© Hitachi High-Technologies Corporation All rights reserved. 2018[45] THE HITACHI SCIENTIFIC INSTRUMENT NEWS 2018 Vol.11 Model LA8080 High-Speed Amino Acid Analyzer SCIENTIFIC INSTRUMENT NEWS Technical Explanation Technical magazine of Electron Microscope and Analytical Instruments. 11 Vol. SEPTEMBER 2018 Fig. 1 Historical evolution of Hitachi’s amino acid analyzers Introduction 1 . Masahito Ito* 1 , Ikuko Narimatsu* 2 , Minling Pei* 1 , Atsuki Morisaki* 1 , Hiroshi Suzuki* 2 , Masato Fukuda* 1 , Takashi Yagi* 1 , Shigeo Otsuki* 1 , Kazuya Seki* 1 , Kosaku Toyosaki* 1 From the moment of the announcement of the first amino acid analyzer by D. H. Spackman, W. H. Stein, and S. Moore in 1958 1) , these instruments have been widely used to analyze the amino acid content of proteins and peptides and to study the amino acids and related substances present in pharmaceuticals, biological solutions, and other samples. When the original instrument was introduced, the use of ion-exchange chromatography and ninhydrin reactions— cutting-edge technologies at that time—made possible the analysis of amino acids with high selectivity, high speed, and high accuracy. Moreover, the successful use of automation techniques in the field of liquid chromatography paved the way for revolutionary new developments, establishing the introduction of amino acid analysis as a key milestone in the history of research. More recently, the development of large numbers of peptide-based pharmaceuticals has spurred a renewed appreciation for the value of this analytical technique, which is well-suited to meet the analytical needs of fields of research that have long placed a premium on high data reliability—including research and quality control for foodstuffs, animal feed, and health supplements, as well as pharmaceutical and taste research. It is against this backdrop that Hitachi’s line of amino-acid analyzers 2) has evolved in a continuous progression from 1962 to the present day (Figure 1). 1962 1971 1977 1986 1997 2005 Model KLA-2 Column (particle size 40 μm) Analysis time: 22 hours Model KLA-5 Column (particle size 17 μm) Analysis time: 2 hours Equipped with autosampler Model 835 Column (particle size 5 μm) Analysis time: 50 minutes Equipped with microcomputer Model L-8800 Control unit equipped with PC Uses reaction-column method Model L-8900 Uses the cartridge-type TDE² reactor Industry’s first Model L-8500 Column (particle size 3 μm) Ninhydrin reagent (2 solutions) Analysis time: 30 minutes

Upload: others

Post on 11-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Model LA8080 High-Speed Amino Acid Analyzer · amino acids with approximately 20 constituents, while lithium-citrate buffer liquids are used for amino acids with approximately 40

© Hitachi High-Technologies Corporation All rights reserved. 2018[45]THE HITACHI SCIENTIFIC INSTRUMENT NEWS ー 2018 Vol.11

1962

2005

1997

1986 1977 1971

Model LA8080 High-Speed Amino Acid Analyzer

SCIENTIFIC INSTRUMENT NEWSTechnical ExplanationTechnical magazine of E lectron Microscope and Analyt ical Instruments.

11Vol. S E P TE M B E R2 0 1 8

Fig. 1 Historical evolution of Hitachi’s amino acid analyzers

Introduction1.

Masahito Ito*1, Ikuko Narimatsu*2, Minling Pei*1, Atsuki Morisaki*1, Hiroshi Suzuki*2, Masato Fukuda*1, Takashi Yagi*1, Shigeo Otsuki*1,

Kazuya Seki*1, Kosaku Toyosaki*1

 From the moment of the announcement of the first amino acid analyzer by D. H. Spackman, W. H. Stein, and S. Moore in 19581), these instruments have been widely used to analyze the amino acid content of proteins and peptides and to study the amino acids and related substances present in pharmaceuticals, biological solutions, and other samples. When the original instrument was introduced, the use of ion-exchange chromatography and ninhydrin reactions—cutting-edge technologies at that time—made possible the analysis of amino acids with high selectivity, high speed, and high accuracy. Moreover, the successful use of automation techniques in the field of liquid chromatography paved the way for revolutionary new developments, establishing the introduction of amino acid analysis as a key milestone in the history of research. More recently, the development of large numbers of peptide-based pharmaceuticals has spurred a renewed appreciation for the value of this analytical technique, which is well-suited to meet the analytical needs of fields of research that have long placed a premium on high data reliability—including research and quality control for foodstuffs, animal feed, and health supplements, as well as pharmaceutical and taste research. It is against this backdrop that Hitachi’s line of amino-acid analyzers2) has evolved in a continuous progression from 1962 to the present day (Figure 1).

1962

1971 1977 1986

1997

2005

Model KLA-2 Column (particle size 40 μm)Analysis time: 22 hours

Model KLA-5 Column (particle size 17 μm)Analysis time: 2 hoursEquipped with autosampler

Model 835Column (particle size 5 μm)Analysis time: 50 minutesEquipped with microcomputer

Model L-8800Control unit equipped with PCUses reaction-column method

Model L-8900 Uses the cartridge-type TDE² reactor

Industry’sfirst

Model L-8500 Column (particle size 3 μm)Ninhydrin reagent (2 solutions)Analysis time: 30 minutes

Page 2: Model LA8080 High-Speed Amino Acid Analyzer · amino acids with approximately 20 constituents, while lithium-citrate buffer liquids are used for amino acids with approximately 40

© Hitachi High-Technologies Corporation All rights reserved. 2018[46]THE HITACHI SCIENTIFIC INSTRUMENT NEWS ー 2018 Vol.11

Key Features of the LA80802. The LA8080 high-speed amino acid analyzer is a dedicated instrument for ion-exchange chromatography via the method of post-column derivatization using ninhydrin reagents; its key features are the following 3 attributes (Figure 2).

(1) The LA8080 is comfortable.The ergonomic design of the instrument carefully accounts for the operator’s working lines and procedural motions. The use of a new technique for ninhydrin reagents—in which two liquids are blended immediately before the derivatization reaction—eliminates the need for refrigeration. Standard commercially available products may be used for the buffer liquid (eluent) and the reaction reagent.

(2) The LA8080 is compact.AminoSAAYA is the first instrument of its type to adopt a desktop configuration, an innovative redesign of previous standalone models. Moreover, the new design occupies some 30% less surface area than its predecessors. The instrument features convenience for storing solution bottles and a vial rack, and its front-access design facilitates operations such as exchanging columns and seals.

(3) The LA8080 is reliable.AminoSAAYA is compatible with the L-8800 and L-8900, to support many reliable data of ion-exchange chromatographic analytical methods. High-stability ninhydrin reactions ensure high precision in analyses of protein hydrolyzate and physiological fluid.

2

Fig. 2 The LA8080 high-speed amino acid analyzer

Page 3: Model LA8080 High-Speed Amino Acid Analyzer · amino acids with approximately 20 constituents, while lithium-citrate buffer liquids are used for amino acids with approximately 40

© Hitachi High-Technologies Corporation All rights reserved. 2018[47]THE HITACHI SCIENTIFIC INSTRUMENT NEWS ー 2018 Vol.11

Product Overview3. The columns of the LA8080 are packed with a strongly acidic cation exchange resin formed by introducing sulfonate groups into a polystyrene divinylbenzene-based polymer. As eluents, sodium-citrate buffer liquids are used for amino acids with approximately 20 constituents, while lithium-citrate buffer liquids are used for amino acids with approximately 40 constituents. For both eluents and reaction reagents, bottles of standard commercial products may be installed as-is, with no modification required. For both analytical methods, stepwise or gradient elution methods involving several types of buffer liquid are used to elute large numbers of constituents. Figure 3 depicts the flow-line configuration of the analyzer. Various eluents are selected by an solenoid valve, including regenerating liquid, pump 1 can deliver up to 6 liquids. To suppress chromatogram baseline noise, a one-stroke 10 μL micropump is used to reduce pressure pulsation. The autosampler can accommodate up to 120 test specimens in 1.5 ml vials. For cases required cold storage, such as specimen containing asparagine (AspNH2) or glutamine (GluNH2), the instrument is equipped with a cooling rack with capacity for 100 specimens. The column oven uses a Peltier element to maintain separation columns at constant temperature. The temperature may be varied over the range 20-90°C and is time-programmable. The two ninhydrin reagent liquids are mixed immediately before delivering. The various amino acid constituents eluted from the column mix with reagents delivered by pump 2 and are heated to 135°C by the reaction unit. A cartridge-type TDE3 (cubed) reactor method is used to reduce peak broadening to the greatest extent possible during liquid-mixture flow, as well as to ensure efficient reactions. Amino acids are converted into the reaction products known as Ruhemann’s purple3) and detected at the visible wavelength of 570 nm—where their light absorption is maximized—in the form of chromatogram peaks that may be measured and quantified. The imino acids proline and hydroxyproline do not exhibit absorption peaks in the visible regime, and are instead detected at 440 nm. Visible light at a wavelength of 700 nm is used as a detection reference, with compensating the fluctuations of a tungsten halogen lamp. As a chromatographic data system (CDS), the LA8080 uses OpenLAB CDS Ver. 2 to control the data-acquisition apparatus and perform qualitative and quantitative data analysis of chromatograms.

Nitrogen gascontrol unit

Reagent box

Detector

Reaction unit

Column oven

Autosampler

Ninhydrin reagent

Liquid delivering pump

*Waste tank accessory

Pump 1 (for eluent)

Ammonia filtering column #2650L

Autosampler

20 μL injected

570 nm440 nm

Separation column

20 -90℃

Pump 2 (for reaction

reagent)

Ninhydrin reaction reagent

Cleaning liquid

TDE3 reactor

125 -140℃

Column oven Reaction unit VIS detector

Citric acid buffer liquid

Regeneration solution

Nitrogengas

Fig. 3 Internal schematic and flow-line diagram of the LA8080

Page 4: Model LA8080 High-Speed Amino Acid Analyzer · amino acids with approximately 20 constituents, while lithium-citrate buffer liquids are used for amino acids with approximately 40

© Hitachi High-Technologies Corporation All rights reserved. 2018[48]THE HITACHI SCIENTIFIC INSTRUMENT NEWS ー 2018 Vol.11

4. Illustrative Analyses

 As a standard method of analysis, Figure 4 shows a chromatogram of protein hydrolyzate. A sodium-citrate eluent is transmitted at a flow rate of 0.40 ml/min, while a cation-exchange resin column #2622 (i. d. 4.6 mm × 60 mm) with 3 μm particle is maintained at a temperature of 57°C. Ninhydrin reagent is delivered at a flow rate of 0.35 ml/min; after producing color at a reaction temperature of 135°C, the absorption at the visible wavelength of 570 nm is detected. 20 μL of an amino-acid standard mixture is injected. After 30 minutes of analysis, resolution of 1.2 or greater were observed in all cases. For aspartic acid (Asp), the detection limit is 2.5 pmol or below (signal-to-noise ratio of 2) with an excellent peak-area repeatability (2 nmol) within 1.0% RSD.

 To analyze the amino-acid composition of proteins, one typically performs a hydrolysis treatment using hydrochloric acid; however, cysteine, cystine, and methionine are easily oxidized. For this reason, a widely used alternative method of analysis is to perform hydrochloric-acid hydrolysis treatment after oxidizing with performic acid, then analyze the resulting cysteic acid (CySO3H) and methionine sulfone (MetSON). We also present the results of a study in which CySO3H and MetSON alone were analyzed for short periods of time (Figure 5).

4

Pro

カラム: 3µm陽イオン交換樹脂4.6mmi.d.×60mm

溶離液:  0.40ml/min 4液クエン酸ナトリウム緩衝液+再生液

反応検出:  135℃ニンヒドリン反応

Asp

Thr

Ser

Glu

Ala

Gly Val

Cys

570nm

440nmMet

Leu Tyr

Phe

TrpNH3

Lys

His

Arg

Ile

0 15 30

440 nm

570 nm

Fig. 4 Sample analysis of protein hydrolyzate

Fig. 5 Performic-acid-oxidized protein hydrolyzate and an example of a short-time analysis

Column: 3 μm cation-exchange resin, 4.6 mm i. d. × 60 mmEluent: 0.40 ml/min 4-liquid sodium-citrate buffer liquid + RG solutionDetection method: 135ºC ninhydrin postcolumn derivatization

Cysteic acid, methionine sulfone

Short-time measurement

example

Page 5: Model LA8080 High-Speed Amino Acid Analyzer · amino acids with approximately 20 constituents, while lithium-citrate buffer liquids are used for amino acids with approximately 40

© Hitachi High-Technologies Corporation All rights reserved. 2018[49]THE HITACHI SCIENTIFIC INSTRUMENT NEWS ー 2018 Vol.11

5. Conclusions

 The continuing availability of the method introduced by Moore et al. may be attributed to at least two factors—one associated with the separation system, and one with the reaction-detector system. The match between polystyrene polymer ion-exchange resins and multi-constituent amino acids is excellent, while for neutral amino acids of aromatic species such as phenylalanine and tyrosine there is a hydrophobic interaction with the base polymer material that also contributes to achieving good separation. On the other hand, the method of post-column derivatization using ninhydrin has the key features of high selectivity with respect to contaminants in specimens; as long as one properly performs deproteinization and filtering, one can expect highly reliable analytical results. We are grateful indeed for the accomplishments of our predecessors in developing this outstanding method of analysis that has remained useful for many, many years.

About the authors

*1 Masahito Ito, Minling Pei, Atsuki Morisaki, Masato Fukuda, Takashi Yagi, Shigeo Otsuki, Kazuya Seki, Kosaku ToyosakiOptical Instruments Design DepartmentNaka DivisionHitachi High-Tech Science Corporation

*2 Ikuko Narimatsu, Hiroshi SuzukiApplied Technology DivisionBusiness GroupHitachi High-Tech Science Corporation

References

1)D. H. Spackman, W. H. Stein, S. Moore, Anal. Chem., 1190, 30 (1958).2)Masahito Ito, Bunseki, March Issue 2010, 145–146.3)Shinichi Ozawa, Hiroshi Miyano, Masahito Ito, S.I. NEWS, Vol. 58 No. 1, 4968–4977 (2015).4)Masahito Ito, Wakojunyakujiho, Vol. 86 No 3, 15–17 (2018).