1 Journal of Refractive Surgery • Vol. xx, No. x, 201X NEW CONCEPTS lthough congenital cataract in children is a rare disorder, it is considered to be the most common cause of treatable blindness worldwide. The prev- alence in newborns is approximately 1 to 13 cases in every 10,000 births. 1,2 Accurate diagnosis, ideal intervention tim- ing, and treatment choice are extremely important to ensure visual rehabilitation, avoid amblyopia, and minimize the risk of complications. One of the most crucial decisions when performing pedi- atric cataract surgery is the choice between aphakia or IOL implantation. The type of IOL to implant and the dioptric power of the IOL used are critical to avoid extreme refractive surprises postoperatively. The concept of a multi-component adjustable IOL (InfiniteVision Optics, Strasbourg, France) of- fers a new perspective in the future management of infantile cataracts. Patients who have not yet completed their biologi- cal ocular growth may still have the possibility of an enhance- ment surgery, allowing for short-term and long-term refrac- tive adjustments, as they are required, as the patient matures. IOL IMPLANTATION: CURRENT STATUS The available options when performing cataract surgery in children are aphakia followed by immediate contact lens fitting versus primary IOL implantation. IOL implantation is the preferred approach in children who have completed their second year of life, but surgeons remain hesitant to implant an IOL in infants and children younger than 2 years. The ad- vantages of IOL implantation include amblyopia prevention, lower rate of certain complications, and higher compliance if contact lens fitting is impossible. Ledoux et al. 3 conducted a 14-year retrospective study of children with unilateral or A ABSTRACT PURPOSE: The multi-component intraocular lens (IOL) (IVO; SAS, Strasbourg, France) is a novel approach to the treatment of pediatric cataract. Because the re- fractive requirements for pediatric eyes often change over time, current IOL technology does not easily allow refractive adjustments after the primary surgical inter- vention. The multi-component IOL concept allows easy, surgical refractive adjustments to the initial surgical im- plantation at any postoperative time period. Thus, both surgical implantation and enhancement surgery have been successfully accomplished in adult patients. METHODS: A novel surgical approach to pediatric cat- aract surgery is described. At the time of the primary surgery, a two component IOL was implanted. At any postoperative time period, the front lens component, located in front of the capsular bag, could be easily sur- gically exchanged because the dioptric power require- ments of the pediatric eye changed over time. RESULTS: Both primary and enhancement surgeries have been done in adult patients with good results. Im- plantations have occurred uneventfully in all cases with no intraoperative or postoperative complications. There was no statistically significant difference in the endothe- lial cell density, anterior chamber depth, and pachym- etry readings preoperatively and 2 years postoperatively. There was no interlenticular fibrosis present. CONCLUSION: The multi-component IOL should pro- vide a unique and greatly needed surgically adjustable approach to the treatment of pediatric cataract. [J Refract Surg. 20XX;XX(X):XX-XX.] From the Institute of Vision and Optics, University of Crete, Crete, Greece. Submitted: December 22, 2012; Accepted: August 5, 2013; Posted online: October 30, 2013 Dr. Portaliou has received support for travel and meeting expenses related to InfiniteVision Optics presentations. The remaining authors have no financial or proprietary interest in the materials presented herein. Correspondence: Dimitra M. Portaliou, MD, Institute of Vision and Optics, University of Crete, Medical School, 71003 Heraklion, Crete, Greece. E-mail: [email protected] doi:10.3928/1081597X-20131023-01 Multi-Component Adjustable Intraocular Lenses: A New Concept in Pediatric Cataract Surgery Dimitra M. Portaliou, MD; George D. Kymionis, MD, PhD; Ioannis G. Pallikaris, MD, PhD

Upload: others

Post on 20-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Multi-Component Adjustable Intraocular Lenses: A …m3.wyanokecdn.com/b5d922c96e9de68c781cb581540628d0.pdf2 Crig SLC Irre New Concept in IOL Implantation/Portaliou et al bilateral

1Journal of Refractive Surgery • Vol. xx, No. x, 201X

N E W C O N C E P T S

lthough congenital cataract in children is a rare disorder, it is considered to be the most common cause of treatable blindness worldwide. The prev-

alence in newborns is approximately 1 to 13 cases in every 10,000 births.1,2 Accurate diagnosis, ideal intervention tim-ing, and treatment choice are extremely important to ensure visual rehabilitation, avoid amblyopia, and minimize the risk of complications.

One of the most crucial decisions when performing pedi-atric cataract surgery is the choice between aphakia or IOL implantation. The type of IOL to implant and the dioptric power of the IOL used are critical to avoid extreme refractive surprises postoperatively. The concept of a multi-component adjustable IOL (InfiniteVision Optics, Strasbourg, France) of-fers a new perspective in the future management of infantile cataracts. Patients who have not yet completed their biologi-cal ocular growth may still have the possibility of an enhance-ment surgery, allowing for short-term and long-term refrac-tive adjustments, as they are required, as the patient matures.

IOL IMPLANTATION: CURRENT STATUSThe available options when performing cataract surgery

in children are aphakia followed by immediate contact lens fitting versus primary IOL implantation. IOL implantation is the preferred approach in children who have completed their second year of life, but surgeons remain hesitant to implant an IOL in infants and children younger than 2 years. The ad-vantages of IOL implantation include amblyopia prevention, lower rate of certain complications, and higher compliance if contact lens fitting is impossible. Ledoux et al.3 conducted a 14-year retrospective study of children with unilateral or

AABSTRACT

PURPOSE: The multi-component intraocular lens (IOL) (IVO; SAS, Strasbourg, France) is a novel approach to the treatment of pediatric cataract. Because the re-fractive requirements for pediatric eyes often change over time, current IOL technology does not easily allow refractive adjustments after the primary surgical inter-vention. The multi-component IOL concept allows easy, surgical refractive adjustments to the initial surgical im-plantation at any postoperative time period. Thus, both surgical implantation and enhancement surgery have been successfully accomplished in adult patients.

METHODS: A novel surgical approach to pediatric cat-aract surgery is described. At the time of the primary surgery, a two component IOL was implanted. At any postoperative time period, the front lens component, located in front of the capsular bag, could be easily sur-gically exchanged because the dioptric power require-ments of the pediatric eye changed over time.

RESULTS: Both primary and enhancement surgeries have been done in adult patients with good results. Im-plantations have occurred uneventfully in all cases with no intraoperative or postoperative complications. There was no statistically significant difference in the endothe-lial cell density, anterior chamber depth, and pachym-etry readings preoperatively and 2 years postoperatively. There was no interlenticular fibrosis present.

CONCLUSION: The multi-component IOL should pro-vide a unique and greatly needed surgically adjustable approach to the treatment of pediatric cataract.

[J Refract Surg. 20XX;XX(X):XX-XX.]

From the Institute of Vision and Optics, University of Crete, Crete, Greece.

Submitted: December 22, 2012; Accepted: August 5, 2013; Posted online: October 30, 2013

Dr. Portaliou has received support for travel and meeting expenses related to InfiniteVision Optics presentations. The remaining authors have no financial or proprietary interest in the materials presented herein.

Correspondence: Dimitra M. Portaliou, MD, Institute of Vision and Optics, University of Crete, Medical School, 71003 Heraklion, Crete, Greece. E-mail: [email protected]

doi:10.3928/1081597X-20131023-01

Multi-Component Adjustable Intraocular Lenses: A New Concept in Pediatric Cataract SurgeryDimitra M. Portaliou, MD; George D. Kymionis, MD, PhD; Ioannis G. Pallikaris, MD, PhD

Page 2: Multi-Component Adjustable Intraocular Lenses: A …m3.wyanokecdn.com/b5d922c96e9de68c781cb581540628d0.pdf2 Crig SLC Irre New Concept in IOL Implantation/Portaliou et al bilateral

2 Copyright © SLACK Incorporated

New Concept in IOL Implantation/Portaliou et al

bilateral cataract who underwent IOL implantation, in which approximately 75% of children achieved 20/40 or better visual acuity postoperatively.

Despite the advantages, deciding on the IOL type and power remains challenging because the anatomy of children’s eyes is unique in certain aspects. A child’s ocular system is characterized by steeper corneas, shallower anterior chamber depth, and shorter axial length,4 making correct biometry and accurate IOL cal-culations almost impossible. In a study by Moore et al.,5 the difference between predicted refraction with the usual biometry calculating formulas (Holladay I, SRK II, SRK/T) and actual postoperative refraction was found to be 1.08 ± 0.93 diopters. Children younger than 2 years had the greatest discrepancy of results. Trivedi et al.6 and Kekunnaya et al.7 conducted studies comparing different IOL power calculation formulas in children to enhance the accuracy of the IOL diop-tric power chosen. Trivedi et al.6 concluded that even when preoperative refraction data are not available, the Holladay 2 formula has the least prediction error and absolute prediction error.

It seems that immersion A-scan biometry provides greater accuracy in axial length measurement in chil-dren than contact biometry.8 Both the IOLMaster9 (Carl Zeiss Meditec, Jena, Germany) and Lenstar10 (Haag-Streit AG, Koeniz, Switzerland) have been used in children, but their use is limited due to poor coop-eration. Also, these biometric devices are not useful in the measurement of dense cataracts more commonly encountered in children.

McClatchey11 evaluated a computer program to pre-dict the pseudophakic refraction of a child at any age. A computer program written for Windows 95 calculated the initial postoperative pseudophakic refraction of a child using Holladay’s formula, given the axial length and keratometry readings. The logarithmic model was used to predict the ultimate refraction at age 20 years and chart the predicted curve of refractive error with standard deviations. Despite the relative accuracy of this model, the need for a special pediatric IOL formula is evident and remains to be discovered. Also, because the refraction changes over time, the primary issue is to establish a clear focus at all times. This requires occa-sional changes in the optical system irrespective of ac-curacy at any given time. Also, these small adjustments tend to be accurate based on refractive measures.

Predicting the refractive change of a patient over time is another challenge for pediatric ophthalmolo-gists. Trivedi and Wilson12 concluded that there is a continuous myopic shift in refraction as age advances. The higher difference in axial length occurs during the first 2 years of life. In particular, there is rapid growth

of 0.62 mm/month during the first 6 months and se-quentially 0.19 mm/month in the time frame between the 6th and 18th month of life. The growth continues with a much lower rate (0.01 mm/month) until adult-hood. Of course, these general rules do not always ap-ply because the natural variance in growth of the eye is unpredictable. Astle et al.13 demonstrated in a retro-spective review that myopic shift in refraction can be as high as -10.0 to -12.0 D in children younger than 4 years, making IOL power calculation in infants a dif-ficult task. Most surgeons aim for remaining hyperopia when implanting an IOL, expecting a myopic shift dur-ing the growth process that will lead to emmetropia by the age of 18 years. Others aim for a slight myopia that will facilitate children’s near vision and possibly prevent amblyopia.

The type of IOL to implant is an important issue. In-the-bag implantation is preferable and both poly-methylmethacrylate (PMMA) and acrylic hydrophobic lenses are considered materials of choice. Wilson et al.14 believed the single-piece Acrysof SA (Alcon Lab-oratories, Fort Worth, TX) series was the best option available at the moment for in-the-bag implantation in children. Gupta et al.15 reported a retrospective study of primary IOL implantation after congenital cataract surgery with the use of the SRK II formula and im-plantation of either rigid PMMA non-foldable lenses (Ocular Vision SF 102; Eye Care, Vadodara, India) or acrylic hydrophobic foldable lenses (Acrysof SA60 AT, MA60AC, or SN60WF; Alcon Laboratories) with a low rate of manageable complications (such as opaci-fication of the visual axis and decentration of the IOL) and no reports of glaucoma, endophthalmitis, or reti-nal detachment at the last follow-up.

Certain investigators have debated the use of multi-focal IOLs in children.16 Multifocal IOLs can give the opportunity of spectacle independence in all distances, but at the same time their use remains problematic be-cause accurate biometry calculations are a prerequisite to avoid halos and preserve near visual acuity benefit.17

Tassignon et al.18 reported the outcome of a surgical procedure called ‘bag-in-the-lens’ in pediatric catarac-tous eyes. With this technique, the anterior and pos-terior capsules are placed in the groove of a specially designed IOL after a capsulorhexis of the same size is created in both capsules. The principle behind this IOL design is to ensure a clear visual axis by mechani-cally tucking the two capsules into the IOL, thereby preventing any migration of proliferating lens epithe-lial cells. In theory, this technique could be adapted to the multi-component lens system to isolate the vitre-ous (to simplify potential enhancement procedures), thus allowing easy access to the front lens.

Page 3: Multi-Component Adjustable Intraocular Lenses: A …m3.wyanokecdn.com/b5d922c96e9de68c781cb581540628d0.pdf2 Crig SLC Irre New Concept in IOL Implantation/Portaliou et al bilateral

3Journal of Refractive Surgery • Vol. xx, No. x, 201X

New Concept in IOL Implantation/Portaliou et al

MULTI-COMPONENT ADJUSTABLE IOLThe possibility of an adjustable lens represents

a solution for many of the refractive difficulties that arise during IOL implantation in pediatric cataract surgery.19 The InfiniteVision Optics lens is an acrylic foldable posterior chamber optical system that enables customized correction of all degrees of sphere and cylinder in the primary surgery. This lens allows for continual correction of all residual refractive errors postoperatively with the surgical exchange of one of the lens components. The use of a multifocal lens front component is also possible.

The InfiniteVision Optics lens ocular system con-sists of two components. The base lens or posterior component has a plate-haptic configuration, contains only spherical correction, and can be implanted in the

capsular bag, analogous to a conventional posterior chamber IOL. The front lens assembly has two haptics extending off of the optic. These haptics are designed to fit into the two small bridges that project off the anterior surface of the base lens to secure the front lens assem-bly to the base lens (Figure 1). In case of posterior cap-sule opacification, which is rather common in pediatric cataract surgery, YAG laser can be performed without compromising the safety of an enhancement surgery if it is judged as necessary. It is highly desirable to house the base lens in the capsular bag and allow capsule fi-brosis to surround the base lens peripherally. When fi-brosis is occurring, if the central area of the posterior capsule is involved, a small central capsular opening can be made, still allowing isolation of the vitreous body behind the base lens. This ensures that repeated vitreous surgery would not be required if a front lens exchange is necessary at some time point in the future. The front lens assembly allows for spherical, cylindri-cal, and multifocal primary corrections for immediate and long-term adjustments. Implantation of the two components can be effected through a 1.8- to 2.8-mm incision, with ease of assembly inside the eye.20

The base lens component is first implanted and then rotated 90° so that the attachment openings are perpen-dicular to the insertion point. The front lens assembly is then inserted through the same incision into the an-terior chamber and also rotated by 90°. The haptics of the front lens are then inserted into the bridge capture openings of the base lens. A technique called capsule capture is an important step of the surgical procedure, in which the haptics of the front lens are placed in front of the peripheral anterior capsule, capturing the capsule between the haptics of the front lens and the plate-haptic of the base lens. This surgical maneuver places the front lens component in front of the capsule so that when capsule fibrosis occurs it will only affect the base lens because the front lens is outside the cap-sular bag. Without the occurrence of fibrosis, the front component can be easily removed or exchanged surgi-cally if necessary. Capsule capture of the base lens is important, but assembly and disassembly of the lenses can be performed even if proper capsule capture is not achieved in the primary surgery.

This multi-component lens is of great use in case of undercorrections or overcorrections after cataract ex-traction and IOL implantation because the front lens is easily surgically exchangeable.

The 2-year follow-up of the initial clinical evalua-tion of the InfiniteVision Optics multi-component IOL included 6 eyes of 6 patients with a mean age of 63.3 ± 12.7 years (range: 49 to 82 years) that underwent cataract surgery and multi-component IOL implanta-

Figure 1. The (A) base InfiniteVision Optics lens (InfiniteVision Optics, Strasbourg, France), (B) the front lens, and (C) the assembled front and base lens.

A

B

C

Page 4: Multi-Component Adjustable Intraocular Lenses: A …m3.wyanokecdn.com/b5d922c96e9de68c781cb581540628d0.pdf2 Crig SLC Irre New Concept in IOL Implantation/Portaliou et al bilateral

4 Copyright © SLACK Incorporated

New Concept in IOL Implantation/Portaliou et al

tion.20 The aim of this feasibility study was to evalu-ate the manufacturing expediency of the lens, the ease of the surgical implantation technique, the incidence of interlenticular fibrosis, and the rate of complica-

tions. All surgeries were uneventful and no complica-tions occurred. At 2 years postoperatively, interlen-ticular fibrosis had not developed and there was no statistically significant difference in the endothelial cell density, anterior chamber depth, and pachymetry readings compared to preoperative values. Mean un-corrected and corrected distance visual acuity (UDVA and CDVA) improved significantly (P < .01) from 0.11 ± 0.06 (range: 0.05 to 0.2) (decimal scale) to 0.68 ± 0.11 (range: 0.5 to 0.8) and from 0.28 ± 0.13 (range: 0.1 to 0.5) to 0.83 ± 0.16 (range: 0.6 to 1), respectively. No pa-tient lost lines of CDVA during the follow-up period. The study concluded that multi-component IOL im-plantations were easy and safe to perform without the risk of intraoperative or postoperative complications (Figure 2).

DISCUSSIONPredicting the axial growth and the refractive change

that accompanies it remains one of the key challeng-es in the follow-up of pediatric cataract surgery. The multi-component IOL lens concept can facilitate pe-diatric ophthalmologists by customizing the primary surgery, by allowing a small variety of lenses in their inventory to cover a large refractive range of sphere and cylinder, and by easing the burden of IOL choice.

The multi-component IOL in pediatric cataract sur-gery should be designed with a smaller diameter to fit in a child’s capsular bag and to avoid anterior chamber shallowing and glaucoma development.

A case report of an enhancement performed with the multi-component lens system has been previously described (Gianluca Rubiollini, personal communica-tion, Milan, Italy, 2009).21 The 78-year-old patient in question had a postoperative residual refractive error of 2.25 diopters (due to a biometry error) following routine cataract surgery and multi-component IOL im-plantation even though the target refraction was em-metropia. The patient underwent a minimally invasive enhancement surgery 9 months after the primary sur-gery by removal and exchange of the front lens assem-bly. Using the same incisions created in the primary surgery, the front lens assembly haptics were detached from the bridges of the base lens and the front lens was removed. Then another front lens was inserted in the anterior chamber and attached to the small bridges that project off the anterior surface of the base lens. This re-sulted in a final refraction of +0.25 with a final UDVA of 1.0 at 1 week after the enhancement procedure.

Clearly this case demonstrates the feasibility and safety of such a procedure, backing up the proof of the enhancement concept, which is so essential for the pediatric cataract population. Repeated surgeries can

Figure 2. (A) The InfiniteVision Optics base lens (InfiniteVision Optics, Strasbourg, France), (B) the assembled lens, and (C) slit-lamp photo-graph of a patient with the assembled lens.

A

B

C

Page 5: Multi-Component Adjustable Intraocular Lenses: A …m3.wyanokecdn.com/b5d922c96e9de68c781cb581540628d0.pdf2 Crig SLC Irre New Concept in IOL Implantation/Portaliou et al bilateral

5Journal of Refractive Surgery • Vol. xx, No. x, 201X

New Concept in IOL Implantation/Portaliou et al

worsen the prognosis, but enhancement surgery is a minimally invasive procedure in which the refractive benefit for the patient far outweighs the risk taken.

Pediatric cataract surgery is essentially different from adult cataract surgery. In pediatric cataract sur-gery, additional risks are present and postoperative complications may jeopardize an otherwise successful surgical intervention. One of the main considerations is the ongoing postoperative evolution of the power of IOL needed to correct children’s eyes. A solution to the problem is provided by the development of the new, innovative, multi-component InfiniteVision Op-tics IOL that will allow postoperative adjustments for postoperative changes in refraction.

AUTHOR CONTRIBUTIONSStudy concept and design (DMP, IGP, GDK); drafting of the manu-

script (DMP); critical revision of the manuscript (IGP, GDK); supervi-

sion (IGP, GDK)

REFERENCES 1. Foster A, Gilbert C, Rahi J. Epidemiology of cataract in child-

hood: a global perspective. J Cataract Refract Surg. 1997;23(sup-pl 1):601-604.

2. SanGiovanni JP, Chew EY, Reed GF, et al. Infantile cataract in the collaborative perinatal project: prevalence and risk factors. Arch Ophthalmol. 2002;120:1559-1565.

3. Ledoux DM, Trivedi RH, Wilson ME, Payne JF. Pediatric cata-ract extraction with intraocular lens implantation: visual acuity outcome when measured at age four years and older. J AAPOS. 2007;11:218-224.

4. Lin AA, Buckley EG. Update on pediatric cataract surgery and intraocular lens implantation. Curr Opin Ophthalmol. 2010;21:55-59.

5. Moore DB, Zion B, Neely DE, et al. Accuracy of biometry in pediatric cataract extraction with primary intraocular lens im-plantation. J Cataract Refract Surg. 2008;34:1940-1947.

6. Trivedi RH, Wilson ME, Reardon W. Accuracy of the Holladay 2 intraocular lens formula for pediatric eyes in the absence of preoperative refraction. J Cataract Refract Surg. 2011;37:1239-1243.

7. Kekunnaya R, Gupta A, Sachdeva V, Rao HL, Vaddavalli PK, Om Prakash V. Accuracy of intraocular lens power calculation

formulae in children less than two years. Am J Ophthalmol. 2012;154:13-19.

8. Trivedi RH, Wilson ME. Prediction error after pediatric cataract surgery with intraocular lens implantation: contact versus im-mersion A-scan biometry. J Cataract Refract Surg. 2011;37:501-505.

9. Lenhart PD, Hutchinson AK, Lynn MJ, Lambert SR. Partial co-herence interferometry versus immersion ultrasonography for axial length measurement in children. J Cataract Refract Surg. 2010;36:2100-2104.

10. Gursoy H, Sahin A, Basmak H, Ozer A, Yildirim N, Colak E. Lenstar versus ultrasound for ocular biometry in a pediatric population. Optom Vis Sci. 2011;88:912-919.

11. McClatchey SK. Choosing IOL power in pediatric cataract sur-gery. Int Ophthalmol Clin. 2010;50:115-123.

12. Trivedi RH, Wilson ME. Biometry data from Caucasian and Af-rican–American cataractous pediatric eyes. Invest Ophthalmol Vis Sci. 2007;48:4671-4678.

13. Astle WF, Ingram AD, Isaza GM, Echeverri P. Paediatric pseu-dophakia: analysis of intraocular lens power and myopic shift. Clin Experiment Ophthalmol. 2007;35:244-251.

14. Wilson ME, Trivedi RH. Choice of intraocular lens for pediatric cataract surgery: survey of AAPOS members. J Cataract Refract Surg. 2007;33:1666-1668.

15. Gupta A, Kekunnaya R, Ramappa M, Vaddavalli PK. Safety pro-file of primary intraocular lens implantation in children below 2 years of age. Br J Ophthalmol. 2011;95:477-480.

16. Hug D. Update on pediatric cataracts and cataract surgery. Int Ophthalmol Clin. 2010;50:107-113.

17. Jacobi CP, Dielein TS, Konen W. Multifocal intraocular lens implantation in pediatric cataract surgery. Ophthalmology. 2001;108:1375-1380.

18. Tassignon MJ, Gobin L, Mathysen D, Van Looveren J, De Groot V. Clinical outcomes of cataract surgery after bag-in-the-lens intraocular lens implantation following ISO standard 11979-7:2006. J Cataract Refract Surg. 2011;37:2120-2129.

19. Portaliou DM, Pallikaris IG. Adjustable intraocular lens for use in pediatric ophthalmology. Am J Ophthalmol. 2013;155:199.

20. Portaliou DM, Grentzelos MA, Pallikaris IG. Multicomponent intraocular lens implantation: two-year follow-up. J Cataract Refract Surg. 2013;39:578-584.

21. Portaliou DM, Kymionis GD, Pallikaris GI. First enhancement surgery following multicomponent IOL implantation: early re-sults. Poster presented at the 2010 AAO/MEACO Joint Meeting; Chicago, Illinois; October 16-19, 2010.