New RF technologies for very high field and very low field magnetic resonance imaging

Andrew Webb

C.J. Gorter Center for High Field MRI

LEIDEN UNIVERSITY MEDICAL CENTER

19-Apr-181

Page 2: New RF technologies for very high field and very low field

Outline

1. How does RF technology fit into MRI?

Similarities and differences to radar/antennas

2. Advantages but also RF challenges of very high frequency MRI

Interactions with the body

3. What do we use very high frequency MRI for in the hospital?

4. High frequency components, current and future….

Expert advise welcome!

5. And now for something completely different….

Ultra low field MRI, why and how

Page 3: New RF technologies for very high field and very low field

Magnet (DC), gradient coils (100s kHz), RF coil (100s MHz)

Page 4: New RF technologies for very high field and very low field

Superconducting wire (tens of km, cooled with liq He)

Earths magnetic field: 50 mTesla

Typical clinical scanner: 3 Tesla

Ultra high field human scanner: 10 Tesla

High field NMR spectrometer: 30 Tesla

Page 5: New RF technologies for very high field and very low field

Shielded superconducting magnet

Page 6: New RF technologies for very high field and very low field

Why do we need a very strong magnet?

B0=3 Tesla f=128 MHz

Page 7: New RF technologies for very high field and very low field

dV -

dt

Faradays’law

Effects of a pulse of RF energy

Page 8: New RF technologies for very high field and very low field

What frequencies are we dealing with?

Page 9: New RF technologies for very high field and very low field

Similarities and differences with radar

scattering

FAR FIELD

z

yx

E

B

v

NEAR FIELD

dimensions <<l

Page 10: New RF technologies for very high field and very low field

Multiple coil arrays

circularly polarized

transmit

multiple channel

receive

Page 11: New RF technologies for very high field and very low field

Quadrature

mixer

I

Q

(A/D)1

(A/D)2

preamplifier

filters

variable-gain

amplifiers

s(t) demodulator

LPF

Mixer

sout(t)

Demodulator

sdemod(t)

Quadrature mixer

Oscillator LPF

LPF

90o

phase

shifter

IFcos t

sI(t)

sR(t)

Mixer

Mixer

s(t)

sdemod(t)

0 IFcos t

low-pass filter

Superheterodyne detector

Page 12: New RF technologies for very high field and very low field

Analogue-to-digital converters

SAR

16

14

12

10

8

6

4

2

104 105 106 107 108 109 1010

DS

Pipeline

folding/

interpolating

flash

resolution

(bits)

samples per second

Page 13: New RF technologies for very high field and very low field

High current (400 Amps) gradient power supplies

Page 14: New RF technologies for very high field and very low field

x

y

z

Bz

z

Bz

y

Bz

x

(x,y,z) = gB(x,y,z)

Imposing spatial information on precessional frequency

Page 15: New RF technologies for very high field and very low field

What are the main differences to radar and antennas?

Operate in the near field – so coils not antennas

Relatively low frequency so scattering is not relevant

Spatial resolution is not wavelength-related but depends upon

the gradient strength, since we impose a frequency/phase

difference onto the sample

Magnetic field produces the signal

Electric field produces heating!!

Page 16: New RF technologies for very high field and very low field

Part 2

Advantages and disadvantages of higher fields

Page 17: New RF technologies for very high field and very low field

The move to higher fields

Higher signal-to-noise

Higher spatial resolution

Higher spectral resolution

Increased image contrast

Page 18: New RF technologies for very high field and very low field

LUMC

Page 19: New RF technologies for very high field and very low field

LUMC

C.J.Gorter Center for High Field MRI (established 2008)

Mission: Develop new technology for clinical, medical research and

biomedical research in MRI

Technique – application in volunteers – application in patients

Rapid translation of hardware and software – weeks not years

Most rapid – eye coil construction to first patient ~2 weeks

Page 20: New RF technologies for very high field and very low field

Examples of 7T MRI projects in Leiden

MR angiography

small vessel diseaseCartilage for early

stage osteoarthritisCoronary artery

early stenosis

Page 21: New RF technologies for very high field and very low field

Does it work first time?

Page 22: New RF technologies for very high field and very low field

22

Effects of finite RF wavelength in tissue

1tissue

rfl

r~60

MuscleWavelength (cm)

frequency (MHz)

10

20

30

40

50

60

1.5 Tesla

3 Tesla

7 Tesla

Page 23: New RF technologies for very high field and very low field

RF inhomogeneityconstructive/destructive interference

23

l~12 cm

Page 24: New RF technologies for very high field and very low field

Image non-uniformities at high field

24

3T 4T 5T 6T 7T 8T 10T 12T

Page 25: New RF technologies for very high field and very low field

Increased SAR and heating at 7T

25

300 MHz128 MHz300 MHz128 MHz

SAR Temperature rise

Page 26: New RF technologies for very high field and very low field

EM modelling

Page 27: New RF technologies for very high field and very low field

EM modelling

T (C)Geometry B1+ (mT) SAR (W/kg)

Page 28: New RF technologies for very high field and very low field

Uveal melanoma and treatment planning

LUMC is Netherlands Center for uveal melanoma

200-300 patients per year

ultrasound

19-Apr-1828

Part 3

Page 29: New RF technologies for very high field and very low field

Treatment options

29

Enucleation

Radiotherapy

External beam

therapy

Page 30: New RF technologies for very high field and very low field

Ruthenium plaque therapy

For a patient to be eligible for ruthenium plaque brachytherapy the

maximum tumour prominence is 7 mm.

For tumour prominences greater than 7 mm the eye is removed.

19-Apr-1830

Page 31: New RF technologies for very high field and very low field

Ultrasound

31

Page 32: New RF technologies for very high field and very low field

Ultrasound

32

• Only 2D cross-sections

• Oblique cross-sections typically overestimate tumour size

• Inadequate for large, complexly, shaped tumours

• Low contrast between tumour and sclera

3D imaging technique with better contrast needed

Page 33: New RF technologies for very high field and very low field

High resolution imaging of the eye

19-Apr-1833

Page 34: New RF technologies for very high field and very low field

Uveal melanoma patients

19-Apr-1834

Page 35: New RF technologies for very high field and very low field

Two patients

19-Apr-1835

Page 36: New RF technologies for very high field and very low field

Two patients

Patient 1 – MRI-estimated tumour prominence was 6.2 mm

Patient 2 - MRI-estimated tumour prominence was 9.2 mm

For patient 1, this resulted a substantial change in treatment plan

from eye removal (based on the UBM scan) to treatment with

ruthenium plaque therapy

1-2 patients per week are now being referred to the 7T scanner

for treatment planning

A combined evaluation of all the different sequences

showed no extra-scleral extension for either patient.

19-Apr-1836

Page 37: New RF technologies for very high field and very low field

3T 7T

Imaging patient before cochlear implant

Page 38: New RF technologies for very high field and very low field

RF and image inhomogeneities

Part 4

Page 39: New RF technologies for very high field and very low field

B1 inhomogeneity – parallel transmit approach

19-Apr-1839

Page 40: New RF technologies for very high field and very low field

Parallel transmit approach

19-Apr-1840

A1, 1A2, 2

A3, 3

Page 41: New RF technologies for very high field and very low field

Multi-element dipole antennas

19-apr-1841

Page 42: New RF technologies for very high field and very low field

First attempts in vivo on the 7 Tesla

Page 43: New RF technologies for very high field and very low field

Parallel transmit approach

19-Apr-1843

Page 44: New RF technologies for very high field and very low field

Improving the RF amplifiers?

Class A/B amplifiers

Broadband

Class D?

Class E?

Class H?

On coil amplifiers to

avoid enormous power

losses in cables

Page 45: New RF technologies for very high field and very low field

Different types of amplifiers

Page 46: New RF technologies for very high field and very low field

Passively vs. actively fed dipoles

19-apr-1846

Page 47: New RF technologies for very high field and very low field

Passively vs. actively fed dipoles

19-apr-1847

Page 48: New RF technologies for very high field and very low field

5. Ultra-low field sustainable MRI for developing countries

19-Apr-1848

Page 49: New RF technologies for very high field and very low field

First prototype system 2018

Page 50: New RF technologies for very high field and very low field

The system

Page 51: New RF technologies for very high field and very low field

A Halbach magnet – NdBFe material

Page 52: New RF technologies for very high field and very low field

First iteration modelled

Central frequency 2.5 MHz

Page 53: New RF technologies for very high field and very low field

And built

Page 54: New RF technologies for very high field and very low field

First software written – open source platform

Page 55: New RF technologies for very high field and very low field

How to spatially encode i.e. get an image

Two ring magnet and rotate the outside one

Page 56: New RF technologies for very high field and very low field

Collaboration with signal processing from TU Delft

Page 57: New RF technologies for very high field and very low field

19-Apr-1857

Steps to improve performance in next 5 years

Page 58: New RF technologies for very high field and very low field

19-Apr-1858

Steps to improve performance in next 5 years

Page 59: New RF technologies for very high field and very low field