nice NeÖtech Medical Systems Pvt. Ltd

Page 2: Oxygen analyser

Presentation Flow

COMPANY OVERVIEW

CLINICAL ASPECTS OF nice OXYGEN ANALYSER

PRODUCT OVERVIEW

TROUBLESHOOTING

Page 3: Oxygen analyser

Company INFO

Established in the year 1997 Group of highly skilled professionals 4000+ highly satisfied customers Deep penetration in NICU and respiratory domain User friendly range of products With deep skills and high end technology , we always strive

for growth and delighted customers

Page 4: Oxygen analyser

Timeline-nice Neotech

2005•New Factory Opened

•Increase in production Line

2001•Reg as Pvt ltd

•Production started for different models

2013•New unit opened

1999• Neotech Logo

Designed

• Designed first incubator & open

care prototype

2003• Export to

Vietnam

2009-2011

• CE certification for products

2015• 125+skilled

employee strong firm

• Multiple range of NICU products

1997•Started Neotech Medical Systems with 2 employees

•First Infant Radiant Warmer

Page 5: Oxygen analyser
Page 6: Oxygen analyser

What is an Oxygen Analyser?

The oxygen analyser is a device with an oxygen sensor intended to measure the percentage of oxygen in the gas or gas mixture delivered to the patient.

Page 7: Oxygen analyser

Why Oxygen Analyser is Used?

Measure percentage of Oxygen

Used along with incubator and oxygen hood

Oxygen Analyser

Page 8: Oxygen analyser

Introducing nice 5040 Oxygen Analyser

Designed to perfection

Page 9: Oxygen analyser

Features

Microprocessor controlled for enhanced performance

Large easily readable Hi- bright LED display

Battery Low Indicator

Large sensor cable

Sensor disconnect and failure alarm

20 hours battery back up

Page 10: Oxygen analyser

Operation

Control Panel Operation AlarmApplicationAnesthetic gases

Page 11: Oxygen analyser

Control Panel Operation

Page 12: Oxygen analyser

Alarm

O2 % High alarm: By using the Select key choose the high alarm setting. By using the increase and decrease key until the desired value is displayed in the LED Screen. Press the select or wait for seven second to exit.

O2 % Low alarm: By using the Select key choose the low alarm setting. By using the increase and decrease key until the desired value is displayed in the LED Screen. Press the select or wait for seven second to exit.

Sensor Error If the sensor defects or disconnect, the display screen shows ER1 with INOP LED will be activated

Calibration Error If the calibration fails, the display screen shows ER2 with INOP LED will be activated

Page 13: Oxygen analyser

Application

ProcedureNote: Prior to use, always test the batteries and alarms. Also check calibration, the sensor for leaks and damage, and the alarm settings.

The oxygen analyser can be used to measure a gas mixture for oxygen in two basic modes:

In the inhalation side of breathing circuit ahead of antibacterial filters, humidifiers and medicating devices.

In confined volumes such as incubators, oxygen hood or oxygen tents.When monitoring for oxygen in breathing circuits, the flow diverter must be used. The diverter should be screwed onto the threaded front end of the sensor. A tee adapter should be placed into the circuit, and theabove sensor assembly plugged into the tee adapter.

Page 14: Oxygen analyser

Anesthetic GasesGases that induce reading error

Oxygen Reading Error in a Mixture of Anesthetic Gas

Page 15: Oxygen analyser

TROUBLESHOOTING

AND MAINTENANCE

Page 16: Oxygen analyser

Cleaning & Maintenance

Caution: Use the cleaning solution sparingly on a cloth when cleaning the oxygen analyser. Do not saturate theunit - excessive solution causes damage to internal components.

Caution: Use of nonstandard components: Consult the manufacturer for repair and replacement of components. Use of incorrect component can adversely affect Safety, performance and/or damage the equipment performance

Caution: Do not autoclave or gas sterilize oxygen analyser and oxygen sensor.

Cleaning- The external surfaces of the oxygen analyser body may be cleaned by wiping them with a cloth moistened with a mild detergent solution.

Disinfecting – The diverter and tee adapter may be disinfected by washing them with isopropyl alcohol.

Page 17: Oxygen analyser

Cleaning & Maintenance (Cont;d…….)

Sensor & Cable

Cleaning – The external surfaces of the oxygen sensor and of the cable may be cleaned by wiping them with a cloth moistened with a mild detergent solution.

Disinfecting – The diverter and tee adapter may be disinfected by washing them with isopropyl alcohol or Cidex (per manufacturer’s instructions). The parts must be thoroughly dry before they are used.

Sterilizing – The diverter and tee adapter may be sterilized using Cidex, steam or ethylene oxide (per manufacturer’s instructions).

Page 18: Oxygen analyser

Do’s and Don’ts

Do’s

Read all of the directions before using for the first time. Calibrate every 8 hours or before every use. Visually inspect the sensor for leakage before each use. Calibrate using 100% oxygen and check in air. Check the HI and LO alarm settings prior to each use. Test batteries regularly and charge the device when battery indicator shows low battery Keep the unit, sensor and connections dry, or on the dry side of the breathing circuit. Recalibrate after replacing the batteries. Recalibrate after replacing the sensor. Make sure the sensor is properly attached. Remove the plastic flow diverter only when using the tee adapter. Remove and save the plastic flow diverter when using the sensor in non-flowing applications(incubators, oxygen hood, tents, etc.) Clean the case with isopropyl alcohol or mild detergent only

Page 19: Oxygen analyser

DON’T

Use this analyser if you suspect any malfunction. Use the device in the presence of flammable gases. Use alkaline batteries. Autoclave or freeze the sensor or device. Open or try to repair a leaking or broken sensor. Immerse the unit or sensor in any liquid. Pass hot or cold gas mixtures over the sensor. Expose the unit to devices that produce high levels of radio, short wave, microwave, x-ray, or highfrequency interference. Use cleaning agents or liquids in the cable receptacles or around the battery compartment Place the unit itself in a water vapor-saturated environment. Expose the unit to a condensing water environment such as a mist tent.

Page 20: Oxygen analyser

How nice Oxygen Analyser fare over competitors?

nice Oxygen Analyser

LED displayLi-ion Battery- rechargeable

Calibration using either air or oxygen

Page 21: Oxygen analyser

Trouble ShootingSystem Failure

Page 22: Oxygen analyser

Trouble Shooting( Cont’d…..)

Page 23: Oxygen analyser

Maintenance Intervals:

Always disinfect and clean the unit and accessories before any maintenance – even when returning theunit to the supplier for repair. Always disconnect power supply before any maintenance. Use only Manufacturer’s original parts for maintenance.

Warning: Periodically check the insulation and the connection of the charger cable it may cause fire because of poor insulation and short circuits due to aging.

Warning: Don’t misaligned the EMI Shielding and the beads it may cause the EMI interference to the equipment

Page 24: Oxygen analyser

Certifications and Approvals

Electrical & Product Safety : IEC 60601-1

EMC Safety : IEC 60601-1-2

Particular standard : ISO 7767

Page 25: Oxygen analyser

Contact Us

nice Neotech Medical Systems Private Limited

No 85, Krishna Industrial Estate

Mettukuppam, Vannagaram,

Chennai – 600095, Tamil Nadu,

INDIA

Tel: +91-44-24762594,24764608.

Fax: +91-44-24766920

Web: www.niceneotech.com.

E-mail: [email protected] , [email protected]

Overseas Enquiry: [email protected]

Toll Free : +91-1800-425-2594.