optical spectroscopyweb.mnstate.edu/marasing/chem380/documents/uv-vis 2016.pdf · uv-vis...

20
Optical Spectroscopy Ultraviolet-Visible Techniques involved in the study of the Interaction of matter (molecules) with electromagnetic radiation ~ spectroscopy. Current classification of optical spectroscopy refers to all spectroscopic techniques that use non-ionizing radiation. Covers the region from the far ultraviolet (UV) to the far infrared (IR), the terahertz region. Basic facts of electromagnetic radiation (wave) – All electromagnetic radiation can be considered as waves Characteristics of electromagnetic waves: Wavelength (λ), frequency (ν), wavenumber (ν) (‘color’ of light) – wavenumber = 1/ λ (number of waves / unit length) Blue light ~ 400 - 450 nm wavelength, 25 000 – 22 222 cm -1 Red light ~ 600 - 700 nm wavelength, 16 667 – 14 286 cm -1 – Polarization: direction of transversal vibration of the wave – linearly polarized light vs. randomly polarized – Speed: in vacuum (n=1); c = 2.99 × 10 8 m/s Considered as a particle : photon (energy packet) of energy hν. A wave: periodic change of a property in space and/or time. Instantaneous snap shot of a linearly polarized wave with graphic of field intensities. Basic facts of electromagnetic radiation (wave) EM radiation is a travelling wave; has mutually perpendicular magnetic and electrical fields and these fields fluctuate periodically in time and space . hc photon energy h = = ν λ xy plane, plane of polarization

Upload: others

Post on 19-Jul-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Optical Spectroscopy

Ultraviolet-Visible

Techniques involved in the study of the Interaction of matter (molecules) with electromagnetic radiation ~ spectroscopy.

Current classification of optical spectroscopy refers to all spectroscopic techniques that use non-ionizing radiation.

Covers the region from the far ultraviolet (UV) to the far infrared (IR), the terahertz region.

Basic facts of electromagnetic radiation (wave)

– All electromagnetic radiation can be considered as waves

Characteristics of electromagnetic waves:– Wavelength (λ), frequency (ν), wavenumber (ν) (‘color’ of light)– wavenumber = 1/ λ (number of waves / unit length)

Blue light ~ 400 - 450 nm wavelength, 25 000 – 22 222 cm-1

Red light ~ 600 - 700 nm wavelength, 16 667 – 14 286 cm-1

– Polarization: direction of transversal vibration of the wave– linearly polarized light vs. randomly polarized– Speed: in vacuum (n=1); c = 2.99 × 108 m/s

Considered as a particle: photon (energy packet) of energy hν.

A wave: periodic change of a property in space and/or time.

Instantaneous snap shot of a linearly polarized wave with graphic of field intensities.

Basic facts of electromagnetic radiation (wave)

EM radiation is a travelling wave; has mutually perpendicularmagnetic and electrical fields and these fields fluctuate periodically in time and space.

hcphoton energy h= =ν

λ

xy plane, plane of polarization

Page 2: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

λν=c

Distance between two consecutive similar pointsalong the wave (with same phase and amplitude) is λ.

c

If the vector of the electric field (measured at a fixed point of space) oscillates along a straight line then the waves are called plane-polarized or linearly polarized waves.

Animation of a pencil of polarized light.

View a linearly polarized electromagnetic wave traveling towardsyou. Then you would observe the amplitude variation of the wave in one plane – polarized light – plane of polarization.

The above animation presents a wave that is plane-polarized in a vertical plane (it can be any plane).

Natural light: consists of pencils of light with their planes of polarization in all possible directions.

Plane Polarizer

θ

Polarizer - an optical filter that passes light of a specific polarization and blocks waves of other polarizations.

Polarizer 1 Polarizer 2

Polarizers:Planes of polarization parallel

Polarizer 1 Polarizer 2

No light

Polarizers:crossed

Page 3: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

EtrErot

diatomicmolecule

model

Energy of a molecule; motion, electronic, nuclear etc.

Electromagnetic radiation covers a large energy range

The energy (E) of a molecule consists of translational, rotational, vibrational and electronic energy (major terms).

Emolecule = Enuclearspin + E electronspin + Etranslational +Erotation +Evibration +Eelectronic

Translation – movement of the molecule through spaceSpin – nuclear and electron spinRotational – rotation of the moleculeVibrational – vibration of the atoms with respect to each otherElectronic – mutual interactions of the electrons and nuclei

All these energy states (stationary states) are quantizedexcept translational energy.

Quantized states have precisely assigned energy values.

Optimum attractionpoint

Bond Energy

Bond length

Electronic ground state

In molecules electrons reside molecular orbitals, σ, π, σ*, π* and non-bonding orbitals

Morse curveshape depend on the moleculargeometry ofthe state.

Electronic ground state

Ene

rgy

Inter-atomic distanceBond length

Morse curve

Electronic energy σ

Ene

rgy

Bond distance

Bonding MO

Anti-bonding MO

σ

σ*

The difference in energy between molecular bonding, non-bonding and anti-bonding orbitals ranges from 125-650 kJ/mole

For every covalent bonding orbital in a molecule there is a corresponding anti-bonding orbital of higher energy.

not to scale

Page 4: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Electronic energy levels and transitions

________________

________________

__________________

__________________

___________________

σ *

π *

n

π

σ↑↓

↑↓

↑↓

∆E ~ visiblephotons

Chromophoreshave n and πelectrons.

Each electronic state is associated with a set of vibrational states; and each vibrational state is associated with a set of rotational states.

The energy differences between rotational and vibrational levels are much smaller.

Vibrational levels

Rotational levels

Electronic level

Vibrational levels

Rotational levels

An electronic level

O

Covalent molecules: Bonding involves molecular orbitals.

Electrons are in molecular orbitals σ,π,δ,..and lone pairs (generally many molecular orbitals exist, most important here HOMO and LUMO).

In addition σ*,π*,δ*,..(anti-bonding molecular orbitals are formed).

Most molecules in ground state have electron spins paired, no net spin, singlet states.

C C

CC C O CC NN

C Br C OH C NH2 NHC

σ bonding

π bonding

Non-bonding

Ground State Molecular Orbitals of molecules: Chromophores

A functional group capable of having characteristic electronic transitions (>200nm) is called a chromophore.

Photon and molecule interactions

When electromagnetic radiation (photon) ’interacts' with a molecule, scattering or absorption occurs at the outset.

Photon and molecule interactions: Types of scattering:

1. Elastic Scattering – all energy of the colliding photon is preserved, wavelength of light remains unaltered (Rayleigh scattering – elastic scattering).

2. Inelastic scattering means that the frequency of photons ofinteracting light changes upon interaction with a sample.

Page 5: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

UV-VIS Spectroscopy

Photons in the region of UV-VIS are of energy comparable toenergy required for transitions between electronic states.

UV-VIS absorption arises from electronic transitions.

Atoms → molecules : atomic orbitals → molecular orbitalsElectronic transitions ↔ molecular orbitals.

Ions, metal ion complexes → d-d transitions and/or charge transfer complexes (internal redox reaction)

The difference in energy between bonding, non-bonding and anti-bonding orbitals ranges from 125-650 kJ/mol; corresponds to EM radiation in UV 100-350nm, and VIS 350-700nm.

σ bonding

σ* anti-bonding

π* anti-bonding

π bonding

n non-bonding: lone pairsEnergy

Absorption in UV-VIS RegionGeneral electronic energy hierarchy in molecules

Excited states

∆E1

∆E2

∆E3 ∆E4

∆E1 > ∆E2 > ∆E3 > ∆E4

ν1 > ν2 > ν3 > ν4

λ1 < λ2 < λ3 < λ4

C

C C

C C

C

O

C C

σ → σ* 135nm

σ→ σ* 150nmn → σ* 183nm

π → π* 175nm

π → π* 254nm

n → π* 290nm

Some transitions

OH

Chromophores

Conjugation: Spreading of π-MOs over the molecularframework. Leads to narrowing of HOMO-LUMOgap. Increases λmax - bathochromic effect.

π

π*

HOMO

LUMO

p p

Conjugation: Spreading of π-MOs over the molecularframework. Leads to narrowing of HOMO-LUMOgap. Increases λmax - bathochromic effect.

π

π*

π

π*

π

π*

π

π*

HOMO

LUMO

Conjugation: Spreading of π-MOs over the molecular framework. Leads to narrowing of HOMO-LUMO gap.

Energy

http://www2.chemistry.msu.edu/faculty/reusch/VirtTxtJml/Spectrpy/UV-Vis/spectrum.htm

CH2

CH2

CH2

CH2

CH2

CH2

Page 6: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

C CN

N

N

λmax=180-200nmε=10000

λmax=217nmε=21000

λmax=258nmε=35000

λmax=250nm

λmax=290nm

λmax=360nm

A lc k cλλ λε= =

∆E1 > ∆E2 > ∆E3 > ∆E4

ν1 > ν2 > ν3 > ν4

λ1 < λ2 < λ3 < λ4

Photon energy, frequency and wavelength relationships

HOMO

LUMO

Intramolecular charge transfer transitions absorbs radiation in the electronic spectra as well.

N+

NH2O

O-

NH2+

N

O-

O

IE EA

h

−ν =

1st excited state (electronic)S1

S0

Singlet electronicground state

σ

σ*

Allowed transitions Forbidden transitions

Page 7: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Fine structure

I Gas phase, room temperature

II In isopentane-methylcyclohexane matrix, 77K

III In cyclohexane, room temperature

IV In water, room temperature

1,2,4,5 tetrazine

http://131.104.156.23/Lectures/CHEM_207/uv-vis.htm

Effect of the environmenton the spectral shape.

Condensed phase spectra do not have much fine structuredue to significant line broadening.

Gas phase UV-VIS spectra has fine structure.

Lambert-Beer Law (all absorptionSpectroscopies)

Aλ=ελlcAbsorbance (optical density)

ελ = molar absorptivity(extinction coefficient)

l = path length (cm)c =concentration (M)

Page 8: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Lambert-Beer Law: Derivation

A beam of monochromatic light falling on an absorbing and homogeneous medium loses the intensity of the beam (Power, P) as it traverses through the medium. The loss of the power of the beam, dP, for a traveled path length, dx, is proportional to the concentration of the absorbing species, power of the beam and the length dx.

∝dP c

dP P∝

Therefore,

dP cPdx= −β

P0 Pout

b

c

x=o x=b

dxP P+dP

λ

αdP dx

x

dPcdx

Pβ= −

Integrating within limits;

λλ

= − β ⇒ = − β = −β

= −β = = =

=

β ε ε

ε

∫ ∫ ∫ ∫out

0

P bout

0P 0

out 010 10

0 out

PdP dPcdx cdx ln bc

P P P

P P bc2.30

2.3033log bc log bc A bc

P P

A cb

which is

Absorbance is directly proportional to c.

molar absorptivity at λ

αdP Pcdx

bcout

bcout

PP

ePPλ

λ

ε

β

=

=

100

0

Other forms: out

0

out

0

0

out

10 10

PT

P

P%T 100

P

PA log

P

1A log (T) log

T

=

= ×

=

= − =

molar absorptivities vary by orders of magnitude:104-106 are termed high intensity absorptions103-104 are termed low intensity absorptions0 to 103 are the absorptions of forbidden transitions

1 max

2 max

3 shoulder

Some molecules, depending on the electronic energy leveldifferences can have multiple maxima in UV-VIS spectra.

A

c

1

2

Sensitivity of 1 > 2Data Scatter from 3 very high

Calibration curve:Absorption of a molecule is usually characterized by itsmolar absorptivity.

Aλ=ελbc

ελ = absorbance of a solution of unit concentration ina unit path length at wavelength λ.It is a measure of the transition probability.

For mixtures absorbing at the same λ;

Aλ=Σ ελ,ibci

Page 9: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

010

out

P log A bc

P= = ε

Single beam spectrometer: Block Diagram:

1. Select a wave length2. Calibration P0

3. Measurement P

−ε= bcout 0 P P 10 Note the form

“monochromaticlight”

Entrance and Exit slits ~nm

Po or Pout

Radiation (distribution) froma perfect blackbody

Xenon lamp emission spectrum

UV-VIS Sources:

UV - Deuterium lampVIS - Tungsten lamp

Note the intensity distribution(uneven)

Fastie-Ebert Configuration

A monochromator configuration:

Entrance slit Exit slit

A monochromator is an optical device that transmits a selectable narrow band of wavelengths of light or other radiation chosen from a wider range of wavelengths available at the input.

Page 10: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

010

out

Plog A bc

P= = ε

Double beam spectrometer: Block Diagram

iD

iS

iR

R D10

S D

i ilog A

i i− =−

DOUBLE-BEAM to correct for noise, drift and other Instabilities.

Double Beam Spectrophotometer Different Design.

Optical system of a double-beam spectrophotometer

Non-linearity of calibration curve: instrument contributioncomes from the slit width

Band width ↔ physical (instrument) slit width

A range of wave lengths (more than one in number, for sure) enters the sample - polychromatic beam.

Monochromator Band Width

Rays exiting the monochromator is strictly not monochromatic.Only nearly monochromatic, has a wavelength distribution around (band width) λ, commonly referred to as monochromatic radiation wavelength. The smaller the band width (FWHM) of the beam, the better is the system.

λ

FWHM ~ bandwidth

Page 11: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Spectral slit width is defined as the total spread of wavelengthsrepresented by the blue lines.

The spectral bandwidth will always be narrower than the spectralslit width.

When a monochromator is set to a particular wavelength,light with a Gaussian intensity distribution of wavelengthsemerges from the exit slit. The heavy black line represents the distribution of light that reaches the sample.The spectral bandwidth is defined as the FWHM.

FWHMFor a monochromatic beam at λ,

log (Po/P) = A = εbc = kc (11)Po/P = 10εbc (12)P=P010-εbc

Similarly, at another λ’, Po/P'= 10ε’bc (13)P’=P010-ε’bc

The radiant power of two wavelengths passing through the solvent is given by (Po + P’o), and that passing through the solution containing absorbing species by (P + P’).

Non monochromatic radiation leads calibration plots to deviate from linearity. Illustration: (beam with two wavelengths – simplestnon monochromatic beam)

Po

P’o

λ

λ’

P = Po 10 -εbc

P’ = P’o 10 -ε’bc

P = Po 10 -εbc

log (Po/P) = log(10 εbc) = εbc

linear

The absorbance due to both wavelengths is

A = log [(Po + P’o)/( P + P’)] (14)

Substituting for P and P', we obtain; A = log [(Po + P’o)/( Po 10-εbc+ P’o10-ε’bc)] (15)

Non-linearity arises because molar absorptivity is differentfor the two wavelengths; seen clearly at high c values.

−ε −ε

= +

=

'

bc ' 'bc0 0

'o

kbco o

P (P P )

P = (P 10 P 10 ) a sum !!

not in the Bee P P 10 )r's Law form ( even if P;

+

P

=

A

c

Deviation from linearityincreases with|ε - ε’|

DetectorResponse

An incident beam with more than two wavelengths wouldresult in a deviation quickly on the c axis.

The absorbance due to both wavelengths is

A = log [(Po + P’o)/( P + P')] (14)

Substituting for P and P', we obtain; A = log [(Po+ P’o)/( Po10-εbc+ P’o10-ε’bc)] (15)

Non-linearity of calibration plots arise because molar absorptivities ε ≠ ε’ for the two wavelengths; if ε ≡ ε’it would be linear!!

If ε ≡ ε’ ;A = log [(Po+ P’o)/( Po10-εbc+ P’o10-εbc)] (15)

A = log [(Po+ P’o)/ [(Po+ P’o) 10-εbc)]

A = log [10εbc] = εbc :Beer’s Law !!

Page 12: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Slope = sensitivity

Detector Response Detector Response

The dynamic range of a detector is that concentration range over which a concentration dependent output is produced.

The minimum of the range will be the concentration at which the output is at the detection limit and the maximum that concentration where the detector no longer responds to a concentration increase.

The dynamic range is usually given as a concentration ratio and is thus, dimensionless.

The linear dynamic range of a detector is that concentration range over which the detector output is linearly related to solute concentration.

A

c

Deviation from linearityincreases with|ε - ε’|

DetectorResponse

Flatter maxima has a lower probability of a large |ε - ε’| and therefore produces a desirable larger linear dynamic range in calibration plots.

Pick a moderately large flatter peaks for quantitation.

Narrowpeak

Broadpeak

‘Flatter’ peaksyield better linearity

‘Flatter’ peaksyield better linearity

Page 13: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Bandwidth vs Resolution

Resolution: ~ Ability to distinguish consecutive peaks.

Typical Optical Path: Monochromator

from source

Czerny-Turner design

Monochromator bandwidth is directlyrelated to the slit widths in addition to grating characteristics.

to samplecompartment

Czerny-Turner design

Turning the grating around it’s center allows “different monochromatic λ” s to leave the exit slit of monochromator.

Diffracted beams;energy in different orders are different.

First ordern=1 Second order, n= 2Incident beam

Czerny–Turner monochromator.

http://en.wikipedia.org/wiki/Monochromator

1d

#lines per unit length=

Energy maximum for λ diffracted such that α=β.

Grating surface

Tilt of rulings

http://www.shimadzu.com/products/opt/oh80jt0000001uz0.html

Page 14: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

Angular Dispersion of Gratings

Dispersion is a ‘measure’ of the angle the grating must be rotated to change wavelength of the the exiting light by aunit wavelength.

i.e. angular separation dφ obtained for two wavelengths separated by dλ.

φλ φ

∆ =∆ d

ncos

Smaller d values of the grating generate larger dispersions leading to higher resolutions.

Monochromator Band Width

A ray exiting the monochromator is strictly not monochromatic.Only nearly monochromatic, has a wavelength distribution (band width) about a, λ, commonly referred to as monochromatic radiation wavelength. The smaller the band width (FWHM) of the beam, the better is the monochromator.

λ

FWHM

Two consecutive peaks with ∆λ having a valleyof 5% are said to be tworesolved peaks.

λλ

= =∆

R nN

λ

λλ

= =∆

R nN

λ∆

Spectral resolution is a measure of the ability of an instrumentto differentiate between two adjacent wavelengths λ and λ+∆λ.

Resolutionλλ

= =∆

R nN

n = orderN= # grooves illuminated

Resolution is the ability of an instrument to separate light into finite, distinct wavelength regions and distinguish these finite regions from each other. It is primarily governed by the physical slit width of the instrument.

In combination with the inherent angular dispersion of light from the exit of the monochromator to the detector of theinstrument determines resolution.

Illuminated regionof the grating

Page 15: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

http://people.whitman.edu/~dunnivfm/FAASICPMS_Ebook/CH2/2_2_9.html

Detectors:

Phototube

http://kids.britannica.com/comptons/art-53788/Phototube

Detectors:

PMT

Note: uneven response

Detectors:

Diodes (semiconductor)

-Si-Si-Si-Si-Si-Si-| | | | | |

-Si-Si-Si-Si-Si-Si-| | | | | |

Si-Si-Si-Si-Si-Si-| | | | | |

Si-Si-Si-Si-Si-Si-

Si Crystal

-Si-Si-Si-Si-Si-Si-| | | | | |

-Si- P-Si-Si-Si-Si-| | | | | |

Si-Si-Si-Si- Si-Si-| | | | | |

Si-Si-Si-Si-Si-Si-

n-type Si

-Si-Si-Si-Si-Si-Si-| | | | | |

-Si-Si-Si-Si-Si-Si-| | | | | |

Si-Ga Si-Si-Si-Si-| | | | | |

Si-Si-Si-Si-Si-Si-

p-type Si

Slight excess of e’s Slight deficit of e’s

p-type n-type

_+

Foward biasi ≠ 0

p-n junction

p-type n-type

+_

Reverse biasi = 0

Depletion region

Page 16: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

p-type n-type

+_

Reverse bias

+ e

Current (signal) produced is proportional number ofphotons falling (i.e. intensity of light) at the junction.

Depletion regionPhotosensitive

hνPhoton moves an electron from the VB to CB, generates a current.

Detectors:

Diode Array

Diode-Array Spectrophotometer

UV-VIS in Quantitative Analysis

Inherently UV-VIS absorbing speciesOrganic chromophores, transition metal ionsoxoanions, elemental halogens.

Non-absorbing species may be converted to absorbingspecies quantitatively.

Fe+3 + SCN- → FeSCN+2

Cr+3 → CrO4-2

Ni+2 → Ni-DMGPb+2 → Pb-Diphenyl-thiocarbazone

General Procedure (Single species)

a. select a λ more sensitive but less susceptible to slight drifts in λ(better reproducibility).

b. choose the solvent (medium) to disperse the ‘test’ and standards, dust free.

solvent, pH, temperature, ionic strength and interferents.c. matched cells (clean, scratch-less) d. generate a (external) calibration curve (A vs. c) : A = mc + b

If matrices of ‘tests’ and standards are not comparable it is bestto use standard addition method, not the external calibration curve for quantitation.

Page 17: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

The color exhibited by a solute and/or the intensity of the peaksmay vary with the nature of the solvent in which the chromophore is dissolved in.

Bathochromic effect Hypsochromic effect

Hyperchromic effect

Hypochromic effect

λ

Solvent Effects: Solvent Effects:

gs µg

es µe µg > µeµg < µe

Gas phase/non-polar solvent

Polar solventPolar solvent

Note: In general other factors in addition to the polarity of the solvent molecules are involved in solvent effects (solvatochromism) leading to wavelength shifts.

Solvent Effects:

gs µg

es µe µg > µeµg < µe

T

S

If one or more of T, S and p is colored, spectrophotometric titrations are possible.

Sample reaction mixture after each addition of each aliquot (v), measure Aobs.

Spectrometric/Photometric Titrations

Analyte (S) + Titrant (T) → Products (p)

Corrected Absorbance for dilution;

v+= obs

VA A

V

V

v

Abs

orba

nce

Vol. titrant

S + T p

εT>0

Abs

orba

nce

Vol. titrant

εP>0

Abs

orba

nce

Vol. titrant

S + T p

εS> εT

Abs

orba

nce

Vol. titrant

S + T p

εT> εp

Abs

o rba

nce

Vol. titrant

S + T p

εT< εp

S+T p Advantages of Spectrophotometric Titration:

1. Calculations based on multiple measurements, minimizeserrors.

2. Change of Absorbance is monitored, interferents do not contribute to the final calculation.

Abs

orba

nce

Vol. titrant

with interferentwithout interferent

Page 18: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

3. Extrapolation: to determine equivalence point. Reactionswith relatively low K can be analyzed.

Abs

orba

nce

Vol. titrant

S + T p

Large K

Small K

4. Successive titrations possible. e.g. Bi+2, Cu+2, vs EDTA.

KBi-EDTA> KCu-EDTA

@745nm

Error in absorbance measurementsis dependent on the value of the absorbanceitself. Note the region of least error. Advisableto dilute samples so that absorbance falls in this region for quantification.

Jobs Method: Empirical Formula of binary compounds/complexes using the method of continuous variation or Jobs Method.

P + X = PaXb

The amount of one of the components is varied from a small value to a large value while the amount of the other component is varied from large to small. The amount of the product formed is monitored via the absorbance of the product.

P + k X = m D k=b/a m= d/a

If a series of solutions is made, each containing the same total number of moles of P and X, but a different mole ratio, R = (moles X/moles P); then the maximum amount of product, D, is obtained in the solution in which R = k (the stoichiometric ratio, b/a).

Note: the a, b, k as defined above.

When R > k, there is an excess of reagent X, so reagent P is the limiting reagent and X is in excess, in R > k region.

when R < k, there is an excess of reagent P, and X is limiting reagent and P is the limiting reagent, in R < k region.

a P + b X = d D = Pa Xb

Page 19: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

a P + b X = dD = Pa Xb

X limiting P limiting

b/a = 1/3 1/1 2/1

Isobestic Point:

A ↔ B (Inter-convertible species. Both generate spectra.)A B + H+

This term is usually employed with reference to a set of absorption spectra, plotted on the same chart for a set of solutions in which the sum of the concentrations of two principal absorbing components, A and B, is constant.

The curves of absorbance against wavelength (or frequency) for such a set of mixtures often all intersect at one or more points, called isobestic points.

Methyl Orange

Acidic to alkaline solutions (left to right).

http://en.wikipedia.org/wiki/Methyl_orange

H+

Bromothymol-blue

acidic, neutral, and alkaline solutions (left to right).

http://en.wikipedia.org/wiki/Bromothymol_blue

Molar absorptivity of A and B is the same at the isobestic point.

Isobestic points are commonly encountered when electronic spectra are taken;

(a) on a solution in which a chemical reaction is in progress (in which case the two absorbing components concerned are a ‘reactant’ and a ‘product’),

or (b) on a solution in which the two absorbing components are in equilibrium and their relative proportions are controlledby the concentration of some other component,

In all these examples, A (and/or B) may be coming from a single chemical species or a mixture of chemical species present in invariant proportion.

A B H ++⇌ XH X H− ++⇌

Page 20: Optical Spectroscopyweb.mnstate.edu/marasing/CHEM380/Documents/UV-VIS 2016.pdf · UV-VIS Spectroscopy Photons in the region of UV-VIS are of energy comparable to energy required for

A → B → C (all absorb)

Evidence of an intermediate of a reaction.

If absorption spectra of the types considered above intersect at many isosbestic points, this is prima facie evidence in case;

(a) for the formation of a reaction intermediate in substantial concentration,

(b) for the involvement of a third absorbing species in theequilibrium, e.g.

A B C→ →

2A B H C H+ ++ +⇌ ⇌