E5070B/E5071B Adjustment Program<Agilent Service Center Use Only>
Operation ManualFirst Edition
Agilent Part No. N/AOctober 2002
Printed in Japan
NoticesThe information contained in this document is subject to change without notice.
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies.
Agilent Technologies Japan, Ltd.
Componet Test PGU Kobe
1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan
Copyright © Agilent Technologies Japan, Ltd. 2002
Manual Printing HistoryThe manual’s printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates that are incorporated at reprint do not cause the date to change.) The manual part number changes when extensive technical changes are incorporated.
October 2002 First Edition
2
Contents
1. InstallationWarm-up for Performance Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Preparation for using the Adjustment Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Required Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Installing a GPIB Card (82341C/D or 82350A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Installing Agilent VEE for Personal Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Installing Adjustment Program into Your PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Equipment Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Setting Talker/Listener GPIB address of the E5070B/E5071B . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Registering Power Sensor Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2. OperationHow to perform the E5070B/E5071B Adjustment Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Main Window Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Model Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Adjustment Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Result Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3
Contents
4
1 Installation
This chapter provides information for the installation of E5070B/E5071B adjustment program.
5Agilent Service Center Use Only
InstallationWarm-up for Performance Test
Warm-up for Performance TestAllow the E5070B/E5071B to warm up for at least 30 minutes before you execute any of the adjustments.
Required EquipmentThe required test equipment for the adjustments is listed in Table 1-1. Use only calibrated equipment when adjusting the E5070B/E5071B.
6 Chapter 1 Agilent Service Center Use Only
InstallationRequired Equipment
Table 1-1 Recommended Equipment
Equipment Critical specifications Recommended Model Qty.
Frequency Counter Frequency: 1 MHz to 3 GHz
Accuracy: < 2.5 ppm53131/32A Opt.010 & 030*1
or 53181A Opt.010 & 030*2
1
Frequency Standard Frequency: 10 MHz, Time Base
Error: � � 1� 10–10/year
5071A 1
Power Meter No Substitute E4419A/B 1
Power Sensor No Substitute 8482A
E4412A
1
1
Calibration Kit No Substitute 85032F 1
Fixed Attenuator (6dB)
50 �, N(m)-N(f), VSWR � 1.015 8491A Opt.006,H60 1
Cable BNC(m)-BNC(m) Cable, 61 cm
Coaxial Cable with Type-N(m) connectors, 61 cm (24 in)
p/n 8120-1839
N6314A ( p/n 8120-8862)
1
1
Adapter N(m)-BNC(f) Adapter p/n 1250-0780 1
Torque Wrench Size: 3/4 inchTorque: 136 N-cm
p/n 8710-1766 1
*1.Opt.050 and Opt.124 can be substituted for Opt.30. In this case, a N(m)-BNC(f) adapter is necessary.*2.Opt.050 and Opt.124 can be substituted for Opt.030. In this case, a N(m)-BNC(f) adapter is necessary.
Chapter 1 7Agilent Service Center Use Only
InstallationPreparation for using the Adjustment Program
Preparation for using the Adjustment ProgramTo use the Adjustment Program, some preparation is required. This section describes how to proceed.
Required Controller
The following controller system is required to run the performance test program.
Windows PC PC-AT Compatible, RAM:�64MBytes, CPU Pentium 200 MHz or faster, HDD: 10 Mbytes or move free space
OS Microsoft� Windows NT� (4.0), Windows 95�, Windows 98�, Windows 2000�
Software Agilent VEE (Rev 5.0x or 6.0x)
GPIB Card 82350A, 82340B, 82341C/D
Installing a GPIB Card (82341C/D or 82350A)
Install a GPIB Card into your computer (see the GPIB Card manuals). The select code of the GPIB Card should be set to “7”.
Installing Agilent VEE for Personal Computer
Install the Agilent VEE (Rev 5.0x or 6.0x) into your computer (see the Agilent VEE manuals).
Installing Adjustment Program into Your PC
Step 1. Make a copy of the E5070B/E5071B adjustment program named ADJE507071B.EXE in a directory of the hard disk drive in your PC.
Step 2. Double-click on the filename on the Windows’ Explorer to start extracting the self-extracting achieve.
Step 3. You will be prompted to enter folder name for installing the program files. Enter the folder name, then click on UNZIP. The default folder is C:\ADJE507071B
Figure 1-1 Prompt of Self-extracting Achieve
Step 4. Confirm the message that you have successfully extracted the files and click OK and
8 Chapter 1 Agilent Service Center Use Only
InstallationPreparation for using the Adjustment Program
Close.
Figure 1-2 Message Window
Chapter 1 9Agilent Service Center Use Only
InstallationPreparation for using the Adjustment Program
Equipment Setup
Performing adjustments requires the system described in this section. The Hardware Setup is shown in Figure 1-3.
Figure 1-3 Adjustment Hardware Setup
Setting Talker/Listener GPIB address of the E5070B/E5071B
To control the E5070B/E5071B in the adjustment program, the talker/listener GPIB address must be set. Check and set the GPIB address according to the following procedure.
Step 1. Press - Misc Setup - GPIB Setup - Talker/Listener Address. If the address “17” has not been set, proceed to the next step.
Step 2. Enter the GPIB address “17” using the ENTRY block keys on the front panel.
Registering Power Sensor Calibration Data
The calibration data of the 8482A power sensor must be registered before running the adjustment program. Execute the Agilent EditCal Calibration Factor Editor to register it.
Step 1. Start the EditCal to click on Start button, Programs and Agilent EditCal on Windows.
Step 2. The top window appears as shown in Figure 1-4. Choose Calibration Data, then press Start.
10 Chapter 1 Agilent Service Center Use Only
InstallationPreparation for using the Adjustment Program
Figure 1-4 EditCal Top Window
Step 3. The calibration data menu appears as shown in Figure 1-5. Choose 8482A and the serial number to edit, then click on Edit. If the serial number is not registered, Click on Add and choose 8482A.
Figure 1-5 Calibration Data Menu
Step 4. The 8482A calibration data window appears as shown in Figure 1-6. Input the following items and click on OK.
• Serial number (for addition)• Calibration Date• Calibration Due Date• Calibration Data
NOTE Delete the calibration points which are not described on the calibration report because the adjustment program interpolates the calibration factor for the frequency which calibration factor is not registered.
Chapter 1 11Agilent Service Center Use Only
InstallationPreparation for using the Adjustment Program
Figure 1-6 8482A Calibration Data Window
12 Chapter 1 Agilent Service Center Use Only
2 Operation
This chapter provides information for using the E5070B/E5071B adjustment program.
13Agilent Service Center Use Only
OperationHow to perform the E5070B/E5071B Adjustment Program
How to perform the E5070B/E5071B Adjustment Program
Step 1. Start the Agilent VEE (Rev 5.0x or 6.0x).
Step 2. Load the adjustment program file into the Agilent VEE as follows.
1. Pull down the File menu from the Agilent VEE window and select Open.
2. Navigate to the folder that the program installed, then choose ADJE507071B.VEE and click Open.
Step 3. You may be asked to add drivers for the equipment during the program loading. Click on OK and the GPIB address for each equipment. Enter 0 as the address for each equipment which is not used for the adjustment. (See Figure 2-1).
Figure 2-1 Direct I/O configuration
Step 4. The main window appears as shown in Figure 2-2. Click on START button to proceed.
Figure 2-2 Main Window
Step 5. The ETE Configuration window menu appears as shown in Figure 2-3. Input all columns
14 Chapter 2 Agilent Service Center Use Only
OperationHow to perform the E5070B/E5071B Adjustment Program
and click on Next.
NOTE The Home Directory means the directory where ADJE507071B.VEE is stored.
Figure 2-3 ETE Configuration Window
Step 6. The Emvironment window appears as shown in Figure 2-4. Input the data and click on NEXT STEP.
Chapter 2 15Agilent Service Center Use Only
OperationHow to perform the E5070B/E5071B Adjustment Program
Figure 2-4 Environment Information Window
Step 7. The main menu appears as shown in Figure 2-5. The description of the button on the window is described in Table 2-1.
Figure 2-5 Main Menu
Table 2-1 Description of Menu Buttons
Item Description
All Adjustment Performs all adjustment items. After pushing this button, the program checks connection of ETE and start adjustment.
Spot Adjustment Performs a part of adjustment items according to the replaced assembly. After pushing this button, the program proceed to Step 8.
16 Chapter 2 Agilent Service Center Use Only
OperationHow to perform the E5070B/E5071B Adjustment Program
Step 8. The spot adjustment menu appears as shown in Figure 2-6 when you pushing Spot Adjustment. After clicking on a button, The adjustment items is performed according to the replaced assembly. Refer to Table 2-2 for detail.
NOTE Execute "All Adjustment” if two or more spot adjustment items are selected.
Figure 2-6 Spot Adjustment Window
Input Environment Data Show the environment Information window. After this button, the program proceeds to Step 6.
Program End Quits the adjustment program.
Table 2-1 Description of Menu Buttons
Item Description
Chapter 2 17Agilent Service Center Use Only
OperationHow to perform the E5070B/E5071B Adjustment Program
Table 2-2 Required Adjustment Item
Select Button Adjustment Item
Initi
aliz
e C
orre
ctio
n C
onst
ants
Wri
ting
ID
Freq
uenc
y R
efer
ence
Adj
ustm
ent
Ove
n R
efer
ence
Adj
ustm
ent (
Opt
. 1E
5)
IF R
ange
Gai
n/Ph
ase
Adj
ustm
ent
Sour
ce P
ower
Adj
ustm
ent
Test
Por
ts C
hara
cter
istic
s Adj
ustm
ent
Stor
ing
Cor
rect
ion
Con
stan
ts
Che
ckin
g C
orre
ctio
n C
onst
ants
All Adjustment � � � � � � � � �
Spot Adjustment: Source Board � � � � � �
Spot Adjustment: Receiver (RF) Board � � � � �
Spot Adjustment: Receiver (IF) Board � � � � � �
Spot Adjustment: RF Switch � � � � �
Spot Adjustment: Crystal Oven (for 1E5) �
Spot Adjustment: Att (for 214/314/414) � � � � �
Spot Adjustment: Hard Disk Drive �
Spot Adjustment: Analog Interface � � �
Spot Adjustment: Level Vernier � � � � �
Spot Adjustment: Power Amp Module � � � � �
18 Chapter 2 Agilent Service Center Use Only
OperationMain Window Information
Main Window InformationFigure 2-7 shows the displayed information.
Figure 2-7 Displayed Information in Main Window
Model Number
The model number of the equipment to be adjusted.
Serial Number
The serial number of the equipment to be adjusted.
Option
The installed option in the equipment to be adjusted.
Adjustment Name
The adjustment executed now.
Result Summary
This area shows the summary of the results. The information displayed in each column is described in Table 2-3.
Table 2-3 Column in Test Result Summary
Column Description
ADJUSTMENT NAME Name of Adjustment
JUDGE Result of Adjustment. O means “pass.” X means “fail.” - means “not performed.”
DATE TIME Time when Adjustment is finished.
Chapter 2 19Agilent Service Center Use Only