TRANSCRIPT

Page 1: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

M.Jabari 2005M.Jabari 2005

Page 2: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Pipette

Graduated (Measuring )

To deliver (EX)

To contain (IN) (Wash with solution )

I Line

II Line

Volumetric

Page 3: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

% CV must be less than 0.3 %

% Bias must be less than 0.6%

Ideal % Bias 0 – 0.6% is acceptable

Page 4: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Expected value

Expected value

Observed value_

%Bias= X 100

(x-M)2

n - 1SD=

%CV = X 100MeanSD

Page 5: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

GRAVIMETRY METHOD

COLORIMETRY METHOD

Page 6: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

First weight a dish . Then with a pipette that you want caliber it

transfer DW into the dish and weight that . Repeat 10 time this action and record weight . Caculate each time pure weight of DW that

transfer with pipette. Caculate SD , %CV and % Biass .

Page 7: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

110.100.040.0016

210.150.090.0081

310.250.190.0361

410.120.060.0036

510.0600

610.090.030.0009

79.85-0 .210.0441

810.070.010.0001

99.99-0.070.0049

109.91-0140.0196

(X –mean)2

Mean = 10.06

(x-M)2

n - 1SD=

%CV = X 100Mean

SD

0.1191.0710.06

(x-M)2 = 0.119

= 0.16 = 0.4

= 0.04

Acceptable 0.04 < 0 .3

Page 8: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Mean of Expected weight

Mean of Expected weight

Mean of Observed weight

_

%Bias= X 100

110.100.040.0016

210.150.090.0081

310.250.190.0361

410.120.060.0036

510.0600

610.090.030.0009

79.85-0 .210.0441

810.070.010.0001

99.99-0.070.0049

109.91-0140.0196

(X –mean)2

0.1191.0710.06

10

10

10.06

%Bias= X 100

_

= 0.006

_

Acceptable 0.006 < 0 .6

Page 9: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Methyl orange 492 nmMethyl orange 492 nm

Phenol red 405 nmPhenol red 405 nm

Page 10: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

For 1 ml pipette : In a ballon class A . Prepare 0.1 gr /100 ml (stock ) . Dilute 1 : 11 and measure absorbance . Select 10 tube and add 10 ml DW in each

tube . (for 2 ml add 20 ml DW ) Add 1 ml stock reagent into each tube . Measur absorbance and record it .

Page 11: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Mean of Expected OD

Mean of Expected OD

Mean of Observed OD

_

%Bias= X 100

442

442

435

%Bias= X 100

_

= 1.58

_

1425-10100

2440525

3431-416

4430-525

5428-749

6441636

743500

8440525

9444981

1043611

(X –mean)2

3580435

Absorbance of 1:11 in class A ballone is 442

1.58 > 0.6No acceptable

Page 12: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Mean = 435

(x-M)2

n - 1SD=

%CV = X 100Mean

SD

(x-M)2 = 358

= 39 = 6.2

= 1.4

1425-10100

2440525

3431-416

4430-525

5428-749

6441636

743500

8440525

9444981

1043611

(X –mean)2

3580435 1.4 > 0.3No acceptable

Page 13: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

GRAVIMETRY METHOD

COLORIMETRY METHOD

Page 14: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Each sampler shoud assese 3 – 4 time annually .

Acceptable % CV up to 3%

Acceptable % Biass up to 5%

Page 15: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

First weight a dish .

Then with a sampler that you want caliber it transfer

DW into the dish and weight that .

Repeat 10 time this action and record weight .

Caculate each time pure weight of DW that transfer

with sampler .

Caculate SD , %CV and % Biass .

Page 16: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

1100-0.70.49

21021.31.69

31021.31.69

41032.35.29

599-1.72.89

699-1.72.89

71010.30.09

8100-0.70.49

9101030.09

10100-0.70.49

(X –mean)2

Mean = 100.7

(x-M)2

n - 1SD=

%CV = X 100Mean

SD

16.1100.7

(x-M)2 = 16.1

= 1.78 = 1.33

= 1.3

1.3 < 3Acceptable

Page 17: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Mean of Expected weight

Mean of Expected weight

Mean of Observed weight

%Bias= X 100

100

100

100.7

%Bias= X 100

_

= 0.7

_

1100-0.70.49

21021.31.69

31021.31.69

41032.35.29

599-1.72.89

699-1.72.89

71010.30.09

8100-0.70.49

9101030.09

10100-0.70.49

(X –mean)2

16.1100.7 0.7 < 5Acceptable

Page 18: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Methyl orange 492 nm

Phenol red 405 nm

Page 19: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

ll For 10-100 ml sampler : In a ballon class A . Prepare 0.1 gr /100 ml (stock ) . Dilute 1 : 101 and measure absorbance . Select 10 tube and add 1 ml DW in each tube .

( for 20 l ,50l ,100l add 2 , 5 , 10 ml DW ) Add 10 l stock reagent into each tube . Measur absorbance and record it .

Page 20: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Mean of Expected OD

Mean of Expected weight

Mean of Observed OD

_

%Bias= X 100

455

455

458.4

%Bias= X 100

_

= 0.74

_

14601.62.56

24612.66.76

34590.60.36

4458-0.40.16

5455-3.411.56

6458-0.40.16

74590.60.3684601.62.56

94590.60.36

10455-3.411.56

(X –mean)2

36.40458.4

Absorbance of 1:101 in class A ballone is 455

0.74 < 5Acceptable

Page 21: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

Mean = 458.4

(x-M)2

n - 1SD=

%CV = X 100Mean

SD

(x-M)2 = 36.4

= 4.04= 2.01

= 0.43

14601.62.56

24612.66.76

34590.60.36

4458-0.40.16

5455-3.411.56

6458-0.40.16

74590.60.3684601.62.56

94590.60.36

10455-3.411.56

(X –mean)2

36.40458.40.43 < 3Acceptable

Page 22: M.Jabari 2005. Pipette Graduated (Measuring ) To deliver (EX) To contain (IN) (Wash with solution ) I Line II Line Volumetric

ll

For more than 100 ml sampler :

In a ballon class A .

Dilute 1 : 11 from stock reagent .