optical performance of 3 intraocular lens designs in the
TRANSCRIPT
Optical performance of 3 intraocular lensdesigns in the presence of decentration
Griffith E. Altmann, MS, MBA, Louis D. Nichamin, MD, Stephen S. Lane, MD,Jay S. Pepose, MD, PhD
Purpose: To study the theoretical optical performance of 3 intraocular lens (IOL)designs in the presence of IOL decentration.
Setting: Optics Center, Bausch & Lomb, Rochester, New York, USA.
Methods: A ray-tracing program was used to evaluate the effect of IOL decentrationon the optical performance of 3 silicone IOLs (LI61U, Bausch & Lomb; Tecnis Z9000,Advanced Medical Optics; and a new aberration-free IOL [SofPort AO, Bausch &Lomb]) in an experimental model eye. The study was done using pupil diameters of3.0 mm, 4.0 mm, and 5.0 mm and IOL decentrations of 0 mm, 0.25 mm, 0.50 mm,0.75 mm, and 1.00 mm. The modulation transfer functions were computed andplotted. A Monte Carlo simulation analysis with 1000 trials with IOL decentrationrandomly varying for each pupil size was performed.
Results: Decentration of LI61U and Tecnis Z9000 IOLs led to asymmetricalhigher-order aberrations that adversely affected the optical performance of themodel eye; performance was not affected with the aberration-free IOL because itlacks inherent spherical aberration. Optical performance with the aberration-freeIOL was better than with the LI61U IOL as the former has less spherical aberrationand did not introduce other aberrations when decentered. Performance with theaberration-free IOL was better than with the Tecnis Z9000 IOL for 3.0 mm, 4.0 mm,and 5.0 mm pupils when decentration exceeded 0.15 mm, 0.30 mm, and 0.38 mm,respectively. Performance with the LI61U IOL was better than with the Tecnis Z9000IOL for 3.0 mm, 4.0 mm, and 5.0 mm pupils when decentration exceeded 0.3 mm,0.5 mm, and 0.5 mm, respectively. Monte Carlo simulations showed the expectedpostoperative results of the LI61U IOL and aberration-free IOL would be repeatableand predictable, whereas the outcomeswith the Tecnis Z9000 IOLwould vary widely.
Conclusions: The optical performance of the model eye was not affected bydecentration of an aspheric IOL designed to have no inherent spherical aberration.With decentration, the performance with the new IOL was better than witha conventional spherical IOL and an aspheric IOL designed to offset the sphericalaberration of an average cornea.
J Cataract Refract Surg 2005; 31:574–585 ª 2005 ASCRS and ESCRS
The recent incorporation of wavefront aberrometry
into clinical ophthalmology practice has acceler-
ated the convergence of refractive and cataract surgery.
Standard intraocular lenses (IOLs) generally do not
correct the corneal spherical aberration of eyes having
cataract surgery, resulting in a suboptimal optical
transfer function. This has created renewed interest in
aspheric IOLs and the approval of the first IOL (Tecnis
Z9000, AdvancedMedical Optics) designed to offset an
� 2005 ASCRS and ESCRS
Published by Elsevier Inc.
average amount of corneal spherical aberration in
patients who have cataract surgery.
A basic tenet of medicine is primum non nocere.Whereas well-centered IOLs with an aspheric design
have the potential to improve optical performance and
contrast sensitivity by reducing postoperative spherical
aberration, numerous factors can cause IOL decentra-
tion with respect to the visual axis. Basic principles of
optics suggest that IOLs designed to correct corneal
0886-3350/05/$-see front matter
doi:10.1016/j.jcrs.2004.09.024
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Table 1. Optical characteristics of the theoretical pseudophakic model eye with a 22.0D LI61U IOL.
Surface Radius (mm) Conic Constant Thickness (mm)RefractiveIndex
Object — — Infinity 1.0
Cornea 7.575 �0.14135 3.6 1.3375
Iris — — 0.9 1.336
Anterior lens surface 8.234 0 1.202 1.427
Posterior lens surface �8.234 0 16.996 1.336
Retina — — — —
spherical aberration (eg, Tecnis) might result in a
significantly lower optical transfer function than with
equally decentered standard IOLs, particularly at higher
spatial frequencies, in part because of induced 2nd- and
3rd-order aberrations such as astigmatism and coma. If
this were the case, degradation in the quality of visual
function may necessitate explantation or surgical re-
positioning of decentered aspheric IOLs.
We describe an aspheric silicone IOL (SofPort AO,
Bausch & Lomb) that maintains an excellent optical
transfer function, is neutral to the induction of positive
or negative spherical aberration, and does not induce
significant higher-order aberrations (HOAs), even when
the lens is decentered. We compared the theoretical
optical performance of this IOL with that of 2 other
IOL models when various amounts of decentration
from the visual axis were simulated.
Materials and MethodsThe optical performance of 3 IOLs was evaluated in
a theoretical model of a pseudophakic eye using a commer-
Accepted for publication July 9, 2004.
From the Optics Center, Bausch & Lomb (Altmann), Rochester,New York; the Laurel Eye Clinic (Nichamin), Brookville, Pennsylvania;and the Pepose Vision Institute (Pepose) and the Department ofOpthalmology & Visual Sciences, Washington University School ofMedicine (Pepose), St. Louis, Missouri, USA.
Supported in part by Bausch & Lomb, Inc., and the Midwest CorneaResearch Foundation, St. Louis, Missouri, USA.
Mr. Altmann is an employee of Bausch & Lomb. Dr. Nichamin isa medical monitor for Bausch & Lomb. Dr. Lane has no financial orproprietary interest in any material or method presented. Dr. Peposehas received research support from Bausch & Lomb.
Reprint requests to Griffith E. Altmann, Bausch & Lomb, A54W,1400 North Goodman Street, Rochester, New York 14609, USA.E-mail: [email protected].
J CATARACT REFRACT SUR
cially available ray-tracing program (Zemax, Focus Software).The theoretical model eye1 was the model used to develop theTecnis Z9000 silicone IOL with 1 exception: A Gaussianapodization filter was placed in the entrance pupil to simulatethe Stiles-Crawford effect.2 According to van Meeteren,3 thetransmission, T, as a function of normalized radial position inthe entrance pupil of such a filter, was calculated by theequation
TðrÞ ¼ expð�ar2Þ
where r is the normalized radial pupil coordinate anda equals 0.054. Although the benefits of the Stiles-Crawfordeffect on spatial visual performance are small,4 the effect wasincluded for completeness.
The theoretical model eye was used in this study for 3reasons. First, the positive spherical aberration of the single-surface model cornea matches the average value measured inrecent clinical studies.5–12 Using the recently standardizedZernike polynomials and double-index format for thecoefficients,13 the Z4
0 Zernike coefficient for sphericalaberration of the average cornea is approximately 0.28 mmover a 6.0 mm central zone. Second, the anterior chamberdepth of 4.5 mm matches the measurements of IOL axialpositioning in pseudophakic eyes.14–17 Third, this is the best-case scenario for evaluation of the Tecnis Z9000 IOL. Table 1shows the optical characteristics of the model eye.
Three 22.0 diopter (D) silicone IOLs with differentoptical designs were studied: a conventional IOL withspherical anterior and posterior surfaces (LI61U, Bausch &Lomb), an advanced IOL with a prolate anterior surface anda spherical posterior surface (Tecnis Z9000), and a newaberration-free IOL with aspheric anterior and posteriorsurfaces (SofPort AO). Each IOL model was assumed to haveanterior and posterior optical zone diameters of 6.0 mm. TheLI61U IOL, like any IOL with spherical surfaces, has positiveinherent spherical aberration. The Tecnis IOL has negativespherical aberration designed to offset the positive sphericalaberration of the average cornea. The aberration-free IOL hasno spherical aberration. With the exceptions of asphericanterior and posterior surfaces and a 360-degree square edge,the aberration-free IOL is similar to the LI61U IOL. The Z4
0
575G—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Table 2. Optical characteristics and technical specification of the 3 IOL models.
Optical Characteristic LI61U Tecnis Z9000 SofPort AO
Power (D) 22.0 22.0 22.0
Optic material Silicone Silicone Silicone
Refractive index 1.427 1.458 1.427
Lens shape Equiconvex Equiconvex Biconvex
Anterior surface Sphere 6th-order asphere Conic asphere
Radius (mm) 8.234 11.043 7.285
Conic constant 0 �1.03613 �1.085657
4th-order constant 0 �0.000944 0
6th-order constant 0 �0.0000137 0
Posterior surface Sphere Sphere Conic asphere
Radius (mm) �8.234 �11.043 �9.470
Conic constant 0 0 �1.085657
Center thickness (mm) 1.202 1.164 1.206
A-constant 118.0 119.0 118.0
Optic size (mm) 6.0 6.0 6.0
Overall length (mm) 13.0 12.0 13.0
Haptic material PMMA PVDF PMMA
Haptic angulation (degrees) 5 6 5
PMMA Z poly(methyl methacrylate); PVDF Z polyvinylidene fluoride
coefficient for the inherent spherical aberration of the 22.0 DLI61U, Tecnis, and aberration-free IOLs over a 6.0 mm zoneis 0.32 mm, �0.72 mm, and 0 mm, respectively. Table 2shows the optical characteristics and technical specificationsof the 3 IOLs. One of the authors (G.E.A.) provided theoptical characteristics of the LI61U and aberration-free IOLs.The optical characteristics of the 22.0 D Tecnis Z9000 IOLwere taken directly from the U.S. patent.1
Each IOL was evaluated by centering it in the theoreticalmodel eye so that the anterior surface of the IOL was 0.9 mmbehind the iris. For each combination of IOL model andpupil diameter, the distance between the posterior surface ofthe IOL and the retina was optimized to attain the best opticalperformance for an on-axis object located at infinity ata wavelength of 555 nm. When an IOL is perfectly centered,only axial aberrations (eg, spherical aberrations) of the modelcornea and the IOL itself degrade the image on the modelretina. The IOL was then successively decentered in the tan-gential plane by 0.25mm, 0.50mm, 0.75mm, and 1.00mm.The optical performance of the model eye was evaluated foreach combination of IOL model, pupil diameter, and IOLdecentration. The distance from the IOL to the retina was notreoptimized for each iteration of decentration. The cornea,pupil, and retina were always centered on the optical axisof the theoretical model eye. Pupil diameters of 3.0 mm,4.0 mm, and 5.0 mm were studied. An array of 512 � 512
576 J CATARACT REFRACT SUR
(262 144) rays was traced, and the modulation transferfunction (MTF) was computed for each simulation. Theresulting tangential and sagittal MTF curves over a spatialfrequency range of 0 to 60 cycles per degree (cpd) wereplotted for each simulation.
For each combination of IOLmodel and pupil diameter,a Monte Carlo simulation of 1000 trials was performed. Inthe analysis, the amount of IOL decentration was randomlyselected from a probability distribution, which was derivedfrom a compilation of several recent IOL decentrationstudies.18–24 Descriptive statistics showed the mean of theprobability distribution used in the Monte Carlo analysis was0.36 mm 6 0.22 (SD).
ResultsThe performance of each IOL model in the
presence of decentration was analyzed for 3 pupil
diameters. Figures 1 to 3 show a total of 9 MTF plots
(3 pupil diameters � 3 IOL decentrations).
3.0 mm PupilFor a 3.0 mm pupil, the adverse effects of the
spherical aberration of the cornea and the IOL were
G—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Figure 1. The MTF curves for a theoretical pseudo-
phakic model eye with a 3.0 mm pupil. A: The IOLs are
perfectly centered, and the adverse effects of corneal
and IOL spherical aberration are minimal. B: The IOLs are
decentered 0.5 mm. Induction of asymmetrical HOAs
degraded the performances of LI61U and Tecnis IOLs,
causing the MTF curves to droop and separate. C: The
IOLs are decentered 1.0 mm, further degrading perfor-
mance of the LI61U and Tecnis IOLs but not the
aberration-free IOL (S Z sagittal, T Z tangential).
577J CATARACT REFRACT SURG—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Figure 2. The MTF curves for a theoretical pseudo-
phakic model eye with a 4.0 mm pupil. A: Although the
IOLs are perfectly centered, corneal spherical aberration
decreases the performance of the LI61U and aberration-
free IOLs. B: The IOLs are decentered 0.5 mm. Induced
asymmetrical HOAs cause the MTF curves of the LI61U
and Tecnis IOLs to droop and separate but negligibly
affect the curves of the aberration-free IOL. C: The IOLs
are decentered 1.0 mm. Performance with the LI61U IOL
and Tecnis IOL further degrades; the aberration-free IOL is
unaffected (S Z sagittal, T Z tangential).
578 J CATARACT REFRACT SURG—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Figure 3. The MTF curves for a theoretical pseudo-
phakic model eye with a 5.0 mm pupil. A: The IOLs are
perfectly centered. The adverse effects of corneal
spherical aberration cause poorer performance with
LI61U and aberration-free IOLs. B: The IOLs are
decentered 0.5 mm. Induced asymmetrical HOAs
significantly degrade performance of the Tecnis IOL,
causing its MTF curves to droop and separate sub-
stantially. Degradation with the LI61U IOL is less pro-
nounced because spherical aberration dominates the
induced aberrations. The MTF curves for the aberration-
free IOL are negligibly affected. C: The IOLs are
decentered 1.0 mm (S Z sagittal, T Z tangential).
579J CATARACT REFRACT SURG—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
small. The Z40 coefficient for corneal spherical
aberration was 0.016 mm. The performance of the
model eye with no decentration was near diffraction-
limited with all 3 IOL designs (Figure 1, A). As the IOLs
decentered (Figure 1, B and C ), the performance
degraded with the LI61U and Tecnis IOLs, but not with
the aberration-free lens. The inherent spherical aberra-
tion of the LI61U and the Tecnis IOLs created
asymmetrical HOAs when the IOLs were decentered.
Astigmatism and coma were the primary induced
asymmetrical aberrations. The aberration-free IOL,
which does not have inherent spherical aberration, did
not create HOAs when it decentered.
The model eye performed better with the aberra-
tion-free IOL than with the LI61U IOL at all spatial
frequencies and all IOL decentrations and than with the
Tecnis IOL when decentration exceeded 0.15 mm.
Performance was better with an aberration-free IOL
decentered by 1.0 mm than with a perfectly centered
LI61U IOL at all spatial frequencies and a Tecnis IOL
decentered by 0.15 mm. The performance was better
with the Tecnis IOL than with the LI61U IOL when
decentration was less than 0.3 mm.
4.0 mm PupilFor a 4.0 mm pupil, the adverse effects of spherical
aberration of the cornea and the IOL were more
problematic. The Z40 coefficient for corneal spherical
aberration was 0.051 mm. When the IOLs were
perfectly centered (Figure 2, A), the performance of
the model eye with the Tecnis IOL was diffraction-
limited by design. The performance with the aberra-
tion-free IOL was reduced by the spherical aberration of
the cornea, and the performance with the LI61U IOL
was further reduced by the lens’ inherent positive
spherical aberration. As the IOLs decentered (Figure 2,
B and C ), the performance of the model eyes degraded
with the LI61U and Tecnis IOLs but not with the
aberration-free IOL.
The model eye performed better with the aberra-
tion-free IOL than with the LI61U IOL at all spatial
frequencies and all IOL decentrations and than with
the Tecnis IOL when decentration exceeded 0.30 mm.
Performance was better with an aberration-free IOL
decentered by 1.0 mm than with a perfectly centered
LI61U IOL at all spatial frequencies and than a Tecnis
IOL decentered by 0.30 mm. Overall, the performance
580 J CATARACT REFRACT SUR
with the Tecnis IOL was better than with the LI61U
IOL when decentration was less than 0.50 mm. With
IOL decentration of 0.50 mm, the performance with
the Tecnis IOL was better than with the LI61U IOL
at spatial frequencies less than 20 cpd.
5.0 mm PupilThe adverse effects of spherical aberration of the
cornea and the IOL were most significant for a 5.0 mm
pupil. The Z40 coefficient for corneal spherical
aberration was 0.130 mm. When the IOLs were
perfectly centered (Figure 3, A), the performance of
the model eye with the Tecnis lens was diffraction-
limited by design. The performance of the model eye
with the aberration-free IOL was reduced by spherical
aberration of the cornea, and the performance with the
LI61U IOL was further reduced by its inherent
spherical aberration. As the lenses decentered (Figure 3,
B and C ), the performance of the model eye degraded
with the LI61U and Tecnis IOLs but not with the
aberration-free IOL.
The model eye performed better with the aberra-
tion-free IOL than with the LI61U IOL at all spatial
frequencies and all IOL decentrations and than with
the Tecnis IOL when decentration exceeded 0.38 mm.
Performance was better with an aberration-free IOL
decentered by 1.0 mm than with a perfectly centered
LI61U IOL and a Tecnis IOL decentered by 0.38 mm.
Overall, the performance with the Tecnis IOL was
better than with the LI61U IOL when decentration was
less than 0.50 mm. With decentration of 0.50 mm, the
performance with the Tecnis IOL was better than with
the LI61U IOL at spatial frequencies less than 15 cpd.
Monte Carlo AnalysisThe averages of the tangential and sagittal MTF
curves for a 3.0 mm, 4.0 mm, and 5.0 mm pupil
diameter are shown in Figures 4, 5, and 6, respectively.
The MTF curves for the worst 10% of cases, the best
10% of cases, and the median cases for each IOL model
are shown. Because the performance with the aberra-
tion-free IOL is independent of IOL decentration, the
worst 10%, best 10%, andmedianMTF curves lie upon
one another. Because the LI61U andTecnis IOL designs
have inherent spherical aberration, their performances
are dependent on IOL decentration; thus, the worst
10%, best 10%, and median MTF curves are separated.
G—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Figure 4. From a Monte Carlo analysis, the best 10%,
median, and worst 10% MTF curves for a theoretical
pseudophakic model eye with a 3.0 mm pupil, showing
the results with the LI61U and aberration-free IOLs are
expected to be excellent and repeatable. Excellent
results would also be expected with the Tecnis IOL in
all but 10% of cases.
Figure 5. From a Monte Carlo analysis, the best 10%,
median, and worst 10% MTF curves for a theoretical
pseudophakic model eye with a 4.0 mm pupil, showing
the results with the LI61U IOL are expected to be good
and repeatable and the results with the aberration-free
IOL, even better. The results with the Tecnis IOL are
expected to vary widely; the average result is expected
to be better than with the LI61U IOL but worse than with
the aberration-free IOL.
Figure 6. From a Monte Carlo analysis, the best 10%,
median, and worst 10% MTF curves for a the theoretical
pseudophakic model eye with a 5.0 mm pupil, showing
the results with the LI61U IOL are be expected to be good
and repeatable and the results with the aberration-free
IOL, even better. The results with the Tecnis IOL are
expected to vary widely, with exceptional results in 10%
of the cases. Average low spatial frequency (contrast
sensitivity) results with the Tecnis IOL are to be better
than with the LI61U and aberration-free IOL; however,
better mid and high spatial frequency (visual acuity)
results are expected with the LI61U and aberration-free
IOLs.
581J CATARACT REFRACT SURG—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
Greater separation between the worst 10% and best 10%
MTF curves indicates less repeatability and predictabil-
ity in postoperative outcomes. Thus, in general, repeat-
able and predictable postoperative results would be
expected with the LI61U and aberration-free IOLs and
widely varying results with the Tecnis IOL.
For a 3.0 mm pupil (Figure 4), all MTF curves for
the aberration-free IOL lie above the MTF curve for
a perfectly centered LI61U IOL and nearly coincidewith
the best 10% MTF curve for the Tecnis IOL. Thus,
optical performance with the aberration-free IOL was
better than with the LI61U IOL in 100% of cases and
than with the Tecnis IOL in slightly less than 90% of
cases. The median MTF curves of the LI61U IOL
and Tecnis IOL lie almost upon one another; thus,
the median performance with the 2 lenses was nearly
equivalent.
For a 4.0 mm pupil (Figure 5), all MTF curves for
the aberration-free IOL lie above the MTF curve for
a perfectly centered LI61U IOL and the median MTF
curve for the Tecnis IOL. Thus, optical performance
with the aberration-free IOL was better than with the
LI61U IOL in 100% of cases and than with the Tecnis
IOL in most cases. Similar to the aberration-free IOL,
the worst 10%, best 10%, and median MTF curves for
the LI61U IOL are similar to one another, showing that
IOL decentration only slightly affects optical perfor-
mance. In other words, for a pupil diameter of 4.0 mm
or greater, the spherical aberration of the cornea is the
dominant aberration. The median MTF curve of the
Tecnis IOL lies above the MTF curves of the LI61U
IOL, showing performance with Tecnis IOL was better
than with the LI61U IOL in most cases.
For a 5.0 mm pupil (Figure 6), all MTF curves for
the aberration-free IOL lie above the MTF curve for a
perfectly centered LI61U IOL; thus, performance with
the aberration-free IOL was better than with the LI61U
IOL in 100% of cases. In most cases, the performance
with the aberration-free IOL was better than with the
Tecnis IOL at spatial frequencies greater than 17 cpd,
and vice versa. The performance with the Tecnis IOL
was better than with the LI61U IOL in the most cases,
especially at spatial frequencies lower than 23 cpd.
DiscussionThe introduction of the Tecnis Z9000 silicone IOL
has generated much interest in lenses with aspheric
582 J CATARACT REFRACT SUR
surfaces. The Tecnis IOL is designed to provide
improved retinal image quality by compensating for
the average corneal spherical aberration of the pop-
ulation. A recent theoretical study25 primarily evaluated
the on-axis performance of the Tecnis IOL and
downplayed the effects of IOL decentration and tilt.
However, perfect centration is rare for many reasons
including in–out of the bag placement, incongruency
between bag diameter and overall IOL diameter, a large
capsulorhexis, asymmetrical capsule coverage, IOL
placement in sulcus, capsule fibrosis, capsule phimosis,
and radial bag tears.26 Even if the IOL is perfectly
centered, the other optical components of a human eye
are rarely, if ever, centered on the visual axis or on any
common axis.
The ray-tracing results show that IOL decentra-
tion adversely affects the optical transfer function
in pseudophakic eyes with IOLs that have inherent
spherical aberration. Theoretical studies of conventional
lenses with spherical surfaces show IOL decentration
induces defocus, astigmatism, and coma and the
magnitude of the aberrations depends on the magnitude
of the inherent spherical aberration.27–30 Because it is
uncorrectable with spectacles, coma may be the pre-
dominant monochromatic source of retinal image
degradation31 and may be related to the glare and halos
that have been observed with decentered IOLs.
In our eye model, performance with the aberration-
free IOL was always better than with conventional
IOLs with spherical surfaces for 2 reasons. First, the
aberration-free IOL has no spherical aberration, so
a pseudophakic eye with the aberration-free IOL would
be expected to have less spherical aberration. Second,
the aberration-free IOL does not induce other aberra-
tions when decentered, so a pseudophakic eye with this
IOL would be expected to have less induced aberration.
Even when it was decentered 1.0 mm or more, the
optical performance with the aberration-free IOL was
better than with a perfectly centered conventional IOL.
In the Monte Carlo analyses, the aberration-free IOL
always performed better than the LI61U IOL, regardless
of IOL decentration or pupil diameter.
Performance with the aberration-free IOL was
better than with the Tecnis IOL for 3.0 mm, 4.0 mm,
and 5.0 mm pupils when IOL decentration exceeded
0.12 mm, 0.30 mm, and 0.38 mm, respectively. Based
on the Monte Carlo analysis, for patients with a
G—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
postoperative pupil diameter smaller than 5.0 mm, the
aggregate visual performance with the aberration-free
IOL would be expected to be better than that with the
Tecnis IOL. For a 5.0 mm pupil, the average perform-
ances would be similar.
Our results agreewith those ofHolladay et al.,25 who
report the theoretical performance of the Tecnis IOLwas
better than that of a conventional IOLwhen decentration
was less than roughly 0.4 to 0.5 mm. In particular, for
3.0 mm, 4.0 mm, and 5.0 mm pupils, the performance
with the Tecnis IOL was better than with the LI61U
IOL when decentration was less than 0.3 mm, 0.5 mm,
and0.5mm, respectively.Our results also agreewith clin-
ical observations32–34 that show eyes with a Tecnis IOL
have better contrast sensitivity than eyes with various
conventional lenses, although visual acuity is similar
between the 2 lens types. For moderate IOL decentration
and pupil diameters, the MTF curves of the Tecnis IOL
exceeded those of the LI61U IOL at spatial frequencies
less than 20 cpd but were nearly equivalent at higher
spatial frequencies. Contrast sensitivity is typically
measured at spatial frequencies lower than 20 cpd,
usually at 1.5, 3, 6, 12, and 18 cpd.35 For visual acuity
measurement, spatial frequencies greater than 20 cpd are
more important when acuity exceeds 20/40.
According to Holladay (J. Holladay, MD, ‘‘De-
fining and Measuring Quality of Vision and the
Theoretical Basis for the Tecnis Modified Prolate
IOL,’’ presented at the XX1st Congress of the European
Society of Cataract & Refractive Surgeons, Munich,
Germany, September 2003), there are contraindications
to Tecnis Z9000 IOL implantation because of the lens’
optical design. These include difficulty centering the
IOL, a small pupil, and a superprolate cornea. Another
contraindication to Tecnis IOL implantation is a decen-
tered visual axis, which can occur in patients with
keratoconus, keratoglobus, pellucid marginal degen-
eration, and corectopia and in those who have had
decentered corneal refractive surgery. There are no con-
traindications to the new aberration-free IOL resulting
from its optical design.
Optical performance should be measured along the
visual axis, not along the axis that passes through the
center of the pupil. Studies typically measure IOL
decentration relative to the center of the pupil.
However, the visual axis is generally offset from the
center of the pupil. Rynders and coauthors36 report that
J CATARACT REFRACT SUR
the mean displacement between the achromatic axis
and the center of the pupil in a series of young adult eyes
was 0.376 0.24 mm. Thus, the true decentration of an
IOL is probably greater than the values measured in the
aforementioned recent studies. Even an IOL perfectly
centered in the capsular bag may be significantly
decentered with respect to the visual axis. Furthermore,
the visual axis may shift given changes in pupil size and
shape under photopic,mesopic, and scotopic conditions.
Important questions remain unanswered. For
example, is correcting the average corneal spherical
aberration of the population the optimal target for an
IOL? It is common for cataract surgeons to target
emmetropia, and positive spherical aberration may help
patients with hyperopic postoperative refractions.
Modest amounts of spherical aberration may mitigate
the adverse effects of chromatic aberration37 and higher-
order monochromatic aberrations. Spherical aberration
can be compensated for with spectacle correction;
however, asymmetrical aberrations such as coma
cannot. Is the tradeoff of less spherical aberration for
more asymmetrical aberrations worthwhile? Because
most eyes, including those with exceptional vision, have
positive spherical aberration, the brain appears adept at
interpreting retinal images with positive spherical
aberration. With the Tecnis IOL, negative postopera-
tive spherical aberration is expected in 50% of patients.
Finally, left–right mirror symmetry has been observed
in corneal and total ocular aberrations. Because the
Tecnis IOL will likely introduce asymmetrical aberra-
tions arbitrarily, mirror symmetry of retinal aberrations
may be disrupted. Will the brain be as adept at
interpreting the retinal images with negative spherical
aberration and less mirror symmetry?
Other factors that affect the optical performance of
any pseudophakic eye were not assessed in the study.
These include residual postoperative refractive error,
corneal aberrations, IOL tilt, chromatic aberration, IOL
manufacturing imperfections, ocular pathology, and
scatter of optical media. Clinical evaluation of the
aberration-free IOL is required to verify theoretical
predictions of improved quality of vision and fewer
complications after cataract surgery.
References1. Norrby S, Artal P, Piers PA, et al, inventors; Pharmacia
Groningen BV, assignee. Methods of obtaining
583G—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
ophthalmic lenses providing the eye with reduced aber-rations. US patent 6 609 793. August 26, 2003
2. Stiles WS, Crawford BH. The luminous efficiency of raysentering the eye pupil at different points. Proc R SocLond [Biol] 1933; 112:428–450
3. van Meeteren A. Calculations on the optical modulationtransfer function of the human eye for white light. OpticaActa 1974; 21:395–412
4. Atchison DA, Joblin A, Smith G. Influence of Stiles-Crawford effect apodization on spatial visual perfor-mance. J Opt Soc Am A Opt Image Sci Vis 1998; 15:2545–2551
5. Guirao A, Redondo M, Artal P. Optical aberrations ofthe human cornea as a function of age. J Opt Soc Am AOpt Image Sci Vis 2000; 17:1697–1702
6. Artal P, Berrio E, Guirao A, Piers P. Contribution of thecornea and internal surfaces to the change in ocular aber-rations with age. J Opt Soc Am A Opt Image Sci Vis2002; 19:137–143
7. Wang L, Dai E, Koch DD, Nathoo A. Optical aberra-tions of the human anterior cornea. J Cataract RefractSurg 2003; 29:1514–1521
8. Mrochen M, Jankov M, Bueeler M, Seiler T. Correla-tion between corneal and total wavefront aberrations inmyopic eyes. J Refract Surg 2003; 19:104–112
9. Vinciguerra P, Camesasca FI, Calossi A. Statistical anal-ysis of physiological aberrations of the cornea. J RefractSurg 2003; 19:S265–S269
10. Kiely PM, Smith G, Carney LG. The mean shape of thehuman cornea. Optica Acta 1982; 29:1027–1040
11. Smith G, Cox MJ, Calver R, Garner LF. The sphericalaberration of the crystalline lens of the human eye. VisionRes 2001; 41:235–243
12. He JC, Gwiazda J, Thorn F, Held R. Wave-front aberra-tions in the anterior corneal surface and the whole eye.J Opt Soc Am A Opt Image Sci Vis 2003; 20:1155–1163
13. Thibos LN, Applegate RA, Schwiegerling JT, Webb R.Standards for reporting the optical aberrations of eyes;VSIA Standards Taskforce Members. OSA Trends inOptics and Photonics 2000; 35:232–244
14. Findl O, Drexler W, Menapace R, et al. Changes in in-traocular lens position after neodymium:YAG capsulot-omy. J Cataract Refract Surg 1999; 25:659–662
15. Landau IME, Laurell C-G. Ultrasound biomicroscopyexamination of intraocular lens haptic position after pha-coemulsification with continuous curvilinear capsulo-rhexis and extracapsular cataract extraction with linearcapsulotomy. Acta Ophthalmol Scand 1999; 77:394–396
16. Olsen T, Corydon L, Gimbel H. Intraocular lens powercalculation with an improved anterior chamber depthprediction algorithm. J Cataract Refract Surg 1995; 21:313–319
584 J CATARACT REFRACT SURG
17. Holladay JT, Maverick KJ. Relationship of the actualthick intraocular lens optic to the thin lens equivalent.Am J Ophthalmol 1998; 126:339–347
18. Hayashi K, Harada M, Hayashi H, et al. Decentrationand tilt of polymethyl methacrylate, silicone, and acrylicsoft intraocular lenses. Ophthalmology 1997; 104:793–798
19. WangM-C,Woung L-C,HuC-Y, KuoH-C. Position ofpoly(methyl methacrylate) and silicone intraocular lensesafter phacoemulsification. J Cataract Refract Surg 1998;24:1652–1657
20. Hayashi K, Hayashi H, Nakao F, Hayashi F. Compari-son of decentration and tilt between one piece and threepiece polymethyl methacrylate intraocular lenses. Br JOphthalmol 1998; 82:419–422
21. Kimura W, Kimura T, Sawada T, et al. Postoperativedecentration of three-piece silicone intraocular lenses.J Cataract Refract Surg 1996; 22:1277–1280
22. Hwang IP, Clinch TE, Moshifar M, et al. Decentrationof 3-piece versus plate-haptic silicone intraocular lenses.J Cataract Refract Surg 1998; 24:1505–1508
23. Schwenn O, Kottler U, Krummenauer F, et al. Effect oflarge positioning holes on capsule fixation of plate-hapticintraocular lenses. J Cataract Refract Surg 2000; 26:1778–1785
24. Taketani F, Matsuura T, Yukawa E, Hara Y. High-order aberrations with Hydroview H60M and AcrySofMA30BA intraocular lenses; comparative study. J Cata-ract Refract Surg 2004; 30:844–848
25. Holladay JT, Piers PA, Koranyi G, et al. A new intraoc-ular lens design to reduce spherical aberration of pseudo-phakic eyes. J Refract Surg 2002; 18:683–691
26. Walkow T, Anders N, Pham DT, Wollensak J. Causesof severe decentration and subluxation of intraocularlenses. Graefes Arch Clin Exp Ophthalmol 1998; 236:9–12
27. Atchison DA. Refractive errors induced by displace-ment of intraocular lenses within the pseudophakic eye.Optom Vis Sci 1989; 66:146–152
28. Atchison DA. Optical design of intraocular lenses. III.On-axis performance in the presence of lens displace-ment. Optom Vis Sci 1989; 66:671–681
29. Kozaki J, Takahashi F. Theoretical analysis of imagedefocus with intraocular lens decentration. J CataractRefract Surg 1995; 21:552–555
30. Korynta J, Bok J, Cendelin J, Michalova K. Computermodeling of visual impairment caused by intraocular lensmisalignment. J Cataract Refract Surg 1999; 25:100–105
31. Howland HC, Howland B. A subjective method for themeasurement of monochromatic aberrations of the eye.J Opt Soc Am 1977; 67:1508–1518
32. Packer M, Fine IH, Hoffman RS, Piers PA. Prospectiverandomized trial of an anterior surface modified prolateintraocular lens. J Refract Surg 2002; 18:692–696
—VOL 31, MARCH 2005
LABORATORY SCIENCE: EFFECT OF DECENTRATION ON IOL OPTICAL PERFORMANCE
33. Mester U, Dillinger P, Anterist N. Impact of a modifiedoptic design on visual function: clinical comparativestudy. J Cataract Refract Surg 2003; 29:652–660
34. Kershner RM. Retinal image contrast and functionalvisual performance with aspheric, silicone, and acrylicintraocular lenses; prospective evaluation. J CataractRefract Surg 2003; 29:1684–1694
35. Ginsburg AP. Next generation contrast sensitivitytesting. In: Rosenthal BP, Cole RG, eds, Functional
J CATARACT REFRACT SUR
Assessment of Low Vision. St Louis, MO, Mosby, 1996;77–88
36. Rynders M, Lidkea B, Chisolm W, Thibos LN. Statisti-cal distribution of foveal transverse chromatic aberra-tions, pupil centration, and angle w in a population ofyoung adult eyes. J Opt Soc Am A 1995; 12:2348–2357
37. McLellan JS, Marcos S, Prieto PM, Burns SA. Imperfectoptics may be the eye’s defense against chromatic blur.Nature 2002; 417:174–176
585G—VOL 31, MARCH 2005