DESCRIPTION

MProbe VisHR (http://semiconsoft.com/html/download/brochure/MProbeVisVisHR_brochure2014.pdf) is used to measure coating thickness of sheet metal. Typical coatings includes, anti-finger, anti-rust and oil coating

TRANSCRIPT

Page 1: Measurement of coating Thickness on sheet_metal

83 Pine Hill Rd. Southborough, MA 01772 Phone +1.617.388.6832 Fax. +1.508.858.5473

email: [email protected] http://www.semiconsoft.com

Thin Film Measurement solution Software, sensors, custom development and integration

Measurement of coatings on sheet metal.

Varieties of specialized coatings are deposited in the sheet metal production.

Some of the most common are anti-rust, anti-finger and oil coating. Traditionally, the thickness of these coating is being measured using IR absorption. This method has two serious limitations: sensitivity to the thin films (<2 μm) is low and substrate color variation can affect the measurement results. MProbe system is free from above limitation and can easily measure thickness in these applications Example 1. Measurement of anti-rust coating thickness on zinc-dipped steel sheet. Measurements were done using MProbe VisHR system in 700-1000nm range.

Fig.1 Measurement of the thin anti-rust coating. Fit of the model to measured data Thickness 1.33µm

Page 2: Measurement of coating Thickness on sheet_metal

83 Pine Hill Rd. Southborough, MA 01772 Phone +1.617.388.6832 Fax. +1.508.858.5473

email: [email protected] http://www.semiconsoft.com

Fig.2 Measurement of the thick anti-rust coating. Peak position indicates film thickness Thickness: 10.35µm

Page 3: Measurement of coating Thickness on sheet_metal

83 Pine Hill Rd. Southborough, MA 01772 Phone +1.617.388.6832 Fax. +1.508.858.5473

email: [email protected] http://www.semiconsoft.com

Example 2. Measurement of the anti-finger film on stainless steel. The measurements were taken using MProbe Vis system (400nm -1000nm wavelength range)

Fig. 3 Reflectance spectra of two samples. Reflectance spectra are offset due to a different reflectivity of the substrate. However, the film thickness are almost the same (1.943 μm and 1.945 μm). Using IR absorption will yield significantly different results…

Page 4: Measurement of coating Thickness on sheet_metal

83 Pine Hill Rd. Southborough, MA 01772 Phone +1.617.388.6832 Fax. +1.508.858.5473

email: [email protected]

Fig. 4 Results of the measurement. Thickness:1945nm