mri relaxation measurements in rats and mice · mri relaxation measurements in rats and mice...

30
MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University of Gothenburg Department of radiation physics

Upload: others

Post on 31-May-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

MRI relaxation measurements in rats

and mice

Master’s Thesis, 30hp

Jesper Lindberg

Spring 2008

Supervisor:

Michael Horn

University of Gothenburg

Department of radiation physics

Page 2: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Abstract

The work presented focus on the measurement of proton relaxation times, T1 and T2for organs in mice and the brain in rat at 7 T. The knowledge of T1 and T2 relaxationtimes are necessary for optimizing contrast in imaging methods like functional MRI andperfusion imaging. Relaxation times can be an important way to investigate propertiesof tissues and separate normal tissue from diseased, e.g. cancerous tissue.

The pulse scheme used for the relaxation measurements is IR-TrueFISP, a sequencewhich allows measuring both T1 and T2 within the same scan.

Initial experiments were done in a phantom to investigate the behavior of the pulsesequence. Parameters tested were number of averages, number of frames, ROI size, multi-slice acquisition, intervariability and slice orientation dependencies. Conclusions from thephantom measurements were that ROIs should be larger then 0.27 cm2 and single slicemeasurements are preferred.

Relaxation measurements for mice were performed in the strain C57/BL6. Measure-ments were performed in kidneys, front leg muscle and liver. Relaxation times were(Organ:T1:T2 in ms) kidneys:1559:54, front leg muscle:1190:39 and liver:985:60. Furthermeasurements were performed to investigate the ability of the method to discriminatewhite and brown adipose tissue by relaxation time. Differences could be seen, however,further measurements are needed to establish the method for discrimination of the fatdepots.

Measurements in the rat brain were performed in the strain Sprague Dawley. Re-gions measured were cerebellum, medulla oblongata, cerebral cortex, hippocampus, thala-mus and corpus callosum. Relaxation times were (location:T1:T2 in ms) in cerebel-lum:1343:80, medulla oblongata:1560:68, cerebral cortex (right side):1246:76, cerebral cor-tex (left side):1635:78, hippocampus (right side):1249:171 and hippocampus (left side):1072:254.

Problems that occurred and in some extend affected the results were the lack of trig-gering, eddy current compensation, and magnetic inhomogeneities.

Page 3: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Contents

1 Introduction 4

2 MR theory 4

2.1 Behavior in an external magnetic field . . . . . . . . . . . . . . . . . . . . . . . 4

2.2 Measuring the net magnetization . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2.3 Relaxation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2.4 Inversion recovery technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

2.5 Gradient, slice selection and image encoding . . . . . . . . . . . . . . . . . . . . 7

2.6 Echo creation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2.6.1 Gradient echo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2.6.2 Spin echo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

2.7 Relaxation measurements with IR-True-FISP . . . . . . . . . . . . . . . . . . . 9

3 Materials and Methods 10

3.1 Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

3.1.1 Number of averages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

3.1.2 Number of frames . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3.1.3 Size of the ROI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3.1.4 Single vs. multislice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3.1.5 Intervariability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

3.1.6 Slice orientation dependencies . . . . . . . . . . . . . . . . . . . . . . . . 12

3.2 Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

3.2.1 Organs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

3.2.2 Brown and white adipose tissue . . . . . . . . . . . . . . . . . . . . . . . 13

3.3 Rat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

4 Results 14

4.1 Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

4.1.1 Number of averages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

4.1.2 Number of frames . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

4.1.3 Size of the ROI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

4.1.4 Single vs. multislice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

4.1.5 Intervariability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

4.1.6 Slice orientation dependencies . . . . . . . . . . . . . . . . . . . . . . . . 18

4.2 Mouse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Page 4: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

4.2.1 Organs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

4.2.2 Brown and white adipose tissue . . . . . . . . . . . . . . . . . . . . . . . 19

4.3 Rat . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

5 Discussion 20

5.1 Sequence evaluation with phantom experiments . . . . . . . . . . . . . . . . . . 20

5.2 Relaxation measurements in mice . . . . . . . . . . . . . . . . . . . . . . . . . . 22

5.2.1 Relaxation measurements in organs and anatomical structures . . . . . . 22

5.2.2 Brown and white adipose tissue . . . . . . . . . . . . . . . . . . . . . . . 23

5.3 Relaxation measurements in rat brain . . . . . . . . . . . . . . . . . . . . . . . 23

Acknowledgement 25

References 26

Appendix 28

Page 5: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

1 Introduction

The big advantage of MRI compared to other imaging techniques is the fact that it’s non-

invasive, opposite to x-ray no ionizing radiation is used. MR is a great tool for investigating

living systems, this means that animal studies can decrease the number of animals, no need

for dissection when a MR image could give the same information. MR has the ability to

make images of soft tissues, both structure and function. Images can be acquired with high

resolution and even in 3D. Another benefit is that the image slices can be placed in any

direction. [1]

Measurement of relaxation times in vivo is an important way to investigate properties of

tissues, one way to separate normal tissue from cancerous tissue. [2] The knowledge of T1 and

T2 relaxation times are necessary for optimizing contrast in imaging methods like functional

MRI and perfusion imaging. [3]

The aim of this study was to measure relaxation times (T1 and T2) of the organs in mice

and the brain in rats. We investigated as well if there are any differences in relaxation times

of white- and brown adipose tissue on mice.

2 MR theory

MRI, magnetic resonance imaging is derived from NMR, nuclear magnetic resonance, hence

the properties of the nucleus in the magnetic field is the importance. Nuclei used in MRI are

proton (1H), carbon (13C), and phosphorus (31P). The nuclei must have a non-homogenous

charge distribution.

All fundamental particles posses a property called spin, which can be explained as a ro-

tation of the nucleus around its own axis. Due to spin and the non-homogenous charge

distribution (i.e. a moving charge) the nucleus produces a magnetic moment.

MRI signal depends on the number of nucleus and their magnetic moment, therefore the

best choice (for regular imaging) is the proton, 1H, because of its high abundance and its high

(compared to other nucleus) magnetic moment. [4, 5, 6]

2.1 Behavior in an external magnetic field

The proton behaves like a small bar magnet when it’s placed in a static magnetic field, B0.

It aligns almost parallel to the direction of the field (i.e. the same way a compass aligns to

the magnetic field of the earth). Because of the spin and the laws of quantum mechanics

the behavior is a bit different from a compass needle. The proton does not align perfectly to

the field. Due to the torque it experience from the magnetic field it will precess around the

4

Page 6: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

field axis, as shown in figure 1. The frequency of precession (known as the Larmor frequency)

can be derived from both classical and quantum mechanics. The Larmor frequency is given

in equation 1, where ω is the Larmor frequency, γ is the gyromagnetic ratio and B0 is the

amplitude of the Magnetic field (for derivation see appendix).

Figure 1: The proton aligns parallel or anti-parallel and precess in an magnetic field. Image retrievedfrom http://www.easymeasure.co.uk/ [7]

ω = 2πν = γ · B0 (1)

In an external magnetic field the numbers of possible values for the angular momentum are

2I +1. The 1H have a spin quantum number, I = 12 . That means it can have two orientations

(energy levels) in the field, spin up or spin down (i.e. parallel and anti-parallel to B0). These

two energy levels have slightly different energies. Population of the two levels is dependent

on both magnetic field and temperature. At 37 ◦C and 1.5 T there is a ratio of 1.000004, so

slightly more protons with spin up. The average of all proton magnetization will give a net

magnetization aligned exactly along the B0 field.

It’s possible to induce transitions between the energy levels. This can be achieved by

an additional oscillating magnetic field, B1 perpendicular to the static field, generated by a

radiofrequency (RF) pulse. The frequency of the RF pulse must match the energy difference

between the two states of the proton. The relation between energy and frequency, de Broglies

equation is given in equation 2.

∆ε = � · ω (2)

Where ∆ε is the energy difference, � is Planck’s constant divided by 2π and ω is the

associated frequency. [4, 5, 6]

5

Page 7: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

2.2 Measuring the net magnetization

The net magnetization is flipped perpendicular to B0 field, this makes it possible to measure

the magnetization. Applying an RF pulse will make the magnetic moment of the protons to

change state, due to the higher number of spin-up protons more of them will change state and

the net magnetization will decrease. With the right strength and duration of the RF pulse the

net magnetization will be zero in the longitude view, this is called a 90◦ pulse. Fortunately

protons get into phase with each other during the RF pulse, otherwise the net magnetization

would be zero in the transverse plane too. So the RF pulse makes the net magnetization to

flip down from the z-axis and make the protons precess in phase.

Magnetization in the transverse plane will generate a current in the receiver coil (according

to Faradays law). Over time the signal decays while the net magnetization return to its

equilibrium state, as shown in figure 2. The induced signal is called the free induction decay

(FID). [4, 5, 6]

Figure 2: The transverse magnetization induces a current in the receiver coil and the signal is collected.And overtime the signal decays. Image retrieved from http://www.easymeasure.co.uk/ [7]

2.3 Relaxation

There are two mechanisms that make the net magnetization return to its equilibrium value,

spin-lattice relaxation, spin-spin relaxation.

The return of magnetization in longitude direction (for example after a 90◦ pulse) is called

spin-lattice-, longitudinal- or T1 relaxation. Spin-lattice relaxation is loss of energy from the

spinning nuclei to the surroundings (lattice). This relaxation is characterized by the time

constant, T1. T1 is defined as the time between a complete 90◦ pulse and the relaxation back

to 63 % (1-1/e) of its original value. This means that protons with different surroundings will

relax with different T1 times and give a contrast between different compositions. T1 has a

great dependency of the magnetic field strength. [4, 5, 6]

The return of the transverse magnetization is called spin-spin-, transverse- or T2 relaxation.

The name spin-spin comes from the exchange of energy between the nucleuses. Spin-spin

relaxation is loss of phase coherence, this is due to inhomogeneities of the magnetic field.

6

Page 8: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

This inhomogenities descends from both the protons itself and other magnetic active nuclei

(molecular interactions) and from variations in the B0 field. The time constant, T2* (T2 star)

that characterize the spin-spin relaxation is defined as the time between a complete 90◦ pulse

and a decay to 37 % (1/e) of its original value. The time constant T2 have the same definition

as T2* aside from that it not includes the B0 variations. T2* is always shorter or equal T2.

[4, 5, 6, 8]

2.4 Inversion recovery technique

In the beginning of the pulse sequence the net magnetization is flipped by a 180◦ pulse. The

net magnetization will immediately start to recover and after a time TI (time of inversion) a

90 ◦ pulse is applied, see figure 3. IR (inversion recovery) is used to measure the T1 relaxation

time or to enhance contrast in T1-weighted images. Furthermore, it can also be used for fat

suppression, when the TI is equal the time of relaxation between the 180◦ and 90◦ for fat there

will not be any net magnetization to measure. [4, 6]

Figure 3: This image shows an inversion recovery sequence with spin echo. At first the RF pulse flipsthe net magnetization by 180◦, after a time TI the 90◦ pulse flips the net magnetization again. Thenext RF pulse is needed for the echo creation.

2.5 Gradient, slice selection and image encoding

In order to produce small variations in B0, gradient coils are implemented in a MRI system

(in three spatial dimensions). These variations are used to locate where the acquired signal

originates.

The slice selection gradient, Gz is a linear gradient applied orthogonal to the desired slice.

When the slice selection gradient is turned on, the effective magnetic field at the position z is:

Beff = B0 + Gzz (3)

7

Page 9: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

And the resonance frequency is then:

νeff =γ

2π(B0 + Gzz) (4)

Thus the resonance frequency of the nuclei is dependent of the position along the z-axis.

Hence the slice thickness is determined by the bandwidth of the RF-pulse. The slice thickness

is given in equation 5. [4, 6]

∆z =2π∆ν

γGz(5)

Next gradient to be applied is the phase-encode gradient, Gx, which is oriented perpen-

dicular to the slice gradient. When the gradient is turned on, protons positioned where the

gradient strength is highest (i.e. at the highest magnetic field), will precess faster then the

protons on the other side and the protons at the middle precesses with the same frequency

as before. After the gradient is turned off all protons within the slice returned to the original

larmor frequency, but because of the gradient they have now shifted phases in different extent.

Protons in the high magnetic region is a little ahead the ones in the middle and protons in

the lower magnetic region are a little bit behind. This means they are positioned in columns

precessing with different phases. [5]

The final gradient is the frequency-encode gradient, Gy, (also known as the read gradient)

this one is applied perpendicular to both slice and phase gradient, to form a three dimensional

coordinate system. The gradient is turned on while the signal is collected at the receiver coil.

Now a single line in frequency domain (k-space) is collected. To collect all lines needed for

the image it’s necessary to start over again with a new phase-encode gradient followed by the

read gradient which is repeated until all lines are collected.

After acquiring of data of k-space in the frequency domain, a Fourier transformation is

performed to yield data in time domain. [5]

2.6 Echo creation

The signal collected in the receiver is not the real FID, as the gradients need time for switching,

thus an echo of the FID is acquired. The creation of this echo can be done in two ways, spin

echo or gradient echo.

2.6.1 Gradient echo

After the initial RF pulse a negative gradient is applied, this will dephase the transverse

magnetization. Next a positive gradient is applied and due to this the net magnetization will

8

Page 10: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

rephase until maximum and then dephase again. The gradient echo does not compensate for

dephasing according to field inhomogeneities or spin-spin relaxation. In other word, the decay

of the echo is dependent on T2*.

2.6.2 Spin echo

The spin echo sequence starts with an RF pulse, making the net magnetization to flip 90◦.

Initially the spins are all in phase and are now allowed to dephase. After a time TE/2 (time

of echo divided by two) a 180◦ pulse is applied. This pulse will reverse the phase angle of the

spins, it will make the spins refocus back and form an echo (i.e. dephase backwards). The

echo will appear at a time equal the time between the initial RF pulse and the 180◦ pulse (i.e.

at the time TE). Field inhomogeneities have less effect in spin-echo, assuming the proton is

in the same position at all times of the measurement.

2.7 Relaxation measurements with IR-True-FISP

FISP is an acronym for Fast Imaging with Steady-state Precession. When all gradients used

in the sequence are rewound the sequence is called True-FISP, also known as fully rewound

gradient echo. Steady state imaging means that net magnetization reaches a steady state after

some RF pulses, i.e. the decrease of net magnetization due to excitation is the same as the

relaxation. When the sequence starts with an inversion recovery module we get IR-TrueFISP.

[6, 3]

IR-TrueFISP has the possibilities to measure both T1 and T2 in the same scan. The signal

intensity of an IR-True-FISP sequence is given in equation 6, where Sstst is the steady state

signal, for large inversion times, INV is the ratio of initial (S0) and steady state signal, Sstst.

T1* is the apparent T1 relaxation, T1* describes the recovery back to steady state. [3]

S(t) = Sstst ·(1 − INV · e(−t/T1∗)

)(6)

Schmitt et al. [9, 10] showed that the relaxation times can be calculated with equations

7 and 8. The T1* can be calculated by fitting equation 6 to a series of IR-TrueFISP images

with different TI times.

T1 = T1∗ cosα

2(INV − 1) (7)

T2 = T1∗ sin2 α

2

(1 − cos α

2

INV − 1

)−1

(8)

9

Page 11: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

3 Materials and Methods

All measurements were done on a Bruker MR system, (Bruker BioSpin, Ettlingen, Germany)

located at Center for Mouse Physiology and Bio-Imaging (CPI), with a magnetic flux density

of 7 T. Acquisition of images was done with Paravision 4.0 (Bruker BioSpin). Slice thickness

of 1mm was selected to avoid partial volume artifacts. Minimum pixel size is dependent on

organ size.

Determination of SNR (signal to noise ratio) was done according to equation 9, where

Smean is the mean signal in the ROI (region of interest) placed on the measured object and

Nmean is mean signal of the ROI placed outside of the measured object.

SNR =Smean

Nmean(9)

Pulse scheme used for relaxation measurements were IR-TrueFisp (FISP-T1+T2-map),

with help of the macro IrTrueFispMap (supplied by Bruker). The macro generates a phase

corrected series of images. On the generated images a ROI is placed, with use of the ISA tool

(Image Sequence Analysis) a curve fit is performed and the values of T1 and T2 are given for

the selected ROI.

3.1 Phantom

For transmitting radiofrequency pulses and receiving MR data a volume coil, inner diameter

of 72 mm was used (Bruker BioSpin, Ettlingen, Germany). All measurements on the phantom

were done to optimize the measurement time, investigate the behavior of the sequence and to

learn how to handle the system. Scan time is preferred to be as short as possible in animals

due to anesthesia. A homemade phantom was used, built of five small plastic tubes containing

water with different concentrations of agarose and gadolinium, see figure 4. Scan repetition

time must be kept higher than 5 times T1, because the signal needs to recover. A couple of

parameters (i.e. averages, frames, ROI size, multiple slices, intervariability and slice selection)

were altered to see there effect on the measured value. The different variation schemes are

given below (3.1.1 to 3.1.6).

3.1.1 Number of averages

A sequence with TE: 1.5 ms TR: 3 ms Scan Repetition time: 10 s and matrix 128x128 was

run multiple times with different number of averages (1, 2, 3, 4 and 8). The scan was repeated

three times with different order (i.e sequential, reversed and mixed) there was also a delay

between the scans of 1 h, to avoid effects from heating up the phantom. T1 and T2 were

10

Page 12: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Figure 4: The phantom used for sequence evaluation.

measured for three different ROIs. More averages will lead to better SNR.

3.1.2 Number of frames

To test how many frames (i.e. number of points on the curve) needed for a sufficient curve fit

a sequence with TE: 1.5 ms, TR: 3.0 ms, scan repetition time: 10 s, number of averages: 4,

size of matrix: 128x128 was run three times with different number of frames (20, 40 and 60).

T1 and T2 were measured for three different ROIs.

3.1.3 Size of the ROI

To investigate if the size of the ROI has any influence in the result one image with TE: 1.5

ms, TR: 3.0 ms, scan repetition time: 10 s, number of averages: 1, size of matrix: 128x128

was acquired. T1 and T2 were measured 8 times with different size of the ROI (in cm2 0.0005,

0.0022, 0.0088, 0.0352, 0.1100, 0.27, 0.42, 0.64), see figure 5

3.1.4 Single vs. multislice

In the Bruker manual [3] it’s recommended to run only a single slice at the time. Testing how

much it affected the results a series of measurements were done with different number of slices

(1, 3, 4, 5, 20) and also with and without gap in between them. T1 and T2 were measured at

the same position and in the most middle slice.

3.1.5 Intervariability

To test if the results are stable over time a scan with TE: 1.5 ms, TR: 3.0 ms, scan repetition

time: 10 s, 4 averages and size of matrix 128x128 was repeated 10 times with 1 hour in between

the measurements. Relaxation times were measured at two positions.

11

Page 13: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Figure 5: Different sizes of ROIs were tested. The ROIs were placed centered to each other in thesame position of the phantom.

3.1.6 Slice orientation dependencies

To ensure that the results were not dependent on slice orientation a scan with TE: 1.5 ms, TR:

3.0 ms, scan repetition time: 10 s, 4 averages and size of matrix 128x128 was repeated 3 times

with different slice orientations (axial, sagittal and coronal). The phantom used was different

from the other measurements, this one contained water, gadolinium and sodium chloride in a

single compartment.

3.2 Mouse

The mouse strain used was C57/BL6. Initial measurements were done on dead mice, to learn

the anatomy. Relaxation measurements were done in vivo on living mice.

As the system could not be triggered in senseful mode, measurements of relaxation times

in organs which are moved during respiration and/or cardiac cycle could not be established.

3.2.1 Organs

First a multi slice multi echo scan was acquired for orientation purposes. Thereafter FISP

images (TE: 1.5 scan repetition time: 10 s, averages 4 and matrix 128x128) for each individual

organ (kidney, liver and front leg muscle) were acquired.

12

Page 14: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

3.2.2 Brown and white adipose tissue

The aim of this measurement was to investigate if there are any differences in relaxation times

between brown and white adipose tissue. The differences should be used to discriminate fat

depots for questions within obesity research. The coil used was a 50 mm quadrature volume

coil (Rapid Biomedical, Rimpar, Germany), used for both transmit and receive. The mouse

were in supine position with head first. Brown adipose tissue is located in the back between

the scapulae, so supine position reduces motion artifacts.[11] A MSME (multi slice multi

echo) sequence were used to locate the fat deposit and then a 3 slice, 4 averages FISP image

with TE of 1.7 ms and a scan repetition time of 10 s, matrix size 64x96 and a resolution of

0.0235x0,0302 cm/pixel was acquired. Relaxation times were measured at different positions

within the fat deposit, see figure 6.

Figure 6: How the ROIs where placed for the differentiation of brown and white adipose tissue.

3.3 Rat

Aim of this experiment was to measure relaxation times of different regions of the rat brain,

i.e. (cerebellum, medulla oblongata, thalamus, hippocampus, corpus callosum and cerebral

cortex). Rat strain used was Sprague Dawley and all measurements were done in vivo. Coils

used were a 72 mm volume coil for transmission and a quadrature rat brain coil (Rapid Bio-

medical, Rimpar, Germany) for receiving. First a 20-slice, high resolution (in plane resolution

13

Page 15: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

of 50x68 µm/pixel), T2-weighted, GE image was acquired for orientation purposes. FISP

images were acquired at appropriate positions with guidance from the high resolution image

and Paxinos and Watsons brain atlas [12].

4 Results

4.1 Phantom

4.1.1 Number of averages

Number of averages affect the SNR as well as the relaxation times as shown in Figure 7 for

T1 and in Figure 8 for T2. On T1 there is no big variation at medium (5.9 %) and high (3.8

%) SNR, in the low SNR an increase (10.2 %) of T1 time is seen after 2 averages.

There were a big variation of the T2 values at the low SNR (35.6 %). At the medium SNR

the variation were 12.7 % between the highest and lowest value. The high SNR value had a

variation of 10.7 %. The values of SNR is shown in Table 1. The increase of SNR between 1

and 8 averages were approximate 60 % for all ROIs.

Figure 7: T1 relaxation time as a function of number of averages.

Table 1: Signal to noise ratio are changing with number of averages.Averages Low SNR Medium SNR High SNR

1 1.4 4.1 5.22 1.8 5.7 7.13 2.2 7.8 94 2.5 8.6 10.68 3.2 11.0 13.4

14

Page 16: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Figure 8: T2 relaxation time as a function of number of averages.

4.1.2 Number of frames

Result of the number of frames are shown in figures 9 and 10. The T1 at high SNR is almost

a horizontal line, medium SNR is quite flat with a 11.3 % increase at 60 frames and the low

SNR have a big decrease (55.3 %) with an increased number of frames. T2 times at high and

medium SNR follow the same trend, an increase between 20 and 40 frames and a decrease

between 40 and 60 frames. The T2 values in the low SNR areas are increasing (28.7 %) with

number of frames.

Figure 9: T1-relaxation time as a function of number of frames.

15

Page 17: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Figure 10: T2-relaxation time as a function of number of frames.

4.1.3 Size of the ROI

The relaxation time as a function of ROI size is shown in figure 11. Some variations (3.9 %

for T1 and 12.2 % for T2) at the smallest ROIs and with larger ROIs a small decrease (8.7 %)

of the T1 times and an increase (17.4 %) of the T2 times.

Figure 11: T1 and T2 relaxation times as a fuction of ROI size.

4.1.4 Single vs. multislice

Results of measurements with different number of slices are shown in figures 12 and 13. The

high SNR T1 is decreasing with more slices and narrower gaps. Medium and low SNR show

a decrease between 1 and 5 slices with the large gap and then an increase with 5 slices (0.05

16

Page 18: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

mm gap) and 4 and 20 without gaps. The T2 time is decreasing with more slices and narrow

gaps for the low SNR, while the high SNR have an increase of time between 1 and 5 slices

with the large gap and decrease with reduction of gap size or number of slices. No values for

the medium SNR were achieved due to too long relaxation time.

Figure 12: T1-relaxation time as a function of number of slices.

Figure 13: T2-relaxation time as a function of number of slices.

4.1.5 Intervariability

Results from the intervariability measurement are shown in figure 14. T1 for the first ROI

is stable within 4.7 % between scan number 1 and 7, between scan 1 and 10 the variation is

15.7 %. T1 for the second ROI is stable within 0.5 % between the first 7 scans and have a

variation of 12.2 % between scan 1 and 10. There was no times calculated for T2 on the first

ROI until scan number 9 and the variation between scan 9 and 10 are 9.1 %. T2 times for

the first ROI was stable within 1.3 % between scan 1 and 7 and between scan 1 and 10 the

17

Page 19: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

variation was 37.6 %.

Figure 14: Intervariability of relaxation time measurement.

4.1.6 Slice orientation dependencies

Results from the slice orientation dependencies measurement are shown in figure 15. There

were only small variations between axial, sagittal and coronal slices in T1 (1.6 %). The T2

values between axial and sagital has a variation of only 0.4 % but between sagital and coronal

the variation were 9.7 %.

Figure 15: Slice orientations dependecies on relaxation times.

18

Page 20: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

4.2 Mouse

4.2.1 Organs

The measured relaxation times for the organs in the mouse are given in Table 2. The shortest

T1 value is seen in the liver and the highest in the kidneys. The shortest T2 value is seen in

the front leg muscle and the highest in the liver.

Table 2: Relaxation times of organs in the mouse.Location T1 [ms] T2 [ms]Kidneys 1259 54Front leg muscle 1190 39Liver 985 60

4.2.2 Brown and white adipose tissue

T1 values for the different regions are shown in figure 16. As we see the lowest value are

recorded in the far back of the mouse, and increasing with more lateral location.

Figure 16: Relaxation times for the white and brown adipose tissue, locations of the ROIs can beseen in figure 6 in chapter 3.2.2.

4.3 Rat

The results from the relaxation measurements of the rat brain are given in Table 3. As can

be seen the relaxation times of thalamus and medulla oblongata are almost equal, difference

of T1 by 1.9 % and T2 by 0.15 %. Corpus callosum has the shortest T1 time, and the second

19

Page 21: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

highest T2 time. Hippocampus and cerebral cortex have different values for the right and left

side.

Table 3: Relaxation times of several locations in the rat brain.Location T1 [ms] T2 [ms]Cerebellum 1343 80Corpus Callosum 1080 128Thalamus 1530 68Medulla oblongata 1560 68Cerebral cortex R/L 1246/1635 76/78Hippocampus R/L 1249/1072 171/254

5 Discussion

5.1 Sequence evaluation with phantom experiments

Careful evaluation of the pulse sequence is necessary to ensure valid results when used in

animals. Properties tested were the number of frames (i.e. data points for the fit curve),

size of ROI, number of averages, single vs. multislice acquisition, intervariability and slice

orientation dependencies.

Many data points are needed for a reliable curve fit, therefore a large number of frames is

recommended. A higher number of data points will lead to more exact values of the fit in case

the fit model is chosen correct for the data. In return, a wrong fit model or problems with

the measurements can be identified easier with a higher number of data points. Fewer frames

will give a higher standard deviation of the fit. As the test with varying the frames were all

done at the same total scan time, we decided for 60 frames, aiming for high accuracy. Initial

tests (data not shown) with a lower number of frames, thus less data points, provided still

acceptable curves for a fit. However, as a complete relaxation has to be reached in a FISP

sequence prior to acquisition of the next point in k-space, the shortage in the sequence did

not yield any advantage.

Organs and other anatomical structures can limit the size of ROIs, as they are not always

big enough for placing a sufficient large ROI. Due to small size of the ROI, the average and

the results might be influenced. There is no doubt that large ROIs, covering more tissue will

give an average with lower errors than small ROIs. The largest possible ROI that cover the

organ is therefore recommended, while taking care to average only uniform structures of the

organ. The results shows that ROIs larger then 0.27 cm2 are sufficient in size (i.e. no more

changes when increasing the area) and will provide valid results. If the ROI is smaller than

20

Page 22: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

0.27 cm2, the results are prone to error. Mainly, low accuracy does not provide good data for

the fit and standard error of the fit is increasing dramatically. In the phantom the difference

between a ROI of size 0.035 cm2 and 0.42 cm2 were 5.8 % for the T1 and 9.1 % for the T2.

Organs in mice which are big enough to be covered by a ROI of sufficient size are liver, kidney,

muscles and fat. In principle, even the brain has a sufficient size to be covered by a single

voxel, however, the brain consist of several compartments, i.e. grey and white matter, CSF,

which are known to show different relaxation times. [13]

Variation of the signal to noise ratio, SNR (i.e. number of averages) did affect measured

results for the relaxation time, the largest variations were at the low SNR regions. Some images

indicated that there were susceptibility problems in the high-SNR ROI, probably caused by

an air bubble. The air bubble is seen as signal void in the upper compartment of the phantom

in Figure 5. This can explain the variation of relaxation times in our measurement. In theory

the increase of SNR should go by the square root of number of averages. Comparison between

measurement and theory are shown in Figure 17. Values achieved in the measurements were

aligned quite well to the theory. The standard value used in the sequence is four averages,

thus if the time was not critical in animal measurements 4 averages were used.

Figure 17: This figure shows a comparison between measured SNR and theoretical values.

Some early thoughts about the relaxation measurements were to cover the whole animal

with relaxation measurements and then pick out the values of the most interesting areas

afterwards. The resulting scan time was unacceptable long (>3 h) so complete body scans

could not be performed. Furthermore single vs. multislice measurements indicated that the

values were not identical when changing the number of slices and the gap in-between. This

could be an effect of inaccurate slice selections, and is in concert with Brukers recommendation

for single slice usage of FISP-T1-T2 protocol.

The results should not be time dependent, thus intervariability measurements were per-

21

Page 23: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

formed. The results are constant until scan number 8. As the measurements were performed

automatically over night, something could have happened beyond operator control. Repetition

of the experiment was not possible due to time constraints.

Slice orientation should not affect the results. Almost all values were identical, and inde-

pendent of slice orientation. Remarkably, the coronal T2 value does not align to the others.

Because of the usage of a one-compartment phantom this is unlikely caused by susceptibility.

Earlier results (not shown) were dependent on slice selection, this was before a correction and

adjustment of eddy current compensation were performed. Eddy currents can induce field

which is different for each slice orientation. Corrections and tests of the eddy currents are of

great importance for reliable values in the in vivo situation.

5.2 Relaxation measurements in mice

Measurements on living animals are a lot different from measuring a phantom, as the animal

needs to be monitored and anesthetized. A living animal is moving, due to both cardiac

function and respiration and without triggering there might be some problems. To avoid

motion artefacts frequency- and phase encoding directions must be properly configured.

The protocol provided by Bruker has the possibility to use a triggered mode of the data

acquisition. Unfortunately, the trigger is setup such that an initiating event, e.g. heart beat

or respiration phase, is taken as the trigger point for the whole sequence. I.e. once the trigger

point is reached, the whole measurement is acquired and the motion is not frozen by a triggered

data acquisition. Besides the question what this type of trigger should be useful for, the T1-

T2 measurements could be reprogrammed as SSFP (steady state free precession) sequence.

We have placed this proposal at the manufacturer’s software department. Programming from

scratch would by far exceed the scope of this work.

5.2.1 Relaxation measurements in organs and anatomical structures

As mentioned above, the ROIs to be investigated have to have a certain size in order to get

reproducible values. In the mouse, where body weight is around 20 gram in an adult mouse,

organs are small and are likely to fall beyond the defined size of the ROI. In addition, the

inadequate trigger handling, as mentioned in chapter 5.2, does not allow meaningful data

acquisition of moving organs, i.e. the heart and the thorax, as well as structures next to them.

T1 increases in the order of liver, muscle and kidney and T2 values increases in the order

of muscle, kidney and liver. One might speculate where differences in relaxation times could

be originated. As specific weight of the different organs is nearly constant [14], cell density can

be ruled out. However, different levels of blood in the various organs could be one source for

22

Page 24: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

the differences. [15] Blood, especially when oxygenated contains substantial amounts of Fe3+,

which is paramagnetic. Thus, relaxation times should be shorter with higher regional blood

content of the organs. As the order of T1 relaxation times aligns very well to the regional

blood volume, we assume that Fe3+ content contributes to the differences in the relaxation

times. However, other factors might influence the relaxation times as well.

5.2.2 Brown and white adipose tissue

Non-invasive in-vivo differentiation of brown and white adipose tissue would be helpful in

obesity research, a these fat pools act differently. [16, 17] Brown adipose tissue is supplied

with substantial blood flow, which causes the coloration of the fat. In contrast, white adipose

tissue has limited blood flow. As mentioned in 5.2.1, presence of oxygenated blood which

contains paramagnetic Fe3+, will shorten the relaxation times. During this work, we tested,

whether differences in T1 and T2 relaxation might be helpful to discriminate brown and

white adipose tissue. Motion artifacts due to respiration rule out a clear differentiation. As

stated in 5.2, a SSFP sequence could help to eliminate the problem. Currently, we can state

that differences of 12 % can be seen, however, the differences are not statistically significant.

Further work with an improved sequence should clarify the potential role of MRI in this field

of obesity research.

5.3 Relaxation measurements in rat brain

T2-weighted high resolution images of the rat brain were acquired. Thereafter, positions and

different structures in the rat brain could be assigned to the data of an anatomical brain

atlas. As the size of the ROI is a limiting factor, we could place ROIs only in medulla

oblongata, thalamus, cerebellum, cerebral cortex, hippocampus and corpus callosum, while

regional variations within the different locations could not be evaluated. The values for medulla

oblongata and thalamus are identical (1560/68 ms and 1530/68 ms). At the recent Annual

Scientific Meeting of the ISMRM (International Society for Magnetic Resonance in Medicine),

T1 values of marmoset at a field strength of 7 T were presented by Bock et al.. T1 values

in white and grey matter were 1350 ms and 1750 ms respectively. These values correlate

very well with our own results in the cerebellum (1343 ms). To our surprise, values measured

in the Corpus callosum (1080 ms) are lower, although both structures belong to the white

matter. A difference was seen for structures in the grey matter as well, as medulla oblongata,

thalamus and cerebral cortex show lower values. As brain structure and myelination differ

between species, differences in the observed values in comparison to marmoset could be based

on differences in morphology.

23

Page 25: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

We have seen left-right variation in relaxation times of hippocampus and cerebral cortex.

The differences are in the order of 14 % and 24 % for hippocampus and cerebral cortex, respec-

tively and cannot be neglected. As we exclude a physiological background of the differences,

they might arise from differences in the phase information acquired during the FISP mea-

surements. Even with a reconstruction which reports magnitude values, which should exclude

the phase problem, the values obtained for the left and right hemisphere were not identical.

Field inhomogeneities are one possible reason for the inaccuracy of the phase information.

With non-uniform field strength over the rat brain, precession of the spins will be different

and relaxation times are likely to be influenced. These difficulties show the need for a critical

evaluation of the data acquired by the FISP sequence.

In our measurements, the ratio of the T1 relaxation times between grey and white matter

are approximately 1.15. The ratio is quite low in comparison to values for the human brain

at a field strength of 1.5 T [18], where the ratio is about 2. Thus, we can conclude that the

high field is not beneficial for the imaging contrast of grey and white matter. Discrimination

of grey and white matter is easier at lower field strength, however, the advantages of the high

field strength including higher resolution should not be forgotten.

24

Page 26: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Acknowledgement

I would like to give my deepest gratitude to my supervisor Michel Horn.

I will also thank Maria Enge for letting me borrow some of her fat mice, and to Erik Schéle

for the lesson and dissection of brown adipose tissue.

Measurements where done at Center for Mouse Physiology and Bio-Imaging (CPI), a core

facility of the University of Gothenburg.

25

Page 27: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

References

[1] Radiological Society of North America (RSNA). Magnetic Resonace Imaging (MRI) -

Body. Retrieved March 10, 2008 from:

http://www.radiologyinfo.org/en/info.cfm?pg=bodymr.

[2] Damadian Raymond. Tumor detection by nuclear magnetic resonance. Science, 171:1151–

1153, 1971.

[3] Bruker BioSpin MRI GmbH. ParaVision 4.0 document set / Application. Germany, 2006.

[4] Gadian David G. NMR and it’s applications to living systems. Oxford University Press,

Oxford, 1995.

[5] Hendee William R. Medical imaging physics. Wiley-Liss Inc., New York, 2002.

[6] McRobbie Donald W. MRI From Picture to Proton. Cambridge Univeristy Press, Cam-

bride, 2007.

[7] Puddephat Mike. Principles of Magnetic Resonance Imaging. Retrieved April 28, 2008

from: http://www.easymeasure.co.uk/principlesmri.aspx.

[8] Hornak Joseph P. The Basics of MRI. Retrieved March 27, 2008 from:

http://www.cis.rit.edu/htbooks/mri/inside.htm.

[9] Schmitt Peter et al. Inversion recovery truefisp: Quantification of t1, t2, and spin density.

Magnetic Resonance in Medicine, 51:661–667, 2004.

[10] Schmitt Peter et al. Erratum to "inversion recovery truefisp: Quantification of t1, t2,

and spin density". Magnetic Resonance in Medicine, 52:698, 2004.

[11] Austgen Laura. Brown Adipose Tissue. Retrieved May 18, 2008 from:

http://www.vivo.colostate.edu/hbooks/pathphys/misc_topics/brownfat.html.

[12] Paxinos George and Watson Charles. The Rat Brain in Stereotaxic Coordinates. Academic

Press Limited, 1997.

[13] Fagan Andrew J. et al. Serial postmortem relaxometry in the normal rat brain and

following stroke. Journal of Magnetic Resonance Imaging, 27:469–475, 2008.

[14] Tang Haiying et al. High-resolution magnetic resonance imaging tracks changes in organ

and tissue mass in obese and aging rats. Am J Physiol Regul Integr Comp Physiol,

282:R890–R899, 2001.

26

Page 28: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

[15] Horn Michael et al. 31p-nuclear magnetic resonace spectroscopy of blood: A species

comparsion. Journal of Cardiocascular Magnetic Resonance, 2(2):143–149, 2000.

[16] Hansen Jacob B. and Kristiansen Karsten. Regulatory circuits controlling white versus

brown adipocyte differentiation. Biochemichal journal, 398:153–268, 2006.

[17] York David and Bouchard Claude. How obesity develops. Endocrine, 13:143–154, 2000.

[18] Steinhoff S. et al. Fast t1 mapping with volume coverage. Magnetic Resonance in Medi-

cine, 46:131–140, 2001.

27

Page 29: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

APPENDIX

Derivation of the Larmor frequency, classical mechanics.

The protons magnetic moment is given in equation A-1, where �µ is the magnetic moment, γ

is the gyromagnetic ratio and �B is the magnetic field.

�µ = γ · �B (A-1)

If an external magnetic field is applied, the magnetic moment experiences a torque and pre-

cesses around the magnetic field axis.

∆µ = µ sin θ · ω∆t (A-2)

Where �µ = µ sin θ and small angles are assumed.

Angular momentum, equation of motion:

d�S

dt= �µ × �B (A-3)

Combine equation A-1 and A-3.

d�µ

dt= γ�µ× �B (A-4)

From equation A-4:

∆µ = γ∣∣∣�µ × �B

∣∣∣ · ∆t = γµ · B sin θ · ∆t (A-5)

Set equation A-2 equal to A-5

µ sin θ · ω∆t = γµ · B sin θ · ∆t (A-6)

We now have the Larmor equation:

ω = γ · B (A-7)

28

Page 30: MRI relaxation measurements in rats and mice · MRI relaxation measurements in rats and mice Master’s Thesis, 30hp Jesper Lindberg Spring 2008 Supervisor: Michael Horn University

Derivation of the Larmor frequency, quantum mechanics.

The protons magnetic moment:

�µ = γ�I = γ · �S (A-8)

In an external magnetic field there is 2I +1 possible values for the angular momentum. Thus,1H have two possible enery levels ±1

2 .

Energy states:

ε = �µ · �B = γ · � · �I · �B (A-9)

Energy difference between the two states:

∆ε =[(

12

)−

(−1

2

)]γ · � · B = γ� · B (A-10)

De Broglie’s wave equation:

∆ε = � · ω (A-11)

Set equation A-10 equal to A-11:

γ�B = �ω (A-12)

And now we have the larmor equation:

ω = γ · B (A-13)

29