Water Vapour Permeation Analysers7000

Barrier films PET bottles Containers Closures Flexible Pouches Bags

Applications

Features & Benefits

Modular systems for precision water vapour analysis of packaging film barriers

Analytical Systems Manufacturedtraceable to NIST.

System validation with certified gas orfilm for speed and convenience.

Over 25 yrs experience of ProprietaryCoulometric P2O5 sensor.

Absolute moisture measurement - Nocalibration required.

Wide measurement range

Flow, temperature and humidity controlfor ultimate responsiveness andrepeatability.

Intuitive Windows based software.

For medical and pharmaceuticalpermeation testing, we can now offersoftware which conforms to 21CFRPart 11.

No liquid coolants, catalysts or specialgas mixtures required.

Conforms to: ASTM F-1249* ISO 15106-3 ISO 15105-2 DIN 53122-2

Page 2: Modular systems for precision water vapour … · Modular systems for precision water vapour analysis ... technology to detect water vapour transmission ... headspace analysers for

Our test laboratory will perform your Permeation Testing and Headspace Analysis. Whetheryou are developing innovative materials and packages or validating that your supplier ismeeting specification. We can exceed your expectations with:

Competitive Prices

Fast Turnaround

Independent non-biased results

30 Years Experience

Laboratory Testing Services

All Systech Illinois analysers are certified traceable to NIST. In addition, analyticalperformance is validated using NIST certified gases and NIST traceable films. A set ofvalidation films and a spare P2O5 sensor comes as standard with all of the water vapourpermeation analysers.

These analysers offer precision temperature and humidity flow control providing ultimateresponsiveness and repeatability.

Manufacture traceable to NIST

The two most common sensor types for measuring moisture are IR (infra-red) and P2O5

(phosphorous pentoxide). The P2O5 sensor is more sensitive and stable than IR and does notrequire calibration. P2O5 is the primary method for absolute moisture measurement. SystechIllinois have over 25 years experience of using P2O5 sensor technology.

*The Systech Illinois 7000 analysers comply with ASTM standard F-1249 with the exception ofthe sensor technology. The standard relates to an infra-red sensor whilst the Systech Illinoisanalyser uses a coulometric sensor - a dedicated method of moisture analysis.

P2O5 Sensor Technology

Precision control

The 7000 series offer a wide measurementrange providing research grade flexibility.

Measurement range of 0.002-1000 g/m2/day with masking.

Quality Assurance orientatedspeed and agility.

Up to five expansion modulesavailable to increase testingthroughput.

Test gas and carrier flow gas controlledby premium electronic mass flowcontrollers.

Wide sample temperature range of 5 to50°C (41 to 122°F).

Accurate relative humidity range from20% to 90%.

Wide measurement range

Page 3: Modular systems for precision water vapour … · Modular systems for precision water vapour analysis ... technology to detect water vapour transmission ... headspace analysers for

Start up

Sample Test Chamber

Actual data

The intuitive Windows basedsoftware offers:

Easy input and recall of operating parameters and test protocols.

User-friendly data tracking, searches, sorts, storage and output capabilities.

Graphical representation of measurement data in real time.

Auto-stop feature stops test when samples have reached equilibrium or by user entered elapsed time value.

Complete system diagnostics.

Utilising our proprietary sensortechnology to detect water vapourtransmission rates, samples areclamped or attached to a diffusionchamber. Wet nitrogen is thenintroduced into the upper half of thechamber while a moisture-free carriergas flows through the lower half.

Molecules of water diffusing throughthe sample into the lower chamber areconveyed to the sensor by the carriergas.

This allows a direct measurement ofthe water vapour without usingcomplex extrapolations. Water vapourtransmission rate of the test sample isdisplayed as either g/m2/day org/100in2/day.

To achieve an absolute measure, thetechnology draws upon a fundamentalprinciple of physics.

The phosphorous pentoxide (P2O5)moisture sensor consists of a dualplatinum winding formed around aquartz tube.

The change in the resistance acrossthe windings creates a change in themeasured current. According toFaraday’s Law this is directlyproportional to the amount of moisturein the gas stream.

Software

Principle of Operation

TEST FILM

VENT

N2/H2O OUT

WET N2 IN

DRY N2 IN

Set up

Quartz tube

PlatinumWindings

P2O5

Coating

P2O5 Sensor

Page 4: Modular systems for precision water vapour … · Modular systems for precision water vapour analysis ... technology to detect water vapour transmission ... headspace analysers for

Illinois Instruments, Inc (U.S)2401 Hiller Ridge RoadJohnsburg, Illinois 60051U.S.ATel: +1 815 344 6212Fax: +1 815 344 6332E-mail: [email protected]

Systech Instruments Ltd (UK)17 Thame Park Business Centre,Wenman Road, Thame, Oxfordshire OX9 3XATel: +44 (0)1844 216838 Fax: +44 (0)1844 217220E-mail:[email protected]

Illinois Instruments (Thailand)26/6 Ladprao 23, Jatujak, Bangkok 10900 ThailandTel: +66 (0)2030 5851Fax: +66 (0)2030 5850E-mail: [email protected]

Systech Illinois (China)Room 1107-1108 Forte BuildingNo. 910 Quyang Rd, Hongkou districtShanghai, China 200437Tel: +86 21 65533022Fax: +86 21 65539651Email: [email protected]

Measurement Range

7001 Unmasked 0.002 to 10 g/m2/dayMasked 0.02 to 70 g/m2/dayPackage 0.00001 - 0.05 g/pack

7002 Unmasked 0.002 to 70 g/m2/dayMasked 0.02 to 1000 g/m2/dayPackage 0.00001 - 0.05 g/pack at 10 cc flow,

up to 0.5 g/pack at 50 cc flow

Test Temperature Range 5°C to 50°C (41°F to 122°F)

Test RH Range 20 to 90% RH

Sample Size 50cm2, adaptors available for smaller samples

Operating Conditions Standard laboratory environment

Power Requirements 100-240 VAC, 50/60Hz, 840 VA (max)

Supply Pressure 1.7 bar regulated

Gas Fittings 1/8 in. Swagelok (supplied)

Enclosure Epoxy coated heavy gauge steel

Dimensions 533 x 533 x 305 (mm)

Weight 23.6kg

Options

Expansion Modules 7011 Available for simultaneous WVTR analysis of up to 12 samples

Environmental Chambers For finished package testing

Systech Illinois have over 30 years experience of providing analysis solutions for a wide rangeof industries. From our manufacturing plants in the UK and U.S. we produce gas analysers forindustrial process industries, headspace analysers for monitoring gas flushing of food products,and our range of permeation analysers.

Systech Illinois reserve the right to change specifications without notice. 05/2017

Technical Specifications

Systech Illinois’ range meets the requirement for the testing of any application.

Systech Illinois’ 7002 Water VapourTransmission Analyser delivers the same highperformance as the 7001 but with an extendedmeasurement range for more demandingapplications.