2 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
PRODUCT FEATURES General
The EEG-1200 Electroencephalograph is designed for both clinical and research use.
Fields of application are:
Routine EEG with common provocation tests like photic stimulation and/or hyperventilation
EEG monitoring on intensive care units incl. emergency EEG and brain death diagnostic
Long term monitoring e.g. for epilepsy diagnostics
Polysomnography
Typical configuration of EEG-1200K for clinical use:
Main unit incl. isolation transformer
Shuttle PC with minimum
Intel Core 2 Duo, 3.0 GHz
2 GB RAM
500 GB HDD
38 channel amplifier JE-921A incl. Stand
Photic stimulation incl. control unit and Xenon flash lamp
19 TFT screen
Accessory kit
Software package for acquisition and review of EEG waveforms
Polaris-one database
Digital video option with camera
Cart with optional screen and camera mounting
Typical configuration of EEG-1200K for intensive care use:
Fanless medical grade Panel PC with17 touch screen display
Dual Core CPU, 1.5 GHz
2 GB RAM
160 GB HDD
38 channel amplifier JE-921A with build in SpO2 and CO2
Polaris.one database
Light weight cart designed for ICU environment
Video option available
QP-160AK Trend Program with aEEG is recommended for neonatal ICU
(For details of the composition in regards of standard and optional equipment please see technical data section)
3 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Acquistion
Amplifiers for every requirement EEG-1200 can be delivered with a wide range of amplifiers, from a 32 channel routine amplifier to a 256 channel
amplifier for intracranial long term recordings.
Multifunction junction box for clinical routine
The electrode junction box JE-921A has 25 EEG, 14 bipolar (7 pairs), 4 DC, SpO2 and CO2 connectors. Its sibling
JE-910A comes with 25 EEG, 4 Extra and 3 bipolar connectors.
PSG junction box (option)
The PSG amplifier JE-912AK with comfortable mini junction box JE-915A is an easy to use and flexible solution for
polysomnographic recordings.
Multi-channel amplifier (option)
The multi channel amplifier JE-120A with up to 256 channels and high speed sampling rates of up to 10 kHz is the
ideal solution for long term recordings or research purposes.
Standard amplifier PSG amplifier Mini junction boxes for
JE-921A JE-912AK incl. JE-915A multichannel amplifier
Input box converter (standard with multi channel amplifier) The input box converter QI-122AK provides LAN connectivity so it is easy to expand your EEG system. It connects all
amplifiers to the EEG-1200 via standard CAT5 Ethernet cabling. QI-122AK provides 16 additional DC inputs (multi
channel amplifier only) and connection for photic stimulator.
Wireless input unit, “Air-EEG” WEE-1000K (option) Up to 64 channels EEG data can be sent to the EEG unit by wireless
transmission. Wireless reduces stress to the patients and lets physicians
monitor the EEG data while the patients are moving freely. All options
(including digital video) of the EEG-1200 are available with the wireless
input unit. Innovative roaming functions and an optional SpO2 probe that
can be connected directly to the telemetry unit makes it suitable for all
kind of EEG- and PSG recordings.
EEG Recording Navigation You can set several measurements as one routine in advance and record the series of EEG data automatically. You
can program the pattern, recording time, photic stimulation mode and other items for each stage.
4 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Data review during acquisition Comparison mode: The EEG Scope function lets you look back into
and review previous epochs of EEG while simultaneously monitoring
current EEG acquisition. One side of a split screen shows previous EEG
epochs and the other side shows the current EEG.
On-line frequency mapping The EEG data can be shown as a frequency map during acquisition, so
you can easily check the frequency distribution.
Review Easy to use software interface EEG-1200 software has an easy to operate graphical user interface
which can be adapted to user requirements. Standard settings provide
an intuitive software layout making it easy to learn even for beginners.
Individual configuration of the software layout allows adapting EEG-
1200 to every need:
flexible arrangement of menu bars
colouring of waveforms
individual montage settings
Review of EEG waveforms Waveforms can be displayed in a variety of methods.
40 individual montages with up to 250 traces
Pre-defined or individual high / low cut filters.
Manual or automatic paging forwards and backwards is the standard
way of reviewing EEG waveforms. Other ways are the event bar or
DSA trend graph to navigate to interesting parts of the EEG recording.
Several measurement tools help to go into details.
A part of the waveforms can be copied into the note waveform
window to compare them to the current waveforms.
DSA trend graphs Frequency components of EEG waveforms and the amplitudes of each frequency are displayed as a DSA (Density
Spectral Array) trend graph on the review screen and EEG Scope. The DSA lets you find epileptic seizures of a
specific EEG frequency band over a long period of time at a glance. Up to 8 trend graphs can be displayed, each
with its own settings in regards of the electrodes shown.
5 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
3D voltage mapping Whole head maps provide a complete overview and a better interpretation
of the topography of EEG abnormalities. Just click on a detected pattern to
obtain 3D maps of the whole head. Click on a particular view to obtain a
series of maps showing the change over time.
EEG mapping The EEG mapping module QP-220AK provides amplitude and frequency
mapping of EEG waveforms
Real-time mapping during acquisition
Off-line mapping during review for selected intervals
Up to eight frequency maps (seven power/voltage maps at seven different frequency bands and one map of all
frequency bands)
Power/Voltage spectra for up to 32 channels of EEG waveform data
Edge frequency, average frequency, median frequency or peak frequency for each spectrum is indicated with a
mark.
Potential mapping Spectral analysis Frequency mapping
Reporting NeuroReport is the common reporting tool for all
neurological systems from Nihon Kohden. Various predefined
templates can be used to create customized reports for
different examination types, physicians or other criteria.
You can also create an EMU examination report which has
screen comments automatically assigned to seizure events.
A powerful auto-text function helps to create individual
reports. Saved within the database in a license free format,
the reports will be automatically converted into PDF after
finalizing.
6 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Digital video system (option) Patient image including sound can be recorded and reviewed
synchronized with the EEG waveform.
A wide range of different video formats (MPEG1 and MPEG4)
and resolutions (up to 720*576 pixel) are available.
Real time and fast forward mode (up to 20 times) for reviewing
the video
Comfortable clipping and snapshot functions allow reducing
the video data by creating new video EEG clips or attaching
the clipped video file to the original EEG recording.
BESA Epilepsy (option) BESA Epilepsy is a software for fast detection, visualization,
and evaluation of epileptiform EEG activities recorded non-
invasively from the human scalp.
Automated detection and clustering of brief epileptiform
discharges in 2 h epochs
It allows a very fast evaluation of long-term EEG data using
a new hyperclustering technique
Evaluation and reporting can be done within 5 minutes for
a 24 h interval
Advanced software options On-line and off-line Spike- and seizure detection software QP-251AK based on Persyst®
EEG Examination Software QP-150AK including spike- and seizure detection, voltage mapping of intracranial
recordings, and EEG TrendScope for remote review of recordings
Polysmith sleep analysis software
QP-251AK QP-150AK Polysmith
7 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Data integration and management Polaris.one Polaris.one allows you to easily manage EEG data and examination schedules.
SQL based database for EEG, EMG/EP, and ECG data
User interface can be scaled according to user needs
Calendar view and to-do lists
Copies of EEG data can be stored for personal or global use
Presentation functionality for case conferences
Auto-proposing for text input fields
Manual, semi-automatic, and automatic transfer and archive as background service
Integrated NeuroReport incl. PDF conversion
Offline functionality for acquisition and review stations
Multi-client-capability with user-related filtering of patients
Extensive user rights management (via groups)
Activity log records all changes to medical data
Integrated DVD/CD burning
EEG PortaView
EEG PortaView integrates review software into EEG data files and saves them on a portable storage device, so you
can review the EEG data on any Windows PC without special software. Just carry the EEG data and you can review
anytime and anywhere.
8 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
TECHNICAL DATA
DATA ACQUISITION
JE-921A/AG electrode junction box
Number of input jacks
EEG input jacks: 25
Extra inputs: 4
Bipolar input jacks (pairs):
7
The extra jacks (X1-X4) can be used as bipolar jacks. To select
extra jack or bipolar jack, refer to Section 5 System Programs .
DC input jacks: 4
SpO2 connector: 1
CO2 connector: 1
Input impedance
EEG input/extra input: 100 MΩ
DC input: 1.5 MΩ
Input circuit current: 5 nA or less
Internal noise level
EEG input/extra input: 1.5 μVp-p or less (0.53 to 60 Hz)
DC input: 10 mVp-p or less
CMRR
EEG input/extra input: 105 dB or greater (60 Hz)
Bipolar input: 100 dB or greater (60 Hz)
Low-cut filter: 0.08 Hz (Time constant: 2 s)
High-cut filter: 300 Hz (-18 dB/oct)
Offset tolerance: ±600 mV
AD conversion: 16 bits (97 nV/LSB)
Sampling and hold: All electrodes at the same time
Sampling frequency 100, 200, 500, 1000, selectable
9 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
JE-910A/AG electrode junction box
Number of input jacks
EEG input jacks: 25
Extra inputs: 4
Bipolar input jacks (pairs): 3
Input Impedance
EEG input/extra input: 100 MΩ
Input circuit current: 5 nA or less
Internal noise level
EEG input/extra input: 1.5 μVp-p or less (0.53 to 60 Hz)
CMRR
EEG input/extra input: 105 dB or greater (60 Hz)
Low-cut filter: 0.08 Hz (Time constant: 2 s)
High-cut filter: 300 Hz (−18 dB/oct)
Offset tolerance: ±600 mV
A/D conversion: 16 bits (97 nV/LSB)
Sampling and hold: All electrodes at the same time
Sampling frequency: 100, 200, 500, 1000 Hz, selectable
JE-120A electrode junction box (option)
Number of input jacks The JE-120A electrode junction box must be used with the
following mini junction box combinations.
Mini junction box
(Number of input jacks)
Total number of
input jacks
JE-208A (64) or
JE-225A (64)
64
64
JE-210A (128) or
JE-225A (64) and JE-226A (64)
128
128
JE-225A (64) and JE-226A (64)
and JE-227A 192
JE-225A (64) and JE-226A (64)
and JE-227A (64) and JE-228A (64) 256
Bipolar inputs 4
provided with JE-125A (alternative to JE-225A) or JE-208A
DC inputs 16
Additional inputs SpO2 and etCO2
10 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Maximum number of recorded
referential electrodes and sampling
frequency (clinical line)
256 (2 kHz), 128 (5 kHz), 64 (10 kHz)
Maximum number of recorded
referential electrodes and sampling
frequency (extended line for research)
256 (10 kHz)
Input range: ±3.2mV or ±12,8mV (selectable)
Input impedance: 200 MΩ
Internal noise level: Less than 1.5 μVp-p (0.53 to 120 Hz)
CMRR: 110 dB or greater (60 Hz) (EEG inout)
Time constant: 2 or 10 s, selectable
High-cut filter: 3000 Hz (−18 dB/oct), depending on the sampling frequency
Offset tolerance: ±1000 mV
A/D conversion: 24 bits
Sampling and hold: All electrodes at the same time
Sampling frequency: 10000 Hz (max)
Number of input jacks (for all amplifiers / input boxes):
Model /Inputs EEG EXTRA (X1-) Bipolar Respiration Reference (Z) DC
JE-208A 25 33 8 (4ch) 1
JE-213A / JE-214A 64 1
JE-215A / JE-226A /
JE-227A / JE-228A 64 1
JE-921A/AG 25 14 (7ch) * 3 (X5, X6, X7) 1 6 **
JE-910A/AG 25 4 3
* X1(+) to X4(+) inputs jacks can be used as unipolar inputs
**4 x DC + 1 x SpO2 input + 1 x CO2 input
Note: The extra jacks and the RESP jack for each item cannot be used simultaneously.
DATA PROCESSING
Sensitivity
EEG INPUT: OFF, 1, 2, 3 (2.5), 5, 7, 10, 15, 20, 30, 50, 75, 100, 150,
200 μ V/mm
DC INPUT: OFF, 10, 15, 20, 30, 50, 75, 100, 150, 200 mV/mm
Time constant (Low-cut filter):
0.001, 0.003, 0.03, 0.1, 0.3, 0.6, 1.0, 2.0, 5.0, 10.0 s
0.016, 0.03, 0.08, 0.16, 0.27, 0.53, 1.6, 5.3, 53, 159 Hz
(−6 dB/oct)
11 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
High-cut filter:
JE-921A electrode junction box
15, 30, 35, 60, 70, 120 (−12 dB/oct), 50 (RAPID), 300 Hz (− 18 d
B/oct)
JE-120A electrode junction box (option)
15, 30, 35, 60, 70, 120, 300, 600, 1200 (−12 dB/oct),
50 (RAPID), 3000 Hz (−18 dB/oct)
AC filter: 50 or 60 Hz, (rejection ratio: 1/25 or more)
Calibration waveform
Waveform shape: 0.25 Hz step wave or 10 Hz sine wave
Voltage: 2, 5, 10, 20, 50, 100, 200, 500, 1000 μV (×1000 for DC input
signal)
ECG elimination filter: Available in acquisition and review programs
Impedance check
Indication on the screen: All electrodes are displayed on the screen in electrode position
layout. Impedance for each electrode is displayed and
electrodes with impedance higher than the preset impedance
threshold are highlighted.
Indication on LED: LEDs on the electrode junction box with impedance higher than
the preset impedance threshold light.
Impedance threshold: 2, 5, 10, 20 and 50 kΩ
Pattern: 36 sets of programmable montages combined with
programmable individual amplifier settings
Reference electrode selector: A1 A2, A1 A2, A1 A2, A1 + A2, VX, AV (with
unsuitable electrode deletion function), Aav, Org, SD, BN and
OFF
Marking signal: Photic stimulation mark, Hyperventilation mark
DISPLAY
Display resolution: Up to 1600 dots × 1200 lines
Number of display channels:
Up to 64 and one mark channel
When multi-channel electrode junction box and mini junction
boxes are used:
Up to 250 and one mark channel
Display modes: Overwrite and page-by-page
Waveform display color: 16 colors
Waveform display on/off: Provided
Waveform position adjustment: Provided
12 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Waveform freeze: Provided
Patient image display: Available when the optional QI-120A Camera Capture Unit and
video camera are used.
Waveform sweep speed: 0.1, 0.2, 0.5, 1, 2, 5, 10, 15, 20, 30, 60 s or 5 min/page
Timing mark: 0.1, 1 s
Time scale: off, 0.2, 1 s
Event mark: Displays at the bottom of the screen
EEG scale: Provided
Dual display: Available
ACQUISITION PROGRAM
Timer function: Manual timer, recording timer, HV timer
Data storage device: Hard disk drive (standard)
PHOTIC SIMULATOR
Optional photic control unit: LS-120A
Flash lamp: LS-703A
Maximum flash energy: 1.28 J/Single
Stimulation modes: 3 automatic (30 steps, programmable), manual, and single
Mode of operation: Continuous operation with intermittent loading
Duty cycle: Max. 5 minutes continuous operation in 30 minutes
Automatic stimulation
Stimulus rate: 0.5, 1 to 33 (1 Hz steps), 50 and 60 Hz
Stimulation period: 1 to 99 seconds in 1 second steps
Pause period: 1 to 30 seconds in 1 second steps
Manual stimulation Manually set frequency and stimulation period
Photic frequency: 0.5 Hz, 1 to 33 Hz in 1 Hz steps, 50 and 60 Hz
Stimulation time: 1 to 99 s in 1 second steps and continuous stimulation
(FREE: Max. 5 min)
Pulse mode: Normal, random, and double
Random stimulation: 1 to 33 Hz in 1 Hz steps within ± 50%
Single stimulation: Manual key operation single stimulation or automatic single
stimulation by external trigger signal
Trigger input: TRIGGER IN connector (1 to 5 V)
Trigger output: TRIGGER OUT connector (3 V or more)
13 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
DIGITAL VIDEO UNITS (OPTOIN)
MPEG4 encoding: NF-QV4/STD
Video resolution: up to 640 × 480 pixels (NTSC), up to 720 × 576 pixels (PAL)
Digital video software: QP-110AK
Software encoded MPEG4:
Windows Media 9
Time delay between EEG waveform and video image: within 100
ms in still mode
Camera capture unit: QI-120A
Video storage device: usage of existing system/data hard disk
External USB video capture unit: QI-120A
USB I/F: USB 2.0 high speed
Video input:
Composite × 1 ch RCA pin jack
S-Video × 1 ch MINI-DIN 4-pin jack
720 × 576, 50 fields/s (PAL)
Data transfer speed 23.1 MB/s maximum
Audio input:
2 ch (stereo mini jack × 1)
16 bit sampling 44,100 Hz
Data transfer speed 177 kB/s maximum
Power: 5 V ± 5%, less than 500 mA (using USB bus power)
Data compression and required disk space (Windows Media 9 encoding):
Resolution Frames File Size Audio Capture
Frame 2 320x240 pixels
(NTSC/PAL) 2 frames/s 0.8 MB/min
Monaural, 48 kbps,
44kHz sample rate
Frame 15 320x240 pixels
(NTSC/PAL} 15frames/s 1.2 MB/min
Monaural, 48 kbps,
44kHz sample rate
Low 320x240 pixels
(NTSC/PAL)
29.97 frames/s (NTSC)
25 frames/s (PAL) 1.5 MB/min
Monaural, 48 kbps,
44kHz sample rate
Medium 320x240 pixels
(NTSC/PAL}
29.97 frames/s (NTSC)
25 frames/s (PAL) 2.25 MB/min
Monaural, 48 kbps,
44kHz sample rate
High 320x240 pixels
(NTSC/PAL)
29.97 frames/s (NTSC)
25 frames/s (PAL) 10 MB/min
Stereo, 192 kbps,
44kHz
Extra High 320x240 pixels
(NTSC/PAL)
29.97 frames/s (NTSC)
25 frames/s (PAL) 20 MB/min
Stereo, 192 kbps,
44kHz
High Res 640x480 pix (NTSC)
720x576 pix (PAL)
29.97 frames/s (NTSC)
25 frames/s (PAL) 24 MB/min
Stereo, 192 kbps,
44kHz
14 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
REVIEW PROGRAM
Changeable items: Montage, sensitivity, high-cut filter, time constant, reference
electrode, and display speed
Jump functions: Specified event, page by page, and specified time
Display modes:
Continuous, high speed, high speed with pause, manually page-
by-page, manually second-by-second and manually waveform
centering
Display information: Event, channel number, montage and comment
ELECTROMAGNETIC COMPATIBLITY
IEC 60601-1-2: 2001
IEC 60601-1-2 Amendment 1: 2004
CISPR11 GROUP 1 CLASS B: 2003
CAN/CSA C22.2 No.60601-1-2-03
POWER REQUIREMENTS
Line voltage: SM-120AK isolation unit: AC 220 to 240 V
Line frequency: 50/60 Hz
Power consumption:
750 VA (for PC unit and display)
1 kVA (PC unit, display, photic stimulator control unit and
analog output unit)
SAFETY
Safety standard IEC 60601-1: 1988 CAN/CSA C22.2
No.601.1B-90
IEC 60601-1
Amendment 1: 1991
CAN/CSA C22.2
No.60601-1-1-02
(R2002)
IEC 60601-1
Amendment 2: 1995
CAN/CSA C22.2
No.60601-2-26-04
IEC 60601-1-1: 2000 EN 60601-1: 1990
IEC 60601-1-2: 2001 EN 60601-1
Amendment 1: 1993
IEC 60601-1-2
Amendment 1: 2004
EN 60601-1
Amendment 2: 1995
15 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
IEC 60601-2-26: 2002 EN 60601-1
Amendment 13: 1996
CAN/CSA C22.2
No.601.1-M90: 1990 EN 60601-1 -1: 2001
CAN/CSA C22.2
No.601.1-1S1-94 EN 60601-2-26:2003
Type of protection against electric
shock: Class I
Degree of protection against electric
shock (JE-921A)
Type BF applied part: Electrode jacks, bipolar jacks, SpO2 connector, CO2 connector
Degree of protection against electric
shock (JE-207A/209A/212A)
Type CF applied part: Electrode jacks, bipolar jacks
Degree of protection against harmful
ingress of water: Non-protected (IPX0)
Degree of safety of application in the
presence of a flammable anaesthetic
mixture with air or oxygen or
nitrous oxide:
Not suitable for use in the presence of a flammable anaesthetic
mixture with air or oxygen or nitrous oxide
Mode of operation: Continuous
DIMENSIONS AND WEIGHT
JE-921A, electrode junction box: 185 (W) × 72 (D) × 167 (H) mm, 1.0 kg
PC unit:
200 (W) × 300 (D) × 185 (H) mm, 6.5 kg
The dimensions and weight differ according to model.
Refer to the Operator’s manual of the PC unit.
SM-120AK, isolation unit: 181 (W) × 320 (D) × 140 (H) mm, 13 kg
LS-120A/AG, photic stimulator
control unit: 155 (W) × 300 (D) × 75 (H) mm, 4.10 kg
QD-120A, analog output unit: 155 (W) × 300 (D) × 75 (H) mm, 2.32 kg
NF-KD1200/1, cart: 565 (W) × 625 (D) × 830 (H) mm, 44 kg
16 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
OPERATION CONDITIONS
Temperature: 10 to 35˚C (50 to 95˚F)
Humidity: 30 to 80% (noncondensing)
Atmospheric pressure: 70 to 106 kPa
TRANSPORT AND STORAGE CONDITIONS
Temperature: -20 to +60˚C (-4 to +140˚F)
Humidity: 10 to 80%
Atmospheric pressure: 70 to 106 kPa
ELECTROMAGNETIC EMISSIONS
The EEG-1200’s essential performances in EMC standard satisfy the following criteria.
EEG-1200 is intended for use in the electromagnetic environment specified below.
The customer or the user of the EEG-1200 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment guidance
RF emissions
CISPR 11 Group 1
The EEG-1200 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11 Class B*
The EEG-1200 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes. *
Harmonic emissions
IEC 61000-3-2 Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
* When the optional JE-207A/209A/212A electrode junction box is used, the compliance for RF emission is class A.
The EEG-1200 with JE-207A/209A/212A is suitable for use in all establishments excluding domestic establishments
and those directly connected to the public low-voltage power supply network that supplies buildings used for
domestic urposes.
17 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Electromagnetic Immunity
The EEG-1200’s essential performances in EMC standard satisfy the following criteria. This Model EEG-1200 is intended for use in the electromagnetic environment specified below. The customer or the user of the EEG-1200 should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality should be
thatof a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 5 cycles
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 s
<5% UT (>95% dip in UT)
for 0.5 cycle
40% UT (60% dip in UT)
for 1 cycle
70% UT (30% dip in UT)
for 2 cycles
(for 1 cycle*1)
<5% UT (>95% dip in UT)
for 5 s
Mains power quality should be
thatof a typical commercial or
hospital environment.
If the user of the EEG-1200
requires continued operation
during power mains
interruptions, it is recommended
that the EEG-1200 be powered
from an uninterruptible power
supply or a battery.*2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: UT is the AC mains voltage prior to application of the test level
*1 When the QI-120A is connected to the EEG-1200.
*2 It is recommended that the EEG-1200 be powered from an uninterruptible power supply or a battery if the power
quality is not stable.
18 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2.5
GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the EEG-1200, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d = 1.2 P
d = 1.2 P 80 MHz to 800 MHz
d = 2.3 P 800 MHz to 2.5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey.*1,
should be less than the compliance level in each
frequency range. *2. Interference may occur in
the vicinity of equipment marked with the
following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
*1 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
the EEG-1200 is used exceeds the applicable RF compliance level above, the EEG-1200 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the EEG-1200.
*2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than for all other
functions.
19 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Avoiding Electromagnetic Interference (Impedance Respiration)
Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological
limitations do not allow immunity levels higher than 1 Vrms for conducted disturbances induced by RF fields.
Electromagnetic fields with conducted disturbances above 1 Vrms may cause measurement error. Do not use
electrically radiating equipment near the impedance respiration measurements.
When the JE-921A electrode junction box is connected to the EEG-1200 electroencephalograph
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment – guidance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
0.3 V/m
The JE-921A that is connected to the EEG-1200
must be used only in a shielded location with a
minimum RF shielding effectiveness and, for each
cable that enters the shielded location, a
minimum RF filter attenuation of 31 dB over the
frequency range 80 MHz to 2.5 GHz. Refer to the
note below this table.
Field strengths outside the shielded location from
fixed RF transmitters, as determined by an
electro-magnetic site survey*, should be less than
3 V/m. Refer to the note below this table.
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection structures, objects and people.
NOTE 2: It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to
assure that they meet the minimum specification.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength outside the shielded location in which the JE-921A that is connected to the
EEG-1200 is used exceeds 3 V/m, observe the JE-921A that is connected to the EEG-1200 to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as relocating the JE-921A that is
connected to the EEG-1200 or using the shielded location with a higher RF shielding effectiveness and filter attenuation.
NOTE
The JE-921A that is connected to the EEG-1200 should be used in a shielded location with an RF filter
attenuation of 31 dB over the frequency range 80 MHz to 2.5 GHz. If the JE-921A that is connected to the
EEG‑1200 is used outside shielded location, use it in a location where the measured field strength is less than
0.3 V/m. To measure shielding effectiveness, consult a service engineer.
20 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
The EEG-1200 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the EEG-1200 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the EEG-1200
as recommended below, according to the maximum output power of the communications.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
System Composition for EMC Test
The EEG-1200 electroencephalograph is tested to comply with IEC 60601-1-2: 2001 with the following composition.
If any part which is not specified by Nihon Kohden is used, the EMC specifications may not comply.
Units Cable length
Electrode junction box JE-921A 5.0 m
Photic stimulator control unit LS-120AJ/AK 0.7 m
Flash lamp assembly LS-703A 5.0 m
Mini electrode junction box JE-922A 5.0 m
Camera capture unit QI-120A 0.8 m
Analog output unit QD-120A/AK 0.7 m
Input box converter QI-122A/AK 10 m
Electrode junction box JE-212A 5.0 m
21 Neurofax EEG-1200K | Product Features / Technical Data (Ver. 120131)
STANDARD COMPOSITION
For routine study:
Main unit, EEG-1200K, PC unit,
Electrode junction box, JE-921A/AG or JE-910A/AG
Accessory Kit (INT or GER version)
For 64 channel monitoring:
Main unit, EEG-1200K, PC unit,
Electrode junction box, JE-207A
Input box converter, QI-122AJ/K
For 128 channel monitoring:
Main unit, EEG-1200K, PC unit,
Electrode junction box, JE-209A
Input box converter, QI-122AJ/K
For 192 channel monitoring:
Main unit, EEG-1200K, PC unit,
Electrode junction box, JE-212A
Input box converter, QI-122AJ/K
STANDARD ACCESSORIES
German (GER) accessory kit International (INT) accessory kit
Description Qty Order code Description Qty Order code
Universal EEG cap 1 BE-0007 Collodion electrodes (12 pcs) 2 H527
Chin strap 1 BE-0008 Ear electrodes 3 NE-313AG
Bridge electrodes 23 BE-005W Fast Clip ECG electrodes 1 H068
Electrode cable 25 2144-012358F Cable for Fast Clip 1 K512A
Electrode felt pad 1 NKD-Filz Cardiocream 1 F010/1
Rubber rings 1 2144-011555 Skinpure 1 F020/1
Ear electrodes 3 NE-313AG Elefix 1 F510
Fast Clip ECG electrodes 1 H068
Cable for Fast Clip 1 K512A
Cardiocream 1 F010/1
Skinpure 1 F020/1
OPTIONAL SOFTWARE
Spike detection and evaluation, BESA Epilepsy
Spike and seizure detector software, QP-251AK,
EEG examination support software, QP-150AK
Trend program, QP-160AK
Analysis software, EEGFocus
Sleep analysis software, Polysmith