E5070B/E5071B Adjustment Program<Agilent Service Center Use Only>

Operation ManualFirst Edition

Agilent Part No. N/AOctober 2002

Printed in Japan

Page 2: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

NoticesThe information contained in this document is subject to change without notice.

This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies.

Agilent Technologies Japan, Ltd.

Componet Test PGU Kobe

1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan

Copyright © Agilent Technologies Japan, Ltd. 2002

Manual Printing HistoryThe manual’s printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The manual part number changes when extensive technical changes are incorporated.

October 2002 First Edition

2

Page 3: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

Contents

1. InstallationWarm-up for Performance Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Preparation for using the Adjustment Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Required Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Installing a GPIB Card (82341C/D or 82350A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Installing Agilent VEE for Personal Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Installing Adjustment Program into Your PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Equipment Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Setting Talker/Listener GPIB address of the E5070B/E5071B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Registering Power Sensor Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

2. OperationHow to perform the E5070B/E5071B Adjustment Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Main Window Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Model Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Adjustment Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Result Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

3

Page 4: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

Contents

4

Page 5: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

1 Installation

This chapter provides information for the installation of E5070B/E5071B adjustment program.

5Agilent Service Center Use Only

Page 6: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationWarm-up for Performance Test

Warm-up for Performance TestAllow the E5070B/E5071B to warm up for at least 30 minutes before you execute any of the adjustments.

Required EquipmentThe required test equipment for the adjustments is listed in Table 1-1. Use only calibrated equipment when adjusting the E5070B/E5071B.

6 Chapter 1 Agilent Service Center Use Only

Page 7: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationRequired Equipment

Table 1-1 Recommended Equipment

Equipment Critical specifications Recommended Model Qty.

Frequency Counter Frequency: 1 MHz to 3 GHz

Accuracy: < 2.5 ppm53131/32A Opt.010 & 030*1

or 53181A Opt.010 & 030*2

1

Frequency Standard Frequency: 10 MHz, Time Base

Error: � � 1� 10–10/year

5071A 1

Power Meter No Substitute E4419A/B 1

Power Sensor No Substitute 8482A

E4412A

1

1

Calibration Kit No Substitute 85032F 1

Fixed Attenuator (6dB)

50 �, N(m)-N(f), VSWR � 1.015 8491A Opt.006,H60 1

Cable BNC(m)-BNC(m) Cable, 61 cm

Coaxial Cable with Type-N(m) connectors, 61 cm (24 in)

p/n 8120-1839

N6314A ( p/n 8120-8862)

1

1

Adapter N(m)-BNC(f) Adapter p/n 1250-0780 1

Torque Wrench Size: 3/4 inchTorque: 136 N-cm

p/n 8710-1766 1

*1.Opt.050 and Opt.124 can be substituted for Opt.30. In this case, a N(m)-BNC(f) adapter is necessary.*2.Opt.050 and Opt.124 can be substituted for Opt.030. In this case, a N(m)-BNC(f) adapter is necessary.

Chapter 1 7Agilent Service Center Use Only

Page 8: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationPreparation for using the Adjustment Program

Preparation for using the Adjustment ProgramTo use the Adjustment Program, some preparation is required. This section describes how to proceed.

Required Controller

The following controller system is required to run the performance test program.

Windows PC PC-AT Compatible, RAM:�64MBytes, CPU Pentium 200 MHz or faster, HDD: 10 Mbytes or move free space

OS Microsoft� Windows NT� (4.0), Windows 95�, Windows 98�, Windows 2000�

Software Agilent VEE (Rev 5.0x or 6.0x)

GPIB Card 82350A, 82340B, 82341C/D

Installing a GPIB Card (82341C/D or 82350A)

Install a GPIB Card into your computer (see the GPIB Card manuals). The select code of the GPIB Card should be set to “7”.

Installing Agilent VEE for Personal Computer

Install the Agilent VEE (Rev 5.0x or 6.0x) into your computer (see the Agilent VEE manuals).

Installing Adjustment Program into Your PC

Step 1. Make a copy of the E5070B/E5071B adjustment program named ADJE507071B.EXE in a directory of the hard disk drive in your PC.

Step 2. Double-click on the filename on the Windows’ Explorer to start extracting the self-extracting achieve.

Step 3. You will be prompted to enter folder name for installing the program files. Enter the folder name, then click on UNZIP. The default folder is C:\ADJE507071B

Figure 1-1 Prompt of Self-extracting Achieve

Step 4. Confirm the message that you have successfully extracted the files and click OK and

8 Chapter 1 Agilent Service Center Use Only

Page 9: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationPreparation for using the Adjustment Program

Close.

Figure 1-2 Message Window

Chapter 1 9Agilent Service Center Use Only

Page 10: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationPreparation for using the Adjustment Program

Equipment Setup

Performing adjustments requires the system described in this section. The Hardware Setup is shown in Figure 1-3.

Figure 1-3 Adjustment Hardware Setup

Setting Talker/Listener GPIB address of the E5070B/E5071B

To control the E5070B/E5071B in the adjustment program, the talker/listener GPIB address must be set. Check and set the GPIB address according to the following procedure.

Step 1. Press - Misc Setup - GPIB Setup - Talker/Listener Address. If the address “17” has not been set, proceed to the next step.

Step 2. Enter the GPIB address “17” using the ENTRY block keys on the front panel.

Registering Power Sensor Calibration Data

The calibration data of the 8482A power sensor must be registered before running the adjustment program. Execute the Agilent EditCal Calibration Factor Editor to register it.

Step 1. Start the EditCal to click on Start button, Programs and Agilent EditCal on Windows.

Step 2. The top window appears as shown in Figure 1-4. Choose Calibration Data, then press Start.

10 Chapter 1 Agilent Service Center Use Only

Page 11: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationPreparation for using the Adjustment Program

Figure 1-4 EditCal Top Window

Step 3. The calibration data menu appears as shown in Figure 1-5. Choose 8482A and the serial number to edit, then click on Edit. If the serial number is not registered, Click on Add and choose 8482A.

Figure 1-5 Calibration Data Menu

Step 4. The 8482A calibration data window appears as shown in Figure 1-6. Input the following items and click on OK.

• Serial number (for addition)• Calibration Date• Calibration Due Date• Calibration Data

NOTE Delete the calibration points which are not described on the calibration report because the adjustment program interpolates the calibration factor for the frequency which calibration factor is not registered.

Chapter 1 11Agilent Service Center Use Only

Page 12: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

InstallationPreparation for using the Adjustment Program

Figure 1-6 8482A Calibration Data Window

12 Chapter 1 Agilent Service Center Use Only

Page 13: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

2 Operation

This chapter provides information for using the E5070B/E5071B adjustment program.

13Agilent Service Center Use Only

Page 14: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

OperationHow to perform the E5070B/E5071B Adjustment Program

How to perform the E5070B/E5071B Adjustment Program

Step 1. Start the Agilent VEE (Rev 5.0x or 6.0x).

Step 2. Load the adjustment program file into the Agilent VEE as follows.

1. Pull down the File menu from the Agilent VEE window and select Open.

2. Navigate to the folder that the program installed, then choose ADJE507071B.VEE and click Open.

Step 3. You may be asked to add drivers for the equipment during the program loading. Click on OK and the GPIB address for each equipment. Enter 0 as the address for each equipment which is not used for the adjustment. (See Figure 2-1).

Figure 2-1 Direct I/O configuration

Step 4. The main window appears as shown in Figure 2-2. Click on START button to proceed.

Figure 2-2 Main Window

Step 5. The ETE Configuration window menu appears as shown in Figure 2-3. Input all columns

14 Chapter 2 Agilent Service Center Use Only

Page 15: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

OperationHow to perform the E5070B/E5071B Adjustment Program

and click on Next.

NOTE The Home Directory means the directory where ADJE507071B.VEE is stored.

Figure 2-3 ETE Configuration Window

Step 6. The Emvironment window appears as shown in Figure 2-4. Input the data and click on NEXT STEP.

Chapter 2 15Agilent Service Center Use Only

Page 16: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

OperationHow to perform the E5070B/E5071B Adjustment Program

Figure 2-4 Environment Information Window

Step 7. The main menu appears as shown in Figure 2-5. The description of the button on the window is described in Table 2-1.

Figure 2-5 Main Menu

Table 2-1 Description of Menu Buttons

Item Description

All Adjustment Performs all adjustment items. After pushing this button, the program checks connection of ETE and start adjustment.

Spot Adjustment Performs a part of adjustment items according to the replaced assembly. After pushing this button, the program proceed to Step 8.

16 Chapter 2 Agilent Service Center Use Only

Page 17: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

OperationHow to perform the E5070B/E5071B Adjustment Program

Step 8. The spot adjustment menu appears as shown in Figure 2-6 when you pushing Spot Adjustment. After clicking on a button, The adjustment items is performed according to the replaced assembly. Refer to Table 2-2 for detail.

NOTE Execute "All Adjustment” if two or more spot adjustment items are selected.

Figure 2-6 Spot Adjustment Window

Input Environment Data Show the environment Information window. After this button, the program proceeds to Step 6.

Program End Quits the adjustment program.

Table 2-1 Description of Menu Buttons

Item Description

Chapter 2 17Agilent Service Center Use Only

Page 18: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

OperationHow to perform the E5070B/E5071B Adjustment Program

Table 2-2 Required Adjustment Item

Select Button Adjustment Item

Initi

aliz

e C

orre

ctio

n C

onst

ants

Wri

ting

ID

Freq

uenc

y R

efer

ence

Adj

ustm

ent

Ove

n R

efer

ence

Adj

ustm

ent (

Opt

. 1E

5)

IF R

ange

Gai

n/Ph

ase

Adj

ustm

ent

Sour

ce P

ower

Adj

ustm

ent

Test

Por

ts C

hara

cter

istic

s Adj

ustm

ent

Stor

ing

Cor

rect

ion

Con

stan

ts

Che

ckin

g C

orre

ctio

n C

onst

ants

All Adjustment � � � � � � � � �

Spot Adjustment: Source Board � � � � � �

Spot Adjustment: Receiver (RF) Board � � � � �

Spot Adjustment: Receiver (IF) Board � � � � � �

Spot Adjustment: RF Switch � � � � �

Spot Adjustment: Crystal Oven (for 1E5) �

Spot Adjustment: Att (for 214/314/414) � � � � �

Spot Adjustment: Hard Disk Drive �

Spot Adjustment: Analog Interface � � �

Spot Adjustment: Level Vernier � � � � �

Spot Adjustment: Power Amp Module � � � � �

18 Chapter 2 Agilent Service Center Use Only

Page 19: Operation Manual - Keysight Network Analyzer Service ...na.support.keysight.com/misc/enab_test/svc_adj_manual.pdf · Operation Manual First Edition Agilent Part No. N/A ... Figure

OperationMain Window Information

Main Window InformationFigure 2-7 shows the displayed information.

Figure 2-7 Displayed Information in Main Window

Model Number

The model number of the equipment to be adjusted.

Serial Number

The serial number of the equipment to be adjusted.

Option

The installed option in the equipment to be adjusted.

Adjustment Name

The adjustment executed now.

Result Summary

This area shows the summary of the results. The information displayed in each column is described in Table 2-3.

Table 2-3 Column in Test Result Summary

Column Description

ADJUSTMENT NAME Name of Adjustment

JUDGE Result of Adjustment. O means “pass.” X means “fail.” - means “not performed.”

DATE TIME Time when Adjustment is finished.

Chapter 2 19Agilent Service Center Use Only