Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2 Quadrupole Mass Spectrometers are gas-specific instruments, the fragmentation pattern may lead to spectral overlap several sources of instabilities how to operate the instrument if the SEM or C-SEM is the major source of drift? spectral overlap of gaseous components lead to MCD-Measurements how to calibrate some (two) examples check whether re-calibration is required

Upload: desiree-lankford

Post on 31-Mar-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, N. Müller, W. Neff, Bled April 2012

Quadrupole Mass Spectrometers are

gas-specific instruments, the fragmentation pattern may lead to spectral overlap

several sources of instabilities

how to operate the instrument if the SEM or C-SEM is the major source of drift?

spectral overlap of gaseous components lead to

MCD-Measurements

how to calibrate

some (two) examples

check whether re-calibration is required

Page 2: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Functional units of every Quadrupole Mass-Spectrometer:

An ion-current, amplified ion-current or a count rate is the original out-put of these instruments.

Faraday

U+Vcost

Electron multiplier

Ion sourcesupply

RFgenerator

HVsupply

Electrometerpreamplifier

Quadrupole controller

Ion sourceMass filter

Page 3: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Ion-Source

The amount of ions generated in the ion source depends on the partial pressure ofthe individual species and on the energy of the electrons used for electron impact ionization.

As with every ionization gauge, the signal generated is gas-specific.

Page 4: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Ionization Probability compared to nitrogen (electron energy : 70 eV)

The same current for He and Ar makes a difference in pressure of a factor of 8.

Page 5: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Stability of Quadrupole Mass Spectrometers: (see*)

The cracking pattern and the resolution versus mass range may change.

The mass-scale may shift.

The amplification of the SEM or C-SEM may drift.

* 8th European Vacuum Congress EVC-8 Berlin,. 23.-26. Juni 2003

43.4 43.6 43.8 44.0 44.2 44.4 44.61E-13

1E-12

sensitivity

peak positionS

igna

l/Ion

Cur

rent

amu

0 10 20 30 40 50 600

20

40

60

80

100

trend: aging of the channeltron

QMG 422 with Faraday Prisma with FaradayPrisma with Channeltron

Pa

rtia

l Pre

ssu

re S

en

sitiv

ity [

arb

uni

ts]

Time [hours]

Page 6: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

change of the cracking pattern (due to the electron energy)

amu

Ar++

Page 7: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

There are different instruments on the market which are used for various applications.

Whether the mass scale, the resolution, the fragmentation pattern and the amplification may drift and at which time scale rather depends on the design of the instrument and the quality of the electronics used to drive the instrument.

No general figures and numbers can be given for all these “RGA’s”

Anyway: Calibration is required because the ionization is gas specific.

Fragmentation leads to an overlap of different species at the same mass.

If the drift of the SEM or C-SEM is the major (only) source of instabilities, is there a method to get rid of these drifts?

Page 8: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

The following considerations are valid for Quadrupole Mass Spectrometers, in which the drift of the amplifier is the major source of instabilities, the other’s are neglegible. (see EVC-8 in 2003)

Calibration:To minimize the effort for calibration, gas mixtures which contain several components in a chemical inert carrier gas are used. The pre-condition is that there is no overlap between the components contained in the mixture.Automatically one then gets the sensitivity of the components relative to the carrier gas and the cracking pattern of each component.

If the overall sensitivity of the instrument changes, then there is no influence on the result as long as the change occurs slow compared to a measurement cycle.

Furthermore if a calibration library exists with Argon and Nitrogen as carrier gas the sensitivity for Ar and for N2 have to be determined only in order to combine both libraries.

Page 9: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

nature is not always kind!

Page 10: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Mass-Spectrum of CO2 recorded at an electron energy of 70 eV.

Page 11: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Model spectrum (Origin of peaks)

0 5 10 15 20 25 30 35 40 45 50

m/e

Hydrogen Nitrogen Oxygen Water

Carbon dioxide Argon Carbon monoxide

Page 12: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

How to determine concentrations within a gas mixture of N2, CO and CO2 ??

a) Measure the concentration for CO2 at m/e = 44b) Subtract the contribution to mass 16 from CO2 (need to know the cracking pattern)c) Determine the concentration of CO at m/e = 16 then (hope there is no O2 in the sample!)d) Subtract the contribution of CO to the spectrum from the measured spectrum.e) Determine the concentration of nitrogen either at m/e=14 or at mass 28 now.

Page 13: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

signal at mass 12 = PN2*SN2(12) + PCO*SCO(12) + PCO2*SCO2(12)

signal at mass 14 = PN2*SN2(14) + PCO*SCO(14) + PCO2*SCO2(14)

signal at mass 16 = PN2*SN2(16) + PCO*SCO(16) + PCO2*SCO2(16)

where PN2 = the pressure of N2 and SN2(12) is it’s sensitivity factor at mass 12 (which is zerofor mass 12) and so on for all the three components.

Three unknown and three equations.

Today’s computers solve such a system of linear equations in fractions of seconds.

Provided there are only the three components present and the sensitivity factors are preciselyknown, one get’s also a reasonable result.

However, here we still need to know the absolute sensitivity factors for the individual components.

Do we really?

Mathematics:

Page 14: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

S12 SN2(12) SCO(12) SCO2(12) PN2

S14 = SN2(14) SCO(14) SCO2(14) X PCO

S16 SN2(16) SCO(14) SCO2(14) PCO2

The vector is the solution for the linear equation defined above.

PN2

PCO

PCO2

If we multiply the coefficients SN2, SCO and SCO2 above by a factor F, then

the vector 1/F X is a solution of the modified system of linear equations.

PN2

PCO

PCO2

Page 15: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Assume that the ratio of the individual sensitivity factors is precisely known,

then the correct ratio of the partial pressures will be achieved:

The result will be F*PN2, F*PCO and F*PCO2 with an undetermined factor F.

The assumption that only these three components are present in the gas leads to

[N2] + [CO] + [CO2] = 100%

which can be easily computed by the measured ratio of the partial pressures.

Page 16: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

first have a look what’sthe residual gas composition

Page 17: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

It’s not only the mass spectrometer whichdetermines the results.The vacuum system may be as important as well.

Subtract backgronund, however how stable is thebackground from the residual gas?

Page 18: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

air – inlet and calibration by the known

composition of air:

Page 19: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

still there is no overlap!

Page 20: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Page 21: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

calibrate CO2 to Ar moreprecisely by using a gasmixture of known composition

Page 22: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Page 23: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

In this way an existing calibration matrixcan be expanded.

Page 24: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Example: gas-mixture specified as

CO2 17.2%CO 18.5%Ar 19.2%

N2 45.1%

There is spectral overlap with this mixture.

Page 25: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Question?:re-calibrate with this mixturewas the previous mixture more precise?

pressure change within a measurement cycle

as there is no N2, CO, CO2, Ar in the sample, the assumptionthat the sum of all is 100% is not justified, result not applicable.

Page 26: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

changes in process pressure do notinfluence the result

real, or due to changingbackground?

Page 27: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

0 2 4 6 8 1090

92

94

96

98

100

The MID-Signal may be modulated by pressure fluctuations, thereforea short cycle time is an advatage.A too short dwell time however may lead to an inacurate measurement of

the ion current (10-12

ampere).

MID-Cycle: preselected masses are measured cyclic during a preprogrammed dwell time

Dwelltime Mass1 Dwelltime Mass2 ..... Dwelltime Mass n

MID-Cycle Time versus Pressure Fluctuations

+/- 1%

Tot

al P

ress

ure

[arb

. uni

ts]

time [arb. units]

Page 28: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

proces pressure about 1*10-7 mbar, that’s too low for realisticmeasurement, because much to near to base pressurere-adjustment of background subtraction

Page 29: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

How to check whether

re-calibration is required?

Page 30: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

record a test-gas spectrumfrom time to time

Page 31: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

Ar

Test-Gas-Mixture

export to excel

Page 32: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

amu

Difference of two spectra, each normalized to 1000 at it’s maximum

difference +/- 3 % ok

Page 33: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

amu

mass scale shift

difference of two spectra squared:

Page 34: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Measurement of Gas Concentrations by Mass Spectrometers

Günter Peter, Bled April 2012

How to operate the instrument if the SEM or C-SEM is the major source of drift

spectral overlap of gaseous components lead to

MCD-Measurements (system of linear equations solved on line)

how to calibrate

examples

Suggestion how to check whether re-calibration is required

Page 35: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Apr 11, 202335 file_name

Page 36: Measurement of Gas Concentrations by Mass Spectrometers Günter Peter, N. Müller, W. Neff, Bled April 2012 Quadrupole Mass Spectrometers are gas-specific

Günter Peter, Bled April 2012