optical sectioning of the cornea with a new confocal in vivo slit-scanning videomicroscope
TRANSCRIPT
Optical Sectioning of the Comea with a New Confocal In Vivo Slit--scanning Videomicroscope
Wolfgang Wiegand, MD, PhD, 1 Andreas A Thaer, PhD, 2 Peter Kroll, MD, 1
Otto-C. Geyer, MD,2 Alberto]. Garcia, MD3
Purpose: The purpose of this article is to introduce a newly developed confocal in vivo slit-scanning microscope for continuous recording and real-time imaging of the various corneal subsegments of the patient's eye with high microscopical resolution and adequate contrast.
Methods: One-dimensional confocal slit-scanning videomicroscopy of the human cornea was performed with an instrument mainly consisting of a scanning module, an image-intensifier video camera, a video monitor, and a synchronization unit for matching optical scan and video cycle with respect to frequency and phase. Light intensity or fluorescence intensity profiles through the cornea could be obtained by microphotometric recording of part of the imaging light. An immersion contact technique using an isotonic tear replacement liquid with thixotropic properties avoids any mechanical contact between the front lens of the microscope objective and the corneal surface.
Results: In normal human eyes, the corneal micromorphology could be made visible with satisfactory lateral and axial resolution and with good contrast. The separately focussed sections of the cornea showed the endothelial cells, the superficial, intermediary, and basal cells of the epithelium, as well as stromal keratocytes and nerves. Even in eyes with significant corneal opacities resulting from corneal edema, the endothelial pathology could be imaged with sufficient contrast.
Conclusion: The in vivo slit-scanning videomicroscopy offers real-time noninvasive and noncontact serial imaging of corneal subsegments with resolution and imaging contrast. Thus, an important step toward using confocal scanning microscopy for corneal diagnosis seems to be done. Ophthalmology 1995;102:568-575
Microstructural analysis is of growing importance for ophthalmologic diagnosis. This has become particularly
true for the anterior segment of the eye, namely for the cornea and its subsegments, because photorefractive keratectomy is performed worldwide. A significant barrier for the in vivo use of microstructural findings in the normal eye, in eyes with corneal diseases, and in eyes after photorefractive keratectomy, however, is the restricted optical resolution and magnification range of the slit lamp being sufficient for the macroscopic range only. Attachments to the slit lamp for biomicroscopy or special ophthalmomicroscopes, which have been used successfully for the examination of the corneal endothelium, do not meet the requirements for an in vivo microstructural analysis of the different corneal sections because light reflected and light scattered by structures above and below the focussing plane of the cornea impair the image contrast significantly. To satisfy these requirements, therefore,
Originally received: August 4, 1994. Revision accepted: December 31, 1994. 1 Department of Ophthalmology, University of Marburg, Germany. 2 Institute for Medical Visual Aid, Wetz1ar, Germany. 3 Ophthalmology Practice, Emden, Germany.
Presented in part as a poster at the American Academy of Ophthalmology Annual Meeting, Chicago, November 1993.
Supported by grants of Bundesministerium ftir Forschung und Technologie (BMFT), Bonn, Germany.
Dr. Andreas A. Thaer is consultant to the company Helmut Hund GmbH, Wetzlar, Germany.
Reprint requests to Wolfgang Wiegand, MD, PhD, Department of Ophthalmology, Philipps University, Robert-Koch-Str 4, D-35033 Marburg, Germany.
568
Wiegand et al · In Vivo Optical Sectioning of the Cornea
J/A
,.···-· ..
FIELD DIAPHRAGM
ILLUMINATED OBJECT AREA RESTRICTED BY
........
many attempts have been made in the last few years to develop confocal microscopes. 1.4-9
The purpose of this article is to introduce a newly developed confocal in vivo slit-scanning microscope for realtime serial imaging of the microstructures of the cornea, making use of the known advantages of this new lightmicroscopic principle with respect to optical resolution, image contrast, and optical sectioning capability.
Instrumentation and Methods
In contrast to conventional microscopy, where the illuminated object area is restricted by a field diaphragm only and the incident light thus covers a large area of the imaged tissue almost homogeneously, confocal microscopy is based on the principle that the incident light is focussed within a small region of the focal plane of the microscope objective leading to a steep axial light gradient in the tissue (Fig l ). This reduction of the illuminated object area min-
I :1
J/AL . . . . . . .
........ ........
SLIT· OR POINT .SHAPED CONFOCAL DIAPHRAGM
Figure 1. Axial light-intensity gradient 0/ A) in an axially extended object such as the cornea in case of ordinary (le&) and confocal (right) light microscopy (schematic).
imizes the influence of light on the image which is scattered by structures outside the focussed object plane. The minimizing effect is supported further by inserting an aperture in the plane of the image (i.e., in a confocal position with respect to the illuminating aperture). Thus, all light reaching the image plane outside of this aperture is prevented from influencing the image of the object and therefore prevented from reducing the image contrast. Both confocal diaphragm apertures scan equidirectionally across the imaged object area with identical speed and amplitude.
In principle, confocal microscopy can be performed either by two-dimensional spot scanning1- 8•10- 12 or by onedimensional slit scanning9•13•14 (Fig 2). In our instrument, the one-dimensional slit scanning instead of the two-dimensional spot scanning is used among other things because of its easy applicability in combination with video equipment. The optical contrast in one-dimensional scanning is only slightly lower than in two-dimensional scanning if the effective slit aperture is small enough.
Figure 2. Optical sectioning of a three-dimensional microscopic object (schematic) by two-dimensional spot· scanning (le&) and one-dimensional slit-scanning (right).
569
Ophthalmology Volume 102, Number 4, April1995
Figure 3. Optical arrangement for in vivo confocal slit-scanning microscopy and microphotometry of the cornea.
J/A
,, '• I 1 :I I : 1 I
CONFOCAL SCANNING SLITS LIGHTSOURCE ,' I I I . · .. _______ .. ': SYNCHRONIZED WITH A VIDEO IMAGE ;ICK-UP CYCLE ~
....... ~
/I I I
SPECTRAL FILTER ! \ FLUO:~~ENCE b'fm'm~ Z-SCAN (1 mm) t I I ~
--,t..:-J-.::-:-~----; ~-+---~- --~(?]~~'~'o ..__~--..,..-r I -- - · I \
PHOTOSENSOR 1 : - - -i- - - - _ t;:. : I t : • -, ---· :!f• : ~ x~ ~EAM SPLITTER
1 I / 50/50 I ' I
: ~ DICHROMATIC BEAMSPLITTER ~ FOR FLUORESCENCE
,-,-----. <4t-r--------~----------
IMAGE INTENSIFIER VIDEO CAMERA
D VIDEO MONITOR
The instrument (Fig 3) mainly consists of a conventional light source (halogen lamp 12 VI 100 W /DC), a scanning module for lateral confocal scanning containing
the illuminating and imaging optics with the respective scanning slits combined in a fixed distance to each other in an electromagnetically driven double-slit arrangement, an image-intensifier video camera, a video monitor, and a synchronization unit for frequency and phase-sensitive matching optical scan and video cycle. The scanning frequency is 25 Hz. A microprocessor-controlled stepping motor can move the focus of the microscope optics in the direction of the optical axis with a velocity of I mm/
---CORNEAL EPITHELIUM
TEAR LIQUID
Figure 4. Friction-free optical contact between the front lens of the microscope objective of the confocal scanning system and the corneal surface by use of a thixotropic gel (VIDISIC-Gel, Dr. Mann Pharma, Berlin).
570
Figure 5. Videomicrograph of the rabbit corneal endothelium without (left) and with (right) synchronization between optical slit-scan and videoimage cycle.
Wiegand et al · In Vivo Optical Sectioning of the Cornea
Figure 6. Normal human corneal endothelium (objective, 25/0.60).
second and thus scan the whole cornea within approximately 0.5 second, the exact value depending on the actual thickness of the cornea. Thus, a sequence of lateral sections at different depths of the cornea (z-scan) can be recorded and stored on a videotape. In addition, the z-scan permits the photometric recording of light-intensity profiles through the whole cornea (or parts of it) without lateral scanning but with an axial resolution given by the focal depth of the objective used. For fluorescence measurements, spectral filters (excitation and barrier filter) and a high-pressure mercury lamp ( l 00 W /DC) instead of the halogen lamp can be added.
For eliminating image blurring due to eye movements and internal (e.g., caused by the scanner) or external ("seismic") vibrations, each half of the full video image is generated by a single scan only. Its duration is 20 mseconds (one of the half images) or 40 mseconds (full image), the scanning frequency being 25 cycles in both cases. The time for imaging a microstructural detail still resolved by the video camera is given by its individual exposure time on the photocathode and thus by the width of the slit
Figure 7. Human corneal endothelium with cellular polymegatism (objective, 25/0.60).
Figure 8. Human corneal endothelium with cellular precipitate on its rear side due to inflammation (objective, 25/0.60).
image and its speed across the photocathode. The normal slit width used corresponds with l/30 of the scanning amplitude or approximately 20 lines of the 600 lines of the video camera. Thus, the exposure time of each resolved microsiructure in each slit-shaped area is not more than 20 mseconds/30 = 0.66 msecond! For using the full video line number, an interline CCD camera is used. A onestep image pre-amplifier with optical fiber coupling to the ceo camera reduces the necessary illumination intensity for keeping the light exposure of the cornea as low as possible. An accurate synchronization between the image pick-up cycle of the video cameraand the optical slit scan is obtained by using the V -signal of the video camera for triggering the scan control unit.
In addition to the microscopic imaging of corneal structures, corneal profiles based on light scattering or on fluorescence can be recorded photometrically. Part of the imaging light (20%) is therefore used for a microphotometric registration of scattered light or fluorescence intensity profiles throughout the cornea (see Fig 3) by means
Figure 9. Human corneal endothelium shows an advanced "cornea guttata" (objective, 25/0.60).
571
Ophthalmology Volume 102, Number 4, April1995
Figure 10. Human cornea; stroma with keratocytic nuclei (objective, 25/ 0.60).
of the z-scan. An immersion contact technique (Fig 4) using a commercial isotonic tear replacement liquid (VIDISIC-Gel, Dr. Mann Pharma, Berlin, Germany) with a refractive index of J.to = 1.350 provides an optically almost homogeneous immersion between the front lens of the objective and the anterior chamber. It also avoids any mechanical contact and shearing forces between the microscope objective and the corneal surface. The low viscosity and thixotropic property (low refractive index difference at the interface gel/corneal surface) of the substance is caused by a 0.2% polyacrylic acid content with a molecular weight of approximately 4 million. Thus, the immersion drop behaves like a ball with very low internal convection.
A 25/0.60 or a 40/0.75 water-immersion objective is used as standard lens ofthe instrument. For highest optical resolution, a 50/1.00 water-immersion objective can be used. The use of the latter is restricted to imaging the anterior corneal subsegments for reason of its smaller freeworking distance. The effective lateral optical resolution
Figure 11. Human cornea; nerve branching in the anterior stromal region (objective, 25/0.60).
572
Figure 12. Human cornea; nerve fibers without myelin close to Bowman membrane (objective, 25/0.60).
of the system is between 1 and 2 J.tm, depending on the objective.
Application and Results
Before using the system for in vivo real-time examinations of human corneas, it has been tested thoroughly in human volunteers and in rabbits. Before examination, an informed consent was signed by the patients. At the beginning of the examination, a drop of topical anesthetic was placed into the patient's eye and the eye then was positioned in front of the objective lens of the instrument. All investigations described in this article were performed with the immersion contact technique and either the 25/0.60 water immersion objectives or the 50/1.00 water immersion objective. Due to the immersion fluid on top of the objective lens, the corneal epithelium was not touched during the investigation, and the cornea thus was not applanated. From all patients, initially a z-scan of the whole
Figure 13. Human corneal epithelium with cells before desquamation (objective, 25/0.60).
Wiegand et al · In Vivo Optical Sectioning of the Cornea
Figure 14. Human corneal stroma with keratocytic nuclei surrounded by cytoplasmic contours usually becoming visible in case of stromal swelling (objective, 50/1.00).
cornea was taken and the interesting corneal sections then were imaged subsequently. All images were stored on videotapes, and Figures 5 through I 7 were taken from the video screen.
It generally can be noticed that all photographs of strongly reflecting sections show the best imaging contrast of the structures in the center of the video screen, whereas the cell contours vanish close to the edges of the photographs. This is due to the normal curvature of the cornea ~hich is not applanated by a contact element. Figure 5 Illustrates the effect of synchronization between optical scan and video signal in the rabbit cornea. Figures 6 to 9 show video micrographs of the human corneal endothelium taken with a 25/0.60 water-immersion objective. The norm~l endothelium (Fig 6) as well as endothelial polymegatism due to extended contact lens wearing (Fig 7) and cellular precipitates on the rear side of the endotheli~~ in a patient with anterior uveitis (Fig 8) are clearly visible. Due to the confocal principle, the corneal endothelium could be imaged even in patients with significant corneal opacities resulting from corneal edema or corneal dystrophies. An example of severe endothelial pathology ("cornea guttata") is presented in Figure 9.
Figure 15. Human corneal stroma; structured keratocytic nuclei with notches (objective, 50/1.00).
Figure 16. Human cornea; basal cell layer of the epithelium (objective 50/1.00). '
Within the co_rneal strom~ keratocytic nuclei (Fig 10), nerve branches m the an tenor stromal regions (Fig II) and close to Bowman membrane (Fig 12) can be delineated us_ing a 25/0.60 water-immersion objective also. In sup~rficial corneal epithelium, the cells before desquamatiOn show a much brighter reflection than the surrounding cells (Fig 13). . The i~p~ove~ lateral resolution and selectivity of op
tical sectwnmg m case of a 50/1 .00 water-immersion objective is _demon~trated in Figures 14 to 17. They show keratocytic nuclei surrounded by cytoplasmatic contours (Fig 14) or structured keratocytic nuclei with notches (Fig 15) .. Some other corneal subsegments like the basal epith~hal cell layer (Fig 16) or the intermediate cell layer (wmg cells) of the epithelium (Fig 17) also become visible at a higher numeric aperture only.
None of the video micrographs shown in the figures are_decreased in contrast or blurred due to eye movement which proves the possibility of real-time in vivo imaging of the human cornea at the cellular level. According to
Figure 17. Human cornea; intermediate cell layer of the epithelium (objective, 50/1.00).
573
Ophthalmology Volume 102, Number 4, April1995
INTERFACE I I
TEARFILM I EPITHELIUM BASAL LAMINA
Figure 18. Microphotometric light intensity profile by use of confocal z-scanning throughout the human epithelium recorded in vivo (objective, 50/1.00).
IMMERSION LIQUID
I l I J
...
-
I 1.. T
50J.Im -;;
I I J
... -~ ~I.. l1~~t r•
'•I
-
llljll r.\j II: A~~r
STROMA ~
Ll .it.ra-' ~ .. "' lr.1 .,.1.
J
our experiences, the short exposure time reduces the portion of images with blurred structural details down to or below 5% (i.e., to approximately 1 frame of 25 frames/ second). This is true for image blurring due to both lateral or axial eye movement. Thus, usually a ratio of video images unaffected by eye motions of more than 90% is gained even for the highest magnification and optical resolution applied. Evidence for a sufficient selectivity of the optical sectioning also is given by the light-intensity profile through the human corneal epithelium (Fig 18) recorded in vivo with a 50/1.00 water-immersion objective and zscan movement of the focus.
Discussion and Conclusion
During the last two decades, confocal scanning light microscopy1- 3 has established the basis for an "optical sectioning" of translucent microscopic objects whose thickness exceeds the depth of focus of the microscope optics by many times. Sequential imaging of small pointshaped or slit-shaped areas of the focussed object planes and the confocal optical arrangement, both led to an excellent image contrast. During the last 10 years, such optical arrangements have been used successfully for a systematic investigation of the corneal microstructure of enucleated eyes and excised corneal tissue.4- 6•8•10- 12 Although confocal scanning microscopy also has been applied to in vivo examination of the cornea,6•9•13•14 its introduction into corneal diagnosis is hampered still by image blurring due to eye movements, unacceptable light intensity of the scanning light beam, problems encountered in optically interfacing microscope lens and eye sur-
574
face, and by problems obtaining laterally or axially arranged image series for direct observation, image storing, and image documentation. Some contributions to the solution of these problems have been discussed previously elsewhere. 15- 17
Essential technical features of our device are the confocal one-dimensional scanning by use of slit-shaped diaphragms in the illuminating and imaging light paths coinciding in the focussed object plane, the use of a sensitive image-intensifier camera, and use of an exact frequency and phase-dependent synchronization of optical scanning and video recording. This allows the imaging of lateral optical sections of the cornea with very high resolution and results in sharp images unaffected by eye motion. Due to the video .equipment, the images can be recorded, stored, and repeatedly displayed. The use of an immersion contact technique greatly diminishes the infection risk compared with the use of contact elements. The device has been very effective and reliable in clinical use.
The real-time in vivo video micrographs shown above prove that an important step toward the use of confocal scanning microscopy for corneal diagnosis has been done. Thus, in vivo corneal "histology" and "histopathology" seem to be feasible not only for diagnostic purposes in humans but also for noninvasive tests in animals. The images can be achieved very quickly (with 25 exposures per second) and without exposing patients or test persons to stress exceeding that of a slit-lamp examination. Concerning optical resolution and contrast, the video micrographs in Figures 6 to 1 7 are comparable approximately with selectively exposed in vivo confocal slit-scanning photomicrographs taken by using a shutter at 1/60-second exposure time, which were published recently.9 However, due to a lower number of pixels per area made available
Wiegand et al · In Vivo Optical Sectioning of the Cornea
by the video cathode, our video micrographs show a more pronounced grain of the background and a field restriction compared with photomicrographs exposed on 35-mm film. The high image sampling rate, on the other hand, compensates for this field reduction.
The possibility of studying the corneal microstructure dynamically without any preparational or staining artifacts adds further important information to corneal microanatomy and corneal· physiology. Processes such as wound healing as a function of time thus can be investigated. The new methods of refractive surgery by anterior or intrastromal laser ablation or by keratomileusis in situ doubtlessly will profit by this possibility. In addition to the microstructural information, in vivo measurements of the scattered light intensity and permeation measurements with nontoxic fluorescent stains can be performed, which give further information on the actual functional state of the cornea and its subsegments.
References
1. Petmn M, Hadravsky M, Egger MD, Galambos R. Tandemscanning reflected-light microscopy. J Opt Soc Am 1968;58: 661-4.
2. Wilson T, Sheppard C. Theory and Practice of Scanning Optical Microscopy. London: Academic Press, 1984.
3. Wilson T. Confocal light microscopy. Ann NY Acad Sci 1986;483:416-27.
4. Lemp MA, Dilly PN, Boyde A. Tandem-scanning (confocal) microscopy of the full-thickness cornea. Cornea 1985-86;4: 205-9.
5. Dilly PN. Tandem scanning reflected light microscopy of the cornea. Scanning 1988;10:153-6.
6. Cavanagh HD, Jester JV, Essepian J, eta!. Confocal microscopy of the living eye. CLAO J 1990;16:65-73.
7. Cavanagh HD, Petroll WM, Alizadeh H, eta!. Clinical and diagnostic use of in vivo confocal microscopy in patients with corneal disease. Ophthalmology 1993;100:1444-54.
8. Masters BR, Paddock S. In vitro confocal imaging of the rabbit cornea. J Microsc 1990;158:267-74.
9. Koester CJ, Auran JD, Rosskothen HD, eta!. Clinical microscopy of the cornea utilizing optical sectioning and a high-numerical-aperture objective. J Opt Soc Am [A) 1993;10:1670-9.
10. Jester JV, Cavanagh HD, Lemp MA. Confocal microscopic imaging of the living eye with tandem scanning confocal microscopy. In: Masters BR, ed. Noninvasive Diagnostic Techniques in Ophthalmology. New York: Springer-Verlag, 1990; chap. 11.
II. Masters BR, Paddock SW. Three-dimensional reconstruction of the rabbit cornea by confocal scanning optical microscopy and volume rendering. Appl Optics 1990;29:3816-22.
12. Masters BR, Kino GS. Confocal microscopy of the eye. In: Masters BR, ed. Noninvasive Diagnostic Techniques in Ophthalmology. New York: Springer-Verlag, 1990; chap. 10.
13. Koester CJ, Roberts CW, Donn A, Hoefle FB. Wide field specular microscopy. Clinical and research applications. Ophthalmology 1980;87:849-60.
14. Koester CJ, Roberts CW. Wide-field specular microscopy. In: Masters BR, ed. Noninvasive Diagnostic Techniques in Ophthalmology. New York: Springer-Verlag, 1990; chap. 7.
15. Masters BR, Thaer AA. Real-time scanning slit confocal microscopy of the in vivo human cornea. Appl Optics 1994;33:695-70 1.
16. Masters BR, Thaer AA. Confocal microscopy of the in vivo cornea. In: Ophthalmic and Visual Optics: Summaries of Papers Presented at the Ophthalmic & Visual Optics Topical Meeting, February 19-20, 1993: Noninvasive Assessment of the Visual System Meeting, February 21-23, 1993. Washington, DC: Optical Society of America, 1993;133-6 (Technical digest series 1993;3).
17. Wiegand W, Thaer AA. In-vivo-Hornhautmikroskopie mit einem neuen konfokalen Spalt-Scanning Video-System. In: Robert YCA, Gloor B, Hartmann Ch, Rochels R, eds. 7. KongreJ3 der Deutschsprachigen Gesellschaft fiir Intraokularlinsen-Implantation. Berlin: Springer-Verlag, 1993;484-8.
575