Maintenance and Calibration
PeriFlux 6000 | peripheral pressure made intelligent
44-00315-01
Disclaimer
The information contained in this document is intended to provide general
information only. It is not intended to be, nor does it constitute, medical advice.
Under no circumstances is the information contained in this document to be
interpreted as a recommendation for a particular treatment for specific
individuals. In all cases it is recommended that clinicians perform their own
interpretations of data in conjunction with the clinical assessment of their patient.
Due to Perimed’s commitment to continuous improvement of our products, all
specifications are subject to change without notice.
All information and content in this document is protected by copyright. All rights
are reserved. Users are prohibited from modifying, copying, distributing,
transmitting, displaying, publishing, selling, licensing, creating derivative works,
or using any information available in or through the document for commercial or
public purposes. All responsibility for any liability, loss or risk, personal or
otherwise, which is incurred as a consequence, directly or indirectly, of the use
and application of any of the material in this document is specifically disclaimed.
• Diagnosing Peripheral Arterial Disease
(PAD), Critical Limb Ischemia (CLI) and Non-
healing Wounds
• Hands on, Tips and Tricks
• Interpreting Curves and Results
• Maintenance and Calibration
Introduction
The aim of this document is to summarize the basic
maintenance requirements for laser Doppler probes and
pressure cuffs used for pressure measurements.
Cleaning laser Doppler Probes
1. Wipe down the probe/holder with a disposable cloth/paper. For
probes and probe holders that are attached using tape, tape
remnants may be removed by repeated applications of clean
adhesive tape, for instance Transpore® or Blenderm®.
2. Immerse the probe holder and/or probe head and cable in a
mild detergent and water solution. Normally approximately 15
minutes is enough.
3. Rinse the probe holder and/or probe head and cable
thoroughly under clean running water. Let dry or wipe dry with
a lint-free cloth.
4. Clean the probe connectors with a lint-free cloth moistened
with alcohol to ensure good optical transmission.
Alternatively, the probes and probe holders may be cleaned
with a cloth moistened with alcohol (<70%).
Disinfection of laser Doppler Probes
The following disinfectants are recommended for the
probes and probe holders:
• Alcohol (70%)
• Rely+On™ PeraSafe™ (DuPont™)
• CIDEX® OPA (Johnson&Johnson)
• MadaCide-FD (MADA Inc.)
• Sekusept Plus (Ecolab)
Note! Do not immerse the probe connectors in any liquid.
Cleaning of Cuffs
• If possible, all straight segmental cuffs should be
washed daily or at minimum once a week. This can
be accomplished by hand wash or by using a
washing machine with a mild detergent and gentle
cycle.
• Using a soft brush is useful to remove any organic or
inorganic material without damaging the cuff fabric.
• If there is a lot of protein contamination, such as
blood, an enzyme detergent will help remove the
organic material.
Cleaning of Cuffs cont.
1. If cuffs are grossly contaminated, keep the cuffs wet or moist to
prevent the material from drying, making cleaning more difficult.
2. Remove the bladders from the cuffs.
3. If the bladder is contaminated, wash it with soapy water and
then rinse.
Do not get water inside the bladder or tubing.
4. Add a laundry soap or mild detergent to either warm or cold
water and put the cuff covers in.
5. After washing, open the end of the cuff, where the bladder is
placed, and hang the cuff up to dry.
6. Make sure both the outside and inside are dry, including the
Velcro, before reinserting the bladders. Do not disinfect until
cuffs are completely dry.
Disinfection of Cuffs
The following disinfectants may be used on straight
segmental and digit cuffs:
• Hydrogen peroxide
• Hydrogen peroxide with silver (Sanosil®)
• Hydrogen peroxide
• Peroxyacetic acid
• Silver (Steriplex™)
• Silver with citric acid (PureGreen24™ & SpectraSan™ 24)
• Octyl decyl dimethyl ammonium chloride with dioctyl dimethyl
ammonium chloride with didecyl dimethyl ammonium chloride with
dimethyl benzyl ammonium chloride (Protex™)
• 99 % isopropyl alcohol
• T-Spray™
• Steriplex SD
Cleaning of Tubing
The tubing (Tygon S-50-HL Hospital and Surgical
tubing) can be cleaned with soap and water, or
sterilized with autoclave (steam 30 minutes at 15 psi,
121 °C) or radiation.
Checking laser Doppler Calibration
To obtain reliable measurements, a monthly check of the calibration of
the probe and the PF 6010 unit is recommended.
1. Start the instrument at least 5 minutes before check of calibration.
2. Select MENU|Maintenance|PF6010 Check Calibration.
3. In the dialog that opens, select the function unit(s) to be checked.
In case of several units, they will be checked in sequence.
Checking laser Doppler Calibration cont.
4. Check that the serial number on the connected probe is the
same as the serial number indicated in the dialog. If not, close
the dialog and go to the Change Probe procedure. Otherwise,
tap Next.
5. Place the probe in the motility standard:
– Shake the bottle containing the motility standard gently. Carefully fill
the glass container with motility standard. Avoid bubbles. Place on
a vibration free surface. •
– Place the probe tip 5 mm into the liquid. Tap cautiously to remove
potential air bubbles at the tip.
Note! When checking calibration of a PROBE 407, the probe
should be placed in the motility with a probe holder attached in
order to achieve an accurate result.
Checking laser Doppler Calibration cont.
6. Tap Start to start the check. Wait for the first step to complete.
7. If the first step is completed successfully, place the probe on
the zeroing disc using a double adhesive tape strip. Tap Start.
8. Wait for the second step to complete.
9. In case of successful verification, continue with the next unit
(step 4–9). When finished with all units, close the wizard by
tapping Close.
10. If the check fails, follow the instructions on the screen.
11. When finished, drain the motility standard back into the bottle
and secure the container lid tightly.
Changing laser Doppler Probe
When swapping laser Doppler probes, the correct probe, identified by
probe must be identified by probe type and serial number, and set,
using the Change Probe procedure.
To change probe:
1. Start the instrument at least 5 minutes before calibration.
2. Insert the probe to be used.
3. Select MENU|Maintenance|PF6010 Change Probe.
4. In the dialog that opens, select the function unit(s) to be changed.
Tap Next.
5. In the dialog that opens, select the serial number of the new probe
using the drop down menu. Tap OK.
6. Perform a check of calibration of the probe by following the
instructions on the screen.
Verification of Pressure Measurements
Enter the PF 6050 Verification Mode.
1. Select MENU|Settings|Advanced.
2. Check the Use verification mode check box. Close the
Advanced dialog by tapping OK.
3. Connect an external pump with manometer to one of the
outlets.
4. Check that the instrument displays 0 ± 0.5 mmHg.
5. Pump up to a value of 150 ± 30 mmHg. Check that the
external manometer and the instrument display the
same value ± 3 mmHg.
Verification of Pressure Measurements cont
6. Pump up to a value of 280 ± 10 mmHg. Check that the
external manometer and the instrument display the
same value ± 6 mmHg.
7. Repeat the procedure for all pressure outlets.
8. Select MENU|Settings|Advanced and un-check the
Verification mode check box.
9. Close the Advanced dialog by tapping OK.
If the external manometer and the instrument display
different values outside the given intervals, contact your
Perimed representative.