olympus vs110-s5 slide scanner – standard operating procedure · olympus vs110-s5 slide scanner...

13
Page | 1 Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. XCite lamp (for fluorescence, 10 min warmup) 2. IX2FCB (Filter wheels for fluorescence) 3. Controller (BXUCB large white box) 4. Computer 5. Login: UserVS110 account (no password) 6. Double click VSASWFL: WAIT FOR INITIALIZATION TO COMPLETE SHUTDOWN SEQUENCE 1. Close VSASWFL software 2. Shut down computer FIRST 3. Turn off Controller (BXUCB) 4. Turn off IX2FCB 5. Turn off XCite lamp 6. Transfer data to external drive and delete from computer ______________________________________________________________________________________________________________________________________________ MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION CONTENTS BRIGHTFIELD (BF) slide scanning – SINGLE Slide .......................................................................................... 2 FLUORESCENCE (FL) slide scanning – SINGLE Slide .......................................................................................... 4 BRIGHTFIELD (BF) slide scanning – BATCH Scans ............................................................................................ 6 FLUORESCENCE (FL) slide scanning – BATCH Scans ......................................................................................... 8 Troubleshooting .......................................................................................................................................... 10 FILTER SPECS Single band (Move XY before switching channels) DAPIFITCTRITCAlx647 Excitation (nm) Emmision (nm) DAPI 375/28 460/50 FITC 480/30 535/40 TRITC 540/25 605/55 Alx647 638/31 694/60

Upload: others

Post on 24-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 1   

Olympus VS110-S5 Slide Scanner – Standard Operating Procedure

STARTUP SEQUENCE 

1. X‐Cite lamp (for fluorescence, 10 min warmup) 

2. IX2‐FCB (Filter wheels for fluorescence) 

3. Controller (BX‐UCB large white box) 

4. Computer    

5. Login: UserVS110 account (no password) 

6. Double click VS‐ASW‐FL:     

WAIT FOR INITIALIZATION TO COMPLETE  

SHUTDOWN SEQUENCE 

1. Close VS‐ASW‐FL software 

2. Shut down computer FIRST 

3. Turn off Controller (BX‐UCB) 

4. Turn off IX2‐FCB 

5. Turn off X‐Cite lamp 

6. Transfer data to external drive and delete 

from computer 

______________________________________________________________________________________________________________________________________________ 

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

CONTENTS

BRIGHTFIELD (BF) slide scanning – SINGLE Slide .......................................................................................... 2 

FLUORESCENCE (FL) slide scanning – SINGLE Slide .......................................................................................... 4 

BRIGHTFIELD (BF) slide scanning – BATCH Scans ............................................................................................ 6 

FLUORESCENCE (FL) slide scanning – BATCH Scans ......................................................................................... 8 

Troubleshooting .......................................................................................................................................... 10 

 

FILTER SPECS 

Single band (Move XY before switching channels) 

DAPI‐FITC‐TRITC‐Alx647 

  Excitation (nm)  Emmision (nm) 

DAPI  375/28  460/50 

FITC  480/30  535/40 

TRITC  540/25  605/55 

Alx647  638/31  694/60    

Page 2: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 2   

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

BRIGHTFIELD (BF) slide scanning – SINGLE Slide 1)  

a. For first time users, click on the “Scan Projects” tab and choose ‘VS110 – Brightfield’ project. This 

will populate the parameters with the most common settings (20x detailed scan, VSI file format). 

You can modify settings during the scan setup. 

b. For all other users, you may apply your own Scan Project as above or click on:  

under Single Slide Scan column. 

2) Click on ‘Load/Unload Slides’ button to put stage in loading position and carefully load slides. Ensure 

slides sit flat and un‐tilted on stage. Click ‘Load/Unload Slides’ button when complete. 

3) Fill out ‘Slide Properties’ left column as needed. 

4) Set naming and saving parameters in ‘Automatic Naming and Saving’ middle column. 

5) Click ‘Overview’ 

6) Once overview image is completed, choose objective magnification for detailed scan. ‘Z‐Scan Mode’ 

should be set to ‘Scan one Z‐plane’. If you require more than one Z‐plane, please see a member of 

staff. 

7) Click ‘Edit Scan Area’   

8) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button   to allow 

the program to automatically detect and draw Scan Regions around your samples or you can choose 

`Remove All Scan Areas`   to draw your own. Use mouse wheel to zoom. 

9) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button) 

option is selected.   

10) Next, click on   to put down a focus map. Choose the Focus Grid Density in the left 

column. You can manually add more Focus points with a mouse click, move existing focus points or 

select and delete those not required. NOTE: The more focus points you put down the longer the scan 

will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus 

points while samples with less contrast and/or uneven surfaces require more points. 

11) When done click on   

12) If ready to proceed, click on ‘Scan Now’. If you require more focusing options, click on ‘Next’ 

 

a. In the ‘Focus Settings’ menu, you may choose the Focus Logic. Options are Automatic 

(Default), Semi‐Automatic (user confirms automatic focusing at each point) and Manual (user 

manually sets each focus point). 

b. You can now click on the ‘Scan Now’ button. 

Page 3: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 3   

 

13) Once completed you have several options: 

 a. You may Add another Detailed Scan to the existing slide 

b. Scan a new slide with the current parameters 

c. Remove/Unload current slide  

d. Save Project (HIGHLY RECOMMENDED*) 

e. Or finish the experiment. 

*To preserve your scan parameters, it is highly recommended that you Save your Project so you do not 

have to worry about accidentally using a previous users settings. 

 

14) If not saving a project, Click ‘Finish’ and save your scans if you have not done so. 

15) To remove the slide, click on   and choose ‘Load/Unload Slides’. 

   

Page 4: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 4   

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

FLUORESCENCE (FL) slide scanning – SINGLE Slide 1)  

a. For first time users, click on the “Scan Projects” tab and choose ‘VS110 – Fluorescence’ to populate 

the parameters with the most common settings (20x detailed scan, VSI file format, typical 

exposure etc.). You can modify settings during the scan setup. 

b. For all other users, you may apply your own Scan Project as above or click on:   

under Single Slide Scan column. 

2) Click on ‘Load/Unload Slides’ button to put stage in loading position and carefully load slides. Ensure slides 

sit flat and un‐tilted on stage. Click ‘Load/Unload Slides’ button when complete. 

3) Fill out ‘Slide Properties’ left column as needed. 

4) Set naming and saving parameters in ‘Automatic Naming and Saving’ middle column. 

5) Click ‘Next’ 

6) Under ‘Overview Type’, choose the method you want to collect your overview. 

a. BF (Brightfield): Quickest method but only if your sample has enough contrast. Click drop down 

arrow ↓ to choose Normal, Dim or Faint Sample to correspond to level of contrast. 

b. Tissue: To scan a fluorescence overview image on tissue samples. Click drop down arrow ↓ to 

choose filter set. 

c. Cells: To scan a fluorescence overview image on tissue samples. Click drop down arrow ↓ to 

choose filter set. 

7) Choose objective lens for overview image (2x is usually sufficient). 

8) If you chose option (a) or (b) above, click ‘Start Live’ button under ‘Exposure Time’ column to set exposure. 

Use joystick to navigate to sample. 

9) Click ‘Overview’ button. 

10) Once overview image is completed, choose objective magnification for detailed scan. ‘Z‐Scan Mode’ should 

be set to ‘Scan one Z‐plane’. If you require more than one Z‐plane, please see a member of staff. Click 

‘Next’ button. 

11) Click ‘Edit Scan Area’   

12) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button   to allow the 

program to automatically detect and draw Scan Regions around your samples or you can choose `Remove 

All Scan Areas`   to draw your own. Use mouse wheel to zoom. 

13) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button) option 

is selected.   

14) Next, click on   to put down a focus map. Choose the Focus Grid Density in the left column. 

You can manually add more Focus points with a mouse click, move existing focus points or select and 

delete those not required. NOTE: The more focus points you put down the longer the scan will take. Flat 

Page 5: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 5   

(fold and bubble free) samples with good contrast do not need a large amount of focus points while 

samples with less contrast and/or uneven surfaces require more points. 

15) When done click on   

16) If you chose one of the preset Scan Projects, the list of filters under ‘Channel Settings’ should be filled with 

the default choices. Otherwise, to add a channel, click   and choose to add a single or multiband filter 

set. To remove a channel, click  . 

17) Select Channel 1 (this is the channel that will be used for focusing so it should be your brightest and most 

widespread) and click   

18) In the Stage Navigator window in the far right, make sure the yellow square is in the middle of your 

Detailed Scan area. If not, click inside your region of interest to move the stage. 

19) Refer to the histogram in the top right window. Change the Exposure Time so that you achieve at least 

1000 counts in the mean intensity reading.    

 

The XM10 fluorescence camera is a 14‐bit camera, allowing for a high dynamic range (16384 bit depth). It 

is recommended to use as much of this range as possible (without excessive exposure times causing 

photo‐bleaching) to capture as much data as possible for potential post‐scan analyses. 

20) If ready to proceed, click on ‘Scan Now’. If you require more focusing options, click on ‘Next’ 

21) In the ‘Focus Settings’ menu, you may choose the Focus Logic. Options are Automatic (Default), Semi‐

Automatic (user confirms automatic focusing at each point) and Manual (user manually sets each focus 

point). 

22) You can now click on the ‘Scan Now’. 

23) Once completed you have several options:   

a. You may Add another Detailed Scan to the existing slide 

b. Scan a new slide with the current parameters 

c. Remove/Unload current slide  

d. Save Project (HIGHLY RECOMMENDED*) 

e. Or finish the experiment. 

*To preserve your scan parameters, it is highly recommended that you Save your Project so you do not 

have to worry about accidentally using a previous users settings. 

24) If not saving a project, Click ‘Finish’ and save your scans if you have not done so. 

25) To remove the slide, click on   and choose ‘Load/Unload Slides’.   

Page 6: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 6   

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

BRIGHTFIELD (BF) slide scanning – BATCH Scans 1) Carefully load slides onto the stage. The stage positions are numbered #1 to #5, left to right. If you 

need more room load slides, use the joystick to move the stage. 

2) To scan multiple brightfield slides, click on:   under Batch Scan column. 

3) Highlight each slide and fill out its properties in the appropriate rows and columns. Then click ‘Next’ 

when done. A template can be used to pre‐fill these properties. Talk to staff if needed. 

4) Under the ‘Slide Loader’ column, you are now able to load one or more previously saved Scan 

Project(s) or use the provided ‘VS110 – Brightfield’ project for typical parameters. 

5) In the ‘Scan Settings’ column and ‘Magnification and Image Document’ tab, select desired 

magnification and Z‐scan mode. 

6) a) Click the ‘Apply settings to all slides’ button    to save these settings to every slide in every 

cassette. All slides in the loader should now be highlighted green if 20x lens is used (10x is labelled 

yellow while 40x is labeled blue). 

b) If you only want to scan a selection of your slides, hold the Ctrl‐key and click the positions in the 

cassette to select only the slides you wish to scan. Once the selection is complete, repeat step 5) and 

click the ‘Apply settings to selection’ button   to save settings to just the selected slides. The 

selected slides in the loader should now be highlighted green (or otherwise). 

c) If you want to scan different slide selections with different parameters/projects, repeat steps 4 – 6 

for each slide selection. 

7) Set naming and saving parameters in ‘Automatic Naming and Saving’ column and click ‘Overview’. 

8) If for whatever reason you need to scan a single slide NOT loaded in the cassettes, click on ‘Priority 

Scan’ to interrupt the batch process. 

9) Once overviews are completed, you can review each slide in the ‘Batch Slide Selection’ column. With a 

slide selected, click on ‘Edit Scan Area’   button. 

10) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button   to allow 

the program to automatically detect and draw Scan Regions around your samples or you can choose 

`Remove All Scan Areas`   to draw your own. Use mouse wheel to zoom. 

11) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button) 

option is selected.   

12) Next, click on   to put down a focus map. Choose the Focus Grid Density in the left 

column. You can manually add more Focus points with a mouse click, move existing focus points or 

Page 7: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 7   

select and delete those not required. NOTE: The more focus points you put down the longer the scan 

will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus 

points while samples with less contrast and/or uneven surfaces require more points. 

13) When done click on   

14) Select next slide in the ‘Batch Slide Selection’ column and repeat steps 10 – 14. 

15) If you do not wish to scan a particular slide(s) in the batch, select the slides in the list and click on 

‘Exclude slides from the batch scan’ button   to exclude from batch scan. 

16) If ready to proceed, click on ‘Scan Now’.    

Page 8: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 8   

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

FLUORESCENCE (FL) slide scanning – BATCH Scans 1) Carefully load slides onto the stage. The stage positions are numbered #1 to #5, left to right. If you 

need more room load slides, use the joystick to move the stage. 

2) To scan multiple fluorescence slides, click on:   under Batch Scan column. 

3) Highlight each slide and fill out its properties in the appropriate rows and columns. Then click ‘Next’ 

when done. A template can be used to pre‐fill these properties. Talk to staff if needed. 

4) Under the ‘Slide Loader’ column, you are now able to load one or more previously saved Scan 

Project(s) or use the provided ‘VS110 – Brightfield’ project for typical parameters. 

5) In the ‘Scan Settings’ column and ‘Magnification and Image Document’ tab, select desired 

magnification and Z‐scan mode. You may also modify the ‘Overview Settings’, ‘Illumination Settings’ 

and ‘Focus Settings’ tabs. However, you will have a chance to modify these settings later. 

6) a) Click the ‘Apply settings to all slides of all cassettes’ button    to save these settings to every 

slide in every cassette. All slides in the loader should now be highlighted green if 20x lens is used (10x 

is labelled yellow while 40x is labeled blue).. 

b) If you only want to scan a selection of your slides, hold the Ctrl‐key and click the positions in the 

cassette to select only the slides you wish to scan. Once the selection is complete, repeat step 5) and 

click the ‘Apply settings to selection’ button   to save settings to just the selected slides. The 

selected slides in the loader should now be highlighted green (or otherwise). 

c) If you want to scan different slide selections with different parameters/projects, repeat steps 4 – 6 

for each slide selection. 

7) Set naming and saving parameters in ‘Automatic Naming and Saving’ column and click ‘Overview’. 

8) Once overviews are completed, you can review each slide in the ‘Batch Slide Selection’ column. With a 

slide selected, click on ‘Edit Scan Area’   button. 

9) Under ‘Scan Area’ in the left column. You can either click on ‘Create Scan Area’ button   to allow 

the program to automatically detect and draw Scan Regions around your samples or you can choose 

`Remove All Scan Areas`   to draw your own. Use mouse wheel to zoom. 

10) Under ‘Detailed Scan Settings’, make sure ‘Scan whole area and focus sample only’ (middle button) 

option is selected.   

11) Next, click on   to put down a focus map. Choose the Focus Grid Density in the left 

column. You can manually add more Focus points with a mouse click, move existing focus points or 

select and delete those not required. NOTE: The more focus points you put down the longer the scan 

Page 9: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 9   

will take. Flat (fold and bubble free) samples with good contrast do not need a large amount of focus 

points while samples with less contrast and/or uneven surfaces require more points. 

12) When done click on   

13) Select next slide in the ‘Batch Slide Selection’ column and repeat steps 10 – 14. 

14) In the ‘Scan Settings’ column, set the parameters required for the ‘Magnification and Image 

Document’ tab, ‘Illumination Settings’ tab and ‘Focus Settings’ tab if required. 

15) If you do not wish to scan a particular slide(s) in the batch, select the slides in the list and click on 

‘Exclude slides from the batch scan’ button   to exclude from batch scan. 

16) If ready to proceed, click on ‘Scan Now’.    

Page 10: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 10   

Troubleshooting Sample not recognized 

For weakly stained samples with poor contrast (for example, brightfield slides with weakly staining dyes, very 

sparse stained areas OR brightfield overviews of fluorescently stained samples). 

During the editing scans   dialog you will see the following settings in the left column: 

 

Figure 1.  

To increase the threshold of sample detection, modify the Sample detection sensitivity slider. In the example 

below (Figure 2) with the slider at ‘0’, you see only a small percentage of the spinal cord tissue sections 

detected by the system as indicated by the white, highlighted background. With the slider set to +10 (Figure 3), 

almost all the sections are highlighted. Sample detection is also important when creating focus maps on scan 

areas. Under ‘Detailed Scan Settings’, if the ‘Scan and focus sample only’ or ‘Scan whole area and focus sample 

only’ (first and second button from left)   is selected. Focus maps will only be placed on 

samples that have been detected (as described above). If your sample is still too faint to be detected, you will 

have to choose the 3rd button ‘Scan and focus whole area’ after manually drawing the scan area and/or 

manually add focus points. 

There are various options in the ‘Sample Detection Settings’ and ‘Advanced Sample Detection Settings’ menu 

to assist in fine tuning detection of your samples. 

When your samples are successfully highlighted, you can now click the ‘Create Scan Area’ button    

which will automatically put a scan area around the detected sections (Figure 4). This is especially useful if you 

have many sections on the one slide to avoid manually drawing scan areas. 

Page 11: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 11   

 

 

Figure 2. Brightfield overview image of fluorescently stained mouse spinal cord cross sections, sample 

detection sensitivity set to Zero. 

 

Figure 3. As above with sample detection sensitivity set to +10. 

 

Figure 4. As above after clicking on ‘Create Sample Area’ button. 

Page 12: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 12   

If you choose Brightfield overview scan for your fluorescent sample, there are various options in the drop 

down arrow menu. Choosing BF (brightfield) you have a choice of Normal, Dim and Faint samples. 

 

Depending on the choice, the program will manipulate the brightness and gamma to improve the contrast of 

the samples on your slide, giving you a greater chance of sample detection (mentioned above). 

  BF – Normal Samples 

  BF – Dim Samples 

  BF – Faint Samples 

   

Page 13: Olympus VS110-S5 Slide Scanner – Standard Operating Procedure · Olympus VS110-S5 Slide Scanner – Standard Operating Procedure STARTUP SEQUENCE 1. X‐Cite lamp (for fluorescence,

   

Page | 13   

Focusing Issues 

After setting focus maps on your samples and clicking ‘Scan Now’, the system will go to each focus point to set 

the focus. If the focus is found, the focus point square will change colour to green. If the focus could not be 

found, the focus point square will change to grey. The scan will proceed if there is at least ONE green focus 

point square. 

Various focusing issues and possible solutions: 

1. Samples are weakly stained (either brightfield or fluorescent). 

o In the case of weakly stained brightfield samples, go into ‘Live’ mode before proceeding with 

the scan to manually set exposure times. 

o In the case of weak fluorescent signals, increase the exposure time of the focusing channel 

(channel on the top of the list) so the mean intensity on the histogram is at least 1500 counts 

(Page 5). Always choose the strongest and most widespread fluorescent channel to be the 

focusing channel. 

 

2. Samples are too thick or too thin. 

o Thin samples <5 microns will have a smaller volume of fluorescent signal. Ensure that the stage 

is close to the focus by going into ‘Live’ mode before proceeding with the scan. 

o Thick samples >50 microns will have many points of focus. The algorithm will usually choose 

the top of the sample. 

 

3. Stage is too far away from sample or focusing algorithm gets ‘lost’. 

o Pertaining to both points 1 and 2, double check in ‘Live’ mode to make sure the stage is at 

least close to the focus of your sample. 

o Avoid putting focus maps on empty areas where there is no sample. Focus maps on a glass 

slide can often bring the stage far from the focus.