mineral grains from ivuna (2004.06.16) short summary
DESCRIPTION
Mineral grains from Ivuna (2004.06.16) Short summary. 4 samples removed from the Ivuna meteorite were imaged at beamline BL47XU of SPring-8 on June 16, 2004. - PowerPoint PPT PresentationTRANSCRIPT
Mineral grains from Ivuna (2004.06.16)Short summary
• 4 samples removed from the Ivuna meteorite were imaged at beamline BL47XU of SPring-8 on June 16, 2004.
• All the samples seem to be carbonate. MZ-IV2-3 must be calcite deducing from its CT value (see Table below). The phases of the other samples cannot be estimated due to the effect of harmonics at 8 keV.
• Voids were seen in three samples (MZ-IV2-1, 2 and 3). Many voids in MZ-IV2-3 (calcite) are closed and might be fluid inclusions. Many voids in MZ-IV2-1 and 3 are connected to the outside and should be healed cracks.
sampleimaging
No.energy
exposuretime
projectionmineral
candidatesCT value at
histogram peakremarks
keV sec cm-1
MZ-IV2-2 040616b 8 1 750 183.529412 heald cracks?MZ-IV2-4 040616c 8 1 750 126.406236 very clear crystal, no voidMZ-IV2-1 040616d 8 1 750 97.56186 healed cracksMZ-IV2-3 040616e 10 1 750 calcite 105.07392 many closed voids
mineral/density(g/cc) LAC at 8 keV LAC at 10 keV
cm-1
cm-1
magnesite/MgCO3/2.98 56.557539 29.187255 dolomite/CaMg(CO3)2/2.86 141.6159 75.743928 calcite/CaCO3/2.72 204.654557 110.27765
MZ-IV2-2 (040616b)
Imaging No.: 040616bSample: MZ-IV2-2; a piece of transparent crystalBeamline: BL20XUX-ray photon energy: 8 keVExposure time for each projection: 1 sec.No. of pixels for each projection image: 512× 530 pels
No. of projection images for dark current : 2 (1 each before and after sample imaging)No. of projection images for incident beam current (I0): 151 (+1; for determining the rotation axis)No. of projection images for transmitted beam current (I): 750 (+1; ditto)Imaging sequence: One I0 imaging after every 5 I imagingPixel size of projection images: 0.23 micron
Voxel size of reconstructed CT images : 0.23 micronNo. of voxels : 512 × 512 × 530Location of sample rotation axis:R0(z)= 0.007793×z +233.936012 ±0.828852 ( z=0 ~ 529 )
Byte images:PV=0: CT value = 0 1/cmPV=180: CT value = 183.529412 1/cm
(peak value)
Cut images: Threshold: PV=74Erosion: 3 layersDilation: 6 layers
Voxel size : 0.23 micronNumber of voxels : 371 × 331 × 509
cut/271.tif (width: 85.33 m)
0
5 104
1 105
1.5 105
2 105
2.5 105
3 105
3.5 105
4 105
0 50 100 150 200 250
byte_hst
N
PV
PV=74 (min)
PV=180
MZ-IV2-2 (040616b): browse image
MZ-IV2-2 (040616b): external shape
+X
+Y
-Z image
+Y
-Z
+X image
+X
-Z+Y image
-Z
+X
+Y
MZ-IV2-2 (040616b): frequency diagram and phases
CT value of calcite: 204.65 cm-1
Estimated LAC of calcite: 181.8 cm-1
CT value of dolomite: 141.62 cm-1
Estimated LAC of calcite: 125.1 cm-1
CT value = 0.88305(651) * LAC
MineralsWhite (CTv >260 cm-1): ?Gray (CTv = 183.5 cm-1): calcite??Black: void, airIdentification of mineral phases from the CT values is difficult due to contribution of harmonics at this energy (8 keV).
voidcarbonate
?
3
3.5
4
4.5
5
5.5
6
6.5
7
0 50 100 150 200 250
byte_hst
CT value /cm-1
183.5 cm -1
cut/271.tif (width: 85.33 m)
MZ-IV2-4 (040616c)
Imaging No.: 040616cSample: MZ-IV2-4; a piece of transparent crystalBeamline: BL20XUX-ray photon energy: 8 keVExposure time for each projection: 1 sec.No. of pixels for each projection image: 512× 280 pels
No. of projection images for dark current : 2 (1 each before and after sample imaging)No. of projection images for incident beam current (I0): 76 (+1; for determining the rotation axis)No. of projection images for transmitted beam current (I): 750 (+1; ditto)Imaging sequence: One I0 imaging after every 10 I imagingPixel size of projection images: 0.23 micron
Voxel size of reconstructed CT images : 0.23 micronNo. of voxels : 512 × 512 × 280Location of sample rotation axis:R0(z)= 0.007785×z +229.295481 ±0.000000 ( z=0 ~ 279 )
Byte images:PV=0: CT value = 0 1/cmPV=180: CT value = 126.406236 1/cm
(peak value)
Cut images: Threshold: PV=100Erosion: 3 layersDilation: 6 layers
Voxel size : 0.23 micronNumber of voxels : 184 × 405 × 280
cut/130.tif (width: 42.32 m)
0
5 104
1 105
1.5 105
2 105
0 50 100 150 200 250
byte_hst
N
PV
PV=100 (min)
PV=180
MZ-IV2-4 (040616c): browse image
MZ-IV2-4 (040616c): external shape
+X
+Y
-Z image
+Y
-Z
+X image
+X
-Z+Y image
-Z
+X
+Y
MZ-IV2-4 (040616c): frequency diagram and phases
CT value of calcite: 204.65 cm-1
Estimated LAC of calcite: 181.8 cm-1
CT value of dolomite: 141.62 cm-1
Estimated LAC of calcite: 125.1 cm-1
CT value = 0.88305(651) * LAC
MineralsWhite (CTv >179 cm-1): ?Gray (CTv = 126.4 cm-1): dolomite-calcite??Black: void, airIdentification of mineral phases from the CT values is difficult due to contribution of harmonics at this energy (8 keV).
carbonate
?
3
3.5
4
4.5
5
5.5
6
6.5
0 40 80 120 160
byte_hst
CT value /cm-1
126.4 cm -1
cut/130.tif (width: 42.32 m)
MZ-IV2-1 (040616d)
Imaging No.: 040616dSample: MZ-IV2-1; a piece of transparent crystalBeamline: BL20XUX-ray photon energy: 8 keVExposure time for each projection: 1 sec.No. of pixels for each projection image: 512× 420 pels
No. of projection images for dark current : 2 (1 each before and after sample imaging)No. of projection images for incident beam current (I0): 76 (+1; for determining the rotation axis)No. of projection images for transmitted beam current (I): 750 (+1; ditto)Imaging sequence: One I0 imaging after every 10 I imagingPixel size of projection images: 0.23 micron
Voxel size of reconstructed CT images : 0.23 micronNo. of voxels : 512 × 512 × 420Location of sample rotation axis:R0(z)= 0.008687×z +226.219933 ±0.643388 ( z=0 ~ 419 )
Byte images:PV=0: CT value = 0 1/cmPV=180: CT value = 97.561860 1/cm
(peak value)
Cut images: Threshold: PV=116Erosion: 3 layersDilation: 6 layers
Voxel size : 0.23 micronNumber of voxels : 201 × 340 × 400
cut/260.tif (width: 46.23 m)
0
5 104
1 105
1.5 105
2 105
0 50 100 150 200 250
byte_hst
N
PV
PV=116 (min)
PV=180
MZ-IV2-1 (040616d): browse image
MZ-IV2-1 (040616d): external shape
+X
+Y
-Z image
+Y
-Z
+X image
+X
-Z+Y image
-Z
+X
+Y
MZ-IV2-1 (040616d): frequency diagram and phases
CT value of calcite: 204.65 cm-1
Estimated LAC of calcite: 181.8 cm-1
CT value of dolomite: 141.62 cm-1
Estimated LAC of calcite: 125.1 cm-1
CT value = 0.88305(651) * LAC
cut/176.tif (width: 46.23 mm)
MineralsWhite (CTv >138 cm-1): ?Gray (CTv = 97.61 cm-1): magnesite-dolomite??Black: void, airIdentification of mineral phases from the CT values is difficult due to contribution of harmonics at this energy (8 keV).
void
carbonate
?
3.5
4
4.5
5
5.5
6
6.5
0 20 40 60 80 100 120 140
byte_hst
CT value /cm-1
97.6 cm -1
MZ-IV2-1 (040616d): voids(1) Voids along healed cracks? We can see many voids. Some of them seem to be closed. But they are located along
lines and seem to be healed cracks.
255.tif 260.tif250.tif
265.tif 270.tif 275.tif
MZ-IV2-3 (040616e)
Imaging No.: 040616eSample: MZ-IV2-3; a piece of transparent crystalBeamline: BL20XUX-ray photon energy: 10 keVExposure time for each projection: 1 sec.No. of pixels for each projection image: 1024× 820 pels
No. of projection images for dark current : 2 (1 each before and after sample imaging)No. of projection images for incident beam current (I0): 76 (+1; for determining the rotation axis)No. of projection images for transmitted beam current (I): 750 (+1; ditto)Imaging sequence: One I0 imaging after every 10 I imagingPixel size of projection images: 0.23 micron
Voxel size of reconstructed CT images : 0.23 micronNo. of voxels : 1024 × 1024 × 820Location of sample rotation axis:R0(z)= 0.006011×z +478.638705 ±0.872422 ( z=0 ~ 819 )
Byte images:PV=0: CT value = 0 1/cmPV=180: CT value = 105.073920 1/cm
(peak value)
Cut images: Threshold: PV=80Erosion: 3 layersDilation: 6 layers
Voxel size : 0.23 micronNumber of voxels : 502 × 199 × 820
cut/418.tif (width: 115.46 m)
0
2 105
4 105
6 105
8 105
1 106
0 50 100 150 200 250
byte_hst
N
PV
PV=80(min)
PV=180
MZ-IV2-3 (040616e): browse image
MZ-IV2-3 (040616e): external shape
+X
+Y
-Z image
+Y
-Z
+X image
+X
-Z+Y image
-Z
+X+Y
MZ-IV2-3 (040616e): frequency diagram and phases
CT value of calcite: 110.28 cm-1
Estimated LAC of calcite: 97.4 cm-1
CT value = 0.88305(651) * LAC
cut/492.tif (width: 115.46 mm)
MineralsWhite (CTv >149 cm-1): ?Gray (CTv = 105.1 cm-1): calciteBlack: void, air
void
calcite
?
2
3
4
5
6
7
8
0 50 100 150
byte_hst
CT value /cm-1
105.1 cm -1
MZ-IV2-3 (040616e): voids(1) Closed voids We can recognize many closed voids in slices of 95-104, 111-115, 127-144, 163?, 210-
211, 269?, 314-317, 353, 382?, 411-424, 536-551, etc.
(2) Voids along healed cracks?
100.tif
316.tif
418.tif
572.tif
Glossary Beamline
Three beamlines (BL20B2, BL20XU and BL47XU) are now available at Spring-8 for X-ray microtomography. We can choose the beamline mainly depending on the sample size and the spatial resolution.
Browse imageSome of a successive CT images of a sample can be seen at a glance by this image. Size scale (cm) and gray scale for CT values (cm-1) are shown in the right and left bottoms of the image, respectively.
Byte imageAn monochromatic image where the brightness of each pixel is expressed by 8 bit (0-255 steps). A series of byte images of an sample stored in the directory named “byte” are formed from original CT images by conversion from CT value to PV (usualy the theoretical LAC of a standard mineral is used for this conversion).
CT valueThis is the value of LAC in each pixel of an CT image. In other words, an CT image is expressed as spatial distribution of CT values. CT values are calculated by CT reconstruction. If we use a monochromatic X-ray beam, CT value is ideally equal to the LAC of a material. However, they are not identical with each other in practice. Their empirical relations were determined at BL20B2 and BL47XU of Spring-8 (see np.29 for the relation at BL47XU). From this relation, we can estimate materials from CT values.
Cut imageThis is a CT image, where unnecessary regions without sample are trimmed. A series of cut images of an sample stored in the directory named “cut” are formed from “byte” images by image processing using threshold for a sample and erosion and dilation procedures.
Imaging No.Imaging number shows the date of imaging and its order. For example, 031204c shows the third experiment (in alphabetical order) on 4 December, 2003.
LAC (Linear attenuation coefficient) (of X-ray)A characteristic properties for X-ray attenuation of a material. This is a function of the bulk density, the chemical composition and X-ray photon energy..
PixelOne of many from which a 2-D image is composed. The shape is usually square.
Projection imageCT images are reconstructed from a series of projection images by rotating a sample by180 degrees. The size of a projection image gives the size of a CT image. For example, projection images of 512 × 220 pixels give 220 CT images of 512 × 512 pixels.
Projection image for dark currentA projection image without X-ray beam (dark image). This gives back ground of the projection images. Usually, 1 (or 2) dark image(s) is (are) taken before and after sample imaging experiment and their average is used for CT reconstruction calculation.
Projection images for incident beam current (I0)A projection image of X-ray beam without sample (I0 image). This is taken at regular intervals of sample imaging (ussually 5 or 10 intervals).
Coordination system
Projection images for transmitted beam current (I)A projection image of sample (I image). The projection , p=ln(I0/I), is calculated and used for CT reconstruction.
Sample rotation axisThe location of a sample rotation axis is required for CT reconstruction. This is called “center value”, which is generally described as a function of the height of a projection image (z: see the coordinate system). For example, the center value is expressed as follows: 249.934 -0.003046 × z ±0.320748( z = 0 - 219) .
PV (Pixel value)This is the value that expresses the brightness of each pixel of a digital image. For example, the pixel is black and white for PV=0 and 255, respectively, in a byte image.
VoxelOne of many from which a 3-D image is composed. The shape is usually cubic.
+X
+Y
+Z
(0,0,0)