medical radiology
TRANSCRIPT
Villa Marketing Division
Villa Sistemi Medicali SpA
Guide lines: Medical radiology
Villa Marketing Division
• Introduction To Radiology• Key Components
• Work Techniques
• Types Of Exams
• Types Of Devices
Contents
Villa Marketing Division
• The radiology is used as diagnostic instrument to visualize anatomical structures, that can not be seen from outside and it is a part of a bigger discipline, called “Diagnostic Imaging”
• The X-rays were discovered accidentally by Wilhelm Conrad Röntgen (German) in 1895
• In the course of the century, technologies, means and materials have become refined and have brought the radiology (or better the Diagnostic Imaging) be a particular branch of the medicine
• A significant evolution of the radiology has been the CT (Computerized Tomography) (Hounsfield et Al, 1972) which still uses collimated X-rays and allows the study of body sections
• Progressively other diagnostic modalities have been improved, as ultrasound, nuclear medicine, magnetic resonance which use different physical principles from X-rays and allow to obtain different information compared to ones given by radiographic images
Introduction To Radiology
Villa Marketing Division
• With the progress of knowledge and of technologies, radiologists have developed a strong specialization in their competences
• Thanks to X-rays, specialized products were designed, addressed directly to clinical specialist in different fields: Mammography, Interventistic, Angiography, Mineralometry, etc.
• Each clinical specialty is equipped with devices, which, limiting the application field, increase the diagnostic efficacy of the product
• Radiological competence can be found also in other specializations as, for example, surgeon, cardiologist, breast specialist, dentists, etc. They use radiological images, but each of them with very specific purpose and needs, that must be understood and supported.
Introduction To Radiology
Villa Marketing Division
• Radiology is based on the use of X-rays, electromagnetic radiations with wave-length of about 10-10 m
• X-rays can easily cross each kind of material: they practically go into the object and come out in the opposite side. When X-rays cross an object, they are attenuated by the object itself: the more dense is an object, the higher is its attenuation
- Air is not very dense: X-rays cross it easily
- Water has a middle density: only a part of X-rays can cross it
- Lead has a very high density: practically X-rays don’t cross it (this is the reason for which the lead is used as protective barrier)
• In the human organism there are: elements that are not very dense (soft tissue, muscles, internal organs) and elements very dense (bones)
Physical principle
Introduction To Radiology
Villa Marketing Division
• X-rays, after having crossed the body, “go out” and hit a film which is very similar to a normal photographic film
• The radiographic film is not very sensitive to X-ray , but it’s much more sensitive to common light: that’s why it is always positioned inside a “cassette”, which has several purposes:
- Protect the film from the light- Increase the sensitivity of the film, getting it in touch with “screens” that
transform X-rays into visible light- Increase the mechanical resistance
• The X-rays are noxious for the organism so it is necessary to limit the exposure of both patients and operators
• Obviously the relation risks/benefits is very favorable, in fact the radiological technique, even if is a century “old”, continues to be used
Physical principleIntroduction To Radiology
Villa Marketing Division
• Introduction To Radiology
• Key Components
• Work Techniques
• Types Of Exams
• Types Of Devices
Contents
Villa Marketing Division
• X-rays are the most popular technology for diagnostic imaging in a wide spectrum of application (bucky tables and remote controlled RF systems)
• A radiological system is mainly made up of the following elements:
- Patient positioner - Table
- X-ray source - Tube
- High voltage Generator
- Image receptor - Film, Image Intensifier, DR
• In addition, there are:
- Grid
- Collimator
- High tension cables
Key Components
Villa Marketing Division
• The table is the element which is mainly useful for support and the positioning of the patient
• Typically the table is also provided with a column, which supports the X-ray tube and allows its positioning as needed
MoviplanApollo
Table
Tube
ColumnTube
Column
Key Components
Villa Marketing Division
The table is equipped with a device to support and position the cassette:• The most simple system is called Cassette Holder and is not used frequently
• A middle system is called Potter Bucky and it is typically used for bone exams (Moviplan, chest stands)
• The most sophisticate system is called Spot Film Device and it is typically used for Gastro-Intestinal exams on R/F tables (Apollo, Vision)
Table
Spot Film Device
Potter Bucky
Key Components
Villa Marketing Division
• The X-ray tube is the component which emits the X-rays
• To emit the X-rays, the tube must be supplied with a voltage from 40 to 150 kV
• There are several models of tube, in function of the kind of the system, of the application and of workflow
• The models are different as a function of several parameters:- Power- Focal spot size- Heat storage capacity
Tube
Key Components
Villa Marketing Division
• The generator generates the high voltage to be applied to the tube to produce X-rays
• There are several models in function of the power and of the application
• The rad-only generators are used with tables as with Moviplan for bones applications
• The R/F generators are used with remote controlled tables (Apollo) and tilting tables (Vision and Viromatic) in gastro-intestinal (GI) and angiographic applications
• With the console it is possible to set three main parameters:- kV (Kilovolt): from 40 to 150- mA (milliampere): from 10 to 1000- s (seconds): from 0.001 to 20
• Changing those three parameters it is possible to adapt X-ray emission to the organ in exam and to the kind of image that the operator wants to obtain
Generator
Key Components
Villa Marketing Division
Image receptor
• The radiographic film, similar to a photographic film, is hit by the X-rays and an image with the crossed tissue density can be obtained
• The Image Intensifier subsystem is used to operate in fluoroscopy mode to convert x-rays into a bright visible light
• Digital X-ray sensors are used in Digital Radiography, instead of traditional film, and are called Flat Panel Detectors. They can be fixed or portable and contain a layer of scintillator material commonly made by Caesium Iodide (CsI) or Gadolinium oxysulfide (Gadox)
Key Components
Villa Marketing Division
• Inside the Potter Bucky or the SFD a component called grid is present and it has the purpose of eliminating the “scattered radiation”, that is a sort of fog generated by the subject in exam, which affect the image quality
• The grid is a sort of filter which cleans the image
Grid
Key Components
Villa Marketing Division
The collimator is present on the tube in the point from which X-rays are emitted, and is used to:• Limit the size of the irradiated field • Project on the patient a luminous square which shows the field size. • Reduce the stray radiation in the environment
Collimator
Collimator
Key Components
Villa Marketing Division
• The high tension cables (HT) are used to connect the generator with the tube
• Their length must be chosen in function of the system composition (14/16 m cables are the most common ones)
HT cables
Key Components
Villa Marketing Division
• Introduction To Radiology
• Key Components
• Work Techniques• Types Of Exams
• Types Of Devices
Contents
Villa Marketing Division
• kV are representative of single photons energy and therefore of their penetration power. A radiation with high kV can cross denser and/or thicker materials
• mA are representative of quantity of emitted X-photons and, under the same kV and object in exam, a radiation with higher mA makes the film darker
• Exposure time (s) can be assimilated to emitted photons quantity: under the same mA, if the exposure time doubles, also the photons quantity doubles
Work TechniquesParameters meaning
Villa Marketing Division
• mA and s are therefore comparable in the final effect (film darkening)
• The mAs (milliampere-second) unit, as result of the product mA*s, is often used
• For example, if 25 mAs are necessary to obtain a correct exposure, it is possible to use 25 mA for one second, or it is possible to use 250 mA for 0.1 s
• It is preferable to use high mA and short times to reduce the artifacts due to the movement of the patient or of the organs
Parameters meaning
Work Techniques
Villa Marketing Division
• Three points technique: the operator can set independently kV, mA and s values
• Two points technique: the operator can set kV and mAs values. The generator sets the highest possible mA value for the kV value selected
• One point technique: the operator can set only the kV value
• Zero points technique: automatic setting of kV, mA, s
Work Techniques
Villa Marketing Division
• Allow to set exposure parameters automatically
• The operator sets the organ to examine, the projection (frontal or lateral) and the patient size
• The system sets the best kV, mA, sec values in function of the exam
Work TechniquesAnatomic programs
Villa Marketing Division
• Generator accessory
• Composed of two parts:- AEC interface (it is a board, which is installed into the
generator)- Measure chamber is placed in the potter bucky or in the
SFD immediately above the cassette
• The X-rays, before hitting the cassette, cross the measure chamber, which emits an electrical signal proportional to the rays quantity which has crossed it
Automatic Exposure Control (AEC)
Work Techniques
Villa Marketing Division
• The signal is sent to interface board, which elaborates it and informs the generator to interrupt the exposure when the radiation dose emitted is sufficient to give the correct film darkening
• Substantially the parameter on which the AEC acts is the exposure time
• It is called also 1 point technique: the operator can set only the kV value, manually or through an anatomic program (in this case a mixed anatomic technique + AEC is used). The mA value is set automatically with the max possible value for that particular kV value
Automatic Exposure Control (AEC)
Work Techniques
Villa Marketing Division
• Conventional generator: the wave-shape of the high tension is directly linked to the 50 Hz mains frequency. The kV have a sinusoidal shape which goes from 0 to 100 kV (in example)
• HF generator: the wave-shape of the high tension is generated by an inverter which works with frequencies of several kHz and it is further “levelled” so that it is almost constant
Conventional and HF generators
t
kV
tt
kV
Conventional X-ray generator
Constant Potential (HF) X-ray generator
X-ray pulses
Work Techniques
Villa Marketing Division
• Higher efficiency (under the same application, it is sufficient a lower power)
• Reduction of the patient dose
• Insensitivity to mains voltage fluctuations
• Reduction of “soft rays”, that are the rays generated in the rising and falling portions of the wave. They don’t contribute to the image formation, but are noxious for the patient
Advantages of HF technology
Work Techniques
Villa Marketing Division
• Introduction To Radiology
• Key Components
• Work Techniques
• Types Of Exams
• Types Of Devices
Contents
Villa Marketing Division
• Radiography: it is comparable to a photo
• Serial radiography: it is comparable to a sequence of photo taken at very short intervals
• Fluoroscopy: it is comparable to a movie
• Pulsed fluoroscopy: it emits X-ray as a series of short pulses instead of a continuous flow, reducing the patient dose
• Digital radiography or fluoroscopy: it allows to operate in each type of exam in digital way
• Tomography: it make a sectional image by moving an X-ray source and the film in opposite directions during the X-ray exposure
Types Of Exams
Villa Marketing Division
• It allows to take static images (chest, bones)
• It puts in evidence fine details
• It is executed on:
- Bucky tables (Moviplan) and chest stands (for chest exams) thanks to the Potter Bucky
- Remote controlled tables (Apollo) thanks to under-table SFD
- Tilting tables (Vision, Viromatic) thanks to under-table potter bucky
Radiography
Types Of Exams
Villa Marketing Division
• It is used typically for dynamic GI (gastro-intestinal) study
• More images (from 2 to 6), called divisions, at intervals of 0.5-1 sec. each other, are taken on the same film
• It is executed on R/F (Radio/Fluoro) tables, tilting or remote controlled, through the SFD
Serial radiography
Types Of Exams
Villa Marketing Division
• It allows to take pictures of dynamic events (gastro-Intestinal)
• It is not useful to put in evidence details, but to see the evolution of a phenomenon
• It is executed on R/F (Radio/Fluoro) tables, tilting or remote controlled and on surgical C-arm
• The image is taken by TV Chain, made up of an II (Image Intensifier) and a CCD camera, and is visualized on the monitor
Fluoroscopy
Types Of Exams
Villa Marketing Division
• Study of dynamic phenomenon/district with middle/high speed (up to 25 img/s)
• It is used typically for digestive and vascular apparatus study
• It allows to execute angiographic examination with DSA (Digital Subtraction Angiography)
• The system is based on a I.I, a high resolution CCD camera and a digital processing system (DIVA-D)
Digital fluoroscopy
Types Of Exams
Villa Marketing Division
• It is used to take internal organs sections
• It is based on the principle of “cancellation” of structures which are outside the slice layer
• The cancellation is obtained with an defocussing procedure, thanks to a relative movement between the film and the structures which are not interesting
• For example it is used for kidney, liver and other exams
• It is often replaced by Computed Tomography and Magnetic Resonance
Tomography
Types Of Exams
Villa Marketing Division
The effect of “cancellation” of the structures which are outside the focal layer is obtained through the movement in opposite directions of the source and the film respect to the organ.
Tomography
Film/screen
Table surface
Patient body
X-ray source movement
Cassette movement
Focal plane
Types Of Exams
Villa Marketing Division
• It can be executed on:
- Remote controlled tables (standard)
- Tilting tables, through optional tomographic column
- Bucky table through optional tomographic column
• The simplest systems use a mechanical bar which synchronizes the tube and the potter movement
• The most sophisticated systems use an electronic system without bar (electronic tomography)
Tomography
Types Of Exams
Villa Marketing Division
Types Of ExamsExposure parameters
Parameter Radiography, serialRadiography, digital
Fluoroscopy
kV 40 – 150 40 – 120
mA 10 – 1000 0,5 – 10
Times 1 ms – 10 sec From few seconds to some
minutes
Villa Marketing Division
• Introduction To Radiology
• Key Components
• Work Techniques
• Types Of Exams
• Types Of Devices
Contents
Villa Marketing Division
Types Of Devices
• Remote controlled tables
• Conventional titling tables
• Bucky tables
• Chest stands
• Ceiling tubestands
• Mobile units
• Surgical “C” arms
• Mammographic units
• U-arm Digital Radiographic Systems
Villa Marketing Division
• It is the device which allows to execute the widest number of R/F and RAD exams
• The control console is positioned in a separate room
• It is typical of Italian and French schools
• It allows to make the best of digital acquisition systems
Remote controlled tables
Types Of Devices
Villa Marketing Division
• It allows to execute R/F exams
• To execute RAD exams, an under-table potter bucky and a second tube on column or ceiling tubestand are needed
• The controls are on the table itself
• It is typical of Anglo-Saxon school
Conventional tilting tables
Types Of Devices
Villa Marketing Division
• It is a simple device which allows to execute a wide number of exams
• The tabletop is “floating” to allows a fast and precise manual positioning
• It is often coupled with a chest stand to make a general radiography room (general rad)
• It is often provided with vertical movement thanks to a lift, which allows to lower the tabletop to about 50 cm from floor, to facilitate the access of patients with motion difficulties (elders or with extremities problems) since the device is often used for first study of casualty ward
Bucky tables
Types Of Devices
Villa Marketing Division
• It is an accessory which completes a radiological system (generally based on Moviplan or Vision/Viromatic, sometimes with remote control)
• It is used to execute chest and spinal column exams with standing patients
• It can also be used for loaded lower extremities studies
Chest stands
Types Of Devices
Villa Marketing Division
• Useful in casualty wards to have 4-side access to the bucky table, without the floor column
• For lateral projections on remote control tables
Ceiling tubestands
Types Of Devices
Villa Marketing Division
• Mobile units: mobile radiographic systems used in trauma and surgery rooms, at patient’s bed and also as backup unit in general diagnostic
• Surgical “C” arms: radiographic and fluoroscopy systems used in general and vascular surgery, traumatology and orthopedics thanks to their positioning flexibility
• U-arm Direct Digital Radiographic Systems: systems based on flat panel detectors that can cover virtually all the applications of a modern radiological room assuring flexibility
• Mammographic units: analog or digital, are available ready for biopsy
Other devices
Types Of Devices
Villa Marketing Division
Radiological tables and applicationsKind RAD
R/FModel Villa
Applications
Remote control R/F Apollo BonesGastrointestinal (GI)Angiography TomographyChest
Conventional Tilting
R/F VisionViromatic
BonesGastrointestinal (GI)Angiography (limited)Tomography
Bucky Table RAD Moviplan BonesTomography
Chest stand RAD Chest stand
ChestSpinal column Under load inf. extremities
Types Of Devices
Villa Marketing Division
Thank you for your attention!