ob-dopplerultrasound

Upload: gayla-v-perillo

Post on 09-Apr-2018

218 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/7/2019 OB-dopplerultrasound

    1/36

    DOPPLER ULTRASOUNDDOPPLER ULTRASOUND

    in Pregnancyin Pregnancy

    Dr. Mohammed AbdallaDr. Mohammed Abdalla

    Egypt, Domiat HospitalEgypt, Domiat Hospital

  • 8/7/2019 OB-dopplerultrasound

    2/36

    First use of DopplerFirst use of Doppler

    ultrasonography to studyultrasonography to study

    flow velocity in the fetalflow velocity in the fetal

    umbilical arteryumbilical arterywas reported inwas reported in 19771977

    Doppler HistoryDoppler History

  • 8/7/2019 OB-dopplerultrasound

    3/36

    Doppler HistoryDoppler History

    Fitzgerald & Drumm.Fitzgerald & Drumm. Umbilical artery studiesUmbilical artery studies 19771977BMJBMJ

    EikEik--Nes et al.Nes et al. Fetal aortic velocimetryFetal aortic velocimetry : Dupplexscanner: Dupplexscanner

    19801980 LancetLancet

    Campbell et al.Campbell et al. UteroUtero--placental circulationplacental circulation: Dupplex: Dupplexscanner 1983scanner 1983 LancetLancet

    Wladimiroff et al.Wladimiroff et al. MCA / UA PI ratioMCA / UA PI ratio 19871987 OGOG

    Kiserud et al.Kiserud et al. Ductus venosus velocimetryDuctus venosus velocimetry 19911991LancetLancet

  • 8/7/2019 OB-dopplerultrasound

    4/36

    Basic principalsBasic principals

    Echoes from stationary tissues are the same fromEchoes from stationary tissues are the same frompulse to pulse.pulse to pulse.

    Echoes from moving objects exhibit slightEchoes from moving objects exhibit slight

    differences in the time for the signal to bedifferences in the time for the signal to bereturned to the receiver.returned to the receiver.

    These differences can measured asThese differences can measured asphase shiftphase shiftfrom which thefrom which the Doppler frequencyDoppler frequencyis obtained.is obtained.

  • 8/7/2019 OB-dopplerultrasound

    5/36

    TT11 : time of omitted signal: time of omitted signal ..

    TT22 : time of returned signal .: time of returned signal .

    TT22 TT11 == time difference or phasetime difference or phase

    shift .shift .

    from phase shift the Dopplerfrom phase shift the Doppler

    frequency is obtained.frequency is obtained.

    AS TIME DIFFERENCE DECREASE THEAS TIME DIFFERENCE DECREASE THEDOPPLER FREQUENCY INCREASE.DOPPLER FREQUENCY INCREASE.

  • 8/7/2019 OB-dopplerultrasound

    6/36

    pulse repetition frequencypulse repetition frequency

    (T2 T1)phase shiftphase shiftwith known beam / flow angle cancalculate flow velocity .

    T1

    T2

  • 8/7/2019 OB-dopplerultrasound

    7/36

    The time difference or phase shiftThe time difference or phase shift

    are then proceeded to produceare then proceeded to produce

    eithereithercolorflow displaycolorflow display or aor a

    Doppler sonogramDoppler sonogram

    Basic PrincipalsBasic Principals

  • 8/7/2019 OB-dopplerultrasound

    8/36

    Basic PrincipalsBasic Principals

    Doppler frequency is obtained by measuringDoppler frequency is obtained by measuring

    the time difference for the signal to be returnedthe time difference for the signal to be returned

    when reflected from moving scatterers .when reflected from moving scatterers .

    Doppler frequency increase if:Doppler frequency increase if:

    1.1. flow velocity increased .flow velocity increased .

    2.2. beam is more aligned to the direction ofbeam is more aligned to the direction of

    flow.flow.

    3.3. higher transducer frequency is used.higher transducer frequency is used.

  • 8/7/2019 OB-dopplerultrasound

    9/36

    (the angle(the angle qq between the beam and the direction ofbetween the beam and the direction of

    flow becomes smaller). This is of the utmostflow becomes smaller). This is of the utmost

    importance in the use of Doppler ultrasound.importance in the use of Doppler ultrasound.

    qq

    The angle of insonation

    Flow velocity

    3

    2

    1

    Factors affecting doppler frequency

  • 8/7/2019 OB-dopplerultrasound

    10/36

    (the angle(the angle qq between the beam and the direction of flowbetween the beam and the direction of flow

    becomes smaller). This is of the utmost importance in thebecomes smaller). This is of the utmost importance in the

    use of Doppler ultrasound.use of Doppler ultrasound.

    beambeam (A)(A) is more aligned thanis more aligned than (B)(B)

    The beam/flow angle atThe beam/flow angle at (C)(C) is almostis almost 9090 and there is a very poor Doppler signaland there is a very poor Doppler signal

    The flow atThe flow at (D)(D) is away from the beam and there is a negative signal.is away from the beam and there is a negative signal.

  • 8/7/2019 OB-dopplerultrasound

    11/36

    AliasingAliasing

    If a second pulse is sent before the first is received, the receiver cannot

    discriminate between the reflected signal from both pulses and aliasing

    occur.

  • 8/7/2019 OB-dopplerultrasound

    12/36

    So to eliminate aliasing TheSo to eliminate aliasing The

    pulse repetition frequency orpulse repetition frequency or

    scale is set appropriately forscale is set appropriately for

    the flow velocitiesthe flow velocities

    AliasingAliasing

  • 8/7/2019 OB-dopplerultrasound

    13/36

    The volume flow in the UAsThe volume flow in the UAs

    increases with advancingincreases with advancing

    gestation. The high vasculargestation. The high vascularimpedance detected in the firstimpedance detected in the first

    trimester gradually decreases. It istrimester gradually decreases. It is

    attributed to growth of placentalattributed to growth of placental

    unit and increase in the number ofunit and increase in the number ofthe functioning vascular channels.the functioning vascular channels.

    Basic PrincipalsBasic Principals

  • 8/7/2019 OB-dopplerultrasound

    14/36

    UsesUsesplays a vital role in the diagnosis ofplays a vital role in the diagnosis of

    fetalfetal cardiac defectscardiac defects ..

    assessment of the hemodynamicassessment of the hemodynamicresponses to fetalresponses to fetal hypoxiahypoxia andand

    anemiaanemia..

    diagnosis of otherdiagnosis of other nonnon--cardiaccardiacmalformations.malformations.

  • 8/7/2019 OB-dopplerultrasound

    15/36

    Blood supply provided byBlood supply provided bythe ovarian and uterinethe ovarian and uterine

    arteriesarteries

    Uterine Arteries: mainUterine Arteries: main

    branches of the internalbranches of the internaliliac arteriesiliac arteries

    Uterine Arteries: AscendUterine Arteries: Ascend

    through the lateral wall andthrough the lateral wall and

    anastomose with theanastomose with the

    ovarian arteriesovarian arteries

    Blood supply provided byBlood supply provided bythe ovarian and uterinethe ovarian and uterine

    arteriesarteries

    Uterine Arteries: mainUterine Arteries: main

    branches of the internalbranches of the internaliliac arteriesiliac arteries

    Uterine Arteries: AscendUterine Arteries: Ascend

    through the lateral wall andthrough the lateral wall and

    anastomose with theanastomose with the

    ovarian arteriesovarian arteries

    AnatomyAnatomyAnatomyAnatomy

  • 8/7/2019 OB-dopplerultrasound

    16/36

    Spiral Arteries: connect the maternal circulation to theSpiral Arteries: connect the maternal circulation to the

    endometriumendometrium

    Responsible for a 10 fold increase in blood flowResponsible for a 10 fold increase in blood flow

    Spiral Arteries: connect the maternal circulation to theSpiral Arteries: connect the maternal circulation to the

    endometriumendometrium

    Responsible for a 10 fold increase in blood flowResponsible for a 10 fold increase in blood flow

    AnatomyAnatomyAnatomyAnatomy

    Arcuate Arteries: Run Circumferentially around the uterusArcuate Arteries: Run Circumferentially around the uterus Uterus: Blood supply to anterior and posterior wallsUterus: Blood supply to anterior and posterior walls

    provided by the Arcuate arteriesprovided by the Arcuate arteries

    Radial Arteries: Extend from the arcuate arteries andRadial Arteries: Extend from the arcuate arteries and

    enter the endometriumenter the endometrium

    Arcuate Arteries: Run Circumferentially around the uterusArcuate Arteries: Run Circumferentially around the uterus Uterus: Blood supply to anterior and posterior wallsUterus: Blood supply to anterior and posterior walls

    provided by the Arcuate arteriesprovided by the Arcuate arteries

    Radial Arteries: Extend from the arcuate arteries andRadial Arteries: Extend from the arcuate arteries and

    enter the endometriumenter the endometrium

  • 8/7/2019 OB-dopplerultrasound

    17/36

    Conversion of small muscular spiral arteries into largeConversion of small muscular spiral arteries into large

    vascular channels transforms the uteroplacentalvascular channels transforms the uteroplacental

    circulation into a lowcirculation into a low--resistanceresistance--toto--flow system. Theseflow system. Thesehave a dilated and tortuous lumen, a complete absencehave a dilated and tortuous lumen, a complete absence

    of muscular and elastic tissue, no continuousof muscular and elastic tissue, no continuous

    endothelial lining.endothelial lining.

    Conversion of small muscular spiral arteries into largeConversion of small muscular spiral arteries into large

    vascular channels transforms the uteroplacentalvascular channels transforms the uteroplacental

    circulation into a lowcirculation into a low--resistanceresistance--toto--flow system. Theseflow system. Thesehave a dilated and tortuous lumen, a complete absencehave a dilated and tortuous lumen, a complete absence

    of muscular and elastic tissue, no continuousof muscular and elastic tissue, no continuous

    endothelial lining.endothelial lining.

    AnatomyAnatomyAnatomyAnatomy

  • 8/7/2019 OB-dopplerultrasound

    18/36

    Umbilical artery

    doppler

  • 8/7/2019 OB-dopplerultrasound

    19/36

    Doppler indices

  • 8/7/2019 OB-dopplerultrasound

    20/36

    UMBILICAL ARTERY FLOW

    characteristic saw-tooth appearance of arterial

    flow in one direction and continuous umbilical

    venous blood flow in the other.

    Umbilical arteryUmbilical artery

  • 8/7/2019 OB-dopplerultrasound

    21/36

    FACTORS AFFECTING UMBILICAL ARTERY DOPPLERFACTORS AFFECTING UMBILICAL ARTERY DOPPLER

    FLOW VELOCITY WAVEFORMS*FLOW VELOCITY WAVEFORMS*

  • 8/7/2019 OB-dopplerultrasound

    22/36

    Benefit of Umbilical Artery EvaluationBenefit of Umbilical Artery Evaluation

    Less experienced operators can achieveLess experienced operators can achieve

    highly reproducible results with simple,highly reproducible results with simple,inexpensive continuousinexpensive continuous--wavewave

    equipment .equipment .

    Umbilical arteryUmbilical artery

  • 8/7/2019 OB-dopplerultrasound

    23/36

    The 40% of the combined fetal ventricular output isThe 40% of the combined fetal ventricular output is

    directed to the placenta by two umbilical arteries.directed to the placenta by two umbilical arteries.The assessment of umbilical blood flow providesThe assessment of umbilical blood flow provides

    information on blood perfusion of the fetoplacentalinformation on blood perfusion of the fetoplacental

    unit .unit .

    Umbilical arteryUmbilical artery

  • 8/7/2019 OB-dopplerultrasound

    24/36

    With advancing gestation,With advancing gestation,

    umbilical arterial Dopplerumbilical arterial Doppler

    waveforms demonstrate awaveforms demonstrate a

    progressive rise in theprogressive rise in the

    endend--diastolic velocity anddiastolic velocity and

    a decrease in thea decrease in the

    pulsatility index.pulsatility index.

    Umbilical arteryUmbilical artery

  • 8/7/2019 OB-dopplerultrasound

    25/36

    Middle cerebral artery

    doppler

  • 8/7/2019 OB-dopplerultrasound

    26/36

    Using color flow imaging, the middle cerebral artery can be

    seen as a major lateral branch of the circle of Willis, running

    anterolaterally at the borderline between the anterior and the

    middle cerebral fossae

    The possible Doppler velocimetry sitesThe possible Doppler velocimetry sites

    Middle cerebral artery

  • 8/7/2019 OB-dopplerultrasound

    27/36

    Middle cerebral arteryMiddle cerebral artery

    The blood velocity increases with advancing gestation,

    and this increase is significantly associated with the

    decrease in PI

  • 8/7/2019 OB-dopplerultrasound

    28/36

    An early stage in fetal adaptationAn early stage in fetal adaptation toto hypoxemiahypoxemia --central redistribution of blood flowcentral redistribution of blood flow

    ( brain( brain--sparing reflex)sparing reflex)

    increased blood flow to protect the brain, heart,increased blood flow to protect the brain, heart,and adrenalsand adrenals

    reduced flow to the peripheral and placentalreduced flow to the peripheral and placentalcirculationscirculations

    Middle cerebral arteryMiddle cerebral artery

  • 8/7/2019 OB-dopplerultrasound

    29/36

    Doppler wave form of early stage ofDoppler wave form of early stage of

    fetal hypoxemiafetal hypoxemia

    increased endincreased end--diastolic flow in the middlediastolic flow in the middle

    cerebral artery (lower MCAcerebral artery (lower MCA pulsatilitypulsatility

    indexindex or resistanceor resistance indexindex))

    Average of both MCAs must be calculatedAverage of both MCAs must be calculated

    for more precise resultfor more precise result

  • 8/7/2019 OB-dopplerultrasound

    30/36

    Middle Cerebral ArteryMiddle Cerebral Artery

    Flow velocity waveform in the fetal middle cerebral arteryFlow velocity waveform in the fetal middle cerebral arteryin a severely anemic fetus atin a severely anemic fetus at 2222 weeks (left) and in aweeks (left) and in anormal fetus (right). In fetal anemia, blood velocity isnormal fetus (right). In fetal anemia, blood velocity isincreasedincreased

  • 8/7/2019 OB-dopplerultrasound

    31/36

    When the fetus is hypoxic, the cerebra arteriesWhen the fetus is hypoxic, the cerebra arteries

    tend to become dilated in order to preservetend to become dilated in order to preserve

    the blood flow to the brain and The systolic tothe blood flow to the brain and The systolic to

    diastolic (A/B) ratio will decrease (due to andiastolic (A/B) ratio will decrease (due to an

    increase in diastolic flow)increase in diastolic flow)

    Middle Cerebral ArteryMiddle Cerebral Artery

  • 8/7/2019 OB-dopplerultrasound

    32/36

    Doppler ultrasound for the fetal assessment inDoppler ultrasound for the fetal assessment in

    highhigh--risk pregnanciesrisk pregnancies

    (Cochrane Review). In:(Cochrane Review). In: The Cochrane Library,The Cochrane Library,

    19991999. Neilson JP and Alfirevic Z. Neilson JP and Alfirevic Z

    Trudinger et alTrudinger et al 19871987

    McParland et alMcParland et al 19881988

    Tyrrell et alTyrrell et al 19901990 Hofmeyr et alHofmeyr et al 19911991

    Newham et alNewham et al 19911991

    Burke et alBurke et al 19921992

    11 Studies Included In Analysis

    Almstrom et al 1992

    Biljan et al 1992

    Johnstone et al 1993Pattison et al 1994

    Nienhuis et al 1997

  • 8/7/2019 OB-dopplerultrasound

    33/36

    Doppler ultrasound for the fetal assessmentDoppler ultrasound for the fetal assessment

    in highin high--risk pregnanciesrisk pregnancies

    NearlyNearly 70007000 patients were includedpatients were included

    The trials compared no Doppler ultrasound toThe trials compared no Doppler ultrasound to

    Doppler ultrasound in highDoppler ultrasound in high--risk pregnancyrisk pregnancy

    (hypertension or presumed impaired fetal(hypertension or presumed impaired fetal

    growth)growth)

    Meta analysis

  • 8/7/2019 OB-dopplerultrasound

    34/36

    Doppler ultrasound for the fetal assessmentDoppler ultrasound for the fetal assessment

    in highin high--risk pregnanciesrisk pregnancies

    A reduction in perinatal deaths.A reduction in perinatal deaths.

    Fewer inductions of labour .Fewer inductions of labour .Fewer admissions to hospital .Fewer admissions to hospital .

    no report of adverse effects .no report of adverse effects .

    No difference was found for fetal distressNo difference was found for fetal distressin labour .in labour .

    No difference in caesarean delivery .No difference in caesarean delivery .

    Main results

  • 8/7/2019 OB-dopplerultrasound

    35/36

    Biophysical profile for fetal assessment inBiophysical profile for fetal assessment in

    high risk pregnancieshigh risk pregnancies

    When compared with conventional fetal monitoring When compared with conventional fetal monitoring

    (usually cardiotocography) biophysical profile(usually cardiotocography) biophysical profiletesting showed no obvious effect (either beneficialtesting showed no obvious effect (either beneficial

    or deleterious) on pregnancy outcome. There wasor deleterious) on pregnancy outcome. There was

    an increase in the number of inductions of labouran increase in the number of inductions of labour

    following biophysical profile in the trial.following biophysical profile in the trial.

    Alfirevic Z, Neilson JP. Biophysical profile for fetal assessment in high risk

    pregnancies (CochraneReview). In: The Cochrane Library, 1995.

  • 8/7/2019 OB-dopplerultrasound

    36/36