microrheological coagulation assay exploiting ...€¦ · microrheological coagulation assay...

15
Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duy, and Martin Hegner* CRANN, School of Physics, Trinity College Dublin, Dublin 2, Ireland * S Supporting Information ABSTRACT: Rheological measurements in biological liquids yield insights into homeostasis and provide information on important molecular processes that aect uidity. We present a fully automated cantilever-based method for highly precise and sensitive measurements of microliter sample volumes of human blood plasma coagulation (0.009 cP for viscosity range 0.53 cP and 0.0012 g/cm 3 for density range 0.91.1 g/cm 3 ). Microcantilever arrays are driven by a piezoelectric element, and resonance frequencies and quality factors of sensors that change over time are evaluated. A highly accurate approximation of the hydrodynamic function is introduced that correlates resonance frequency and quality factor of cantilever beams immersed in a uid to the viscosity and density of that uid. The theoretical model was validated using glycerol reference solutions. We present a surface functionalization protocol that allows minimization of unspecic protein adsorption onto cantilevers. Adsorption leads to measurement distortions and incorrect estimation of the uid parameters (viscosity and density). Two hydrophilic terminated self-assembled monolayers (SAMs) sensor surfaces are compared to a hydrophobic terminated SAM coating. As expected, the hydrophobic modied surfaces induced the highest mass adsorption and could promote conformational changes of the proteins and subsequent abnormal biological activity. Finally, the activated partial thromboplastin time (aPTT) coagulation assay was performed, and the viscosity, density, and coagulation rate of human blood plasma were measured along with the standard coagulation time. The method could extend and improve current coagulation testing. V iscosity and density measurements of biological uids such as whole blood and blood plasma provide important insights on biological processes that regulate health and disease. 13 In particular during blood coagulation the viscosity of plasma increases due to brinogen polymerization. Blood coagulation is the result of a complex series of biological reactions that leads to blood clot formation. The coagulation process is unique, but there are two distinctions: Hemostasis and Thrombosis. Hemostasis is coagulation that occurs in a physiological setting and results in sealing a break in the circulatory system. Thrombosis is coagulation occurring in a pathological context that leads to localized intravascular clotting and potentially to an occlusion of a vessel or embolus formation. 4 Central to the coagulation cascade is the enzyme thrombin. Conversion of brinogen into brin by thrombin creates a lamentous protein network. Two commonly used tests to evaluate coagulation disorders are the aPTT and the prothrombin time (PT). The PT test and the aPTT test evaluate the extrinsic and intrinsic pathways of the coagulation cascade, respectively. In clinical settings either mechanical or optical systems measure international normalized ratio (INR) levels, which are the standard unit for reporting the end point of clotting time. 5,6 In both clinical tests the measured variable is blood clotting time; thus, blood clot physical properties such as density and viscosity are not measured. There are dierent techniques 7,8 to measure viscosity and density of a solution, but they require milliliter samples. Many of these techniques are time-consuming and not suitable for medical environments. Other surface analytical assays have been used to measure coagulation parameters such as quartz crystal microbalance (QCM), 911 surface plasmon resonance (SPR), 10 and bulk acoustic wave sensors. 12 Mostly these methods are single-sensor techniques and do not separate the eects of the mass adsorption onto the sensor from solution bulk viscosity and density. Recent work by Cakmak et al. 13 shows a microelectromechanical system (MEMS) technique for measurements of aPTT and PT. The method presented did not evaluate physical properties of the blood clot and did not include a functionalization procedure that can minimize unspecic protein adsorption onto the measurement sensors. We believe that there is a demand for assays that can reliably track viscosity and density changes during plasma coagulation in addition to the coagulation time. This information could improve coagulation tests that are conducted presurgery and for anticoagulant therapy by providing absolute physical parame- ters to the user. Abnormal blood viscosity has been reported to Received: August 25, 2016 Accepted: November 29, 2016 Published: November 29, 2016 Article pubs.acs.org/ac © 2016 American Chemical Society 751 DOI: 10.1021/acs.analchem.6b03347 Anal. Chem. 2017, 89, 751758

Upload: others

Post on 13-Aug-2020

21 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

Microrheological Coagulation Assay Exploiting MicromechanicalResonatorsFrancesco Padovani, James Duffy, and Martin Hegner*

CRANN, School of Physics, Trinity College Dublin, Dublin 2, Ireland

*S Supporting Information

ABSTRACT: Rheological measurements in biological liquids yieldinsights into homeostasis and provide information on importantmolecular processes that affect fluidity. We present a fully automatedcantilever-based method for highly precise and sensitive measurementsof microliter sample volumes of human blood plasma coagulation (0.009cP for viscosity range 0.5−3 cP and 0.0012 g/cm3 for density range 0.9−1.1 g/cm3). Microcantilever arrays are driven by a piezoelectric element,and resonance frequencies and quality factors of sensors that change overtime are evaluated. A highly accurate approximation of the hydrodynamicfunction is introduced that correlates resonance frequency and qualityfactor of cantilever beams immersed in a fluid to the viscosity and densityof that fluid. The theoretical model was validated using glycerol referencesolutions. We present a surface functionalization protocol that allowsminimization of unspecific protein adsorption onto cantilevers. Adsorption leads to measurement distortions and incorrectestimation of the fluid parameters (viscosity and density). Two hydrophilic terminated self-assembled monolayers (SAMs)sensor surfaces are compared to a hydrophobic terminated SAM coating. As expected, the hydrophobic modified surfacesinduced the highest mass adsorption and could promote conformational changes of the proteins and subsequent abnormalbiological activity. Finally, the activated partial thromboplastin time (aPTT) coagulation assay was performed, and the viscosity,density, and coagulation rate of human blood plasma were measured along with the standard coagulation time. The methodcould extend and improve current coagulation testing.

Viscosity and density measurements of biological fluids suchas whole blood and blood plasma provide important

insights on biological processes that regulate health anddisease.1−3 In particular during blood coagulation the viscosityof plasma increases due to fibrinogen polymerization. Bloodcoagulation is the result of a complex series of biologicalreactions that leads to blood clot formation. The coagulationprocess is unique, but there are two distinctions: Hemostasis andThrombosis. Hemostasis is coagulation that occurs in aphysiological setting and results in sealing a break in thecirculatory system. Thrombosis is coagulation occurring in apathological context that leads to localized intravascular clottingand potentially to an occlusion of a vessel or embolusformation.4 Central to the coagulation cascade is the enzymethrombin. Conversion of fibrinogen into fibrin by thrombincreates a filamentous protein network. Two commonly usedtests to evaluate coagulation disorders are the aPTT and theprothrombin time (PT). The PT test and the aPTT testevaluate the extrinsic and intrinsic pathways of the coagulationcascade, respectively. In clinical settings either mechanical oroptical systems measure international normalized ratio (INR)levels, which are the standard unit for reporting the end pointof clotting time.5,6 In both clinical tests the measured variable isblood clotting time; thus, blood clot physical properties such asdensity and viscosity are not measured.

There are different techniques7,8 to measure viscosity anddensity of a solution, but they require milliliter samples. Manyof these techniques are time-consuming and not suitable formedical environments. Other surface analytical assays havebeen used to measure coagulation parameters such as quartzcrystal microbalance (QCM),9−11 surface plasmon resonance(SPR),10 and bulk acoustic wave sensors.12 Mostly thesemethods are single-sensor techniques and do not separate theeffects of the mass adsorption onto the sensor from solutionbulk viscosity and density. Recent work by Cakmak et al.13

shows a microelectromechanical system (MEMS) technique formeasurements of aPTT and PT. The method presented did notevaluate physical properties of the blood clot and did notinclude a functionalization procedure that can minimizeunspecific protein adsorption onto the measurement sensors.We believe that there is a demand for assays that can reliablytrack viscosity and density changes during plasma coagulationin addition to the coagulation time. This information couldimprove coagulation tests that are conducted presurgery and foranticoagulant therapy by providing absolute physical parame-ters to the user. Abnormal blood viscosity has been reported to

Received: August 25, 2016Accepted: November 29, 2016Published: November 29, 2016

Article

pubs.acs.org/ac

© 2016 American Chemical Society 751 DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

Page 2: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

be involved in long-standing retinal vein occlusion,14,15 cerebralthromboembolic events,16 and ischemic heart disease.17 Despitethis evidence, blood viscosity is a parameter that is notcommonly evaluated. Currently each lab has to run its ownstandardization, and therefore certain variability within theclinical analytics exists.Cantilever sensors in dynamic mode proved to be an

excellent candidate for rheological measurements of liquidsusing small volumes.18−23 We use an array of 8 microcantileversto measure viscosity and density changes in real time duringhuman blood plasma coagulation. This method also providesthe aPTT measurement. An approximation to the exactanalytical solution24 of the hydrodynamic load is presented,and the measurement device has been calibrated using glycerolreference solutions. Utilizing the availability of in situdifferential analysis, we have established a functionalizationprocedure for the sensor surfaces that minimizes the massadsorption that is inevitable when the sensor is in contact withblood plasma proteins.25,26 Mass adsorption onto the hydro-phobic surfaces11,27,28 induces conformational changes ofproteins, which result in subsequent abnormal activity measure-ments. It also leads to erroneous calculations of the physicalproperties of the liquid surrounding the vibrating structure.Finally, we evaluated the performance of the platform using astandardized aPTT test. Along with the coagulation time, threeadditional parameters could be extracted: density, viscosity, andcoagulation rate. The presented assay allows coagulation testsfollowing clinical guidelines and the measured aPTT (∼47 s)was within the clinical range. The volume achieved for the testsis 20 μL. The measurement chamber has a volume of 4 μL;therefore, the required total volume of plasma could be furtherreduced with microfluidic optimization.

■ MATERIALS AND METHODSUnless otherwise noted, chemicals were purchased from Sigma-Aldrich, Ireland. Glycerol solutions ranging between 5% and40% have been prepared by mixing glycerol (>99%) withnanopure water. Three heterobifunctional organochemicalcompounds that form SAMs on gold surfaces have been usedin the experiments: 1-octadecanethiol for monolayers exposinga hydrophobic (CH3) terminus; (11-mercaptoundecyl)tetra-(ethylene glycol) for a hydrophilic, neutral terminus (PEG);and 11-mercapto-1-undecanol for a hydrophilic, neutralterminus (OH). Human serum albumin (HSA, lyophilizedpowder) was reconstituted in 1 mM HEPES (≥99.5%,) bufferat pH 6.8. Thrombin (from human plasma, lyophilized powder)was reconstituted in 0.1% HSA solution to a concentration of350 NIH units/mL and diluted down to 70 NIH units/mLwhen needed. In all of the experiments, thrombin concentrationhas been kept constant at a physiological concentration of 70NIH units/mL. Fibrinogen (lyophilized powder, from humanplasma) was reconstituted with 0.9% NaCl in 1 mM HEPESbuffer, pH 6.8, at a final concentration of 10 mg/mL anddiluted down to 2.5 mg/mL when needed. Human bloodplasma controls and aPTT kits (HemoSIL Normal ControlAssayed and HemoSIL aPTT-SP liquid, InstrumentationLaboratory, U.S.A.) were purchased from Brennan & Co,Dublin, Ireland. Human blood plasma was reconstituted in 1mL of nanopure water and gently swirled for 2 min to ensurefull reconstitution. The aPTT kit contains a 25 mM calciumchloride (CaCl2) solution and a colloidal silica dispersion withsynthetic phospholipids, buffer, and preservatives (ready touse).

Experimental Setup. For details about the fully automizedexperimental setup, we refer the reader to ref 29. Briefly, anarray of 8 cantilevers is mounted into a 4 μL measurementchamber that allows liquid exchange through 4 syringe pumpsand a microdispensing valve. The cantilevers were oscillated bya piezoelectric stack mounted underneath the array, and theirvibration is recorded using an optical beam detection system.The amplitude and phase spectra of the oscillations wereevaluated from the recorded data using a custom LabVIEWsoftware interface. Upgrades on the previously presenteddevice29 include the following: (1) digitizer with higherresolution (National Instruments PCI-5105, 60 MS/s, 12-bit)to better resolve individual voltage levels of the positionsensitive detector (PSD) response, (2) expanded and improvedtemperature-controlled enclosure, and (3) four remotelycontrolled syringe pumps housed inside the temperature-controlled enclosure. The normalized differential signal of thePSD is acquired at 10 MS/s, and each driving signal frequencyis applied for 1 ms. For detailed information on resonancefrequency and quality factor evaluation, see SupportingInformation.

Theoretical Background. To evaluate both density andviscosity of the liquid surrounding the vibrating cantilever, bothquality factor and resonance frequency were experimentallyevaluated. The resonance frequency f R,n and the quality factorQn (at mode number n) are correlated to the hydrodynamicload24 as follows:

βκ

= ·+ Γ

f Ck

m Re m( , )n nn

R, cal,fc r

fl (1)

κκ

=+ ΓΓ

Q Cm Re m

Re m( , )

( , )n Qn

ncal,

c rf

l

if

l (2)

where βn = Cn2/2π 3 with Cn = 1.875, 4.694, 7.854, 11, ..., π(n

− 0.5);30 k = 3EI/l3 is the cantilever spring constant, which iscalculated from the Young’s modulus E and the length l of thecantilever; while I = bh3/12 is the moment of inertia of arectangular beam with b and h, respectively, being the widthand thickness of the cantilever beam. The parameters mc and mlare, respectively, the mass of the cantilever and a “virtual mass”of inviscid liquid comoved by the cantilever when κn = 0.24

Both masses are evaluated from the density of the cantilever ρcand the density of the liquid ρl as mc = ρcbhl and ml = πρlb

2l/4.The Reynold’s number Re = 2πfρlb

2/η expresses theimportance of inertial forces in the liquid relative to viscousforces (η is the viscosity of the liquid). The calibration factorsCcal,f and Ccal,Q are introduced to account for internal losses inthe cantilever structure and uncertainties in the spring constantof the cantilever. These factors are determined through acalibration step where a liquid with known viscosity and densityis introduced into the measurement chamber. In all theexperiments presented, we indicate the specific density andviscosity for the calibration step. The hydrodynamic load isexpressed through the hydrodynamic function Γf(Re,κn) =Γrf(Re,κn) + iΓi

f(Re,κn) where the superscript f refers to theflexural mode while r and i refer to real and imaginary parts,respectively (i = −1 ). The hydrodynamic function dependson the Reynold’s number and on the normalized mode numberκn = Cnb/l. We introduce an approximation to the analyticalsolution of the hydrodynamic function24 that is valid over a

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

752

Page 3: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

large range of Reynold’s numbers Re = [10−4, 105] and over alarge range of normalized mode numbers κn = [0, 20]

Γ = ++ ·

Re AC G

( )1

( 10 )xrf

1.68 1/4(3)

Γ = − + + +Re( ) 10 B x a x a x xif ( [( )(sgn( ) 1)])H H H2

23

3

(4)

where x = log10Re and xH = x − H. Equations 3 (modifiedlogistic function) and 4 (10 to a third order polynomial) havebeen fitted to the analytical solution and the fitting parametersA, C, G, B, a2, a3, and H have been evaluated in the range κn =[0, 20] through nonlinear least-squares fitting. Tabulated valuesof the fitting parameters are reported in Table S1 (SupportingInformation). Note that both functions satisfy the requirementsin an inviscid fluid, that is, Re → +∞. In particular, theimaginary part tends to 0 as long as a3 < 0, while the real parttends to A. Therefore, A was calculated using the inviscidtheory,32 and a3 was kept negative throughout the fittingprocess. The maximum error in the approximation is 0.6%;thus, it can be confidently used within the range Re = [10−4,105]. Figure 1 shows the calculated and approximated values forboth the real and imaginary parts of the hydrodynamic functionfor different κn values. The exact analytical solution of thehydrodynamic function24 was computed numerically using

Mathematica 10, and the values obtained converged at the sixthdigit. Convergence of the solution requires a square matrix sizeof M = 52, as discussed in ref 33. Substituting eqs 3 and 4 intoeqs 1 and 2, we obtain a dependency of the resonancefrequency f R,n and the quality factor Qn on the Reynold’snumber. Expressing the Reynold’s number as a function of theliquid’s density ρl and viscosity η, we numerically computedboth liquid parameters.

Cantilever Preparation. We used arrays consisting of 8microcantilevers (IBM Zurich Research Laboratory, Switzer-land) with length, width, and thickness of 500, 100, and 1 μm,respectively. They were cleaned with subsequent immersions inthree different solvents: (1) 10 min acetone (CHROMA-SOLV), (2) 5 min ethanol (CHROMASOLV), (3) 1 minnanopure water, and (4) 5 min acetone. The solvent cleaningwas followed by plasma cleaning (Diener Electronic GmbHPICO Barrel Asher, 0.3 mbar oxygen (O2), 160 W, 40 kHz for3 min) and a final immersion in ethanol. The arrays were thencoated with 3 nm of titanium (deposition rate 0.2 Å/s) on bothsides and with 23 and 33 nm of gold (deposition rate 0.5 Å/s)on the top and bottom sides, respectively, using electron beamevaporator (Temescal FC-2000, U.S.A.). After metal coating,the arrays were stored under argon to prevent chemicaldegradation of the gold layers. Before use the array was cleanedagain with the above solvent cleaning sequence followed by aUV cleaning step to activate the gold layers, and it was thenfunctionalized with 1 mM solutions (in ethanol) of differentSAMs for 60 min using a capillary technique.34 The aboveconcentration and incubation time provide optimal coverageand ensure good quality of the SAM.35 Finally the array isrinsed with ethanol and stored in 1 mM HEPES buffer untiluse.

Glycerol Solution Experiments. Reference solutions wereprepared by mixing different concentrations of glycerol (v/v)with nanopure water. The concentration range was 5−40%,which corresponds to a viscosity range of 1−3 cP and a densityrange of 1−1.1 g/cm3. Table S2 (Supporting Information)reports the expected viscosities and densities of solutions withdifferent concentrations of glycerol.31 These ranges cover theviscosity and density range of many biological fluids, includinghuman blood plasma.36 The temperature for these experimentswas kept constant at 23 ± 0.01 °C. The array was firstimmersed in water (viscosity 0.9532 cP, density 0.9977 g/cm3),and then the data acquisition began. Then 500 μL of referencesolutions were flushed automatically through the chamber at100 μL/min. The flow was then stopped, and for each solution∼100 spectra (amplitude and phase) were acquired percantilever in 20 min. The average resonance frequency andquality factor were then used for the computation of theviscosity and density of each solution.

SAMs Performance Experiments. The array of 8 sensorswas selectively functionalized; three cantilevers were function-alized with OH-terminated SAM, three with PEG-terminatedSAM, and two with CH3-terminated SAM. PEG and OH SAMshave been reported to resist protein adsorption,37,38 while CH3provides a hydrophobic surface that increases adsorption ofmost of the proteins.39,40 A selective functionalization allowsthe evaluation of different SAMs’ resistance to proteinadsorption using the same array. The array was stabilized at37 °C in water (viscosity 0.6913 cP, density 0.9933 g/cm3; seeref 41) for 10 min; then 100 μL of 1 mM HEPES buffer (pH6.8) was flushed at 10 μL/min and left in the chamber for 15min. Note that these experiments were performed at higher

Figure 1. Hydrodynamic function plot for different Reynold’snumbers and different normalized mode numbers κ. Plot (a) is forthe real part and (b) is for the imaginary part. For the imaginary part,only 3 normalized mode numbers are plotted to enhance visibility.Markers are values calculated from the exact analytical solutionaccording to literature,24 while the lines are the approximation valuescalculated from eqs 3 and 4. The accuracy of the approximation iswithin 0.6%.

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

753

Page 4: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

temperature than previous experiments; therefore, the viscosityand density of water is lower. Subsequently, 500 μL offibrinogen was flushed in at 50 μL/min, and finally, afteranother 10 min, 25 μL of fibrinogen and thrombin solution wasinjected at 250 μL/min. The relatively high injection speed isrequired for a fast exchange (6 s) of the liquid; otherwise, thereaction between thrombin and fibrinogen would occur beforeentering the measurement chamber. The mixing occurredautomatically 10 μL before the chamber, and this volume has tobe pushed 15 μL further to enter the chamber (see design in ref29). To ensure optimal mixing of the solution, differentvolumes and injection speeds were tested with two differentcolored solutions, and the final color (a mix of the two) wasrecorded and analyzed with image-analysis software (see videoin ref 29). During the whole experiment, amplitude and phasespectra were recorded every ∼20 s.Human Blood Plasma Coagulation Experiments. To

study the coagulation of human blood plasma, we followed thestandard aPTT clinical protocol. In an aPTT test citric acid isadded to the blood plasma sample to prevent spontaneouscoagulation. The lyophilized human blood plasma alreadycontains citric acid. Reconstitution is ensured by gentle swirlingfor 2 min after addition of 1 mL of nanopure water. Next a 25mM calcium chloride solution (CaCl2) is added to the bloodplasma together with a phospholipid suspension that containssilica particles. The excess of calcium is required to counteractthe anticoagulant activity of citric acid, while the silica(negatively charged surface) acts as an activator, inducingcontact activation.42 To perform the procedure described aboveand to adhere to clinical timing guidelines, we have followedand automated the same aPTT protocol. The cantilever arraywas stabilized in 1 mM HEPES buffer at 37 °C. After 30 min,100 μL of human blood plasma was injected into the chamberat 10 μL/min. After another 20 min a 25 μL mixture ofcolloidal silica solution, CaCl2 solution, and human bloodplasma was injected simultaneously into the chamber at 250μL/min. The relatively high speed is required to ensure propermixing of the solutions. The reaction mix is fully exchanged inthe chamber after ∼6 s, allowing a real-time measurement ofthe biochemical reactions. Throughout the whole experimentamplitude and phase spectra were recorded every ∼3 s.

■ RESULTS AND DISCUSSIONA microrheological measurement of the viscosity and density ofhuman blood plasma during coagulation using a microcanti-lever array approach is presented. The method enables smallvolumes (20 μL) to be used and achieves high sensitivity (0.009cP for the viscosity and 0.0012 g/cm3 for the density, 3 sigmas)along with a full characterization of biologically relevant fluidsas long as they behave as Newtonian fluids. Three differenttypes of experiments are presented: (1) the validation of thetheoretical model using solutions with known density andviscosity, (2) fibrinogen polymerization triggered by thrombin,and (3) a clinical aPTT coagulation test.Theoretical Model Validation Using Glycerol Refer-

ence Solutions. To test the validity of the theoretical modelfor the calculation of density and viscosity of a liquid, we usedsolutions with different concentrations of glycerol. Figure 2shows a comparison between the expected values and themeasured values of both density and viscosity. The accuracy ofthe measurement is very much dependent on the quality factor.The SHO model is valid for Qn ≫ 1.24 Liquid viscosity isdirectly proportional to dissipative effects. With increasing

dissipative effects (hydrodynamic function imaginary part), thequality factor decreases. Deviations from reference values are inthe range 0.2−11% for the viscosity and 0.1−2.2% for thedensity. To take into account these deviations that depend onthe quality factor, we have used the results (Figure 2) as acalibration curve for the measurements involving fibrinogenpolymerization and blood plasma coagulation. Other workevaluated the rheology of liquids21,22 using cantilevers vibratingat the first two oscillation modes. Such an approach can begreatly enhanced when measuring at higher oscillationmodes.19,24 Driving beams clamped at one end directly with apiezo electric element underneath the array at higher modeslead to an increase in the quality factor from ∼2 at first mode to∼25 at the 10th mode. While increasing the percentage ofglycerol from 0% to 40% (with steps; see Figure 2), the qualityfactor drops accordingly from the highest value of 24.26 (0%)to the lowest value of 13.5 (40% glycerol). The measuredvalues of quality factors still surpass the previously reportedvalues31 by 14-fold.

Fibrinogen Polymerization: Protein Adsorption onDifferent SAMs Surfaces. Fibrinogen polymerization istriggered by the protein thrombin. When thrombin andfibrinogen are mixed, the latter is cleaved into fibrin, whichcreates a polymer-like structure, and the liquid viscosityincreases. Cantilever-based measurements require the sensorsto be fully immersed in the solution. Fibrinogen can bespontaneously adsorbed onto gold surfaces,43,44 increasing the

Figure 2. Experimental (red dots) and reference (black squares)viscosity and density values for different glycerol concentrations (%, v/v) as a function of resonance frequency and quality factor (mode 10,average of 8 cantilevers) at 23 °C. The discrepancy between referenceand experimental values is in the range 0.2−11% for the viscosity and0.1−2.2% for the density. The best accuracy is reached with higherquality factors because the simple harmonic oscillator model used toevaluate resonance frequency and quality factor is valid for Qn ≫ 1.The accuracy in the range 0.9−1.4 cP is >94%. The signal-to-noise(SNR) ratio of the device is high enough to clearly evaluate viscosityand density of 1% change in the glycerol concentration (differencebetween 14% and 15% glycerol solutions).

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

754

Page 5: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

cantilever’s overall mass. Unspecific mass adsorption should beminimized; otherwise, the convolution of the individualphysical properties (mass increase versus bulk fluid propertychanges) would lead to an incorrect estimation of the viscosityand density properties. Furthermore, for subsequent testing ofbiological fluids, it is necessary that the sensor is renderedbiocompatible. The performances of three different SAMs andnonfunctionalized gold-coated surfaces (called simply gold)have been tested in terms of relative protein adsorption. Wehave investigated two SAMs that prevent protein adsorption(PEG- and OH-terminated SAMs) and a hydrophobicterminated SAM (CH3). Figure 3 shows resonance frequencyshift, quality factor, and viscosity and density change (aftercalibration) for the three different SAMs and gold duringfibrinogen polymerization. After initial stabilization in water, 1mM HEPES buffer is flushed through the chamber followed byan injection of fibrinogen (2.5 mg/mL). Due to the differentdensities of buffer and the fibrinogen solution compared towater, the resonance frequency decreases while the qualityfactor remains constant. The buffer used for fibrinogenpolymerization and human blood plasma coagulation experi-ments was always 1 mM HEPES buffer at pH 6.8. We havemeasured the density and viscosity of the buffer, and we haveused it as calibration solution in the following experiments forthe calculation of the calibration parameters Ccal,f and Ccal,Q (seeeqs 1 and 2). The calculated density and viscosity of 1 mMHEPES buffer at 37 °C were 0.9949 ± 0.0004 g/cm3 and 0.692± 0.003 cP, respectively. After another 10 min, thrombin (70NIH units/mL) and fibrinogen solutions are mixed in thechamber. Thrombin causes fibrinogen to polymerize into afibrin network, and the overall viscosity of the solution

increases. By calculating density and viscosity of thepolymerized fibrin solution, we obtained different density andviscosity values between hydrophilic and hydrophobic coatingsof the sensor. In particular with a hydrophobic surface, theresonance frequency and quality factor shifted to lower valuesand to higher values, respectively, compared to a hydrophilicsurface. This further decrease must be caused by nonspecificprotein adsorption on the hydrophobic surfaces that can lead toa conformational change of the proteins.45 If we supposeuniform mass adsorption on both sides of the cantilever, theadsorbed mass can be evaluated as an increase of thecantilever’s overall mass.30 Uniform mass adsorption resultsin an increase of the quality factor (see eq 2, where mc becomesmc + Δm after mass adsorption) while the resonance frequencydecreases. Resonance frequency also decreases if the mass is notuniformly distributed and concentrated in particular loca-tions.46,47 In Figure 3 after thrombin injection the resonancefrequency for OH-coated sensors shifted to lower values thanPEG-coated sensors while the quality factor remained in thesame range (∼24). This resonance frequency differenceindicates a mass adsorption onto the OH-coated sensors48

while a similar quality factor indicates that this mass adsorptionis not uniform. Nonuniform mass adsorption suggests thatclusters of fibrin are adsorbed onto the OH-coated surfaces.These clusters are far enough apart that they do not interactwith each other to form a uniform mass. The gold-coated,nonfunctionalized surfaces present higher uniform massadsorption of fibrinogen than hydrophobic surfaces. Adsorbedfibrinogen might undergo a conformational change, andsubsequent abnormal activity was measured by a suddendecrease in both quality factor and resonance frequency when

Figure 3. (a) Resonance frequency and quality factor change; (b) correlated density and viscosity change over time for three different SAMs andnonfunctionalized, gold-coated cantilevers: CH3-terminated (purple dots, hydrophobic, average of 2 cantilevers), OH-terminated (blue dots,hydrophilic, average of 3 cantilevers), PEG-terminated (black dots, hydrophilic, average of 3 cantilevers), and gold-coated without functionalization(red dots, average of 8 cantilevers) at 37 °C. The cantilever array was stabilized in water. At minute 10, 500 μL of 1 mM HEPES buffer was flushedthrough the chamber at 100 μL/min (5 min, white dashed area). After another 10 min buffer was exchanged with 100 μL of 2.5 mg/mL fibrinogensolution at 10 μL/min (yellow dashed area). Finally after 45 min (yellow plain area) 25 μL of a thrombin (70 NIH units/mL) and fibrinogen (2.5mg/mL) 1:1 mixture was injected. This requires 6 s; thus, it is too short to be visible on the plot. Resonance frequencies and quality factors duringthe reaction between thrombin and fibrinogen were recorded for the last 10 min (green plain area).

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

755

Page 6: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

no other solutions were injected (see minute 38 in Figure 3). Asimultaneous change of both resonance frequency and qualityfactor indicates that the firmly adsorbed fibrinogen attractsother molecules from the surrounding solution. Thesemolecules are loosely bound to the surface and to each other,causing a local, abnormal increase in viscosity. When theconversion of fibrinogen into fibrin was triggered by thrombin,further decreases in both resonance frequency and qualityfactor were measured. The formed clot showed stability for thenext 10 min. The stability of the fibrin mesh suggests that thepreviously loosely bound fibrinogen did not change itsconformation and maintained its normal activity. As expected,the gold-coated, nonfunctionalized surfaces trigger an irrever-sible protein mass adsorption compared to hydrophilic, protein-resistant surfaces.37 Figure 4 shows the difference in mass

adsorption between the hydrophobic and the hydrophilicsurfaces. The final mass of polymerized fibrin adsorbed ontothe hydrophobic surfaces is on the order of ∼20 ng, whichcorresponds to a layer with thickness of ∼160 nm (assumingprotein density to be 1.22 g/cm3; see ref 49). To correctlycalculate density and viscosity of a solution that containsproteins, a hydrophobic coating and nonfunctionalized, gold-coated surfaces have to be avoided.11 The performances ofPEG- and OH-terminated SAMs compared to CH3-terminatedSAM were tested further by subsequent injections of fibrinogenand thrombin mixtures. While on the PEG- and OH-coatedsurfaces no further quality factor and resonance frequencydecrease were observable, the CH3-coated sensors showed acontinuous shift, indicating further unspecific mass adsorption(data not shown). Surfaces that resist protein adsorption haveto present the following characteristics: (a) hydrophilic, (b)

include hydrogen-bond acceptors, (c) neutral, and (d) do nothave hydrogen-bond donors.48 Surfaces functionalized withPEG are therefore optimal for rheological measurements ofbiological liquids.

Human Blood Plasma Coagulation. The performance ofthe sensor array device as described above has been comparedto one of the most commonly used coagulation assay in clinicallaboratories, the aPTT test. The measured variable in astandard aPTT test is coagulation time, and no otherinformation is evaluated. Currently there are different methodsused to measure the aPTT time, and they include the following:(a) amperometric (electrochemical) measurement,6 (b) opticaldetection of mechanical motion,50,51 and (c) optical readout oflight scattering.52 The volumes of plasma required varydepending on whether the instrument is portable or not. Inthese methods the minimum volume required is ∼20 μL(portable, fingerstick) and ∼1 mL (nonportable). By measuringresonance frequencies and quality factors of cantilevers (PEG-coated) changing over time, we can determine not only theaPTT time but also the instantaneous coagulation rate plus thedensity and viscosity of the final clot, therefore providingadditional coagulation parameters. Figure 5 shows (a)frequency and quality factor change and (b) the calculateddensity and viscosity change during blood plasma coagulation.The resonance frequency and the quality factor shift to lowervalues when human blood plasma is injected into the chamberbecause plasma is both denser and more viscous than buffer.After oscillating the sensors for 10 min in plasma, all thereagents required for the aPTT test that trigger the coagulationcascade are injected into the chamber within 6 s, and a furtherdecrease of both resonance frequency and quality factor isdetected. Note that given the few nanometer displacements ofthe microcantilever, the total energy released into thesurrounding liquid is negligible. Details about calculation ofthe displacements and comparison of energy levels with otherclot-disrupting techniques are provided in the SupportingInformation. As mentioned above from these clinical assays, it ispossible to extract more parameters than aPTT testing such as(1) the viscosity and the density of both blood plasma and finalclot, (2) the aPTT time, and (3) the coagulation rate(measured in cP/s). These parameters provide importantinformation about the overall process of the blood plasmacoagulation, and they can be used to assess whether it isoccurring normally or abnormally. At present each clinicallaboratory has to determine and normalize the performances oftheir devices using an internal control (normal coagulation) inorder to evaluate the INR. The INR is then used as a target toreach when the patient is undergoing anticoagulation therapy.By providing absolute values (density, viscosity, and rate ofcoagulation), there is no need for an internal control and theanticoagulant therapy can be further tuned toward personalizeddiagnostics. Other work53,54 has extensively reviewed self-monitoring as an effective option for anticoagulation tests, andit has been concluded that self-monitoring is feasible only for aportion of patients. We believe that this portion could beincreased by measuring absolute values along with the usual PT,aPTT, and INR. Figure 6 shows detailed analysis of theparameters that can be extrapolated from measuring theviscosity change during coagulation. The measured viscosityof plasma is 0.967 ± 0.003 cP (physiological range36). Whenthe coagulation cascade is triggered, viscosity starts to increaseat a rate of 0.015 cP/s and then stabilizes at 1.832 ± 0.003 cP.The aPTT time is calculated at 47.2 s, which is in the typical

Figure 4. Differential mass adsorption between hydrophilic SAMs(PEG- and OH-terminated) and hydrophobic SAM (CH3-terminated)at 37 °C. The sequence of injections is the same as in Figure 3. Thedifference between mass adsorbed by hydrophobic and hydrophilicsurfaces when only fibrinogen solution is in the chamber was ∼2.5 ng.Fibrinogen is not adsorbed permanently on the surface and it ispossible to recover the same resonance frequency if the chamber isflushed again with buffer (data not shown). When thrombin is injectedtogether with fibrinogen, the latter polymerizes into a fibrin mesh thattends to adsorb onto the hydrophobic surfaces. The differential massadsorbed of fibrin mesh was ∼20 ng for Δ(CH3−PEG) and ∼10 ngfor Δ(CH3−OH) and Δ(OH−PEG), which corresponds to a layerwith thickness of ∼160 nm and ∼80 nm, respectively (consideringuniform adsorption on both sides of the cantilever and an averageprotein density of 1.22 g/cm3).49 The fibrin layer on the OH coatedsensor is not uniformly adsorbed onto the underlying surfaces.

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

756

Page 7: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

range for normal control plasma.55 The analysis has beenperformed on the viscosity values because they are morerepresentative of the coagulation process.

■ CONCLUSIONSWe introduced a method that exploits vibrating microcanti-levers to measure the coagulation time of human blood plasmaalong with viscosity, density, and coagulation rate. Theseparameters are monitored in real time, with high precision, andin small volumes. The sensitivity is 0.009 cP for a viscosityrange 0.5−3 cP and 0.0012 g/cm3 for density range 0.9−1.1 g/cm3. To the best of our knowledge, this is the highest viscosity

sensitivity value reported with micromechanical resonators inthe low-viscosity range. The device has a fluid volume chamberof 4 μL, and the volume required for complete volumeexchange within the measurement chamber is 20 μL. Theprocedure is fully automated. Manual pipetting or dispensing ofthe liquids is not required, increasing repeatability andminimizing operator errors. The sensors have been function-alized with SAMs that minimize mass adsorption andsubsequent protein conformational change, avoiding possibleabnormal activity and providing biocompatible protein-resistantsurfaces. The exact analytical solution of the hydrodynamicfunction provided by ref 24 has been approximated with amaximum deviation of 0.6% over a wide range of Reynold’snumbers (10−4−105). The device has been tested and thetheoretical model validated with glycerol reference solutionsand then utilized to monitor human blood plasma coagulationin real time. The small volumes achieved, the speed, and thereliability of the analysis make the presented device ideal for themicrorheological measurements of coagulation parameters indiagnostics.

■ ASSOCIATED CONTENT

*S Supporting InformationThe Supporting Information is available free of charge on theACS Publications website 0at DOI: 10.1021/acs.anal-chem.6b03347.

Resonance frequencies and quality factors evaluation,shear stress calculation and comparison of energy levelsbetween the microcantilever method and ultrasound-assisted lipoplasty (UAL) used for clot disruption;tabulated fitting parameters for the hydrodynamicapproximation, viscosity, and density parameters ofvarious glycerol solutions (PDF)

Figure 5. (a) Resonance frequency and quality factor change and (b) density and viscosity change during human blood plasma coagulation (mode10, average of 2 cantilevers). The cantilever array was stabilized in 1 mM HEPES buffer for 30 min (white area). After stabilization 100 μL of bloodplasma was flushed through the chamber at 10 μL/min (yellow dashed area). The flow was then stopped for 10 min (yellow plain area). Finally a 25μL mixture of plasma, colloidal solution of silica particles, and 25 mM CaCl2 (mixture 1:1:1) was injected, and the coagulation was monitored in realtime.

Figure 6. Viscosity change during human blood plasma coagulation.The plot shows the final part of the viscosity plot in Figure 5.Indicated: plasma viscosity before coagulation (yellow area) and aftercoagulation (orange area), initial rate of coagulation (slope in cP/s),and aPTT defined as the time required to reach a stable value of theviscosity considering a constant coagulation rate (0.015 cP/s). Thecalculated time and viscosity values are in the expected physiologicalrange.

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

757

Page 8: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

■ AUTHOR INFORMATION

Corresponding Author*E-mail: [email protected]. Phone: +353 (0) 1 896 2285.Fax: +353 (0) 1 896 3037.

ORCIDMartin Hegner: 0000-0003-1514-3437NotesThe authors declare no competing financial interest.

■ ACKNOWLEDGMENTS

We thank Prof. I. Dufour (Univ. Bordeaux) and Prof. J. Sader(Univ. Melbourne) for fruitful discussions and Patrick Murphy(Mechanical Workshop, TCD) for the manufacture of thetemperature stabilized enclosure. The work was supported byScience Foundation Ireland under the IvP scheme SFI/09IN/1B2623, the CSET scheme SFI/10/CSET/B1821 and the SFI/15/IA/3023.

■ REFERENCES(1) Reinhart, W. H. Biorheol. 2001, 38, 203−212.(2) Yilmaz, F.; Gundogdu, M. Y. Korea-Austr. Rheol. J. 2008, 20,197−211.(3) Evans, P.; Hawkins, K.; Lawrence, M.; Williams, R.; Barrow, M.;Thirumalai, N.; Williams, P. Med. Eng. Phys. 2008, 30, 671−679.(4) Key, N.; Makris, M.; O’Shaughnessy, D.; Lillicrap, D. PracticalHemostasis and Thrombosis; John Wiley & Sons: Chapel Hill, NC,2009.(5) Evans, P.; Hawkins, K.; Williams, P. Rheol. Rev. 2006, 255.(6) Cosmi, B.; Palareti, G.; Moia, M.; Carpenedo, M.; Pengo, V.;Biasiolo, A.; Rampazzo, P.; Morstabilini, G.; Testa, S. Thromb. Res.2000, 100, 279−286.(7) Da Luz, L. T.; Nascimento, B.; Rizoli, S. Scand. J. Trauma, Resusc.,Emerg. Med. 2013, 21, 29.(8) Rosencranz, R.; Bogen, S. A. Pathol. Patterns Rev. 2006, 125,S78−86.(9) Guhr, G.; Kunze, R.; Martin, G.; Schmidt, H.; Weihnachr, M.;Gehrisch, S.; Siegert, G. In IEEE Ultrasonics Symposium, 2005; IEEE:2005; pp 58−61.(10) Vikinge, T. P.; Hansson, K. M.; Sandstrom, P.; Liedberg, B.;Lindahl, T. L.; Lundstrom, I.; Tengvall, P.; Hook, F. Biosens.Bioelectron. 2000, 15, 605−613.(11) Lakshmanan, R. S.; Efremov, V.; Cullen, S. M.; Killard, A. J. Sens.Actuators, B 2014, 192, 23−28.(12) Shi-Hui, S.; Yuan-Jin, X.; Li-Hua, N.; Shou-Zhuo, Y. J. Biochem.Biophys. Methods 1996, 31, 135−143.(13) Cakmak, O.; Ermek, E.; Kilinc, N.; Bulut, S.; Baris, I.; Kavakli, I.;Yaralioglu, G.; Urey, H. Lab Chip 2015, 15, 113−120.(14) Trope, G.; Lowe, G.; McArdle, B.; Douglas, J.; Forbes, C.;Prentice, C.; Foulds, W. Br. J. Ophthalmol. 1983, 67, 137−142.(15) Williamson, T. H.; Rumley, A.; Lowe, G. Br. J. Ophthalmol.1996, 80, 203−208.(16) Strand, T.; Asplund, K.; Eriksson, S.; Hagg, E.; Lithner, F.;Wester, P. Stroke 1984, 15, 980−989.(17) Lowe, G.; Lee, A.; Rumley, A.; Price, J.; Fowkes, F. Br. J.Haematol. 1997, 96, 168−173.(18) Boskovic, S.; Chon, J.; Mulvaney, P.; Sader, J. J. Rheol. 2002, 46,891−899.(19) Ghatkesar, M. K.; Rakhmatullina, E.; Lang, H.-P.; Gerber, C.;Hegner, M.; Braun, T. Sens. Actuators, B 2008, 135, 133−138.(20) Shih, W. Y.; Li, X.; Gu, H.; Shih, W.-H.; Aksay, I. A. J. Appl.Phys. 2001, 89, 1497−1505.(21) Youssry, M.; Lemaire, E.; Caillard, B.; Colin, A.; Dufour, I.Meas.Sci. Technol. 2012, 23, 125306.(22) Lemaire, E.; Heinisch, M.; Caillard, B.; Jakoby, B.; Dufour, I.Meas. Sci. Technol. 2013, 24, 084005.

(23) Bircher, B. A.; Duempelmann, L.; Renggli, K.; Lang, H. P.;Gerber, C.; Bruns, N.; Braun, T. Anal. Chem. 2013, 85, 8676−8683.(24) Van Eysden, C. A.; Sader, J. E. J. Appl. Phys. 2007, 101, 044908.(25) Gessner, A.; Waicz, R.; Lieske, A.; Paulke, B.-R.; Mader, K.;Muller, R. Int. J. Pharm. 2000, 196, 245−249.(26) Rodrigues, S. N.; Goncalves, I. C.; Martins, M.; Barbosa, M. A.;Ratner, B. D. Biomaterials 2006, 27, 5357−5367.(27) Zeng, H. Polymer adhesion, friction, and lubrication; John Wiley &Sons: Edmonton, Alberta, Canada, 2013.(28) Wilson, C. J.; Clegg, R. E.; Leavesley, D. I.; Pearcy, M. J. TissueEng. 2005, 11, 1−18.(29) Walther, M.; Fleming, P. M.; Padovani, F.; Hegner, M. EPJ.Technol. Instrument. 2015, 2, 1.(30) Braun, T.; Barwich, V.; Ghatkesar, M. K.; Bredekamp, A. H.;Gerber, C.; Hegner, M.; Lang, H. P. Phys. Rev. E: Stat. Phys., Plasmas,Fluids 2005, 72, 031907.(31) Cakmak, O.; Elbuken, C.; Ermek, E.; Mostafazadeh, A.; Baris, I.;Alaca, B. E.; Kavakli, I. H.; Urey, H. Methods 2013, 63, 225−232.(32) Van Eysden, C. A.; Sader, J. E. J. Appl. Phys. 2006, 100, 114916.(33) Van Eysden, C. A.; Sader, J. E. Phys. Fluids 2006, 18, 123102.(34) Jensen, J.; Maloney, N.; Hegner, M. Sens. Actuators, B 2013, 183,388−394.(35) Wagner, P.; Hegner, M.; Kernen, P.; Zaugg, F.; Semenza, G.Biophys. J. 1996, 70, 2052.(36) Kesmarky, G.; Kenyeres, P.; Rabai, M.; Toth, K. Clin. Hemorheol.Microcirc. 2008, 39, 243−246.(37) Love, J. C.; Estroff, L. A.; Kriebel, J. K.; Nuzzo, R. G.;Whitesides, G. M. Chem. Rev. 2005, 105, 1103−1170.(38) Michel, R.; Pasche, S.; Textor, M.; Castner, D. G. Langmuir2005, 21, 12327−12332.(39) Prime, K. L.; Whitesides, G. M. J. Am. Chem. Soc. 1993, 115,10714−10721.(40) Malmsten, M. J. Colloid Interface Sci. 1998, 207, 186−199.(41) Wagner, W.; Kretzschmar, H.-J. International Steam TablesProperties of Water and Steam Based on the Industrial FormulationIAPWS-IF97; Springer-Verlag: Berlin/Heidelberg, 2008.(42) Fritsma, G. A.; Dembitzer, F. R.; Randhawa, A.; Marques, M. B.;Van Cott, E. M.; Adcock-Funk, D.; Peerschke, E. I. Am. J. Clin. Pathol.2012, 137, 904−908.(43) Tegoulia, V. A.; Rao, W.; Kalambur, A. T.; Rabolt, J. F.; Cooper,S. L. Langmuir 2001, 17, 4396−4404.(44) Hemmersam, A. G.; Foss, M.; Chevallier, J.; Besenbacher, F.Colloids Surf., B 2005, 43, 208−215.(45) Krishnan, A.; Liu, Y.-H.; Cha, P.; Allara, D.; Vogler, E. A. J. R.Soc., Interface 2006, 3, 283−301.(46) Hanay, M. S.; Kelber, S. I.; O’Connell, C. D.; Mulvaney, P.;Sader, J. E.; Roukes, M. L. Nat. Nanotechnol. 2015, 10, 339−344.(47) Maloney, N.; Lukacs, G.; Jensen, J.; Hegner, M. Nanoscale 2014,6, 8242−8249.(48) Ostuni, E.; Chapman, R. G.; Holmlin, R. E.; Takayama, S.;Whitesides, G. M. Langmuir 2001, 17, 5605−5620.(49) Quillin, M. L.; Matthews, B. W. Acta Crystallogr., Sect. D: Biol.Crystallogr. 2000, 56, 791−794.(50) Carter, A. J.; Hicks, K.; Heldman, A. W.; Resar, J. R.; Laird, J. R.;Coombs, V. J.; Brinker, J. A.; Blumenthal, R. S. Catheter. Cardiovasc.Interv. 1996, 39, 97−102.(51) Thomas, O.; Schott, U. J. Cardiovasc. Dis. Res. 2013, 1, 8−12.(52) D’Angelo, A.; Seveso, M. P.; D'Angelo, S. V.; Gilardoni, F.;Macagni, A.; Manotti, C.; Bonini, P. Am. J. Clin. Pathol. 1989, 92, 321−328.(53) Heneghan, C.; Ward, A.; Perera, R.; Bankhead, C.; Fuller, A.;Stevens, R.; Bradford, K.; Tyndel, S.; Alonso-Coello, P.; Ansell, J.; et al.Lancet 2012, 379, 322−334.(54) Sharma, P.; Scotland, G.; Cruickshank, M.; Tassie, E.; Fraser, C.;Burton, C.; Croal, B.; Ramsay, C. R.; Brazzelli, M. BMJ. Open 2015, 5,e007758.(55) Takemoto, C. M.; Streiff, M. B.; Shermock, K. M.; Kraus, P. S.;Chen, J.; Jani, J.; Kickler, T. Am. J. Clin. Pathol. 2013, 139, 450−456.

Analytical Chemistry Article

DOI: 10.1021/acs.analchem.6b03347Anal. Chem. 2017, 89, 751−758

758

Page 9: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S1

Supporting Information for:

Microrheological Coagulation Assay

Exploiting Micromechanical Resonators

Francesco Padovani1, James Duffy1 and Martin Hegner1*

1CRANN, School of Physics, Trinity College Dublin, Dublin, Ireland.

*Corresponding Author

E-mail: [email protected]

Abstract This Supporting Information provides details on resonance frequencies and quality factors

evaluation, shear stress calculation and comparison of energy levels between the

microcantilever method and Ultrasound-assisted lipoplasty (UAL) used for clot disruption.

Tabulated fitting parameters for the hydrodynamic approximation, viscosity and density

parameters of various glycerol solutions that are cited in the main text are provided at the end

of this document.

Page 10: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S2

Resonance frequency and quality factor evaluation. According to the simple harmonic oscillator modelS1 the transient solution of the cantilever

response diminishes exponentially with time constant 𝜏𝜏 = 2𝑄𝑄/𝜔𝜔0 where 𝑄𝑄 is the quality

factor and 𝜔𝜔0 is the radial resonance frequency. In all of the following experiments the

quality factor was in the range 10-30 and the resonance frequency was in the range 300-400

kHz which results in 𝜏𝜏 < 0.05 ms. This time constant is much lower than the driving time (1

ms for each frequency in the spectrum), therefore the steady state solution is reached. The

amplitude and phase of the response signal were evaluated for each driven frequency. The

total time required for the acquisition of a single spectrum can be adjusted by changing the

number of driven frequencies. For fast tracking of the resonance frequencies and quality

factors a total of 1000 frequencies were chosen, resulting in an overall driving time of 1 s for

each sensor analysis. The time resolution is therefore in the order of 2-3 s, thus quick changes

in viscosity and/or density can be tracked in real-time.

The recorded amplitude and phase spectra were analysed using a custom LabVIEW

programme where the Simple Harmonic Oscillator (SHO) model is fitted to the data through

non-linear least square fitting. The well-known SHO equations correlate amplitude and phase

spectra to resonance frequency 𝑓𝑓𝑅𝑅,𝑛𝑛 and quality factor 𝑄𝑄𝑛𝑛 as follows:

𝐴𝐴0(𝑓𝑓) =

𝐴𝐴𝑑𝑑𝑓𝑓𝑅𝑅,𝑛𝑛2

��𝑓𝑓𝑅𝑅,𝑛𝑛2 − 𝑓𝑓2�

2+ �

𝑓𝑓𝑅𝑅,𝑛𝑛𝑓𝑓𝑄𝑄𝑛𝑛

�2

+ 𝐴𝐴𝑏𝑏𝑏𝑏 (1)

Δ𝜙𝜙(𝑓𝑓) = tan−1

𝑓𝑓𝑅𝑅,𝑛𝑛𝑓𝑓𝑄𝑄𝑛𝑛�𝑓𝑓𝑅𝑅,𝑛𝑛

2 − 𝑓𝑓2�+ 𝜙𝜙𝑏𝑏𝑏𝑏

(2)

where 𝐴𝐴𝑑𝑑 and 𝑓𝑓 are the amplitude and frequency of the driving signal respectively, 𝐴𝐴0(𝑓𝑓) is

the amplitude spectrum of the cantilever response signal, Δ𝜙𝜙(𝑓𝑓) is the phase spectrum of the

difference between the driving signal’s phase and the cantilever response signal while 𝐴𝐴𝑏𝑏𝑏𝑏

and 𝜙𝜙𝑏𝑏𝑏𝑏 are the white noise floor amplitudeS2 and a phase constant respectively. The phase

constant 𝜙𝜙𝑏𝑏𝑏𝑏 is introduced to consider the offset of the phase. Note that compared to previous

workS3 no slope correction parameters are required. These slope parameters would in fact

distort the SHO equations and could lead to an incorrect quality factor evaluation. The

constants 𝐴𝐴𝑏𝑏𝑏𝑏 and 𝜙𝜙𝑏𝑏𝑏𝑏 increase the fitting accuracy but do not affect quality factor and

resonance frequency evaluation.

Page 11: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S3

Shear stress on the microcantilever surface. The microcantilever mode shape (Figure S1) for the displacement is a function 𝑢𝑢(𝑥𝑥, 𝑡𝑡) that

can be expressed using the Euler-Bernoulli beam theory as follows:

⎩⎪⎨

⎪⎧

𝑢𝑢(𝑥𝑥, 𝑡𝑡) = 𝑢𝑢(𝑥𝑥) ∙ sin(𝜔𝜔𝑡𝑡)

𝑢𝑢(𝑥𝑥) = 𝑢𝑢1 �cosh �𝐶𝐶𝑛𝑛𝑥𝑥𝑙𝑙� − cos �

𝐶𝐶𝑛𝑛𝑥𝑥𝑙𝑙� + 𝐵𝐵1 �sin �

𝐶𝐶𝑛𝑛𝑥𝑥𝑙𝑙� − sinh �

𝐶𝐶𝑛𝑛𝑥𝑥𝑙𝑙���

𝐵𝐵1 =cos(𝐶𝐶𝑛𝑛) + cosh(𝐶𝐶𝑛𝑛)sin(𝐶𝐶𝑛𝑛) + sinh(𝐶𝐶𝑛𝑛)

(3)

where 𝑙𝑙 is the length of the cantilever. The force generated by the cantilever reaches a

maximum in time when sin(𝜔𝜔𝑡𝑡) = 1. We can then calculate the acceleration ��𝑢(𝑥𝑥, 𝑡𝑡) in this

particular 𝑡𝑡 = 𝑡𝑡𝑚𝑚𝑚𝑚𝑚𝑚 as:

��𝑢(𝑥𝑥) = 𝑢𝑢(𝑥𝑥) ∙ 𝜔𝜔2 ∙ sin(𝜔𝜔𝑡𝑡𝑚𝑚𝑚𝑚𝑚𝑚) (4)

The total force can be evaluated with the well-known equation of motion (where the damping

term is included in F(𝑥𝑥, 𝑡𝑡)):

F(𝑥𝑥, 𝑡𝑡) = 𝑘𝑘 ∙ 𝑢𝑢(𝑥𝑥, 𝑡𝑡) + 𝑚𝑚 ∙ ��𝑢(𝑥𝑥, 𝑡𝑡) (5)

The stress (𝑢𝑢 direction, see Figure S1) on the surface of the cantilever can be related to the

total force through the following relationship:

𝑑𝑑F(𝑥𝑥, 𝑡𝑡)𝑑𝑑𝑥𝑥

∙1𝑏𝑏

= σ(𝑥𝑥, 𝑡𝑡) = �𝑘𝑘 ∙𝑑𝑑𝑢𝑢(𝑥𝑥)𝑑𝑑𝑥𝑥

+ 𝑚𝑚 ∙𝑑𝑑𝑢𝑢(𝑥𝑥)𝑑𝑑𝑥𝑥

∙ 𝜔𝜔2 ∙ sin(𝜔𝜔𝑡𝑡)� ∙1𝑏𝑏

(6)

where 𝑏𝑏 is the width of the cantilever. We can now calculate the shear stress (component of

the force that is tangent to the cantilever motion) at 𝑡𝑡 = 𝑡𝑡0 as:

𝜏𝜏(𝑥𝑥) = 𝜎𝜎(𝑥𝑥) ∙ sin �tan−1 �

𝑑𝑑𝑢𝑢(𝑥𝑥)𝑑𝑑𝑥𝑥

�� (7)

Figure S1 – Cantilever mode shape 𝑢𝑢(𝑥𝑥) (see equation (3)) and decomposition of the stress vector 𝜎𝜎(𝑥𝑥) into normal 𝜎𝜎𝑛𝑛(𝑥𝑥) and tangent 𝜏𝜏(𝑥𝑥) components to the surface of the cantilever.

Page 12: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S4

Equation (7) requires that 𝑢𝑢(𝑥𝑥) is fully determined, thus 𝑢𝑢1 has to be measured. Figure S2

shows a schematic representation of the optical readout geometry that we used to determine

𝑢𝑢(𝑥𝑥).

Figure S2 – Geometrical representation of the optical readout for the determination of 𝑢𝑢(𝑥𝑥).

The displacement Δ𝑑𝑑 of the laser spot on the PSD reflected from the cantilever surface is

geometrically related to the rotation angle 𝛼𝛼 of the tangent to the cantilever surface in the

node as follows:

𝛼𝛼 =

12

tan−1 �Δ𝑑𝑑𝑑𝑑𝑃𝑃𝑃𝑃𝑃𝑃

� (8)

To estimate the maximum amplitude of the oscillation 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚 first we need to calculate the

position 𝑥𝑥𝑛𝑛 of the node impinged by the laser beam and then from the measurement of 𝛼𝛼

(equation (8)) we can calculate 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚 as follows:

�𝑑𝑑𝑢𝑢(𝑥𝑥)𝑑𝑑𝑥𝑥

=𝐶𝐶𝑛𝑛𝑙𝑙𝑢𝑢1 �sinh �

𝐶𝐶𝑛𝑛𝑥𝑥𝑛𝑛 𝑙𝑙

� + sin �𝐶𝐶𝑛𝑛𝑥𝑥𝑛𝑛 𝑙𝑙

� + 𝐵𝐵1 �cos �𝐶𝐶𝑛𝑛𝑥𝑥𝑛𝑛 𝑙𝑙

� − cosh �𝐶𝐶𝑛𝑛𝑥𝑥𝑛𝑛 𝑙𝑙

���

𝑢𝑢(𝑥𝑥𝑛𝑛) = 0 (9)

From the set of equations (9) we can calculate 𝑢𝑢1 and 𝑥𝑥𝑛𝑛 and subsequently the maximum

displacement 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚. Evaluation of 𝑥𝑥𝑛𝑛 requires a numerical method. Numerical values for a

cantilever with 𝑙𝑙 = 500 µm are summarized in Table S1.

Maximum shear stress 𝜏𝜏𝑚𝑚𝑚𝑚𝑚𝑚 was calculated by combining equations (6), (7) and (9).

dPSD [mm] Δd [µm] 𝑥𝑥𝑛𝑛 [µm] 𝛼𝛼 [deg] 𝑢𝑢1 [nm] 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚 [nm] 𝜏𝜏𝑚𝑚𝑚𝑚𝑚𝑚

50 19.050 328.95 ~0.011 ~2.3 ~4.5 0.52

Table S1– Numerical values obtained using equations (1) to (9).

The obtained value of 0.52 Pa is ~280-fold smaller than the critical shear stress for fibrin

clots disruptionS4. Figure S3 shows the comparison between the measured normalized

differential signal and the expected 𝑑𝑑𝑢𝑢(𝑥𝑥)/𝑑𝑑𝑥𝑥 along the cantilever length (𝑥𝑥 direction).

Page 13: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S5

Figure S3 – Comparison between the measured and expected derivative of the mode shape 𝑢𝑢(𝑥𝑥) at mode 10. The experimental data points were evaluated combining equations (8) and (9). Each maxima and minima correspond to nodes in 𝑢𝑢(𝑥𝑥). Note that 𝑥𝑥=500 is the tip of the cantilever. To calculate 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚 the measured value 0.00381 at the node 𝑥𝑥𝑛𝑛=328.95 was inserted into equation (S9) to obtain 𝑢𝑢1. According to equation (S3) the maximum displacement is at 𝑥𝑥=500 and is 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚 = 2𝑢𝑢1. These displacements are consistent with typical displacements of piezo-electric stack actuatorsS5.

Ultrasound-assisted lipoplasthy versus microcantilever energies. Previous works have reported clot disruption with ultrasound techniquesS6-S10. These methods

exploit two different physical phenomena: cavitation and mechanical vibration of a probe. In

order to compare these methods to our presented study we can assume the vibrating

microcantilever as a structure that generates acoustic waves into the surrounding media. The

principle is similar to ultrasound-assisted lipoplasty (UAL)S7,S10. In UAL a probe is placed

with the help of a catheter in the vicinity of or directly into a thrombus. The probe is then

driven by an ultrasound energy source and probe vibrations eventually cause disassociation of

the clot. By comparing frequencies, amplitudes and energies between the microcantilever

method and UAL we can determine if the vibrating microcantilever can break the fibrin clot.

The microcantilever displacement 𝑢𝑢𝑚𝑚𝑚𝑚𝑚𝑚 obtained with the above method is in the order of 4-

5 nm (mode 10, ~360 kHz). In UAL typical displacements of the probe tip are in the order of

60-100 µm, while the frequency of the vibrations is 20 kHz. These two parameters provide a

quick comparison between total energy levels of the two techniques (microcantilever and

UAL). The sound intensity of the acoustic waves generated by the vibrating structures is

directly proportional to the square of the frequency and amplitudeS11. While the

Page 14: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S6

microcantilevers resonance frequency is ~18-fold higher than the UAL probe operating

frequency the displacements are ~16000-fold smaller resulting in a total acoustic intensity

almost 6 orders of magnitude smaller with the microcantilever method. Furthermore the

typical driving time in a UAL is 30 ms while the microcantilevers array is driven for 1 ms.

List of tables cited in the main text. Table S2 – Fitting parameters for the approximation of the hydrodynamic function (see

equations (3) and (4) in the main text).

Mode 𝑛𝑛 (𝑏𝑏/𝑙𝑙=1/5)

𝜅𝜅𝑛𝑛 A G C B a2 a3 H

1 0.375 0.976404 0.001790 0.000020 0.803558 0.050012 -0.003017 -1.457784

2 0.9388 0.890066 0.007717 0.006190 0.985864 0.071633 -0.006330 -0.501259

3 1.5708 0.777928 0.014362 0.055505 1.137829 0.075309 -0.007009 -0.042264

4 2.2 0.677486 0.025123 0.631616 1.237636 0.082840 -0.008646 0.344548

5 2.827433 0.594123 0.042516 2.503898 1.295157 0.083365 -0.008832 0.550529

6 3.455752 0.525971 0.068367 8.437011 1.356503 0.083736 -0.009024 0.737802

7 4.084070 0.470279 0.095842 20.825717 1.413807 0.082939 -0.008935 0.886045

8 4.712389 0.424376 0.123064 43.040196 1.464434 0.082127 -0.008821 1.011039

9 5.340708 0.386121 0.159901 84.373280 1.509711 0.081177 -0.008648 1.120454

10 5.969026 0.353875 0.193957 147.589550 1.550751 0.080433 -0.008512 1.217325

11 6.597344 0.326397 0.229068 242.169180 1.588196 0.079353 -0.008261 1.301415

12 7.225663 0.302745 0.241368 349.714580 1.622697 0.077590 -0.007794 1.369993

13 7.853982 0.282198 0.231637 453.807132 1.654686 0.075406 -0.007185 1.426968

14 8.4823 0.264199 0.246054 625.765033 1.684469 0.074279 -0.006862 1.488896

15 9.110618 0.248314 0.287703 915.450851 1.712284 0.073421 -0.006596 1.550510

16 9.738937 0.234198 0.370438 1439.098030 1.738420 0.073541 -0.006614 1.616738

N/A 10 0.228786 0.425679 1746.575342 1.748830 0.073958 -0.006735 1.646025

N/A 20 0.120913 0.959381 39383.023573 2.029474 0.056486 0.000000 2.178497

Page 15: Microrheological Coagulation Assay Exploiting ...€¦ · Microrheological Coagulation Assay Exploiting Micromechanical Resonators Francesco Padovani, James Duffy, and Martin Hegner*

S7

Table S3 – Density (g/cm3) and viscosity (cP) of glycerol solutions at 23°CS12.

Glycerol % Viscosity (cP) Density (g/cm3)

0 0.9532 0.9977

5 1.0642 1.0095

10 1.1698 1.0214

12 1.2323 1.0266

14 1.2685 1.0312

15 1.304 1.0337

20 1.4838 1.0462

30 2.2843 1.0717

40 3.0221 1.0983

References (S1) Bhusan, B. Springer Handbook of Nanotechnology; Springer-Verlag Berlin Heidelberg, 2010. (S2) Chon, J. W.; Mulvaney, P.; Sader, J. E. J. Appl. Phys. 2000, 87, 3978-3988. (S3) Bircher, B. A.; Duempelmann, L.; Renggli, K.; Lang, H. P.; Gerber, C.; Bruns, N.; Braun, T. Anal. Chem. 2013, 85, 8676-8683. (S4) Longstaff, C.; Varjú, I.; Sótonyi, P.; Szabó, L.; Krumrey, M.; Hoell, A.; Bóta, A.; Varga, Z.; Komorowicz, E.; Kolev, K. J. Biol. Chem. 2013, 288, 6946-6956. (S5) Jensen, J.; Maloney, N.; Hegner, M. Sens. Actuators B : Chem. 2013, 183, 388-394. (S6) Everbach, E. C.; Francis, C. W. Ultrasound Med. Biol. 2000, 26, 1153-1160. (S7) Ariani, M.; Fishbein, M. C.; Chae, J. S.; Sadeghi, H.; Michael, A. D.; Dubin, S.; Siegel, R. Circulation 1991, 84, 1680-1688. (S8) Ensminger, D.; Bond, L. J. Ultrasonics: fundamentals, technologies, and applications; CRC press, 2011. (S9) Rosenschein, U.; Bernstein, J. J.; DiSegni, E.; Kaplinsky, E.; Bernheim, J.; Rozenzsajn, L. A. J. Am. Coll. Cardiol. 1990, 15, 711-717. (S10) Wylie, M. P.; McGuinness, G. B.; Gavin, G. P. In 2009 Ann. Intl. Conf. IEEE Eng. Med. Biol. Soc.; IEEE, 2009, pp 282-285. (S11) Elert, G. The Physics Hypertextbook, July 18, 2006; Vol. 2016. (S12) Cakmak, O.; Elbuken, C.; Ermek, E.; Mostafazadeh, A.; Baris, I.; Alaca, B. E.; Kavakli, I. H.; Urey, H. Methods 2013, 63, 225-232.