monitor each load containing an implantand the load should be quarantined until the BI resultis available. A. True B. False

7. Steam sterilization processes should be routinelymonitored with a BI PCD weekly, preferably daily(each day the sterilizer is used).A. True B. False

8. Ethylene oxide sterilization processes should beroutinely monitored with a BI PCD in each load.A. True B. False

9. Hydrogen peroxide and ozone sterilization processesshould be monitored with a BI PCD at least daily, but preferable in every sterilization cycle.A. True B. False

10. When a BI PCD, a CI PCD, or other CIs, or physicalmonitors detect a sterilization process failure the loadshould be quarantined, the sterilizer removed fromservice, an investigation of the cause initiated, and allloads back to the last negative BI recalled if thereason for the cause of the failure is not immediatelyidentified (usually an operator error).A. True B. False

Many thanks to the team at 3M Health Care forworking with healthVIE.com to provide the followingaccredited course. IAHCSMM has awarded 1.5 contactpoints for completion of this continuing education lessontoward IAHCSMM recertification. The CBSPD haspreapproved this inservice for 1.5 contact hours for aperiod of five (5) years from the date of publication, andto be used only once in a recertification period. Thisinservice is 3M Health Care Provider approved by theCalifornia Board of Registered Nurses, CEP 5770 for 1contact hour. This form is valid up to five (5) years fromthe date of publication. Instructions for submittingresults are on page 94.

healthVIE.com and 3M Health Care will be workingcollaboratively to provide continuing education coursesin monthly editions of healthVIE.com.

ObjectivesAfter completion of this self-study activity, the learner will

be able to: 1. Define load control monitoring and process challenge

devices.2. Develop policies and procedures for using biological

indicator process challenge devices for routine release ofnonimplant and implant loads.

3. Develop policies and procedures for using biological indicator process challenge devices for routine sterilizerefficacy monitoring.

4. Discuss the value of using a biological indicator processchallenge device in every load for load control monitoring.

Test Questions1. The advantage of using a biological indicator process

challenge device (BI PCD) in each load for load control is to reduce the risk and cost of healthcare-associated infections (HAIs) by not releasing the loads until the BIresults are negative.A. True B. False

IntroductionLoad control monitoring is the process by which a load is

monitored and released based on the result of a biological indi-cator (BI) in a process challenge device (PCD). Only a BI candetect the actual killing of microbial spores by the sterilizationprocesses. (ANSI/AAMI ST79 Section 10.5.3.1)1

Using a self-contained biological indicator with a one-,three- or four-hour readout in a BI PCD in each load allowsquarantining of all loads pending BI results, especially thosecontaining implants, and reduces variability and chance forerrors. The advantages of using a BI PCD in each load for loadcontrol include: Reducing the risk and cost of healthcare-associated

infections (HAIs) because you know the load is notreleased until the BI is negative;

Providing a universal standard of patient care by treatingall loads the same with the equal chance of identifying asterilization process failure before the patient becomesinvolved;

Detects and controls variability in the process; Reducing the cost and impact of a recall; To be certain all implants, including those in loaners, are

appropriately monitored and quarantined until the BI isnegative;

Ensure every type of steam sterilization cycle used ismonitored; and

Ensure every type of packaging used in flash or immediateuse sterilization is monitored.

For monitoring of nonimplant loads, a Class 5 integratingindicator or Class 6 emulating indicator PCD may be used asload control by providing “additional information about theattainment of the critical parameters of the sterilizationprocess.” (ANSI/AAMI ST79 Section 10.5.2.1 and 10.5.3.2)1

Using a CI PCD does not provide the same information asusing a BI PCD because the performance of Class 5 integratingand Class 6 emulating CIs “do not contain spores and do notdirectly measure the lethality of a sterilization cycle.” (ANSI/AAMI ST79 Section 10.5.3.2)1

This inservice will discuss: Process challenge devices; BI PCDs used for:

Routine release of nonimplant loads;Routine release of implant loads;Routine sterilizer efficacy monitoring;

What to do when the load control monitoring devicedetects a problem.

Recommended practices from the Association for theAdvancement of Medical Instrumentation (AAMI) and theAssociation of periOperative Registered Nurses (AORN) were

used to prepare this inservice. This inservice updates the infor-mation in Load Control Monitoring Using Process ChallengeDevices published in healthVIE.com June 2010 by providinginformation on the usage of Class 6 emulating indicators.

Process Challenge DevicesANSI/AAMI ST79 defines in Section 2.100 a process

challenge device (PCD):“Item designed to constitute a defined resistance to a

sterilization process and used to assess performance of theprocess.”1

This PCD should provide “a challenge that is equal to or greater than the challenge posed by the most difficult item routinely processed.” (ANSI/AAMI ST79 Section10.5.4)1 The type of PCD used will depend upon the type ofsterilization process, cycle, and parameters and the type of loadbeing monitored. ANSI/AAMI ST79 Section 10.5.4 states aPCD for steam sterilization processes may include thefollowing monitors: A BI; A BI and a Class 5 integrating indicator; A Class 5 integrating indicator; A Class 6 emulating indicator.1

A PCD may be user-assembled or a commercially available, disposable, pre-assembled pack. These commer-cially available BI PCDs should be FDA cleared and theirperformance should be comparable to the published standarduser-assembled or reference BI PCD described in ANSI/AAMIST79 in Section 10.7.2.1 for steam sterilizers larger than 2cubic feet and in ANSI/AAMI ST41 in Section10.7.2 forethylene oxide (EO) sterilizers.1,2 Presently there are nopublished standard user-assembled or reference BI PCDs forhydrogen peroxide or ozone sterilization. (AAMI TIR31Section 6.4 and 6.5)3

The BI PCDs for table-top sterilizers (ANSI/AAMISection 10.7.3) and flash or immediate use steam sterilization(ANSI/AAMI Section 10.7.4) are user-assembled and shouldbe representative of the load.1

Practical Application• All BI PCDs should contain the appropriate BI

and CI for the sterilization process and cyclebeing tested.

• All BI PCDs should be representative of the loadcontents and create a challenge equal to or greaterthan the most difficult item processed.

• The instructions for use from the manufacturer ofthe BI PCD and CI PCD should be followed toensure accurate results.

72 February 2011healthVIE.com

Professional Education & Training

Practical Application• Use physical monitors, chemical and biolog-

ical indicators to release nonimplant loads.• Every load monitoring with a BI PCD and

quarantining until the BI is negative ensuresthat improperly sterilized medical devicesare not released for patient use.

• Competency testing of employees ensurescorrect interpretation of monitoring tools.

Routine Load Release of Implant LoadsRoutine release of implant loads should also be an

active decision based on the evaluation of all availabledata. An experienced and knowledgeable person shouldmake that decision at the end of the steam sterilizationcycle after evaluating the: Physical monitors; External chemical indicators; Internal chemical indicators; BI PCD containing a Class 5 integrating indicator.

(ANSI/AAMI ST79 Section 10.4)1

AAMI and AORN state that the load containing theimplant should be quarantined until the results of the BItesting are available. (ANSI/AAMI ST79 Section 10.6.3,ANSI/AAMI ST41 Section 10.6.3, ANSI/AAMI ST58Section 9.5.4.3, and AORN Sterilization XVI)1,2,4,5 Thisrecommendation is based on reducing the patient riskassociated with implants.

For steam sterilization ANSI/AAMI ST79 states inSection 10.6.3 in a documented emergency situation itmay be necessary to release the implant based on theresults of the Class 5 integrating indicator in the BI PCD.1This information should be fully traceable to the patient.“Releasing implants before the BI results are known isunacceptable and should be the exception, not the rule.”(ANSI/AAMI ST79 Section 10.6.3)1 This is a patientsafety issue.

Class 6 emulating indicators can be part of therelease criteria for loads containing implants. (ANSI/AAMI ST79 Table 7)2 The results cannot be used torelease the implant in emergency situations before the BIresults are available as you can do with a Class 5 CI. A Class 5 or Class 6 CI cannot be used as a replacementfor the BI because they do not directly measurer thelethality of a cycle. (ANSI/AAMI ST79 Section 10.5.3)1

For loads run in EO the BI PCD should contain aClass 4 multi-variable or a Class 5 integrating CI. (ANSI/AAMI ST41 Section 10.4 and 10.5.2.2.2)2

74 February 2011healthVIE.com

Let’s discuss the types of BI PCDs to use for routine loadrelease of nonimplant and implant loads and for routine sterilizerefficacy monitoring.

Routine Load Release of Nonimplant LoadsRoutine release of nonimplant loads should be an active

decision based on the evaluation of all available data. An experi-enced and knowledgeable person should make that decision at theend of the steam sterilization cycle after evaluating the: Physical monitors; External chemical indicators; Internal chemical indicators; A PCDs containing either a

• BI,• BI and Class 5 integrating indicator,• Class 5 integrating indicator,• Class 6 emulating indicator. (ANSI/AAMI ST79 Section

10.4)1

The Class 6 CI PCD should only be used in the cycle for whichit is labeled. (ANSI/AAMI ST79 Section 10.5.2.2)1 There is nouniversal Class 6 CI or CI PCD.

The use of a PCD in each nonimplant load is optional based onthe ANSI/AAMI ST79 recommended practices but the advantage ofusing a BI PCD in each load is discussed in the introduction sectionabove. How will you determine that the sterilization process is effec-tive at killing microorganisms if you do not monitor with a BI PCD?

For EO sterilizers the following should also be evaluated by an experienced and knowledgeable person: Physical monitors; External chemical indicators; Internal chemical indicators; BI PCDs containing a chemical indicator (CI) in each load.

(ANSI/AAMI ST41 Section 10.4)2

The AAMI recommendations for monitoring other gaseouschemical sterilization processes (e.g., hydrogen peroxide and ozone)include the use of physical monitors, chemical indicators and a PCDwith a BI or a BI and a CI. (AAMI ST58 Section 9.5)4 The usershould also be appropriately trained and knowledgeable about theperformance and interpretation of results of the physical, chemicaland biological indicators and competency should be assessed.

AORN recommends in the 2011 Perioperative Standards andRecommended Practices for Sterilization and DisinfectionRecommendation XVI that the following be evaluated for all steamand low temperature sterilization processes (e.g., EO, hydrogenperoxide, and ozone): Physical monitors; External chemical indicators; Internal chemical indicators; BIs.5

Professional Education & Training

For hydrogen peroxide and ozone, the PCD may contain a BIor a BI and CI. (ANSI/AAMI ST58 9.5.4.1)4

When documented medical exceptions occur for lowtemperature sterilization and an early release is unavoidable,the release of the device before the BI is know should be documented. (ANSI/AAMI ST58 section 9.5.4.3)2,4

Practical Application• Use physical monitors, chemical and biological

indicators to release implant loads.• Every implant load should be monitored with a

BI PCD and quarantined until the BI is negativeto ensure improperly sterilized items are notreleased for patient use.

• Competency testing of employees ensures correctinterpretation of monitoring tools.

Routine Sterilizer Efficacy Monitoring Routine sterilizer efficacy monitoring establishes a

regular pattern of testing the efficacy of the sterilizationprocess. All sterilizers should be routinely monitored usingappropriate BI PCDs. If a BI PCD is used in each load thismonitoring meets the recommended practice requirements forroutine sterilizer efficacy monitoring and takes the guess workout of whether all implants, every type of steam sterilizationcycle and every type of packaging used in flash or immediateuse sterilization is monitored with a BI. Every load monitoringwith a BI PCD ensures consistency and detects operator errors which are the main reason for sterilization process failures. Let’s discuss the routine sterilizer efficacy monitoringrequirements for each type of sterilization process.

STEAM STERILIZERSSteam sterilizers should be monitored with a BI PCD

weekly, preferably daily (each day the sterilizer is used).(ANSI/AAMI ST79 Section 10.4)1 ANSI/AAMI ST79 Section10.7.1 states that if a steam sterilizer is designed to be used formultiple types of cycles then: each type of cycle should be routinely tested because

each cycle creates a different challenge to air removal andsteam penetration;• gravity-displacement at 132°C to 135°C [270°F to

275°F],• gravity-displacement at 121°C [250°F],• dynamic-air removal at 132°C to 135°C [270°F to

275°F],• flash at 132°C to 135°C [270°F to 275°F],• flash or immediate use with single wrapper or other

packaging.1

76 February 2011healthVIE.com

Professional Education & Training

A note states that if you are running both a 4 and a 10minute dynamic-air removal sterilizer at 132°C to 135°C[270°F to 275°F], then only the shortest sterilization timeneeds to be tested.1

Steam sterilizers greater than 2 cubic feet The BI PCD should be an AAMI 16 towel pack (routine

test pack) (see Figure 1) or a commercially available BI PCDof equivalent performance. A biological indicator containingGeobacillus stearothermophilus spores is used. The CI shouldbe either a Class 4 multi-variable, a Class 5 integrating indicator, or a Class 6 emulating indicator. One BI PCD isplaced on the bottom shelf of the sterilizer, over the drain, andin a full load to create the creates challenge to air removal andsteam penetration. (ANSI/AAMI ST79 Section 10.7.2)1 SeeTable 1 for examples of BI PCDs for the load contents.Remember to also test each type of cycle used in a sterilizer asdescribed under steam sterilizers above and to use the Class 6CI only in the cycle for which it is labeled.

Figure 1. Instructions to Assemble and Run an AAMI Routine16-Towel BI PCD for Steam Sterilizers Greater than 2 CubicFeet (ANSI/AAMI Section 10.7.2)1

Components: One or more BIs (one or two test vials and one

control vial from the same lot). Sixteen clean, preconditioned, reusable huck or

absorbent surgical towels, in good condition, eachapproximately 16 in x 26 in (41-cm x 66-cm).

Preparation: Fold each towel lengthwise into thirds and then fold

widthwise in the middle. Stack towels one on top of another, with folds opposite each other, to form a stack that is approximately 9 in wide, 9 in long,and 6 in high (23 cm x 23 cm x 15 cm). (See AAMIFigure 10.)

Place the BI(s) between the eighth and ninth towelsin the approximate geometric center of the pack.One additional BI from the lot used for testingshould be left unexposed to the sterilant and used asa positive control.

Tape the pack in a manner that will yield the packapproximately 6 in (15 cm) high.

Label as a BI PCD.

Test Procedure: Place the test pack flat in a full chamber on a rack or

shelf near the drain. (See AAMI Figure 11.)

Table 1. BI PCDs for Routine SterilizerEfficacy Monitoring of 270°F to 275°F (132°C to 135°C) Dynamic-Air-Removal or250°F (121°C) Gravity Displacement SteamSterilization Cycles*

*Check with the medical device or sterilizer manu-facturer for correct times for the items being processed.

**Check with the sterilizer and biological indicatormanufacturer to make sure you are using the correct BIfor the cycles being tested.

***Perforated or mesh bottom tray.

Practical Application• Test each type of cycle used in a

steam sterilizer with a BI PCD that isrepresentative of that load.

Table-top steam sterilizersThe BI PCD will be user-assembled and

should be representative of the same type ofpackage or tray to be routinely processed and the

78 February 2011healthVIE.com

Professional Education & Training

AAMI Figure 11—Placement of the 16 towel PCD (BI challenge test pack)for routine biological monitoring of sterilizers larger than 2 cubic feet

BI PCD**

BI in wrapped surgical tray(include porous items if in patient care tray) or an AAMI 16-towel BI PCD orcommercially available BI PCD of equivalent performance if appropriate forcycle parameters

BI in AAMI 16-towel PCD orcommercially available BI PCDof equivalent performance

BI in protective organizingcase in area(s) that create the greatest challenge to air removal and sterilant penetration or an AAMI 16-towel PCD or commerciallyavailable BI PCD of equivalentperformance if shown in product testing to be appropriate

BI in rigid sterilizationcontainer in area(s) that create the greatest challengeto air removal and sterilantpenetration or an AAMI 16-towel PCD or commerciallyavailable BI PCD of equivalentperformance if shown inproduct testing to be appropriate

Type of TrayConfiguration

or Load

Wrappedsurgical tray,***with or withoutporous item(towel, foampad, etc.)

Mixed wrappedload

Protectiveorganizing case

Rigid sterilizationcontainer

Run the load according to the sterilizer manufacturers’ instruc-tions.

At the end of the cycle, cool the test pack according to the BImanufacturers’ instructions.

Incubate the BI test vial and a control vial from the same loteach day a test vial is incubated in each incubator or auto-reader.

Read and record the results.

AAMI Figure 10—Preparation of the 16 towel PCD (BI challengetest pack)

Reprinted from ANSI/AAMI ST79:2010 and A1:2010 (Consolidated Text)with permission of the Association for the Advancement of Medical Instrumentation,Inc. (C) 2010 AAMI www.aami.org. All rights reserved. Further reproduction ordistribution prohibited.

Reprinted from ANSI/AAMI ST79:2010and A1:2010 (Consolidated Text) withpermission of the Association for theAdvancement of MedicalInstrumentation, Inc. (C) 2010 AAMIwww.aami.org. All rights reserved.Further reproduction or distributionprohibited.

Practical Application• Test each type of cycle used in a steam

sterilizer with a BI PCD that is representa-tive of that load.

Flash or immediate use sterilization For routine biological monitoring of immediate use

sterilization ANSI/AAMI ST79 Section 10.7.4.1 states: each type of tray configuration in routine use

should be tested because each creates a differentchallenge to air removal and steam penetrationduring the sterilization process:• perforated, mesh-bottom, open surgical tray;• rigid sterilization container system;• protective organizing case;• single-wrapped surgical tray.1

Remember to also test each type of cycle used in a sterilizer as described under steam sterilizers above.The BI PCD will be user-assembled and consist of a BI and a Class 4, 5, or 6 CI placed inside each trayconfiguration to be tested. Remember to use the Class 6CI only in the cycle for which it is labeled. See Figure 3and 4 for examples. A biological indicator containingGeobacillus stearothermophilus spores is used. AORNrecommends “Class 5 chemical integrating indicatorsshould be used within each sterilizer container or tray.”(AORN IV.d.)5 The BI PCD should be placed on thebottom shelf of the sterilizer, over the drain, in an emptyload to create the greatest challenge to air removal andsteam penetration. (ANSI/AAM ST79 Section 10.7.4)1

Limiting the number of tray configurations used willminimize the amount of routine testing needed.

80 February 2011healthVIE.com

Professional Education & Training

most difficult to sterilize.(See Figure 2.)(ANSI/AAMI ST79Section 10.7.3)1 A biological indicator containing Geobacilusstearothermophilus spores is used. The CI should be either a Class4 multi-variable, a Class 5 integrating indicator, or a Class 6emulating indicator. The Class 6 CI should be used in the cycle forwhich it is labeled. One BI PCD is placed on edge if it is a smallpack or flat if it is a tray or large pack in a full load which isnormally the front of the sterilizer. Table 2 describes appropriate BIPCDs for the load contents. Remember to also test each type ofcycle used in a sterilizer as described under steam sterilizers above.

Table 2. BI PCDs for Routine Sterilizer Efficacy Monitoringof Table-Top Steam Sterilizers

Temperature

270°F-274°F(132°C-135°C)

270°F-274°F(132°C-135°C)

250ºF(121ºC)

Time*

≥3 min

≥4 min

≥30 min

BI PCD**

BI in unwrappedperforated instrumenttray or glassware

BI in a wrapped trayor peel pouch andinclude porous items(e.g., towel, foampad, etc) if applicable

BI in wrapped packthat is representativeof the load, includeporous items ifappropriate

Program/Load

Unwrappedinstruments on a perforatedinstrument tray***or glassware

Wrapped trays ofinstruments,instruments inpeel pouches

Packs, wrapped

*Check with the medical device or sterilizer manufacturer forcorrect times for the items being processed.

**Check with the sterilizer and biological indicator manufacturer tomake sure you are using the correct BI for the cycles being tested.

***Perforated or mesh bottom.

Figure 2. Examples of BI PCDs for table-top sterilizers

“Class 5 chemical integratingindicators should be used withineach sterilizer container or tray.”

82 February 2011healthVIE.com

Professional Education & Training

Table 3. BI PCDs for Different Cycle/Tray Configurations for Flash or Immediate Use Sterilization Cycles*

Type of Tray Configuration

Open surgical tray***

Wrapped surgical tray, with or withoutporous item (towel, foam pad, etc.)

Protective organizing case

Rigid sterilization container

Open surgical tray with or withoutporous item (towel, foam pad, etc.)

Wrapped surgical tray, with or withoutporous item (towel, foam pad, etc.)

Protective organizing case

Rigid sterilization container

BI PCD**

BI in open surgical tray

BI in wrapped surgical tray (include porous items if inpatient care tray)

BI in protective organizing case in area(s) that create thegreatest challenge to air removal and sterilant penetration

BI in rigid sterilization container in area(s) that create thegreatest challenge to air removal and sterilant penetration

BI in open surgical tray (include porous items if in patientcare tray)

BI in wrapped surgical tray (include porous items if inpatient care tray)

BI in protective organizing case in area(s) that create thegreatest challenge to air removal and sterilant penetration

BI in rigid sterilization container in area(s) that create thegreatest challenge to air removal and sterilant penetration

Type of Cycle

Gravity

Gravity

Gravity

Gravity

Dynamic-air-removal

Dynamic-air-removal

Dynamic-air-removal

Dynamic-air-removal

*Check with the medical device or sterilizer manufacturer for correcttimes for the items being processed.

**Check with the sterilizer and biological indicator manufacturer tomake sure you are using the correct BI for the cycles being tested.***Perforated or mesh bottom trays.

Open perforatedtray

Rigid container

Figure 3. BI PCDs for Flash or Immediate Use Sterilization in Gravity Sterilizers

Figure 4. BI PCDs for Flash or Immediate use Sterilization in Dynamic-air Removal sterilizers

Open perforatedtray

Rigid container

Practical Application• Test each type of cycle used in each flash

sterilizer and each type of packaging used witha BI PCD that is representative of that load.

LOW TEMPERATURE STERILIZATIONEthylene oxide sterilizers

The routine BI PCD recommended in Section 10.7 ofANSI/AAMI ST79 or a commercially available BI PCDequivalent should be use in each sterilization cycle.2 TheBI PCD contains a BI with Bacillus atophaeus spores anda Class 4 multi-parameter or Class 5 integrating indicator.One BI PCD should be placed in the center of each loadunless otherwise indicated by the sterilizer manufacturerbecause this creates the greatest challenge to humidity andEO penetration.

Figure 5. Instructions to Assemble an AAMIRoutine BI PCD for Ethylene Oxide Sterilization(ANSI/AAMI ST41 Section 10.72

84 February 2011healthVIE.com

Professional Education & Training

Components:1. Two BIs (one test vial and one control vial

from the same lot) and one Class 4 or Class 5CI. (See AAMI Figure 3.)

2. One plastic syringe (approximately 20 cc).3. A clean surgical towel, woven, 100% cotton.4. A peel pouch or wrapper large enough to

contain the test pack contents.

All components should be held at roomtemperature [18°C to 24°C (65°F to 75°F)] and arelative humidity of 35% for at least 2 hours priorto making the BI PCD.

Preparation:1. Place one BI, according to the BI manufactur-

ers’ instructions, inside a plastic syringe ofsufficient size that the plunger diaphragm doesnot touch the BI when the plunger is insertedinto the barrel of the syringe. If there is aplastic protective tip guard on the syringe,remove it. One additional BI from the lot usedfor testing should be left unexposed to thesterilant and used as a positive control.

2. Place the syringe with the BI and the CIbetween the folds of the clean surgical towel,which has been folded lengthwise into thirdsand then in thirds again to create nine layers.

AAMI Figure 3—Preparation of the PCD (routine BI test pack)(drawing not to scale)

3. Place these items inside one peel pouch or wrapperlarge enough to contain the test pack components.

4. Label as a BI PCD.

Test Procedure:1. Place in the center of the load. 2. Upon completion of the cycle, the BI PCD should be

handled according to the healthcare facility’s protocolfor minimizing worker exposure to EO (see sterilizerand BI PCD manufacturers’ instructions).

3. Read the CI and record the results.4. Incubate the BI test vial and a control vial from the

same lot each day a test vial is incubated in each incu-bator or auto-reader. Read and record the results.

Reprinted from ANSI/AAMI ST41:2008 with permission of the Association for the Advancement of Medical Instrumentation, Inc. (C) 2010 AAMI www.aami.org.All rights reserved. Further reproduction or distribution prohibited.

NOTE 1—Place the BI in the syringe according to the BImanufacturer’s instructions. The correct orientation of the BIin the syringe ensures that any vent in the BI faces toward the needle end of the syringe. (Paper strip BIs may be used inany orientation.)

NOTE 2—A CI should be placed in the folds of the towel.

Practical Application• Test each EO load with an AAMI routine BI PCD

or a commercial available BI PCD of equivalentperformance.

Hydrogen peroxide sterilizersA BI PCD should be used at least daily, but preferably in

every sterilization cycle.(ANSI/AAMI ST58 Section 9.4.4.3)4

A BI PCD for the hydrogen peroxide sterilization processconsists of a BI containing Geobacillus stearothermophilusspores and a CI. “The BI PCD should create a challenge to thesterilization process that is representative of the most difficultitem in the load being processed.” (ANSI/AAMI TIR 31Section 6.4)3 Commercial BI PCD kits are available. Thesekits are specific to the sterilizer model so ensure you are usingthe correct kit for the sterilizer being monitored. (ANSI/AAMI TIR30 Section 6.4)3 A healthcare facility prepared BI PCD should include a BI and CI in a package or trayroutinely processed that is representative of the most difficultto sterilize item in the load being processed. (ANSI/AAMITIR 31 Section 6.4.4.2)3 Consult with the sterilizer manufac-turer for material selection, assembly of, and placement of the BI PCD.

Practical Application• Test each hydrogen peroxide load with a BI PCD.

Ozone sterilizersA BI PCD should be used at least daily, but preferably in

every sterilization cycle. (ANSI/AAMI ST58 Section 9.4.4.3)4

A BI PCD for the ozone sterilization process consists of abiological indicator containing Geobacillus stearother-mophilus spores and a CI. “The BI PCD should provide achallenge to the sterilization process that is representative ofthe most difficult item to sterilize in the load being processed.”(ANSI/AAMI TIR30 Section 6.5.2)3. The manufacturerrecommends the preparation of a syringe pack. The BI isplaced with the cap towards the opening or bevel of the syringewith the tip removed. The syringe containing the BI and a CI(outside of the syringe) is placed inside a peel pouch. Contactthe sterilizer manufacturer for further information about thematerial selection, assembly of, and placement of the BI PCD.

Practical Application• Test each ozone load with a BI PCD.

What to Do When the Monitoring ProductsDetect a Problem

When the BI PCD, CI PCD, other CIs, or physical monitors detect a sterilization process failure a healthcarefacility needs to take action. ANSI/AAMI ST79 provides adecision tree (see AAMI Figure 12) for conducting investiga-tions of steam sterilization process failures for steam sterilizersbut the same process used in this decision tree could be used forlow temperature sterilization processes. (ANSI/AAMI ST79Section10.7.5)1 In addition see AAMI Table 8 for a checklistsfor identifying reasons for steam sterilization process failures.(ANSI/AAMI ST79 Section10.7.5)1

If any of the monitoring tools identify a sterilizationprocess failure the load is quarantined, the sterilizer is removedfrom service, and an investigation of the cause is initiated.(ANSI/AAMI ST79 Section10.7.5)1 The action steps say inANSI/AAMI ST79 section 10.7.5.1 for steam sterilizers:

“b) If the cause of failure is immediately identified(usually operator error) and confined to one load orone item in the load (i.e., an item with a nonresponsiveinternal CI), the cause of the failure should becorrected and the load should be reprocessed. If thecause of the failure is not immediately identified, theload should be quarantined, and all loads back to thelast negative BI should be recalled. Items in theseloads should be retrieved, if possible, and reprocessed(see 10.11). The sterilizer in question should be takenout of service for further investigation of root causes.See Figure 12 and Table 8 for guidance on how toconduct this investigation.”1

ANSI/AAMI ST41 states in Section 10.7.6.1 for EO sterilizers:

“b) If it is determined that the sterilization failurewas not the result of operator error (e.g., selecting theincorrect cycle), items processed in that sterilizer sincethe last negative BI results should be considerednonsterile. They should be retrieved, if possible, andreprocessed. The sterilizer in question should be takenout of service.”2

86 February 2011healthVIE.com

Professional Education & Training

Test each hydrogen peroxide or ozone load with a BI PCD.

88 February 2011healthVIE.com

Professional Education & Training

Figure 12 – Decision tree for conducting investigations of steam sterilization process failures

Reprinted from ANSI/AAMI ST79:2010 and A1:2010 (Consolidated Text) with the permission of Association for the Advancement of MedicalInstrumentation, Inc. (C) 2010 AAMI www.aami.org. All rights reserved. Further reproduction or distribution prohibited.

The same response should occur when any monitoringproducts used in hydrogen peroxide or ozone sterilizers indicatea sterilization process failure. The qualification testing to bedone to place the sterilizers back into use will depend onwhether there was a minor or major repair of the sterilizer or

utilities. This subject was addressed in sterilizer QualificationTesting Using Process Challenge Devices in the healthVIE.comJuly 2010 issue. More information on product recall can befound in the Product Recall Update published November 2009in Managing Infection Control (now called healthVIE.com).

90 February 2011healthVIE.com

Professional Education & Training

Table 8—Checklist for identifying reasons for steam sterilization process failures

Operator Errors

Incorrect use and interpretation of monitoring tools Incorrect physical monitors for the load

Incorrect use of BI or BI PCD• Incorrect selection of BI or BI PCD for the load• Incorrect placement of BI PCD in the load (e.g.,

another pack was placed on top of the PCD)• Incorrect incubation of BI• Misinterpretation of BI result• Incorrect documentation of BI result

Incorrect use of Class 5 integrating CI PCD or Class 6emulating CI PCD.• Incorrect selection of CI PCD for the load• Incorrect placement of CI PCD in the load (e.g.,

another pack was placed on top of the PCD)• Misinterpretation of Class 5 integrating CI result

or Class 6 emulating CI result• Incorrect documentation of Class 5 integrating CI

result or Class 6 emulating CI result

Incorrect use of internal CI• Incorrect selection of internal CI for the load• Misinterpretation of internal CI result• Incorrect documentation of internal CI results

Incorrect storage of any CIs or BIs

Failure to check physical monitors for functionalitybefore running cycle

Use of broken media ampoule or ampoule withmissing spore strip

Use of BI PCD or CI PCD that is missing the BI or CI

Use of defective CI (e.g., a CI that is expired, faded,shows a partial color change because of incorrectstorage, or has been previously exposed to the steri-lant)

Selection of incorrect cycle for load contents (containment device or medical device manufacturer’sinstructions for use not followed)Use of inappropriate packaging materials or packagingtechnique Incorrect packaging or containment device for the

cycle parameters

Incorrect preparation of containment device for use(e.g., incorrect filters, valves, or bottom tray)

Use of a paper-plastic pouch, woven or nonwovenwrapper, or towel in a 270ºF to 275ºF (132ºC to135ºC) gravity-displacement cycle

Use of a tray that does not allow air removal and steampenetration

Use of a wrapper that is too large for the application

Placement of a folded paper-plastic pouch inside anotherpaper-plastic pouch

Placement of a paper-plastic pouch inside a wrapped set orcontainment device without verification of adequate airremoval and steam penetration by product testing

Incorrect placement of basins in set (i.e., basins are notaligned in the same direction)

Failure to use nonlinting absorbent material between nestedbasins

Preparation of textile packs that are too dense to sterilize with the cycle parameters chosen

Inadequate preconditioning of packaging materials (i.e., notholding package materials at 68ºF to 73ºF (20ºC to 23ºC) for2 hours before use)

Incorrect loading of sterilizer Stacking of containment devices if not recommended by

manufacturer

Stacking of perforated instrument trays

Incorrect placement of instrument trays (i.e., not layinginstrument trays flat or parallel to the shelf)

Incorrect placement of paper-plastic pouches (e.g., placingpouches flat instead of on edge; not allowing sufficient spacebetween pouches; not placing pouches with plastic sidesfacing one direction)

Incorrect placement of basins (i.e., not placing basins ontheir sides so that water can drain)

Incorrect placement of textile packs (i.e., not placing them onedge)

Placement of packages too close together, impeding airremoval and sterilant penetration in the load

Sterilizer or Utilities Malfunctions

Poor steam quality or quantity Wet steam

• Improper insulation of steam lines• Malfunction of trap in steam line or no trap in steam line• Malfunction of drain check valve or no drain check

valve

92 February 2011healthVIE.com

Professional Education & Training

• Steam contact with a cold load• Too much water in steam produced at boiler

Superheated steam• Improper heatup of chamber• Desiccated packaging materials (e.g., towels)• Steam pressure too low for the temperature• Excessive reduction of steam pressure too close to sterilizer• Faulty steam control valve or pressure reducer control valve

Other steam problems• Variations in steam pressure because of clogged filter,

poorly engineered piping, or excessive demands• Out-of-calibration pressure gauges and controllers• Clogged steam lines• Clogged steam supply strainer• Clogged chamber drain line, strainer, or chamber drain

screen• Malfunction of valves

Incomplete air removal Inadequate vacuum or vacuum depth or other air removal

system

Clogged chamber drain line, strainer, or chamber drain screen

Clogged vent lines

Leak caused by faulty door gasket

Leak in other areas of chamber

Plugged, faulty or incorrectly adjusted control valves

Low steam pressure

High water temperature

Table 8—Checklist for identifying reasons for steam sterilization process failures (continued from page 90)

SummaryLoad control monitoring is a process in which a load is

monitored and released based on the result of a BI PCD. The BIPCD should contain the appropriate BI for the process being moni-tored and create a challenge equal to or greater than the challengeposed by the most difficult item routinely processed. Competencytesting of employees also ensures correct interpretation of themonitoring tools used to release nonimplant and implant loads andto routinely monitor the efficacy of the sterilization process. Usinga self-contained BI with a one, three- or four-hour readout for loadcontrol in steam or EO sterilization processes allows quarantiningof all loads pending BI results which reduces the risk and cost ofhealthcare-associated infections because loads are not releaseduntil the BI is negative. This is about improving patient safety. ✛

Inadequate water supply pressure

Clogged water supply strainer

Trapping of air by the load

Incorrect cycle parameters for the load

Inadequate cycle temperature Out-of-calibration temperature gauge

Long heatup time for large loads (i.e., heat lag)

Clogged chamber drain line, strainer, or chamberdrain screen

Variations in steam pressure because of cloggedfilter, poorly engineered piping, or excessivedemands on steam supply

Presence of noncondensable gases in steam lineand load

Inadequate steam supply pressure

Clogged steam supply strainer

Insufficient time at temperature Out-of-calibration control timer

Inappropriate cycle parameters for the load beingprocessed

Come-up time of less than 1.5 minutes in a 270ºFto 275ºF (132ºC to 135ºC) gravity-displacementcycle

Oversized load

References1. Association for the Advancement of Medical Instrumentation.

Comprehensive guide to steam sterilization and sterility assurance in health care facilities. ANSI/AAMI ST79:2010and A1:2010 (Consolidated Text).

2. Association for the Advancement of Medical Instrumentation.Ethylene oxide sterilization in health care facilities: Safetyand effectiveness. ANSI/AAMI ST41:2008.

3. Association for the Advancement of Medical Instrumentation.Process challenge device/test packs for use in health carefacilities. AAMI TIR31:2008.

4. Association for the Advancement of Medical Instrumentation.Chemical sterilization and high-level disinfection in healthcare facilities. ANSI/AAMI ST58:2005.

5. Association of perioperative registered Nurses. RecommendedPractices for Sterilization in the Perioperative PracticeSetting. Perioperative Standards and Recommended Practices.2011.

94 February 2011healthVIE.com

Professional Education & Training

1. A2. A3. A4. A5. A

6. A7. A8. A9. A

10. A

Nursing CEU Application FormThis inservice is approved by the California Board of Registered Nurses, CEP

5770 for 1 contact hour. This form is valid up to five (5) years from the date of publication.1. Make a photocopy of this form.2. Print your name, address and daytime phone number and position/title.3. Add the last 4 digits of your social security number or your nursing license number.4. Date the application and sign. 5. Answer the true/false CE questions. KEEP A COPY FOR YOUR RECORDS.6. Submit this form and the answer sheet to:

TKMK Media, LLChealthVIE.comPO Box 25310, Scottsdale, AZ 85255-9998

7. For questions, contact [email protected]. Participants who score at least 70% will receive a certificate of completion

within 30 days of healthVIE.com’s receipt of the application.

ApplicationPlease print or type.

Name______________________________________________________________

Mailing Address______________________________________________________

City, State, Country, Zip _______________________________________________

Daytime phone ( )__________________________________________

Position/Title_______________________________________________________

Social Security or Nursing License Number ________________________________

Date application submitted _____________________________________________

Signature __________________________________________________________

Offer expires February 2016On a scale of 1-5, 5 being Excellent and 1 being Poor, please rate this program

for the following:

1) Overall content ___________________

2) Met written objectives ______________

3) Usability of content ________________

Sterile Process and Distribution CEU InformationCEU Applicant Name _________________________________________________

Address___________________________________________________________

City____________________________ State________ Zip Code ______________

The CBSPD (Certification Board for Sterile Processing and Distribution) has pre-approved this inservice for 1.5 contact hours for a period of five (5) years from the date of publication. Successful completion of the lesson and post test must bedocumented by facility management and those records maintained by the individualsuntil recertification is required. DO NOT SEND LESSON OR TEST TO CBSPD.

For additional information regarding Certification contact: CBSPD, 148 Main St.,Lebanon, NJ, 08833 or call 908-236-0530 or 800-555-9765 or visit the Web site atwww.sterileprocessing.org.

IAHCSMM has awarded 1.5 Contact Points for completion of this continuing education lesson toward IAHCSMM recertification.

ANSWERS

<02/2011>

Martha Young, BS, MS, CSPDT,is president of Martha L. Young, LLC,providing SAVVY sterilization solu-tions to healthcare manufacturers andfacilities and a consultant for 3M. Sherecently retired from the 3M InfectionPrevention Division, St. Paul, Minn.after 31 years and has more than 28years of experience in the specialtyarea of sterilization and disinfection.Ms. Young has lectured around theworld, has numerous publications oninfection prevention with an emphasison how to improve the performance of the sterilization process, and is atechnical advisor for healthVIE.com.She is a member of IAHCSMM,AORN (Past Professional/PracticeIssues Chair for AORN SpecialityAssembly for Sterilization Processingand Materials Management from2006-2010), APIC and a certifiedCentral Sterile Processing andDistribution Technician. Additionally,Ms. Young is a member of severalAAMI working groups developingrecommended practices. In 2007 HPNacknowledged her as one of the “30Pros Worth Knowing” who are themost influential in healthcare sterileprocessing. Ms. Young can be reachedat [email protected].

3M Sterilization Assurance, Attn HC4160RR Donnelly Fulfillment Services585 Hale Ave N, Oakdale, MN 55128-9935