pacing code. cardiac pacing - ihs fileserver · cardiac pacing sami viskin tel aviv, 1998. pacing...
TRANSCRIPT
![Page 1: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/1.jpg)
1
Cardiac PacingSami ViskinSami Viskin
Tel Aviv, 1998.Tel Aviv, 1998.
Pacing code.Chamber
PacedChamber
SensedResponse
VAD
VAD
ITD
PR
O O O
ProgramRate
modulation
Programmable parameters in single chamber pacing.
• Lower rate limit• Sensitivity• Output
VoltagePulse width
SensitivitymV
5
2
8 Undersensing
Good sensing
OversensingHigh
Low
Sensitivity.Temporary pacemaker
Demand Asynch.Demand(less sensitive)
Programmable parametersPacing output.
Vol
tage
Pulse width
![Page 2: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/2.jpg)
2
87654321
0.2 0.4 0.6 0.8 1.0 1.4 1.6Pulse width (msec)
Volts
Threshold
2X
3X
Pacing threshold.Strength-duration curve.
Basic timing cycles: VVI pacing.
Lower rate limit
V:V R V
Basic timing cycles: VVI pacing.
V:Refractory
Oversensing.T wave oversensing
V:Myopotential oversensing
V:
Ventricular undersensing.
V:
Pseudo-undersensing.
![Page 3: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/3.jpg)
3
Failure to pace.No output
No capture
Failure to pace
No output:• Pacemaker failure• Total battery
depletion• Lead fracture• Lead disconnection• Oversensing
No capture• Partial battery
depletion.• Lead dislodgment• Lead insulation
brake• Threshold rise
Undersensing / oversensing.• Lead dislodgment• Poor lead position• Lead insulation break• Inappropriate
programming
Basic timing cycles: DDD pacing.
AV interval
VA interval
A
V
Lower rate limit = AV interval + VA intervalLower rate limit = AV interval + VA interval
Pacemaker-mediated tachycardia (DDD mode)
Tracking of a fast atrial rhythm.
End-less loop tachycardia
Initiation of end-less loop tachycardia.
Retrograde P
Retrograde P
PVC
Atrial non-capture
![Page 4: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/4.jpg)
4
Timing cycles (DDD) Refractory
AV
PVARP
Blanking periodSafety pacing
![Page 5: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/5.jpg)
5
Referred to emergency room because of this electrocardiog
What to do?
![Page 6: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/6.jpg)
6
P
QT=680 msec
Same Holter
![Page 7: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/7.jpg)
7
V-rate 38/min R-R=1600 msec; QT = 600 msec
Magnet-rate 98/min QT = 7200 msec
R-R=620 msec R-R=1320 msec
PPP PP PPP A
720 600
*
P P P P P PP PP P
A
A
P P
A woman (80 years old) with femur fracture in need for surgery
![Page 8: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/8.jpg)
8
Soon after surgery……….
Toran called because of pacemaker malfunction
What to do?
Toran places magnet on pacemaker………………..
Everything is fine after the magnet… o’Kus po’Kus?
Soon after surgery……….
![Page 9: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/9.jpg)
9
QT = 600 msec
QT = 480 msec
A
V
Stored electrogram in a dual chamber ICD: Is this VT?
A
V
A
V
A
V
![Page 10: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate](https://reader030.vdocuments.net/reader030/viewer/2022040110/5e0761f220c0e13b0f3b8235/html5/thumbnails/10.jpg)
10
Appropriate detection of VT 180/min
VT detection rate > 180/min
Antitachycardia pacing
VT = 179/min not detected