N9005 and N9094 OPERATOR MANUAL

Upload: others

Post on 24-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

  • N9005 and N9094OPERATOR MANUAL

  • 1

    CONTENTSPAGE

    General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Calibration,CertificationandService . . . . . . . . . . . . . . . . . . . . . . 41. IntroductiontotheN9005andN9094 . . . . . . . . . . . . . . . . . . . . . 52. N9005andN9094ThermometerswithBioCote®Protection . . . . 53. OverviewofDisplaySymbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64. ConnectingProbestotheThermometer . . . . . . . . . . . . . . . . . . . . 65. SwitchingOnandOff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76. DataHold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77. CountdownTimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78. AutoSwitchOff/PowerSavingMode . . . . . . . . . . . . . . . . . . . . . . 79. SetUpMenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.1Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.2CountdownTimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.3Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 9.4Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.5HiAlarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.6LoAlarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.7AutoPowerOff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.8ClockAlarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 9.9Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 910.BeforeUsingYourThermometer . . . . . . . . . . . . . . . . . . . . . . . . . 1011. TakingTemperatureMeasurement . . . . . . . . . . . . . . . . . . . . . . . . 1112.ProbeSelection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1113.CareoftheThermometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1414.SealedforLifeBattery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    15. SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    FIGURES 1—PX22LPenetrationProbe . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 2—AK28MFastAirProbe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3—ST23LBetweenPackProbe . . . . . . . . . . . . . . . . . . . . . . . . 12 4—SK21MSurfaceProbe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 5—IT21LDeepFatProbe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 6—ST38LHeavyDutyPalletProbe . . . . . . . . . . . . . . . . . . . . . . 13 7—DX31LFoodSimulantProbe . . . . . . . . . . . . . . . . . . . . . . . . 13

  • 2

  • 3

    GENERAL INFORMATION

    Itisrecommendedthatyoureadthesafetyandoperationinstructionsbeforeusingthisinstrument.

    WARNINGTOAVOIDELECTRICSHOCKDONOTALLOWANYPROBEORSENSORTOCOMEINTOCONTACTWITHLIVEELECTRICALPOWERCONDUCTORSWITHVOLTAGESINEXCESSOF30VACRMSOR60VDC.

    TOAVOIDDAMAGEORBURNS,DONOTMAKETEMPERATUREMEASUREMENTSINMICRO-WAVEOVENS.

    CAUTIONTemperature Measurement ProbesThisprecisioninstrumenthasbeendesignedforusewiththeextensiverangeofComarktemperatureprobes.Theuseofotherprobesmayimpairtheperformanceandaccuracyoftheinstrument.FulldetailsofComarkprobesandsensorscanbeobtainedfromComarkCustomerSupportdepartmentoryourlocaldistributor.

    ThisinstrumentismanufacturedinaccordancewiththeCompany’sISO9000:2000QualityApprovedSystem.

    ThisinstrumentcomplieswiththeElectromagneticCompatibilityDirectiveEN61326-1.

    DeclarationsofConformityavailable.ContactComarkCustomerSupportoryourlocalDistributor.

    Inlinewithitspolicyofcontinuousdevelopment,ComarkInstrumentsreservestherighttoaltertheinstrumentspecificationwithoutpriornotice.FurtherinformationisavailablefromComarkInstrumentsoryourdistributor.

    ISO 9001FM 29700

  • 4

    CALIBRATION, CERTIFICATION AND SERVICE

    CertificationComarkcanprovidecertificatesofcalibrationforitswholeproductrange,tosuitISO9000andotherqualityassuranceprocedures,foodhygieneregulations,HACCPsandenvironmentalregulations.ComarkcertificatesareproducedbyindependentqualitycontrolledprocesseswhichcompareproductperformanceagainstagreedNationalStandards.ForpeaceofmindandbestpracticeComarkrecommendannualcertification/recalibration.

    Twolevelsofcertificationareavailableforinfra-redtemperatureandnontempera-tureinstruments,excludinghumidity:-UKAScertificatesviaanexternalaccreditedlaboratoryandNPLtraceablecertificatesfromtheComarkcalibrationlaboratory.

    Threelevelsofcertificationareavailableforcontacttemperatureproductsandthesearedetailedhere:

    a) UKAS Temperature CertificationTheComarkUKAS(UnitedKingdomAccreditationService)accreditedtemperaturecalibrationlaboratoryisoneofthefinestintheUK.ComarkUKAScertificatescanofferthelowestuncertaintyof0.01°Candprovideindependentproofofcorrectcalibrationusingequipmentandproce-duresauditedbyUKASinspectors.TheequipmentusedisfullytraceabletotheNationalPhysicalLaboratory.

    b) UKAS Humidity CertificationInadditiontotheComarktemperaturelaboratory,thehumiditylaboratorycontinuesthetraditionofhighaccuracycertificationandawiderange(25%to90%RH)withuncertaintiesof2.8%ofreading.ThisrangealsoincludesDewpointmeasurements.

    c) NPL Traceable CertificationComarkNPLcertificatesaretraceabletotheNationalPhysicalLabora-toryandcanofferuncertaintyaslowas0.3°C.

    ConformanceCertificatesofconformancecanbesuppliedfornew,servicedandrecalibratedinstruments.Theseconfirmthatinstrumentsarewithintheiroriginalmanufacturedspecification.

    Service/RepairsRegularservicingandanyrequiredrepairs,underwarrantyorafter,areavailablefromtheComarkServiceDepartment.

    FormoreinformationonallComarkcertification,calibrationandservicefacilitiespleasecallComarkCustomerSupportorcontactyourlocaldistributor.

    No. 0451

  • 5

    1. INTRODUCTION TO THE N9005 AND N9094

    TheN9005andN9094handheldthermometersrepresentaleapforwardinspecificationandperformanceandhaveanumberofnewfeatures.

    BothinstrumentshavetheexistingComarkpremiumcaseforultimatestrengthanddurability.Thisoffersthehighestlevelofstrength,durabilityandprotectionagainstdustandwateringress.TheinstrumentsarecertifiedtomeetIP68.Featuresincludeaclock,countdowntimer,timeralarms,highandlowtemperaturealarms,selectablesensortype,variableauto-offanddatahold.

    2. N9005 AND N9094 THERMOMETERS WITH BIOCOTE® PROTECTION

    BioCote®isasilverbasedanti-microbialagentthatisimpregnatedintotheinstrumentcaseduringmoulding.BioCote®effectivelyinhibitsthefunction,growthandreproductionofawiderangeofmicro-organismsanditsprotectionisincreasinglyacceptedaspartofHACCP,duediligenceandhealthandsafetyprocedurestoreducecrosscontamination.BioCote®protectionlastsforthelifeoftheinstrument,becausetheagentispresentthroughoutthecaseplasticandcannotrubofforbewashedorleachedout.

    Biocote®activeingredientsareregisteredwiththeUSEnvironmentalProtectionAgency(EPA).

    AlthoughBioCote®inhibitsthegrowthofmicro-organismsontheinstrumentcase,itdoesnotprotectindividualsagainstsuchharmfulorganismsorremovetheneedtomaintainthehigheststandardsofpersonalandproducthygieneandcleanliness.

  • 6

    3. OVERVIEW OF DISPLAY SYMBOLS

    A–IndicatesthatHOLDhasbeenselected.B–SettingmenuitemtoON/OFFC–IndicatesLowBattery.D–TheSetUpmenuisactive.E–Indicatesprobetype.F–MainDisplayforTemperatureG–ClockIndicator.H–Displayfortimeandcountdowntimer.I–Indicatescountdownactive.J–Alarmindication

    4. CONNECTING PROBES TO THE THERMOMETERS

    TheN9094thermometerhasaLumbergsocketforprobeconnectionandofferaclassleadingchoicebetweenComarkthermistor,typeTthermocoupleortypeKthermocoupleprobes.

    TheN9094thermometerwillautomaticallydetectthesensortypeoftheprobeconnected.TheLCDwillthenindicatethermistorprobesas“PST”,andtypeTorKthermocoupleprobesas“T”or“K”.

    TheN9005thermometerhasasub-miniatureconnectorforprobeconectionandoffersthechoicebetweentypeTandtypeKthermocoupleprobes.

    Note:TypeTorKmustbeselectedinthesetupmenu,seeSection9.

    A B C D E

    F

    G H I

    J

  • 7

    5. SWITCHING ON AND OFF

    Display Test/Mode

    The‘ON/OFF’buttononthekeypadisusedtoswitchonthethermometer.Afterashortpausethedisplaywillshowthecurrentmodeloftheinstrumentandwillthenreverttonormaltemperaturedisplay.ThecurrenttimeisalsopermanentlydisplayedontheLCD.

    6. DATA HOLD

    TheN9005andN9094thermometershaveadataholdfacility.Ifthe‘II’buttonispressedatanytimeduringnormaltemperaturemode,thecurrenttemperaturewillbeheldonthedisplay.Toreleasepress‘II’.

    7. COUNTDOWN TIMER

    TheN9005andN9094thermometershaveacountdowntimer.ThisfunctionisactivatedfromtheSetUpMenu–seesectionbelow,andcanbecalibratedinseconds/minutesorhours.Pressingthecountdowntimerbuttonwillstartthisfunction.Pressingthebuttonagainduringthecountdownperiodwillstopthecount.Theinstrumentwillbeepfor30secondsattheendofacountdownperiodtoindicatetotheuserthatithasbeencompleted.

    8. AUTO SWITCH OFF/POWER SAVING MODE

    TheN9005andN9094thermometershaveaselectableauto-switchofffeature.Thisissetto3minutesfromthefactory.PleaserefertoSetUpMenuforinstructionsonchangingthissetting.

  • 8

    9. SET UP MENU

    TheN9005andN9094thermometershaveahiddenmenustructurethatisusedtosetupanumberofparameterswithintheinstrument.

    TheSetUpMenucanbeenteredwhentheinstrumentisswitchedon.

    Iftheinstrumentisalreadyon,itmustbeturnedoff.

    Pressthecountdownbuttonandkeepthishelddownwhilepressingtheonbuttontoswitchtheinstrumenton.Afterthethermometerhascompleteditsself-checks,itwillenterthemenu.

    N.B.AftereachmenuitemusetheON/OFFbuttontosaveandmovetothenextmenuitem.

    Clockselectionisthefirstiteminthemenu:

    9.1 Clock – the display will show ‘CLOC’

    Usethecountdowntimerbuttonandtheholdbuttontoadjusttheclocktimeupordown.Theclockitselfisa12hourclockwithsettingforAMorPM.Movingthrough12noonwillchangetheclockfromAMtoPMandsoon.

    9.2 Countdown Timer – the display will show ‘Cd’

    Usethecountdowntimerbuttonandtheholdbuttontoselecttimeronoroff.Usetheholdandcountdowntimerbuttonstoscrollupanddowninsecondsupto5minutesthenin1minuteintervalsupto24hours.

    9.3 Scale – the display will show ‘SCAL’

    AfterenteringtheSCALEmenu,usethecountdowntimerbuttonandtheholdbuttontoscrollbetweentheavailablescales.

  • 9

    9.4 Type – the display will show ‘tYPE’

    UseeitherthecountdowntimerbuttonortheholdbuttontoscrollbetweenthermistorsensorspluseithertypeTorKthermocouples.ThedefaultisthermistorandtypeT,withthelatterforprobeswithbrownorblueleads.ThermistorandtypeKisavailableforprobeswitheitherredorgreenleads.

    9.5 High Alarm – the display will show ‘HIAL’

    Usethecountdownbuttonandtheholdbuttontoselecthighalarmonoroff.UsetheholdandcountdowntimerbuttonstoscrollupordownindegreesCorFtothelimitsofthechosenthermocoupletype.

    9.5 Low Alarm – the display will show ‘LOAL’

    Usethecountdownbuttonandtheholdbuttontoselectlowalarmonoroff.UsetheholdandcountdowntimerbuttonstoscrollupordownindegreesCorFtothelimitsofthechosenthermocoupletype.

    9.7 Auto Power OFF – the display will show ‘AOFF’

    Usethecountdowntimerbuttonandtheholdbuttontoscrollbetweentheavailableoptionsof3,10,30minutes.

    9.8 Clock Alarms

    Usethecountdownbuttonandtheholdbuttontoselectclockalarm1,onoroff.Useholdandcountdowntimerbuttonstosetthealarmtime-12hourclock.Repeatforalarms2and3.

    9.9 Resolution

    Usethecountdownandtheholdbuttontoselectordeselecttemperaturedisplayinhighresolution0.1°orlowresolution1°.

  • 10

    10. BEFORE USING YOUR THERMOMETER

    Beforeusingthethermometermakesureyouhavetherightprobeforthejob.Therearespecificprobesformeasuringproduct,air,‘betweenpack’,surfaceanddeepfattemperatures.Forpermanentinstallationthereisalsoachoiceof‘damped’sensorsandfoodsimulants.

    TheN9094iscompatiblewiththermistorandtypeTandKthermocouplesensorstogiveultimatesystemaccuracyandfastestspeedofresponse.ItiscompatiblewithallexistingN9091/N9092probesandthermistortestcaps.Itselfconfigurestotheparticularsensortype.Allyouhavetodoischoosetherightprobeforthejob.

    TheN9005iscompatiblewithallindustrystandardtypeTandtypeKprobeswithsub-miniatureconnectors.

    BoththeN9005andtheN9094canbetestedusingaC9040thermocouplesimulator.(C9040KtypeKsub-miniature,C9040TMtypeTsub-miniatureandC9040TLtypeTLumberg).

    Beforemeasuringproducttemperaturesmakesurethattheprobeisadequatelycleaned.Usebactericidalwipes(PW175T-Tubof175probewipesorPW70T-tubof70probewipes)orimmersetheshaftinasuitablefoodqualitysterilisingsolution.Remembertoreadtheinstructionscarefullyandallowthespecifiedtime.

    Thermometersneedtobecheckedatregularintervalstoconfirmthattheiraccuracyiswithinspecification.Itisrecommendedthatthermometersareregularlychecked(everythreemonths).TheN9094canbecheckedbycomparingthemeasuredreadingwithacertifiedtestcap.

    ComarktestcapsallowyoutocheckN9094instrumentaccuracyatspecificpointstoimproveyourconfidenceinthemeasurement.Thesecanbeusedonafrequentbasisifrequired.

    N9005andN9094thermometersandtheirprobescanalsoberegularlycheckedusingmeltingice.Simplyuseabeakerfulloficecubes,addcoldwaterandstirforafewminutes.Insertthepenetrationprobeandstiritforafurthertwominutes.Thethermometershouldthenreadbetween–0.5°Cand+0.5°C.TheaccuracycanbecheckedagainstareferencethermometersuchastheKM20REF.Makesureyouusesufficienticecubesandstirlongenoughforthewatertocooldownproperly.

  • 11

    Shouldyouwishtochecktheaccuracyofyourthermometeratothertemperaturesitismoredifficultwithoutsophisticatedequipment.

    IfyouintendtouseyourthermometertocheckotherequipmentorprocessesaspartofaBS5750/ISO9000procedureortocheckcompliancewithothercodesofpractice,youmayrequireeitherUKASorNationalStandardscertification.Comarkcanoffercalibrationtobothstandards.PleasecontactourUKASaccreditedlaboratoryforfurtherdetails.

    11. TAKING A TEMPERATURE MEASUREMENT

    Positiontheprobeintheplaceyouwishtotakethemeasurementandrecordareadingwhenthetemperaturehasstabilised.

    Note: Donotexposethermistorprobestotemperaturesoutsidetherange–80°Cto+150°Candavoidthermallyshockingprobesbyrapidheatingorcooling.

    12. PROBE SELECTION

    Penetration ProbesForproducttemperaturesusethestainlesssteel‘penetration’probe.Inserttheprobeatleast25mmintotheproductorpreferablytothecentreoftheproduct.Allowsufficienttimeforthereadingtostabilise.Typicallyyouwillneedtowait30secondsinaliquidandupto2minutesinadrypowder.Trysomeexperimentsyourself.

    Note:Toavoidcrosscontaminationproperlycleanyourprobeaftereachmeasurement. Order: Thermistor - PX22L (White cap), PX23L (Red cap), PX24L (Green cap) or PX25L (Blue cap) Thermocouple - PK24M, PT22M

    FIG.1 PX22L PENETRATION PROBE (MAX. TEMP. +150°C)

  • 12

    Fast Air ProbeForairtemperaturesalwaysusethefastresponseairprobes.Trytoplacetheprobeinamovingairflow,orifyoucanwavetheprobearoundtospeeduptheresponse.(Stillairisaverygoodthermalinsulatorsoinaclosedrefrigeratoryoumayneedtowaitminutesforthereadingstofinallystabilise).Order: AK28M

    FIG.2 AK28M FAST AIR PROBE (MAX. TEMP. +250°C)

    Between Pack ProbeForfasttemperaturechecksofpackagedgoodswithoutdamagingthepack,thisflexibleprobeshouldbeplacedbetweentwosimilarpacks.Itcanbeshapedtoexactlyfittherequiredsurfaces. Order ST23L

    FIG.3 ST23L BETWEEN PACK PROBE(MAX. TEMP. +70°C)

    Surface ProbeFormeasuringthetemperatureofflatsurfaceslikechilleddisplaycasesorhotplates.Ensurethatthesurfaceiswipedcleanbeforeuse.Takeextracarewhenmeasuringhotsurfaces.Order: SK21M

    FIG.4 SK21M SURFACE PROBE(MAX. TEMP. +250°C)

  • 13

    Deep Fat ProbeFortemperaturemeasurementindeepfatfryers.Thisprobeismadefromaveryflexiblematerialthatcanbereshapedtimeandtimeagain.Cranktheshaftoftheprobesothatthefattemperaturecanbemeasuredwithoutyourhandneedingtobeabovethefatitself.Nevertakethemeasurementwithyourhanddirectlyabovethefat.Takeadequateprecautions.Remembertheprobewillbeveryhotonwithdrawalfromthefat.Order: IT21L

    FIG.5 IT21L DEEP FAT PROBE(MAX. TEMP. +400°C)

    Pallet ProbeThisprobeallowsthemeasurementofthetemperatureofcartonsonapallet.Bycarefullyslidingtheflattenedsurfacesoftheprobebetweenadjacentcartonstheirtemperaturecanbesampledwithouttheneedtounstackthepallet. Order: ST38L

    FIG.6 ST38L HEAVY DUTY PALLET PROBE(MAX. TEMP. +250°C)

    Food Simulant ProbeThesimulanthasbeendesignedtoreplicatethetemperaturecharacteristicsofthemosttemperaturesensitivepackagedfood.Leftpermanentlyinplaceitgivesultimateconfidencethatthemostrepresentativemethodofmonitoringisbeingused.Order: DX31L

    FIG.7 DX31L FOOD SIMULANT PROBE(MAX. TEMP. +70°C)

  • 14

    13. CARE OF THE THERMOMETER

    WARNING

    TheN9005andN9094thermometersaredustandwaterproofasstatedinthespecificationsandwillwithstandharshenvironments.Useadampclothorwarmsoapywatertoremovedepositsandpreventthemfromhardeningorbecomingsticky.Donotusesolvent-basedcleanersormethylatedspirits,etc.

    Caution:Donotplaceanythermometerinadishwasher.

    Never use the thermometer/probes:-

    Tomeasurethetemperatureoflivepowerconductors

    Tolift,leverorhammer

    Inamicrowaveoven…thedoorwillnotcloseproperlyandthemicrowaveradiationcoulddamagethethermometerandoven

    Inanormaloven…theplasticpartsmaymelt

    Don’ttrytocleantheplasticcaseorprobehandlewithanyformofsolvent...itmayattacktheplasticorcauseittobecomebrittle.Itisjustaseffectivetouseadampclothandnormalkitchencleaner…regularly.

    Don’tforgettoregularlycheckyourthermometer.Trytobuildcrosschecksintoyoureverydayoperationssothatpotentialproblemsareimmediatelyidentified.

    Always:

    Keepyourinstrumentandprobeclean.

    Usethecorrectprobefortheparticularjob.

    Removethebatteryifyouarestoringyourinstrumentforlongperiods.

    Takecarewheninsertingthepenetrationprobeintohardsubstances.Thesharppointcaneasilypenetrateflesh.keeprecordsofthetemperaturesyouhavemeasured.Itmayprovideusefulevidenceifthereiseveraproblem.

  • 15

    14. SEALED FOR LIFE BATTERY

    TheN9005andN9094havealonglifeLithiumcellfittedinsidethesealedbatterycompartmentonthereverseoftheinstrument.Thisisdesignedtolastupto10yearsinnormalusage.HowevershouldthesymbolBATappearsonthedisplaywhenthebatteryvoltagehasreduced.Batteryreplacementisrecommendedatthistime.ThiscanonlybecarriedoutbyComarkorApprovedDistributorsopleasereturntheinstrumentasappropriateforbatteryreplacement.

    Donotattempttochangethebatteryyourselfasthiswillresultindamagetotheinstrumentandalossofwaterproofrating.

  • 16

    15. SPECIFICATION

    SensorTypesThermistor(PST) TypeTThermocouple TypeKThermocouple MeasurementRange Thermistor–50°Cto+150°C,–58°Fto+302°FTypeT –200°Cto400°C,–328°Fto+752°FTypeK–200°Cto+1372°C,–328°Fto+2501°F

    Scales °Cand°F DisplayedResolution >-100°<+1000°C 0.1° ≤-100°≥+1000°C 1°

    SystemAccuracy(Thermistor)* *At+23°C/+73°F0°Cto+70°C,+32°Fto+158°F

  • Comark Instruments52 Hurricane Way,Norwich, Norfolk, NR6 6JB United KingdomTel: +44 844 815 6599Fax: +44 844 815 6598Email: [email protected]

    Website: www.comarkinstruments.com

    Comark InstrumentsPO Box 500, Beaverton, OR 97077 USATel: 503 643 5204Fax: 503 644 5859Email: sales@comarkusacom

    Website: www.comarkusa.com

    © Comark Instruments Stock No. 18128 Issue 7 (11/12)