meander-line structure for eddy-current sensing of magnetic nanoparticles
DESCRIPTION
Oral presentation in ISMANAM 2013 in Torino (Italy)TRANSCRIPT
![Page 1: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/1.jpg)
Meander-line structure for eddy-current sensing of
superparamagnetic nanoparticles
M. Rivas, D. Lago-Cachón, J. C. Martínez-García, J. A. GarcíaNanobioanalysis Research Group, University of Oviedo, Spain
Nano
Bio
Análisis
![Page 2: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/2.jpg)
Magnetoimpedance-baseddetection(1/4)
![Page 3: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/3.jpg)
Surface modified amorphous ribbon based magnetoimpedance biosensor
Magnetoimpedance (MI) changes due to surface modification of the sensitive element caused by human urine, were studied with the aim of creating a robust biosensor working on a principle of electrochemical magnetoimpedance spectroscopy. A biosensor prototype ...
Cited by 8 Related articles All 5 versions Cite
Microwire array for giant magneto-impedance detection of magnetic particles for biosensor prototype
A new giant magneto-impedance (GMI) biosensor prototype based on a glass-coated microwires array is reported. The optimal measurement conditions and the influence of a liquid suspension of commercially available polymer-based magnetic microparticles on ...
Cited by 20 Related articles All 5 versions Cite
Giant magnetoimpedance-based microchannel system for quick and parallel genotyping of human papilloma virus type 16/18
... Quick and parallel genotyping of human papilloma virus (HPV) type 16/18 is carried out by aspecially designed giant magnetoimpedance (GMI) based microchannel system. Micropatternedsoft magnetic ribbon exhibiting large GMI ratio serves as the biosensor element. ...
Cited by 19 Related articles All 9 versions Cite
Surface modified amorphous ribbon based magnetoimpedance biosensor.
Microwire array for giant magneto-impedance detection of magnetic particles for biosensorprototype. ...
Cited by 6 Related articles All 6 versions Cite
Magneto-impedance sensor for biomedical applications
... A giant-magnetoimpedance-based biosensor prototype was designed and tested by using simpleglass ... the geometry of the dc excitation external magnetic field on the magneto-impedance
![Page 4: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/4.jpg)
Magnetoimpedance
Co70Fe5Si10B15
ribbon
![Page 5: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/5.jpg)
Samples and methods
Test fixture ribbon to impedance analyser(1-110 MHz) + Helmholz coils (dc field)
Superparamagnetic 10 nmFe3O4 nanoparticles
NPs deposited onto blotting paper strips
![Page 6: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/6.jpg)
NPs detection with Co (GMI)
![Page 7: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/7.jpg)
NPs detection
Magnetoimpedancecurve (detail)Co70Fe5Si10B15 ribbon
Increase of impedance640 mg Fe3O4
![Page 8: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/8.jpg)
Cu-based detection(2/4)
![Page 9: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/9.jpg)
NPs detection with Cu
![Page 10: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/10.jpg)
NPs detection with Cu
Cu-ribbon
Co-ribbon
![Page 11: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/11.jpg)
Origin of the detection
Changing magnetic field
Induced electromotiveforces
Switching magnetic moment
![Page 12: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/12.jpg)
Printed Circuit Board(3/4)
![Page 13: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/13.jpg)
Printed Circuit Board
![Page 14: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/14.jpg)
NPs detection with PCB1
PCB1
Cu-ribbon
x 5
![Page 15: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/15.jpg)
A frequency-related detection
![Page 16: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/16.jpg)
Skin effect
400 mm
35
mm
![Page 17: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/17.jpg)
Skin effect
Active area
![Page 18: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/18.jpg)
Meander-line circuit
![Page 19: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/19.jpg)
NPs detection with meander-structure
Meander
PCB1
x 3
![Page 20: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/20.jpg)
Planar spiral-like circuit
![Page 21: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/21.jpg)
NPs detection with spiral-sensor
PCB1
Spiral
x 7
![Page 22: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/22.jpg)
Proximity effect
Cross-section of three parallel traces
![Page 23: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/23.jpg)
Proximity effect
Cross-section of three parallel traces
![Page 24: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/24.jpg)
Planar spiral-like circuit
Ratio over
![Page 25: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/25.jpg)
Planar spiral-like circuit
![Page 26: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/26.jpg)
NPs detection with 5-turn spiral
5-turns
PCB1
x 100
![Page 27: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/27.jpg)
Conclusions(4/4)
![Page 28: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/28.jpg)
Detection with copper: highly sensitive, low-cost and simple fabrication/handling.
Nanoparticles detection
![Page 29: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/29.jpg)
Thinner, narrower, numerous traces.
Higher frequency.
Spiral-like configuration.
Best choices
![Page 30: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/30.jpg)
The physical origin of the detection is the induction of emf in the copper/any conductor.
Interpretation
![Page 31: Meander-line structure for eddy-current sensing of magnetic nanoparticles](https://reader033.vdocuments.net/reader033/viewer/2022042816/558b4aedd8b42a272a8b45ce/html5/thumbnails/31.jpg)
Yes, we can detect just with copper.
Widening applications?
Challenges