operative technique variax locking plates locking plates select an appropriate sized implant. note:...

2
11 VariAx Hand Locking Plate Module Operative Technique VariAx Locking Plates Select an appropriate sized implant. Note: The 1.7mm VariAx Locking Plates are not indicated for use on the metacarpals. Step 1 Cut and/or Contour the plate if necessary. Step 2 Place the appropriate Locking Drill Guide in the hole of the plate and aim the appropriate drill in the desired position. This drill guide will not allow for drilling past 10 degrees. Note: This drill guide must be used with all locking plates for application of bone screws and locking screws. Step 3 20° ‘locking cone’

Upload: doananh

Post on 12-Jun-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Operative Technique VariAx Locking Plates Locking Plates Select an appropriate sized implant. Note: The 1.7mm VariAx Locking Plates are not indicated for use on the metacarpals. Step

11

VariA

xH

and

Lockin

g P

late

Module

Operative Technique

VariAx Locking Plates

Select an appropriate sized implant.

Note: The 1.7mm VariAx LockingPlates are not indicated for use on the metacarpals.

Step 1

Cut and/or Contour the plate if necessary.

Step 2

Place the appropriate Locking DrillGuide in the hole of the plate and aimthe appropriate drill in the desiredposition. This drill guide will notallow for drilling past 10 degrees.

Note: This drill guide must be usedwith all locking plates forapplication of bone screws and locking screws.

Step 3

20° ‘locking cone’

Page 2: Operative Technique VariAx Locking Plates Locking Plates Select an appropriate sized implant. Note: The 1.7mm VariAx Locking Plates are not indicated for use on the metacarpals. Step

12

Operative Technique

VariAx Locking Plates, continued

Measure for the length of the screwand then insert the proper lockingscrew.

Bone screws can also be used in anyholes if desired

Step 4