menami oxytrace 01 - ТЕХНОАНАЛИТ

94
AMI Oxytrace AMI Oxytrace QED Version 4.22 and later A-96.250.531 / 081128 Operator’s Manual

Upload: others

Post on 27-Apr-2022

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace AMI Oxytrace QEDVersion 4.22 and later

A-96.250.531 / 081128

Ope

rato

r’s M

anua

l

Page 2: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

Customer SupportSWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest SWAN representative, or the manufacturer:

SWAN ANALYTISCHE INSTRUMENTE AGStudbachstrasse 138340 HinwilSwitzerland

Internet: www.swan.ch

E-mail: [email protected]

Document Status

Title: Monitor AMI Oxytrace Operator’s Manual

ID: A-96.250.531

Revision Issue

00 Jan 2008

01 Nov 2008 including AMI Oxytrace QED

© 2008, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved

subject to change without notice

Page 3: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)

AMI Oxytrace QED - Operator’s Manual

PrefaceThe operator’s manual provides thorough information about the installation, operation and the maintenance of the instrument for the user and service technician.This document also contains the Profibus and Modbus instructions and the Material Safety Data Sheets.

Qualification,Training

To be qualified for instrument installation and operation, you must:

read and understand the instructions in this manual as well as the Material Safety Data Sheetsknow the relevant safety rules and regulations.

A-96.250.531 / 081128 1

Page 4: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)

Table of Contents

1. Safety Instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51.1 General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . 51.2 Warning Notices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2. Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . 82.1 Description of the system. . . . . . . . . . . . . . . . . . . . . . . . . 82.2 Instrument Specification. . . . . . . . . . . . . . . . . . . . . . . . . . 10

3. Instrument Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 123.1 Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . . . . . . 153.2 Flow cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 163.2.1 QV-Flow PMMA OTG. . . . . . . . . . . . . . . . . . . . . . . . . . 163.2.2 QV-Flow PMMA OTG-QED . . . . . . . . . . . . . . . . . . . . . 173.2.3 B-Flow SS316L OTG . . . . . . . . . . . . . . . . . . . . . . . . . . 183.3 Display, Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193.4 Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213.5 Changing parameters and values . . . . . . . . . . . . . . . . . . 22

4. Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.1 Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234.2 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.3 Connecting sample and waste . . . . . . . . . . . . . . . . . . . . . 244.4 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 254.4.1 Electrical connection scheme AMI Oxytrace . . . . . . . . 264.4.2 Electrical connection scheme AMI Oxytrace QED . . . . 274.4.3 Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284.4.4 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 294.5 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304.6 Relay Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304.6.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 304.6.2 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . 304.7 Current Signal Outputs. . . . . . . . . . . . . . . . . . . . . . . . . . . 324.7.1 Current Signal Outputs 1 and 2 . . . . . . . . . . . . . . . . . . 324.7.2 Current Signal Output 3 (Optional). . . . . . . . . . . . . . . . 324.8 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 334.8.1 Interface RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 334.8.2 Interface RS485 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 334.9 Sensor Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

2 A-96.250.531 / 081128

Page 5: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)

5. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 355.1 Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . 355.2 Stop Operation before Maintenance . . . . . . . . . . . . . . . . 355.3 Maintenance of Oxygen Sensor. . . . . . . . . . . . . . . . . . . . 365.3.1 Electrolyte exchange . . . . . . . . . . . . . . . . . . . . . . . . . . 365.4 Cleaning of Flow Cell and Sensor . . . . . . . . . . . . . . . . . . 375.5 Maintenance of the Faraday electrode . . . . . . . . . . . . . . 385.6 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 395.7 Zero-Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 405.8 Faraday Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

6. Error list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 426.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 426.2 Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

A-96.250.531 / 081128 3

Page 6: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)

A. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

B. Profibus (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48B.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48B.2 Cyclic Data Telegram. . . . . . . . . . . . . . . . . . . . . . . . . . . . 49B.3 Customizing the Cyclic Data Telegram . . . . . . . . . . . . . . 50B.4 Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51B.5 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55B.5.1 Device Master Files (GSD). . . . . . . . . . . . . . . . . . . . . . 55B.5.2 GSD File Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55B.6 Acyclic Data Communication . . . . . . . . . . . . . . . . . . . . . . 56

C. Modbus (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57C.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57C.2. Data and Control Functions . . . . . . . . . . . . . . . . . . . . . . . 59C.3 Function Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

D. Program List and Explanations . . . . . . . . . . . . . . . . . . 651 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 652 Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 653 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 674 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 685 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70

E. MSDS Filling electrolyte . . . . . . . . . . . . . . . . . . . . . . . . 82

F. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86

G. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88

4 A-96.250.531 / 081128

Page 7: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Safety Instructions

AMI Oxytrace - Operator’s ManualThis document describes the main steps for instrument setup, operation and maintenance.

1. Safety InstructionsThis chapter contains general information assuring safe operation of the instrument. More safety instructions are given throughout this manual, at the respective locations where observation is most important.Strictly follow all safety instructions in this publication. For safe instrument installation and operation you must read and understand the instructions in this manual.Only SWAN trained and authorized personnel shall perform the tasks described in this document.

1.1 General Safety RegulationsThe instructions included in this section explain the potential risks associated with the instrument operation and provide important safety practices designed to minimize these risks.If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environment.

LegalRequirements

The user is responsible for proper operating according to all local, state and federal laws that may apply. All precautions must be followed to ensure safe operation of the instrument.

GeneralInspection

The SWAN instrument shall only be operated in optimal running condition. The preventive maintenance procedures listed in this manual must be performed.In case of safety relevant deficiencies, SWAN must immediately be informed.

Spare Parts Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.

Modifications Modifications and instrument upgrades shall only be carried out by an authorized Service Technician. SWAN will not accept

A-96.250.531 / 081128 5

Page 8: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Safety Instructions

responsibility for any claim resulting from unauthorized modification or alteration.

MSDS Material Safety Data Sheet(s) listed in this manual must be read and understood. All safety and protection issues must be followed.

Electricalshock hazard

If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.

To prevent from electrical shock, always make sure thatthe ground wire is connected.Service shall be performed by authorized personnel only.Whenever electronic service is required, disconnectinstrument power and power of devices connected to• relay 1• relay 2• alarm relay

ManualLocation

The AMI Operator’s Manual shall be kept in proximity of the instrument.

6 A-96.250.531 / 081128

Page 9: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Safety Instructions

1.2 Warning NoticesThe symbols used for safety-related notices have the following significance:

WARNINGSymbols

Warning

Generally, the triangular warning symbol indicates the possibility of personal injury or even loss of life if instructions are not followed, e.g.

electrical shock hazard

caustic

AttentionSymbols

AttentionThe attention symbol indicates the possibility of equipment damage, malfunctions or incorrect process results if instructions are not followed.

WarningFor safe instrument installation and operation you must read and understand the instructions in this manual.Only SWAN trained and authorized personnel shall perform the tasks described in this document.

A-96.250.531 / 081128 7

Page 10: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Product Description

2. Product DescriptionThis manual describes the functions of the monitors AMI Oxytrace and AMI Oxytrace QED. Both monitors are mainly identical except that AMI Oxytrace QED includes a faraday verification. AMI Oxytrace cannot be upgraded with the faraday verification.

2.1 Description of the systemApplication AMI Oxytrace is used to measure low levels of oxygen in high-

purity water. Especially in power plant water cycles (e.g. feedwater), a very low level of oxygen is needed to prevent corrosion.

Measuringprinciple

Clark principle:The sensor consists of one noble metal electrode (e.g. platinum or gold), a reference electrode (mostly Ag/AgCl) and optionally a metal guard electrode.The Clark-type electrode is the most used frequently oxygen sensor for measuring oxygen dissolved in a liquid. The basic principle is that there is a cathode and an anode submersed in an electrolyte and a voltage is applied between the two parts. Oxygen enters the sensor through a permeable membrane by diffusion, and is reduced at the cathode, creating a measurable electrical current.O2 + 2 e- + 2 H2O --> H2O2 + 2 OH-

There is a linear correlation between the oxygen concentration and the electrical current. The guard electrode mounted circularly around the cathode is on the same voltage level but there is no current measurement. Oxygen which diffuses from the electrolyte to the cathode is consumed by the guard. As a consequence, residual oxygen in the electrolyte will no more disturb the measurement signal and the response time to low oxygen levels will be shorter.

Temperaturecompensa-

tion

The measuring signal depends on temperature, but is automatically compensated to 25°C. The sample temperature is determined continuously by a temperature sensor inside the oxygen electrode.

On-lineoperation

Swansensor Oxytrace G combined with QV-flow PMMA OTG flow cell: The sample enters the flow cell at the sample inlet, passes the flow regulating valve, the faraday electrode and fills the flow cell. The oxygen concentration will be determined. Exiting the flow cell,

8 A-96.250.531 / 081128

Page 11: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Product Description

the sample drives the rotor of the flowmeter (for flow monitoring) and the sample leaves the flow cell through the pressure-free outlet.

FaradayVerification

When a direct current is passed through water by means of two electrodes, electrolysis of the liquid takes place, according to the laws of Michael Faraday. Water is converted to molecular oxygen and hydrogen. Thus, by controlling the current, a fixed, known amount of oxygen, which is independent of temperature and pressure, may be generated. If the sample flow is known, an exact oxygen concentration increment can be generated at low levels. This increment is used to verify the flawless operation of the whole system. Thereby the response characteristics (incremental change and response time) of the sensor are taken into account. Sensor failures (loss of electrolyte etc.) can be detected very easily as shown in the plot below:.

Unusual sensor performance is recognized immediately and gives also a maintenance indication to the operator/user. The faraday verification is therefore an excellent QA/QC tool.

A-96.250.531 / 081128 9

Page 12: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Product Description

2.2 Instrument SpecificationSample

Require-ments

Flow rate: approx. 6-25 l/hSample pressure inlet: QV-flow: 0.2 - 1 bar

B-flow: max. 3 barSample outlet: QV-flow: pressure-free

B-flow: max. 3 barTemperature range: 15 - 45 °C (41 - 113 °F)No oil, no grease, no sand.

On-siteRequire-

ments

Sample inlet: Swagelok connection for stainless steel tube 1/4”.Drain: QV-flow: 14x20 mm (1/2”) hose nozzle which must

end in convenient atmospheric waste of sufficient capacity.B-Flow: Swagelok connection for stainless steel tube 1/4”.

Power: 85-265 VAC, 47-63 Hz or 24 VDC, isolated, power consumption max. 20 VA.

MeasuringRange

Oxygen

Range: 0-20 ppm or 0-200 % saturation (at 25°C)Accuracy: 0.3%, if cal temp. = measurement temp

1.5%, at +/- 10°C deviation to cal temp.Response time: t90 < 30 sec (rising concentration)

MeasuringRange

Temperature

Measuring range: up to 60°CResolution: 0.1°C

Sample FlowMeasurement

QV-flow: With Swan sample flow meter and alarm in case of insufficient sample flow.

B-flow: None.

SafetyFeatures

No data loss after power failure, all data is saved in non-volatile memory. Overvoltage protection of in- and outputs. Galvanic separation of measuring inputs and signal outputs.

AlarmRelay

One potential free contact for summary alarm indication for programmable alarm values and instrument faults.Maximum load: 1A / 250 VAC

Input One input for potential-free contact. Programmable hold or remote off function.

Relays Two potential-free contacts programmable as

10 A-96.250.531 / 081128

Page 13: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Product Description

limit switches for measuring valuescontrollerstimer for system cleaning with automatic hold functionRated load: 1A / 250VAC

SignalOutputs

Two programmable signal outputs for measured values (freely scalable, linear, bilinear or logarithmic) or as continuous control output (control parameters programmable).Current loop: 0/4 - 20mAMaximum burden: 510ΩThird signal output with the same specifications as option available.

Communi-cation Inter-

face (option)

RS232 interface for logger download with HyperTerminal or RS485 interface with Fieldbus protocol Modbus or Profibus DP V1.

ControlFunction

Relays or current outputs programmable for 1 or 2 pulse dosing pumps, solenoid valves or for one motor valve. Programmable P, PI, PID or PD control parameters

A-96.250.531 / 081128 11

Page 14: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3. Instrument OverviewAMI Oxytrace

Mounting panel

Transmitter

Swansensor Oxytrace G

Flowmeter

QV-Flow PMMA OTG:Flow cell mounted on steelangle

Sample outlet

Sample inlet

12 A-96.250.531 / 081128

Page 15: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

AMI Oxytrace QED

Mounting panel

Transmitter

Swansensor Oxytrace G

Flowmeter

QV-Flow PMMA OTG-QED:Flow cell with faraday electrode, mounted on steel angle

Sample outlet

Sample inlet

Faraday control unit

A-96.250.531 / 081128 13

Page 16: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

Dimensions

Mounting: Panel: Distance between mounting holes: 254x824 mmScrews: 6 mm in diameter

14 A-96.250.531 / 081128

Page 17: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.1 Swansensor Oxytrace GThe Swansensor Oxytrace G is used for the measurement of dissolved oxygen in high purity water and is designed for low level measurement.

Cable with sleeves

Sensor head (black)

Sensor cap (white)

Measuring range: 0 - 20 ppm O2 (at 25°C)0 - 200% saturation

Accuracy: 0.3%, if cal temp. = measurement temp1.5%, at +/- 10°C deviation to cal temp.

Temperature sensor: NT5kOperating temp: 0 - 50 °COperating pressure: 0 - 3 barSensor mounting: threadMaterial: shaft: polyacetal copolymer

cathode/anode/guard: gold/silver/silvermembrane: fluoropolymer

Response time: Approx. 120 sec. from air to 100 ppbApprox. 270 sec. from air to 10 ppbApprox. 30 min. from air to 1 ppb

Minimal sample flow: 50 cm/s

A-96.250.531 / 081128 15

Page 18: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.2 Flow cells3.2.1 QV-Flow PMMA OTG

Flow cell made of acrylic glass with integrated flow sensor. Sample temperature: max. 45 °CPressure inlet QV-Flow : max. 1 bar The sample outlet has to be pressure-free against atmosphere. Sample flow QV-Flow: 6 - 25 l/hProcess connection inlet: Swagelok connection for 1/4” tubeOutlet: SERTO angle +6mm flexible tubeDimensions: see picture below

Note: QV-Flow PMMA OTG is delivered with AMI Oxytrace.

Sample inlet

Sample outlet

Flowmeter

Oxygensensor

Sample inlet

Sample outlet

Flowmeter

Oxygensensor

Sample outlet

Flowmeter

Oxygensensor

16 A-96.250.531 / 081128

Page 19: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.2.2 QV-Flow PMMA OTG-QEDFlow cell made of acrylic glass with integrated flow sensor.QV-Flow PMMA OTG-QED includes a faraday electrode to perform a faraday verification (see Faraday Verification, p. 41 for more details).Sample temperature: max. 45 °CPressure inlet QV-Flow: max. 1 bar The sample outlet has to be pressure-free against atmosphere. Sample flow QV-Flow: 6 - 25 l/hProcess connection inlet: Swagelok connection for 1/4” tubeOutlet: SERTO angle +6mm flexible tubeDimensions: see picture below

Note: QV-Flow PMMA OTG-QED is delivered with AMI Oxytrace QED.

O-Ring Øi 32 x 2,5, Vi ton

80

75

13817

8.5

80 Sample outlet

Flow meter

Oxygen sensor

Faradayelectrode

Faraday control unit

Sample inlet

Sensor mountingscrew

A-96.250.531 / 081128 17

Page 20: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.2.3 B-Flow SS316L OTGB-Flow SS316L OTG is made of stainless steel and can be used for higher operating pressures and temperatures.There is no flow sensor.Sample temperature: 0 - 50 °CPressure inlet: max. 5 bar at 50 °CProcess connection inlet and outlet:

Swagelok connections for 1/4” tubeDimensions: see picture below:

Sample outlet

Sample inlet

Sample outlet

Sample inlet

18 A-96.250.531 / 081128

Page 21: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.3 Display, KeysDisplay ofMeasured

Value backlit LCD

Menu navigation keys

upper / lower limit not yet reachedupper / lower limit reachedcontrol upw. / dnw.: no action

control upw. / dnw.: active, dark bar indicates control intensity

timertimer: timing active (hand rotating)

Relay Status

motor valve: valve closed

l/h

motor valve: open, dark bar indicates approx. position

°C

Status Time

R1 ProcessValuesR2

blinking: fatal errorerror

HOLD input closed or cal delay: Instrument on hold (shows status of signal outputs). Add. info see Program List and Explanations, p. 65.

OFF input closed: Control/limit is interrupted (shows status of signal outputs). Add. info see Program List and Explanations, p. 65.

RUN normal operation

sample temperature

sampleflow

dissolved oxygen xx.x ppb

A-96.250.531 / 081128 19

Page 22: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

Keys

ProgramAccess, Exit

to exit a menu or command (without accepting)to move back to the previous menu level

to move DOWN in a menu list and to decrease digits

to move UP in a menu list and to increase digits

to open a selected sub-menuto accept an entry

Exit

Enter

25.4°C

RUN

7.2 l/h

14:10:45R1

5.3 ppbR2

1

InstallationOperation

DiagnosticsMessages

Maintenance

Main MenuEnter

Exit

20 A-96.250.531 / 081128

Page 23: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.4 Software Structure

1

Messages

OperationMaintenanceDiagnostics

Main Menu

Menu 4: OperationSubset of menu 5 - installation, but process-related. User relevant parameters that might need to be modified during daily routine. Normally password protected and used by the process-operator.

Menu 1: MessagesReveals pending errors as well as an event history (time and state of events that have occurred at an earlier point of time) and maintenance requests. It contains user relevant data.

Menu 2: DiagnosticsProvides user relevant instrument and sample data.

Menu 5: InstallationFor initial instrument set up by SWAN authorized persons, to set all instrument parameters. Can be protected by means of a password.

Menu 3: MaintenanceFor instrument calibration and verification, elektrolyte exchange, relay and signal output simulation and to set the instrument time. It is used by the service personnel.

Installation

1.1

Pending ErrorsMessages

2.1

InterfaceI/O StateSample

IdentificationSensors

Diagnostics

3.1

Calibration

Simulation

Maintenance

Set Time 01.01.05 16:30:00

4.1

LoggerRelay ContactsSensorsOperation

5.1

InterfaceMiscellaneousRelay Contacts

SensorsSignal Outputs

Installation

Maintenance ListMessage List

Faraday VerificationElektrolyte

A-96.250.531 / 081128 21

Page 24: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Instrument Overview

3.5 Changing parameters and valuesChanging

parametersThe following example shows how to change the current loop from 4-20 mA to 0-20 mA (see Program Overview, p. 45 Installation):

Changingparameter

values

Highlight the menu item indicating the parameter you want to change.

Press .

Press or to highlight the required parameter.

Press to confirm the selection (or [Exit] to keep the previous parameter).

The selected parameter is indicated (but not saved yet).

Press .Yes highlighted, press to save the new parameter. The system reboots, the new parameter is set.

Enter

Enter

ExitEnter

5.2.1Signal output 1

Current loop.Parameter Oxygen

Scaling 1 MinFunction Free chlorine

Curr. loop.

4-20 mA

5.2.1Signal output 1

Current loop 0-20 mAParameter Oxygen

Function linNo

Save ?Yes

5.2.1Signal output 1

Function lin

Parameter OxygenCurrent loop 4-20 mA

5.2.1Signal output 1

Function lin

Parameter OxygenCurrent loop 0-20 mA

Scaling

0-20 mA

Scaling

Scaling

5.3.1.1

Alarm High 1000 ppbAlarm Oxygen

Alarm Low 0.00 ppbHysteresis 1.0 ppbDelay 5 Sec

5.3.1.1Alarm Oxygen

Alarm Low 0.00 ppbHysteresis 1.0 ppbDelay 5 Sec

Alarm High 5.0 ppb

Select parameter and press .

to set required value.

to confirm new value.

to exit the menu.

Enter

Enter

Exit

22 A-96.250.531 / 081128

Page 25: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4. InstallationThe first part of this chapter describes the preparing and placing of the system in position for use. The second part describes the wiring and the properties of the signal outputs and relay.Restrictions on useIf exposed to weather, install a cover to protect the instrument from insolation and precipitation. See Instrument Specification, p. 10 for temperature limitations.

4.1 Installation Checklist

Check Instrument’s specification must conform to your AC power ratings. Do not turn on power until instructed to do so.

On site requirements

85-265 VAC, 47-63 Hz or 24 VDC power outlet with ground connection and 20 VA.Sample line with 6-14 l/h and 0.2 - 3 bar.Waste line with atmospheric drain.See Connecting sample and waste, p. 24 andElectrical Connections, p. 25

Installation Mount the instrument in vertical position. Display should be at eye level.Connect the sample and waste line.

Electrical Wiring

Connect all external devices like limit switches, current loops and pumps. Connect power cord; do NOT switch on power yet!

Electrode The sensor is delivered with prefilled electrolyte chamber and a mounted cap with sodium sulfite solution. Before installation, remove the protective cap and clean the sensor surfaces with water. Afterwards install oxygen sensor and connect the cable

A-96.250.531 / 081128 23

Page 26: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.2 MountingFor ease of operation, mount the instrument in a way that the display is on eye level. For the panel size, see Dimensions, p. 14

4.3 Connecting sample and wasteSample inlet:Connect the 1/4” stainless steel tube to the sample line following the instructions of Swagelok fittings:Insert the tubing into the Swagelok fitting. Make sure that the tubing rests firmly on the shoulder of the fitting and that the nut is finger-tight.While holding the fitting body steady with a backup wrench, tighten the nut 1 1/4 turns.

Waste:QV-Flow: Connect the 1/2” tube to the waste nozzle and place it into the atmospheric drain.B-flow: Swagelok connection 1/4”, identical to sample inlet.

Turn-on power

Switch on power.First, the analyzer performs a self test, displays the firmware ver-sion and then starts normal operation

Run-in PeriodCalibration

Leave the sensor at the air. The sensor has to run-in for at least 30 min, better 1 h. During this time it has to be at air with a dry and clean membrane and has to be connected to the powered instru-menty. Calibrate the sensor after this time (Air Calibration Proce-dure, p. 39) and mount it into the flow-cell afterwards.

Open Sample Flow

Open the flow regulating valve to allow sample flow into the flow cell and into the waste. Sample must always overflow into waste

Programming Program all parameters for external devices (interface, etc.). Set all parameters for instrument operation (limits, alarms).

24 A-96.250.531 / 081128

Page 27: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.4 Electrical ConnectionsWarningAlways turn off AC power before manipulating electric parts.Grounding requirements: Only operate the instrument from an power outlet which has a ground connection.Make sure the power specification of the instrument corresponds to the power on site.

Cablethick-

nesses

In order to comply with IP66, use the following cable thicknesses:

Wire Power, RelaysUse max. 1.5 mm2 / AWG 14 stranded wire with end sleeves.

Signal Outputs, InputUse 0.25 mm2 / AWG 23 stranded wire with end sleeves.

WARNINGMake sure that the devices to be connected to:

relay 1 & relay 2alarm relay

are disconnected from the power before resuming installation.

WARNING

To prevent from electrical shock, do not connect the instrument to the power unless the ground wire (PE) is connected.Do not connect unless specifically instructed to do so.

PG 9 fittings: Cable Øouter 4 - 8 mm

PG 11 fitting: Cable Øouter 5 - 10 mm

PG 7 fittings: Cable Øouter 3 - 6.5 mm

Protect unused glands.

A-96.250.531 / 081128 25

Page 28: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.4.1 Electrical connection scheme AMI Oxytrace

Use only the terminals shown in this diagram, and only for the men-tioned purpose. Use of any other terminals will cause short circuits withpossible damage to material and personnel.

26 A-96.250.531 / 081128

Page 29: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.4.2 Electrical connection scheme AMI Oxytrace QED

Use only the terminals shown in this diagram, and only for the men-tioned purpose. Use of any other terminals will cause short circuits withpossible damage to material and personnel.

A-96.250.531 / 081128 27

Page 30: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.4.3 Terminals

AMITransmitter

The AMI transmitter has a numbered terminal block inside. Connect all parts like sensor, flow meter, communication boards etc. to the correct terminal as described in Electrical connection scheme AMI Oxytrace, p. 26 or Sensor Connections, p. 34

L NPowersupply

Relay1

Relay2

Alarmrelay - + +

Signal output1 2

Input

Sampleflow

Signal output 3 - +OR

BUS interface PA/BPB/AGround

PROFIBUS

28 A-96.250.531 / 081128

Page 31: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

Faradaycontrol unit

The faraday control unit has a numbered terminal block inside. Connect faraday electrode as described inSensor Connections, p. 34

4.4.4 Power Supply

Terminal 1 Phase LTerminal 2 Neutral N

Attention: It is not allowed to power external loads out of the AMI transmitter.

Connection toAMI Transmitter

Connection tofaraday electrode

power supply connector

Neutral

Power

The protective earth cable (ground) MUST be connected to the provided grounding screw.

A-96.250.531 / 081128 29

Page 32: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.5 InputNote: Use only potential-free (dry) contacts.Terminals 16/42For programming see Program Overview, p. 45 Menu Installation

4.6 Relay Contacts4.6.1 Alarm Relay

Note: Max. load 1 A (time delay) / 250 VACAlarm output for system errors. Error codes see Error list, p. 42.Open during normal operation. Closed on error and loss of power.Terminals 10/11.

4.6.2 Relay Contacts 1 and 2

Note: Rated load 1 A (time) / 250 VACRelay 1: Terminals 6/7Relay 2: Terminals 8/9For programming see Program Overview, p. 45, Menu Installation

AttentionHeavy inductive or direct controlled loads (solenoid valves, dosing pumps) must be switched by external power relays. In this case, use the AMI relay box which is available as an option.Small inductive loads (max 0.1 A) as for example the coil of a power relay can be switched directly. To avoid noise voltage in the AMI it is mandatory to connect an snubber circuit in parallel to the load (this procedure is not necessary if a AMI Relaybox is used)

30 A-96.250.531 / 081128

Page 33: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

Resistive loads (max. 1 A) and control signals for PLC, impulse pumps and so on can be connected without further measures.

Actuators are using both relays: One relay contact is used for opening (relay 1), the other for closing (relay 2) the valve, i.e. with the 2 relay contacts available, only one motor valve can be controlled.Motors with loads bigger than 0.1 A must be controlled via external power relays or an AMI Relaybox.

A-96.250.531 / 081128 31

Page 34: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.7 Current Signal Outputs4.7.1 Current Signal Outputs 1 and 2

Note: Max. burden 510 Ω.If signals are sent to two different receivers, use signal isolator (loop isolator).

Signal output 1: Terminals 14 (+) and 13 (-)Signal output 2: Terminals 15 (+) and 13 (-)For programming see Program Overview, p. 45, Menu Installation

4.7.2 Current Signal Output 3 (Optional)Requires the additional board for thethird signal output (0/4 - 20 mA PCB

Note: Max. burden: 510 Ω.)The third signal output is installed in the upper holder on the main board. You can operate either the third signal output or the communication interface, not both!Terminal 38 (+) and 37 (-).For programming see Installation Signal output 3 in Program Overview, p. 45.

Third signal output 0/4 - 20 mA PCBinstalled on main board

32 A-96.250.531 / 081128

Page 35: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.8 Interface4.8.1 Interface RS232

Terminal 50, 52, 53The AMI Interface RS232 Print Circuit Board (=PCB) is used for Logger download and Firmware upload. For detailed information see the corresponding manual “AMI RS232 Interface”.

RS232 Interface PCB

4.8.2 Interface RS485Terminal 37 PB, Terminal 38 PATo connect several instruments by means of a network or to configure a PROFIBUS DP connection. Use appropriate network cable.

.

RS485 PCB

Note: Bus termination only by means of ON/OFF switch on the RS485 PCB. If only one instrument is installed, or on the last instrument in a bus, the switch must be ON.

on/off switch

A-96.250.531 / 081128 33

Page 36: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Installation

4.9 Sensor Connections SwansensorOxytrace G

Connect the sensor as given bellow:

Swanflow meter

Connect the Swan flow meter of the QV-Flow cell as given below.

Faradayelectrode

Connect the faraday electrode to the faraday control unit (only Oxytrace QED) as given below:

For further details see:Electrical connection scheme AMI Oxytrace, p. 26Electrical connection scheme AMI Oxytrace QED, p. 27

Black Brown Green White

19 20 21 22

Black White Brown

39 40 41

White Green

6 5

34 A-96.250.531 / 081128

Page 37: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

5. MaintenanceMaintenance frequency depends strongly on the water quality. The AMI Oxytrace is designed for determination of low level of dissolved oxygen in high purity water. It is NOT suitable for the measurement of dissolved oxygen in waste water.

5.1 Maintenance Schedule

*A change of membrane and electrolyte is recommended:if indicated in the maintenance list (remaining amount <10%)if the response of the sensor is slowif the sensor can not be calibrated any more and/or the instrument shows a corresponding error messageif the sensor signal is very unstable.

5.2 Stop Operation before MaintenanceStop sample flow.Shut off power of the instrument.To take the sensor out of the flow cell, turn the sensor mounting screw counter clockwise. Unscrew it completely. Open the sample valve slightly for a moment. This will loosen the tight

Weekly Check sample flow.Make manual measurement.If necessary, correct.

Monthly If necessary, perform an air calibration.

Half a year Clean Oxytrace G membrane with a soft tissue.Clean faraday electrode

Yearly If necessary, replace filling electrolyte.If the sensor is exposed to air frequently and during long time intervals, the electrolyte and membrane may have to be changed earlier (see below*).Clean flow cell and flow meter, if dirty.

Every 2nd year

Replace Oxytrace G electrode membrane by using a new, prefilled sensor cap.

A-96.250.531 / 081128 35

Page 38: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

seal of the sealing ring, and the electrode may be taken out of the flow cell easily. Do NOT pull the sensor with force!

5.3 Maintenance of Oxygen SensorFrom time to time it is necessary to exchange the electrolyte and membrane.

Warning: The filling solution is alkaline and caustic. It contains less than 1% of potassium hydroxide.

Do not ingest. Wear protective goggles and gloves during handling. Avoid contact with clothes.In case of accidental contact with the eyes, wash immediately with clear water and contact a physician. Show him the label of the bottle or this section of the manual.Short contact with skin is harmless, nevertheless wash with lots of water.

5.3.1 Electrolyte exchangeAn electrolyte exchange is indicated in the maintenance list as soon as the remaining amount is below 10%.

E

A

C

D

A GrooveB ThreadC AnodeD Cathode

Guard electrodeE Sensor cap

B

36 A-96.250.531 / 081128

Page 39: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

Switch of power.Screw off the sensor cap. Exhaust the remaining electrolyte.Refill the sensor cap with fresh electrolyte.There is a groove in the electrode thread, where excessive air and electrolyte can escape while mounting the sensor cap. Hold the sensor in upright position, sensor tip pointing downwards, with the groove on the upper side.Slowly fix the electrode head and the sensor cap to allow the excess electrolyte to escape without building up too much pressure inside the electrode. Fasten sensor cap tightly.Clean the electrode thoroughly and dry the sensing membrane with a soft tissue.Switch on power. Let the sensor run-in at air for at least 30 min, better 1 h. Afterwards perform an air calibration and install the sensor into the flow cell again.Reset the counter for remaining electrolyte (menu 3.2.5) by selecting New Filling - YES.

5.4 Cleaning of Flow Cell and SensorDepending on the water quality, the sensor and the flow cell will necessitate a cleaning.Before cleaning, stop operation as described in Stop Operation before Maintenance, p. 35. Afterwards:

Dismount the oxygen sensorClean the sensor with a soft tissue and rinse it with water afterwardsUse a soft brush to remove the dirt which sticks on the walls of the flow cell.Flush the flow cell with clean water. Remount the oxygen sensor and start sample flow.

A-96.250.531 / 081128 37

Page 40: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

5.5 Maintenance of the Faraday electrodeTurn off the instrument and close sample valve.Disconnect Faraday electrode cable (A) from faraday control unit.Loose black screw (B) completely and pull out Faraday electrode (do not pull cable).Strip of washer (C) and o-ring seal (D).

Attention: Take care when disassembling the electrode. Do not bend nor force.

Unscrew electrode tip containing the hollow electrode (G).Clean the inner electrode (E) with a tissue, the hollow electrode with a pipe cleaner. The electrode surfaces should be shining metallic after cleaning. Use probably a polishing detergent or a small amount of toothpaste.Rinse well with water.Reassemble hollow electrode finger-tight and place into flow-through cell. Replace o-ring, washer and black screw, if necessary.Tighten screw (B) firmly.Connect electrode cable (see Sensor Connections, p. 34) and turn instrument on.Turn on sample flow. Check for air bubbles and set flow to roughly 8 l/h.

A CableB Screw (black)C WasherD O-Ring sealE Inner electrodeF O-Ring sealG Hollow electrode

38 A-96.250.531 / 081128

Page 41: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

5.6 CalibrationAir

CalibrationProcedure

If the instrument is in operation, turn off the sample flow by closing the appropriate valve (sample inlet).Take the electrode out of the flow cell:

– Turn the fastening of the sensor counter clockwise. Unscrew it completely.

– Open the sample valve slightly for a moment.This will loosen the tight seal of the sealing ring, and the electrode may be taken out of the flow cell easily.Avoid pulling the electrode with force.

Dry the electrode's body and sensing membrane with a soft paper tissue. This is very important!Place the electrode slightly tilted into the flow cell, so that the sensor cap of the electrode rests on the rim for the sealing ring.In the wet flow cell, the atmosphere will be saturated with water vapour. This atmosphere will produce the most accurate calibration results. The sensing part of the electrode must not be in direct contact with water!Choose menu 3 (maintenance), then calibration (point 3.1). Press Enter. Follow instructions on the screen.The calibration is now done automatically. As soon as the reading is stable, the microprocessor will store the calibration data in the memory.The required time for a calibration depends mainly on the difference between temperature and oxygen content in the sample and in the air. It can take 15 - 20 minutes. This is also the case, if the electrolyte was exchanged. The end of the calibration is indicated on the display. Push Enter to confirm the calibration.Insert the electrode back into the flow cell, push down completely. Tighten the sensor fastening.Turn the sample flow on and off until all air bubbles are flushed out.

A-96.250.531 / 081128 39

Page 42: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

Hold aftercalibration

During calibration, control is interrupted. The signal outputs are frozen if hold after calibration has been programmed (menu 4.3.4.2). Otherwise the outputs track the measuring value. Hold after calibration is indicated by Hold in the display.For additional explanations see Error Codes, p. 43

Calibrationverification

The calibration value should be verified monthly. Leave the sensor with clean and dry membrane at air. Set the display to % saturation and check the calibration value. The air pressure is automatically acquired during calibration. Therefore the influence of altitude and pressure variations because of the weather are eliminated.

5.7 Zero-VerificationSWANSENSOR OXYTRACE G for the measurement of low oxygen content (< 1 ppb).

Prepare a 5%-sodium sulfite solution with demineralized water.If necessary, install the electrode according to the description in the manual (Maintenance of Oxygen Sensor, p. 36), i. e. place a measuring membrane into the head screw cap of the electrode and fill it with electrolyte.Calibrate according to the manual (Air Calibration Procedure, p. 39).Put the electrode into the sodium sulfite solution afterwards. Assure that there are no air bubbles in front of the sensor.The measured value should now be < 1 ppb.Depending on the state of the electrode this process can take several hours. In case of an electrode-refill, this can take days until the measured value is lower than 1 ppb.

40 A-96.250.531 / 081128

Page 43: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Maintenance

5.8 Faraday VerificationIt is possible to program an automatic or to perform a manual faraday verification.

AutomaticVerification

Per default the instruments performs an automatic faraday verification every 3 hours. To change the settings for automatic verification go to menu Faraday Parameter- Timer mode, see menu 4.1.3.1, p. 68 for details. Possible settings are: off, interval, daily or weekly.

ManualVerification

To start a manual verification, go to menu 3.3 and choose Faraday verification. The verification then starts immediately.

Results are saved in the Verification history menu 2.2.1.5

3.3

Progress

Manual Verification

<Enter> to stop

Current Value 1.62 ppbFaraday Conc. 12.85 ppb

Wait till the progress bar indicates the end

3.3

Progress

Manual Verification

Done

Efficiency 91.5 %Faraday Conc. 12.85 ppb

EnterPress enter to finish the manual verification

A-96.250.531 / 081128 41

Page 44: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Error list

6. Error list

6.1 Error messagesThe menu 1 - Messages - reveals pending errors as well as the message list with text, time and state of errors that have occurred before.

Error Nonfatal error. The system continues to operate normally, except for the defective part.

Fatal error (blinking symbol)If used for control relay 1, 2 as well as the Signal Outputs are deactivated, i. e. any control is interrupted. The indicated measured values are incorrect. Depending on the error the system may stop operating.

To solve the problem, please consult Error Codes, p. 43

If an error is corrected, by intervention or by the system itself, it is listed in the Message List (Menu 1.2). The symbol disappears.The last 64 messages are stored for viewing. They are deleted if SET Default Values (Menu 5.4.2) is carried out.

1.1.5Pending ErrorsError Code E010

Sample Flow low

<Enter> to Acknowledge

Errors remain in the Pending Errors window until corrected, even if acknowledged.Acknowledging the error only deactivates the Alarm Relay, i.e. it switches off an alarm device if available.

42 A-96.250.531 / 081128

Page 45: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Error list

6.2 Error CodesE0xx (bold and red) are fatal errors.

Error Error text Corrective action

E001 Oxygen Alarm high check process

E002 Oxygen Alarm low check process

E003 Saturation Alarm high check process

E004 Saturation Alarm low check process

E007 Sample Temp. high check sample temperature

E008 Sample Temp. low check sample temperature

E009 Sample Flow high check sample inlet pressure, check sample flow, check alarm value in Installation/Relays/Alarm relay/Sample

E010 Sample Flow low establish sample flow, check sample inlet pressure, check alarm value in Installation/Relays/Alarm relay/Sample,clean instrument

E011 Temp. shorted check wiring of temperature sensor, check temp. sensor

E012 Temp. disconnected check wiring of temperature sensor, check temp. sensor

E013 Case Temp. high check environment temperature

E014 Case Temp. low check environment temperature

E017 Control time-out check control device or programming in Installation, Relay contact, Relay 1/2

E018 Faraday Efficiency clean Faraday electrodedo an air calibrationmaintenance of the oxygen sensor

E024 Input active Information that the Input is active (see program-ming Installation, Input, Fault “Yes”)

A-96.250.531 / 081128 43

Page 46: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Error list

E025 IC MK41T56 call service

E026 IC LM75 call service

E027 IC PCF8574 call service

E028 EEProm Microcon call service

E029 EEProm Motherboard call service

E030 EEprom Front-End call service

E031 Calibration RecOut call service

E032 Wrong Front-End call service

E033 Power-on none, status message

E034 Power-down none, status message

Error Error text Corrective action

44 A-96.250.531 / 081128

Page 47: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program Overview

A. Program Overview

Menu Messages may be protected by a password. Menu Diagnostics is always accessible for everybody. No password protection. In both menus, no settings can be modified.

Messages (Main Menu 1)

Pending Errors 1.1* Maintenance List 1.2* Message List 1.3*

Diagnostics (Main Menu 2)

Identification Designation 2.1* 2.1* Version 2.1* Factory Test 2.1.3*

Operating Time 2.1.4*

Sensors Oxytrace G Current Value 2.2.1* 2.2* 2.2.1* Raw Value 2.2.1*

Saturation 2.2.1* Cal. History 2.2.1.4* Ver. History 2.2.1.5*

Miscellaneous Case Temp. 2.2.2.1* 2.2.2* Air pressure 2.2.2.1*

Sample Sample ID 2.3.1* 2.3* Temperature [°C] 2.3.1* Temperature Nt5k [Ohm] 2.3.1* Sample Flow [l/h] 2.3.1* Sample flow raw value [Hz] 2.3.1*

I/O State Alarm Relay 2.4.1 & 2.4.2* 2.4* Relay 1 & 2 2.4.1 & 2.4.2*

Input 2.4.1 & 2.4.2* Signal Output 1 & 2 2.4.1 & 2.4.2*

Interface Protocol 2.5.1* (only with RS485 or RS232 interface board)2.5* Baud rate 2.5.1*

* Menu numbers

A-96.250.531 / 081128 45

Page 48: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program Overview

Menu Maintenance is for service: Calibration, simulation of outputs and set time/date. Please protect with password.Menu Operation is for the user, allowing to set limits, alarm values, etc. The presetting is done in the menu Installation (only for the System engineer). Please protect with password.

Maintenance (Main Menu 3)

Calibration 3.1*

Elektrolyte Last Filling [date] 3.2.* 3.2* Remaining amount [%] 3.2.* Remaining time [d] 3.2.* New Filling [yes/no] 3.2.5*

Faraday Verification Current Value [ppb] 3.3* 3.3* Faraday conc. [ppb] 3.3* Progress 3.3*

Simulation Alarm Relay 3.4.1* 3.4* Relay 1 & 2 3.4.2* & 3.4.3*

Signal Output 1 & 2 3.4.4* & 3.4.5*

Set Time 3.5*

Operation (Main Menu 4)

Sensors Filter Time Const. 4.1.1* 4.1* Hold after Cal. 4.1.2*

Faraday Parameter 4.1.3* Timer Mode 4.1.3.1* [off, interval, daily, weekly]

Relay Contacts Alarm Relay Alarm Oxygen 4.2.1.1* 4.2* 4.2.1* Alarm Saturation 4.2.1.2*

Relay 1 & 2 (Parameters depend on chosen 4.2.2* & 4.2.3* settings in menu Installation)

Input Active 4.2.4.1* 4.2.4* Signal Outputs 4.2.4.2* Output/Control 4.2.4.3* Fault 4.2.4.4* Delay 4.2.4.5*

Logger Log Interval 4.3.1* 4.3* Clear Logger 4.3.2*

* Menu numbers

46 A-96.250.531 / 081128

Page 49: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program Overview

Menu Installation: Defining assignment of all inputs and outputs, measuring parameters, interface, passwords, etc. Menu for the system engineer. Password strongly recommended

Installation (Main Menu 5)

Sensors Miscellaneous Flow 5.1.1.1* 5.1* 5.1.1* O2 Offset 5.1.1.2*

Signal Outputs Signal Output 1 & 2 Parameter 5.2.X.1*5.2* 5.2.1* & 5.2.2* Current Loop 5.2.X.2*

Function 5.2.X.3* Scaling / Control Parameters 5.2.X.4X*

Relay Contacts Alarm Relay Alarm Oxygen 5.3.1.1* 5.3* 5.3.1* Sample Flow 5.3.1.2* Sample Temp 5.3.1.3* Alarm Saturation 5.3.1.4* Case Temp. 5.3.1.5*

Relay 1 & 2 Function 5.3.X.1* 5.3.2* & 5.3.3* Parameter 5.3.X.20*

further settings depends on chosen function

Input Active 5.3.4.1* 5.3.4* Signal Outputs 5.3.4.2* Output/Control 5.3.4.3* Fault 5.3.4.4* Delay 5.3.4.5*

Miscellaneous Language 5.4.1*5.4* Set defaults 5.4.2*

Load Firmware 5.4.3*Password 5.4.4*Sample ID 5.4.5*

Interface Protocol 5.5.1* (only with RS485 or RS232 interface)5.5* Baud rate 5.5.2*

* Menu numbers

A-96.250.531 / 081128 47

Page 50: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

B. Profibus (option)

B.1 IntroductionThe AMI family of transmitters provides a common base for a number of analytical instruments with application specific sensor interfaces. Each instrument has a basic structure with identical functionality.

AMI instruments can (optionally) be equipped with a PROFIBUS DP-V1 interface. The interface supports the PROFIBUS-PA Profile 3.0. This part of the manual describes the functionality of the interface and the integration of the AMI family of transmitters into an automation system.

Hardware Wiring, bus terminals and connectors are not part of this manual. It should, however, be noted that the observation of the corresponding standards is of paramount importance to proper operation. Setting the transmission rate to reasonable values will help avoid communication errors.

GSD File forAMI Oxytrace

The PROFIBUS system requires a description of the device parameters, e. g. output data, input data, data format, data volume and supported transmission rate, so that it can integrate the field devices into the bus system. These data are contained in a Device Master File (GSD file) which is placed at the disposal of the PROFIBUS master while communication system is being commissioned.

Device bitmaps can also be integrated. These appear as icons in the network tree.The manufacturer-specific GSD file SW17E1.gsd is available for download from http://www.swan.ch

48 A-96.250.531 / 081128

Page 51: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

B.2 Cyclic Data Telegram AMI Oxytrace makes the following modules available as input data for the cyclic data exchange:

1 Main Process Value (ppb)2 Main Temperature 3 Main Flow (l/h)4 2nd Process Value (%)5 Control AMI6 Diagnostic Values (5 values)

a Raw Current Temperature Compensated in nA b Raw Current in nAc Air Pressure in hPad Remaining electrolyte in %e Case Temperature in °C

The input data is transferred from AMI Oxytrace in the following structure :

Figure 1: Input data structure of AMI OxytraceIndex input data

Data Access Data format / comments Configuration data

0 ... 4 Analog Input block 1 Main Process Value

read Measured value (32-bit floating point number) Status byte

0x42, 0x84, 0x08, 0x05 or 0x42, 0x84, 0x81, 0x81 or 0x94

5 ... 9 Analog Input block 2 Main Temperature

read Measured value (32-bit floating point number) Status byte

0x42, 0x84, 0x08, 0x05 or 0x42, 0x84, 0x81, 0x81 or 0x94

10 ... 14 Analog Input block 3 Main Flow

read Measured value (32-bit floating point number) Status byte

0x42, 0x84, 0x08, 0x05 or 0x42, 0x84, 0x81, 0x81 or 0x94

15 ... 19 Analog Input block 4 2nd Process Value

read Measured value (32-bit floating point number) Status byte

0x42, 0x84, 0x08, 0x05 or 0x42, 0x84, 0x81, 0x81 or 0x94

20 ... 39 Diagnostic Values read 5 Measured values (32-bit floating point numbers)

0x45, 0x93, 0x08, 0x08, 0x08, 0x08, 0x08, 0xE9

A-96.250.531 / 081128 49

Page 52: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

The output data is transferred to AMI Oxytrace in the following structure:

B.3 Customizing the Cyclic Data TelegramYou can customize the cyclic telegram to better meet the requirements of a process. The tables above represent the maximum contents of the cyclic data telegram. If you do not want to use all the cyclic data of AMI Oxytrace, you can use the device configuration (Chk_Cfg) to eliminate individual data blocks from the cyclic telegram via the PLC software. Shortening the telegram improves the data throughput rate of a PROFIBUS system. You should only keep those blocks active which you process further in the system. You can do this by means of a "negative" selection in the configuration tool. To achieve the correct structure of the cyclic data telegram, the PROFIBUS master must send the identification FREE_PLACE (00h) for the non-active blocks.

Index output data

Data Access Data format / comments Configuration data

0 ... 1 Control AMI (SP_D)

write Byte 0 Bit Description 0 Relay 1

0 = open 1 = close

1 Relay 2 0 = open 1 = close

2 Not used 3 Not used 4..7 Not used

Byte 1

Bit Description 0 Hold 1 Control Off 2..7 Not used

0x82, 0x81, 0x05, 0x05 or 0x82, 0x81, 0x84, 0x82 or 0xA1

50 A-96.250.531 / 081128

Page 53: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

B.4 Data FormatsEach measured value has the same format: one four-byte (32-bit) floating-point value in IEEE 754 Short Real Number format and a status code byte referring to the measured value. There are four quality states, 16 sub-status values (not all defined) for each quality state, and four limit states for the measured value indicated in its associated status byte (See Figures 2,3, 4 and 5).

Figure 2: Out Value + Status

Figure 3: IEEE 754 floating point number:

Formula: Value = (-1) sign * 2(exponent - 127) * (1 + mantissa) Example: 40 F0 00 00 h = 0100 0000 1111 0000 0000 0000 0000 0000 b Value = (-1)0 * 2(129 - 127) * (1 + 2-1 + 2-2 + 2-3) = 1 * 22* (1 + 0.5 + 0.25 + 0.125) = 1 * 4 * 1.875 = 7.5

Figure 4: Status Code Byte

Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Analog values in IEEE 754 Format Status

Byte n Byte n+1 Byte n+2 Byte n+3 Bit 7 Bit 6 Bit 0 Bit 7 Bit 6 Bit 0 Bit 7 Bit 0 Bit 7 Bit 0 Sign 27 26 25 24 23 22 21 20 2–1 2–2 2–3 2–4 2–5 2–6 2–7 2–8 2–9 2–10 2–11 2–12 2–13 2–14 2–15 2–16 2–17 2–18 2–19 2–20 2–21 2–22 2–23

Exponent Mantissa Mantissa Mantissa

7 6 5 4 3 2 1 0 Quality Sub-status Limits Meaning

0 0 bad 0 1 uncertain 1 0 good

0 0 ok

0 1 low limited 1 0 high limited

The sub-status has a different quality

depending on the coding

1 1 constant

A-96.250.531 / 081128 51

Page 54: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

Figure 5: Status codes of the measured value

Figure 6: DP Diagnostics

Status code

Device status Meaning Limits

0x08 BAD Not connected (Temp Sensor not connected)

0x0C BAD device failure 0x10 BAD sensor failure (Temp Sensor Short Circuit) 0x47 UNCERTAIN last usable value (HOLD) CONST 0x50 UNCERTAIN sensor conversion not accurate (No Flow) OK 0x64 0x67

UNCERTAIN sensor calibration sensor calibration, hold

OK CONST

0x80 GOOD ok OK 0x89 0x8A

GOOD Warning Low Warning High

LOW_LIM HIGH_LIM

0x8D 0x8E

GOOD Alarm Low Alarm High

LOW_LIM HIGH_LIM

Byte Description 0 Status 1 1 Status 2 2 Status 3 3 Master address 4 Ident number (high byte) 5 Ident number (low byte)

Standard DP Diagnostics

6 Header (block length incl. Header byte) = 14 7 Status Type (0x81) 8 Slot Number (0x00) 9 Status Specifier

Status coding according to DP/V1

10 ..13 Diagnosis 14 ..19 Diagnosis Extension

52 A-96.250.531 / 081128

Page 55: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

The diagnosis object consists of 4 Bytes

Byte Bit Meaning when “1” Abbreviation 0 0 Hardware failure electronics DIA_HW_ELECTR 1 Hardware failure mechanics DIA_HW_MECH 2 - 3 Electronic temperature too high DIA_TEMP_ELECTR 4 - 5 Measurement failure DIA_MEASUREMENT 6 - 7 -

1 0 - 1 - 2 - 3 - 4 - 5 Maintenance required DIA_MAINTENANCE 6 - 7 -

2 0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 -

3 0 - 1 - 2 - 3 - 4 - 5 - 6 - 7 Extension Available EXTENSION_AVAILABLE

A-96.250.531 / 081128 53

Page 56: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

The diagnosis extension object consists of 6 Bytes.

Byte 5 is reserved for future extensions.

Byte Bit Meaning when “1” 0 0 Oxygen Alarm high - 1 Oxygen Alarm low - 2 Saturation Alarm high - 3 Saturation Alarm low - 4 - - 5 - - 6 Sample Temp. high DIA_MEASUREMENT 7 Sample Temp. low DIA_MEASUREMENT

1 0 Sample Flow high DIA_MEASUREMENT 1 Sample Flow low DIA_MEASUREMENT 2 Temp. shorted DIA_HW_ELECTR 3 Temp. disconnected DIA_HW_ELECTR 4 Case Temp. high DIA_TEMP_ELECTR 5 Case Temp. low DIA_TEMP_ELECTR 6 - 7 -

2 0 Control Timeout - 1 Faraday Efficiency DIA_MEASUREMENT 2 - - 3 - - 4 - - 5 - - 6 - - 7 Input Active DIA_HW_ELECTR

3 0 IC MK41T56 DIA_HW_ELECTR 1 IC LM75 DIA_HW_ELECTR 2 IC PCF8574 DIA_HW_ELECTR 3 EEprom Microcon DIA_HW_ELECTR 4 EEProm Motherboard DIA_HW_ELECTR 5 EEProm Frontend DIA_HW_ELECTR 6 Cal. Recout DIA_HW_ELECTR 7 Wrong Frontend DIA_HW_ELECTR

4 0 Electrolyte depleted DIA_MAINTENANCE 1 - 2 - 3 - 4 - 5 - 6 - 7 -

54 A-96.250.531 / 081128

Page 57: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

B.5 Configuration

B.5.1 Device Master Files (GSD)The device is ready for system integration once commissioning has been effected via the local display or the Class 2 master. The PROFIBUS system requires a description of the device parameters, e.g. output data, input data, data format, data volume and supported transmission rate, so that it can integrate the field devices into the bus system. These data are contained in a Device Master File (GSD file) which is placed at the disposal of the PROFIBUS master while the communication system is being commissioned. Device bitmaps can also be integrated. These appear as icons in the network tree. The Profile 3.0 Device Master File (GSD) allows field devices from various manufacturers to be exchanged without having to reconfigure.

B.5.2 GSD File TypesPrior to configuration, decide which GSD you want to use to operate the system.You can change the setting by means of a Class 2 master (under Physical Block-Parameter Ident_Number_Selector).

The following two Device Master Files with different functionalities are available:

Manufacturer-specific GSD with Profile 3.0 functionality (default setting):

This GSD guarantees the unlimited functionality of the field device. Device-specific process parameters and functions are therefore available.

Profile GSD's: If a system is configured with profile GSD's, it is possible to exchange devices that are supplied by various manufacturers. It is, however, essential that the cyclic process values follow the same sequence. The transmitter supports the profiles Analyzer and Multivariable (IEC 61158-2). These GSD's comprise Analog Input blocks (AI).

A-96.250.531 / 081128 55

Page 58: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Profibus (option)

GSD files for AMI Oxytrace:

Working with GSD files The GSDs must be integrated into the automation system. Depending on the software that is being used, the GSD files can be copied to the program-specific directory or can be read into the database using the import function within the configuration software.

B.6 Acyclic Data CommunicationThe AMI Series of analytical instruments is available for a wide range of parameters. Instrument configuration depends on connected sensors and application. Therefore, configuration must be performed by trained personnel with a background in analytical methods.

It should be noted that there is a clear distinction between instruments that are configured and calibrated in an application (flow, level, pressure) and instruments that are calibrated with a known standard (pH, chlorine). AMI instruments are of the second type.

To ensure reliability and consistency of the measurement, remote configuration has not been enabled at this time.

Device name Ident_number_ Selector

ID number GSD Bitmaps

Manufacturer-specific GSD with Profile 3.0 functionality:

AMI Oxytrace 1 17E1Hex SW17E1.gsd SW17E1_D.bmp SW17E1_N.bmp SW17E1_D.bmp

Profile 3.0 GSD Analyzer:

AMI Oxytrace 0 9750 Hex PA039750.gsd PA_9750n.bmp

Profile 3.0 GSD Multivariable:

AMI Oxytrace 3 9760 Hex PA039760.gsd PA_9760n.bmp

56 A-96.250.531 / 081128

Page 59: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

C. Modbus (option)

C.1 IntroductionThe AMI Oxytrace is an electronic transmitter / controller for the measurement of dissolved oxygen in high purity water. It can be operated as an intelligent slave device on the Modbus/JBUS using twisted pair RS485 lines.

Modbus/Jbus The Modbus / Jbus Protocol communicates using a master-slave technique, in which only one device (the master) can initiate transactions (queries). The other devices (the slaves) respond by supplying the requested data to the master, or by taking the action requested in the query. Typical master devices include host processors, PLCs etc. Typical slaves can be PLCs or intelligent transmitter/controller like the AMI family of the instruments.

The Query The function code in the query tells the addressed slave device what kind of action to perform. The data bytes contain any additional information the slave will need to function. For example, function code 03 will query the slave to read holding registers and respond with their contents. The data field must contain the information telling the slave which register to start at and how many registers to read.

The Re-sponse

If the slave makes a normal response, the function code in the response is an echo of the function code in the query. The data bytes contain the data collected by the slave.

SerialTransmission

modes

Modbus supports two transmission modes: ASCII or RTU. The AMI Instruments support Modbus RTU (Remote Terminal Unit) mode only.

RTU Mode In RTU mode, each 8-bit byte in a message contains two 4-bit hexadecimal characters. The main advantage of this mode is that its greater character density allows better data throughput than ASCII for the same baud rate. Each message must be transmitted in a continuous stream. Users select the desired serial port communication parameters (baud rate, parity mode, etc). The serial parameters must be the same for all devices on a Modbus network.

A-96.250.531 / 081128 57

Page 60: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

Coding SystemEight bit binary1 start bit8 data bits, least significant bit sent first1 bit for even / odd parity, no bit for no parity1 stop bit if parity is used, 2 bits if no parityCyclic Redundancy Check (CRC)RTU framing: messages start with a silent interval of at least 3 ½ character times. A similar interval of at least 3 ½ character times marks the end of the message.

A typical message frame is shown below

Start Slave Address Function Data CRC END

T1-T2-T3-T4 8 Bits 8 Bits n x 8 Bits 16 Bits T1-T2-T3-T4

58 A-96.250.531 / 081128

Page 61: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

C.2. Data and Control FunctionsData address Data Addresses in Modbus Messages

All data addresses in Modbus Messages are referenced to zero. The first occurrence of a data item is addressed as item number zero. For example:

Holding Register 40001 is addressed as register 0000 in the data address field of the message. The function code field already specifies a holding register operation. Therefore the 4x reference is implicit.

Functioncodes

Modbus Function CodesThe list below shows the function codes supported by the AMI Instruments. Codes are listed in decimal. Function codes not supported return an exception response 02 to the master.Table 1: Modbus function codes

Data re-presentation

Data representationThe process- and parameter data of the AMI Instruments are represented by float (IEEE 754), signed- or unsigned integer values. Depending on the range of the value, float, integer8, unsigned8, integer16, unsigned16, integer32 or unsigned32 representation is used.

Code Name Description 02 Read Input Status Read Alarm & Status Bits 03 Read Holding Registers Read Parameter Data 04 Read Input Registers Read Process Data 05 Force Single Coil Set Relays, fake Keystrokes etc. 16 Preset Multiple Registers Preset Parameter Data (SWAN Use only)

A-96.250.531 / 081128 59

Page 62: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

Discrete InputsAlarm &Status bits are packed as one input per bit of the data field.

Definition: big-endian and little-endianBig-endian and little-endian are terms that describe the order in which a sequence of bytes are stored in computer memory. Big-endian is an order in which the "big end" (most significant value in the sequence) is stored first (at the lowest storage address). Little-endian is an order in which the "little end" (least significant value in the sequence) is stored first. For example, in a big-endian computer, the two bytes required for the hexadecimal number 4F52 would be stored as 4F52 in storage. In a little-endian system, it would be stored as 524F.Attention: big-endian format is used !!!

60 A-96.250.531 / 081128

Page 63: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

C.3 Function DescriptionInput status Read Input Status (Alarm &Status Bits), Function 02

Up to 32 bits of Alarm are available as summarised in the table below:Table 2: Alarm and Status addresses

Addr. Type Description 0 Alarm Oxygen Alarm high 1 Alarm Oxygen Alarm low 2 Alarm Saturation Alarm high 3 Alarm Saturation Alarm low 4 Alarm - 5 Alarm - 6 Alarm Sample Temp. high 7 Alarm Sample Temp. low 8 Alarm Sample Flow high 9 Alarm Sample Flow low 10 Alarm Temp. shorted 11 Alarm Temp. disconnected 12 Alarm Case Temp. high 13 Alarm Case Temp. low 14 Alarm - 15 Alarm - 16 Alarm Control Timeout 17 Alarm Faraday Efficiency (QED) 18 Alarm - 19 Alarm - 20 Alarm - 21 Alarm - 22 Alarm 23 Alarm Input Active 24 Alarm IC MK41T56 25 Alarm IC LM75 26 Alarm IC PCF8574 27 Alarm EEprom Microcon 28 Alarm EEprom Motherboard 29 Alarm EEprom Front-end 30 Alarm Cal. Recout 31 Alarm Wrong Front-end 32 Alarm - 33 Alarm - 34 Alarm - 35 Alarm - 36 Alarm - 37 Alarm - 38 Alarm - 39 Alarm - 40 Status Electrolyte depleted 41 42 43 44 45 46 47

A-96.250.531 / 081128 61

Page 64: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

Input register Read Input Registers (Process data): Function 04The process data are made up of the measuring values of the AMI Powercon.

Table 3: Input registers

Single coilfunction

Force Single Coil: Function 05This function is used to set relays and reset the instrument. Relays can only be forced if set to fieldbus function. Table 4: single coil function

Byte-Offset

Data type Description Range of physical value

0 Float Oxygen 0.. 20’000 ppb 4 Float Temperature -30 … 120 ºC 8 Float Flow 0... 50 l/h

12 Float Saturation 0… 200 % 16 Float Air Pressure 500... 1200 hPa 20 Float Raw Oxgen temp. compensated 0... 50000 nA 24 Float Raw Oxgen uncompensated 0... 50000 nA 28 Float Raw Temp. (NT5k) 150 … 66’000 Ω 32 Float Raw Flow 0.. 100 Hz 36 Float Remaining Electrolyte 0.. 100 % 40 Float Case Temperature -55... 125 ºC 44 Float Signal Output 1 0.. 20 mA 48 Float Signal Output 2 0.. 20 mA 52 INSTRUMENT_STATUS Instrument Display Status

Addr. Description OFF (0) ON (1) 0.. 1 Relay Contacts 1..2 open close

2 Hold no yes

3 Remote Off no yes

15 Reset Instrument (like hardware reset) reset

62 A-96.250.531 / 081128

Page 65: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

Holdingregister

Read Holding Registers (Parameter data): Function 03Read Parameter data (see Parameter Table). All objects of type Record are defined in the Manufacturer specific tables following the parameter table. Note: The Address is used in a proprietary way to select the designated record.see parameter table on the next page.

Table 5: parameter table

Specific data Manufacturer-specific Data Types & RecordsDATE_STAMP Format

Address Modbus Addr Parameter Object Type Data Type Description Acc. 2300 1151 Version Record FIRMWARE_REC Firmware Version r 2400 1201 Num_Evt Simple Unsigned16 No. of Event Records r 2402 1202 Num_Log Simple Unsigned16 No. of Log Records r 3000 1501 Cal_History_1 Array Array of CAL_HIST_1[15] Cal. History Oxytrace r 3001 1502 Vfy_History_1 Array Array of VFY_HIST_1[15] Vfyl. History Oxytrace r 4000.. 4126 2001.. 2064 Event_Rec Record EVENT_REC Get Event Record

64 Records max. r

5000.. 8070 2501.. 4036 Log_Rec Record LOG_REC Get Logger Record

1536 Records max. r

31.. 26 25..22 21..17 16..12 11.. 6 5.. 0

YYYYYY MMMM DDDDD HHHHH MMMMMM SSSSSS

Year 0... 63

Month 1... 12

Date 1... 31

Hour 0... 23

Minutes 0... 59

Seconds 0... 59

A-96.250.531 / 081128 63

Page 66: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Modbus (option)

Parameter Size Data Type Description Range

INSTRUMENT_STATUS 32 Record Instrument Display Status

Language 1 Unsigned8 Language Selection

0 = German 1 = English 2 = French 3 = Spanish

RunMode 1 Unsigned8 Run Mode

0 = Run 1 = Hold 2 = Off 3 = Grab

AlaMode 1 Unsigned8 Alarm Mode 0 = Off 1 = on 2 = Blinking

MaintMode 1 Unsigned8 Maintenance Mode 0 = Off 1 = on 2 = Blinking

RelMode1 1 Unsigned8 Relais 1 Display Mode 0 = None 1 = Limit Hi On 2 = Limit Hi Off 3 = Limit Lo On 4 = Limit Lo Off 5 = Timer Off 6 = Timer On 7 = Percent Control8 = Valve 9 = Bus

RelMode2 1 Unsigned8 Relais 2 Display Mode 0 = None 1 = Limit Hi On 2 = Limit Hi Off 3 = Limit Lo On 4 = Limit Lo Off 5 = Timer Off 6 = Timer On 7 = Percent Control8 = Valve 9 = Bus

PercDos1 2 Unsigned16 Percent Dosage Relais 1 0.. 100 % PercDos2 2 Unsigned16 Percent Dosage Relais 2 0.. 100 % MaintValue 2 Unsigned16 Maintenance Value 0.. 100 % AlarmsActive 4 Unsigned32 Alarms Pending AlarmsActive AlarmsLatched 4 Unsigned32 Alarms no Acknowledged AlarmsLatched AlarmsActiveE 4 Unsigned32 Extended Alarms Pending AlarmsActiveE AlarmsLatchedE 4 Unsigned32 Extended Alarms not Acknowledged AlarmsLatchedE

TimeStamp 4 DATE_STAMP Current Time end Record

64 A-96.250.531 / 081128

Page 67: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

D. Program List and Explanations1 Messages1.1 Pending Errors

Provides the list of active errors. If an active error is acknowledged, the alarm relay opens again. Cleared errors are moved to the Message list.

1.2 Maintenance ListAs soon as the remaing amount of electrolyte is below 10% a maintenance of the oxygen sensor is recommended.

1.3 Message ListShows the error history: Error code, date and time of issue, and status (active, acknowledged, cleared).64 errors are memorized. Then the oldest error is cleared to save the newest error (circular buffer).

2 DiagnosticsIn diagnostics mode, the values can only be viewed, not modified.

2.1 IdentificationDesignation: instrument: e.g. AMI Oxytrace QEDVersion: Firmwareversion of instrument, e.g. V4.21-25/08

2.1.3 Factory test: Testdate of the Instrument-, Motherboard- and Frontend- QC factory test.

2.1.4 Operating time: years, days, hours, minutes, seconds

2.2 Sensors2.2.1 Oxytrace G

Current value: in ppbRaw value: raw value in nA, uncompensated.Saturation: Current saturation in %

2.2.1.4 Calibration history: Review the diagnostic values of the last calibration of the oxygen sensor. Only for diagnostic purpose. Max. 64 data records are memorized.

A-96.250.531 / 081128 65

Page 68: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

2.2.1.5 Verification History: Review the diagnostic values of the last faraday verification of the oxygen sensor. Only for diagnostic purpose. Max. 64 data records are memorized.

2.2.2 MiscellaneousCase Temperature: Actual temperature in °C inside the transmitter.Air pressure: Displaying of the current air pressure in hPa

2.3 Sample2.3.1 Sample ID: Review the programmed code. The code is defined by

the user to identify the sample point in the plant.Temperature: Current sample temperature in °C and additionally the resistance of the Nt5k temperature probe in Ohm.Sample Flow: Current sample flow in l/h and additionally the raw value of the flow meter in Hz.

2.4 I/O State2.4.1- 2.4.2 Reveals the actual status of all in and outputs:

Alarm relay: Open or closedRelay 1 and Relay 2: Open or closedInput: Open or closedSignal output 1, 2 (and 3, if option is installed): Actual current in mA.

2.5 InterfaceOnly available if optional interface is installed.Review programmed communication settings.

66 A-96.250.531 / 081128

Page 69: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

3 Maintenance3.1 Calibration

3.1.1 Start a calibration and follow the instructions on the screen. Displayed values are saturation in % and the saturation current in nA. The indication bar shows the progress.

3.2 ElektrolyteLast Filling: Date of the last filling.Remaining amount: Remaining amount of electrolyte in %.Remaining time: Remaining time in days until electrolyte exchange recommended.

3.2.5 New Filling: Select after electrolyte exchange to reset the counter.

3.3 Faraday VerificationStart a manual faraday verification. Displayed values are current value in ppb and the faraday concentration in %. The indication bar shows the progress.Current Value: in ppb.Faraday concentration: in %.Progress: bar shows the progress.

3.4 Simulation3.4.1-3.4.5 Select alarm relay, relay 1 /2 or signal output 1/2 to be tested with

the <Enter> key and press the up/down key to change the value. After confirming the setting with the Enter key, the value is simulated by the relay/signal output.At the absence of any key activities, the instrument will switch back to normal mode after 20 min.If you quit the menu, all simulated values will be reseted.

3.5 Set TimeAdjust date and time.

A-96.250.531 / 081128 67

Page 70: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

4 Operation4.1 Sensors

4.1.1 Filter time constant: Used to damp noisy signals. The higher the filter time constant, the slower the system reacts to changes of the measured value.Range: 5-300 sec

4.1.2 Hold after calibration: To allow the instrument to stabilize again after calibration. During calibration and hold time the signal outputs are frozen, alarms and limits are not active.Range: 0 - 6000 sec

4.1.3 Faraday Parameter

4.1.3.1 Timer Mode: Select interval, daily or weekly and set desired time/delay for automatic verfication or select off for manual faraday verification with menu 3.3 Faraday Verification, p. 67.

4.2 Relay Contacts 4.2.1 Alarm Relay: The alarm relay is used as cumulative error indicator.

Under normal operating conditions, the contact is open.The contact closes upon:

power lossdetection of system faults like defective sensors or electronic parts high case temperatureprocess values out of programmed ranges

See chapter 5 for a complete list and description of parameters. Program process value alarm levels here or in menu 5.3.1

4.2.2 and4.2.3

Relay 1 and 2 The function of relay contacts 1 or 2 are defined by the user:Start by defining the function of each relay in menu 5.3.2 and 5.3.3.These functions may be selected: limit switch, controller, timer or fieldbus driven.Then enter the necessary data in this menu or in the submenus of 5.3.2 / 5.3.3. The description of each function is given in chapter 5.

68 A-96.250.531 / 081128

Page 71: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

4.2.4 Input: The functions of the relays and signal outputs can be defined depending on the position of the input contact, i.e. no function, closed or open. See menu 5.3.4, p. 79 for parameter description

4.3 LoggerThe instrument is equipped with an internal logger. The data can be downloaded to a PC by Hyperterminal if option “HyperTerminal interface” is installed.The logger can save approx. 1500 data records. Records consists of: Date, time, alarms, measured value, measured value uncompensated, temperature, flow.

4.3.1 Log Interval: Select a convenient log interval. Consult the table below to estimate the max logging time. When the loggin buffer is full, the oldest data record is erased to make room for the newest one. (circular buffer)

4.3.2 Clear logger: If confirmed with yes, all data is erased and a new data series is started.

Interval 1 s 5 s 1 min 5 min 10 min 30 min 1 h

Time 25 min 2 h 25 h 5 d 10 d 31 d 62 d

A-96.250.531 / 081128 69

Page 72: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

5 Installation5.1 Sensors

5.1.1 Miscellaneous5.1.1.1 Flow: If a flow cell without flow measurement (e.g. B-Flow) is used,

choose none. With flow measurement select Q-Flow

5.1.1.2 O2 Offset: Manual, small correction of the offset. Range 0-3 ppb

5.2 Signal outputs5.2.1 - 5.2.2 Signal output 1, Signal output 2, Signal output 3:

Signal output 3 is only available if option 3rd signal output has been installed.Assign process value, the current loop range and a function to each signal output.

5.2.1.1 Parameter: Assign one of the process values to the signal output.Available values: Oxygen, Temperature, Sample flow and Saturation.

5.2.1.2 Current loop: Select the current range of the signal output. Make sure the connected device works with the same current range. Available ranges: 0-20 mA or 4-20 mA.

5.2.1.3 Function: Define if the signal output is used to transmit a process value or to drive a control unit. Available functions are:

linear, bilinear or logarithmic for process values.See As process values:, p. 71Control upwards or control downwoards for drive control.See As control output:, p. 72

70 A-96.250.531 / 081128

Page 73: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

As process values:The process value can be represented in 3 ways: linear, bilinear or logarithmic. See graphs below

.

5.2.1.40 Scaling: Enter beginning and end point of the linear or logarithmic scale. In addition, the midpoint for the bilinear scale.

0.0 0.1 0.2 0.3 0.4 0.5 0.6 unit

0-20 mA: 104-20 mA: 12

0/4

20

mA

meas. value

bilinear

linear

mA20

0-20 mA:104-20 mA: 12

0/41 10 100 1000 100002 3 6 30 value0 1 2 3 4 log

logarithmic

A-96.250.531 / 081128 71

Page 74: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

As control output:Signal outputs can be used for driving control units. We distinguish different kinds of controls:

P-controller: The controller action is proportional to the deviation from the setpoint. The controller is characterized by the P-Band. In the steady-state, the setpoint will never be reached. The deviation is called steady-state error.Parameters: setpoint, P-BandPI-controller: The combination of a P-controller with an I-controller will minimize the steady-state error. If the reset time is set to zero, the I-controller is switched off.Parameters: setpoint, P-Band, reset timePD-controller: The combination of a P-controller with a D-controller will minimize the response time to a fast change of the process value. If the derivative time is set to zero, the D-controller is switched off.Parameters: setpoint, P-Band, derivative time.PID-controller: The combination of a P-, an I- and a D-controller allows a proper control of the process.Parameters: setpoint, P-Band, reset time, derivative time

Ziegler-Nichols method for the optimization of a PID controller:

The point of intersection of the tangent with the respective axis will result in the parameters a and L .Consult the manual of the control unit for connecting and programming details. Choose control upwards or downwards.

parameters: setpointP-Band: Xp = 1.2 / areset time: Tn = 2 Lderivative time: Tv = L / 2

L

response on maximumcontrol output

time

tangent on the inflection

a

mea

sure

parameters: setpointP-Band: Xp = 1.2 / areset time: Tn = 2 Lderivative time: Tv = L / 2

L

response on maximumcontrol output

time

tangent on the inflection

a

mea

sure

LL

response on maximumcontrol outputresponse on maximumcontrol output

time

tangent on the inflectiontangent on the inflection

aa

mea

sure

72 A-96.250.531 / 081128

Page 75: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

Control upwards/downwards:

5.2.x.43.10 Setpoint: User-defined process value (Oxygen, Temperature, Sample flow or Saturation).

5.2.x.43.20 P-Band: Range below (upwards control) or above (downwards control) the set-point, within the dosing intensity is reduced from 100% to 0% to reach the set-point without overshooting.

5.2.X.43.3 Reset time: The reset time is the time till the step response of a single I-controller will reach the same value as it will be suddenly reached by a P-controller.Range: 0 - 9000 sec.

5.2.X.43.4 Derivative time: The derivative time is the time till the ramp response of a single P-controller will reach the same value as it will be suddenly reached by a D-controller.Range: 0 - 9000 sec.

5.2.X.43.5 Control timeout: If a controller action (dosing intensity) is constantly over 90% during a defined period of time and the process value does not come closer to the setpoint, the dosing process will be stopped to safety reasons. Range: 0 - 720 min.

A-96.250.531 / 081128 73

Page 76: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

5.3 Relay contacts5.3.1 Alarm relay

The alarm relay is used as cumulative error indicator.Under normal operating conditions the contact is open.The contact closes upon:

power lossdetection of system faults like defective sensors or electronic partshigh case temperatureprocess values out of programmed ranges.

Program alarm levels, hysteresis values and delay times for the following parameters:Oxygen, Sample Temp, Sample Flow, Saturation and Case temp.The alarm values of Oxygen and Saturation can also be programmed in menu 4.2.1.1 or 4.2.1.2

5.3.1.1 Alarm Oxygen:

5.3.1.1.1 Alarm high: If the measured value rises above the alarm high value, the alarm relay closes and E001 is displayed in the message list. Range: 0-10 ppm

5.3.1.1.23 Alarm low: If the measured value falls below the alarm low value, the alarm relay closes and E002 is displayed in the message list. Range: 0-10 ppm

5.3.1.1.33 Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relais contacts when the measured value fluctuates around the alarm value. Range. 0-10 ppm

5.3.1.1.43 Delay: During run time plus the delay time the signal and control outputs are held in the operating mode. Range: 0-28800 sec

5.3.1.2 Alarm Sample Flow:

5.3.1.2.1-5.3.1.2.36

Flow Alarm high (E009) and low (E010)Range alarm high: 10-50 l/hRange alarm low: 0.5-9 l/h

5.3.1.3 Alarm Sample Temperature:

5.3.1.3.1-5.3.1.3.23

Alarm Sample Temperature high (E007) and low (E008)Range alarm high: 30-100 °CRange alarm low: -10-20 °C

74 A-96.250.531 / 081128

Page 77: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

5.3.1.4 Alarm Saturation:

5.3.1.4.1 Alarm high: If the measured value rises above the alarm high value, the alarm relay closes and E003 is displayed in the message list. Range: 0-200 %

5.3.1.4.23 Alarm low: If the measured value falls below the alarm low value, the alarm relay closes and E003 is displayed in the message list. Range: 0-200 %

5.3.1.4.33 Hysteresis: Within the hyst. range, the relay does not switch. This prevents damage of relais contacts when the measured value fluctuates around the alarm value. Range. 0-200 %

5.3.1.4.43 Delay: During run time plus the delay time the signal and control outputs are held in the operating mode. Range: 0-28800 sec

5.3.1.4-5.3.1.5 Case temperature alarm high (E013) and low (E014)Range alarm high: 30 - 75 °CRange alarm low: -10 - 20 °C

A-96.250.531 / 081128 75

Page 78: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

5.3.2 - 5.3.3 Relay 1 and 2The function of relay contacts 1 or 2 are defined by the user:First select the functions as: limit switch, controller, timer or fieldbus driven.Then enter the necessary data depending on the selected function. The same values may also be entered in menu 4.2.2 and 4.2.3.

5.3.X.1 FunctionSelect:

Limit upper, Limit lower; see Limit upper/lower:, p. 76Control upwards, Control downwards,see Control upwards / downwards:, p. 76Timer; see Timer, p. 78Fieldbus; see Fieldbus, p. 78

Limit upper/lower:5.3.X.20 Parameter: Select Oxygen, Temperature, Sample flow or

Saturation

5.3.X.300 Set-point: User-defined process value (Oxygen, Temperature, Sample flow or Saturation).

5.3.X.400 Hysteresis: Within the hysteresis range, the relay does not switch. This prevents damage of relais contacts when the measured value fluctuates around the alarm value.

5.3.2.50 Delay: uring run time plus the delay time the signal and control outputs are held in the operating mode. Range: 0-28800 sec

Control upwards / downwards:The relays may be used to drive control units such as solenoid valves, membrane dosing pumps or motor valves. When driving a motor valve both relays are needed, relay 1 to open and relay 2 to close the valve.

5.3.X.22 Parameterselect the process parameter to be controlled (Oxygen, Temperature, Sample flow or Saturation)

5.3.X.32 SettingsChoose the respective actuator:Time proportionalExamples of metering devices that are driven time proportional are solenoid valves, peristaltic pumps.

76 A-96.250.531 / 081128

Page 79: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

Dosing is controlled by the operating timeParameters: Cycle time: duration of one control cycle (on/off change).Range: 10-600 sec.Response time: Minimal time the metering device needs to reactRange: 0-240 sec.

Frequency Examples of metering devices that are puls frequency driven are the classic membrane pumps with a potential free triggering input. Dosing is controlled by the repetition speed of dosing shotsParameters:Pulse frequency: Max. pulses per minute the device is able to respond to. Range: 20-300/min.

Motor valveDosing is controlled by the position of a motor driven mixing valve. Parameters: Run time: Time needed to open a completly closed valveRange: 5-300 sec.Neutral zone: Minimal response time in % of the of the runtime. If the requested dosing output is smaller than the response time, no change will take place. Range: 0-240 sec.

5.3.X.32.31 Control ParametersDefine the control parameters for the respective actuator

5.3.X.32.31.10 Set-point:User-defined process value (Oxygen, Temperature, Sample flow or Saturation)

5.3.X.32.31.20 P-Band: Range below (upwards control) or above (downwards control) the set-point, within the dosing intensity is reduced from 100% to 0% to reach the set-point without overshooting.

5.3.X.32.31.3 Reset time: The reset time is the time till the step response of a single I-controller will reach the same value as it will be suddenly reached by a P-controller.Range: 0-9000 sec.

5.3.X.32.31.4 Derivative time: The derivative time is the time till the ramp response of a single P-controller will reach the same value as it will be suddenly reached by a D-controller.Range: 0 - 9000 sec.

5.3.X.32.31.5 Control timeout: If a controller action (dosing intensity) is constantly over 90% during a defined period of time and the process value

A-96.250.531 / 081128 77

Page 80: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

does not come closer to the setpoint, the dosing process will be stopped to safety reasons. Range: 0 - 720 min.

Timer The relay will be closed repetitively depending on the programmed time scheme.This function is intended mainly to run Swan's cleaning modules or to program an automatic filter flush.

5.3.X.24 Operation mode: Select the operation mode (interval, daily, weekly) of the signal outputs.

5.3.X.340 Interval: Set interval time. range 1-1440 min.

5.3.X.44 Run Time: Time the relay stays closed. Range: 5-6000 sec.

5.3.X.54 Delay: during run time plus the delay time the signal and control outputs are held as programmed below .

5.3.X.6 Signal Outputs: Select the behavior of the signal outputs when the relay closes.Continuous: Signal output continues normally.Hold: Signal output continues on the last valid value.Off: Signal output is switched off and goes to 0/4 mA

4.2.2.6 Output / Control: Select the behavior of the control outputs when the relay closes.Continuous: Controller continues normally.Hold: Controller continues on the last valid value.Off: Controller is switched off.

FieldbusThe relay will be switched via the Profibus input. No further parameters are needed.See section Profibus (option), p. 48 for more details.

78 A-96.250.531 / 081128

Page 81: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)

5.3.4 Input5.3.4.1 Define when the input should be active:

No: Input is never activeWhen closed: Input is active if the input relay is closedWhen open: Input is active if the input relay is open

5.3.4.2 Signal output: Select the operation mode of the signal outputs when the relay is active (closed):Continuous: Signal outputs continue to issue the measured value.Hold: Signal outputs issue the last valid measured value.Off: set to 0 or 4 mA respectively.

5.3.4.3 Output/Control (relay or signal output): Continuous: Controller continues normally..Hold: Controller continues on the last valid value.Off: Controller is switched off.

5.3.4.4 Fault: No: No message is issued and the alarm relay does not close when input is active.Yes: Message E024 is issued and alarm relay closes when input is active.

5.3.4.5 Delay: Time which the instrument waits, after the input is deactivated,before returning to normal operation. Range: 0 - 6000 s

A-96.250.531 / 081128 79

Page 82: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

5.4 Miscellaneous5.4.1 Language: Set the desired language.

Available settings: Deutsch / English / Français / Español

5.4.2 Set default values: Reset the instrument to factory default values in three different ways:Calibration: Sets calibration values back to default. All other values are kept in memory.In parts: Communication parameters are kept in memory. All other values are set back to default values.Completely: Sets back all values including communication parameters.

5.4.3 Load firmware: Firmware updates should be done by instructed service personnel only.

5.4.4 Password: Select a password different from 0000 to prevent unauthorized access to the menus “Messages”, “Maintenance”, “Operation” and “Installation”. Each menu may be protected by a different password.If you forgot the passwords, contact the closest SWAN representative.

5.4.5 Sample ID: Identify the process value with any meaning full text, such as KKS number.

80 A-96.250.531 / 081128

Page 83: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Program List and Explanations

5.5 Interface5.5.1 Select one of the following communication protocols. Depending on

your selection, different parameters must be defined.Profibus: Device Address: 0 - 126

ID Number: Analyzer, Manufacturer, MultivariableLocal operation: enable, disablesee Profibus (option), p. 48 for details.

Modbus RTU: Device Address: 0 - 126Baud Rate: 2400 - 115200Parity:none, even, oddsee Modbus (option), p. 57 for details

HyperTerminal:Baud Rate: 2400 - 115200see separate RS232 Operation Manual for Logger download

Webserver: Device Address: 0-126see separate Webserver Manual.

A-96.250.531 / 081128 81

Page 84: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)MSDS Filling electrolyte

E. MSDS Filling electrolyte

1 Identification of the substance/preparation and of the company/undertaking

Identification of the product: Catalogue No.: 102751.10 Product name: Filling solution 1ALK for oxygen sensor

Use of substance/preparation: Sensor electrolyte, no public product, only for laboratory use

Manufacturer/supplier identification: Company: Swan Analytical Instruments AG, CH – 8340 Hinwil Tel.: +41 44 / 943 63 00, Fax: +41 44 / 943 63 01

2 Composition/information on ingredients

Synonyms: Sodium hydroxide solution

Hazardous ingredients: Sodium hydroxide solution CAS-No.: 1310-73-2 EC No: 215-185-5 EC Index No: 011-002-00-6 Classification: C; R35 Content: ~0.2 mol/l

3 Hazards identification

Irritating to eyes and skin.

4 First aid measures

After inhalation: fresh air After skin contact: wash off with plenty of water. Remove contaminated clothing. After eye contact: rinse out with plenty of water with the eyelid held wide open. Call in ophthalmologist. After swallowing: Immediately make victim drink plenty of water. Call in physician.

5 Fire-fighting measures

Suitable extinguishing media: In adaptation to materials stored in the immediate neighbourhood.

Special risks: non-combustible. Ambient fire may liberate hazardous vapours.

Special protective equipment for fire fighting: DO not stay in dangerous zone without self-contained breathing apparatus. In order to avoid contact with skin, keep safely distance and wear suitable protective clothing.

6 Accidental release measures: Person-related precautionary measures: Do not inhale vapours / aerosols. Avoid substance contact

Procedures for cleaning / absorption: Dilute with water and take-up with liquid-absorbent material.

82 A-96.250.531 / 081128

Page 85: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)MSDS Filling electrolyte

7 Handling and storage:

Handling: Can not be stored indefinitely. No further requirements.

Storage: Tightly closed. At +15°C to +25 °C.

Requirements for storage rooms / container: No aluminum, tin or zinc containers.

8 Exposure controls/personal protection

Personal protective equipment: Respiratory protection: Required when vapours / aerosols are generated. Eye protection: Required. Hand protection: Required

Industrial hygiene: Immediately change contaminated clothing. Apply skin-protective barrier cream. Wash hands and face after working with substance

9 Physical and chemical properties

Form: liquid Colour: colourless Odour: odourless

pH value (20°C): ~13

Melting point: not available Boiling point: near 100°C

Ignition temperature: not available Flash point: not available

Explosion limits: lower not available upper not available

Density (20°C) 1.01 g/cm3

Solubility in water (20°C): soluble

10 Stability and reactivity

Conditions to be avoided: --

Substances to be avoided: The constituents may react with: metals, light metals: Formed could be hydrogen (risk of explosion!), acids, nitriles, alkaline earth metals in powder form, ammonium compounds, cyanides, magnesium, organic nitro compounds, organic combustibles substances, phenols and oxidizable substances

Hazardous decomposition products: none

Further information: Unsuitable working materials: Various metals. Hydrogen may form upon contact with light metals

A-96.250.531 / 081128 83

Page 86: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)MSDS Filling electrolyte

11 Toxicological information

Acute toxicity: No toxicological data available

Further toxicological information: After inhalation of vapours / aerosols: Mucosal irritation After skin contact: Irritations After eye contact: Irritation After swallowing: Irritations of mucous membranes in the mouth, pharynx, oesophagus and gastrointestinal tract

12 Ecological information

Ecotoxic effects: --

Biological effects: Toxic effect on fish and plankton. Harmful effect due to pH shift. Does not cause biological oxygen deficit. Neutralization possible in waste water treatment plants

Further ecologic data: None

13 Disposal considerations

Product: There are no uniform EC regulations for the disposal of chemicals or residues. Chemical residues generally count as special waste. The disposal of the latter is regulated in the EC member countries through corresponding laws and regulations and in Switzerland also by the cantons. We recommend that you contact either the authorities in charge or approved waste disposal companies which will advise you on how to dispose of special waste.

Packaging: Disposal in compliance with official regulations. Handle contaminated packaging in the same way as the substance itself. If not officially specified differently, non-contaminated packaging may be treated like household waste or recycled.

14 Transport information

Road & Rail ADR, RID UN 1824 NATRIUMHYDROXIDLOESUNG, 8, III

Inland waterway ADN, ADNR (not tested)

Sea IMDG-Code UN 1824 SODIUM HYDROXIDE SOLUTION, 8, III Ems F-A S-B

Air CAO, PAX UN 1824 SODIUM HYDROXIDE SOLUTION, 8, III

84 A-96.250.531 / 081128

Page 87: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)MSDS Filling electrolyte

15 Regulatory information

Labelling according to EG Directives Symbol: Xi Irritant R-phrases: 36/38 Irritating to eyes and skin S-phrases: 26 In case of contact with eyes, rinse immediately with plenty of water and seek medical advice

16 Other information

Date of issue: 11.04.2008 The information contained herein is based on the present state of our knowledge. It characterizes the product with regard to the appropriate safety precautions. It does not represent a guarantee of the properties of the product.

A-96.250.531 / 081128 85

Page 88: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Default Values

F. Default ValuesOperation:

Sensors: Filter time constant: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 sHold after cal: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 300 sTimer mode faraday verification: . . . . . . . . . . . . . . . . . interval = 3h

Relay 1,Relay 2, Input:

same as in Installation

Log. Interval:Clear Logger

30 minno

Installation:

Sensors: Flow: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . NoneO2 Offset: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0 ppb

Signal Outputs: Parameter Output 1: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OxygenParameter Output 2: . . . . . . . . . . . . . . . . . . . . . . . . . . TemperatureCurrent loop: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20 mAFunction: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . linearScaling:Oxygen range low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0 ppbOxygen range high: . . . . . . . . . . . . . . . . . . . . . . . . . . . .10.00 ppmTemperature range low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0 °CTemperature range high: . . . . . . . . . . . . . . . . . . . . . . . . . . . 50.0 °CFlow range low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 .0 l/hFlow range high: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50.0 l/hSaturation range low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.0 %Saturation range high: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 %

Alarm Relay: Alarm Oxygen high: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 ppmAlarm Oxygen low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.0 ppbHysteresis: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100 ppbDelay: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 sFlow alarm: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . yesFlow alarm high: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20.0 l/hFlow alarm low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 l/hSample temp. alarm high: . . . . . . . . . . . . . . . . . . . . . . . . . . .100 °CSample temp. alarm low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0 °CSaturation alarm high: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 %Saturation alarm low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0.0 %Hysteresis: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 %Delay: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 s

86 A-96.250.531 / 081128

Page 89: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Default Values

Case temp. high: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 °CCase temp. low: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 °C

Relay1/2 Function: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . limit upperParameter: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OxygenSetpoint: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 ppmHysteresis: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100.0 ppbDelay: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 sControl: Actuator: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FrequencyControl: Frequency: Frequency: . . . . . . . . . . . . . . . . . . . . . 120/minControl: Time prop.: Cycle time: . . . . . . . 60s, Response time: 10 sControl: Motor valve: Run time: . . . . . . . . . .60s, Neutral zone: 5%Control: Control parameter Setpoint: . . . . . . . . . . . . . . . . . 10 ppmControl: Control parameter P-band: . . . . . . . . . . . . . . . . 100.0 ppbControl: Control parameter Reset time: . . . . . . . . . . . . . . . . . . . 0 sControl: Control parameter Derivative time: . . . . . . . . . . . . . . . . 0 sControl: Control parameter Control time-out: . . . . . . . . . . . . . 0 minTimer: Mode: Interval: Timer: Interval:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 minTimer: Mode: daily/weekly: . . . . . . . . . . . . . Starting time: 00.00.00 Timer: Run time: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10sTimer: Delay: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5sTimer: Signal output: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . cont.Timer: Output/Control: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . cont.

Input Active: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . when closedSignal Outputs: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . holdOutput/Control: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . offFault: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . noDelay: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 s

Miscellaneous Language: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . EnglishSet default: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . noLoad firmware: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . noPassword: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . for all modes 0000Sample ID: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . --

Interface Protocol: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HyperterminalBaud Rate: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115’200 Baud

A-96.250.531 / 081128 87

Page 90: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Index

G. IndexAAlarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 68Attention Symbols . . . . . . . . . . . . . . . . . . . . . . . . . 7

BB-Flow SS316L OTG . . . . . . . . . . . . . . . . . . . . . . 18

CCalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Calibration verification . . . . . . . . . . . . . . . . . . . . . 40Cyclic Data Telegram . . . . . . . . . . . . . . . . . . . . . . 49

DData Communication, acyclic . . . . . . . . . . . . . . . . 56Data Communication, cyclic . . . . . . . . . . . . . . . . . 49Data logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . 86Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74, 75Description of the system . . . . . . . . . . . . . . . . . . . 8Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

EElectrical Connections . . . . . . . . . . . . . . . . . . . . . 25Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43Error list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

FFaraday Verification . . . . . . . . . . . . . . . . . . . . . . . 41Filter time constant . . . . . . . . . . . . . . . . . . . . . . . . 68

GGSD File Types . . . . . . . . . . . . . . . . . . . . . . . . . . 55

HHold after calibration . . . . . . . . . . . . . . . . . . . . . . . 40Hysteresis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74, 75

88 A-96.250.531 / 081128

Page 91: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Index

IInput . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 30, 69Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Installation Checklist . . . . . . . . . . . . . . . . . . . . . . . . 35Instrument Specification . . . . . . . . . . . . . . . . . . . . . 10Interface RS232 . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Interface RS485 . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

MMaintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35, 67Maintenance of Oxygen Sensor . . . . . . . . . . . . . . . 36Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . 35Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . 8Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 15Modbus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57MSDS Filling electrolyte . . . . . . . . . . . . . . . . . . . . . 82

OOn-line operation . . . . . . . . . . . . . . . . . . . . . . . . . . 8On-site Requirements . . . . . . . . . . . . . . . . . . . . . . 10Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

PPower Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29Profibus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48Program List and Explanations . . . . . . . . . . . . . . . 65Programm Overview . . . . . . . . . . . . . . . . . . . . . . . . 45

QQV-Flow PMMA OTG . . . . . . . . . . . . . . . . . . . . . . . 16

RRelay Contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10, 68

SSample Requirements . . . . . . . . . . . . . . . . . . . . . . 10Sensor Connections . . . . . . . . . . . . . . . . . . . . . . . . 34Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . .11, 32, 70, 74Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . 15

A-96.250.531 / 081128 89

Page 92: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Index

TTerminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

WWarning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . 7Warning Symbols . . . . . . . . . . . . . . . . . . . . . . . . . 7

ZZero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . 40

90 A-96.250.531 / 081128

Page 93: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Index

A-96.250.531 / 081128 91

Page 94: MenAMI Oxytrace 01 - ТЕХНОАНАЛИТ

AMI Oxytrace (QED)Index

SWANis represented worldwide by subsidiary companiesand distribuitors.

cooperates with independent representativesall over the world.

SWAN ProductsAnalytical Instruments for:

High Purity Water

Feedwater, Steam and Condensate

Potable Water

Pool and Sanitary Water

Cooling Water

Waster Water and Effluents

Made in Switzerland.

92 A-96.250.531 / 081128