Page 1: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

1

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Scientific Program and Requirements

What type of Spectrometer is MACS?

Which experiments is it good for?

Specifications to MACS-imize science output

Page 2: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

2

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Goals in Neutron Spectroscopy

• A central tool in condensed matter physics– Unique information about dynamic correlations– Model independent access to interaction strength– Access microscopic structure of dynamic systems

• Limited scope on current instruments– Need cm3 sized crystals– Need weeks of beam time– Need to be neutron scattering expert

• Increased sensitivity will broaden impact– Smaller samples earlier in new materials cycle– Impact in a wider range of science– Parametric studies– Comprehensive surveys for tests of theory

Page 3: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

3

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Q and E resolved spectroscopy requires 11.01.0 aQJE

To probe solids with low energy scales─ Reduce E by cooling neutrons─ Increase angular divergence to match Q

Energy scale J varies more than length scale a

Page 4: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

4

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Comparing TOF to TAS

• Large detector solid angle is possible

• E-scan without moving parts

• Can use spallation source peak flux

• Large detector solid angle is possible

• E-scan without moving parts

• Can use spallation source peak flux

• Can focus neutrons with Bragg optics

• Freely select range of energy transfer

• Can use reactor CW flux

• Can focus neutrons with Bragg optics

• Freely select range of energy transfer

• Can use reactor CW flux

TAS like

TOF like

Page 5: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

5

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Holes in a Quantum Spin Liquid

Y3+

Ni

O

AFMFM

Ca2+

Xu, Broholm et al. Science (2000)

Page 6: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

6

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

TOF Data from MAPS/ISIS TAS Data from NIST

Page 7: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

7

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Elements of Scientific Program for MACS Expand the scope for Inelastic scattering from crystals:

– 0.5 mm3 samples

– Impurities at the 1% level

– Complete surveys to reveal spin-wave-conduction electron interactions

– Extreme environments: pressure and fields to tune correlated systems

Probing short range order– Solid ionic conductors, spin glasses, quasi-crystals, ferroelectrics,

charge and spin polarons, quantum magnets, frustrated magnets.

Excitations in artificially structured solids– Spin waves in magnetic super-lattices

– Magnetic fluctuations in nano-structured materials

Weak broken symmetry phases– Incommensurate charge, lattice, and spin order in correlated electron

systems

Matsunada et al. (2002)

Saxena et al. (2000)

Lee et al. (2002)

Schreyer et al (2000)

Page 8: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

8

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Overall Requirements for MACS Maximize sensitivity

– Maximize flux on sample E≈0.2 meV, Q≈0.1Å-1

– Maximize detection solid angle at fixed Ef

– Minimize background

Optimize performance for users– Robust and reliable soft- and hard-ware– Standardized dynamic “finger prints” of sample– Versatility cannot compromise basic mode– Streamline experimental process

Start Commissioning in 3 years from now

Page 9: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

9

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Top Level Specification

MACS List of Requirements,

NCNR NIST Center for Neutron Research Building 235 Gaithersburg, Maryland 20899

MACS List of Requirements This document defines the top-level specification for Multi Axis Crystal Spectrometer, MACS, at NG0 of the NIST reactor (NBSR). The document becomes effective only if all parties whose names appear below have signed it. Changes to the signed document can be initiated by each of those parties, or their replacements, at which time a new document including the proposed revision is prepared. Changes become effective only after all parties sign the new document. Newly modified sections shall be marked as such. The old document remains part of the record. Date: December 4, 2001. Revision: A Scientist: C. Broholm …………….……………., Date ………... Engineering: C. Brocker …………….……………., Date ………... NCNR Director: P. Gallagher …………….……………., Date ………...

MACS List of Requirements,

Requirements for MACS.(1) 1. Beam extraction system 1.1 Beam tube dimensions (modified)

The aperture at La=1654 mm from the source shall be left open. The beam tube opening following this aperture shall be minimized while ensuring that the wm x hm = 441 mm x 357 mm monochromator when in monochromatic focusing geometry is illuminated as much as possible throughout the range of incident energies considering the afore mentioned aperture. For this optimization, the active part of the source shall be taken to be the geometrical optical image on the source of a 20 mm wide by 40 mm tall sample located on the sample table.

1.2 Beam line shielding and atmosphere (New) From source to sample the beam path shall be evacuated or filled with 4He wherever feasible to minimize air scattering losses. All space which is not part of the active beam tube as specified by 1.1, shall to the extent feasible, be filled with neutron shielding material. There are two exceptions to this: Shielding around the monochromator that can be viewed from the sample position shall be recessed so that it is not illuminated by the reactor beam. Furthermore, the location and characteristics of the reactor beam dump shall be chosen to minimize its contribution to the detector count rate.

1.2 Shutter (modified) The beam shutter shall be outside the biological reactor shielding. It must reduce

radiation incident on the monochromator sufficiently to allow extraction of a monochromator for repairs while the reactor is operating. It must also reduce radiation on the sample to less than 5 mR/hr when closed. The time to open or close shall be less than 15 sec. A clearly visible sign shall indicate the position of the shutter and this information shall be available to the instrument control computer. The shutter position shall be controlled from a single switch that is interlocked with a radiation exclusion zone, details of which are to be specified by NIST health physics.

1.3 Pre monochromator filters (modified) There shall be a 3-position filter exchanger immediately following or part of the

shutter mechanism. All filters shall be large enough in their transverse dimensions to accommodate the full beam as specified in 1.1. Shielding material shall surround each filter such that neutrons either pass through the filters or are absorbed in the surrounding shielding. All filters shall be cooled to 77 K or colder and remain at that temperature when fully illuminated by the reactor beam. Filter changes shall be effectuated from the instrument control computer. Filter options shall be:

MACS List of Requirements,

1.3.1 (modified) Single crystalline sapphire grade B4 or better from Crystal Systems Inc. with a beam path length of 80 mm. The orientation of the single crystalline material shall be uniform throughout the filter to within 10 degrees. However, the average crystal orientation with respect to the beam direction is unimportant and shall be chosen to minimize cost.

1.3.2 (modified) Beryllium grade I-220-H from Brush Wellman with a beam

path length of 100 mm. 1.3.3 (modified) Pyrolytic Graphite grade ZYH from Advanced Ceramics with a

total beam path length of 80 mm. The c-axis shall be oriented to within 2 degrees of the local beam direction back to the center of the source.

2. Monochromating system 2.1 General Principle (modified)

The monochromator design is based on a system in which the crystal slides along the white beam, while it rotates simultaneously. At the same time, a shielding drum holding a converging super-mirror guide rotates around an axis that is located on the line connecting the sample rotation axis and the current monochromator rotation axis. The sample position axis is permanently attached to this drum. The drum shall be tightly sandwiched between two saddle shields, yet be free to rotate. We denote the location of the monochromator at 2=900 as the reference position. The distance from the reference position to the source, to the center of the drum, and to the sample shall be minimized while maintaining all other specifications.

2.2 Pre-monochromator collimators

2.2.1 (modified) Immediately following the filter exchanger shall be a 4-position collimator exchanger. Collimator changes shall be effectuated from the control computer. It shall take less than 30 sec. to exchange collimators. All collimators shall be embedded in neutron shielding material so that neutrons either pass through the window of the collimator or are absorbed. This shielding shall have high Boron content. 2.2.2 (modified) There shall be four different collimators. All shall be radial collimators with the source as their focal point. The window of the collimator shall match the size of the beam as specified in 1.1. The spacing between blades shall be given by

crr

r

LL

Ld

0

0 ,

where Lcr is the distance from the center of the monochromator at its reference position to the down stream and broadest side of the collimator, L0r is the distance from monochromator reference position to source, and 400 mm is the length

MACS List of Requirements,

of the collimator blades. The thickness of the blades shall be 0.2 mm or less. The effective beam divergence, shall be 20', 40', 60', and 120' for the 4 collimators respectively.

2.2.3 (modified) The focal point of the collimators shall coincide with the

brightest part of the source to within 10 mm in the transverse directions and 100 mm in the longitudinal direction. The collimators shall be parallel to each other to within 0.03 degree. A line parallel to the central blade of any collimator shall pass through the monochromator rotation axis to within 5 mm.

2.3 Variable Reactor Beam aperture (modified)

As close as possible and no more than 1100 mm from the center of rotation of the active monochromator shall be a slit capable of closing the reactor beam from the dimensions specified in 1.1 to wmin=30 mm and hmin=50 mm. The aperture shall not be within line of sight of the sample in normal operation. The aperture shall be 100 mm thick, made from B4C, and covered with LiF towards the reactor. The slit shall be centered with respect to the line connecting the monochromator rotation axis to the center of the source to within 2 mm.

2.4 Monochromator exchanger (modified)

This device shall enable computer-controlled selection between two different single crystal monochromator assemblies. Exchange shall take less than 1 minute. During typical operation, exchange will occur approximately once a week. The filling of this requirement may be combined with a translation stage specified in 2.1. It shall be possible to remove either one of the monochromators from the incident beam line for service during reactor operation.

2.5 Monochromating assemblies (modified) There will be two doubly focusing monochromators. One based on pyrolytic graphite the other based on MICA, Heussler, Silicon, or Germanium to be specified at a later stage. The mechanical assemblies for each monochromator are specified separately. It shall be possible to extract both monochromating assemblies from the instrument for service while the reactor is operating. A web camera with appropriate lighting shall enable remote viewing of the monochromators for diagnostic purposed when they are driven to a specific position along the translation stage.

2.6 Monochromator shield 2.6.1 Range of take-off-angles shall be 35 to 140 . It shall take less than 1 min.

to change the take-off-angle from one extreme to the other. The monochromator translation stage and the drum rotation shall provide a setting accuracy of 0.03o for the monochromator take off angle.

2.6.2 (modified) Allowable radiation dose rate, and background, outside of

monochromatic beam: ALARA. In addition, the instrument shall lie within an

MACS List of Requirements,

interlocked radiation exclusion zone that shall be separately specified in collaboration with NIST health physicists.

2.7 Monochromator to sample super-mirror guide. (modified) From monochromator to sample shall be a converging super-mirror guide with the

largest practical critical angle of order 3Nic . The guide shall extend from as close

to the monochromator as possible until 250 mm before the sample. The inside height of the guide as a function of the distance, x, from the sample shall be given by

rL

xs

hm

hshxh1

)(

where hs=40 mm is the sample height, hm=357 mm is the monochromator height, and L1r is the monochromator to sample distance at the 2=90o reference position. The sample end of the guide shall have an inside width of 18 mm and be centered with respect to the reference line that connects the sample rotation axis to the monochromator rotation axis to within 0.5 mm. The angle between the guide sides and the reference line shall be independently variable under computer control from 0 to 2.5o with an accuracy of 0.03o. On either side of the guide where it protrudes from the monochromator drum shall be shielding that moves with the sides of the guide to function as beam defining apertures. On the sample end of the guide this shielding shall extend until the end of the guide and on the monochromator side it shall be as long as possible. The incoherent scattering cross section of materials that are illuminated by the monochromator and visible from the sample shall be minimized.

2.8 Beam optics between super mirror guide and sample

The following items shall be permanently mounted just after the super mirror guide. Their total thickness shall be less than 50 mm.

2.8.1 (modified) A monitor with a sensitivity no greater than 10-5 at 5 meV. The sensitivity shall be proportional to wave length. 2.8.2 (modified) An attenuation exchanger with four positions and capable of introducing three different planar objects into the beam under computer control. Two of the positions shall provide 10 times and 100 times attenuation respectively at 3.7 meV. These attenuators shall be permanently installed in the exchanger. The third position shall be an auxillary slot that can hold a plate with a thickness between 1 mm and 10 mm, width 40 mm and height 50 mm. When selected by the attenuation exchanger the plate shall be held in the center of the beam to within 1 mm. The attenuation exchanger shall be controlled from the main instrument control computer.

2.8.3 A computer controlled thermal neutron aperture with variable opening from closed to the full width and height of the beam. The aperture shall be centered in

MACS List of Requirements,

the beam to within 0.5 mm and its degrees of freedom shall only be the width and height of the opening. The positioning accuracy shall be better than 0.5 mm.

3. Sample table 3.1 Location

The distance from the sample rotation axis to the monochromator rotation axis shall be minimized. When the spectrometer is in its 2M=90o reference position this distance shall not be greater than 2500 mm.

3.2 Degrees of freedom provided. (modified)

3.1.1 Rotation of sample 0-360o with accuracy of 0.005 degrees. 3.1.2 Tilt of sample table +/- 12o about two mutually perpendicular horizontal axis. Accuracy better than 0.1o for loads in the range specified in 3.3. The effective rotation axes shall lie within 20 mm of beam height. 3.1.3 Elevator +/- 20 mm about beam center. Accuracy better than 1 mm. 3.1.4 Horizontal translation +/- 15 mm along two mutually perpendicular horizontal directions. Accuracy better than 0.5 mm. Translation shall occur along the sample tilt axes.

3.3 Dimensions and load capacity. (modified) The mounting surface shall lie at least 152 mm below the beam center. Load capacity shall be 400 kg on axis. Max horizontal torque shall be 4 102 Nm and shall result in less than a 0.1o tilt of the sample rotation axis from the vertical.

4. Detection system 4.1 Specification for detector bank as a whole:

The detection system shall consist of at least 20 identical and equidistant detection "channels" which view the sample with a relative offset in scattering angle that shall be minimized and shall not exceed 8o. The detection system shall cover 160o in the horizontal plane. 4.1.1 The bank of detectors must be able to rotate as a whole around the sample such that the central detector covers the range of scattering angles from -45o to 45o. The detector bank rotation shall be concentric with the sample rotation to within 0.5 mm. The setting accuracy shall be better than 0.01o

4.1.2 The effective scattering angle detected by each channel shall be within 0.03o of the nominal value

MACS List of Requirements,

4.1.3 It shall take less than 1 minute to rotate the detector bank between its extreme positions. 4.1.4 The direct beam will always be incident on a part of the detector bank. Therefore, there must be a primary beam stop between the sample and the detector bank. The beam stop shall be as close to the detector bank as possible so it cannot scatter neutrons into active detection channels. The illuminated beam stop shall produce as little neutron and hard gamma radiation as possible. The width of the beam stop shall be minimized. 4.1.5 (modified) A video camera with scintillating plate and image intensifier shall be available to position in front of any of the detection channels for radiography of the sample region or viewing Bragg reflected beams from the sample. The sensitivity shall allow real time imaging of 10,000-500,000 neutrons per second or integrated imaging of 1000 counts per second. The width of the camera at beam height shall be minimized. The images shall be available to the main computer for analysis and printing.

4.2 Specification of individual detection channel: Each detection channel shall view a 20 mm wide by 40 mm tall sample with a horizontal divergence of at least 2 degrees and a vertical divergence of 8o. While the filters and collimator need not be independently selectable, the final energy setting of each channel must be. Each channel shall consist of the following items: 4.2.1 (modified) Computer controlled selection of one of three different filters. All filters shall be cooled to a temperature of 77 K or less. Absorbing spacers shall prevent horizontal thermal neutron propagation by more than a channel width perpendicular to the beam. Filter options shall be 150 mm Thermalox 995 BeO machining stock from Brush Wellman. 100 mm Be with BW specification I220H Rev A type I from Brush Wellman. 50 mm Pyrolytic Graphite grade ZYH from Advanced Ceramics (PG). When in position each filter shall completely block the beam paths for all detection channels. Positioning of the filters with respect to each of the channels shall be reproducible to within 1 mm in all directions. PG filters must be oriented so that their crystallographic c-direction is within 1 degree of the average beam direction for the corresponding channel. The orientation of the PG filters shall be reproducible to within 0.1 degrees.

4.2.2 (modified) computer controlled selection of three different collimators with divergence angles to be specified. The collimators shall be after the filters. The transmission of the collimators shall be greater than 90%. 4.2.3 (modified) Double crystal PG(002) analyzer system covering the energy range from 2.5 meV to 15 meV. The width of the blades shall be 60 mm. The mosaic of each of the crystals remains to be specified. Starting from the sample, the first analyzer shall have a thickness of 4 mm and the second analyzer shall

MACS List of Requirements,

have a thickness of 2 mm. Both analyzers shall have fixed vertical focusing with a radius of curvature of 500 mm and their height shall be 180 mm. 4.2.4 (modified) The angular accuracy for rotation shall be better than 0.03o , which shall also be the reproducibility after 5000 motor movements. Translation of the analyzers shall be parallel to within 0.15o. The direction of translation for the first analyzer crystal shall coincide with the nominal centerline for the corresponding detection channel to within 0.03o and 0.5 mm. The direction of translation for the second crystal shall be parallel to this direction to within 0.03o. The axis of rotation for each analyzer shall pass through the center of mass of the crystals to within 1 mm. The accuracy in placement of crystals on the cylindrically focusing surface shall be better than 0.3o for analyzer number 1 and better than 0.15o for analyzer number 2. 4.2.5 (modified) It shall take less than 10 sec to change the final energy of all channels between their extreme limits and less than 2 sec to change the final energy of all analyzers by a FWHM of their energy resolution. 4.2.6 There shall be two cylindrical 3He detectors associated with each channel. One immediately following the first crystal and viewing the sample through the collimator. One following the second crystal. Both detectors shall have a partial 3He pressure and thickness to achieve 90% detection efficiency for 15 meV neutrons over the full width of the detection channel. The height of the detectors shall be minimized under the constraint that they shall be tall enough to yield 8o vertical acceptance taking into account the fixed vertical focusing of the crystals. 4.2.7 (modified) Shielding shall be sufficient to yield a fast neutron background (measured by blocking the entrance to the channel with a cadmium sheet) of no more than 30 counts per hour per channel in the second detector over the full range of incident energies and configurations of the monochromating and analyzing system. The background in the first detector shall be less than 240 counts per hour under similar conditions. 4.2.8 The energy resolution, integrated intensity, and fast neutron background shall vary by less than 10% from channel to channel. Background and integrated intensity shall vary less than 5% and mean energy less than 5 % of FWHM resolution following continuous use of the instrument for one month.

5 System Wide Requirements. 5.1 Hard and soft limits. All degrees of freedom shall be equipped with soft and/or

hardware limits that prevent any collisions but allow the full range of angles physically achievable. Sample rotation is a special case where hardware limits must be variable because of the different constraints associated with different sample environments.

MACS List of Requirements,

5.2 Automated alignment. There shall be an automated alignment protocol for all

degrees of freedom of the instrument. 5.3 Permanent electrical wiring. All wiring shall be permanently installed and

comply with applicable industry standards. 5.4 The instrument will require a dedicated and integrated software package to plan,

execute, and analyze data to take optimal advantage of the doubly focusing monochromator and the multi-channel analyzing system. Details are to be specified separately.

5.5 Radiation Safety Exclusion zone. In accordance with ALARA, the instrument

shall be within an interlocked exclusion zone. No personnel can be in this zone when the beam is on. Details are to be specified by Health Physics.

Page 10: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

10

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Strategy and Components for Incident Beam line

Goals for incident beam line

Active beam line components

Projected Performance

Page 11: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

11

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Goals for Incident beam line

• Maximize monochromatic (E=0.2 meV) flux on sample

• Minimize fast neutrons leaving enclosure

• Decouple energy and wave vector resolution

• Rapid and accurate setting of EI

• Robust design, no need to realign.

Page 12: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

12

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Increase brightness at fixed neutron production

MACS Beam

New cold source installed in 2002 has increased flux by 1.8

K 45

/n/cm 108.4 213

effT

s

Page 13: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

13

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Symmetric horizontal focusing

Sample

Sample

Source

Source

Mon

ochr

omato

r

Mon

ochr

omato

r

Focal point for monochromator

-distribution is on its bisector (symmetric reflection geometry)

Focal point for monochromator

-distribution is on its bisector (symmetric reflection geometry)

Page 14: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

14

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Space constraints require L0≈3L1

L0=5400 mm

L1=1475 mm

Page 15: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

15

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Solutions to asymmetry problem

• Create virtual source point with super-mirror guide+ Can close fast neutron aperture

– E-dependent angular divergence limited to mc ≈ 20

– Loss of efficiency when guide cannot get close to source

• Use focusing monochromator for L0≠L1

+ Can use full solid angle view of source at all E+ Can vary energy and angular distribution independently– Greater aperture for fast neutrons

Page 16: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

Asymmetric horizontal focusing

Focal point for monochromator

-distribution is not on bisector(Miscut geometry)

Focal point for monochromator

-distribution is not on bisector(Miscut geometry)

Sample

Sample

SourceSource

Mon

ochr

omat

or

Mon

ochr

omat

or

Page 17: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

17

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Rotate full device to satisfy tangency conditionRotate individual blades to achieve focusingNeutrons that go through the cracks do so to satisfy Liouville

Top view

Page 18: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

18

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Other Venetian blind monochromator configurations

SourceSource

SampleSample

resolution function can be manipulated through choice of tangent to Rowland circle

─ Good energy resolution─ Coarser transverse Q-resolution

─ coarse energy resolution─ good transverse Q-resolution

other tradeoffs with values in between

Page 19: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

19

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Two Radial Collimators that Control energy resolution

Two Radial Collimators that Control energy resolutionTranslating MACS Monochromator Translating MACS Monochromator Variable aperture that

controls Q-resolution

Variable aperture that controls Q-resolution

Cooled Be, PG, and Al2O3Cooled Be, PG, and Al2O3

Beam ShutterBeam Shutter

Overview of MACS

Design by T. D. Pike and C. Brocker

21 channel crystal analyzer21 channel crystal analyzer

super mirror guide channelsuper mirror guide channel

Page 20: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

20

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Incident beam filters

• PG filter (8 cm)– Order suppression at 13.7 meV and 14.7 meV– Fast neutron suppression E<15 meV

• Be filter (10 cm)– Order suppression E<5 meV– Fast neutron suppression E<5 meV

• Sapphire filter (8 cm)– Fast neutron suppression 15<E<20 meV

Page 21: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

40’-60’

40’

60’

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Radial Collimatordefines source size

Y. Qiu (2002)

Page 22: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

22

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Collimators in series

AA L

d

LA

NdA

BB L

dB

LA

BA

BA

BBeff LL

d

A B

Page 23: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

23

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Y. Qiu (2002)

Page 24: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

24

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

5 cm f=4 m

1428 cm2 PG

Smee et al. (2002)

Page 25: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

25

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Flux versus mosaic at fixed 0.2 meV energy resolution

Page 26: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

26

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Guide increases angular size of sample

Page 27: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

27

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Qiu and Broholm (2002)

Monte Carlo Simulation of MACS

Y. Qiu (2002)

Page 28: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

28

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Strategy and Components for Detection System

Goals for Detection System

Active Components in detector system

Modes of Operation

Page 29: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

29

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Goals for Detection System

• Maximize solid angle at fixed EF

• Minimize background count rate• Variable E and Q resolution• Alignment once a year at most• Maximize rejection ratio for EI neutrons• Energy integrating “two-axis” detectors

operating simultaneously

Page 30: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

30

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Twenty-one Channel Analyzer System

Design by C. Brocker

Page 31: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

BeO filterBeO filter

Be filterBe filter

PG filterPG filter

Collimator 1Collimator 1

8o vertically focusingAnalyzer crystals

8o vertically focusingAnalyzer crystals

“TAS” detector“TAS” detector

Energy integratingDetector

Energy integratingDetector

Design by C. BrockerCollimator 2Collimator 2

Page 32: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

32

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Neutrons detected left and right

Length of principal axes of projection of resolution function on scattering plane

EF=3.7 meV60’-60’-20’-20’

Left at analyzerRight at analyzer

Page 33: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

34

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Double analyzer is “compound lens” efficient vertical focusing

Fixed vertical focusing of analyzers

Page 34: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

35

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

meV 1 meV 7.3 fEConstant energy transfer slice

ki

kf

Q

Page 35: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

September 24, 2002 Collin L. Broholm Timothy D. Pike

36

MACS –a New High Intensity Cold Neutron Spectrometer at NIST

Assembling slices to probe Q-E volume

2 meV2 meV

1 meV1 meV

0 meV0 meV

Page 36: MACS –a New High Intensity Cold Neutron Spectrometer at NIST September 24, 2002Collin L. Broholm Timothy D. Pike 1 Scientific Program and Requirements

• Powder sample Q-E map– 8 x 30 pts x 0.2 min. = 1:00

• Single crystal constant-E slice– 8 x 100 pts x 0.5 min. = 6:40

• Single crystal complete Q-E Volume– 8 x 100 x 10 pts x 0.5 min.= 3 days

CuHpCl

SCGO

MACS –a New High Intensity Cold Neutron Spectrometer at NIST