long-term durability of carpentier-edwards magna …vzs0037/ats_edwards_paper.pdf · long-term...

7
Long-term Durability of Carpentier-Edwards Magna Ease valve: A One Billion cycle In-Vitro Study Vrishank Raghav, PhD a , Ikechukwu Okafor, BS a , Michael Quach, BS b,** , Lynn Dang, BS b,** , Salvador Marquez, BS, MBA b,** , Ajit P. Yoganathan, PhD a,* a Georgia Institute of Technology, Atlanta, GA b Edwards Lifesciences, Irvine, CA Abstract Background and Aim of Study Durability and hemodynamic performance are top considerations in selecting a valve for valve replacement surgery. This study was conducted in order to evaluate the long-term mechanical durability and hydrodynamic per- formance of the Carpentier-Edwards R PERIMOUNT R Magna Ease TM Bioprostheses, through one billion cycles (equivalent to 25 years). Materials and Methods In-vitro valve hydrodynamic performance, durability, and quantitative flow visualization were conducted in accordance with ISO 5840:2005 heart valve standard. The study valves were subjected to accelerated valve cycling to an equivalent of 25 years of wear. Hydrodynamic evaluations at intervals of 100 million cycles (2.5 years) were performed on the study valves. New un-cycled Magna Ease valves were used as hydrodynamic controls in this study. A quantitative assessment of the fluid motion downstream of the control and study valves was performed using Particle Image Velocimetry. The results between the test and control valves were compared to assess valve performance after an equivalent of 25 years of wear. Results All study valves met the ISO5840 requirements for eective orifice area, 1.81±0.06 cm 2 and 2.06±0.17 cm 2 , and regurgitant fraction, 1.11±0.87% and 2.5±2.34%, for the 21mm and 23mm study valves, respectively. The flow characterization of the control valves and the billion- cycle valves demonstrated that the valves exhibited similar flow characteristics. The velocity and shear stress fields were similar between the control and study valves. Conclusions The Magna Ease valves demonstrated excellent durability and hydrodynamic performance after an equivalent of 25 years of simulated in-vitro wear. All study valves successfully endured one billion cycles of simulated wear, five times longer than the standard requirement for a tissue valve as stipulated in ISO5840. Keywords: bioprosthesis valve durability, tissue valve durability, valve, tissue valve, long-term performance 1. Introduction The use of bioprosthetic heart valves (BHV) for valve re- placement surgeries has increased in the recent past. In 2009, Brown et al. reported a 10-year study between 1996 and 2006 demonstrating a dramatic shift away from mechanical heart valves towards BHVs. In this work, Brown et al. also showed that outcomes, including rates of death and stroke, have not only improved but are quite low for isolated aortic valve re- placement surgery with BHVs (1). With the increasing use of BHVs, long-term durability and good hydrodynamic perfor- mance remain the paramount considerations in valve replace- ment surgery. Primary failure modes of BHVs take on many * Wallace H Coulter Department of Biomedical Engineering Georgia Institute of Technology and Emory University Technology Enterprise Park, Suite 200 387 Technology Circle, Atlanta, GA 30313 ** Conflict of Interest Statement: Lynn Dang, Michael Quach, and Salvador Marquez are employed by Edwards Lifesciences. Wear testing for this study was conducted at Edwards Lifesciences, Irvine, California. Ajit Yoganathan, PhD is a consultant for Edwards Lifesciences Email address: [email protected] (Ajit P. Yoganathan, PhD) forms such as mechanical wear and tear, as well as calcific, non-calcific and fibrotic degeneration (2, 3), with calcification and mechanical modes accounting for the largest proportion of failures (4). Vesely et al. studied these two failure modes in BHVs and concluded that majority of the damage was occurring strictly due to mechanical reasons (5). Roselli et al. (2) conjec- tured that the advancements in the fixation techniques currently available reduce calcification in the biological milieu and hence mechanical wear could be a dominant mechanism for structural valve deterioration. In this work, we focus on this dominant failure mechanism: mechanical wear and tear. An in vitro study can be used to isolate and evaluate the long-term mechanical durability of the bioprosthetic valve in the absence of calcification. Past stud- ies have shown encouraging correlations between durability ob- served in in vitro experiments and clinical in vivo durability of prosthetic heart valves (6, 7). Currently, the International Or- ganization for Standardization (ISO) has mandated durability testing of BHVs in an in vitro setting up to 200 million cycles (equivalent to 5 years in vivo), as specified in the ISO 5840 In- ternational Heart Valve Standard (8). This mandate was estab- lished based on previous clinical data that demonstrated that in Accepted Postprint submitted to Annals of Thoracic Surgery May 1, 2016

Upload: hamien

Post on 04-Jun-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

Long-term Durability of Carpentier-Edwards Magna Ease valve: A One Billion cycleIn-Vitro Study

Vrishank Raghav, PhDa, Ikechukwu Okafor, BSa, Michael Quach, BSb,∗∗, Lynn Dang, BSb,∗∗, Salvador Marquez, BS, MBAb,∗∗,Ajit P. Yoganathan, PhDa,∗

aGeorgia Institute of Technology, Atlanta, GAbEdwards Lifesciences, Irvine, CA

Abstract

Background and Aim of Study Durability and hemodynamic performance are top considerations in selecting a valve for valvereplacement surgery. This study was conducted in order to evaluate the long-term mechanical durability and hydrodynamic per-formance of the Carpentier-Edwards R© PERIMOUNT R© Magna EaseTM Bioprostheses, through one billion cycles (equivalent to 25years).Materials and Methods In-vitro valve hydrodynamic performance, durability, and quantitative flow visualization were conductedin accordance with ISO 5840:2005 heart valve standard. The study valves were subjected to accelerated valve cycling to anequivalent of 25 years of wear. Hydrodynamic evaluations at intervals of 100 million cycles (2.5 years) were performed on thestudy valves. New un-cycled Magna Ease valves were used as hydrodynamic controls in this study. A quantitative assessment ofthe fluid motion downstream of the control and study valves was performed using Particle Image Velocimetry. The results betweenthe test and control valves were compared to assess valve performance after an equivalent of 25 years of wear.Results All study valves met the ISO5840 requirements for effective orifice area, 1.81±0.06 cm2 and 2.06±0.17 cm2, and regurgitantfraction, 1.11±0.87% and 2.5±2.34%, for the 21mm and 23mm study valves, respectively. The flow characterization of the controlvalves and the billion- cycle valves demonstrated that the valves exhibited similar flow characteristics. The velocity and shear stressfields were similar between the control and study valves.Conclusions The Magna Ease valves demonstrated excellent durability and hydrodynamic performance after an equivalent of 25years of simulated in-vitro wear. All study valves successfully endured one billion cycles of simulated wear, five times longer thanthe standard requirement for a tissue valve as stipulated in ISO5840.

Keywords: bioprosthesis valve durability, tissue valve durability, valve, tissue valve, long-term performance

1. Introduction

The use of bioprosthetic heart valves (BHV) for valve re-placement surgeries has increased in the recent past. In 2009,Brown et al. reported a 10-year study between 1996 and 2006demonstrating a dramatic shift away from mechanical heartvalves towards BHVs. In this work, Brown et al. also showedthat outcomes, including rates of death and stroke, have notonly improved but are quite low for isolated aortic valve re-placement surgery with BHVs (1). With the increasing useof BHVs, long-term durability and good hydrodynamic perfor-mance remain the paramount considerations in valve replace-ment surgery. Primary failure modes of BHVs take on many

∗Wallace H Coulter Department of Biomedical EngineeringGeorgia Institute of Technology and Emory UniversityTechnology Enterprise Park, Suite 200387 Technology Circle, Atlanta, GA 30313∗∗Conflict of Interest Statement: Lynn Dang, Michael Quach, and Salvador

Marquez are employed by Edwards Lifesciences. Wear testing for this studywas conducted at Edwards Lifesciences, Irvine, California. Ajit Yoganathan,PhD is a consultant for Edwards Lifesciences

Email address: [email protected] (Ajit P.Yoganathan, PhD)

forms such as mechanical wear and tear, as well as calcific,non-calcific and fibrotic degeneration (2, 3), with calcificationand mechanical modes accounting for the largest proportion offailures (4). Vesely et al. studied these two failure modes inBHVs and concluded that majority of the damage was occurringstrictly due to mechanical reasons (5). Roselli et al. (2) conjec-tured that the advancements in the fixation techniques currentlyavailable reduce calcification in the biological milieu and hencemechanical wear could be a dominant mechanism for structuralvalve deterioration.

In this work, we focus on this dominant failure mechanism:mechanical wear and tear. An in vitro study can be used toisolate and evaluate the long-term mechanical durability of thebioprosthetic valve in the absence of calcification. Past stud-ies have shown encouraging correlations between durability ob-served in in vitro experiments and clinical in vivo durability ofprosthetic heart valves (6, 7). Currently, the International Or-ganization for Standardization (ISO) has mandated durabilitytesting of BHVs in an in vitro setting up to 200 million cycles(equivalent to 5 years in vivo), as specified in the ISO 5840 In-ternational Heart Valve Standard (8). This mandate was estab-lished based on previous clinical data that demonstrated that in

Accepted Postprint submitted to Annals of Thoracic Surgery May 1, 2016

an in vivo setting, the calcification cascade begins after 5 years,at which point mechanical testing would no longer be clinicallyrelevant. However, in the light of new evidence that demon-strates the primary cause of valve deterioration as mechanicalwear (5), it is now very important to perform durability tests ofBHVs well past 200 million cycles. This need for long-termdurability testing is further emphasized based on clinical stud-ies that have demonstrated integrity and hemodynamic stabilityof BHVs up to 20 years post-implantation in patients (9, 10).

Considering the increasing use of BHVs, the objective ofthis study was to assess the simulated long-term durability andhydrodynamic performance characteristics of the Magna Easevalve in an in vitro setting. This study evaluates the valvesup to 1 billion cycles which is equivalent to 25 years of sim-ulated in vivo wear. Three tests were conducted: a) Acceler-ated Wear Testing (AWT), b) Hydrodynamic evaluation, and c)Particle Image Velocimetry (PIV). The AWT enables the simu-lation of 25 years of mechanical wear (excluding calcificationand other biological processes) on the valve within a relativelyshort amount of time by accelerating the beat rate. Hydrody-namic evaluation was used to quantify the performance of thevalves and PIV was used to understand the specific flow fieldfeatures of the valves. Hydrodynamic performance and flowfield results of the wear tested study valves were compared tothose of uncycled-control valves of same sizes.

2. Materials and Methods

2.1. Valve Selection

The valve selected for this study was the Magna Ease aor-tic valve which is one of the most commonly implanted andcommercially available aortic valves. Three samples each of21mm and 23mm of clinical quality valves were used as thestudy valves. These two sizes were selected because they rep-resent the majority of implanted valves in the patient popula-tion. To determine how the performance of the study valvesmight change over time, three samples of the Magna Ease aorticvalve of same sizes in the uncycled condition were used as thecontrols valves for hydrodynamic comparison to the 1-billioncycled study valves.

2.2. Valve Description

The Magna Ease model 3300TFX valve is a stented bovinepericardial tissue valve fixed using glutaraldehyde solution andtreated with the proprietary ThermaFix processes. This is a pro-prietary fixation method to reduce calcification in the pericar-dial tissue. The stent is made from cobalt chromium coveredwith PTFE cloth.

2.3. Experimental Setup and Testing Methodology

The study met the requirements for test equipment, set-up,and test intervals as specified for hydrodynamic, durability, andflow visualization testing specified in the FDA ReplacementHeart Valve Guidance and ISO 5840 International Heart ValveStandard. The accelerated wear tester (AWT) was used in thisstudy to simulate wear of the valves within a relatively short

amount of time by accelerating the beat rate. In this study, thetester was set to operate at a frequency between 800-1100 cpm(cycles per minute) for a minimum of one billion cycles. Thetester was tuned to peak differential closing pressure of greaterthan 95mmHg, where 5% of each cycle must be above this min-imum pressure for 95% of all cycles per ISO5840 requirements.The AWT was also tuned to achieve complete valve openingand closing for each cycle, similar to the valve opening andclosing observed in the pulse duplicator tuned to physiologicconditions. In order to meet both of these conditions, the av-erage peak transvalvular differential pressure the valves weresubjected to was approximately 120 mmHg.

Leaflet kinematics were evaluated using high-speed videoduring AWT and during hydrodynamic performance testing.Hydrodynamic performance testing was conducted using apulse duplicator system consisting of an atrium, mitral section,ventricular drive, aortic section and peripheral compliance andresistance. The pulse duplicator system can be adjusted to sim-ulate various physiologic flow and pressure conditions. Valveswere evaluated for pressure drop, leakage, and Effective Ori-fice Area (EOA). The valves were subjected to hydrodynamicevaluation at nominal physiologic flow and pressure conditions(Cardiac Output (CO) 5 LPM, Heart Rate (HR) 70 BPM, andMean Aortic Pressure (MAP) 100 mmHg). The study valvetesting and valve inspections were conducted at specific inter-vals of 100 million cycles from time zero up to one billion cy-cles.

In order to characterize the fluid flow characteristics of thevalves, Particle Image Velocimetry (PIV) was performed onthe study valves and control valves. PIV is an instantaneous,non-intrusive, whole-field, optical flow diagnostic technique tomeasure global velocity field. Fluid mechanic metrics such asturbulence and shear stress fields can be derived using this ve-locity field information. The PIV experiments were conductedin the Cardiovascular Fluid Mechanics Laboratory at the Geor-gia Institute of Technology.

2.4. Pulsatile Left Heart SimulatorThe Georgia Tech Left Heart Simulator (GTLHS) was used

as the pulsatile flow loop for the PIV studies. This loop con-sists of a fluid reservoir, a mechanical mitral valve, a bladderpump, a valve chamber, a compliance chamber, and a resis-tance element. The bladder pump, driven by compressed air andsolenoid valves is used to simulate ventricular (11). A water-glycerin solution containing 36% glycerin by volume was usedas a blood analog fluid to match the density and viscosity ofblood (kinematic viscosity of 3.51 cP) and to minimize opticaldistortion during PIV experiments.

The flow conditions used for these experiments were thenominal physiological cardiac output and stroke volume of 5.0L/min and 70 mL per beat at a mean arterial pressure of 100mmHg. The cardiac output, ventricular and aortic pressuresthrough the cardiac cycle were measured at 500 Hz for allvalves, using a custom LabView program for 200 cycles beforeand after each experimental run. Measurements of the velocityfield were acquired at 20 time points during the cardiac cycle.A higher temporal resolution during systole was chosen with

2

Figure 1: A) Sample hemodynamic plot show aortic flow, aortic pressure, and ventricular pressure. The blue region and vertical dashed lines represent time pointsat which PIV data were acquired. B) Sample velocity vector field for the 21 mm 0 cycle valve indicating the locations at which the velocity profiles were extractedC) Orientation of PIV plane relative to the aortic valve tested

a uniform time spacing of 25 ms (over 400 ms centered aboutpeak systole), yielding 16 time points and 4 other discrete timepoints were chosen during diastole (Figure 1).

2.5. PIV InstrumentationPIV is a quantitative, optical based method of flow analysis

which has been extensively used to study velocity and shearstress fields of aortic valve prosthesis in in vitro simulators (11,12). Two dimensional PIV was used to understand the flowfield downstream of the valve models. The time spacing be-tween laser pulses and image acquisition was optimized at eachtime point during the cardiac cycle to ensure that the averagedisplacement of particles between images was approximately25% of the final interrogation window size. Velocity measure-ments were acquired along the center line of the valve assem-bly such that one commissure post and line of leaflet coaptationwas aligned with the imaging plane and the opposing leafletwas aligned perpendicular to the same plane (Figure 1C).

2.6. Definition of Calculated Quantities- Principal ReynoldsShear Stress

Shear stress has been correlated to blood cell damage and canserve as a predictor of potential areas where platelet activationor damage may occur (12). Reynolds shear stresses (RSS) isa measure of the average momentum flux in a flow field dueto temporal variations in the velocity field, which arises dueto cycle-to-cycle variations as well as fluctuations in turbulentflow. The principal RSS is defined as follows:

RS S = ρ

√(u′u′ − v′v′

2

)2

+ ( ¯u′v′)2 (1)

Please refer to (13) for more details on the computation andinterpretation of this metric.

3. Results and Discussion

3.1. AWT and Hydrodynamic PerformanceRepresentative opening and closed images of a 23mm study

valve in the pulsatile flow tester and AWT at zero and one bil-

lion cycles test interval are shown in Figure 2. In general, theonly observable difference between the images is the dimin-ishing size of the central gap caused by leaflet relaxation overtime. The valves were taken out of the AWT at 100 million cy-cle test intervals for detailed inspection. Minor tears in the clothwere observed near the top of the stent posts on two of the size23mm valves. These tears are known to be a test artifact. Clothtears are not observed clinically, likely because the depositionof blood proteins and eventual ingrowth of host tissue eliminatethe direct contact between the leaflet and cloth.

One of the size 23mm valves developed a pin hole in theleaflet after approximately 500 million cycles (equivalent to ap-proximately 12.5 years) caused by the loose cloth from the clothtear that came into contact with the open leaflet a few millime-ters below the commissure tip immediately adjacent to the stentpost (Figure 3). This valve remained fully functional throughone billion cycles, meeting the acceptance criteria for effectiveorifice area and regurgitant fraction in ISO 5840. X-ray evalu-ation of the study valves verified that all structural componentsfor all valves remained intact over the one billion cycle durationof the AWT study. Valve kinematic evaluations verified leafletfunction from full opening to closure throughout the duration ofthe study. Hydrodynamic testing results of the study and con-trol valves, EOA and RF, are presented in Table 1. EOA showsapproximately 12% increase for the 23mm study valves and ap-proximately 10% increase for the 21mm study valves throughone billion cycles. The increase in EOA of the study valvesmay be due to leaflets becoming more flexible from the repeatedpressure loading in the accelerated wear tester over time (14).Both the study and control valves display EOA values (1.81cm2 and 1.60 cm2 for the 21 mm, respectively, and 2.06 cm2

and 1.70 cm2 for the 23 mm valves) well above the ISO mini-mum EOA requirement of 0.85 cm2 for 21 mm valves and 1.00cm2 for 23 mm valves. The supplementary movie submittedalong with the paper qualitatively illustrates the similarities inperformance between the two valves.

The average measured regurgitant fraction (RF) of the studyvalves and control valves are presented in Table 1. The resultsgenerally show similar regurgitant fractions between the study

3

Figure 2: Representative 23mm control and study valves opening and closing in the pulsatile tester, and photo inspection of 23mm study valve at time 0 and at 1billion cycles

Figure 3: 23mm study valve with hole in leaflet near right commissure

and control valves. Both the study and control valves displayregurgitant fraction values well below the minimum regurgitantfraction requirement in ISO5840.

The results presented above show that the Carpentier Ed-wards Magna Ease valves are able withstand the mechanical en-vironment of aortic valves for up to 25 years and perhaps evenlonger. This in-vitro AWT study demonstrates that the size 21and 23 mm Magna Ease study valves (N=3) maintained struc-tural integrity and good hydrodynamic performance up to 25years of simulated accelerated wear. These results support theclinical findings of Johnston et al. who showed that the cumula-tive incidence of explant/year for structural valve deteriorationwas just 5.45% at 20 years (10).

The next major hurdle to overcome for BHVs is calcifica-tion. Several clinical and in vitro studies have reported aboutthe different means calcification can develop on BHVs (15, 16).Kheradvar et al. (15, 16) discuss that calcification can occur onBHVs not only via chemical/biological mechanisms but alsodue to mechanical reasons such as micro-injury and stress. Theability to reduce calcification clinically has the potential to ex-tend the mechanical durability of bioprosthetic surgical valvesin patients.

3.2. PIV Results

Figure 1B illustrates a sample velocity vector field for the 21mm control valve at peak systole (T = 230 ms). It also indicatesthe locations at which the velocity profiles were extracted (15,25, 35, 45, and 55 mm downstream from the valve annulus) forall the valves tested in this work.

Figure 4 illustrates the velocity profiles at 5 locations down-stream from the valve annulus measured at peak systole for the21 mm and 23 mm valves. For the 21 mm valve a compari-son between the control and study valve shows that the velocityprofiles and peak velocity magnitude are quantitatively simi-lar (difference 0.1 m/s) proximal to the valve annulus. How-ever, the velocity fields of the two valves differ with respect tothe velocity profile (difference 0.5 m/s) distal from the valve(Figure 4B). This could imply that the momentum of the flowthrough the study valve was not sustained at the downstreamlocations when compared to the control valve. The increase inEOA observed for the study valve could be responsible for theobserved reduced velocity. For the 23 mm valves (control and

4

Table 1: Summary of Effective Orifice Area and Regurgitant Fraction Results, EOA = effective orifice area; ISO = International Organization for Standardization;RF = regurgitant fraction

Control Valve (0 cycles) Study Valve (1 Billion cycles)

Valve size EOA (cm2) RF (%) EOA (cm2) RF(%)

21 mm (n=3) 1.60±0.07 1.59±0.80 1.81±0.06 1.11±0.87

ISO 5840:200521 mm minimum requirement ≥0.85 ≤10% ≥0.85 ≤10%

23 mm (n=3) 1.70±0.06 2.08±0.90 2.06±0.17 2.52±2.34

ISO 5840:200523 mm minimum requirement ≥1.00 ≤10% ≥1.00 ≤10%

Figure 4: Velocity profile at 5 downstream locations from the valve annulus at peak systole in the center plane: A) 21 mm 0 cycle B) 21 mm 1 billion cycles C)23mm 0 cycle D) 23mm 1 billion cycles.

5

Figure 5: Principal RSS at peak systole for the valves in the center plane: A) 21mm 0 cycles B) 21mm 1 billion cycles C) 23 mm valve 0 cycles D) 23mm 1 billioncycles.

study) a greater plug flow velocity profile was observed whencompared to the 21 mm valves (Figure 4). The combination ofvalve size and chamber size (here the chamber size was heldfixed, while the valve size changed), could be responsible forthis observation. Furthermore, the 23 mm valves exhibit thesame trends as that of the 21 mm valves with respect the varia-tion of velocity profile at distal locations from the valve.

Figure 5 illustrates the principal RSS fields for the 21 mmand 23 mm valves at peak systole (T = 230 ms). For both the 21mm and 23 mm valves, a comparison between the control andstudy valve shows that peak magnitudes of RSS(∼250 N/m2)are comparable across the whole flow field. However, the spa-tial distribution of the RSS field is different between the controland the study valves (in both 21 mm and 23 mm). It should benoted that the outer edges of the flow field in PIV data measure-ments are usually noisy and hence should be taken into accountwhile drawing any conclusions.

Both study valves exhibited larger RSS values close to theleaflets. This is expected due to fluctuations/flutter of the in-dividual leaflet tips (due to wear) during systole. The fluctua-tions of the leaflets thus translate into higher turbulence in theflow field, which manifest themselves as higher RSS values. Itis interesting to note, that the larger spatial extent of the RSSfield (Figure 5A and 5B) appears to be correlated to the notice-able reduction in the phase average velocity profile (Figure 4Aand 4B) at distal locations from the annulus of the valve.

The PIV analysis performed in this work was used to evalu-ate the platelet damage potential of the study valve and to de-

termine if it was quantitatively different from the control valve.RSS is an extensively used metric that has been experimentallycorrelated to the shear induced damage to platelets. All valvestested fell far below the determined threshold for platelet dam-age between the range of 600 N/m2 and 200 N/m2 for exposuretimes between 5 and 15 ms (12). This indicates there was nodifference between the study and control valve with respect tocell damage potential. Prior in vitro studies on a porcine bio-prosthetic aortic valve by Lim et al. (17) also reported similarmaximum principal RSS values in the range of 250 - 500 N/m2.The blood damage potential of the valves observed in this studyappears to be within the standard reported and accepted ranges.

4. Study Limitations

The primary limitation of this study is the in vitro nature ofthe study. The pulse duplicators used in this work were fabri-cated using rigid tubing with lumped compliance and resistanceelements. The ascending aorta and sinus was axisymmetric andidealized, but were designed based on previous anatomic mea-surements. Similarly, the AWT system used for this study isidealized; however, the hemodynamic environment in which thevalves were tested matched the worst case scenarios of in vivoconditions. Furthermore, planar PIV measurements do not pro-vide the third velocity component; however, in the bulk jet flowfrom the aortic valve the third component should be minimal.

6

5. Summary and Conclusions

With recent clinical evidence of BHVs maintaining structuralintegrity and hydrodynamic stability to around 20 years in pa-tients, it is imperative that these valves be tested for mechani-cal wear and tear. The results from this work demonstrate thatthe study valves remained functional through the completionof one billion cycles (∼25 years) under an average transvalvu-lar pressure of approximately 120 mmHg; approximately fivetimes longer and at a pressure 25% higher than the standard re-quirements for a tissue valve in ISO 5840. All the structuralcomponents of the study valves remained intact and the hydro-dynamic results were functionally equivalent to the zero cycledcontrol valves. At the time of this publication, the study valvesare continuing to be cycled beyond 1.2 billion cycles withoutfailure. All the valves also met the acceptance criteria for re-gurgitant fraction and effective orifice area per ISO5840 guide-lines at the completion of one billion cycles. Furthermore, PIVresults indicate that the study valves exhibited similar flow fieldcharacteristics as the control valves. The above results demon-strate that in the absence of calcification the study valves exhib-ited superior long-term mechanical durability and stable hydro-dynamic performance. Therefore, the authors speculate that theability to reduce calcification clinically has the potential to fur-ther extend the mechanical durability of bioprosthetic surgicalvalves in patients.

6. Acknowledgments

The authors gratefully acknowledge the contributions ofMichael Truong, and Hao-Chung Yang. Edwards and MagnaEase are trademarks of Edwards Lifesciences Corporation. Ed-wards Lifesciences, the stylized E logo, Carpentier-Edwards,PERIMOUNT, Magna, and ThermaFix are trademarks of Ed-wards Lifesciences Corporation and are registered in the UnitedStates Patent and Trademark Office. The authors would liketo acknowledge the members of the Cardiovascular Fluid Me-chanics Laboratory for their assistance and feedback. The workat the Cardiovascular Fluid Mechanics Laboratory at the Geor-gia Institute of Technology was funded through discretionaryfunds available to the Principal Investigator through the Wal-lace H Coulter Endowed Chair.

References

[1] James M Brown, Sean M O’Brien, Changfu Wu, Jo Ann H Sikora, Bart-ley P Griffith, and James S Gammie. Isolated aortic valve replacement innorth america comprising 108,687 patients in 10 years: changes in risks,valve types, and outcomes in the society of thoracic surgeons nationaldatabase. The Journal of thoracic and cardiovascular surgery, 137(1):82–90, 2009.

[2] EE Roselli, NG Smedira, and EH Blackstone. Failure modes of thecarpentier-edwards pericardial bioprosthesis in the aortic position. TheJournal of heart valve disease, 15(3):421–7, 2006.

[3] Jagdish Butany and Richard Leask. The failure modes of biological pros-thetic heart valves. Journal of long-term effects of medical implants,11(3&4), 2001.

[4] I Vesely. The evolution of bioprosthetic heart valve design and its impacton durability. Cardiovascular Pathology, 12(5):277–286, 2003.

[5] I Vesely, JE Barber, and NB Ratliff. Tissue damage and calcificationmay be independent mechanisms of bioprosthetic heart valve failure. TheJournal of heart valve disease, 10(4):471–477, 2001.

[6] Richard E Clark, WM Swanson, John L Kardos, Ronald W Hagen, andRichard A Beauchamp. Durability of prosthetic heart valves. The Annalsof thoracic surgery, 26(4):323–335, 1978.

[7] TG Mackay, GM Bernacca, AC Fisher, CS Hindle, and DJ Wheatley. Invitro function and durability assessment of a novel polyurethane heartvalve prosthesis. Artificial organs, 20(9):1017–1025, 1996.

[8] Cardiovascular implants - cardiac valve prostheses. part 2: Surgical im-planted heart valve substitutes.

[9] Thierry Bourguignon, Rym El Khoury, Pascal Candolfi, Claudia Loardi,Alain Mirza, Julie Boulanger-Lothion, Anne-Lorraine Bouquiaux-Stablo-Duncan, Fabien Espitalier, Michel Marchand, and Michel Aupart. Verylong-term outcomes of the carpentier-edwards perimount aortic valve inpatients aged 60 or younger. The Annals of thoracic surgery, 100(3):853–859, 2015.

[10] Douglas R Johnston, Edward G Soltesz, Nakul Vakil, Jeevanantham Ra-jeswaran, Eric E Roselli, Joseph F Sabik, Nicholas G Smedira, Lars GSvensson, Bruce W Lytle, and Eugene H Blackstone. Long-term durabil-ity of bioprosthetic aortic valves: implications from 12,569 implants. TheAnnals of thoracic surgery, 99(4):1239–1247, 2015.

[11] Hwa Liang Leo, Lakshmi Prasad Dasi, Josie Carberry, Helene A Simon,and Ajit P Yoganathan. Fluid dynamic assessment of three polymericheart valves using particle image velocimetry. Annals of biomedical en-gineering, 34(6):936–952, 2006.

[12] G Colantuoni, JD Heliums, JL Moake, and CP Alfrey Jr. The response ofhuman platelets to shear stress at short exposure times. ASAIO Journal,23(1):626–630, 1977.

[13] Neelakantan Saikrishnan, Choon-Hwai Yap, Nicole C Milligan, Niko-lay V Vasilyev, and Ajit P Yoganathan. In vitro characterization of bicus-pid aortic valve hemodynamics using particle image velocimetry. Annalsof biomedical engineering, 40(8):1760–1775, 2012.

[14] Caitlin Martin and Wei Sun. Simulation of long-term fatigue damage inbioprosthetic heart valves: effects of leaflet and stent elastic properties.Biomechanics and modeling in mechanobiology, 13(4):759–770, 2014.

[15] Arash Kheradvar, Elliott M Groves, Lakshmi P Dasi, S Hamed Alavi,Robert Tranquillo, K Jane Grande-Allen, Craig A Simmons, Boyce Grif-fith, Ahmad Falahatpisheh, Craig J Goergen, et al. Emerging trends inheart valve engineering: Part i. solutions for future. Annals of biomedicalengineering, 43(4):833–843, 2015.

[16] Arash Kheradvar, Elliott M Groves, Ahmad Falahatpisheh, Mohammad KMofrad, S Hamed Alavi, Robert Tranquillo, Lakshmi P Dasi, Craig ASimmons, K Jane Grande-Allen, Craig J Goergen, et al. Emerging trendsin heart valve engineering: Part iv. computational modeling and exper-imental studies. Annals of biomedical engineering, 43(10):2314–2333,2015.

[17] WL Lim, YT Chew, TC Chew, and HT Low. Pulsatile flow studies of aporcine bioprosthetic aortic valve in vitro: Piv measurements and shear-induced blood damage. Journal of biomechanics, 34(11):1417–1427,2001.

7