optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source...

12
Optimizing the accuracy of a helical diode array dosimeter: A comprehensive calibration methodology coupled with a novel virtual inclinometer Jakub Kozelka Sun Nuclear Corp., Melbourne, Florida 32940 Joshua Robinson Department of Physics, University of South Florida, Tampa, Florida 33612 Benjamin Nelms Canis Lupus LLC, Sauk County, Wisconsin 53561 Geoffrey Zhang Division of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612 Dennis Savitskij a) Sun Nuclear Corp., Melbourne, Florida 32940 Vladimir Feygelman Division of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612 (Received 6 April 2011; revised 6 July 2011; accepted for publication 15 July 2011; published 11 August 2011) Purpose: The goal of any dosimeter is to be as accurate as possible when measuring absolute dose to compare with calculated dose. This limits the uncertainties associated with the dosimeter itself and allows the task of dose QA to focus on detecting errors in the treatment planning (TPS) and/or delivery systems. This work introduces enhancements to the measurement accuracy of a 3D dosim- eter comprised of a helical plane of diodes in a volumetric phantom. Methods: We describe the methods and derivations of new corrections that account for repetition rate dependence, intrinsic relative sensitivity per diode, field size dependence based on the dynamic field size determination, and positional correction. Required and described is an accurate “virtual inclinometer” algorithm. The system allows for calibrating the array directly against an ion cham- ber signal collected with high angular resolution. These enhancements are quantitatively validated using several strategies including ion chamber measurements taken using a “blank” plastic shell mimicking the actual phantom, and comparison to high resolution dose calculations for a variety of fields: static, simple arcs, and VMAT. A number of sophisticated treatment planning algorithms were benchmarked against ion chamber measurements for their ability to handle a large air cavity in the phantom. Results: Each calibration correction is quantified and presented vs its independent variable(s). The virtual inclinometer is validated by direct comparison to the gantry angle vs time data from machine log files. The effects of the calibration are quantified and improvements are seen in the dose agreement with the ion chamber reference measurements and with the TPS calculations. These improved agreements are a result of removing prior limitations and assumptions in the calibration methodology. Average gamma analysis passing rates for VMAT plans based on the AAPM TG-119 report are 98.4 and 93.3% for the 3%/3 mm and 2%/2 mm dose-error/distance to agreement thresh- old criteria, respectively, with the global dose-error normalization. With the local dose-error nor- malization, the average passing rates are reduced to 94.6 and 85.7% for the 3%/3 mm and 2%/2 mm criteria, respectively. Some algorithms in the convolution/superposition family are not suffi- ciently accurate in predicting the exit dose in the presence of a 15 cm diameter air cavity. Conclusions: Introduction of the improved calibration methodology, enabled by a robust virtual incli- nometer algorithm, improves the accuracy of the dosimeter’s absolute dose measurements. With our treatment planning and delivery chain, gamma analysis passing rates for the VMAT plans based on the AAPM TG-119 report are expected to be above 91% and average at about 95% level for c(3%/3 mm) with the local dose-error normalization. This stringent comparison methodology is more indica- tive of the true VMAT system commissioning accuracy compared to the often quoted dose-error nor- malization to a single high value. V C 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3622823] Key words: VMAT QA, rotational therapy, dose distribution comparison, three-dimensional dosimetry, diode array calibration 5021 Med. Phys. 38 (9), September 2011 0094-2405/2011/38(9)/5021/12/$30.00 V C 2011 Am. Assoc. Phys. Med. 5021

Upload: others

Post on 25-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

Optimizing the accuracy of a helical diode array dosimeter: A comprehensivecalibration methodology coupled with a novel virtual inclinometer

Jakub KozelkaSun Nuclear Corp., Melbourne, Florida 32940

Joshua RobinsonDepartment of Physics, University of South Florida, Tampa, Florida 33612

Benjamin NelmsCanis Lupus LLC, Sauk County, Wisconsin 53561

Geoffrey ZhangDivision of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612

Dennis Savitskija)

Sun Nuclear Corp., Melbourne, Florida 32940

Vladimir FeygelmanDivision of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612

(Received 6 April 2011; revised 6 July 2011; accepted for publication 15 July 2011; published 11

August 2011)

Purpose: The goal of any dosimeter is to be as accurate as possible when measuring absolute dose

to compare with calculated dose. This limits the uncertainties associated with the dosimeter itself

and allows the task of dose QA to focus on detecting errors in the treatment planning (TPS) and/or

delivery systems. This work introduces enhancements to the measurement accuracy of a 3D dosim-

eter comprised of a helical plane of diodes in a volumetric phantom.

Methods: We describe the methods and derivations of new corrections that account for repetition

rate dependence, intrinsic relative sensitivity per diode, field size dependence based on the dynamic

field size determination, and positional correction. Required and described is an accurate “virtual

inclinometer” algorithm. The system allows for calibrating the array directly against an ion cham-

ber signal collected with high angular resolution. These enhancements are quantitatively validated

using several strategies including ion chamber measurements taken using a “blank” plastic shell

mimicking the actual phantom, and comparison to high resolution dose calculations for a variety of

fields: static, simple arcs, and VMAT. A number of sophisticated treatment planning algorithms

were benchmarked against ion chamber measurements for their ability to handle a large air cavity

in the phantom.

Results: Each calibration correction is quantified and presented vs its independent variable(s). The

virtual inclinometer is validated by direct comparison to the gantry angle vs time data from

machine log files. The effects of the calibration are quantified and improvements are seen in the

dose agreement with the ion chamber reference measurements and with the TPS calculations. These

improved agreements are a result of removing prior limitations and assumptions in the calibration

methodology. Average gamma analysis passing rates for VMAT plans based on the AAPM TG-119

report are 98.4 and 93.3% for the 3%/3 mm and 2%/2 mm dose-error/distance to agreement thresh-

old criteria, respectively, with the global dose-error normalization. With the local dose-error nor-

malization, the average passing rates are reduced to 94.6 and 85.7% for the 3%/3 mm and 2%/2

mm criteria, respectively. Some algorithms in the convolution/superposition family are not suffi-

ciently accurate in predicting the exit dose in the presence of a 15 cm diameter air cavity.

Conclusions: Introduction of the improved calibration methodology, enabled by a robust virtual incli-

nometer algorithm, improves the accuracy of the dosimeter’s absolute dose measurements. With our

treatment planning and delivery chain, gamma analysis passing rates for the VMAT plans based on

the AAPM TG-119 report are expected to be above 91% and average at about 95% level for c(3%/3

mm) with the local dose-error normalization. This stringent comparison methodology is more indica-

tive of the true VMAT system commissioning accuracy compared to the often quoted dose-error nor-

malization to a single high value. VC 2011 American Association of Physicists in Medicine.

[DOI: 10.1118/1.3622823]

Key words: VMAT QA, rotational therapy, dose distribution comparison, three-dimensional

dosimetry, diode array calibration

5021 Med. Phys. 38 (9), September 2011 0094-2405/2011/38(9)/5021/12/$30.00 VC 2011 Am. Assoc. Phys. Med. 5021

Page 2: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

I. INTRODUCTION

The goal of any dosimeter is to be as accurate as possible in

reporting absolute dose that is compared with the calcula-

tions, thus allowing the physicist to focus on detecting and

mitigating errors in the TPS and/or the delivery system as

opposed to the uncertainties associated with the dosimeter

itself. By this rationale, if a dosimeter has better accuracy, it

would cease to be a major source of uncertainty in dose QA.

ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

dosimeter intended primarily for commissioning and pre-

treatment dose verification of IMRT and volumetric modu-

lated arc treatments (VMATs). The design of the device,

particularly the geometry of its detector array, is a departure

from other commercially available array dosimeters, which

typically consist of one1–4 or two5 flat detector planes. ARC-

CHECK has the detectors arranged on a helical grid placed on

an approximately cylindrical surface. Such an arrangement

requires thorough investigation of the array calibration and

performance both in terms of pure geometrical effects and

physical propertied of the device.

Some fundamental differences between the helical grid

and a biplanar array were reported by Feygelman et al.6

They are primarily geometrical in nature and related to the

hypersensitivity of the peripherally positioned diodes to the

small changes in arc parameters. Two earlier papers

described methods for array calibration and application of

the correction factors for the ARCCHECK,7,8 but in both of

these papers a prototype device was studied. More impor-

tantly, certain simplifying assumptions were made that could

potentially affect the dosimetric results. In particular, field

size dependence of the diode sensitivity was not considered,

and the calibration methodology did not consistently rely on

the direct comparison with the ion chamber. While the nu-

merical effect of those assumptions may be well hidden

within the currently acceptable uncertainty of IMRT meas-

urements,9 it is highly desirable to develop a “reading to

dose” calibration and correction formalism as accurate and

representative of actual physical processes as possible.

Letourneau et al.7 and Yan et al.8 made an important ob-

servation regarding the relative calibration of the array, in

that, the position of the diode with respect to the isocenter is

of importance. However, it was assumed that the ARCCHECK,

which employs the same diode detectors as its planar prede-

cessor, MAPCHECK, should exhibit similar negligible field size

dependence.1,2 This assumption was later refuted6 and, as it

turns out, the field size dependence has to be included into

the calibration formalism. Yan et al.8 also pointed out that

Letourneau et al.7 defined a diode response as a function of

gantry angle by finding the ratio of the individual diode

response to the mean of the response curve. A more tradi-

tional approach favored by Yan et al.8 would be normalizing

the diode response to the reading from an independent detec-

tor (ion chamber). However, the directional response

approach of Yan et al.8 is based on the calculations in a ho-

mogeneous phantom rather than the direct measurements. In

this paper, we set out to develop a relative calibration

method based on direct comparison with the ion chamber

and applicable to both hollow and homogeneous configura-

tions of the phantom.

Neither group has evaluated the possible accelerator dose

rate dependence in terms of monitor units per minute (repeti-

tion rate), previously reported for the MAPCHECK.7 This is par-

ticularly important for VMAT measurements, where the

repetition rate can fluctuate widely. In this work, we deter-

mine and incorporate a repetition rate correction.

Since both papers7,8 emphasize the importance of correct-

ing for the diode relative response as a function of gantry

angle, both groups reported means for determination of that

angle during beam delivery. Letourneau et al.7 employed an

independent physical inclinometer attached to the gantry. This

requires extra set-up time in the room, as well as a radiation-

hard, fast commercial inclinometer which in itself may be a

challenge. Yan et al.8 took the alternative path of developing

a “virtual” inclinometer based on the beam outline projection

onto the phantom. In that paper, the virtual inclinometer was

successfully tested with fixed-gantry IMRT beams. While the

concept of virtual inclinometer is advantageous, our tests of

the algorithm in the ARCCHECK software with a variety of arcs

indicated suboptimal performance. In this paper, we describe

a much more precise virtual inclinometer algorithm.

By coincidence, all three papers describing the agreement

between the calculated and measured dose in ARCCHECK

(Refs. 6–8) used the same treatment planning system

(TPS)—PINNACLE (Philips Radiation Oncology Systems,

Fitchburg, WI). As the device is intended for use with a vari-

ety of different TPS, it is instructive to at least briefly evalu-

ate how accurate different algorithms, even from the family

of sophisticated ones presumed to handle inhomogeneities

with acceptable accuracy,10 are in the presence of a sizable

air cavity in the middle of the ARCCHECK phantom.

II. MATERIALS AND METHODS

II.A. General

II.A.1 ARCCHECK description

Prior publications describe the prototype7,8 and commer-

cial6 versions of the ARCCHECK. In brief, the current version

features a 2D array of 1386 point detectors mounted on 22

printed circuit board (PCB) sections approximating a cylin-

drical surface at a distance of 10.4 cm from the center of a

doughnut-shaped phantom. There are three diode columns

per width of every board (Fig. 1). The phantom has an outer

diameter of 26.6 cm and an inner cavity diameter of 15.1

cm. The buildup and backscatter thickness is 2.9 cm

PMMA. The diodes form a helical pattern. They are posi-

tioned 1 cm apart along both the cylindrical length and cir-

cumference. This detector pattern is designed to accomplish

two goals: first, to reduce the rotational response depend-

ence by making the detector array nearly cylindrically sym-

metrical; and second, to increase the apparent detector

density in the beam’s eye view (BEV) by shifting the exit

diodes with respect to the entrance ones. An optional

PMMA plug is used to eliminate the central cavity inhomo-

geneity when desired.

5022 Kozelka et al.: Optimizing the accuracy of a helical diode array 5022

Medical Physics, Vol. 38, No. 9, September 2011

Page 3: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

II.A.2. ARCCHECK calibration measurements

All calibration measurements, except for the absolute cali-

bration, were performed on a Varian TrueBeam linear accelera-

tor (Varian Medical Systems, Palo Alto, CA) with four photon

energies (6, 10, 15 MV, and 6 MV flattening filter free). This

selection of energies was useful for measuring parameters

expected to be energy-dependent, most notably field size and

positional dependencies. While the multienergy data are neces-

sary for clinical implementation of the dosimeter, only the 6

MV results are reported here for conciseness. The linac can

produce dose rates for a 6 MV beam from 5 to 600 MU/min.

II.A.3. Treatment planning parameters

All plans used in this study were calculated on a 2 mm dose

grid. All arcs were calculated at 2� angular intervals. The primary

TPS was Philips PINNACLE v. 9.0. Additional studies were also

performed with CMS XiO v. 4.5 and Monaco v. 3.0 (both from

Elekta CMS Software, Maryland Heights, MO) and Varian

Eclipse v. 10.0. A bulk relative density of 1.147 (Ref. 11) was

used to represent the PMMA phantom. All field sizes are reported

as X�Y, as defined by the IEC beam coordinate system.

II.A.4. Evaluation measurements

All verification measurements were performed on a dif-

ferent Varian linac—a Trilogy with a 120 leaf Millennium

MLC. Unless stated otherwise, an ARCCHECK phantom with

an air cavity was used for verification measurements. ARC-

CHECK absolute calibration was performed on this accelera-

tor. This follows the intended procedure where all relative

calibration parameters are configured at the factory, and only

absolute calibration is performed in the user’s beam.

II.B. Benchmarking TPS accuracy with the ionchamber in a PMMA phantom

We have manufactured a homogeneous PMMA shell mim-

icking the dimensions of the ARCCHECK phantom. Longitudinal

holes were drilled to place an A1SL 0.06 cm3 ion chamber

(Standard Imaging Inc., Middleton WI) at some diode loca-

tions, approximately 15� apart. The center of the chamber

coincided with the diode’s stated effective point of measure-

ment (10.45 cm from the center). The shell was centered on

the room lasers so that the chamber center was in the isocenter

plane. The ratio of chamber reading at 180� and 0� (exit to en-

trance dose ratio) per 200 MU was recorded for different

square filed sizes. These ratios were measured with and with-

out the PMMA plug in place. The exit to entrance dose ratios

were then calculated with Pinnacle Adaptive Convolve, XiO

Convolution and Multigrid Superposition, Eclipse Analytical

Anisotropic, and Monaco Monte Carlo algorithms. For the lat-

ter, the number of histories corresponded to the statistical var-

iance of 1.5%. The difference in measured and calculated

ratios was tabulated.

The same shell was used to generate circumferential rela-

tive dose profiles. The PMMA shell with inserted ion chamber

was positioned on the specially constructed “rotisserie” device

equipped with a stepping motor and irradiated under the con-

tinuous beam. The reference ion chamber attached to the ac-

celerator head was used to correct for the dose rate

fluctuations. Both chambers were connected to a dual-channel

electrometer with a 20 Hz sampling rate (PC Electrometer,

Sun Nuclear) and the ratio of ionization readings vs time was

recorded on a portable computer connected to the electrome-

ter. The readings from two full revolutions were averaged.

These high angular resolution (0.5�) data were validated and

augmented by integrating ion chamber measurements at 13

points along the array circumference. Calculated profiles

along the array surface were extracted on a 1 mm interpolated

grid by the ARCCHECK software (v. 5.02) from the DICOM RT

dose files transferred from the TPS.

II.C. Calibration formalism and procedures

II.C.1. General calibration equation

Expanding upon the methods described previously for the

ARCCHECK prototype,7,8 the equation relating the absolute

dose to the raw reading of a diode in the primary beam, at an

arbitrary ith position in the phantom, is

Di ¼ RiNrefCTCRRCrel

i CFSi Cpos

i ; (1)

where

Ri (reading) is the ith diode raw reading after background

subtraction.

Nref (Gy/Reading) is the absolute dose calibration factor

for a normalization (reference) diode in the array, at a known

accelerator dose rate (preferably 300 MU/min). The normal-

ization diode is the one close to the central axis, on top of

the phantom in its normal measurement orientation. By

choosing the calibration dose rate, we effectively assign the

value of unity to the repetition rate correction (described

below) for that rate. Three hundred monitor units per minute

correspond approximately to the midrange of the repetition

rate correction values. As a result, the error is minimized

even if the correction is not applied. The absolute calibration

factor is best determined by calculating the dose to the

FIG. 1. The schematic cross-section of the ARCCHECK diode array demonstrat-

ing diode triplets positioned on 22 printed circuit boards (PCBs). Note that

lateral diodes on each PCB are not facing the center of the phantom.

5023 Kozelka et al.: Optimizing the accuracy of a helical diode array 5023

Medical Physics, Vol. 38, No. 9, September 2011

Page 4: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

reference detector with a TPS.6,12 Although, in principle, the

reference dose could be determined with a calibrated ion

chamber replacing the diode, absolute dose measurements in

PMMA with a chamber calibrated to dose to water are not

straightforward.

CT is a phantom temperature correction factor accounting

for the possible difference between the phantom temperature

at the time of calibration and measurement. This correction

is relatively small and is obviated if a cross-calibration read-

ing is taken prior to measurements. Such cross-calibration

procedure typically consists of exposing the dosimeter to an

arrangement of simple open beams and finding a correction

factor minimizing the difference between the measured and

calculated doses for a number of centrally located diodes.6

The last four corrections will be discussed in detail,

including their measurement conditions and numerical appli-

cation method, following their definitions:

CRR is an array correction for the accelerator pulse repetition

rates.

Creli is the diodes intrinsic sensitivity correction relative to

the reference diode in the array.

CFSi is the field size correction.

Cposi is the diode position correction in the phantom with

respect to the radiation beam.

II.D. Repetition rate correction CRR

CRR accounts for the change in diode response with the

change in the accelerator repetition rate (monitor unit per mi-

nute) and is applied uniformly to every diode in this work,

since at any given time all the diodes are exposed to the same

repetition rate. Possible variations of this correction factor

based on the diode position are discussed in detail in Secs. III

and IV. The repetition rate effect was originally reported for

the MAPCHECK array that uses the same diodes as the ARC-

CHECK.2 The correction was measured in this study by examin-

ing the average response ratio of the diodes exposed in a

40� 25 cm2 field to the constant number of monitor units at

different repetition rates (40 MU for the repetition rates of 5

and 10 MU/min, 100 MU for 20–600 MU/min). To account

for the possible accelerator output variation with repetition

rate, the diode reading was normalized by the simultaneously

acquired integrated reference Farmer chamber signal. The

change in the ion chamber collection efficiency was negligible

(<0.1%) as measured with a standard two-voltage tech-

nique.13 The measurements were done with and without a

PMMA plug. The results were analyzed in two ways: (i) for

all 1386 diodes in aggregate and (ii) for two separate groups,

each limited to 254 detectors situated within 65 cm from the

Y axis along the beam entrance and exit points.

To address a possibility that repetition rate correction

might differ between different accelerator designs from the

same manufacturer, or between different manufacturers, a

comparison study was made. The additional accelerators

included Varian Trilogy and Siemens Oncor (Siemens Medi-

cal Solutions USA, Inc., Malvern, PA). Published data2 for

the MAPCHECK with an Elekta SL20 accelerator was also

included in comparison.

To apply the repetition rate correction, the accelerator

repetition rates at the time of calibration and measurement

must be known. In this work we used a simplified approach

based on the average repetition rate for dynamic (open arcs

and VMAT) deliveries. The average repetition rate was

defined as the total monitor units divided by the total treat-

ment time.

II.D.1. Intrinsic relative sensitivity Cirel

Cirel accounts for the intrinsic sensitivity of an arbitrary

diode relative to the reference detector in the array. For the

two-dimensional planar arrays, these corrections are typi-

cally determined by a wide field calibration technique

employing a series of measurements with array shifts and

rotations, followed by a solution of a system of linear equa-

tions.14 The ARCCHECK detector geometry and the rotisserie

apparatus are conducive to separating this two-dimensional

problem into a sequence of two sequential one-dimensional

steps. First, the phantom was translated longitudinally in 1

cm increments under a relatively narrow beam (aperture

length Y¼ 5 cm). An integrated reading for a set number of

monitor units (100) was collected for each of the diodes

aligned parallel to the Y axis on top of the unit (“top”

diodes), as each detector was positioned sequentially in the

center of the beam. This direct procedure yields a diode with

known intrinsic sensitivity relative to the normalization

diode in each of the quasiaxial planes defining a single de-

tector “ring” of 66 diodes. Strictly speaking, the helical de-

tector pattern results in no two diodes situated exactly in any

single axial plane, but the longitudinal displacements for a

single ring are neglected in the wide beam.7 Then, the phan-

tom was rotated on a rotisserie under the continuous wide

radiation beam (27� 25 cm2) with a 10 s period. The data

from two revolutions were averaged. An external reference

diode attached to the gantry head was used to account for the

small variations in the accelerator dose rate. ARCCHECK has a

special input connector for this reference diode, which

allows synchronizing the readings of the internal and exter-

nal detectors. The ratio of the reading of any diode in the

ring to the one previously calibrated in the translational step,

when occupying the same position in the beam, relates the

intrinsic sensitivity of every diode in the ring to a single

“top” diode. The “top” diodes, in turn, were calibrated

against the normalization diode in the translational step.

Thus the Cirel value for each diode in the array in relation to

the normalization diode can be easily established.

II.D.2. Field size dependence CiFS

CiFS accounts for the changes in diode response relative to

the energy-independent detector (ion chamber) with variation

in field size.15,16 Field size correction is dependent on the

position (i) of the diode relative to the beam, as the scatter

conditions change around the phantom for any given beam ge-

ometry. For the same reason, the field size correction can

change when the central plug is added. CiFS is defined as

5024 Kozelka et al.: Optimizing the accuracy of a helical diode array 5024

Medical Physics, Vol. 38, No. 9, September 2011

Page 5: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

CFSi ¼

RFSi

�IFSi

R10�10i

�I10�10i

; (2)

where RiFS and Ri

10� 10 are the diode readings in the arbitrary

field at the ith location and in the 10� 10 cm2 reference field

at the same location, respectively. Ii are the corresponding

ion chamber readings. The field size is defined at the point of

measurement.

We have previously demonstrated that the filed size cor-

rection is more pronounced in the ARCCHECK (Ref. 6) then in

the MAPCHECK,1,2 although both devices use the same diode.

Originally, a flat-phantom model was used as an approxima-

tion for ion chamber measurements to prove the effect.6

Now, we have refined the measurements with the cylindrical

PMMA shell. Field size dependence was measured with a

single diode facing toward the source (0�), away from the

source (180�), and at three positions in between (88.8, 121.1,

and 142.6�). The values for other angular positions were

interpolated. Asymmetric Y jaws in conjunction with the

collimator rotated to 90� were used to define the fields away

from the 0 or 180� positions. A set of measurements was per-

formed with and without the central plug.

Numerically, the field size value is approximated dynami-

cally for each measurement update based on the number of

directly-irradiated diodes surrounding the detector in ques-

tion. This alleviates the requirement of a priori knowledge

of a segment field size as described previously for a different

dosimeter.5 Given an estimated field size, an energy-specific

correction from a lookup table is applied.

II.D.3. Positional correction Cipos

Finally, positional correction Cipos accounts for the intrin-

sic anisotropy of the diode response and the additional ani-

sotropy introduced by the inhomogeneous phantom structure

at an arbitrary angular position i. Cipos also accounts for the

variation in diode response due to the different instantaneous

dose rates (i.e., dose per pulse) (Ref. 17) at different loca-

tions in the phantom. It is assumed here that the measure-

ment position of the phantom is similar to calibration.8

While strictly speaking, a diode response is a function of its

three-dimensional position, and it was demonstrated6 that

the effect of the longitudinal displacement from the central

axis is negligible. The correction factor can be determined

by making a measurement with the ion chamber and the

same diode at two different angular locations: reference (0�)and arbitrary ith one in question. By definition, at the refer-

ence location C0pos¼ 1.0. Then, taking the ratio of doses at

those two positions according to Eq. (1),

Cposi ¼ IiR0CFS

0

I0RiCFSi

; (3)

where I’s are ion chamber readings at the ith and 0� positions,

R’s are readings for the same diode taken at those positions,

and CFS’s are field size correction factors from Eq. (2), again

at those two positions. The ratio of the field size corrections

must be included because of the beam divergence and differ-

ence in the effective scattering volumes at various locations in

the phantom.

Cpos values were measured in a wide (27� 25 cm2) beam

with and without the PMMA plug.

The rotational ARCCHECK data from the intrinsic sensitivity

measurements and the corresponding ion chamber data from

the TPS accuracy test were synchronized by aligning the 50%

intensity points on the response curves. Cpos values were

determined by applying Eq. (3). When different diodes on the

PCB occupy the same nominal angular position, their orienta-

tions with respect to the beam are different, leading to segre-

gation into three groups (Fig. 1): middle (M), left lateral (LL),

and right lateral (RL). In other words, the diodes are not

“focused” toward the center of the phantom (insert in Fig. 1).

Consequently, Cpos values were determined separately for

each group using the reflection symmetries associated with

each group. The three center rings out of the 21 rings along

the cylinder axis were used in the analysis. The three rings

provide a set of triplet detectors at any angular position. These

triplets are averaged. There are 22 angular positions for each

triplet in each symmetry group M, LL, and RL. The M group

response curve was mirrored with respect to the sagittal plane

and averaged with the original one. The LL and RL groups

have a symmetry exchange across the sagittal plane, and they

were mirrored and averaged accordingly, resulting in a pair of

asymmetric curves that are a reflection of each other. This

symmetry-based procedure helps to reduce noise and to

account for minor misalignments during the measurements.

Once the positional corrections are tabulated, the gantry

angle relative to the phantom angular orientation has to be

known to apply them during actual measurements. To that

end, we implemented a virtual inclinometer described below.

II.E. Virtual inclinometer

II.E.1. Algorithm

The prior virtual inclinometer algorithm (we will call it

“VI-old”) based on finding the 50% intensity lines of a diver-

gent beam was described by Yan et al.8 They validated the

algorithm for the static IMRT beams. We have confirmed

their results with the commercial software implementation,

even for artificially low (1 or 2) monitor units per segment.

However, the angle readings are noisy for the VMAT plans

(Fig. 2). An alternative virtual inclinometer algorithm (we

will call this “VI-new”) was therefore required, and its design

is described here. VI-new is currently validated only for the

nominal couch angle. To describe VI-new, consider a virtual

projection plane that cuts the cylindrical phantom in half

(Fig. 3). Approximately, the same configuration of detectors

is positioned above and below this plane. This projection

plane is used to create two dose images produced by the same

beam—one of them is created by projecting the dose distribu-

tion measured by the diodes in the “upper” half of the phan-

tom and another is related to the “lower” half. The projection

is based on the known position of detectors in space, known

orientation of the projection plane and estimated position of

the radiation source. Beam attenuation alters the projection in-

tensity and must be taken into account. Different experimental

5025 Kozelka et al.: Optimizing the accuracy of a helical diode array 5025

Medical Physics, Vol. 38, No. 9, September 2011

Page 6: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

attenuation tables are used for different beam energies and

plugged vs hollow phantom configurations. If the estimated

position of the source is in coincidence with the actual

position [Fig. 3(a)], then the two images will be very similar

qualitatively, with some expected deviation due to depth-

dependence of scatter in phantom. If on the other hand the

estimated position deviates from the real one [Fig. 3(b)], the

two images will be substantially different. A difference

between the two images as a function of beam angle a can be

expressed as

D að Þ ¼X

i

Xj

image1 að Þi;j � image2 að Þi;j� �2

; (4)

where image(a)i,j is the pixel value of a given image and

summation is performed for each pixel of the image. In the

ideal case, when the central axis is exactly perpendicular to

the projection plane [Fig. 3(a)], D(a)¼ 0. In reality, we

numerically search for a global minimum of D(a), which

corresponds to the best estimate of the beam angle.

II.E.2. Validation

The performance of the algorithm can be easily verified

by comparison with the accelerator dynamic log (dynalog)

files.18,19 The dynalog files for dynamic arcs contain gantry

angle information in 50 ms increments. The virtual incli-

nometer information is also updated every 50 ms. The time

stamps are not synchronized, so the plot of the virtual incli-

nometer vs dynalog gantry angle can be randomly shifted by

up to 100 ms, leading to a maximum error of approximately

0.5� at the highest gantry speed. The virtual inclinometer

and dynalog data were compared for a number of simple

arcs and also for a set of VMAT plans with different degree

of complexity, as described later.

II.F. Dose verification measurements

II.F.1. Cross-calibration

To eliminate the effect of the accelerator output variation

and some other residual uncertainties (e.g., the uncorrected

temperature effect or diode sensitivity degradation between

scheduled periodic recalibrations), the ARCCHECK was irradi-

ated with a four-field box consisting of 10� 10 cm2 fields.

Absolute dose in the central 6� 6 cm2 area was extracted

and a calibration coefficient was determined to achieve best

agreement between measured and calculated dose over 144

detectors, similar to the previously described method.6

Assuming for the moment a rather high 3% random uncer-

tainty in the individual diode calibration, using an average of

144 readings reduces the overall random error by a factor of

12 to 0.25%.

II.F.2. Agreement between ARCCHECK and calculateddose – simple open fields

With the ARCCHECK phantom centered on the lasers, 100

MUs were delivered with a series of symmetric beams rang-

ing in size from 5� 5 to 27� 25 cm2, with the gantry angle

at 0�. Three fields from 3� 3 to 5� 5 cm2 in size, produced

by independent Y jaws and offset laterally by 10.4 cm, were

also used. In addition, a 27� 25 cm2 aperture was used to

create an arc spanning 356� and delivering 400 MU in 1.52

min. The difference between the measured and calculated

doses was evaluated by gamma analysis in two ways: with

the global maximum [“Van Dyk on” (Ref. 9)] and local

dose-error normalization. Two sets of threshold criteria for

gamma analysis were used: dose-error 3% with distance to

agreement (DTA) 3 mm [c(3%/3 mm)], and dose-error 2%

with DTA 2 mm [c(2%/2 mm)]. Thus four passing rates are

reported for each plan. The couch attenuation was modeled

for the arc dose calculations and all VMAT plans.

II.F.3. Agreement between ARCCHECK and calculateddose—VMAT

Four cases were planned following the methodology of

AAPM TG-119 Report9—Multi-Target, Mock Prostate,

Mock Head-and-Neck, and C-shape (Easy).6,20 Volumetric

modulated arcs were substituted for the beam arrangements

FIG. 3. Projection plane with simulated source position: (a) in coincidence

and (b) not in coincidence, with the real source position. In the new virtual

inclinometer methodology, all simulated source positions are searched for

an optimal match (minimum difference) of projected entrance and exit rela-

tive dose images.

FIG. 2. Deviation of the virtual inclinometer gantry angle vs time for two

methods, using the Varian Dynalog file as the reference. Data shown here

are for a representative VMAT treatment based on the AAPM TG 119 (Ref.

9) Mock Prostate planning objectives.

5026 Kozelka et al.: Optimizing the accuracy of a helical diode array 5026

Medical Physics, Vol. 38, No. 9, September 2011

Page 7: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

in the Report. Single-arc plans were generated for every case

and additional dual-arc plans for the H&N and C-shape.6 Ion

chamber measurements in a Plastic Water Cube phantom

(CIRS Inc, Norfolk, VA) indicated an average deviation

from the calculated dose of 0.4 6 0.7% at one standard devi-

ation (1 SD) across all points of interest for all plans. Point

dose-error was normalized to the prescription value.9

Gamma analysis was performed as described above, again

with both global and local dose-error normalization. No

measurement uncertainty was applied, unlike in TG-119,

where a 3% stated dose-error criterion is effectively inflated

to 4% for the MAPCHECK analysis.

III. RESULTS

III.A. Benchmarking TPS accuracy with the ionchamber in a PMMA phantom

Exit to entrance dose ratio results are presented in Table I.

The differences between the measured and calculated doses

in the homogeneous phantom for the field sizes up to

20� 20 cm2 indicate reasonably well modeled depth doses in

PMMA for all algorithms. With the air cavity, XiO Convolu-

tion and Eclipse AAA algorithms deviate from the measured

values by up to 11 and 9%, respectively. A greater error in a

homogeneous phantom was noted for the Pinnacle calcula-

tions for the 27� 25 cm2 field. The progressive deviation

trend with increased thickness of the intervening material is

confirmed by a dose profile in Fig. 4.

III.B. Calibration formalism and procedures

III.B.1. Repetition rate correction CRR

Graphs of the repetition rate correction are presented in

Fig. 5. With no plug, the ratio of reading at 10, and

�20 MU/min to that at 600 MU/min were essentially the

same, whether calculated for the entire diode population or

separately for the entrance and exit subgroups (maximum

difference of 0.2% at 10 MU/min and <0.1% above 20

MU). With the plug in place, some differences were noted at

the lower repetition rates and the sensitivity ratios were plot-

ted out separately for the entrance and exit detectors [Fig.

5(a)]. Comparison of repetition rate sensitivity dependence

for different accelerators is presented in Fig. 5(b).

III.B.2. Intrinsic relative sensitivity Cirel

The results of this calibration step are not presented in

isolation but rather as an integral part of the overall dosimet-

ric agreement tests below.

III.B.3. Field size dependence CiFS

Field size dependence of the diode response as a function

of position and phantom configuration is presented in Fig. 6.

As expected, the largest difference between the hollow and

plugged configurations is observed for the larger filed sizes,

as the scattering volume increases giving rise to the low

energy photons responsible for the detector over-response

compared to the ion chamber.16

III.B.4. Positional correction Cipos

The plots of positional correction vs the detector angular

position in the phantom are presented in Fig. 7.

III.C. Virtual inclinometer validation

A typical example of a plot of the VI-new angle reading

against the dynalog gantry angle as a function of time is

presented in Fig. 2. The worst case average error over the

entire arc for six VMAT plans of varying complexity

was� 0.5 6 0.6� (1SD). For an open arc, the worst average

error was 0.4 6 0.7�.

III.D. Dose verification measurements

The gamma analysis results for the open static beams and

an open arc and the VMAT plans are presented in Tables II

and III, respectively. The average repetition rates for the

TABLE I. Difference in measured (ion chamber) vs calculated exit to entrance dose ratios for different TPS algorithms. 6 MV. Data for homogeneous and air

cavity phantom configurations are presented.

Exit to entrance ratio: Measured—Calculated (%)

TPS/algorithm Pinnacle adapt. conv. XiO Superpos. XiO Convolution Eclipse AAA Monaco Monte Carlo

Eq. Sq. field size 5 10 20 25 5 10 20 5 10 20 5 10 20 5 10 20

Homogeneous PMMA 0.0 �0.9 1.2 2.6 �0.2 0.2 1.1 �1.1 �2.0 0.2 �2.5 0.4 �0.3 �2.6 �0.1 �1.2

15 cm air cavity 0.8 �0.1 �0.5 0.1 �0.6 �0.6 �2.0 6.4 11.5 12.1 5.3 9.2 6.6 0.6 1.5 0.6

FIG. 4. Measured (ion chamber—IC) vs Pinnacle calculated relative dose

profiles for 27� 25 cm2 fields, with and without the plug.

5027 Kozelka et al.: Optimizing the accuracy of a helical diode array 5027

Medical Physics, Vol. 38, No. 9, September 2011

Page 8: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

VMAT plans ranged from 255 to 507 MU/min, resulting in a

repetition rate correction factor close to unity.

IV. DISCUSSION

IV.A. Benchmarking TPS accuracy with the ionchamber in a PMMA phantom

Calculation of dose without the plug is a nice stress test

of a TPS’s performance in the presence of large air-density

heterogeneity. The failure of the XiO Convolution algorithm

under these conditions is not surprising.22,23 Similar prob-

lems with the Eclipse AAA algorithm are perhaps less

known but nonetheless were reported in the literature.21–23

It must be mentioned that a new deterministic algorithm

based on the solution of the Boltzmann radiation transport

equation (Acuros XB) was recently implemented in

Eclipse.24–27 We did not have an opportunity to evaluate this

algorithm with the ARCCHECK. The closest indication of its

performance with a hollow phantom was reported by Bush

et al.24 They compared the new algorithm with Monte Carlo

(MC) simulations on a unit density rectangular phantom con-

taining a 10 cm air cavity. Acuros XB was found to be in

agreement with MC within the calculation uncertainty, while

with AAA, an underestimate of secondary dose buildup of

8%–45% was observed. While verification is certainly

needed, this is a good indication that Acuros XB would be

sufficiently accurate with a hollow ARCCHECK phantom.

The rest of the algorithms studied in this work demon-

strated adequate agreement for the field sizes up to

20� 20 cm2. It is therefore prudent to use the PMMA plug

when calculating ARCCHECK QA plans with the AAA algo-

rithm in Eclipse and to use Superposition algorithm in XiO.

The effect of the incorrect exit dose is somewhat mitigated

by the fact that it contributes roughly one-third to the com-

posite sum of the entrance and exit doses, assuming a diode

is irradiated by opposing beams.

The data show progressively increasing disagreement

between the measured and calculated (Pinnacle) PDDs in a

solid PMMA phantom for the largest field size. Examination

FIG. 5. (a)Average repetition rate sensitivity for 1386 diodes with and without the PMMA plug. The data with the plug are further subdivided into the entrance

and exit subgroups of 254 detectors each. (b) Comparison of average repetition rate sensitivity measured with the ARCCHECK on different accelerators. MAPCHECK

results for an Elekta SL20 from Ref. 2 are included for comparison.

FIG. 6. Field size dependence of the diode response at different positions in

the phantom for: (a) hollow and (b) plugged phantom configurations.

FIG. 7. Positional correction factor (Cpos) as a function of detector angular

position in the phantom, for different detector positions on the PCB, with

and without the plug. Insert: Field size correction CFS interpolated as a func-

tion of the detector angular position in a 27� 25 cm2 field, with and without

the plug. Cpos and CFS are related through Eq. (3).

5028 Kozelka et al.: Optimizing the accuracy of a helical diode array 5028

Medical Physics, Vol. 38, No. 9, September 2011

Page 9: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

of the TPS model data in water revealed no excessive PDD

disagreement for the entire range of filed sizes (<1% local

error). The results were confirmed in a different PMMA shell

phantom with a different ion chamber (Wellhofer IC13, IBA

dosimetry, Bartlett, TN). Also, we previously noted no such

disagreement when using the A1SL chamber in a cylindrical

30 cm diameter Solid Water phantom.6 This large-field PDD

error has limited practical implications for the use of ARC-

CHECK, as the device is primarily intended for measurements

of the segmented beams.

The relative dose profile agreement with Pinnacle for the

hollow phantom is generally good, under 0.5%, but the error

rises above 1% in a 615� interval, peaking at 2.5% around

�135� (the minimum single X jaw opening to irradiate a

diode at this position would be 6.9 cm). These results were

as well confirmed on a different accelerator and with a dif-

ferent phantom/chamber combination as described above.

We attribute it to the ray lines being nearly tangential to the

air cavity circumference. This introduces additional uncer-

tainty in comparing the measured and calculated doses for

the detectors in that region.

IV.B. Calibration formalism and procedures

IV.B.1. Intrinsic relative sensitivity Cirel

The direct translational calibration procedure in a narrow

beam reduced the error at the edges of the array along the Y

axis by about 1.0% compared to the current method used for

commercial units, wherein the top diodes are calibrated with

the wide field procedure.14 The only requirement for the se-

quential narrow-beam procedure is the stable beam output

on the central axis. In our case, the accelerator was stable to

better than 0.1%. The narrow-beam procedure does not

require additional side scatter to properly equalize the

response of the peripheral detectors.28,30

IV.B.2. Repetition rate correction CRR

Nowadays, detector sensitivity to the accelerator repeti-

tion rate is more important than before2 because within

VMAT treatments, the dose rate may vary substantially. For

repetition rates of 40 MU/min and up, the standard deviation

of the corresponding correction factors for all individual ARC-

CHECK diodes in a hollow phantom configuration does not

exceed 0.3%, and is below 0.1% above 200 MU/min. With

no difference between the entrance and exit diode groups,

application of a single average correction factor is justified.

Since the magnitude of the correction is relatively small, par-

ticularly when the unit is calibrated at 300 MU/min, in this

work, we applied the dose rate correction based on the calcu-

lated average dose rate. All subsequent dosimetric verifica-

tion measurements that depend on this correction were

performed with a hollow phantom. At first glance, it would

seem that the situation becomes complicated when the plug

is used, as the entrance and exit dependencies diverge some-

what (Fig. 5), with the former being closer to the hollow

phantom and the latter essentially flat. However, with the

plug in place and a 6 MV beam, the exit dose is small in

comparison with the entrance dose (�one-third) and it is

therefore advisable to use the entrance side correction values

for the whole phantom. This strategy introduces an addi-

tional uncertainty of only 0.25% at the repetition rate of 20

MU/min and less than 0.1% above 60 MU/min. These errors

would be negligible in practice, especially weighted against

the gains in accuracy of the higher (entrance) doses.

The diode sensitivity variation with the linac repetition

rate was reported before by Letourneau et al.2 Their sensitiv-

ity ratio between 50 and 600 MU/min (0.984) for a single

MAPCHECK diode measured with an Elekta SL20 accelerator

compares favorably with our average ARCCHECK value of

0.987. The same small difference of 0.3% is seen between

TABLE II. Gamma analysis passing rates for simple open fields calculated by the Pinnacle TPS. All comparisons were for absolute dose and using both Global

(normalized to dose maximum) and Local percent dose-error normalization. The low threshold dose for analysis was 10% of the maximum dose.

Field size (cm2) c (3%/3 mm) Global, pass % c (2%/2 mm) Global, pass % c (3%/3 mm) Local, pass % c (2%/2 mm) Local, pass %

27� 25 100 99.6 99.6 95.7

10� 10 100 99.6 100 95.4

5� 5 100 98.4 98.4 89.1

5� 5, offset laterally by 10.4 cm 100 100 100 100

Full arc 27� 25 100 100 100 100

TABLE III. VMAT dose QA gamma analysis passing rates using both Global (normalized to maximum) and Local percent dose-error normalization. All com-

parisons were for absolute dose. The low threshold dose for analysis was 10% of the maximum dose.

c (3%/3 mm) Global, pass % c (2%/2 mm) Global, pass % c (3%/3 mm) Local, pass % c (2%/2 mm) Local, pass %

Prostate 99.5 96.8 95.8 88.5

Multi-Tgt 97.3 91.1 91.8 84.0

H&N, 1 Arc 99.5 95 97.3 88.1

H&N, 2 Arcs 98.3 93.1 94.9 85.9

C-shape, 1 Arc 99.2 93.9 95.3 85.9

C-shape, 2 Arcs 96.8 90.1 92.4 81.7

Average 98.4 93.3 94.6 85.7

5029 Kozelka et al.: Optimizing the accuracy of a helical diode array 5029

Medical Physics, Vol. 38, No. 9, September 2011

Page 10: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

the Varian Trilogy and TrueBeam accelerators at 100 MU/

min. In the range of average dose rates encountered in this

work (250–600 MU/min), the difference between all acceler-

ators does not exceed 0.2% [Fig. 5(b)] and can be considered

negligible in practice. The mechanism of the repetition rate

dependency in general, and of the difference between the en-

trance and exit diodes with the plug in place in particular, is

still unknown. Its investigation is a complex undertaking in

solid state physics, beyond the scope of this work.

IV.B.3. Positional correction Cipos

There is a clear difference between the positional correc-

tion values around 180� with and without the plug. The

amount of scatter, and consequently the necessary field size

correction, reaches its maximum at the bottom of the homo-

geneous phantom. Since the “lateral” curves were made mir-

ror images of each other, only one is shown for clarity for

each phantom configuration. The “lateral” curves are asym-

metric and differ from the symmetric “middle” curves. The

complex behavior of the plots at 90� and 270� is similar for

both plugged and hollow configurations. Based on visual

inspection of the drawings and the CT scan of the phantom,

those are attributable to the intrinsic structure of the PCB

boards and air cavities around them. The 1.5% spike seen at

�225� (and its omitted mirror image at �135�) for the

“lateral” diodes in the hollow phantom has to be attributed

to the air cavity, as the irregularities disappear when the plug

is inserted. It is not clear, however, whether the curve shape

abnormality is real or caused by directly comparing readings

of the different size detectors in the high gradient area.

Absent clear evidence to the contrary, we chose to use the

unaltered experimental values. The effect is relatively small

both in magnitude and in number of diodes affected in any

given beam orientation.

The dosimetric verification data presented below confirm

that the average positional corrections measured around the

center of the phantom are valid throughout the entire length

of the device. In addition, it was also confirmed by direct ion

chamber measurements at a position 7.5 cm off-axis in the

longitudinal direction.

It is worth noting that the positional corrections include

as a subcomponent what is commonly called “angular” or

“anisotropy” corrections. Regarding the diode detectors

studied here, Jursinic et al.29 have quantified the angular de-

pendence of the planar diode array, which shares some quali-

tative characteristics of angular dependence with the helical

one, but is inherently different due to the different phantom

configurations.

IV.C. Virtual inclinometer validation

The difference function D(a) has a rather sharp global

minimum and is therefore not sensitive to the decrease in

signal to noise ratio (SNR) during the low repetition rate por-

tions of VMAT deliveries. These drops in SNR greatly

affects the VI-old, 50% intensity projection algorithm

(Fig. 2). In addition, the accuracy of the old VI depends on

the beam aperture size and shape and is reduced when the

central plug is not used. If the beam cross-section is larger

than the array, the 50% intensity points cannot be found and

the algorithm fails. The new VI algorithm is robust and capa-

ble of sufficient accuracy, comparable to a physical incli-

nometer used with the ARCCHECK prototype previously,7 or to

the expected gantry angle tolerance of a typical linear accel-

erator. While the algorithm is formally validated only for the

nominal couch angle, preliminary data suggest that some

couch kick can be accommodated.

IV.D. Dose verification measurements

IV.D.1. Agreement between ARCCHECK and calculateddose—simple open beams

A good agreement is seen for all static symmetrical

beams at the c(2%/2 mm) level with local dose-error percent

normalization, which is more stringent than the global per-

cent difference method traditionally used for TPS commis-

sioning.9,30 Over 95% of the detector-measured absolute

dose points in the open 27� 25 cm2 field are within 2%

(local) or 2 mm of the TPS dose. When analyzed with the

same criterion used by Letourneau et al.7—straight 2% dose-

error or 1 mm DTA without gamma analysis31—the passing

rate for the wide open field is essentially the same as in that

work, at 91%. This level of agreement would be considered

satisfactory for Pinnacle under ideal measurement condi-

tions, i.e., single ion chamber measurements in the water

phantom.30,32 It is as good as could be reasonably expected

for a nonrectangular, PMMA, heterogeneous phantom with

multiple detectors. The errors tend to be localized in the

angular interval where the Pinnacle calculations disagree

with the ion chamber by 1.5 to 2.5%. When compared to the

XiO Superposition data, the deviation from the ion chamber

increases by another 2%. In the absence of clear evidence to

the contrary, it is prudent to rely on the experimental data

for calibration. This dose-difference will ultimately have to

be resolved by Monte Carlo calculations. In the meantime,

the dose agreement with the TPS for an open filed is reduced

because the number of detector are deliberately calibrated

not to agree with the TPS. As a point of reference, both com-

mercial 3D diode arrays have a small area of increased mea-

surement uncertainty. For the DELTA4, an additional

uncertainty of >2% exists within 65� angular intervals cen-

tered on the detector boards.12 The practical effect of those

errors was shown to be negligible.12 For the ARCCHECK, with

the full wide open arc, which is more representative of the

intended use of the device, the overall effect of the dosimet-

ric error diminishes even further (Table II). 88.8% of the

detectors are within 1% (local dose-error) of the calculated

dose for the open arc, in addition to the 100% c(2%/2 mm)

(local dose-error) passing rate. This is in contrast to the orig-

inal data from the commercial unit with no angular and field

size corrections, where no diode agreed with the calculated

absolute dose to within 2% (global dose-error).6

Yan et al.8 reported excellent agreement (100% at the 1%

dose-error level) between measured and calculated dose for

a similar wide field. It was not specified if the dose-error

5030 Kozelka et al.: Optimizing the accuracy of a helical diode array 5030

Medical Physics, Vol. 38, No. 9, September 2011

Page 11: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

normalization was global or local. But more importantly,

their results are applicable only to the homogeneous phan-

tom, where better dosimetric agreement is typically

expected, and the angular response correction is based on the

“true dose profile” extracted from the TPS. An agreement

with the TPS is thus forced. We chose to accept a higher

deviation from the TPS in favor of using the ion chamber

data as the gold standard.

IV.D.2. Agreement between ARCCHECK and calculateddose – VMAT

Unlike with the wide open fields, the average global pass-

ing rates has shown little difference compared to the previ-

ously reported results for the same plans.6 This is consistent

with our previous ozbservation6 that the plans in question

use relatively small field sizes, for which absence of the

(small) positional corrections did not appreciably influence

the passing rates. Passing rates in the range of 91.8%–97.3%

(average 94.6%) were achieved or the local dose-error c(3%/

3 mm) analysis. We believe that local agreement is a supe-

rior indicator of the TPS accuracy, although the de-factostandard practice is to normalize dose-error to the global

maximum.9,30,33 The goal to reliably achieve >90% passing

rates for the local c(2%/2 mm) criterion so far remains elu-

sive. An average rate of 85.7% was obtained. Comparison of

the passing rates for open and VMAT fields clearly indicates

the deterioration of agreement for the modulated beams, par-

ticularly with the 2% local dose-error criterion.

V. CONCLUSIONS

A comprehensive, robust calibration system was devel-

oped and validated for the commercial ARCCHECK helical

diode array to improve the accuracy of its reported dose

measurements. The system allows for calibrating the array

directly against an ion chamber signal collected with high

angular resolution. A robust novel virtual inclinometer with

0.5� angular accuracy complements this positional calibra-

tion system. In the single wide field, 95% of the measured

absolute dose points pass the c (2%/2 mm) test with the local

dose-error. This is close to what one may reasonably expect

for TPS commissioning under the ideal conditions (i.e., ion

chamber measurements in the water phantom) and is a very

good agreement for a heterogeneous cylindrical phantom

with multiple diode detectors. The errors tend to be located

in relatively small angular intervals around 135� and 225�.Those errors mimic the ion chamber dose deviation from the

TPS calculation, as the array was calibrated to directly fol-

low the ion chamber. Further study is needed to determine

with certainty if the error is primarily due to the calculation

(e.g., scatter from the wall nearly parallel to the beam) or

measurement (e.g., ion chamber volume averaging) uncer-

tainty. The practical effect of these errors is reduced by a

factor of 3 when a sum of exit and entrance dose is consid-

ered. As a result, 88.8% of the detectors are within 1% (local

dose-error) of the calculated dose for a wide open full arc

and 100% are within 2%. The VMAT plans based on the

TG-119 objectives show average passing rates in excess of

90% for both globally normalized c(2%/2 mm) and local

c(3%/3 mm). However, the goal of using some of the more

stringent criteria while still maintaining passing rates consis-

tently above 90% proved still elusive in our setup. Some

TPS algorithms, notably Eclipse AAA, are not sufficiently

accurate in the presence of the large air cavity. It is recom-

mended to use the homogeneous phantom with those.

ACKNOWLEDGMENTS

This work was supported in part by a grant from Sun Nu-

clear Corp. The authors are grateful to J.-L. Dumont, S. Mar-

shall, R. Larson (Elekta), A. Langenegger (Varian), and

W. Wedding for help with the different treatment planning

systems.

a)Current address: Elekta/CMS Inc., Maryland Heights, Missouri 63043.1P. A. Jursinic and B. E. Nelms, “A 2-D diode array and analysis software

for verification of intensity modulated radiation therapy delivery,” Med.

Phys. 30, 870–879 (2003).2D. Letourneau et al., “Evaluation of a 2D diode array for IMRT quality

assurance,” Radiother. Oncol. 70, 199–206 (2004).3J. Herzen et al., “Dosimetric evaluation of a 2D pixel ionization chamber

for implementation in clinical routine,” Phys. Med. Biol. 52, 1197–1208

(2007).4B. Poppe et al., “Two-dimensional ionization chamber arrays for IMRT

plan verification,” Med. Phys. 33, 1005–1015 (2006).5R. Sadagopan et al., “Characterization and clinical evaluation of a novel

IMRT quality assurance system,” J. Appl. Clin. Med. Phys. 10, 104–119

(2009).6V. Feygelman et al., “Evaluation of a New VMAT QA Device or: The

“X,” and “O,” Array Geometries,” J. Appl. Clin. Med. Phys. 12, 146–168

(2011).7D. Letourneau et al., “Novel dosimetric phantom for quality assurance of

volumetric modulated arc therapy,” Med. Phys. 36, 1813–1821 (2009).8G. Yan et al., “Calibration of a novel four-dimensional diode array’”,

Med. Phys. 37, 108–115 (2010).9G. A. Ezzell et al., “IMRT commissioning: multiple institution planning

and dosimetry comparisons, a report from AAPM Task Group 119,” Med.

Phys. 36, 5359–5373 (2009).10N. Papanikolaou et al., “Tissue inhomogeneity corrections for megavolt-

age photon beams: Report of Task Group No. 65 of the Radiation Therapy

Committee,” AAPM Report No. 85 (Medical Physics Publishing, Madi-

son, 2004).11F. M. Khan et al., “Clinical electron-beam dosimetry: report of AAPM

Radiation Therapy Committee Task Group No. 25,” Med. Phys. 18, 73–

109 (1991).12V. Feygelman et al., “Evaluation of a biplanar diode array dosimeter for

quality assurance of step-and-shoot IMRT,” J. Appl. Clin. Med. Phys. 10,

64–78 (2009).13P. R. Almond et al., “AAPM’s TG-51 protocol for clinical reference do-

simetry of high-energy photon and electron beams,” Med. Phys. 26, 1847–

1870 (1999).14W. E. Simon, J. Shi, and C. A. Iannello, U.S. patent 6,125,335 (10 April

1998).15M. G. Karlsson et al., “Semi-conductor detectors in output factor meas-

urements,” Radiother. Oncol. 42, 293–296 (1997).16Z. Yin, R. P. Hugtenburg and A. H. Beddoe, “Response corrections for

solid-state detectors in megavoltage photon dosimetry,” Phys. Med. Biol.

3691–3702 (2004).17J. Shi, W. E. Simon and T. C. Zhu, “Modeling the instantaneous dose rate

dependence of radiation diode detectors,” Med. Phys. 30, 2509–2519

(2003).18D. W. Litzenberg, J. M. Moran and B. A. Fraass, “Verification of dynamic

and segmental IMRT delivery by dynamic log file analysis,” J. Appl. Clin.

Med. Phys. 3, 63–72 (2002).19J. G. Li et al., “Validation of dynamic MLC-controller log files using a

two-dimensional diode array,” Med. Phys. 30, 799–805 (2003).

5031 Kozelka et al.: Optimizing the accuracy of a helical diode array 5031

Medical Physics, Vol. 38, No. 9, September 2011

Page 12: Optimizing the accuracy of a helical diode array dosimeter ... · would cease to be a major source of uncertainty in dose QA. ARCCHECK (Sun Nuclear Corp, Melbourne, FL) is a 3D diode

20V. Feygelman, G. G. Zhang and C. W. Stevens, “Initial dosimetric evalua-

tion of SmartArc—a novel VMAT treatment planning module imple-

mented in a multi-vendor delivery chain,” J. Appl. Clin. Med. Phys. 11,

99–116 (2010).21A. Fogliata et al., “On the dosimetric behaviour of photon dose calculation

algorithms in the presence of simple geometric heterogeneities: comparison

with Monte Carlo calculations,” Phys. Med. Biol. 52, 1363–1385 (2007).22R. Kohno et al., “Dosimetric verification in inhomogeneous phantom geo-

metries for the XiO radiotherapy treatment planning system with 6-MV

photon beams,” Radiol. Phys. Technol. 2, 87–96 (2009).23D. Robinson, “Inhomogeneity correction and the analytic anisotropic algo-

rithm,” J. Appl. Clin. Med. Phys. 9, 112–122 (2008).24K. Bush et al., “Dosimetric validation of AcurosVR XB with Monte Carlo

methods for photon dose calculations,” Med. Phys. 38, 2208–2221 (2011).25A. Fogliata et al., “Dosimetric validation of the Acuros XB Advanced

Dose Calculation algorithm: fundamental characterization in water,” Phys.

Med. Biol. 56, 1879–1904 (2011).26T. Han et al., “Dosimetric comparison of Acuros XB deterministic radia-

tion transport method with Monte Carlo and model-based convolution

methods in heterogeneous media,” Med. Phys. 38, 2651–2664 (2011).

27O. N. Vassiliev et al., “Validation of a new grid-based Boltzmann equa-

tion solver for dose calculation in radiotherapy with photon beams,” Phys.

Med. Biol. 55, 581–598 (2010).28T. A. Simon et al., “Wide field array calibration dependence on the

stability of measured dose distributions,” Med. Phys. 37, 3501–3509

(2010).29P. A. Jursinic, R. Sharma and J. Reuter, “MAPCHECK used for rotational

IMRT measurements: Step-and-shoot, Tomotherapy, RapidArc,” Med.

Phys. 37, 2837–2846 (2010).30B. Fraass et al., “American Association of Physicists in Medicine

Radiation Therapy Committee Task Group 53: quality assurance for

clinical radiotherapy treatment planning,” Med. Phys. 25, 1773–1829

(1998).31D. A. Low et al., “A technique for the quantitative evaluation of dose dis-

tributions,” Med. Phys. 25, 656–661 (1998).32G. Starkschall et al., “Beam-commissioning methodology for a three-

dimensional convolution/superposition photon dose algorithm,” J. Appl.

Clin. Med. Phys. 1, 8–27 (2000).33B. E. Nelms and J. A. Simon, “A survey on planar IMRT QA analysis,”

J. Appl. Clin. Med. Phys. 8, 76–90 (2007).

5032 Kozelka et al.: Optimizing the accuracy of a helical diode array 5032

Medical Physics, Vol. 38, No. 9, September 2011