measuring aniseikonia using scattering filters … · measuring aniseikonia using scattering...

105
MEASURING ANISEIKONIA USING SCATTERING FILTERS TO SIMULATE CATARACT By JASON WILSON RODERICK J. FULLARD, COMMITTEE CHAIR DAWN K. DECARLO ROBERT P. RUTSTEIN A THESIS Submitted to the graduate faculty of The University of Alabama at Birmingham, in partial fulfillment of the requirements of the degree of Master of Science BIRMINGHAM, ALABAMA 2011

Upload: truongmien

Post on 06-May-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

MEASURING ANISEIKONIA USING SCATTERING FILTERS TO SIMULATE CATARACT

By

JASON WILSON

RODERICK J. FULLARD, COMMITTEE CHAIR DAWN K. DECARLO

ROBERT P. RUTSTEIN

A THESIS Submitted to the graduate faculty of The University of Alabama at Birmingham,

in partial fulfillment of the requirements of the degree of Master of Science

BIRMINGHAM, ALABAMA

2011

Copyright by Jason Wilson

2011

iii

MEASURING ANISEIKONIA USING SCATTERING FILTERS TO SIMULATE CATARACT

JASON WILSON

DEPARTMENT OF VISION SCIENCE

ABSTRACT

The relationship between anisometropia and aniseikonia (ANK) is not well

understood. Ametropic cataract patients provide a unique opportunity to study this

relationship after undergoing emmetropizing lens extraction. Because light scatter may affect

ANK measurement in cataract patients, its effect should also be evaluated.

The Basic Aniseikonia Test (BAT) was evaluated using afocal size lenses to produce

specific changes in retinal height. Several light scattering devices were then evaluated to

determine which produced effects most similar to cataract. Contrast sensitivity and visual

acuity (VA) losses were measured with each device and compared to those reported in

cataract. After determining the most appropriate light scattering device, twenty healthy

patients with normal visual function were recruited to perform the BAT using the filters to

simulate cataract.

Cataract patients were recruited from Vision America and the University of Alabama

at Birmingham School of Optometry. Patients between 20 and 75 years of age with at least

20/80 VA in each eye, ≥ 2D ametropia, and normal binocular function were recruited.

Stereopsis and ANK were tested and each patient completed a symptom questionnaire.

ANK measurements using afocal size lenses indicated that the BAT underestimates

ANK, although the effect was minimal for vertical targets and darkened surroundings, as

previously reported. Based on VA and contrast sensitivity loss, Vistech scattering filters

iv

produced changes most similar to cataract. Results of the BAT using Vistech filters

demonstrated that a moderate cataract but not a mild cataract may affect the ANK

measurement.

ANK measurements on cataract patients indicated that those with ≥ 2 D ametropia in

each eye may suffer from induced ANK after the first cataract extraction. With upcoming

healthcare reform, unilateral cataract extraction may be covered, but not necessarily bilateral,

depending on patient VA in each eye. However, a questionnaire about symptoms at each visit

in the current study showed that visual comfort did not improve after unilateral, but did

improve after bilateral, cataract extraction. This indicates that quality of life should be better

in bilateral cataract patients only if both cataracts are removed. This is supported by the

findings of other studies.

Keywords: Aniseikonia, Anisometropia, Cataract, Ametropia

v

ACKNOWLEDGEMENTS

I sincerely thank Dr. Fullard for is dedication, guidance, and motivating nature

throughout this process. He has kept on track with this project throughout its entirety, and his

lightheartedness has keep my moral, as well as the moral of everyone else in the lab, up in

rough times.

I also want to thank Drs. Rutstein and DeCarlo. Dr. Rutstein for keeping the Vision

America project alive and kicking, and Dr. DeCarlo for encouraging me and giving very

helpful suggestions for the simulated cataract part of the study.

The staff at Vision America deserves special thanks for working around me as I recruited and

tested their patients. Drs. McCurdy, Batson, and Helton as well as Billie Lively and Joy at the

front desk have been an invaluable asset.

My parents, Larry and Jeannine, deserve so much credit that it cannot be put into

words. Their seemingly never ending faith in their son seems crazy at times, but it has help

me get through this. It will also continue to motivate through optometry school and beyond.

Finally, my sincerest thanks to my fiancé, Peggy. She has packed up her life and moved away

from home to be with me for this. There is nothing more I could ask of her. I thank her for

being there, listening to what I have to say even though she may not understand it, critiquing

presentations, and never flinching at the talk of the long journey to get where I want to be. I

am ever thankful that I will not have to do it alone.

vi

TABLE OF CONTENTS

Page

ABSTRACT ....................................................................................................................... iii ACKNOWLEDGMENTS .................................................................................................. v LIST OF TABLES ............................................................................................................. ix LIST OF FIGURES ............................................................................................................ x LIST OF ABBREVERATIONS ....................................................................................... xii CHAPTER

1 INTRODUCTION ...................................................................................................... 1

Anisometropia and Aniseikonia ................................................................................ 2 Causes of Aniseikonia ............................................................................................... 3 Symptoms of Aniseikonia ......................................................................................... 4 Measuring Aniseikonia ............................................................................................. 4 Vernier Alignment .................................................................................................... 6 Cataract and Aniseikonia .......................................................................................... 8 Other Factors affecting Aniseikonia ......................................................................... 9 Cataract Simulating Devices ................................................................................... 10 Bangerter Foils ................................................................................................... 10 Vistech Scattering Goggles ................................................................................ 10 Tiffen ProMist Black Filters .............................................................................. 10 Optical Defocusing Lenses ................................................................................ 11 Visual Acuity .......................................................................................................... 11 Contrast Sensitivity ................................................................................................. 12 2 AIMS AND RATIONALE ....................................................................................... 14

Null Hypothesis ....................................................................................................... 14 Alternative Hypothesis ............................................................................................ 14 Aims ........................................................................................................................ 14 Specific Aim 1.................................................................................................... 14 Specific Aim 2.................................................................................................... 15

vii

3 EXPERIMENTAL DESIGN ........................................................................................ 16

Validating the Basic Aniseikonia Test .................................................................... 16 Determining appropriate scattering device ............................................................. 16 Criteria for Selecting Study Participants ................................................................. 18 Simulated Cataract Participants ......................................................................... 18 Cataract Patients ................................................................................................. 19 Data Collection........................................................................................................ 19 Simulated Cataract Participants ......................................................................... 19 Cataract Patients ................................................................................................. 21 4 RESULTS ................................................................................................................. 25 Validating Basic Aniseikonia Test .......................................................................... 25 Vernier Alignment .................................................................................................. 28 Determining Appropriate Scattering Device for Subsequent Testing..................... 34 Bangerter Foils ................................................................................................... 34 Optical Defocus Lenses ..................................................................................... 37 Vistech Scattering Filters ................................................................................... 39 Tiffen ProMist Black Filters .............................................................................. 41 Final Selection of Simulated Cataract Filter ........................................................... 46 Simulation of Cataract using Vistech Filters .......................................................... 47 Cataract Patient Study: Specific Aim 2 ................................................................... 52 5 DISCUSSION ........................................................................................................... 63 Key Findings ........................................................................................................... 63 Validating Basic Aniseikonia Test .......................................................................... 63 Vernier Alignment .................................................................................................. 64 Simulated Cataract .................................................................................................. 65 Cataract Patients ...................................................................................................... 67 Correction of Refractive Error for Working Distance of BAT .......................... 68 Limitations and Possible Future Studies ................................................................ 71 Simulated Cataract ............................................................................................. 71 Cataract Patient Study ........................................................................................ 71 Future Studies .................................................................................................... 72 Conclusions ........................................................................................................ 73 LIST OF REFERENCES .................................................................................................. 75 APPENDICES .................................................................................................................. 79

viii

A. INSTITUTIONAL REVIEW BOARD LETTER .............................................. 79

B. CATARACT PATIENT DATA COLLECTION FORMS ................................ 80

C. CATARACT PATIENT DATA ......................................................................... 86

ix

LIST OF TABLES Table Page

1 Details of Scattering Devices ..................................................................................... 18

2 Expected ANK Results using BAT with Afocal Size Lenses .................................... 26

3 Preliminary MARS contrast sensitivity Results of all Filters .................................... 43

4a Preliminary Visual Acuity Results of all Filters ....................................................... 44

4b Preliminary Visual Acuity Results of all Filters ....................................................... 45 5 Comparative Effect of each Filter Type on ANK Measurement ............................... 47 6 VA Results using Snellen Chart for Participants Enrolled in Simulated Cataract Study ........................................................................................... 50 7 CS results using MARS Contrast Sensitivity Test for Participants Enrolled in Simulated Cataract Study ........................................................................ 51 8 Expected ANK Results using BAT on Cataract Patients ........................................... 53 9 ANK Means of Cataract Patients with Ametropia Type, Consistency to Prediction, and Predicted ANK due to SM ....................................... 58 10 Stereopsis Results of Cataract Patients ...................................................................... 59 11 Results of One Way ANOVA and t test on Cataract Patient Questionnaire Results ................................................................................................ 62

x

LIST OF FIGURES

Figure Page 1 Example of BAT Display ............................................................................................. 6

2 Example of Vernier Alignment Task Display .............................................................. 7

3 Spectacle Magnification Formula ................................................................................ 9

4 Formula used to Predict the Contribution of SM to ANK ...................................... 24

5 Normalized Preliminary BAT Results using Afocal Size Lenses in the Absence of

Defocusing or Scattering Devices .............................................................................. 27

6 Vernier Alignment Task Results with “No Filter” Versus Optical

Defocus Lenses .......................................................................................................... 29

7 Vernier Alignment Task Results with “No Filter” Versus Vistech Filters ................ 31

8 Vernier Alignment Task Results with “No Filter” Versus Tiffen

Pro Mist Black Filters 1,2,3,5, and 6.......................................................................... 32

8a Vernier Alignment Task Results with “No Filter” Versus Tiffen

Pro Mist Black Filters 1,2,3,5, and 6 with no Error Bars .......................................... 33

9 Normalized Preliminary BAT Results using Bangerter Foils .................................... 36

10 Normalized Preliminary BAT Results using Optical Defocusing Lenses ................. 38

11 Normalized Preliminary BAT Results using Vistech Scattering Filters .................... 40

12 Normalized Preliminary BAT Results using Tiffen Pro Mist Black Filters .............. 42

13 Normalized BAT ANK Results using Vistech Filters 1 and 2................................... 49

xi

14 BAT Measurements from Cataract Patients ............................................................... 54

15 Vision America cataract patient questionnaire results ............................................... 60

xii

LIST OF ABBREVIATIONS

ANK Aniseikonia ANOVA Analysis of Variance BAT Basic Aniseikonia Test CS Contrast Sensitivity D Diopter IOL Intraocular Lens LOGMAR Logarithmic Minimal Angle of Resolution OD Ocular Dexter OS Ocular Sinister OU Ocular Utirque RCT Repeatability Coefficient Test SD Standard Deviation SEM Standard Error of the Mean Sim K Simulated Keratometry SM Spectacle Magnification TPM Tiffen Pro Mist Black Filter UABSO University of Alabama at Birmingham School of Optometry VA Visual Acuity

1

CHAPTER 1

INTRODUCTION

The goals of this study are to determine whether or not light scattering affects

aniseikonia (ANK) measurement using the Basic Aniseikonia Test (BAT) and to better

understand the effects of emmetropizing cataract surgery on previously ametropic

patients. In particular, do symptoms of ANK develop after surgery in previously

asymptomatic patients? With upcoming healthcare reform that may affect the conditions

required to allow reimbursement of bilateral rather than unilateral cataract extraction,

there is a pressing need for a greater understanding of ANK to provide greater incentive

for clinical testing and treatment of this condition.

The relationship between anisometropia and ANK is not very well understood.

Ametropic patients undergoing non-simultaneous bilateral cataract surgery provide a

unique opportunity to study clinically induced ANK after patients have had the first

cataract extracted.

The original goal of this study was to look at previously anisometropic refractive

surgery patients to determine the effect of creating bilateral emmetropia (therefore

removing anisometropia) after bilateral surgery. Due to a receding economy and

refractive surgery not typically covered by insurance providers, it was not feasible to

obtain definitive data on such a group. Therefore, the effect of unilateral cataract surgery

on patients with at least 2D ametropia in each eye was instead investigated. The rationale

2

was that anisometropia would be induced after the first surgery and subsequently

eliminated if a second surgery was performed on the contralateral eye. An additional

consideration was the possibility that the aniseikonia measuring device itself could be

influenced by the light scattering effect of cataract – in particular unilateral cataract.

Given the relatively modest number of qualifying cataract patients, the study of light

scattering effects on the ANK measurement became the primary emphasis of this thesis.

Anisometropia and Aniseikonia

Anisometropia is the departure from emmetropia in which the two eyes have

ametropia of unequal magnitude. If a sufficient difference exists between the refractive

powers of the patient’s eyes, ANK may develop. ANK is a binocular condition in which

the left and right eyes perceive the same object as having different sizes (Rutstein and

Daum 1998). The visual cortex typically has problems fusing the separate images with a

size difference of 2% or greater. ANK exists in 1% to 3.5% of the population, and can

cause dizziness, headaches, amblyopia, reduction of stereopsis, and other impairments of

visual function (Jimenez et al. 2002). Although ANK is often diagnosed in conjunction

with anisometropia, one can have ANK without being diagnosed with anisometropia

because the relationship between anisometropia and ANK is very poorly understood, and

much more research will be required to better understand this relationship.

The original research on ANK occurred at Dartmouth Eye Institute in 1932 by

Adelbert Ames Jr. and Professor Charles Proctor, of the Department of Physics at

Dartmouth College, with the assistance of the Eastman Kodak Company. Although their

original study involving a camera model to simulate image formation and aberrations of

3

the human eye was not a success, it prompted a ten-year series of studies on the

physiologies of binocular anomalies. In 1940, Walter Lancaster became the head of the

Dartmouth Eye Institute and coined the term aniseikonia. After Ames completed his

research, Kenneth Ogle and his colleagues continued on, making the measurement and

correction of ANK the hallmark of the Dartmouth Eye Institute (Achiron et al. 1997).

Causes of Aniseikonia

ANK has many possible causes including: differences in ocular size, axial length,

refractive error, and retinal or neural distribution of ganglion cells and receptive fields

respectively. Retinotopic mapping allows the formation of images, and if one retina

developed in a different manner due to a larger eye or longer axial length, that image will

be perceived as being larger by the cortex. A long eye will produce a larger retinal image,

but a larger eye may have the receptors placed further apart making cortical receptive

fields larger and may counterbalance the larger retinal image. ANK is often associated

with oblique astigmatism as well as retinal diseases such as epiretinal membrane and

vitreomacular traction. Isometropic ANK is the condition in which one eye is larger than

the other, or has a different ganglion cell density, but both eyes have the same magnitude

of ametropia (Phillips, 1958).

Several other ocular parameters may contribute to ANK including axial length

differences between eyes, location of the crystalline lens (or intraocular lens (IOL)), and

corneal power. Light scattering produced by cataract may affect ANK as well.

4

Symptoms of Aniseikonia

The symptoms of ANK range from physical to optical to neurological. Most

patients with ANK experience asthenopia and headaches. Some have reported

photophobia, amblyopia, excessive tearing, and difficulty reading, along with a host of

other physiological manifestations. In about 10% of cases, patients report mobility

difficulties due to diplopia. Patients will also experience spatial distortions accompanied

by impaired stereoscopic depth perception. Keratometry gives insight into the origin of

anisometropia and the likelihood of ANK with the patient’s habitual correction. If a

patient’s corneal powers are sufficiently different anisometropia is present, it is likely that

the difference in refractive power is the primary cause of anisometropia. However, if

corneal powers are the same, axial length is more likely to be responsible for the

anisometropia (Bannon, 1954). Axial length measurement can confirm this. According to

Knapp’s Law, axial anisometropes should be corrected with spectacle lenses ideally

located at the anterior focus of the eye to produce equal corrected retinal image heights.

Measuring Aniseikonia

The Space Eikonometer (American Optical Corp., Southbane, MA) has for many

years been considered the gold standard in ANK testing. However, it does have its

shortcomings, in particular because it presents an exceptionally difficult visual challenge

for the patient. For this reason, the Space Eikonometer has fallen out of favor and has not

been in production for many decades. Significant training is required to become

accustomed to it, and even then it is heavily influenced by patient judgment (Antona et. al

2006, McCormack et. et al. 1992). Measurements are based on optics research in

5

stereopsis and single binocular vision (McCormack et al. 1992). The New ANK Test

(Handaya, Tokyo, Japan) is commercially available and better accepted by practitioners

than the Space Eikonometer but tends to underestimate the degree of ANK in the patient

(McCormack et. al 1992, Yoshida et. al. 1997). While other methods of ANK testing are

becoming available and more are being developed, the most appropriate ANK test

available appears to be the deWit Basic Aniseikonia Test (Optical Diagnostics,

Culemborg, The Netherlands). While it is easier to operate and offers more time-efficient

testing than the Space Eikonometer, it does tend to underestimate ANK in the vertical

meridian and is inconsistent in the horizontal meridian due to heterophoria. It is also

more accurate when used in a dim room to aid in eliminating peripheral fusion cues

(Fullard et al. 2007). This test uses a computer monitor set at a specific distance (77cm

for the configuration used in the study) to produce the appropriate visual angle of specific

targets. The targets are two vertically aligned semicircles, one red and one green, with a

strong fixation point and a yellow background as illustrated in Figure 1. The subject

wears special red/green anaglyph glasses so that the eyes are dissociated. The subject

then adjusts the size of the right (red) semicircle to match the size of the other. This

procedure is performed twice with the right semicircle starting 25% smaller than the left,

and then the right semicircle starting out 25% larger than the left. The average of these

two tests yields an ANK measurement. The selection of the deWit BAT is the best choice

given the options, but it is far from perfect. A series of evaluation tests were performed as

part of the current project.

6

Figure 1. Example of BAT Display.

Vernier Alignment

Part of the BAT task is a simultaneous vernier alignment between the upper and

lower edges of the semicircles to make sure that the subject is matching semicircle sizes

and not the top or bottom of one relative to the other. Vernier alignment is also a more

accurate gauge of acuity than a typical Snellen test because it is not affected by

differences in legibility of visual acuity (VA) chart letters. Vernier alignment

performance is related to cataract severity (Essock et al. 1984).The test consists of two

horizontal lines, one fixed and one variable, separated by a gap as shown in Figure 2. A

strong fixation point is included to prevent fusion. The participant must indicate whether

the variable horizontal line is aligned above or below the fixed horizontal line. Polarizing

7

the screen and using matching orthogonal polarizing glasses appeared in preliminary

testing to enhance vernier alignment reliability over red-green anaglyph glasses.

Figure 2. Vernier Alignment Task Display.

Using cataract patients may compromise the validity of the BAT due to the light

scattering effects of cataract. Based on discussion with binocular vision research experts

at The University of Alabama School of Optometry (UABSO), the question arose as to

whether the light scattering caused by cataract could affect the ability to resolve the edges

on the semicircles on the BAT. If so, this may influence the perceived retinal image

height of the light scatter-affected image, possibly making it appear larger. Due to

increased spread in the image, vernier alignment may also be impeded by the light

scattering effect. The patient may not be able to make an accurate judgment of the

potentially thicker appearing upper and lower edges of the semicircles due to the light

scattering caused by cataract.

8

Cataract and Aniseikonia

Cataract surgery presents a unique opportunity to study clinically induced ANK.

Even though these are not typical anisometropes, having one phakic and one

pseudophakic lens in a previously ≥2D bilateral ametrope produces a difference in

refractive power (induced refractive anisometropia) of acute onset. This does not leave

the visual system much time to adapt to the new conditions, or provide long term data

about the ANK. Emmetropizing cataract surgery is the most cost effective and commonly

performed surgery in the world. Cataract has been shown to improve visual function as

well as quality of life (Mangione et al. 1994). Nuclear and cortical cataracts are the most

common types of cataracts seen in clinic (Tan et al. 2006). With cataract surgery

becoming more commonplace, clinically induced ANK in anisometropes from unilateral

surgery or in ametropes in the period between the first and second surgery is inevitable.

The degree of ANK, as well as the plasticity of the patients’ visual system, varies

depending on the visual stimuli or visual task presented to the patient (Troutman, 1962).

The degree of ANK produced by unilateral emmetropizing cataract surgery needs

to be accurately determined due to healthcare reform. Bilateral cataract surgery may not

be insured if a sufficient quality of life is determined by healthcare regulators to be

accomplished by performing unilateral cataract surgery. This would leave the patient in

an anisometropic condition that may induce ANK. Based on literature reports, it appears

that unilateral pseudophakes have a higher incidence of ANK symptoms than bilateral

pseudophakes (Kramer et al 1999). Bilateral ametropic cataract patients could be utilized

in order to clinically induce ANK and measure it using the BAT.

9

Other Factors Affecting Aniseikonia

For patients wearing spectacles, performing the BAT is affected by the spectacle

magnification of their lenses. Spectacle magnification (SM) is calculated by using the

shape factor and power factor of the spectacle lens as shown in Figure 3. The shape factor

depends on the front surface power and the thickness of the lens while the power factor

depends on the back vertex power and the distance of the vertex to the cornea (Ogle

1972). The spectacle magnification of the cataract patients’ spectacles will be taken into

account.

Figure 3. Spectacle Magnification Formula. t is the thickness of the lens, n is the

refractive index of the material used, F1 is the base curve of the lens, d is the vertex

distance, and F’v is the lens back vertex power expressed as equivalent sphere.

Placement of the IOL inside the eye will affect the total refractive power.

According to the literature, IOLs in the anterior chamber produce more retinal image

magnification than IOLs placed in the posterior chamber. If IOL power does not produce

emmetropia and a residual spectacle correction is required, each diopter of spectacle

overcorrection at a vertex distance of 12 mm causes 2% retinal image magnification (plus

lenses) or minification (minus lenses) (Atebera et al. 2009). IOLs intrinsically produce a

1

1 11

1 V

spec lens

Spectacle Magnificationd Ft F

n

= ×′−

− ×

10

change in retinal image magnification relative to the crystalline lens that can be a

confounder in ANK measurement.

Cataract Simulating Devices

Bangerter Foils

Bangerter foils are plastic devices with embedded microbubbles of varying

densities. They have been available since the 1960’s and are typically used to treat

amblyopia by reducing retinal image quality in the unaffected eye to a predicted level.

Different Bangerter foils should produce different degrees of light scatter. However,

studies have shown that the rated VA loss caused by the foils is inaccurate (Perez et al.

2010).

Vistech Scattering Goggles

Scattering goggles produced by Vistech Inc. now Stereo Optical are used to

illustrate to the family of a cataract patient how the affected person sees through the

cataracts. They have also been used in prior studies to simulate the light scattering effects

of cataract. These goggles can be combined on top of each other to increase the image

degradation. The goggles have shown to simulate the angular light distribution of cataract

(Elliot et al. 1996).

Tiffen ProMist Black Filters

Tiffen ProMist Black (TPM) filters 1, 2, and 3 are camera filters that are used in

photography to produce a hazy appearance. These filters can also be combined by

11

screwing them together to increase their density. According to reports, these filters

provide a representation of early and mid level cataracts as they cause glare effects

analogous to these conditions (DeWit, et al. 2006).

Optical Defocusing Lenses

Optical defocusing lenses are simple plus sphere lenses out of a trial frame set.

They are not a source of light scatter per se, but are included to allow comparison of the

effects of light scatter and defocus on the ANK measurement. PERG (Pattern

Electroretinogram) studies comparing the impact of both optical defocus of +1.75

through +5.00D and light scattering transparencies indicated that the two scattering

methods cause different effects on the retina (Bach and Matheau, 2004).

Visual Acuity

Visual acuity loss is associated with cataract, although it does not always give the

full picture of how the patient’s vision is affected (Chylack et al. 1993). Snellen VA

typically measures visual acuity loss. However, a plethora of ocular diseases affect the

transparency of the ocular media and retina. Intraocular light scatter can decrease VA. A

simple VA measurement is therefore inadequate due to incomplete assessment of visual

function. A patient can have reasonable VA and severely reduced contrast sensitivity

(CS) that will not show up on a standard Snellen acuity test.

12

Contrast Sensitivity

Increasing severity in all types of cataract has been associated with decreasing VA

and CS (Maraini et al. 1994). It has been found that contrast sensitivity measurement in

both high and low spatial frequencies does provide more information about vision loss

due to diabetic cataract than visual acuity measurement alone (Chylack et al. 1993).

Contrast sensitivity is sometimes clinically used in quantifying vision loss resulting from

cataract (Guyton and Rubin, 1990) as well as a host of other ocular diseases. In early

cataract, CS is not affected as much in the lower spatial frequencies. In previous studies,

the Pelli-Robson test has been used (Klein et al. 2003) due to its ease of administration

and repeatability (Rubin et al. 1997). The Lighthouse Letter Contrast Sensitivity Test,

later named the MARS Letter Contrast Sensitivity Test, was chosen due to its portability

over the Pelli-Robson test as well as the increased accuracy. Contrast sensitivity is

associated with decreased ambulatory mobility (Marron and Baily, 1982), driving

performance (Wood and Troutbeck, 1995), facial recognition, and daily tasks (West et. al

2002). All of these tasks operate in the lower special frequency range (Dougherty et al.

2005). The test operates in the low spatial frequencies at 0.5m, around 1.25 cycles/degree

or 20/480 visual acuity. The BAT also operates in the lower spatial frequency range. The

test uses the same Sloan letter set as the Pelli-Robson test, but decreases contrast

sensitivity by each letter as opposed to each trio of letters in the Pelli-Robson test. Also,

the Mars test includes three tests of the same contrast sensitivity but with different letters.

The Mars test measures 23x35.5 cm and is printed on rigid plastic. It has a white

background and black letters similar to a Snellen chart. It has 48 letters that are 1.75 cm

high in 8 rows of 6 letters each. The advertised contrast varies from 91% (-0.04 log units)

13

to 1.2% (-1.92 log units). Each letter subtends 2 degrees at the testing distance of 0.5m

(Dougherty et al. 2005).

14

CHAPTER 2

AIMS AND RATIONALE

Null Hypothesis

Light scattering does not affect ANK measurement using the BAT in normal

subjects using an appropriate scattering device to simulate different degrees of cataracts.

ANK is not induced by emmetropizing cataract surgery in asymptomatic patients with 2

D or greater presurgical ametropia following initial cataract extraction. Subsequent

removal of the contralateral cataract will also produce no change in symptoms of ANK.

Alternative Hypothesis

Light scattering does affect ANK measurement using the BAT in normal subjects

using an appropriate scattering device to simulate different degrees of cataracts.

Aniseikonia is induced by emmetropizing cataract surgery in asymptomatic patients with

2 D or greater presurgical ametropia following initial cataract extraction. Subsequent

removal of the contralateral cataract will return ANK measurements to baseline.

Aims

Specific Aim 1

Validate the Basic Aniseikonia Test using a group of normal participants and

afocal size lenses to induce size difference between each eye. Determine the appropriate

15

device to accurately simulate cataracts by comparing visual acuity and contrast sensitivity

loss of different scattering filters. Then, the potential influence of cataract-induced light

scatter on perceived retinal image height and therefore ANK will be measured. This will

be accomplished by performing the BAT on a group of normal participants wearing an

appropriate scattering device simulating different degrees of cataracts.

Specific Aim 2

Measure ANK in >2D ametropic cataract patients prior to cataract extraction and

after each cataract is removed and replaced with an emmetropizing intraocular lens,

based on surgeries spaced at least 3 weeks apart.

The changes in light scattering due to lens removal may influence ANK

measurements. The data from Aim 1 will address the influence of light scatter. The

amount of induced ANK from the difference in overall refractive power between

ametropic and emmetropic eyes can then be determined. Factors influencing the changes

included removal of the spectacle lens SM and the potential magnification change

induced by the IOL. This will be used to predict the degree of ANK induced by

emmetropizing surgery and how anisometropia relates to ANK. This may be useful to

predict the effect of emmetropizing cataract surgery on previously anisometropic patients

and could eventually be applied to correct ANK in moderate to high isometropes who

will only obtain unilateral cataract surgery. Determination of the effect of unilateral

cataract extraction on quality of life is important for another reason. If healthcare

providers restrict coverage of second cataract extractions based only on presurgical VA,

this may adversely affect the patient’s quality of life.

16

CHAPTER 3

EXPERIMENTAL DESIGN

Validating the Basic Aniseikonia Test

A group of six participants was used to validate the BAT. This was accomplished

by performing the BAT with just the red/green filters as well as placing afocal size lenses

in front of each eye. Afocal size lenses enlarge an objects size by a determined

percentage without altering refractive error. The afocal size lenses that were used are

+1%, +2%, and +3.5%. The BAT, which was presented on a Dell 2405fpw monitor, was

set at 77cm from the participants, and the left and right arrows of the keyboard adjusted

the size of the right semicircle. The participants’ ANK mean with no size lens present

was used to normalize their size lens data. All subsequent simulated cataract ANK tests

were performed with the Dell 2405 fpw monitor. With an image height of 11cm and a

viewing distance of 77cm, the image subtends 8.1 degrees of the retina. The ANK

measurements were analyzed using linear regression.

Determining Appropriate Scattering Device

Scattering devices were tested for their ability to simulate cataract. They were

also evaluated with the BAT to ensure that ANK testing was possible. Four different

scattering devices were tested. These devices are listed in Table 1. Vistech filters became

available only after testing on the first two devices was completed. Tiffen ProMist filters

17

were available at a later date again. This meant that three separate test sessions were

required to complete this part of the study.

Six participants were used to evaluate each cataract simulator. The BAT was set

at a 77cm test distance, and the left and right keyboard arrows were used to adjust the

size of the right semicircle. To determine if each participant attained sufficient

repeatability on the BAT, the Repeatability Coefficient Test (RCT) was applied. The

mean and standard deviation(SD) of the 14 BAT measurements were taken, and the

standard deviation multiplied by 2.77. This was the subject’s repeatability coefficient.

The difference between consecutive measurements was then evaluated. To reach

appropriate repeatability required a difference between consecutive measurements less

than the repeatability coefficient (Weiss et al. 2010).

Each scattering device was tested using the following procedure. The scattering

device was placed in front of one eye and the BAT measurement repeated 14 times. The

device was then placed in front of the contralateral eye and 14 further measurements

taken. This process was repeated for each grade of scattering device (e.g. 0.3, 0.4 and 0.6

LogMAR Bangerter foils). VA was measured using a Snellen chart and contrast

sensitivity using a MARS contrast sensitivity test. The outcomes of these tests were

used to determine which scattering device most closely simulated visual performance

losses caused by cataracts. All simulated cataract ANK measurements were analyzed

using one way ANOVA and Tukey’s test (normal distribution of Kruskal-Wallis if not

normal). Post-hoc tests were applied only if there were significant differences among

group mean values.

18

Details of Scattering Devices

Table 1.

Details of Scattering Devices

Scattering Device Source Grades Bangerter Foil Ryser Ophthalmologie, St.

Gallen, Switzerland 0.3, 0.4, 0.6 LogMAR*

Optical Defocusing Lenses

Oculus, Wetzlar, Germany +0.50 D, +1.00 D

Vistech Scattering Filters

Stereo Optical, Chicago, IL 1, 2, 3, and 4 filters

Tiffen ProMist Black Filters

Tiffen, New York, NY 1, 2, 3, 5 (filters 2 and 3 combined), and 6 (filters 1, 2 and 3 combined)

*LogMAR = Logarithm of the minimum angle of resolution

Criteria for Selecting Study Patients

Simulated Cataract Participants

Participants were recruited using flyers placed in the UABSO, Worrell Building,

and Shelby Biomedical Sciences Building. Normal binocular vision was required

because heterophoria was shown in preliminary testing to cause inconsistent ANK

measurements. An age range of 20 to 30 years was required to lessen the chance of

ocular media opacities. Slit Lamp examination verification of clear media ensured that

the only sources of light scattering were the simulating devices.

19

Cataract Patients

Patients were selected from Vision America in Birmingham as well as the

UABSO Eye Clinic. Patients were eligible for the study if they were between 20 and 75

years of age. In addition, these patients had 20/80 or better VA in the worse eye. To be

included, patients were required to have at least 2 D ametropia in each eye and normal

binocular vision. Study forms are located in Appendix B.

Data Collection

Simulated Cataract Participants

Qualified patients underwent:

• Slit lamp exam

• Visual Acuity using a Snellen chart (OS, OD, OU)

• Cover test at distance and near

• Contrast Sensitivity using MARS test (OS, OD, OU)

The MARS test was performed in the recommended fashion by setting the test 0.5

m from the participant, and then allowing patients to read left to right. The test was

concluded when the participant responded with two incorrect answers in a row. The test

was performed monocularly and binocularly with no filter present and with Vistech 1 and

2 in front of each eye.

Demonstration of satisfactory repeatability on the BAT was a requirement to

participate in the simulated cataract study. The BAT was set at 77cm from the

participants, and the left and right arrows of the keyboard adjusted the size of the right

semicircle. This was determined by performing the BAT 21 times with no scattering

20

device. The RCT was used on this data to determine repeatability. To determine if the

ANK measurement is altered by the scattering produced by simulated cataract, the

following procedure was used. A Vistech filter was placed in front of one eye,

alternating between OD first and OS first for sequential subjects and BAT testing was

conducted. This process was repeated with each Vistech filter in front of each eye.

Some subjects also performed the vernier alignment test. It has been postulated

that is a similar task as aligning the top and bottom edges of the two semicircles in the

BAT. Vernier alignment was tested by having participants indicate if one variable

horizontal line was above or below a fixed horizontal line on a computer screen using the

up and down arrows on a Bluetooth keyboard. A Bluetooth keyboard was necessary due

to the testing distance. The test was forced choice as the participants could not indicate if

the lines were aligned. A fixation point was included to aid in preventing fusion. The

participants wore TechSpec Linear polarizing laminated film orthogonally with matching

orthogonal polarizing film attached to the screen so only one horizontal line could be

seen with each eye. Initially, the horizontal lines were 75x2 pixels, but unavoidable

fusion into stereo occurred. The lines had to be increased to 150x2 pixels to perform the

task. All scattering devices were tested in front of the left eye, and the test was performed

on a Sony Trinitron Multiscan G400 monitor. A cathode ray tube monitor was necessary

to produce polarizable light. The testing distance was 5m, and the distance from the

bottom edge of the maximum upward displacement of the variable horizontal line and the

top edge of the maximum downward displacement of the variable horizontal line was 3

mm. The angle subtended by the upper and lower edge of the most positive and most

negative position respectively is 0.03 degrees. This test was repeated 55 times per eye

21

with each degree of scattering device. Only the vertical meridian was tested in both the

ANK and Vernier alignment studies. The resulting data expressed as a percentage correct

for each possible offset position was analyzed by comparing the “No Filter” condition to

each filter type’s offset using one way ANOVA and Tukey’s test (normal distribution of

Kruskal-Wallis if not normal): post-hoc applied only if significant difference indicated

among groups based on between group mean values.

Cataract Patients

Vision America cataract patients were approached during their presurgical

consultation and the purpose of the study was explained to them. UABSO clinic patients

were usually contacted the Friday before their presurgical consult. If the patient agreed to

the study, the following tests were performed by Vision America Staff:

• Refraction

• Visual Acuity using a Snellen chart

• Alternate prism cover test at distance and near

• Axial Length

• Corneal topography with stimulated keratometry

• Measurement of spectacle parameters

o Spectacle prescription

The principal investigator then conducted the following series of tests:

• Measurement of spectacle parameters

o Base curve

o Type of material for refractive index

22

o Normal vertex distance using a Pupillary Distance ruler

o Central lens thickness

• Stereopsis using a Randot stereopsis test

• BAT

• Questionnaire

The measurement of spectacle parameters is important in determining the amount

of spectacle magnification the patient experiences. Increasing the base curve, refractive

index, and central thickness increases the magnification. If vertex distance is increased

with a plus lens (or decreased vertex distance with a minus lens), spectacle magnification

is increased. If refractive index is increased, spectacle magnification is decreased. The

serial numbers of the implanted lenses as well as the power were recorded.

The cataract patient questionnaire consisted of 10 visual symptoms that the

patient might experience if ANK was present. Then patient indicated the severity of each

symptom as a number ranging from one to five with one being never and five being

always. The symptoms are those most commonly reported in a prior study of 500

aniseikonic patients (Bannon and Triller, 1944). The same questionnaire was given at

each visit and is located in Appendix B.

The BAT viewing distance was set at 77cm on an identical monitor to the one

used for the simulated cataract study (Dell 2405fpw), and the left and right arrows of the

keyboard adjusted the size of the right semicircle. Additional lenses were not used to

correct for the working distance due to concerns about the change in spectacle

magnification and reduced field of view. The subjects were trained on the BAT before

data collection began. This consisted of the patient simply becoming familiar with the

23

format and what is expected of them. The patients performed the test with their habitual

correction. The BAT test was repeated 10 times. Before the test, each patient was

reminded to examine the top and bottom edges of the semicircles to ensure that they are

properly aligned and stationary. The initial measurements provided a baseline with which

to correlate data. Most patients undergo bilateral cataract surgery. Therefore, two

postoperative tests were performed in which the subject had to return to Vision America

or the UABSO. These visits typically took place at the 30 day post op or the evaluation

for the second cataract extraction. For the first postoperative visit, patients wore their

habitual visual correction in front of the unoperated eye and a Halberg clip with their

current over-correction if their equivalent sphere was ≥ 0.50D when performing the BAT.

Patients were not corrected in the operated eye if their equivalent sphere was <0.50D.

Patients wore a Halberg clip with their current correction if their equivalent sphere was ≥

0.50D as well as a +2.50D add when testing stereopsis. ANK measurements from the

cataract patients were analyzed using one way ANOVA on parametric data, and Kruskal-

Wallis One Way ANOVA with Tukey Test on nonparametric data. During these visits,

Vision America staff performed the following tests:

• Refraction

• Visual Acuity

• Alternate prism cover test at distance and near

Then, the principal investigator conducted the following tests:

• Stereopsis

• BAT

• Questionnaire

24

These visits occurred within one month plus or minus one week of each surgery. Data

collected at each visit is listed in Appendix C.

Figure 4 illustrates the formula used to predict the contribution of spectacle

magnification to ANK. Using the spectacle magnification calculated based on the

spectacle parameters, an estimate of the contribution of SM to ANK was determined. The

ANK value recorded prior to the first surgery was divided by the SM of the operated eye

for an estimated ANK measurement for post surgery 1. Multiplying the estimated ANK

measurement for post surgery 1 by the SM of the contralateral lens produced an

estimation of the ANK post surgery 2.

11 Presurgery ANK MeasurementPredicted ANK PostsurgerySM of unoperated eye

=

2 * SM of spectacle lens from second operated eyePredicted ANK Postsurgery Presurgery ANK MeasurementSM of spectacle lens from first operated eye

=

Figure 4. Formula used to Predict the Contribution of SM to ANK.

25

CHAPTER 4

RESULTS

Validating the Basic Aniseikonia Test

A baseline measurement with no size lenses was obtained from each subject. The

baseline measurements were used to normalize each participant’s data. This method was

used for all simulated cataract BAT tests. Results expected based only on afocal size lens

magnification are listed in Table 2.

Actual results are graphed in Figure 5. Rank Sum Tests on the +1% OS and +1%

OD size lens ANK measurement results showed a significant difference (P<0.001). A

similar significant difference (p<0.001) was found with the +3.5% OD and +3.5% OS

size lens ANK measurements. Because the data was normal for the +2% OD and +2%

OS size lens ANK measurement results, a student’s t test was used. The +2% OD and

+2% OS size lens ANK results differed significantly (p <0.001, Power = 1.00). These

results indicate a significant difference in ANK measurement when the afocal size lenses

are moved from one eye to the other. This is consistent with the fact that unilateral size

lenses change retinal image height in one eye. Absolute differences in retinal image

height were smaller on average than the value expected based on size lens magnification.

The underestimation was similar for size lens in front of OD and size lens in front of OS.

This was consistent with literature reports (Fullard et al, 2007) that the BAT tends to

underestimate ANK.

26

Table 2.

Expected ANK Results using BAT with Afocal Size Lenses.

Eye with lens Retinal Image Change ANK Measurement

OS ↑ Positive OD ↑ Negative

The correlation between afocal size lens magnification and change from baseline

ANK was measured by linear regression analysis. For the three size lenses placed in

front of the left eye, a slope of 0.924 for ANK versus size lens magnification was

obtained (R = 0.765, R2 = 0.585, p<0.001). For the size lenses placed in front of the right

eye a slope of −0.795 for ANK versus size lens magnification was obtained (R = 0.711,

R2 = 0.505, p<0.001).

27

Figure 5. Normalized Preliminary BAT Results using Afocal Size Lenses in the Absence

of Defocusing or Scattering Devices. N=6. Black points represent afocal size lenses in

front of the left eye. Red points represent afocal size lenses in front of the left eye. Error

bars are SD.

28

Vernier Alignment

During the test, it was difficult for the participant to retain dissociated images and

required continued attention to the fixation point. Red-green glasses were therefore tried

as a potential replacement for orthogonal plane polarizers and the colors of the horizontal

lines were changed to match in an attempt to make the task easier on the participant.

However, the screen would not produce the necessary wavelengths of light and the colors

bled through allowing both lines to be seen by one eye. For these reasons, the orthogonal

polarizers were used for all subsequent vernier alignment tests. The line width was

increased from 75 to 150 pixels, and vernier alignment tests were possible using the

+0.50 D and +1.00 D defocusing lenses (Figure 6), Tiffen Pro Mist Black filters

1,2,3,5(TPM 2 and 3 combined),and 6 (TPM 1,2, and 3 combined) ( Figure 7), and

Vistech 1 and 2 (Figure 8 and 8a). Vistech 3 and 4 filters as well as the Bangerter foils

still caused unavoidable fusion even with the 150 pixel line width making them unusable,

and while the test proved challenging for participants, there was no significant difference

between the participants’ data in the absence of any devices and all devices tested.

29

Figure 6. Vernier alignment Task Results with “No Filter” Versus Defocusing

Lenses. N=6. Results are expressed as a percent correct for each position the variable

horizontal line was presented. The defocus lenses did not significantly alter the

percentage of correct indications at each offset position. There were no correct answers

for 0 offset because it is a forced choice test. Error Bars are Standard Error of the Mean

(SEM).

30

Figure 6 illustrates the vernier alignment task results comparing the “No filter”

condition to the optical defocusing lenses. Performing ANOVA and Tukey’s test (normal

distribution, or Kruskal-Wallis test if not normal) on the percent correct at offset position

with the “No Filter” condition and defocusing lenses did not reveal any significant

differences. However, at position −1 (one position lower than 0), a p value of 0.065 was

found, indicating that it was close to being significant. A similar result was found at

position 2 (p=0.092). The powers (P) for ANOVA tests on vernier alignment results for

all positions with normal data were very low (P ≤0.385, 95% confidence interval).

Figure 7 shows the vernier alignment task results with no filter versus Vistech

filters 1 and 2. There is no marked difference in vernier alignment accuracy between the

“No Filter” condition and each Vistech filter. ANOVA and Tukey’s test were used

(normal distribution of Kruskal-Wallis if not normal). Post-hoc tests were applied only

when a significant difference among groups was indicated by ANOVA (based on

between group mean values). For the data that was normal, statistical power was very low

(P ≤0.364, 95% confidence interval), and as for defocusing lenses, the positions that

showed differences closest to being significant were −1 (p=0.078) and +2 (p=0.071). The

reason for this is unclear.

31

Figure 7. Vernier Alignment Task Results with “No Filter” Versus Vistech

Filters. N=6. Vistech filters 1 and 2 were tested. Results are expressed as a percent

correct for each position the variable horizontal line was presented. There were no

significant differences between the percentage correct at each vernier line position with

no filter and with each filter in place. There were no correct answers for 0 offset because

it is a forced choice test. Error Bars are SEM.

32

Figure 8. Vernier Alignment Task Results with “No Filter” Versus Tiffen Pro Mist

Black Filters 1,2,3,5 (TPM 2 and 3 combined), and 6 (TPM 1,2, and 3 combined). N=6.

Results are expressed as a percent correct for each position the variable horizontal line

was presented. There were no correct answers for 0 offset because it is a forced choice

test. Error bars indicate SEM.

33

Figure 8a. Vernier alignment task with Tiffen Pro Mist Black Filters in front of OS

with no Error Bars.

34

Figure 8 shows the vernier alignment task results with the Tiffen Pro Mist Black

1,2,3,5, and 6 filters relative to “No Filter.” The data is reillustrated in Figure 8a with no

error bars. The Tiffen Pro Mist Black filters also do not show any difference in the

percentage of correct choices of vernier offset line direction (above or below) as the filter

density is increased. ANOVA and Tukey’s test were used to determine significance

(normal distribution, or Kruskal-Wallis test if not normal). Post-hoc comparisons of

means applied only when ANOVA showed significant differences among groups (based

on between group mean values). For the data that was normal, statistical power was also

low (P≤0.050, 95% confidence interval). The p values (p≥0.522) indicated the differences

for all positions were not significant.

Determining the Appropriate Scattering Device for Subsequent Testing

Several different scattering devices were evaluated to determine the most suitable

means of simulating different densities of cataract. Ideally, the device should produce

comparable VA loss and CS loss to that of cataract. Visual acuity loss was assessed with

a Snellen chart, and CS was determined using the MARS contrast sensitivity test.

Bangerter Foils

If scattering was influencing the ANK measurement, a possible trend would be an

increase in retinal image height of the eye with the Bangerter foil. This is based on the

concept that the scatter of light produced by the foil makes the edges appear wider and

therefore makes the overall image appear larger. Therefore, if the foil was placed in front

of the right eye, the ANK measurement would be decreased. If the foil was placed in

35

front of the left eye, the ANK measurement would be increased, and this trend would

increase as the density of the foil was increased.

The BAT results using 0.3, 0.4, and 0.6 Bangerter foils (Figure 9) produced

multiple significant differences in the ANK measurement (based on ANOVA and post-

hoc comparisons among means) for different foil densities in front of the right eye versus

the left eye. Data was normalized using the average of the measurements taken using just

the red/green glasses. Regression analysis of the Bangerter foil data interestingly revealed

no significant trend in ANK measurement versus foil density with filters in front of the

left eye (R=0.0741, R2=0.0055, p=0.241, m=0.560). However, a weak, but significant

trend was found when the filters were placed in front of the right eye (R=0.167,

R2=0.0280, p=0.008, m=0.918).

As further evidence that Bangerter may not be the optimal choice of cataract

simulation device; the foils did not produce a graded VA or CS loss. Contrast sensitivity

and VA losses were moderate, but did not increase with Bangerter rating as seen in

Tables 3 and 4. A similar result was found by Odell et al. (2008) in a report on the effects

of Bangerter foils on Visual Acuity. For the above reasons, Bangerter Foils were

considered to be a poor choice for subsequent cataract simulation experiments.

36

Figure 9. Normalized Preliminary BAT Results using Bangerter Foils. N=6. No

significant trend going from low visual acuity ratings to high visual acuity ratings. X axis

is reversed due to the Bangerter foil ratings being LogMAR. Error bars are SD.

37

Optical Defocusing Lenses

Figure 10 illustrates the BAT results using the +0.50D and +1.00D defocusing lenses.

There was a similar trend in the ANK measurements on the BAT to that found earlier

with the afocal size lenses as shown in Figure 5. BAT data was normalized using the

average of the measurements taken with no lens in place. Plus lenses in front of the left

eye produced a larger retinal image in the left eye indicated by the positive measurement

caused by the participants adjusting the right semicircle to be larger. Plus lenses in front

of the right eye produced a larger retinal image in the right eye, and the participants

adjusted the right semicircle to be smaller, producing a negative measurement. This is

consistent with the preliminary results using size lenses. However, dioptric lenses

produce a purely refractive change, and the optical defocusing lenses produce more than

1% change per 1D of refractive error. Therefore, overcorrection may affect ANK more

than what would be assumed. VA and CS were only mildly affected as seen in Tables 3

and 4. These lenses do not prove to be a good simulation of cataract because their

defocusing effect is not analogous to the typical light scattering effect caused by cataract.

Using a PERG, it has been shown that light scattering reduces amplitude more than

defocus. The two methods of image degradation have different effects on the retina (Bach

and Matheau, 2004).

38

Figure 10. Normalized Preliminary BAT Results using Optical Defocusing

Lenses. Black points represent defocusing lenses in front of the left eye. Red points

represent defocusing lenses in front of the right eye. N=6. Error bars are SD.

39

Regression analysis of the defocusing lens BAT ANK measurements revealed a

highly significant trend in both the left and right eye (p<0.001). This is consistent with

the earlier findings using the afocal size lenses (Figure 5). However, the defocusing

lenses produced slopes larger than 1 (−2.617 OD and 2.771 OS) indicating that plus

sphere lenses produce magnification higher than the accepted “rule of thumb” of 1 D

produces 1% magnification (Linksz et al., 1965).

Vistech Scattering Filters

Vistech scattering filters decreased both contrast sensitivity and visual acuity as a

function of the number of stacked filters. CS was affected more than VA as seen in

Tables 3 and 4. The ANK results were normalized using the average of the measurements

taken using just the red/green glasses.

Figure 11 illustrates the normalized ANK measurements from the BAT with

Vistech filters in front of each eye. Performing the BAT using 1 as well as 2, 3, and 4

Vistech scattering filters stacked on top of each other showed a small trend in ANK

measurement using linear regression for 1 and 2 filters stacked in front of the right eye

(p=0.018, m=0.355, R=0.182, R2=0.0331), but not in the left (p=0.240, m=−0.151,

R=0.0.0911, R2=0.0829). However, the subjects were not able to see the semicircle with

the eye in front of which 3 or 4 filters stacked. The test was repeated by reversing the red

green filters and changing the semicircle that is adjusted in an attempt to allow testing of

Vistech 3 and 4. However, the same unsuccessful result was obtained with 3 and 4 filters

stacked.

40

Figure 11. Normalized Preliminary BAT Results using Vistech Scattering Filters.

N=6. The Vistech filters showed a small trend warranting further exploration. Black

points represent filters in front of the left eye. Red points represent filters in front of the

right eye. Error bars are SD.

41

Tiffen Pro Mist Black Filters

An expected trend of ANK measurement using the TPM filters on the BAT is for the

ANK measurement to increase if the filter was placed in front of the left eye and increase

more with increasing filter density. The filters were mounted to the red green glasses

using Halberg clips and a special threaded adapter that allowed the filters to be fixed in

place. Additional filters could be screwed on to the filter attached to the adapter. CS and

VA were only mildly affected by the filters as seen in Tables 3 and 4. CS was affected to

a greater extent than VA. BAT data was normalized using the average of the

measurements taken using just the red/green glasses. The BAT results using the Tiffen

Pro Mist Black filters are illustrated in Figure 12.

Analyzing the BAT results using the TPM filters with a linear regression revealed

that using TPM 1, 2, and 3, 2 and 3 together (TPM5), as well as TPM1, 2, and 3 (TPM6)

did not produce a trend in ANK measurement on the BAT in front of the left eye

(p=0.0311, m=−0.0217, R=0.0498, R2=0.00248). However, as with the Bangerter foils

and the Vistech filters, the TPM filters in front of the right eye produced a produced a

significant trend in the right eye (p<0.001, m=0.0332, R=0.0967, R2=0.00727).

42

Figure 12. Normalized Preliminary BAT Results using Tiffen ProMist Black Filters.

N=6. The TPM filters did not show any significant trend between OD and OS. Error bars

are SD.

43

Table 3. Preliminary MARS Contrast Sensitivity Results of all Filters

Filter Type OD JW

OS JW

Binocular JW

OD LW

OS LW

Binocular LW

OD MS

OS MS

Binocular MS

No filter 1.56 1.44 1.72 1.68 1.68 1.68 1.64 1.60 1.72 1Vistech 1.20 1.20 1.64 1.16 1.12 1.68 1.20 1.20 1.68 2Vistech 0.76 0.76 1.64 0.76 0.76 1.68 0.72 0.72 1.68 3Vistech 0.32 0.32 1.60 0.36 0.36 1.68 0.28 0.24 1.68 4Vistech 0 0 1.44 0 0 1.68 0 0 1.68 0.6 Bangerter 1.20 1.24 1.64 1.08 1.20 1.68 1.16 1.12 1.68 0.4 Bangerter 1.24 1.24 1.64 1.36 1.20 1.68 1.20 1.20 1.68 0.3 Bangerter 1.28 1.28 1.64 1.32 1.28 1.68 1.20 1.20 1.68 +0.5D 1.64 1.64 1.64 1.68 1.68 1.68 1.64 1.56 1.68 +1.00D 1.60 1.56 1.68 1.68 1.64 1.68 1.56 1.44 1.68 TPM1 1.56 1.60 1.68 1.52 1.52 1.68 1.44 1.44 1.68 TPM2 1.56 1.60 1.68 1.52 1.52 1.68 1.44 1.44 1.68 TPM3 1.44 1.44 1.68 1.44 1.52 1.68 1.44 1.44 1.68 TPM5 1.40 1.36 1.68 1.40 1.28 1.68 1.20 1.20 1.68 TPM6 1.20 1.20 1.68 1.20 1.12 1.68 0.96 1.20 1.68

Filter Type OD NG

OS NG

Binocular NG

OD JB OS JB Binocular JB

OD KD

OS KD

Binocular KD

No Filter 1.72 1.72 1.72 1.72 1.72 1.76 1.60 1.64 1.68 1Vistech 1.12 1.12 1.68 1.20 1.20 1.68 1.16 1.16 1.56 2Vistech 0.72 0.72 1.60 0.72 0.72 1.64 0.72 0.76 1.60 3Vistech 0.24 0.44 1.60 0.40 0.20 1.64 0.36 0.24 1.56 4Vistech 0 0 1.60 0 0 1.68 0 0 1.68 0.6 Bangerter 1.16 1.16 1.64 1.16 1.08 1.64 1.12 1.08 1.64 0.4 Bangerter 1.16 1.20 1.64 1.20 1.16 1.68 1.24 1.20 1.56 0.3 Bangerter 1.16 1.24 1.64 1.24 1.24 1.64 1.24 1.2 1.60 +0.5D 1.44 1.56 1.72 1.64 1.56 1.72 1.60 1.64 1.72 +1.00D 1.56 1.6 1.64 1.60 1.56 1.75 1.68 1.60 1.68 TPM1 1.44 1.44 1.68 1.64 1.56 1.72 1.56 1.44 1.72 TPM2 1.44 1.44 1.68 1.48 1.52 1.72 1.60 1.48 1.68 TPM3 1.36 1.44 1.64 1.44 1.44 1.68 1.48 1.44 1.64 TPM5 1.32 1.40 1.68 1.36 1.36 1.72 1.32 1.32 1.68 TPM6 1.16 1.16 1.64 1.28 1.24 1.72 1.20 1.20 1.68

*Binocular indicates that test was performed with both eyes with one filter in front of the right eye.

44

Table 4a.

Preliminary Visual Acuity Results of all Filters

Filter Type OD JW OS JW Binocular JW OD LW OS LW

Binocular LW OD MS OS MS

Binocular MS

No Filter 20/20+1 20/15-1 20/15 20/20-1 20/25+2 20/20-1 20/20+2 20/20 20/20+3 1Vistech 20/20+1 20/20+2 20/15 20/25+3 20/25+2 20/20 20/30+3 20/25-2 20/20+3 2Vistech 20/20 20/20 20/15 20/25-2 20/40+2 20/20-2 20/30+3 20/25-2 20/20+3 3Vistech 20/30-1 20/30+2 20/20+2 20/40+2 20/60 20/20-2 20/50 20/50 20/20+3 4Vistech 20/60+1 20/60 20/20 20/70 20/70-1 20/20-2 20/70+1 20/70 20/20+3 0.6 Bangerter 20/60-1 20/60 20/20+1 20/60+2 20/50 20/20-2 20/70-1 20/70 20/20+3 0.4 Bangerter 20/50-1 20/50 20/20 20/50+2 20/50+2 20/20-1 20/50 20/60 20/15-2 0.3 Bangerter 20/60 20/50 20/20+1 20/40+1 20/60+2 20/20 20/50-2 20/50+1 20/20+3 +0.5D 20/20-1 20/20+3 20/15 20/20+2 20/15-2 20/20-1 20/20 20/20 20/20+3 +1.00D 20/30 20/25 20/15-2 20/25+3 20/25+3 20/25+5 20/20 20/20 20/20 TPM1 20/15 20/15 20/15 20/20-2 20/20-2 20/20-1 20/20 20/20 20/20+2 TPM2 20/15 20/15-1 20/15 20/20-1 20/20-1 20/20-1 20/20 20/20 20/20+3 TPM3 20/15 20/15-3 20/15 20/20 20/30-2 20/20 20/20 20/20 20/20+3 TPM23 20/20-1 20/20-1 20/15 20/20+1 20/20-1 20/20+3 20/25-1 20/25-1 20/20-1 TPM123 20/20-1 20/20+1 20/15 20/20-1 20/30 20/20+1 20/25-1 20/25-2 20/20

*Binocular indicates that test was performed with both eyes and with one filter in front of the right eye.

45

Table 4b.

Preliminary Visual Acuity Results of all Filters

Filter Type OD NG OS NG Binocular NG OD JB OS JB Binocular JB OD KD OS KD

Binocular KD

No Filter 20/20 20/20 20/20 20/20 20/25-1 20/20+2 20/20-1 20/20-2 20/20 1Vistech 20/20-2 20/25 20/20 20/20-1 20/20-1 20/15-2 20/20-1 20/25-1 20/25-1 2Vistech 20/30+3 20/25-2 20/20-2 20/25-2 20/25-1 20/15-2 20/50+2 20/25-2 20/25-2 3Vistech 20/40-2 20/50-1 2020-1 20/50 20/50-1 20/15-2 20/50-1 20/60 20/40-2 4Vistech 20/70-2 20/70-1 20/25+2 20/100+1 20/100 20/15-2 20/80 20/80 20/30-2 0.6 Bangerter 20/80+1 20/70-1 20/20+2 20/60-1 20/50-2 20/20+2 20/80 20/70 20/30-2 0.4 Bangerter 20/60 2/60+1 20/20+2 20/30 20/30 20/15-1 20/60-1 20/70-1 20/40 0.3 Bangerter 20/60-2 20/60+2 20/15+2 20/40 20/40 20/20+1 20/70-1 20/70 20/30-1 +0.5D 20/20-1 20/20-2 20/20+1 20/15-3 20/15-2 20/20+2 20/30 20/25 20/25-1 +1.00D 20/30-2 20/40-2 20/20 20/20-2 20/15-1 20/20+1 20/5-+1 20/25-2 20/25-2 TPM1 20/20-2 20/20-1 20/20+1 20/20-1 20/25+2 20/20+2 20/25 20/25-2 20/25+1 TPM2 20/20-2 20/20-2 20/20+1 20/20-2 20/20 20/20+3 20/25-1 20/25-1 20/20-1 TPM3 20/25+1 20/25+1 20/20 20/20-2 20/20-1 20/20+1 20/25-2 20/25-2 20/25-2 TPM5 20/25+2 20/25+3 20/20+3 20/20-2 20/25 20/20+2 20/25 20/40+2 20/20-1 TPM6 20/25-2 20/25+2 20/20+2 20/25-2 20/25 20/20 20/25 20/40+1 20/20-1

*Binocular indicates that test was performed with both eyes with one filter in front of the right eye

46

Final Selection of Cataract Simulation Filter

Table 5 shows the results of the tests performed to find the effect of each filter on

ANK measurement (using Tukey’s test) and to look for trends in ANK measurement as a

function of filter density to further investigate which filter would most likely produce an

effect on ANK measurement.

Selection of the filter to be used for the comprehensive cataract simulation study

was based on CS and VA measurements. Research performed by others (Chylack et al.

1993) and (Maraini et al. 1994) shows that while VA is not greatly affected by cataract,

CS is substantially reduced. Optical defocusing lenses produce different PERG amplitude

to that resulting from a light scattering device indicating that it has a different effect

(Back and Mathieu, 2004). The optical defocus lenses were therefore eliminated.

Bangerter foils were not chosen because of the uniform effect on VA and CS. They did

not produce a graded loss of VA and CS as the density of the filters increased. The Tiffen

Pro Mist Black filters and the Vistech filters were similar in that they mildly affected VA

and had a moderate affect on CS. This is consistent with the findings of Chylack et al.

(1993) and Maraini et al. (1994). Ultimately, the Vistech filters were chosen because they

had a greater effect on CS than the TPM filters.

47

Table 5.

Comparative Effect of each Filter Type on ANK Measurement

Number of Significant

Differences Subject Reliability TPM Bangerter Vistech JB 0.413 18 4 4 JW 0.759 0 4 1 KD 0.82 4 7 7 LW 0.54 20 5 4 MS 1.89 14 8 1 NG 0.735 10 14 5 Total SDs* 66 42 22

Total Comps 330 126 60

% SDs 20.00% 33.33% 36.67%

*Number of significant differences in ANK between filter combinations (Tukey’s test)

Vistech filters showed the greatest percentage of significant differences in ANK

measurement when all possible filter combinations were compared.

Simulation of Cataract using Vistech Filters

The participant CS and VA data is shown in tables 6 and 7. The results of the VA

measurements indicate that Vistech 1 simulates a mild cataract. The participants’ VA

changed by less than one line. However, CS is affected moderately, decreasing about 0.4

log units. Vistech 2 simulates a more severe cataract, degrading VA by a line, and

contrast sensitivity by about 0.8 log units compared to the “no filter” condition. Binocular

VA and CS were not significantly affected by either filter.

48

Student t tests comparing the ANK measurements obtained using the BAT with 1

Vistech filter in front of the left versus right eye reveal that while the Vistech 1 is not

significantly affecting ANK measurements, Vistech 2 significantly affected BAT

performance. Using a Mann-Whitney sum test for both conditions, comparing 1 filter OD

and OS resulted in a p value of 0.307. Using the same test comparing 2 stacked filters OD

or OS resulted in a p value of <0.001 Participants complained of image alignment

problems and difficulty keeping both semicircles in the same horizontal and/or vertical

position while performing the BAT with Vistech 2 in front of either eye. It is

hypothesized that this is a result of a greater influence of the patient’s heterophoria on

ANK measurement as the image becomes more degraded. The results indicate that the

eye with the filter placed in front of it has a reduced retinal image height. This is contrary

to the previously postulated theory that the blurred edges would make the images seen

through the filter appear larger. Figure 13 illustrates the BAT results of the simulated

cataract study using the Vistech filters to simulate cataract.

Linear regression analysis reveals a significant difference in the measurement of

ANK using the BAT with the filters in front of each eye. For Vistech filters 1 and 2 in

front of the left eye, a slope of 0.467 was obtained (R=0.244, R2=0.0597, p<0.001). For

Vistech filters 1 and 2 in front of the right eye, a slope of -0.396 was obtained (R=0.198,

R2=0.0394, p<0.001). The statistical powers for both tests were very high at P=1.000.

ANOVA tests comparing the “no filter” to Vistech OS and Vistech OD ANK

measurements revealed significant difference (p<0.001). Tukey’s test showed significant

difference (p<0.05) in all comparisons of ANK measurements except 1 Vistech filter in

front of the OS versus “no filter.”

49

Figure13. Normalized BAT ANK Results using Vistech Filters 1 and 2. N=20. Increasing

Vistech density showed a similar trend as the preliminary Vistech data. Black points

represent filters in front of the OS. Red filters represent filters in front of the OD. Error

bars are SD.

50

Table 6.

VA Results using Snellen Chart for Participants Enrolled in Simulated Cataract Study.

PATIENT VA OD VA OS VA OU VA OD V1 VA OS V1 VA OU V1 VA OD V2 VA OS V2 VA OU V2 MS-1 20/20 20/20 20/20-1 20/30+3 20/25-2 20/20+3 20/30+3 20/25-2 20/20+3 JW-2 20/20+1 20/15-1 20/15 20/20+1 20/20+2 20/15 20/20 20/20 20/15 LW-3 20/20-1 20/20-2 20/20-1 20/25+3 20/25+2 20/20 20/25-2 20/40+2 20/20-2 KD-4 20/20-1 20/20-2 20/20 20/20-1 20/25-1 20/25-1 20/50+2 20/25-2 20/25-2 NG-6 20/20 20/20 20/20 20/20-2 20/25 20/20 20/30+3 20/25-2 20/20-2 MC-7 20/15-2 20/15-3 20/15 20/25+3 20/20 20/20+3 20/25+1 20/25 20/20+3 KT-8 20/20-1 20/15-1 20/15 20/20-2 20/20-2 20/20-1 20/25-2 20/25-2 20/20-2 LT-9 20/20-1 20/20-1 20/15-2 20/20-1 20/25+1 20/20-1 20/25-1 20/40+3 20/20-1 AM-10 20/15-1 20/20 20/15-1 20/20-1 20/25+3 20/15-2 20/25+2 20/30-2 20/20 AW-11 20/15-1 20/15 20/15-1 20/20+1 20/20 20/15-2 20/25-1 20/25-1 20/15 TP-12 20/20+2 20/20-2 20/15-3 20/20 20/20 20/20 20/25 20/25-2 20/25-1 LK-13 20/15-2 20/15-2 20/15-2 20/20+3 20/20+2 20/15-1 20/25+2 20/25+2 20/15-1 PC-14 20/20+1 20/20 20/20+2 20/25+3 20/20-1 20/20 20/25-1 20/25-2 20/20+2 WR-15 20/20-3 20/20-2 20/20-3 20/30+3 20/25-3 20/20-2 20/30-2 20/30-2 20/15-2 NR-16 20/20 20/15-2 20/15-2 20/20+2 20-20 20/20 20/25-1 20/25-2 20/20-2 SP-17 20/20 20/20 20/20+1 20/25-2 20/25-1 20/20+1 20/25-2 20/30+3 20/20-2 SO-18 20/20 20/20-2 20/20 20/25-2 20/25-1 20/20-1 20/40-2 20/25 20/20+2 HM-19 20/15-2 20/15-2 20/15+2 20/20 20/20 20/15-2 20/25-2 20/25-1 20/15-1 CC-20 20/15-2 20/20+2 20/15-2 20/25+2 20/20 20/15-1 20/25-1 20/25 20/20+2 LL-21 20/20-2 20/20-1 20/15-1 20/20 20/20-1 20/15-2 20/25 20/25 20/15-1

* In OU, the filter was placed over the right eye with the left eye unoccluded.

51

Table 7.

CS Results using MARS Contrast Sensitivity Test for Participants Enrolled in Simulated Cataract Study.

PATIENT CS OD CS OS CS OU CS OD V1 CS OS V1 CS OU V1 CS OD V2 CS OS V2 CS OU V2 MS-1 1.64 1.60 1.68 1.20 1.20 1.68 0.72 0.72 1.68 JW-2 1.55 1.44 1.72 1.20 1.20 1.64 0.76 0.76 1.64 LW-3 1.68 1.68 1.68 1.16 1.12 1.68 0.76 0.76 1.68 KD-4 1.60 1.64 1.68 1.16 1.16 1.56 0.72 0.76 1.60 NG-6 1.72 1.72 1.72 1.12 1.12 1.68 0.72 0.72 1.60 MC-7 1.68 1.68 1.72 1.20 1.24 1.68 0.76 0.72 1.64 KT-8 1.64 1.64 1.68 1.20 1.20 1.56 0.72 0.72 1.56 LT-9 1.64 1.64 1.68 1.20 1.24 1.64 0.72 0.72 1.64 AM-10 1.64 1.64 1.68 1.20 1.16 1.60 0.76 0.84 1.64 AW-11 1.44 1.68 1.68 1.20 1.16 1.68 0.72 0.72 1.68 TP-12 1.68 1.68 1.72 1.36 1.32 1.68 0.96 0.96 1.68 LK-13 1.68 1.68 1.68 1.40 1.32 1.68 0.96 0.96 1.68 PC-14 1.60 1.64 1.72 1.24 1.28 1.72 0.76 0.68 1.68 WR-15 1.52 1.52 1.60 1.16 1.08 1.60 0.68 0.56 1.56 NR-16 1.60 1.60 1.68 1.12 1.12 1.60 0.72 0.72 1.60 SP-17 1.68 1.44 1.68 1.20 1.20 1.68 0.60 0.64 1.64 SO-18 1.64 1.64 1.68 1.28 1.24 1.64 0.80 0.72 1.64 HM-19 1.64 1.68 1.72 1.36 1.24 1.72 0.92 0.92 1.68 CC-20 1.68 1.68 1.68 1.20 1.20 1.68 0.76 0.76 1.68 LL-21 1.68 1.72 1.76 1.36 1.36 1.68 0.84 0.80 1.60

* In OU, the filter was placed over the right eye with the left eye unoccluded.

52

Cataract Patient Study: Specific Aim 2

Results from cataract patients are illustrated in Figure 14. The results indicated an

aniseikonic condition at the intermediate visit. Patients recorded close to 0% ANK on the

BAT at the preliminary visit. A shift in either the positive or negative direction develops

after the first surgery. Stereopsis was greatly reduced post surgery 1, with some patients

not able to perform the test. The spectacles each patient had at the presurgery visit were

used with the lens in front of the pseudophakic eye removed except where noted.

Stereopsis was restored, and in some cases improved, after post surgery 2. All patients

reported being very pleased with the quality of vision after bilateral cataract extraction.

Symptoms reported by the cataract patient questionnaire showed a reduction in severity

after unilateral cataract extraction and a mild reduction in severity after bilateral cataract

extraction as shown in Figure 15. The expected ANK measurements using the BAT on

cataract patients are listed in Table 8. The direction of the shift depends on if the patient

is hyperopic or myopic as well as which lens is extracted first. If the patient was

hyperopic and the left lens was extracted first, the ANK measurements were more

negative due to the decreased retinal image size in the left eye. If the patient was

hyperopic and the right lens was removed first, the ANK measurements were more

positive due to the smaller retinal image size in the right eye. If the patient was myopic

and had the left lens removed first, the ANK measurements were more positive due to the

increased retinal image size in the left eye. If the patient was myopic and had the right

lens removed first, the ANK measurements were more negative due to the increased

retinal image size in the right eye. ANK measurements return to ~0% at the final visit.

All data was analyzed using one way ANOVA.

53

Table 8.

Expected ANK Results using BAT on Cataract Patients.

Lens Extracted/Refractive Error

Retinal Image Height Change in eye after cataract extraction

ANK Measurement

OS Hyperope OS ↓ Negative OD Hyperope OD ↓ Positive OS Myope OS ↑ Positive OD Myope OD ↑ Negative

54

Figure 14. BAT Measurements from Cataract Patients. N=7.Each data point is from one

collection before extraction, after one extraction, and after 2 extractions.

55

Table 9 shows the results of the BAT measurements for the seven patients from

Vision America and UABSO. The type of ametropia along with the predicted ANK mean

change due to change in SM is shown. Also, whether or not the patients’ ANK mean

moved in the right direction is shown. Figure 14 is a graphical representation of the BAT

means for each study patient. Patients GAK, JKR, and BHS had changes in ANK

measurement consistent with predicted ANK as calculated using SM.

Patient NJC had a significant shift in the amount of ANK post surgery 1 in the

opposite direction than was predicted. However, the ANK results were variable when

comparing the mean ANK values of each visit (Pre surgery=0.150, Post Surgery 1=1.192,

Post Surgery 2=0.270) to the variance (Pre-surgery=1.134, Post Surgery 1=1.121, Post

Surgery 2=0.440). Extraction of the OD cataract caused a statistically significant shift in

the negative direction back to near baseline.

Patient KLN patient had 1.75 D anisometropia presurgery with equivalent sphere

refractive errors −4.25 DS OD and −2.50 DS OS, axial lengths 25.17mm OD and

24.73mm OS, 4 D exophoria, and an ANK measurement of -0.87%. Post surgery 1, the

patient did not have his spectacles, so a trial frame with his spectacle prescription was

used. This patient had 5.00D of anisometropia with equivalent sphere refractive errors of

-4.25 DS OD and +0.75 DS OS, 4 D exophoria, and -18.98% of ANK. This magnitude of

ANK is greater than expected given the presurgical ANK measurement and change of

SM due to the removal of the left spectacle lens. Post surgery 2, this patient had 1.50D

anisometropia with +1.00 DS OD and -0.50 DS OS, 2 D exophoria, and -4.84% ANK. As

with post surgery 1, the magnitude of ANK recorded is greater than predicted, indicating

that some other factors must have been influencing ANK measurement. This patient’s

56

extremely negative measurements may be due a vertical heterophoria, although vertical

heterophoria was not measured. Vertical displacement of the IOL could induce

heterophoria.

Patient BHD did not have her spectacles for the second visit, and reported that she

could not see clearly through the trial frame with her spectacle prescription in place. The

trial frame was adjusted several times and the patient confirmed that she did have an

unobstructed view of the BAT display. After extraction of the OS cataract, there was a

statistically significant shift of the ANK measurement in the positive direction. This is

contrary to expectation because removal of the hyperopic OS cataract should cause a shift

in the negative direction. After the OD cataract extraction, there was a significant shift in

the negative direction.

Patient MRE had an insignificant shift in ANK measurement in the opposite

direction of what would be expected post surgery 1. After having the OS cataract

extracted, the patient had a significant shift in the positive direction. This was also an

unexpected shift.

The calculated contribution of SM to ANK measurement is listed in Table 9.

Spectacle magnification is calculated using the formula in Figure 3 for each eye.

Expected ANK measurement is calculated using the formula in Figure 4. The main

concern of spectacle magnification is during visit 2 (post surgery 1), when the patient has

not had time to adapt to the lack of spectacle magnification in the operated eye. It appears

that spectacle magnification can account for some ANK recorded by the BAT. For four

of seven patients, direction and a degree of magnitude of ANK is predicted by taking into

account the change in SM.

57

Table 10 lists the global and contour stereopsis results for each patient for each

visit. Both global and contour stereopsis were reduced from pre surgery to post surgery 1,

with some patients not even able to perform the test. Stereopsis was improved post

surgery 2, often to presurgery 1 levels or better.

58

Table 9.

ANK Means of Patients with Ametropia Type, Consistency to Prediction, and Predicted ANK due to SM

Pre Surgery Post Surgery 1 Post Surgery 2

Patent Code

Extraction performed first Ametropia type ANK Mean

Predicted ANK Mean

Observed ANK Mean

Direction Consistent with prediction

Predicted ANK Mean

Observed ANK Mean

Direction Consistent with prediction

GAK1 OS Hyperope 0.77% -2.10% -1.13% Yes 0.84% 0.65% Yes NJK3 OS Hyperope 0.15% -4.49% 1.92% No 0.03% 0.27% No JKR5 OS Hyperope 0.35% -5.71% -6.10% Yes 0.22% -4.00% Yes KLN8 * OD Myope -0.87% -5.39% -18.98% Yes 1.20% -4.84% Yes BHS11 OS Myope -0.23% 4.28% 4.77% Yes 1.25% 3.75% Yes BHD12 ** OS Hyperope 0.67% -5.95% 3.36% No -0.05% 0.20% No MRE12 OD Hyperope 0.14% 6.06% -2.45% No -0.59% 2.68% No

*Patient crushed spectacles after pre surgery visit and a trial frame had to be used.

**Patient did not have spectacles and could not see through a trial frame.

Four out of seven patients showed change in ANK consistent with those predicted using SM.

59

Table 10.

Stereopsis Results of Cataract Patients

Patient Code Pre Surgery Post Surgery 1 Post Surgery 2 Global Contour Global Contour Global Contour GAK1 250 100 250 100 250 50 NJC3 250 140 250 70 250 70 JKR5 250 100 500 400 250 140 KLN8 250 40 >500 140 250 70 BHS11 500 140 >500 >400 250 40 BHD12 >500 >400 >500 >400 250 30 MRE13 500 50 >500 100 250 30

The results of the cataract patient questionnaire given at each visit are illustrated

in Figure 15. The results showed that patients experienced a significant reduction in

headaches between pre surgery and post surgery 2. Light sensitivity was reduced

significantly between post surgery 1 and post surgery 2 (p=0.031). Reading difficulty was

significantly reduced between pre surgery and post surgery 2. There were slight, but not

significant, reductions in the following symptoms: vertigo and dizziness, closing one eye

to read, nervousness, vertigo and dizziness, fatigue, and distorted space perception.

60

Figure 15. Cataract Patient Questionnaire Results. N=7. Significant difference was seen in the reduction of the severity of headaches,

light sensitivity, and reading difficulty.

61

Table 11 lists the significance of each symptom asked in the cataract patient

questionnaire as the result of a one way ANOVA test. If the data was nonparametric, a

Kruskal-Wallis one way ANOVA test was used. While no significant differences were

found for any symptom through the ANOVA, headaches, light sensitivity, reading

difficulty, nervousness, vertigo and dizziness, and distorted space perception were the

most significant. These symptoms were further investigated by t test comparing pre

surgery questionnaire scores to post surgery 2 scores. The results of the t tests show that

there is a significant change in the severity of headaches, and reading difficulty between

pre surgery and post surgery 2. A significant reduction was also found in the reduction of

severity of light sensitivity between post surgery 1 and post surgery 2.

62

Table 11. Results of One Way ANOVA and t test on Cataract Patient Questionnaire Results

Means

Pre Surgery Mean Post Surgery 1 Mean Post Surgery 2 Mean ANOVA P Value T Test P Value Between Pre Surgery and Post Surgery 2

Headaches 2.4 1.9 1.1 0.157* 0.047**

Eye ache, pain, or pulling 1.6 1.3 1.4 0.892

Light sensitivity 3.3 3.3 1.7 0.057* 0.051

Reading difficulty 3.9 2.4 1.8 0.002* <0.001**

Closes one eye to read 1.6 2.0 1.1 0.429

Double vision 1.4 1.3 1.1 0.990

Nervousness 1.7 1.7 1.1 0.117* 0.165

Vertigo and dizziness 2.0 1.3 1.1 0.163* 0.099

General fatigue 2.3 1.9 1.6 0.240

Distorted space perception 1.6 1.4 1.0 0.138* 0.209

* Warranted further investigation through t test

** Showed statistical significance in t test

63

CHAPTER 5

DISCUSSION

Key Findings

1. The Basic Aniseikonia Test (BAT) provided a reasonable, but imperfect,

measurement for ANK when applied in the vertical meridian in the dark.

2. Vernier alignment was not significantly affected by any of the tested scattering

devices.

3. Of the devices tested, the Vistech filters were the most accurate representation of

cataract judging by induced changes in VA and CS.

4. ANK measurement using the BAT was affected by the light scattering effects of

the Vistech filters.

5. Previously ametropic cataract patients showed a change in ANK measurement

after the first cataract extraction, with a return to near baseline after the second

extraction. However, there was increased variability between patients when

compared to presurgical ANK measurements.

Validating the Basic Aniseikonia Test

Using size lenses to validate the accuracy of the BAT indicated that the test

underestimates ANK measurement. Results also indicated a significant change in ANK

measurement when the afocal size lenses were moved from one eye to another. Figure 5

64

shows this trend for the afocal size lenses. Overall, the BAT was confirmed to be

underestimating size lens-induced ANK in the vertical meridian in the dark. These

results are in agreement with previous studies (Fullard et al. 2007, Rutstein et al. 2007)

because they both showed slopes for ANK measurement versus size lens

magnification that were less than 1. The lowest amount of underestimation was

previously found in the vertical meridian in the dark (Fullard et al. 2007). However, the

BAT proved to be the best available method for the conditions of the study in agreement

with the conclusions of Fullard et al (2007), that there is no more suitable alternative to

the BAT test for measurement of ANK.

Vernier Alignment

No significant differences were found between each filter type and the “No Filter”

condition results of the vernier alignment test (ANOVA). Figures 6-8a illustrated that all

unilaterally placed filter types and densities maintained the same curve as the “No Filter”

condition indicating that scatter-degraded edges created by simulated cataract are not

affecting the vernier alignment component of the BAT. However, for several reasons, the

vernier alignment test provided an incomplete assessment of the alignment task involved

in the BAT: (1) the challenging nature of the test, (2) high variability of data, (3)

inconsistency between results for upward versus downward displacement of the movable

vernier target, (4) the fact that the subjects were unable to perform the test for some of the

filters, and (5) because the results were expressed as incremental rather than continuously

variable values, the ANOVA test may not have been as robust. The effects of light scatter

on ANK measurement is a result of light scatter on the overall image, not just the top and

65

bottom edges of the semicircles. In addition, while vernier alignment is useful in judging

the alignment of the semicircles, the overall image combined with the strong fixation

point are probably equally, if not more, important.

Simulated Cataract

The ideal cataract simulator is a filter that affects both VA and CS, but has a

greater effect on CS. These effects should also increase with increasing filter density. The

Bangerter foils did not produce a progressive degradation of VA and CS. Thus, they were

eliminated from the study. The optical defocusing lenses were eliminated because that

they did not produce VA and CS loss consistent with cataract. Defocus has also been

reported by others to affect the retinal image by a different mechanism than light

scattering as demonstrated by Bach and Matheau (2004) using PERG. Vistech filters and

Tiffen Pro Mist Black filters both produced cataract-like effects on VA and CS. Of the

two, Vistech filters more closely resembled the CS loss of real cataract. Significant

changes in the ANK measurement with both filter types indicated that the light scattering

effects of cataract may influence ANK measurements on actual cataract patients.

Vistech filters were ultimately selected due to their more cataract-like impact on

CS rather than their effect on visual acuity, which is consistent for cataract according to

literature reports (Hess and Woo, 1978, Shandiz et al. 2011, Chylack et. al 1993.)

Vistech filters have also been found to scatter light with a similar angular distribution to

that of cataract (Elliot et al. 1996).

Twenty subjects were used to investigate how the Vistech filters affect ANK

measurement with the BAT. A Student t test comparing the normalized ANK

66

measurements with each filter on each eye showed that a single filter did not produce a

statistically significant difference in ANK measurement on either eye, while two stacked

filters in front of either eye did show a significant change in ANK measurement.

ANOVA tests comparing ANK measurements with no filter and with one or two filters in

front of either eye showed a significant effect on ANK measurements. Tukey’s test

revealed significant changes in ANK measurement when two filters were placed in front

of either eye and no filter in front of the contralateral eye. Because two filters produced a

significant change in measured ANK, the presence of cataract itself may influence ANK

measurements. This may be of little importance for bilateral cataract because the similar

light scattering effects OD and OS may cancel. However, when measuring ANK after

one cataract has been removed, the situation would more closely resemble that of the

unilateral scattering filters. Because a single Vistech filter showed a less consistent effect

on ANK, a mild cataract may not affect ANK measurement.

As reported by deWarrd et al (1992) forward scattering of light in the eye

produces a “veiling illuminance” superimposed on the retinal image that reduces retinal

image contrast. They cite cataract as a condition in which forward scattering is increased.

It was therefore expected in the current study that this “veiling illuminance” would make

the retinal image of the BAT target appear larger. However, based on the outcomes of the

cataract simulation study, the exact opposite was found. Instead of appearing larger,

ANK measurement results showed that that the image seen through the filter was

perceived as smaller, indicating that simulated cataract caused by Vistech filters has a

minifying effect on retinal image height. The finding of significant minification may have

67

been due in part to the large number of data points, but this does not alter the fact that the

minification effect was significant.

Cataract Patients

There are three possible factors that could be influencing ANK measurement

using the BAT on cataract patients:

1. The light scattering effects of cataract

2. The magnification and placement of the IOL

3. The inherent underestimation of ANK measured by the BAT

Patients showed a trend in changing ANK measurement depending on which lens

was extracted as well as if they were hyperopic or myopic. This is listed in Table 9. The

amount of measured ANK change is very close in magnitude to predicted ANK taking

into account SM. However, not all of the measured ANK can be accounted for based on

SM alone. Other factors may be influencing the ANK measurement. Inconsistent

measurements taken by patients NJC and MRE at post surgery 1 could have been a result

of the test being more challenging for them than others. The extremely negative

measurements recorded by patient KLN could have been due in to exophoria, as

heterophoria has been shown to affect ANK measurement with other devices (Lancaster,

1942). Lancaster, the originator of the term “aniseikonia”, reported that when using a

direct comparison eikonometer in patients with heterophoria, the heterophoria was

measured and not the ANK. In the current study, patient KLN’s ANK measurements may

have been affected by using a trial frame instead of his spectacles. It was necessary to use

a trial frame as his spectacles were irreparably damaged between pre surgery and post

68

surgery 1. Trial frame lenses could produce a different SM than spectacles and thus affect

the ANK measurement. This would be due to shape factor differences. Lens curvatures

and central thickness differ between trial lenses and typical positive meniscus ophthalmic

spectacle lenses. For four of the seven patients, the data corresponds with results from

the literature (Gobin et al. 2008) as the appropriate ANK change took place given which

lens was extracted, and the results were in the direction predicted using SM, although the

magnitude was not the same. Other factors may be having an effect on ANK

measurement such as light scatter or IOL placement and magnification. In the other three

cases, the trend was not as expected. Clearly, with the small number of patients in the

cataract study and the difficulty experienced by three of the seven, a larger patient

number should be investigated to be able to draw conclusions about the impact of

emmetropizing cataract surgery on ANK.

Correction of Refractive Error for Working Distance of BAT

Correction of refractive error for the working distance (77cm) of the BAT was not

performed due to the change in SM the lenses produce. However, preliminary results on

the BAT using defocusing lenses above +1.00D increased the difficulty of the task when

performed by normal healthy subjects, and younger participants would have more

accommodation than the older patients enrolled in the Cataract Patient study. Therefore,

the under corrected subjects in the Cataract Patient study with no correction for the

working distance would have a more difficult time performing the BAT than younger

patients. However, adding the appropriate lens (+1.25D) would change SM, and this

would need to be addressed.

69

Light scattering may be having a minifying effect on retinal image size and

subsequent ANK measurement judging by the results of the simulated cataract study

using Vistech filters. IOL’s do cause a change in retinal image magnification as reported

by Atebara et al (2009), and this is not accounted for in the calculations from the current

study. It is not only IOL power, but anterior-posterior position and tilt that may affect

retinal image height and therefore the subsequent ANK measurement (Atebara et al,

2009).

The cataract patient questionnaire showed a reduction in the severity of headaches

and reading difficulty between presurgery and post surgery 1. The severity of light

sensitivity was reduced between post surgery 1 and post surgery 2. These symptoms were

reported as being prevalent in patients with ANK (Bannon, 1945). Reading difficulty was

reported as being a symptom of contrast sensitivity loss (Whittaker and Lovie-Kitchin,

1993), and loss of contrast sensitivity is reported in the literature as being a result of

cataract (Maraini et al. 1993, Chylack et al. 1994). This could be due to going from a

light scattered condition to an aniseikonic condition, to an emmetropic condition. There

was little change in eye ache, pain, or pulling, double vision, and distorted space

perception. There was an increase in tendency to close one eye to read from presurgery to

post surgery 1, and a reduction in this tendency post surgery 2. After the first surgery, it

makes sense that patients would be more likely to read with one eye closed. This could

either be the unoperated eye, in particular if the eye is myopic, or the pseudophakic eye.

Patients only experienced decreased nervousness post surgery 2, indicating that they were

equally nervous in both the cataract and aniseikonic conditions. Vertigo and dizziness

was higher presurgery, but equal post-surgery 1 and 2. There was no increase in distorted

70

space perception post surgery 1. General fatigue decreased slightly over the course of the

surgeries, possibly due to first having to cope with the cataracts, then having to adapt to

the aniseikonic condition. It has been reported that even in patients with early stage

cataract, improved quality of life can be obtained by performing bilateral cataract surgery

over unilateral cataract surgery (Elliot et al. 2000). Expanding this to patients with

induced ANK after unilateral cataract surgery, the quality of life could be increased if the

detrimental effects of both unilateral cataract surgery and ANK are removed.

Spectacle magnification is of great importance in measuring ANK due to the

influence of SM on retinal image height (Achiron et al, 1997). In all patients except three,

the predicted spectacle magnification change could account for the change in direction of

ANK measurement in post surgery 1 as well as most of the magnitude. The remainder

could be due to inconsistency in ANK measurement, light scattering caused by the

remaining cataract, IOL power, or IOL position. It appears important for the patient to

perform the BAT with their habitual lenses because measurement of ANK using trial

frames was generally less successful.

However, the results of the simulated cataract study using Vistech filters indicate

that using the BAT to test ametropic cataract patients after unilateral cataract extraction

may be compromised due to the light scattering effects of cataract. The underestimation

of ANK measurement and minifying effect of light scatter indicated by the results of the

simulated cataract study suggest that ANK measurement on cataract patients may be

inherently inaccurate.

The decrease in both global and contour stereopsis after unilateral cataract

extraction and the improvement after bilateral cataract extraction are consistent with

71

other studies that found that loss of stereopsis is associated with the presence of ANK

(Lovasik and Szymkiw, 1985).

Limitations and Possible Future Studies

Simulated Cataract

This project had several limitations. Finding a scattering device that accurately

represented the contrast sensitivity and visual acuity loss caused by cataract, and that

could be adjusted in several increments, was challenging. Because the scattering device is

not within the crystalline lens, it may be producing effects that differ from those of

cataract. A simple filter cannot simulate the light scattering effects of all types of

cataract: for example nuclear, cortical, anterior subcapsular and posterior subcapsular.

The BAT has been shown to underestimate ANK measurements in the vertical meridian,

even in the dark (Fullard et al, 2007). It has also been shown to produce an even greater

underestimation in the horizontal meridian, which is the reason this meridian was not

tested in the current study (Fullard et al, 2007). While it is very convenient and portable,

it is not an ideal device for measuring ANK. The effects of short duration light scatter

caused by the various scattering devices may not be the same as the long term light

scatter experienced by cataract.

Cataract Patient Study

The difficulty in recruiting suitable cataract patients decreased the likelihood of

establishing significant trends in this part of the cataract patient study. Placement or

design of the IOL may have a confounding effect on ANK measurements. Data from the

72

simulated cataract study indicates that light scattering may be causing a minifying artifact

when measuring ANK using the BAT. Patients were under corrected for the working

distance of the BAT. Future studies involving cataract patients will include adding a lens

to correct for this distance. However, the preliminary BAT data using optical defocusing

lenses showed that defocusing lenses have an effect on BAT measurement. This will need

to be taken into consideration. Vertical heterophoria was not recorded and should be with

related studies. As with the simulated cataract study, the BAT is inherently flawed as it

has been proven to underestimate ANK in the vertical meridian in the dark and is unable

to accurately measure in the horizontal meridian (Fullard et al. 2007). Because these

patients are induced refractive anisometropes, they may not be experiencing ANK in the

same way as native anisometropes.

Future Studies

More sophisticated cataract simulating devices may provide more complete

information about the apparent minification produced by light scattering. Polymer-

dispersed liquid-crystal scattering devices are being investigated by several groups as

potential cataract simulators (Ozolinsh and Papelba, 2004). With the ability to

continuously vary the amount of scattering, these devices show potential. Measuring the

full contrast sensitivity function using an instrument that tests across several spatial

frequencies, such as Vistech Contrast Sensitivity Charts, would be useful. This approach

could provide a better assessment of parallels between a cataract simulator and real

cataract. Further investigation of the influence of light scatter on vernier alignment would

also be an option for further study.

73

Expanding the cataract study patient numbers to allow more definite conclusions

to be drawn is important. Including a criterion for reliability similar to the Repeatability

Coefficient Test that was used in the simulated cataract study as well as measuring

vertical heterophoria could decrease the number of patients who perform unreliably on

the BAT. Increasing the number of ANK measurements taken by the cataract patients

should not be a problem because the patients did not complain of fatigue after performing

the task.

The Metrovision device (Gobin et al., 2008) is a computer based direct

comparison eikonometer that utilized liquid crystal spectacles that oscillate at 120 Hz, a

frequency undetectable by the human eye, and a synchronized computer screen. As one

spectacle lens becomes transparent, a semicircle is presented, and as the other spectacle

lens becomes transparent, another semicircles is presented. The participant adjusts one

semicircle stepwise until the two semicircles appear the same, similar to the BAT. This

method may be more appropriate as it does not rely on anaglyph glasses. If found to be

more accurate in ANK measurement than the BAT, it would be a more suitable test for

measuring ANK in cataract patients as well as determining the light scattering effects of

simulated cataract on ANK.

Conclusions

In conclusion, the BAT was shown to underestimate ANK measurement in the

vertical meridian in the dark. Vernier alignment was not affected by any of the scattering

devices tested; indicating that some other aspect of the BAT must be influenced by light

74

scattering. Vistech filters showed VA and CS loss similar to losses seen in cataract,

proving that they were the optimal cataract simulator. These filters minified retinal image

height as measured by the BAT. Therefore, ANK measurements in ametropic cataract

patients may be underestimated due to both the BAT as well as light scattering caused by

cataract. Despite these shortcomings, for 4 of 7 cataract patients, ANK measurements

were in the predicted direction showing that ANK is produced by emmetropizing cataract

surgery on ametropic cataract patients. The combination of reduced stereopsis, results of

the questionnaire, and ANK measurements shows the advantage of bilateral over

unilateral cataract extraction in patients where ANK may develop after unilateral

extraction.

75

LIST OF REFERENCES

Achiron L, Witkin N, Primo S, Broocker G. Contemporary Management of Aniseikonia. Survey of Ophthalmology 1997.41: 4, 331-330.

Antona B, Barra F, Barrio A, Gonzalez E, Sanchez I. The validity and repeatability of the

New Aniseikonia Test. Optometry and Vision Science 2006. 83, 903–9. Arditi, A, Improving the design of the letter contrast sensitivity test. Investigative

Ophthalmology 2005.46,6, 2225-2229 Atebara N, Asbell P, Azar D, Ellis F, Faye E, Hoffer K, Wiggins R. Clinical Optics

Section 3, Singapore,.2009; 223. Bach M, Matheau M. Different effect of dioptric defocus vs. light scattering on the

Pattern Electroretinalgram. Documenta Ophthalmolgica 2004. 108; 99-106. Bangerter A. Occlusion in pleoptics and orthoptics. Kiln Monatsbl Augenheilkd 1960.

136, 305–31. Bannon R. Clinical Manual on Aniseikonia 1954. Buffalo, NY: American Optical

Corporation. Bannon R, Triller W. Aniseikonia-A clinical report covering a ten year period. American

Journal of Optometry 1944. 21; 171-182. Benjamin W. Borish’s Clinical Refraction. 1998. 1142–1159. Burian H. Clinical significance of Aniseikonia. Archives of Ophthalmology 1944. 29,

116–33. Chylack L, Padhye N, Khu P, et al. Loss of contrast sensitivity in diabetic patients with

LOCS II classified cataracts. British Journal of Ophthalmology. 1993. 77, 7-11. De Warrd P, IJspeert J, van den Berg T, de Jong P. Intraocular Light Scattering in Age-

Related Cataracts. Investigative Ophthalmology and Visual Science 1992. 33; 3, 618-625.

76

de Wit, G, Franssen, L, Commens, J, van den Berg, T. Simulating the Straylight effects of Cataracts. Cataract Refractive Surgery 2006. 32, 294-300.

Dougherty B, Flom R, Bullimore M. An evaluation of the Mars Letter Contrast

Sensitivity Test. Optometry and Vision Science 2005. 82: 11, 970-5. Elliot D, Bullimore M, Patla A, Whitaker D. Effect of a cataract simulation on clinical

and real world vision. British Journal of Ophthalmology 1996. 80, 799-804. Elliot D, Patla A Furniss M, Adkin A. Improvements in clinical and functional vision and

quality of life after second cataract surgery. Optometry and Vision Science 2000. 77; 1, 13-24

Elliot D, Situ P. Visual acuity versus letter contrast sensitivity in early cataract. Vision

Research 1998. 38, 2047-52. Essock, E, Williams, R, Enoch, J, Raphael, S. Effects of Image Degradation by Cataract

on Vernier Acuity. Investigative Ophthalmology and Visual Science 1984. 24, 1043-1050.

Fullard R, Rutstein R, Corliss D. The Evaluation of Two New Computer-Based Tests for

Measurement of Aniseikonia; Optometry and Vision Science 2007. 84; 12, 1093-1100.

Gobin L, Rozema J, Tassignon M. Predicting refractive aniseikonia after cataract surgery

in anisometropia. Journal of Cataract Refractive Surgery 2008. 35; 1353-1361 Guyton D, Rubin G (Eds). Contrast Sensitivity and glare testing in the evaluation of

anterior segment disease. Ophthalmology 1990. 97, 1233-7. Hess R, Woo G. Vision through cataracts. Investigative Ophthalmology and Visual

Science 1978. 17: 428-435. Jimenez J, Ponce A, delBarco L, Diaz J, Perez-Ocon F. Impact of Induced Aniseikonia

on Stereopsis with Random-dot Stereogram; Optometry and Vision Science 2002.79, 121–5.

Klein B, Moss S, Klein R, Lee K, Cruichshanks, K. Associations of visual function with

physical outcomes and limitation 5 years later in an older population; The Beaver Dam eye study. Ophthalmology 2003. 110, 644-50.

Kramer P, Lubkin V, Pavlica M, Covin R. Symptomatic Aniseikonia in Unilateral and

Bilateral Pseudophakia. A Projection Space Eikonometer study. Binocular Vision and Strabismus Quarterly 1999. 14; 3, 183-190.

77

Lancaster W. A Reply to Criticisms of Aniseikonia. Transactions of the American Ophthalmological Society 1942. 40; 82-117

Lang J. An efficient treatment and new criteria for cure of strabismic amblyopia: reading

and Bangerter foils. Binocular Vision and Strabismus Quarterly 1999. 14: 9–10. Linksz A, Bannon R. Aniseikonia and refractive problems. International Ophthalmology

Clinics 1965.5,515-534. Lovasik J, Svymkiw M. Effects of Aniseikonia, Anisometropia, accommodation, retinal

illuminance, and pupil size on stereopsis. Investigative Ophthalmology and Visual Science 1985.26;5,741-750.

Maraini G, Rosmini F, Graziosi, P, Tomba, M, Bonacini M, Cotichini, R, Pasquini, P,

Sperduto, R. and the, Italian American Cataract Study Group. Influence of Type and Severity of Pure Forms of Age- Related Cataract on Visual Acuity and Contrast Sensitivity. Investigative Ophthalmology & Visual Science 1994. 35, 262-267.

Mangione C, Phillips R, Lawrence M, Seddon J, Orav E, Godman, L. Improved Visual

Function and Attenuation of Declines in Health-Related Quality After Cataract Extraction. Archives of Ophthalmology 1994.112:1419-25.

Marron J, Bailey I. Visual factors and orientation: Mobility Performance. American

Journal of Optometry and Physiological Optics 1982.59; 413-26. McCormack G, Peli E, Stone P. Differences in Tests of Aniseikonia. Investigative

Ophthalmology and Visual Science 1992. 33:2063–7. Miller D, Benedeck G. Intraocular Scattering: Theory and Clinical Application 1973.29-

32, 61-65. Odell N, Leske D, Hatt S, Adams W, Holmes J. The effect of Bangerter filters on

optotype acuity, Vernier acuity, and contrast sensitivity. Journal of AAPOS 2008. 12, 555–559.

Ogle K, Researches in Binocular Vision, New York, Hafner, 1972, 304 Ozolinsh M, Papelba G. Eye cataract simulation using polymer dispersed liquid crystal.

scattering obstacles. Ferroelectrics 2004.51,609-613. Perez G, Archer S, Artal P. Optical Characterization of Bangerter Foils. Investigative

Ophthalmology and Visual Science 2010. 51,609-13. Phillips C. Strabismus, Anisometropia, and Amblyopia. British Journal of

Ophthalmology 1959.43, 449-460

78

Rubin G. Reliability and sensitivity of clinical contrast sensitivity test. Clinical Vision

Science 988.2,169-77. Rubin G, West S, Munoz B Bandeen-Roche K, Zeger S, Schein O, Fried L. A

comprehensive assessment of visual impairment in a population of older Americans, The SEE study. Salisbury Eye Evaluation Project, Investigative Ophthalmology and Visual Science 1997. 38: 557-68.

Rutstein R, Corliss D, Fullard R; Comparison of Aniseikonia as Measured by the

Aniseikonia Inspector and the Space Eikonometer; Optometry and Vision Science 2006. 83, 11, 836-842

Rutstein R, Daum K. Anomalies of Binocular Vision: Diagnosis and Management. St.

Louis: Mosby; 1998. Shandiz J, Derakhshan A, Daneshyar A, Azimi A, Moghaddam H, Yekta A, Yazdi S,

Esmaily H. Effect of cataract type and severity of visual acuity and contrast sensitivity. Journal of Ophthalmic and Vision Research 2011. 6; 1, 26-31.

Tan A, Wang J, Rochtchina E, Mitchell. Comparison of age-specific cataract prevalence

in two population-based surveys 6 years apart. BMC Ophthalmology 2006. 6:17 Troutman R. Artiphakia and Aniseikonia. More evidence of aniseikonia in pseudophakia

from another expert, with a "seconding" of a warning to corneal, refractive and implant surgeons. Binocular Vision and Strabismus Quarterly 1999. 263

Weise K, Marsh-Tootle W, Corliss D. Evaluation of Computer- Based Testing for

Aniseikonia in Children. Optometry and Visual Science 2010.87; 11,883-88. West S, Rubin G, Broman A, Munoz B, Bandeen-Roche K, Turano K. How does visual

impairment affect performance on tasks of everyday life? The SEE Project. Salisbury Eye Evaluation. Archives Ophthalmology 200. 120:774-780.

Whittaker S, Lovie-Kitchin J. Visual Requirements for Reading. Optometry and Visual

Science 1993. 70, 54-65. Wood J, Troutbeck R. Elderly drivers and simulated visual impairment. Optometry and

Visual Science 1995. 72:115-124. Yoshida M, Sato M, Awaya S. Evaluation of the clinical usefulness of the New

Aniseikonia Tests. Nippon Ganka Gakkai Zasshi 1997. 101,718–22.

79

APPENDIX A

INSTITUTIONAL REVIEW BOARD APPROVAL LETTER

80

APPENDIX B

CATARACT PATIENT DATA COLLECTION FORMS

81

Presurgical Evaluation (Visit A) - 1 week pr ior to surgery Date: ______________ Patient Code:____/_____/____/___# Gender : ______ Race: _______ Age:_________

Patient’s habitual refractive correction (what type of correction they wear most of

the time): glasses or contact lenses ______________

Spectacle lens parameters:

Habitual Rx: OD_____________ OS_______________

Base Curve: OD_____ OS______

Center thickness: OD______ OS_______

n’: OD______ OS_____

Vertex distance: OD_______ OS_________

Visual Acuity with Habitual Rx: OD_______________ OS________________

Refractive Er ror : OD________(VA 20/ ) OS__________(VA 20/ )

Alternate pr ism cover test at distance and near : __________ (with Habitual Rx)

Axial Length: OD___________ OS__________

82

Corneal Topography with simulated keratometry: OD_________________

OS___________________________

Stereopsis: Contour_________ Global__________ (with habitual Rx)

Aniseikonia Measurements (with habitual Rx, green filter before OD)

1. Ver tical 1 ____________

2. Ver tical 2 ____________

3. Ver tical 3 ____________

4. Ver tical 4 ____________

5. Ver tical 5 ____________

6. Ver tical 6 ____________

7. Ver tical 7 ____________

8. Ver tical 8 ____________

9. Ver tical 9 ____________

10. Ver tical 10 ___________

Patient must demonstrate reasonable repeatability during preliminary Aniseikonia

testing. If investigator finds that patient demonstrates difficulty, he/she should not be

enrolled in the study.

# First initial/middle initial/last initial/patient number

83

Post Surgical Evaluation (Visit B and Visit C) - 1 month +/- 1 week following surgery Date:___________ Patient code:_____/____/____/__ Surgery (IOL) type/Eye:____________/____

Habitual Visual Acuity: OD______________ OS________________

Refractive Er ror : OD____________(VA 20/ ) OS_______________(VA 20/ )

Alternate cover test with pr ism at distance and near : _____/____ Habitual

spectacle lens before pseudophakic eye is removed. Use Halberg clip when residual

RE > 0.50 D spher ical equivalent (include 2.50 D add before pseudophakic eye

when doing near cover test).

Stereopsis: Contour____________ Global____________ (Use Halberg clip before

pseudophakic eye when residual RE > 0.50 D spher ical equivalent. Include 2.50 D

add before pseudophakic for testing stereopsis)

84

Aniseikonia Measurements: (Use Halberg clip before pseudophakic eye when

residual RE > 0.50 D, green filter before OD)

1. Ver tical 1 ______________

2. Ver tical 2 ______________

3. Ver tical 3 ______________

4. Ver tical 4 ______________

5. Ver tical 5 ______________

6. Ver tical 6 ______________

7. Ver tical 7 ______________

8. Ver tical 8 ______________

9. Ver tical 9 ______________

10. Ver tical 10 ____________

85

Questionnaire for Aniseikonia Study

Patient Code :___/___/___/___(first, middle, last name, and enrollment #)

Date:__________

The following lists 10 visual symptoms that you may have. Please grade them as

1(. never), 2. (a little), 3. (sometimes), 4. (a lot), 5. (always).

1. Headaches-

2. Eye ache, pain, or pulling-

3. Light sensitivity-

4. Reading difficulty-

5. Closes one eye to read-

6. Double vision-

7. Nervousness-

8.Vertigo and dizziness-

9. General fatigue-

10. Distorted space perception-

86

APPENDIX C

CATARACT PATIENT DATA

87

Initial Visit Post Surgery 1 (OS) Post Surgery 2 (OD) Patient Code: G/A/K/01 Gender: Male Race: White Age: 54 Habitual Correction: Glasses Habitual Rx: OD +1.25 OS +1.25 Base Curve: OD +5.25 OS +5.25 Center Thickness: OD 3.2 OS 3 Refractive Index: OD 1.49 OS 1.49 Vertex Distance: OD 15mm OS 15mm

Visual Acuity w/ Habitual: OD 20/80 OS 20/20 OD 20/80 OS 20/20 OD 20/40 OS 20/30

Refractive Error: OD +2.00 20/20 OS +2.00 20/20 OD +2.25 -.050 x175 20/20

OS +.75 -0.75 x 180 20/20

OD +1.00 - 1.50 x 170

OS +0.75 - 1.00 x 180

Alternate Prism Test: Distance Ortho Near Ortho Distance Ortho Near Ortho Distance Ortho Near Ortho Axial Length: OD 22.11mm OS 22.12mm

Corneal Topography:

OD 45.67D/7.29mm @ 83 43.77D/7.71mm @ 173

OS 43.77D/7.71mm@179 44.70D/7.55mm@89

OD 45.49D/7.42mm @87 * 44.00D/7.67mm @177

OS 43.72D/7.72mm @ 8 * 44.41D/7.00mm @ 98

Stereopsis: Global 250 Contour 100 Global 250 Contour 100 Global 250 Contour 50

* Corneal Topography and axial length not routinely taken after surgery

88

Initial Visit Post Surgery 1 (OS) Post Surgery 2 (OD) Patient Code: J/K/R/05 Gender: Female Race: White Age: 74 Habitual Correction: Glasses Habitual Rx: OD +3.00-0.75x75 OS+2.75-0.75x98 Base Curve: OD +8.20 OS+8.20 Center Thickness: OD 4.50 OS 4.75 Refractive Index: OD 1.49 OS 1.49 Vertex Distance: OD 14 mm OS 14 mm

Visual Acuity w/ Habitual: OD 20/20 OS 20/40 OD 20/30 OS 20/20 OD 20/20 OS 20/20

Refractive Error: OD +4.25 - 1.25 x 100 OS +3 - 1.25 x 100

OD +.75-0.25x134

OS +3.50-0.75x83

OD +0.25-0.50x90

OS +0.75-0.50x130

Alternate Prism Test: Distance Ortho Near Ortho Distance Ortho Near Ortho

Distance Ortho Near Ortho

Axial Length: OD 24.18mm OS 24.18mm

Corneal Topography:

OD 40.86/8.26mm@51 41.21/8.19@141

OS 41.26/8.18mm@147 41.82/8.07@57

Stereopsis: Global 250 Contour 100 Global 500 Contour 400 Global 250 Contour 150

89

Initial Visit Post Surgery 1 (OS) Post Surgery 2 (OD) Patient Code: K/L/N/08 Gender: M Race: White Age: 64 Habitual Correction: Glasses

Habitual Rx: OD -4.25-0.25x141

OS -2.25-0.75x150

Base Curve: OD +4 OS+4 Center Thickness: OD 1.8 OS 2 Refractive Index: OD 1.49 OS 1.49 Vertex Distance: OD 12mm OS 12mm Visual Acuity w/ Habitual: 20/40+2 20/25- OD 20/40+2 OS 20/20-1 OD 20/25 OS 20/20-2

Refractive Error: OD -4.25-0.25x141

OS -2.25-0.75x150 OD -4.25

OS +1.-.50x177

OD +1.25-.75x165 OS -.25-.50x155

Alternate Prism Test: Distance 2exo Near 4 exo Distance 4 exo Near 0 Distance 2xp Near 0

Axial Length: OD 25.17mm OS 24.73mm

Corneal Topography: OD 43.10 42.90

OS 44.18 44.35

Stereopsis: Global 250 Contour 40 Global >500 Contour 140 Global 250 Contour 70

90

Initial Visit Post Surgery 1 (OS) Post Surgery 2 (OD) Patient Code: B/H/S/11 Gender: Female Race: White Age: 71 Habitual Correction: Glasses

Habitual Rx: OD -3.25-0.50x117 OS-4.25

Base Curve: OD +6.5 OS +5.8 Center Thickness: OD 1.4 OS 1.8 Refractive Index: OD 1.49 OS 1.49 Vertex Distance: OD 12mm OS 12mm

Visual Acuity w/ Habitual: 20/50-2 20/50-2 OD 20/60 OS 20/30 OD20.25 OS 20/25

Refractive Error: OD -3.25-0.50x117 OS -4.25

OD -3.00-0.75x110

OS 0-0.75x75

OD -1.00 -1.25 x 121

OS 0 -0.75 x75

Alternate Prism Test: Distance Ortho Near Ortho Distance Ortho Near Ortho Distance Ortho Near Ortho Axial Length: OD 23.29mm OS 23.47mm

Corneal Topography: OD 44.50/45.75 @ 42

OS45.50/46.00 @ 77

Stereopsis: Global 500 Contour 140 Global >500 Contour >400 Global 250 Contour 40

91

Initial Visit Post Surgery 1 (OS) Post Surgery 2 (OD) Patient Code: B/H/D/12 Gender: Female Race: White Age: 71 Habitual Correction: Glasses Habitual Rx: OD +4.50 -1.00x85 OS +4.50-0.75x93 Base Curve: OD +7.5 OS+7.75 Center Thickness: OD 3.75 OS3.50 Refractive Index: OD 1.49 OS 1.49 Vertex Distance: OD 12mm OS 12mm Visual Acuity w/ Habitual: OD 20/40 OS 20/50 OD 20/40 OS 20/25-1 OD 20/20 +

OS 20/30 -2

Refractive Error: OD +4.50-1.00x85 OS +4.50-.75x95 OD +4.50-1.00x85 OS PL

OD +0.25-0.25x115 OS +1.00-0.75x90

Alternate Prism Test: Distance Ortho Near Ortho Distance Ortho Near Ortho Distance Ortho Near Ortho

Axial Length: OD 22.98mm OS 22.75mm

Corneal Topography: OD43.32/7.79@123 44.06/7.66@33

OS44.18/7.64@105 44.64/7.56@15

Stereopsis: Global >500 Contour >400 Global >500 Contour >400 Global 250

Contour 30

92

Initial Visit Post Surgery 1 (OD) Post Surgery 2 (OD) Patient Code: M/R/E/13 Gender: M Race: W Age: 75 Habitual Correction: Glasses Habitual Rx: OD OS Base Curve: OD +7.5 OS +7.75 Center Thickness: OD 5 OS 4 Refractive Index: OD 1.49 OS 1.49 Vertex Distance: OD 12mm OS 12mm

Visual Acuity w/ Habitual: OD 20/30 OS 20/30 OD 20/20 OS 20/50 OD 20/25 OS 20/25

Refractive Error: OD+3.00-2.00x090 OS+3.00-1.25x090

OD P+1.00x090

OS+3.00-2.00x90

OD PL-0.75x90

OS +0.50-1.25 x 80

Alternate Prism Test: Distance Ortho Near Ortho Distance Ortho Near Ortho

Distance Ortho Near Ortho

Axial Length: OD 24.23 OS24.28

Corneal Topography: OD41.46/8.14@106, 42.94/7.86 @16

OS 41.31/8.17 @82 42.83/7.88@ 172

Stereopsis: Global 500 Contour 50 Global >500 Contour 100 Global 250 Contour 30