pain stimulator adam plattjen wells biz slyziukkalee mcfadden bryan eckerleapril 13, 2004
TRANSCRIPT
Pain Stimulator
Adam Platt Jen Wells
Biz Slyziuk Kalee McFadden
Bryan Eckerle April 13, 2004
Dr. Stephen Bruehl
Research Interests Endogenous Pain Regulatory Systems,
Psychobiology of Emotions Focus: Understanding endogenous pain
regulatory systems and dysfunction associated with chronic pain.
Interaction between cardiovascular and pain regulatory systems to maintain homeostasis during painful stimuli compared to interaction in chronic pain patients
Blood Pressure
A5 Cell GroupNE Neurons and
α-2 receptors
Rostal Ventrolateral Medulla
NE Neurons and α-2 receptors
Caudal Ventrolateral
MedullaNE Neurons
Rostal Ventromedial Medulla
α-2 receptors
IntermediolateralCell Column
(SNS Preganglionic Neurons)
PeripheralSNS Terminals(α-2Presynaptic Autoreceptors)
Primary Afferent Fiber Terminals α-2 Presynaptic
Dorsal HornLaminae I-IV
α-2 Postsynaptic
Nucleus Tractus Solitarius
NE Neurons and α-2 receptors
Baroreceptor Input
PeriaqueductalGray
α-2 receptors
Locus CoeruleusNE Neurons and
α-2 receptors
Nucleus Raphe Magnus
Spinal Sites of Descending Modulation
Experimental Importance
NIH funded 120 subjects Two research sessions one week apart Stimulation of C-fibers Placebo vs. drug
Measures level of activity of receptors Draw conclusions on natural pain
systems in humans
Experimental Method
Dinamap Compact-T automated oscillometric BP monitor
Finapres 2300 ECG Cannula inserted into dominant arm Blood drawn to assess natural stress hormones
and inflammatory substances Forgione-Barber Finger Pressure Stimulator McGill Pain Questionnaire
Drugs to be Tested
Yohimbine Blocks 2 adrenergic receptors which inhibit release
of norepinephrine Used clinically for erectile dysfunction (orally) For Dr. Bruehl’s purposes, the FDA has only
approved injections
Naloxone Blocks opiate receptors Used clinically for heroine overdose Injected into patient
Problems with Existing Device
No electronic output Oral reading
No accurate measurement of pressure
Lever system with constant weight
No adjustability for height or hand dominance Adjustable only for finger
placement
Our Design
Device
Maintain lever system
Most stable design
Reposition weight to be placed vertically above finger
Adjustable Weight
2 – 5 lbs
Contact Point
Between top two joints on index finger of dominant hand
Middle Phalanx #2
Our Design… Cont.
Table Provided by Dr. Bruehl Transportation Positioning
Adjustable Height Sliding Tracks Forearm length
Slider scale mounted to table displaying “No Pain” to “Worst Pain Imaginable”
Scale mounted on each side of device
Device
(Pressure Transducer)
Slider Scale
A/D Converter
PC Computer
LabVIEW
Schematic
Device with Measurements
Materials
Aluminum Lucite Steel
Device Construction
In process Featheringill machine shop
LabView
Timer Sliding Scale Digital Display Graphical Output Audiovisuals Digital Data Storage
Patient Administrator Date Experimental Data
LabVIEW Interface
Potentiometers
Circuit sensitivity
≈ .62V/cm @ Vs=5V
Potentiometer sensitivity
≈ 1500Ω/cm
Questions?