Dr. Jerry Zweigenbaum Senior Applications Scientist Food Safety Minimize Method Development Time for Pesticide Screening Using LC/MS Application Kit LC/MS/MS Pesticide Analysis October 5,2010 Page 1

Upload: others

Post on 30-Jul-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Dr. Jerry Zweigenbaum

Senior Applications Scientist

Food Safety

Minimize Method

Development Time for

Pesticide Screening

Using LC/MS

Application Kit

LC/MS/MS Pesticide Analysis

October 5,2010Page 1

Page 2: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

LC/MS/MS TRIPLE QUADRUPOLE

Pesticide Application Kit with Dynamic MRM

Database for the

LC/MS/MS Pesticide Analysis

October 5,2010Page 2

SampliQ

Extraction and

Dispersive SPE Kits

Agilent 1260 LC

Agilent 6460 LC/QQQ

Page 3: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Methods for YOUR Pesticides!!!!

LC/MS/MS Pesticide Analysis

October 5,2010Page 3

Page 4: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Copy YOUR list to the Dynamic MRM Database and

Search

LC/MS/MS Pesticide Analysis

October 5,2010Page 4

Page 5: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Search on the METHOD and the Names to get RT

You will Use

LC/MS/MS Pesticide Analysis

October 5,2010Page 5

Page 6: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Database is Directly Linked to OPTIMIZER

LC/MS/MS Pesticide Analysis

October 5,2010Page 6

Page 7: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Many Columns and Info but may be Hidden

LC/MS/MS Pesticide Analysis

October 5,2010Page 7

Page 8: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Test Mix Positive Ion Mode: Quick Start

Verification

LC/MS/MS Pesticide Analysis

October 5,2010Page 8

Page 9: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Test Mix Negative Ion Mode

LC/MS/MS Pesticide Analysis

October 5,2010Page 9

Page 10: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Dynamic MRM Acquisition Parameters –

Customized Methods from Database

Page 10 October 5,2010

LC/MS/MS Pesticide Analysis

Page 11: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Dynamic MRM at the 10 pg level for 100 pesticides

1 min ΔR.T.- based acquisition window for each MRM transition

Page 11 October 5,2010

LC/MS/MS Pesticide Analysis

Page 12: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Dynamic MRM 11 Compounds (22 MRMs) eluting in about 1

min window

Compound Name Precursor Ion Product Ion Retention Time

Cinosulfuron 414.1 183 5.579

Cinosulfuron (Q) 414.1 157 5.579

Chlorotoluron 213.1 72 5.642

Chlorotoluron (Q) 213.1 140 5.642

Atrazine 216.1 174 5.682

Atrazine (Q) 216.1 132 5.682

Carbaryl 202.1 145 5.736

Carbaryl (Q) 202.1 117 5.734

Carboxin 236.1 143 5.836

Carboxin (Q) 236.1 87 5.836

Chlorsulfuron 358.0 167 5.896

Chlorsulfuron (Q) 358.0 141 5.896

Ethiofencarb 226.1 107 5.937

Ethiofencarb (Q) 226.1 164 5.936

Dodemorph 282.3 116 6.073

Dodemorph (Q) 282.3 98 6.074

Diuron 233.0 72 6.101

Diuron (Q) 233.0 160 6.101

Cyprodinil 226.1 108 6.245

Cyprodinil (Q) 226.1 93 6.246

Difenoxurone 287.1 123 6.509

Difenoxurone (Q) 287.1 72 6.509

Page 12 October 5,2010

LC/MS/MS Pesticide Analysis

Page 13: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Typical results: 10 pg Atrazine on-column using Dynamic MRM

6460A QQQ with Jet Stream Technology

•Avg. Signal Height: 15,650

•Avg. Signal Area: 50,966

•RSD: 3.2 %

•Estimated LOQ: 100 fg or less

•6-7 data points above FWHM

•3 sec FWHM, 6 sec @10% valley

•20 data points baseline-to-

baseline

••

•••

••

• • •

•• • •

3 sec FWHM

10% valley

Page 13 October 5,2010

LC/MS/MS Pesticide Analysis

Page 14: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Pesticides LC/MS method

Jet stream conditions

Drying gas temperature 275 ºC

Drying gas flow (nitrogen) 6 L/min

Nebulizer gas pressure (nitrogen) 35 psig

Capillary voltage 4000 V

Sheath gas temperature 325 ºC

Sheath gas flow 12 L/min

Nozzle voltage Off

6460A QQQ settings

MS1 and MS2 resolution Unit

Time Filtering peak width = 0.03 min

Dynamic MRM transitions up to 4000

Constant cycle time 373 ms

Delta EMV 400 V

1200 LC parameters

Column Eclipse Plus-C18, 2.1 x 100mm, 1.8 µm

Column temperature 35 ºC

Injection volume 5.0 mL

Autosampler temp 6 ºC

Needle wash flushport (MeOH:H2O 75:25), 5 secs

Mobile phase A = 0.01% formic acid in water

B =5 mM ammonium formate 0.01% formic acid in 95:5 acetonitrile:water

Flow rate 0.3 mL/min

Gradient 0 min 6% B

15 min 95% B

Stop time 20 min 6% B

Post time 1 min

Page 14 October 5,2010

LC/MS/MS Pesticide Analysis

Page 15: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Dynamic MRM of 300 Pesticides two Transitions

Each

October 5,2010Page 15

LC/MS/MS Pesticide Analysis

Run with 1200-SL LC 2.1 x 100 C18 Eclipse PLUS 1.8 u

Page 16: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

October 5,2010Page 16

LC/MS/MS Pesticide Analysis

Page 17: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Pesticide Mix run with Agilent 1290 Infinity 1260

Bar LC: Benefit of HIGHER PEAK CAPACITY

Page 17

LC 2.1 x 150 C18 Eclipse PLUS HD 1.8 u, 0.6 mL/min, 800 bar

October 5,2010

LC/MS/MS Pesticide Analysis

Page 18: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

3x10

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

1.8

1.9

2

2.1

2.2

2.3

2.4

2.5

2.6

2.7

2.8

2.9

3

1

Counts vs. Acquisition Time (min)0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5

2

3

4

5

6

7

8

9

10

11

12

IS

MRM chromatograms of 50 ng/g fortified sample processed by EN method. Peak identification: 1. Acephate, 2. Pymetrozine, 3. Carbendazim, 4. Thiabendazole, 5. Imidacloprid, 6. Imazalil, 7. Propoxur, 8. Carbaryl, 9. Cyprodinil, 10. Ethoprophos, 11. Penconazole, 12. Kresoxim-methyl, IS: TPP.

Column Agilent Poroshell 120 EC-C18, 2.1 × 100 mm, 2.7 μm

October 5,2010Page 18

LC/MS/MS Pesticide Analysis

Page 19: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

LC/MS/MS Pesticide Analysis

October 5,2010Page 19

Page 20: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

DYNAMIC MRM – CREATE MY

OWN

Using YOUR Chromatography

LC/MS/MS Pesticide Analysis

October 5,2010Page 20

Page 21: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Building a Dynamic MRM Acquisition Method

LC/MS/MS Pesticide Analysis

October 5,2010Page 21

Create a Dynamic MRM method

Import the results generated in the custom report into acquisition method

Specify delta retention time window

Analyze medium level standard mixtures

fixed dwell time for each MRM transition using one time segment

MHW Quantitative Data Analysis gets retention times

Load tabular MRM from Database

Compound names, ISTD (optional)MRM transitions, fragmentor voltages,

collision energies

Page 22: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Building a Dynamic MRM Method – 4.00

October 5,2010Page 22

LC/MS/MS Pesticide Analysis

Page 23: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Select Datafile and Parameters

LC/MS/MS Pesticide Analysis

October 5,2010Page 23

Page 24: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Acquisition is Automatically Populated with

Retention Times and RT Windows

LC/MS/MS Pesticide Analysis

October 5,2010Page 24

Page 25: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Next View the Method

LC/MS/MS Pesticide Analysis

October 5,2010Page 25

Page 26: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

October 5,2010Page 26

LC/MS/MS Pesticide Analysis

Page 27: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

View Method Provides Important Information in

Cohesive Format

October 5,2010Page 27

LC/MS/MS Pesticide Analysis

Page 28: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Finally – Set the Cycle Time

LC/MS/MS Pesticide Analysis

October 5,2010Page 28

Page 29: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Typical results: 10 pg Atrazine on-column using Dynamic MRM

6460A QQQ with Jet Stream Technology

•Avg. Signal Height: 15,650

•Avg. Signal Area: 50,966

•RSD: 3.2 %

•Estimated LOQ: 100 fg or less

•6-7 data points above FWHM

•3 sec FWHM, 6 sec @10% valley

•20 data points baseline-to-

baseline

••

•••

••

• • •

•• • •

3 sec FWHM

10% valley

Page 29 October 5,2010

LC/MS/MS Pesticide Analysis

Page 30: Minimize Method Development Time for Pesticide Screening ... · Pesticide Application Kit with Dynamic MRM Database for the LC/MS/MS Pesticide Analysis Page 2 October 5,2010 SampliQ

Summary

• Dynamic MRM Database provides quick method setup with

compound specific parameters and retention times

• Compounds not in method can be easily added

• New compounds can easily be added via optimizer to method

and database

• Custom methods can be easily developed and Dynamic MRM

readily created with “Update Method”

LC/MS/MS Pesticide Analysis

October 5,2010Page 30