paper ei(2) by pankaj sir

Upload: mahesh-singh

Post on 02-Jun-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    1/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    2/247

    Jainam Solved Paper Electronic Instrumentation

    It is the smallest change in input which an electronic instrument

    is able to detect. Thus, it is the full scale value of the lowest

    voltage range multiplied by the resolution of the meter.

    Sensitivity,

    fs

    S=

    Where,

    f

    ( s)min

    =Lowest full-scale value of digital meter

    R= Resolution in decimal

    c)ACCURACY:

    The accuracy of an instrument is a measure of how close the

    output reading of the instrument is to the correct value. In the

    accuracy specifications, the following two quantities are included.

    A percentage of range.

    A percentage of reading.

    It is an important system design rule that instruments are

    chosen such that their range is appropriate to the spread of

    values being measured, in order that the best possible accuracy is

    maintained in instrument readings.

    d)AVERAGE/TRUE ROOT MEAN SQUARE:

    The standard deviation of an infinite number of data is the Square

    root of the sum of all the individual deviations squared divided by

    the number of readings.

    It may be expressed as

    = d 1

    2+d22+d32 +dn2

    n

    The standard deviation is also known as root mean square

    deviation, and is the most important factor in the statisticalanalysis of measurement data. Reduction in this quantity

    effectively means improvement in measurement.

    Jainam Publication TMPage 2

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    3/247

    Jainam Solved Paper Electronic Instrumentation

    e)CREST FACTOR:

    The crest factor or peak-to-average ratio (PAR) is a measurement of

    awaveform, calculated from thepeak amplitudeof the waveform

    divided by theRMSvalue of the waveform [1]:

    The purpose of the crest factor calculation is to give an analyst a

    quick idea of how much impacting is occurring in a waveform.

    Impacting is often associated with roller bearing wear, cavitation

    and gear tooth wear.

    f)FORM FACTOR:

    Form factor is defined as the ratio of root mean square value to

    the average value of the unit.

    In the case of a sinusoidal wave ie, an analogue wave, the form

    factor is approximately 1.11. In the case of a square wave i.e., a

    digital wave, the RMS and the average values are equal; therefore,

    the form factor is 1.

    Q. 2: What do you understand by the following terms:

    i) Normal mode rejection ratio

    ii) Common mode rejection ratio

    iii)Effective common rejection ratio

    iv)Zero or offset frequency response (CSVTU April- May 2009)

    Ans:

    i) NORMAL MODE REJECTION RATIO:

    Jainam Publication TMPage 3

    http://en.wikipedia.org/wiki/Waveformhttp://en.wikipedia.org/wiki/Peak_amplitudehttp://en.wikipedia.org/wiki/Root_mean_squarehttp://en.wikipedia.org/wiki/Crest_factor#cite_note-0http://en.wikipedia.org/wiki/Peak_amplitudehttp://en.wikipedia.org/wiki/Root_mean_squarehttp://en.wikipedia.org/wiki/Crest_factor#cite_note-0http://en.wikipedia.org/wiki/Waveform
  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    4/247

    Jainam Solved Paper Electronic Instrumentation

    Normal-mode rejection ratio (NMRR) describes the ability of the

    DMM to reject an ACnormal-mode signal, usually at power line

    frequencies. NMRR is given by the following formula:

    NMRR = 20*log(Vin/Verror)

    WhereVerroris the value returned by the DMM for an applied AC

    normal-mode voltageVin.

    NMRR is useful for measurement systems that can eliminate

    signals at a given frequency or over a range of frequencies. NMRR,

    which is often used to indicate the capability of the instrument to

    reject powerline noise of 50 or 60 Hz, is valid only at the specified

    frequency and is useful when making DC measurements.

    ii) COMMON MODE REJECTION RATIO:

    Thecommon-mode rejection ratio(CMRR) of adifferential

    amplifier(or other device) is the tendency of the devices to reject

    the input signals common to both input leads. A high CMRR is

    important in applications where the signal of interest is

    represented by a small voltage fluctuation superimposed on a

    (possibly large) voltage offset, or when relevant information is

    contained in the voltage difference between two signals.

    The CMRR is a measure of how well the device rejects a common-

    mode signal.

    Its simply the ratio of the differential gainAvover the common-

    mode gainAcm.

    Jainam Publication TMPage 4

    http://zone.ni.com/reference/en-XX/help/370384N-01/dmm/normal_mode/http://en.wikipedia.org/wiki/Differential_amplifierhttp://en.wikipedia.org/wiki/Differential_amplifierhttp://zone.ni.com/reference/en-XX/help/370384N-01/dmm/normal_mode/http://en.wikipedia.org/wiki/Differential_amplifierhttp://en.wikipedia.org/wiki/Differential_amplifier
  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    5/247

    Jainam Solved Paper Electronic Instrumentation

    iii) EFFECTIVE COMMON REJECTION RATIO:

    Effective common-mode rejection ratio (ECMRR) is the sum of

    CMRR and NMRR at a given frequency and is only valid for DCmeasurements. It is the effective rejection on a given noise signal

    that is applied to both input leads because it is rejected first by

    the CMRR capability of the instrument and then again by its

    NMRR capability. This specification is useful at powerline

    frequencies, particularly for laboratory and manufacturing floor

    environments. An equivalent equation to represent ECMRR is as

    follows:

    ECMRR = 20*log10(VCM/Verror)

    whereVerroris the value returned by the digital multimeter in

    response to an applied common mode voltageVCM.

    iv) ZERO OR OFFSET FREQUENCY RESPONSE:

    Proportional control action is characterized by a permanent

    residual error in the operating point of the controlled variablewhen a change in load occurs. This error is called the OFFSET.

    The offset can be reduced by selecting higher value of controller

    gain (KP) corresponding to narrow bandwidth.

    Fig.1. Offset Error

    Figure 1, shows how offset error occurs in a proportional

    control action. Let the system error be zero at nominal load withcontroller output P = 50% corresponding.

    Jainam Publication TMPage 5

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    6/247

    Jainam Solved Paper Electronic Instrumentation

    If, however, a transient error occurs, the system tends to adjust

    the controller output so that the point A (corresponding to zero

    error) is reached again. But for this to happen there must be

    change in the load system. This changed controller load produces

    a new value (P new) of the characteristic output, which gives rise

    to point B on the linear characteristic of the proportionalcontroller. The permanent small difference between the percentage

    error values corresponding to point B and A is called the offset

    error of the fcontrol.

    Q. 3: - Explain digital voltmeters?

    (CSVTU Nov-Dec 2010, Nov-Dec 2009, April-May 2009, Nov-Dec

    2008, Nov-Dec 2007)

    Ans:

    DIGITAL VOLTMETERS (DVMS)

    The digital voltmeter (DVM) displays measurements of ac or

    dc voltages as discrete numerals instead of a pointer deflection ona continuous scale as in analog instruments. It is aversatile and

    accurate instrument that is employed in many laboratory

    measurement applications.

    Digital voltmeters (DVMs) are measuring instruments that

    convert analog voltage signals into a digital or numeric readout.

    This digital readout can be displayed on the front panel and also

    used as an electrical digital output signal. Any DVM is capable of

    measuring analog dc voltages. However, with appropriates signal

    conditioners preceding the input of the DVM, quantities such as

    ac voltages, ohms, dc and ac current, temperature, and pressure

    can be measured. The common element in all these signal

    conditioners is the dc voltage, which is proportional to the level of

    the unknown quantity being measured. This dc output is then

    measured by the DVM.

    It is a versatile and accurate instrument that is employed

    in many laboratory measurement applications. Because of

    Jainam Publication TMPage 6

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    7/247

    Jainam Solved Paper Electronic Instrumentation

    development and perfection of IC modules, the size, power

    requirements and costofthe digital voltmeter has been drastically

    reduced and, therefore, DVMs can actively compete with

    conventional analog instruments, both in price and portability.

    The block diagram of a simple digital voltmeteris shown inFig.1.

    The unknownvoltagesignal is fed to the pulse generator

    which generates a pulse whose width is directlyproportionaltothe input unknown voltage. The output of the pulse generator is

    applied to one leg of anANDgate. The input signal to the other leg

    of the AND gate is a train of pulses. The outputof theAND gate is,

    thus, a positive trigger train of durationt seconds and the inverter

    convertsit intoa negative trigger train. The counter, counts the

    number of triggers intsecondswhich mproportional to the

    voltage under measurement. Thus, the counter can be calibrated

    to indicate voltage in volts directly.

    Thus, we see that the DVM described above is an ADC

    which converts an analog signal into a train of pulses, the number

    of which is proportional to the input voltage. So a digital voltmeter

    can be made by using any one of the A/D conversion methods and

    can be represented by a block diagram shown in Fig. 2. Thus the

    DVMs can be classified on the basis of ADCs used.

    Jainam Publication TMPage 7

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    8/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    9/247

    Jainam Solved Paper Electronic Instrumentation

    Fig. 2 shows the,"voltage-to-time conversion"using gated -dock

    pulses.

    At the start of the measuring cycle, a ramp voltage is

    initiated; this voltage can be positive going or, negative goings the

    negative-going ramp, shown in fig.2 is continuously compared

    with the unknown input-voltage. At the instant that the ramp voltage equals the unknown

    voltage, a coincidence circuit, or comparator generates a pulse

    Jainam Publication TMPage 9

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    10/247

    Jainam Solved Paper Electronic Instrumentation

    which opens a gate, the ramp voltage continues to decrease with

    time until finally reaches 0 V and a second comparator generates

    an output pulse which closes the gate. An oscillator generates

    clock pulses which are allowed to pass through-the-gate to a

    number of decades counting units (DCUs) which totalize the

    number of pulses passed through the gate. Thedecimal numberdisplayed by the indicator tubes associated with the DCUs, is a

    measure of the magnitude of the input voltage.

    The sample-rate multivibrator (MV) determines the rate

    at which the measurement cycles are initiated. The sample-rate

    circuit provides an initiating pulse for the ramp generator to start

    its next ramp voltage. At the same time, a reset pulse is generated

    which returns all the DCUs to their zero state, removing the

    display momentarily from the indicator tubes.

    Advantages:

    1)The ramp technique circuit is easy to design and its cost is

    low.

    2)The output pulse can be transmitted over long feeder lines.

    However, the single ramp requires excellent characteristics

    regarding linearity of the ramp and time measurement.

    Disadvantage:

    Large errors are possible when noise is superimposed on the

    input signal. Input filters are usually required with this type

    of converter.

    2)DUAL SLOPE INTEGRATING TYPE DVM:

    In ramp techniques, superimposed noise can cause large errors.

    In the dual ramp technique, noise is averaged out by the positive

    and negative ramps using the process of integration.

    The basic Dual Slope Integrating Type DVM consist of Five basic

    building blocks ,

    1.Op-amp, used as an integrator;2.A level comparator;

    3.A basic block, for generating timing pulses;

    Jainam Publication TMPage 10

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    11/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    12/247

    Jainam Solved Paper Electronic Instrumentation

    This unknown time t is determined by counting timing pulsesfrom

    the clock until the voltage across the capacitor reaches its basic

    reference value (reference may be ground or any other basic

    reference level ).

    Then, from similar triangles of Figure shown above we have :

    The count aftertwhich isproportional to the input voltage Vi is

    displayed as the measured voltage.

    This instrument can be used to measure currents, resistances andA.C. voltages by using appropriate signal conditioners.

    Advantages:

    i)The averaging characteristics and cancellation of errors that

    usually limit the performance of ramp-type DVM are the

    main advantages of such DVMs:

    ii)The integration characteristics provide the average value of

    the input signal during the period of first integration.

    Consequently, disturbance, such as spurious noise pulses,

    areminimized.

    iii) Long-term drifts in the time constant, as may result from

    temperature variations or aging, do not affect conversion

    accuracy. The long-term alterations in clock frequency have

    no effect.

    iv) Higher accuracy and resolution.

    v) Greater speed.

    vi) No parallax.

    vii)Reducedhuman error.

    viii)Compatibility with other digital equipment for further

    processing and recording.

    Q. 5: Discuss digital frequency meters? (CSVTU April-May

    2011, April-May 2010, Nov-Dec 2008)

    Jainam Publication TMPage 12

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    13/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    14/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    15/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    16/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    17/247

    Jainam Solved Paper Electronic Instrumentation

    In the "A.C. voltage mode", the applied input is fed

    through a calibrated, compensated attenuator, to a precision full-

    wave rectifier circuit followed by a ripple reduction filter. The

    resulting D.C. is fed to ADC and the subsequent display system.

    For current measurements, the drop across an internal

    calibrated shunt is measured, directly by the ADC in the "D.C.

    current mode", and after A.C. to D.C. conversion in the "A.C.

    current mode". This drop is often in the range of 200 mV

    (corresponding to full scale).

    Due to the lack of precision in the A.C.-D.C. conversions,

    the accuracy in the A.C. range is in general of the order of 0.2 to

    0.5%. In addition, the measurement range is often limited to

    about 50 Hz at the lower frequency end due to the ripple in therectified signal becoming a non-negligible percentage of the

    display and hence results in fluctuation of the displayed number.

    At the higher frequency end, deterioration of the performance of

    the AC/DC converter limits the accuracy.

    The A.C. measurement, isoften average reading, r.m.s.

    calibrated.

    Jainam Publication TMPage 17

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    18/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    19/247

    Jainam Solved Paper Electronic Instrumentation

    And we can read the rpm of the rotating shaft directly. So,

    the relation between the gate period and the number of pulses

    produced by the pickup isG= 60/P.If we fix the gate period as

    one second (G = 1 s), then the revolution pickup must be capable

    of producing 60 pulses per revolution.

    Figure shows a schematic diagram of a digital tachometer.

    Q. 8: - Discuss digital pH meter? (CSVTU April-May 2010)

    Ans: -

    DIGITAL PH METER:

    The measurement of hydrogen ion activity (pH) in a solution can

    be accomplished with the help of a pH meter. For those unfamiliarwith the terminology, a very brief review is included.

    pH is a quantative measure of acidity. If the pH is less

    than 7, the solution is acidic (the lower the pH, the greater the

    acidity). A neutral solution has a pH of 7 and alkaline (basic)

    solutions have a pH greater than 7.

    The pH unit is defined as

    pH = - log (concentration of H+

    )where H+is the hydrogen or hydronium ion.

    Jainam Publication TMPage 19

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    20/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    21/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    22/247

    Jainam Solved Paper Electronic Instrumentation

    Now as the inputPxwhich has a phase difference with

    respect to P0crosses zero (0) in the positive half cycle, the zero

    detector is activated, causing its output to go high(1). This high

    input in turn toggles the J-K flipflops-2, making its output goes

    high. This output (Q) of flipflops-2 is connected to the clear input

    of flipflops-l forcing the flipflops-l to reset. Hence the output offlipflops-l goes to zero (0). The AND gate is thus disabled, and the

    counter stops counting.

    The number of pulses counted while enabling and

    disabling the AND gate is in direct proportion to the phase

    difference, hence the display unit gives a direct readout of the

    phase difference between the two inputs having the same

    frequency. If the input signal frequency is f, then the clock

    frequency must be 360times the input frequency for accurate

    measurements.

    Q. 10: - What do you mean by digital capacitance meter?

    Ans:

    DIGITAL CAPACITANCE METER:

    Since the capacitance is linearly proportional to the time constant,

    when a capacitor is charged by a constant current source and

    discharged through a fixed resistance, we can use a 555 timer

    along with some digital test equipment to measure capacitances.

    One obvious way is to measure the time period of the

    oscillations. By choosing the right size of charging resistance, we

    can get a reading directly in microfarads ornanofarads.Unlike

    many capacitance measuring schemes, this one easily handles

    electrolytics up to the tens of thousands of microfarads.

    A better way is to measure only the capacitor discharge

    time, as shown in Figure below

    Jainam Publication TMPage 22

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    23/247

    Jainam Solved Paper Electronic Instrumentation

    With this method, any leakage in the capacitor under test

    will make the capacitor appear smaller in value than it actually is,

    and is an effective indicator of how the test capacitor will behave

    in most timing and bypass circuits.

    In this circuit, the 555 timer is used as an astable

    multivibrator. At the peak of the charging curve, a digital counter

    is reset and a clock of 100 kHz pulses is turned on and routed tothe counter. When the discharge portion of the cycle is completed,

    the display is updated and the value of the capacitor is readout.

    By selecting the proper reference frequency and charging currents,

    one can obtain a direct digital display of the value of the

    capacitance.

    Be sure to properly shield the leads and keep them short

    for low capacity measurements, since the 50 Hz hum can causesome slight instability.

    Q. 11: -Explain the advantages of electronic voltmeters over

    conventional type voltmeters are regards(CSVTU Nov-Dec

    2010)

    (i) detection of low level signals

    (ii) power consumption

    Jainam Publication TMPage 23

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    24/247

    Jainam Solved Paper Electronic Instrumentation

    (iii) loading effects

    (iv) frequency range.

    Ans:

    Inall electronics voltmeter circuits the principle involved is

    that an indicator on a permanent magnet moving coil

    instrument portional to the input voltage is obtained by means

    of amplification in one or more stages with a high input

    impedance.

    Being costlier than electrical instrument but it has many

    advantages over conventional ones, as discussed below:

    1.High Sensitivity

    2.Low Power Consumption

    3.High Accuracy

    4.High Frequency range

    5.Low level signal detection

    6.Less loading effect

    7.High input impedance

    Let us explain the advantage of electronic voltmeter with

    example:

    Analog instruments use PMMC movement for indication.

    This movement cannot be constructed with a full scale sensitivity

    of less than 50 A. Any measurement using a P'MMC movement

    must draw a current of 50 A from the measured quantity for its

    operation for full scale deflection if conventional voltmeters are

    used. This would produce great loading effects especially in

    electronic and communication circuits. Electronic voltmeters avoid

    the loading errorsby supplying power required for measurement

    by using external circuits like amplifiers. The amplifiers not only

    supply power for the operation but make it possible for low level

    signals, which produce a current less than 50 A for full scaledeflection, to be detected which otherwise cannot be detected in

    the absence of amplifiers.

    Jainam Publication TMPage 24

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    25/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    26/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    27/247

    Jainam Solved Paper Electronic Instrumentation

    iv) Hysteresis Error:

    The maximum separation due to hysteresis between

    upscale-going and downscale-going indications of a measured

    variable. The delay between the action and reaction of a

    measuring instrument. Hysteresis is the amount of error thatresults when this action occurs.

    The maximum differences in outputs at any measured value

    within the specific range when approaching the point first with

    increasing and then with decreasing input may be termed as

    Hysteresis.

    v)Dynamic Error:

    The Dynamic Error also called measurement error is the

    difference between the true value of the quantity changing with

    Time and the value indicated by the measurement system if no

    static error is assumed.

    vi) Cross Sensitivity:

    It is a factor which is to be taken into account when we

    measure mechanical quantity. It is a situation where the actual

    Jainam Publication TMPage 27

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    28/247

    Jainam Solved Paper Electronic Instrumentation

    quantity is being measured is in one plane and another quantity

    which is subjected to variations is on another plane.

    Q: 13: - A 3 digit DVM has an accuracy specification of

    0.05% of the reading 1 digit.

    (i) What is the error in volt when reading is 2 V on its 10

    V range?

    (ii) What is the error in volt when reading is 5 V on its 10

    V range?

    (iii)What is the % error of reading, when the reading is

    0.1 V on its 10 V range? 7(CSVTU Nov-Dec 2011, April-May 2011)

    Ans:

    1.Number of full digits on display, n=3

    So, resolution, R=1

    10n=

    1

    103 = 0.001

    i.e. the meter cannot distinguish between values that differs

    from each other by less than 0.001 of full scale.

    For full scale range of 1 V, the resolution isfs R i.e.

    1x0.001=0.001V

    For full scale range of 10 V, the resolution is 10 x 0.001 = 0.01 V

    2.The display for 2 V reading on 10 V scale of31

    2 digital

    meter would be 02.00

    The digit in the least significant digit has a value offs R i.e.

    5 x1

    103 = 0.005 V

    Total possible error =0.5

    100 x reading + value of digit in LSD

    Jainam Publication TMPage 28

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    29/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    30/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    31/247

    Jainam Solved Paper Electronic Instrumentation

    2)There are 6 digit faces in 5 display, so 13.46 would be

    displayed as 13.4600 V.

    3)Resolution on 1V range is 0.00001 * 1=0.00001

    Any reading up to the 5thdecimal can be displayed.Hence 0.54867 will be displayed as 0.54867 V.

    4)Resolution on 10V range is 0.00001 * 10=0.0001

    Any reading up to the 4thdecimal can be displayed.

    Hence 0.54867 will be displayed as 0.5486 V.

    Q. 17: - What do you mean by analog and digital systems.

    Write some advantages and limitations of digital over

    analog system?

    (CSVTU Nov-Dec 2011, Nov-Dec2009, April-May 2008)

    Ans:

    Analog System:

    An analog system comprises devices that manipulate the

    physical quantities represented in analog form. Example:

    automobile speedometer.

    Digital System:

    A digital system is a combination of devices for manipulating

    physical quantities or information represented in digital form.

    Example: digital computer.

    Advantages of digital system over analog system:

    1)Easier to design.

    2)Easy storage of information

    3)Greater accuracy and precision.

    4)Operation is programmable.

    Jainam Publication TMPage 31

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    32/247

    Jainam Solved Paper Electronic Instrumentation

    5)Digital circuits are less affected by the noise.

    Limitation of digital technique:

    1)Cost is more than analog

    2)Isolation problems are more.

    3)Construction is complicated.

    Examination, April May, 2012

    Q.1. (a)Define the terms sensitivity& accuracy. 2

    Ans: Refer Answer 1-(b) and(c).

    (b)Compare the characteristic of ramp & dual slope type DVM.

    7

    Ans: Refer Answer 4.

    (c)In the wheatstone bridge shown in figure, the values of

    resistance of various arms are given. The galvanometer has a

    current sensitivity of 10 mm/A & an internal resistance of100. Calculate the deflection of galvanometer & sensitivity of

    bridge in terms of deflection per unit change in resistance. 7

    Ans:

    Resistance of unknown resistance in balance condition

    R=(P/Q)*S

    R=(1000/100)*200=2000

    R=2005-2000=5

    Thevenins source generator emf

    Jainam Publication TMPage 32

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    33/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    34/247

    Jainam Solved Paper Electronic Instrumentation

    (b)Explain the operational principle of a dual slope integrating

    type DVM with help of diagrams. If the DVM has maximum

    range of 255 V, determine the time it will take to read the

    unknown voltageVX=180V , using a clock oscillator of 10 kHz.

    Ans: -

    For dual slope integrating type DVM refer answer.

    Remaining solution:

    Maximum range of DVM Vmax= 255 V

    Unknown voltage Vx= 180 V

    Frequency of clock oscillator = 10 kHz

    Time taken to read unknown voltage

    T= {(Vmax- Vx)/2f Vmax= 4.68 sec

    (c)What is the resolution of a31

    2 digit display on 1V & 10V

    ranges? A31

    2 digital voltmeter has an accuracy specification

    of 0.5 of reading 1 digit. What is the possible error in

    volts, when instrument is reading 2.00 V on 5 V scale.

    Ans: Refer answer 13.

    (d)Explain the merits of digital systems over analog systems.

    Also, discuss the limitations of digital techniques.

    Ans:Refer answer 17.

    Examination, April May, 2011Q.1.(a) What is resolution of 3 digit? 2Ans: Refer answer 13.

    (b)Describe any one method of integrating type DVM with theiradvantage & disadvantage. 7

    Ans: Refer answer 4.

    (c)Explain digital frequency meter. 7Ans: Refer answer 5.

    (d)A 3 digit DVM has an accuracy specification of 0.05% of thereading1 digit.

    Jainam Publication TMPage 34

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    35/247

    Jainam Solved Paper Electronic Instrumentation

    (i) What is the error in volt when reading is 5 V on its 10 V range?(ii) What is the % error of reading, when the reading is 0.1 V on its10 V range. 7

    Ans: Refer answer 13.

    Examination, Nov. Dec., 2010

    Q.I.What are the advantages of dual slope over ramp type DVM.2

    Ans: Refer answer 4.

    Q. II.Explain the advantages of electronic voltmeters over

    conventional type voltmeters are regards 7

    (i) detection of low level signals

    (ii) power consumption

    (iii) loading effects

    (iv) frequency range.

    Ans:Refer answer 3.

    Q. III.Explain the operating principle of a ramp type DVM. 7

    Ans: Refer answer 4.

    Q. IV.A 4 digit voltmeter is used for voltage measurements.7

    i) Find its resolution

    ii) How would 12.98 V be displayed on a 10 V range.

    iii)How would 0.6973 be displayed on 1 V and 10 V ranges.

    Ans:Refer answer 14.

    Examination, April May, 2010

    Q.1.(a)Define resolution. Give the resolution of a 41

    2 digit

    display.2

    Ans: Refer answer 1 and 14.

    (b)Discuss dual slope integrating type digital voltmeters. 7

    Ans: Refer answer 4.

    (c)Explain digital frequency meter. 7

    Jainam Publication TMPage 35

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    36/247

    Jainam Solved Paper Electronic Instrumentation

    Ans: Refer answer 5.

    (d)Describe the digital phase meter. 7

    Ans: Refer answer 9.

    Examination, Nov. Dec., 2009

    Q.1. (a)A Wheatstone bridge requires a change of 7in the

    unknown arm of the bridge to produce a change in deflection of

    3 mm of the galvanometer. Determine the sensitivity. Also

    determine the inverse sensitivity or scale factor. 2

    Ans:Refer answer 15.

    (b)Describe the working principle and block diagram of Dual slopeintegrating type DVM. 7

    Ans:Refer answer 4.

    (c)State the advantages of a DVM (digital voltmeter) over an analog

    meter. Also explain the basic principle of a digital voltmeter.7

    Ans:Refer answer 3 and 17.

    (d)A 41

    2 digit voltmeter is used for voltage measurements: 7

    (i) Find its resolution.

    (ii) How would 0.6973 be displayed on 1 V and 10 V rages.

    (iii)A moving coil voltmeter has uniform scale with 100 divisions.

    The full scale reading is 200 V and 1/10thof scale division can be

    estimated by the instrument. Determine resolution of instrument

    in volts.

    Ans:Refer answer 14.

    Examination, April May, 2009

    Q.1.(a)A 4 digit voltmeter is used for voltmeter measurement,

    find its resolution. 2

    Ans: Refer answer 14.

    Jainam Publication TMPage 36

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    37/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    38/247

    Jainam Solved Paper Electronic Instrumentation

    Examination, April May, 2008

    Q. I.Define Accuracy, Average value, Root mean square value &

    Form factor. 2

    Ans:Refer answer 1.

    Q. II.What are the advantages of Digital instruments over Analog

    instruments? What are the different types of Digital Voltmeters,

    compare their performance. 7

    Ans:Refer answer 17 and 3.

    Q. III.Describe the arrangement of Digital Voltmeter based onDual Slope integration method. What are the advantages of

    integrating type DVM? 7

    Ans:Refer answer 4.

    Q. IV. (a)Define Resolution and Sensitivity as related to digital

    instruments. 2+5

    (b)The lowest range on a 4 digit DVM is 10 mV full scale.

    Determine:

    (i) Sensitivity and resolution of meter

    (ii) How would 0.6789 be displayed on 10 V and 100 V ranges.

    Ans:Refer answer 1 for (a) and 14 for (b).

    Examination, Nov. Dec., 2007

    Q.1 (a)Define sensitivity and resolution of a digital meter.2

    Ans:Refer answer 1.

    (b)Describe the digital meters and its working in detail. 7

    Ans:Refer answer 3.

    (c)Describe the working principle and block diagram of a dual

    slope integrating type digital voltmeter. 7

    Jainam Publication TMPage 38

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    39/247

    Jainam Solved Paper Electronic Instrumentation

    Ans:Refer answer 4.

    (d)A 5 digit voltmeter is used for voltage measurement:

    (i) Find its resolution.

    (ii) How would 13.46 V be displayed on a 10 V range?

    (iii) How would 0.54867 be displayed on 1 V and 10 V ranges.7Ans:Refer answer 16.

    Unit-II Transducers

    Q. 1: - What do you mean by transducer?

    (CSVTU April-May 2008, Nov-Dec 2007)

    Ans:

    A transducer is a device that converts oneform of energyto

    another. Energy types include (but are not limited to)electrical,

    mechanical,electromagnetic(includinglight),chemical,acoustic

    orthermalenergy. Transducers are widely used in measuring

    instruments.

    Q. 2: -Write selection criteria of transducers?Or

    Write down the factors which effect the choice of

    transducers?

    Ans:

    The following is the summary of the factors influencing

    the choice of a transducer for measurement of a physical quantity.

    1.Operating Principle:The transducers are many a timesselected on the basis of operating principle used by them.

    Jainam Publication TMPage 39

    http://en.wikipedia.org/wiki/Form_of_energyhttp://en.wikipedia.org/wiki/Electricityhttp://en.wikipedia.org/wiki/Mechanicshttp://en.wikipedia.org/wiki/Electromagnetismhttp://en.wikipedia.org/wiki/Lighthttp://en.wikipedia.org/wiki/Chemistry#Energyhttp://en.wikipedia.org/wiki/Acousticshttp://en.wikipedia.org/wiki/Heathttp://en.wikipedia.org/wiki/Form_of_energyhttp://en.wikipedia.org/wiki/Electricityhttp://en.wikipedia.org/wiki/Mechanicshttp://en.wikipedia.org/wiki/Electromagnetismhttp://en.wikipedia.org/wiki/Lighthttp://en.wikipedia.org/wiki/Chemistry#Energyhttp://en.wikipedia.org/wiki/Acousticshttp://en.wikipedia.org/wiki/Heat
  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    40/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    41/247

    Jainam Solved Paper Electronic Instrumentation

    11. Usage and Ruggedness:The ruggedness both of

    mechanical and electrical intensities of transducer versus its

    size and weight must be considered while selecting a suitable

    transducer.

    12. Electrical aspects:The electrical aspects that need

    consideration while selecting a transducer include the length

    and type of cable required.

    13. Stability :The transducer should exhibit a high degree

    of stability to be operative during its operation and storage

    life.

    14. Reliability:Reliability should be assured in case of

    failure of transducer in order that the functioning of the

    instrumentation system continues uninterrupted.

    Q. 3: - Discuss transducers characteristics.

    Ans:

    The important characteristics of transducers are

    1.Input characteristics

    i) Type of Input and Operating Range

    ii) Loading Effects

    2.Transfer characteristics

    (i) transfer function

    (ii) error, and

    (iii) response of transducer to environmental influences.

    3.Output characteristics

    i) type of electrical output,

    ii) output impedance,

    iii) useful range.

    Input Characteristics

    a)Type of Input and Operating Range:

    The foremost consideration for the choice of a

    transducer is the input quantity it is going to measure and its

    Jainam Publication TMPage 41

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    42/247

    Jainam Solved Paper Electronic Instrumentation

    operating range. The type of input, which can be any physical

    quantity, is generally determined in advance. However, the choice

    of a particular transducer that depends upon the useful range of

    input quantity ever which the transducer can be used. The useful

    operating range of the transducer may be a decisive factor in

    selection of a transducer for a particular application. The upper

    limit is decided by the transducer capabilities while the lower limit

    of range is normally determined by the transducer error or by the

    unavoidable noise originating in the transducer.

    b)Loading Effects:

    Ideally a transducer should have no loading effect on

    the input quantity being measured. In theory, it is impossible,

    although in practice steps may be taken to reduce the loading

    effects to negligible proportions.

    Transfer Characteristics

    The transfer characteristics of transducers require

    attention of three separate elements,

    Transfer Function:

    Thetransfer functionof a transducer defines a

    relationship between the input quantity and the output. The

    transfer function is,

    q0=f(qi)

    Where,q0andqiare respectively output and input of the

    transducer.

    Thesensitivityof a transducer is defined as the differential

    quotient,

    S=dq0

    dq1

    In general, the sensitivity of transducers is not constant but is

    dependent upon the upon quantityq1.

    Jainam Publication TMPage 42

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    43/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    44/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    45/247

    Jainam Solved Paper Electronic Instrumentation

    The maximum separation due to hysteresis between

    upscale-going and downscale-going indications of a measured

    variable. The delay between the action and reaction of a

    measuring instrument. Hysteresis is the amount of error that

    results when this action occurs.

    The maximum differences in outputs at any measured value

    within the specific range when approaching the point first with

    increasing and then with decreasing input may be termed as

    Hysteresis.

    (2) Dynamic errors are those which produce when input changewith respect to time.

    (3,4) Environmental Response(due to noise and drift and due to

    change in frequency):

    The response of the transducer to environmental

    influences is of a great importance. This is often given insufficient

    attention when choosing the best transducer for a particularmeasurement. This gives rise to results that are not as accurate as

    expected, or, worse, results that are accepted as more accurate

    than they actually are. The performance of the transducer is fully

    defined by its transfer function and errors, provided that the

    transducer is in constant environments and not subject to any

    disturbances like stray electromagnetic and electrostatic fields,

    mechanical shocks and vibrations temperature changes, pressureand humidity changes, changes in supply voltage and improper

    mechanical mountings. If transducers are subjected to the above

    Jainam Publication TMPage 45

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    46/247

    Jainam Solved Paper Electronic Instrumentation

    environmental disturbances, which they are, precautions are

    taken, so that changes in transfer function and resulting errors

    there from do not occur.

    Output Characteristics

    The three conditions in the output characteristics which should

    be considered are,

    (i) type of electrical output,

    (ii) output impedance,

    (iii)useful range.

    Type of Electrical Output:

    The types of output which may be available from the

    transducers may be a voltage, current, impedance or a time

    function of these amplitudes. These output quantities may or may

    not be acceptable to the latter stages of the instrumentation

    system. They may have to be manipulatedi.e.their magnitudes

    changed or they may have to be changed in their format by signalconditioning equipment so as to make them drive the subsequent

    stages of instrumentation system.

    Output Impedance:

    The output impedance, Z0 of a transducer determines to

    the extent the subsequent stages of instrumentation is loaded.

    Ideally, the value of output impedance should be zero if no loadingeffects are there on the subsequent stage. However, the output

    impedance, Zn, cannot made equal to zero and therefore, its value

    should be kept as low as possible to minimize the loading effects.

    The output impedance determines the amount of power

    that can be transferred to the succeeding stages of the

    instrumentation system for a given output signal level. If theoutput impedance is low compared to the forward impedance of

    the system, the transducer has the characteristics of aconstant

    Jainam Publication TMPage 46

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    47/247

    Jainam Solved Paper Electronic Instrumentation

    voltage source(provided the output of the transducer is a voltage),

    while in case the forward impedance is high as compared with the

    output impedance of transducer, it behaves asconstant current

    source.

    When the output impedance of the transducer is equalto that of the following stages of instrumentation system,matching

    takes place and maximum power is transferred from the

    transducer to the succeeding stages. However, it must be

    understood that in case maximum power transfer takes place,

    when the output resistance of transducer, is equal to the

    resistance of the succeeding stages, the efficiency is only 50%.

    Useful Output Range:

    The output range of a transducer is limited at the lower

    end by noise signals which may shroud the desired input signal.

    The upper limit is set by the maximum useful input level. The

    output range can be increased, in some cases, by the inclusion of

    amplifier in the transducer. However, the inclusion of an amplifier

    also increases the noise level and therefore in such situations the

    amplifier may not be of any use at all.

    Q. 4: - Classify transducers.

    (CSVTU April-May 2011, Nov-Dec 2009, April-May 2009, April-

    May 2008)

    Ans:

    The transducers can be classified as follows:

    1.On the basis of transduction form used,

    2.Primary and secondary transducers,

    3.Passive and active transducers,

    4.Analog and digital transducers,

    5.Transducers and inverse transducers.

    On the basis of Transduction:-

    The transducers can be classified on the basis of

    principle of transduction as:

    Jainam Publication TMPage 47

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    48/247

    Jainam Solved Paper Electronic Instrumentation

    I. resistive

    II. inductive

    III. capacitive

    It depends upon how they convert the input quantity into

    resistance, inductance or capacitance respectively. They can beclassified as piezoelectric, thermoelectric, magneto restrictive,

    electro kinetic and optical,

    Primary and secondary transducers:-

    When the input signal is directly sensed by the

    transducer and physical phenomenon is converted into the

    electrical form directly then such a transducer is called the

    primary transducer. For example a thermistor used for the

    measurement of temperature fall in this category. The thermistor

    senses the temperature directly and causes the change in

    resistance with the change in temperature.

    When the input signal is sensed first by some detector or

    sensor and then its output being of some form other than input

    signal is given as input to a transducer for conversion into

    electrical form, then such a transducer falls in the category of

    secondary transducers. For example, in case of pressure

    measurement, bourdon tube is a primary sensor which converts

    pressure first into displacement, and then the displacement is

    converted into an output voltage by an LVDT. In this case LVDT is

    secondary transducer.

    Passive and active transducers:-

    Active transducers:

    They are also known as self-generating type

    transducers. The transducers develop their own voltage or

    current. The energy required for production an output signal is

    obtained from the physical phenomenon being measured.

    Examples: Thermocouples and thermopiles, piezoelectric pick-up,

    photovoltaic cell.

    Jainam Publication TMPage 48

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    49/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    50/247

    Jainam Solved Paper Electronic Instrumentation

    Ans:

    The resistance of a metal conductor is expressed by a

    simple equation that involves a few physical quantities. The

    relationship is given by

    R =

    L

    where, R =Resistance, = Resistivity of conductor materials, Q-m,

    L = Length of conductor, m, and

    A =Cross-sectional area of the conductor, m2.

    Any method of varying one of the quantities involvedin the above relationship can be the designed basis of anelectrical

    resistance transducer.There are a number of ways in which

    resistance can be changed by a physical phenomenon.

    1.Thetranslational and rotational "potentiometers"which work on

    the basis of change in the value of resistance with change in

    length of the conductor can be used for measurement of

    translational or rotary displacements.2."Strain gauges" work on the principle that the resistance of a

    conductor or a semiconductor changes when strained.This

    property can be used for measurement of displacement, force

    and pressure.

    3.The resistivity of materials changes with the change of

    temperature thus causing a change of resistance. This property

    may be used for measurement of"temperature".

    4.In aresistance transduceran indication of measured physical

    quantity is given-bya change in the resistance.It may be

    classified(as discussed above) as follows :

    Mechanically varied resistance Potentiometer

    Thermal resistance change Resistance thermometers

    Resistivity changeResistance strain gauge

    Jainam Publication TMPage 50

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    51/247

    Jainam Solved Paper Electronic Instrumentation

    Q. 6: What are potentiometers, explain their working?

    (CSVTU April-May 2010)

    Ans:

    Potentiometersconvert the linear motion or the angular motion

    of a rotating shaft into changes in resistance.The device is avariable resistor whose resistance is varied by the movement of a

    slider over a resistance element.

    1.Translatory devices have strokes from 2.5 mm to 5 mm.

    2.Rotational devices have full scale from 10 to 60 full turn.

    3.Helipot Potentiometer is there which is combination of both

    translatory type and rotational type.

    The potentiometer shown in Figs. 1 is Linear motion

    potentiometer and Fig 2 Rotary motion potentiometer is form a

    part of the bridge circuit whose output voltage is changed by the

    slider position is shown in fig 3.

    Jainam Publication TMPage 51

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    52/247

    Jainam Solved Paper Electronic Instrumentation

    1.The slider is powered by the mechanical part on which the

    linear displacement or angular measurement are to be made.

    2.Due to arm movement, the slider moves over the resistance

    element and thus shorts out a portion of the resistance.The

    change in resistance in the, potentiometer is then an indication of

    the amount of motion and the direction of movement is indicated

    by whether the resistance is increasing or decreasing.The

    unbalanced voltage is measured directly or fed into an amplifier

    and recorded.

    The output voltage under ideal condition is given by

    e0= (resistance at output terminal/resistance at input terminal)x

    input voltage

    RP(! i

    ! t)

    RP ei

    (!i

    !t)e i

    SensitivityS

    "utput

    input=

    e"

    ! i=

    ei

    !t

    Q. 7: - How can we measure the temperature in industries?

    (CSVTU April-May 2012)

    Jainam Publication TMPage 52

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    53/247

    Jainam Solved Paper Electronic Instrumentation

    Ans:

    Temperature is one of the most widely measured and controlled

    variable in industry, as a lot of products during manufacturing

    requires controlled temperature at various stages of processing.

    A wide variety of temperature transducers and

    temperature measurement systems have been developed for

    different applications requirements. Most of the temperature

    transducers are of Resistance Temperature Detectors (RTD),

    Thermistors and Thermocouples. Of these RTD's and Thermistors

    are passive devices whose resistance changes with temperature

    hence need an electrical supply to give a voltage output. On the

    other hand thermocouples are active transducers and are based

    on the principle of generation of thermoelectricity, when two

    dissimilar metals are connected together to form a junction called

    the sensing junction, an emf is generated proportional to the

    tempera

    ture of the junction.

    Q. 8: -Discuss the following:i) Resistance Temperature Detectors

    ii) Thermistors

    (CSVTU Nov-Dec 2010, April-May 2009)

    Ans:

    i) RTD:

    The resistance of a conductor changes when its

    temperature is changed. This property is utilized for measurement

    of temperature.

    The variation of resistanceR With temperatureT (K)

    can be represented by the following relationship for most of the

    metals as :

    R = R0(1+1T+2T2++ nT

    n+ ...)

    Where, R0= resistance at temperatureT= 0 and 1, 2, n, are

    constants.

    Jainam Publication TMPage 53

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    54/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    55/247

    Jainam Solved Paper Electronic Instrumentation

    coefficient of temperature resistancei.e.their resistance decreases

    with increase of temperature.

    The negative temperature coefficient of resistance can

    be as large as several percent per degree Celsius. This allows the

    thermistors circuits to detect very small changes in temperature

    which could not be observed with an RTD or a thermocouple.

    Thermistors are widely used in applications which

    involve measurements in the range of - 60C to 15C. The

    resistance of thermistors ranges from 0.5 ohm to 0.75 M-ohm.

    Thermistor is a highly sensitive device. The price to be paid off for

    the high sensitivity is in terms of linearity. The thermistor exhibits

    a highly non linear characteristic of resistance versus

    temperature.

    Resistance-Temperature Characteristics of Thermistors:

    The mathematical expression for the relationship

    between the resistance of a thermistor and absolute temperature

    of thermistor is :

    RT2= RT1exp [ (1

    #2 -1

    #1 )]

    where

    RT1=resistance of the thermistor at absolute temperature T1at K,

    RT2= resistance of the thermistor at absolute temperatureT2at K

    = a constant depending Upon the material of thermistor,

    typically 3500 to 4500 K.

    The resistance temperature characteristics of a typical thermistorare given in Fig. The resistance temperature characteristics of Fig.

    show that a thermistor has a very high negative temperature co-

    efficient of resistance, making it an ideal temperature transducer.

    Jainam Publication TMPage 55

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    56/247

    Jainam Solved Paper Electronic Instrumentation

    The characteristics of thermistors are no doubt non-

    linear but a linear approximation of the resistance-temperature

    curve can be obtained over a small range of temperatures. Thus,

    for a limited range of temperature, the resistance of a thermistor

    varies as given by Equation:

    R= R0(1+0).

    Q. 9: - Describe strain gauge in detail. Find the expression

    for gauge factor?

    (CSVTU Nov-Dec 2011, April-May 2010, Nov-Dec 2009)

    Ans:

    The strain gauge is basically a device used for measuring

    mechanical surface strain and is one of the most extensively used

    electrical transducers. Its popularity stems from the fact that it

    can detect and convert force or small mechanical displacements

    into electrical signals. Many other quantities such as torque,

    pressure, weight, and tension etc., which involve effects of force or

    displacement, can also be measured by strain gauges.

    Furthermore, if the mechanical displacements under

    measurement have a time-varying form, such as vibration motion,

    signals with frequencies of up to 100 kHz can be detected or

    measured. The applications of strain gauges may be broadly

    classified into two areas

    Jainam Publication TMPage 56

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    57/247

    Jainam Solved Paper Electronic Instrumentation

    (i)Applications where the gauge measure strain as the primary

    objective of measurement as in the case of stress analysis of

    machines and structures.

    (ii)Applications where measurement of strain is utilized in

    transducers as a measure of another parameter such as load,

    pressure, acceleration, or another force associated variable.

    The working of strain gauge is based on the fact that when

    stress is applied on the metal conductor its resistance changes

    owing to change in length and X-sectional area of the conductor

    (Fig.). Resistance of conductor under stress is also changed due to

    change in resistivity of the conductor, this property called the

    piezo-resistive effect. That is why strain gauges are also called thepiezo-resistive strain gauges.

    If a conductor of length L, area of cross-section A is subjected to

    axial tension, the resistance will change because of change in

    length, area and resistivity of the material.

    Resistance of an unstrained conductor is given by an expression

    R= L /

    Also area of the Wire A= (/4)D2=KD2

    Then R= L /$ % &2

    --------------------------------------------1

    Let under strained conditions resistance of conductor be changed

    by R because of change in length by L, cross-sectional area by

    A and resistivity by p. These quantities can be related with each

    Jainam Publication TMPage 57

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    58/247

    Jainam Solved Paper Electronic Instrumentation

    other by differentiating Resistance expression with respect to

    stress i.e.,

    dR=$ &

    2( % dL+L% d)L(2$& % d&)

    ($ % &2)2 ----------------------------2

    dR=1

    $ % &2 ( % dL+L % d2L(d& /&) )

    dR

    R =dL

    L +d

    - 2d &

    & --------------------------------------3

    Also we know that possions ratio is defined as ratio of lateralstrain to longitudinal strain.

    i.e. =-

    d &

    &

    dL

    L

    -----------------------------------------4

    Putting value of equation 4 in equation 3 we get

    dR

    R =dL

    L +d

    + 2'

    dL

    L

    dR

    R =d

    + (1+2'

    dL

    L ----------------------------------------5

    Dividing both side bydL

    L we get

    dR

    R/

    dL

    L ) = 1+ 2 + (

    d

    /dL

    L ) ------------------------------6

    But we know that Gauge factor is defined as the ratio of per unit

    change in resistance to the per unit change in length.

    i.e. Gf=

    dR

    R/ dLL ) ----------------------------------------7

    Jainam Publication TMPage 58

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    59/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    60/247

    Jainam Solved Paper Electronic Instrumentation

    Stress=

    =1000 -g/*m2

    Modulus of Elasticity = Y = 2106

    -g /*m2

    To Calculate:

    1)% change in resistance

    2)Poissons ratio

    Solution:

    Y =

    stress

    str(in=

    1000

    L

    L

    L

    L =

    1000

    . =

    1000

    2106=5104

    +f=2=

    R

    R

    L

    L

    =

    R

    R

    5104

    RR

    =10104=( R

    R)=10106

    (Ans)

    +f=2=1+2/

    /=0.5 (Ans)

    Q. 12: - A strain gauge is bounded to a beam of 0.1 m long

    and has a cross-sectional area 4 cm2. Youngs modulus for

    steel is 207 GN/m2. The strain gauge has an unstrained

    resistance of 240 and a gauge factor of 22. When a load

    is applied, the resistance of gauge changes by 0.013 .

    Calculate the change in length of the steel beam and the

    amount of force applied to the beam.

    (CSVTU Nov-Dec 2010)

    Ans:

    Jainam Publication TMPage 60

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    61/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    62/247

    Jainam Solved Paper Electronic Instrumentation

    The supply to bridge is of 6 volt and G= 2. The applied

    stress is 300 MN/m2and Youngs Modulus of elasticity of

    cantilever material is 60 gN/m2. Determine

    (a)Strain

    (b)Change in resistance

    (c)The output voltage 7

    (CSVTU April-May 2009)

    Ans:

    Given:

    Resistance of strain gauges. R = 150

    Fixed resistances. R = 150

    Applied stress. P = 300 x 106N/m2

    Modulus of elasticity, E = 60 x 109

    N/m2

    Calculation:

    Strain

    Change in resistance

    The output voltage

    Solution:

    Jainam Publication TMPage 62

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    63/247

    Jainam Solved Paper Electronic Instrumentation

    One of the gauges, sayR1, is subjected to tensile stress and the

    other to compressive stress soR1 increases while R3decreases.

    1.Strain = P/) = (300 x 106N/m2)/(60 x 109N/m2) = 5xl0-

    3

    2.Change in the resistance of strain gauges,

    R = G x x R = 2 x 5xl0 -3x 150 = 1.5

    3.Output voltage when the detector connected across output

    terminals is of infinite resistance

    Vout=V

    2G x =

    6

    2 x2 x 5xl0-3= 30 mV

    4.Output voltage when the detector connected across output

    terminals is of 600

    VOL=[1 /((/Rm)+1)] Vout=

    [1 /((150 /600)+1)] 30 =

    24 mV

    5.Detector current. I = VOL / Rm=24mV/600 =40 n A

    6.Current through strain gauges = V/2R = 6/(2x150)= 20 mA

    7.Change in resistance R2for restoration of balance

    R2 =R4(R1+ R 1)

    R3 R3 =150 (150+1.5)

    1501.5 = 3.03

    Q. 14: - Explain any inductive transducer?

    (CSVTU April-May 2012, Nov-Dec 2011, April-May 2011, Nov-

    Dec 2008, April-May 2008, Nov-Dec 2007)

    Ans:

    Linear-variable-differential transformer (LVDT)

    LVDT is a passive inductive transducer and is commonly employed

    to measure force (or weight, pressure and acceleration etc. which

    depend on force) in terms of the amount and direction of

    displacement of an object.

    Jainam Publication TMPage 63

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    64/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    65/247

    Jainam Solved Paper Electronic Instrumentation

    3.Similarly, when the movable core moves towards coil S1, E1> E2

    andV0= E1-E2and is inphase with E1.

    4.Thus, from above discussion, we find that the magnitude ofV0

    is afunction of the distance moved by the coreand itspolarity or

    phaseindicates as to in which direction it has moved.5.If core is attached to moving object, themagnitudeofV0gives the

    position of that object.

    Advantages:

    Jainam Publication TMPage 65

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    66/247

    Jainam Solved Paper Electronic Instrumentation

    1.It gives a high output and therefore many a times there is no

    need for intermediate amplification devices.

    2.The transducer possesses a high sensitivity as high as 40

    V/mm.

    3.It shows a low hysteresis and hence repeatability is excellent

    under all conditions.

    4.Most of the LVDTs consume a power of less than 1 W.

    5.Less friction and less noise (due to absence of sliding contacts).

    6.These transducers can usually tolerate a high degree of shock

    and vibration without any adverse effects.

    7.It can operate over a temperature range from -265C to 600C.

    8.It is available in radiation-resistant design for operation in

    nuclear reactors.

    Disadvantages:

    1.These transducers are sensitive to stray magnetic fields but

    shielding is possible. This is done by providing magnetic shields

    with longitudinal slots.

    2.Relatively large displacements are required for appreciabledifferential output.

    3.The receiving instrument must be selected to operate on A.C.

    signals or demodulator network must be used if a D.C. output

    is required.

    4.Several times, the transducer performance is affected by

    vibrations.

    5.The dynamic response is limited mechanically by the mass ofcore and electrically by the frequency of applied voltage. The

    frequency of the carrier should be at least ten times the highest

    frequency component to be measured.

    Applications:

    1.Measurement of material thickness in hot strip or slab steel

    mills.2.In accelerometers.

    Jainam Publication TMPage 66

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    67/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    68/247

    Jainam Solved Paper Electronic Instrumentation

    Capacitance C =1

    d =1r 10

    d

    10 =Free space permittivity

    1r

    =Relative Permittivity

    1 =Permittivity of medium

    A=Overlapping Area of Plate

    d = Distance between two plates

    The displacement is measured by measuring the change in

    capacitance brought about by:

    1.Change in area

    2.Change in distance between the plates

    Differential capacitor System:

    The Differential capacitor System is used for the measurement of

    the linear displacement.

    In this system two capacitance C1 and C2 are taken and an

    alternating voltage E is applied to both the capacitor.

    The two capacitor are identical to each other and placed one over

    the other as shown in figure1,at this time

    C=C1 =C2 =1

    d

    Jainam Publication TMPage 68

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    69/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    70/247

    Jainam Solved Paper Electronic Instrumentation

    Advantages:

    1.Require extremely small force for operation (hence very useful

    for use in small systems).

    2.Extremely sensitive.3.Require small power for operation,

    4.High input impedance; therefore, loading effects are minimum.

    5.Frequency response is good.

    6.A resolution of the order of 2.5 x 10-3mm can be obtained.

    7.Can be used for applications where stray magnetic fields render

    the inductive transducers useless.

    Disadvantages.

    1.The metallic parts must be insulated from each other. The

    frames must be earthed to reduce the effects of stray

    capacitances.

    2.They show non-linear behaviour several times on account of

    edge effects;guard ringsmust be used to eliminate this effect.

    3.The cable connecting the transducer to the measuring point is

    also a source of error. The cable may be source of loading

    resulting in loss of sensitivity. Also loading makes the low

    frequency response poor.

    Uses of the capacitive transducers.

    1.To measureboth linear and angular displacements.

    2.Tomeasure force and pressure.

    3.Used as pressure transducers in all those cases where thedielectric constant of a medium changes with pressures.

    4.To measure humidity in gases.

    5.Used in conjunction with mechanical modifiers for

    measurement ofvolume, density, weight, input leveletc.

    Q. 17: - A capacitive transducer uses two quartz diaphragms

    of area 600 mm2separated by a distance of 2.5 mm. A

    pressure of 8 105N/m

    2. When applied to the top diaphragm

    causes a deflection of 0.5 mm. The capacitance is 400 10-

    Jainam Publication TMPage 70

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    71/247

    Jainam Solved Paper Electronic Instrumentation

    12F when no pressure is applied to diaphragms. Determine

    the value of capacitance after the application of pressure.

    (CSVTU April-May 2008)

    Ans:

    Suppose0

    1 and0

    2 are respectively the values of capacitance

    before and after application of pressure. Letd

    1 andd2be the

    values of distance between the diaphgrams for corresponding

    pressure conditions.

    01=

    1

    d1 and0

    2=

    1

    d2

    01

    02=

    d1

    d2

    If,d

    1=2.5

    thend

    2=2.50.5=2.0mm

    Value of capacitance after application of pressure

    02=

    4001012

    2.5

    2=5001012

    Q. 18: - A capacitive transducer circuit used for

    measurement of linear displacement. Suppose a flat

    frequency response with an amplitude ratio within 5% is

    required down to a frequency range of 20 Hz. What is the

    minimum allowable value of time constant?

    Calculate phase shift at this frequency. Area of plates is

    300mm2& the distance between plates is 0.125 mm.

    calculate the value of series resistance R. what is the

    amplitude ratio at 5 Hz with the above time constant?

    (CSVTU April-May 2012)

    Ans:

    For a flat response with 5%, the amplitude ratio is:

    Jainam Publication TMPage 71

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    72/247

    Jainam Solved Paper Electronic Instrumentation

    M=1-0.05 = 0.95

    M = 0.95 = 1/{1+(1/)2}1/2

    So, = 24.2 * 10-3sec = 24.2 msec

    Phase shift = (/2)-tan-1 = 18.20

    Capacitance (C) = A/d = 21.24* 10-12

    Therefore series resistance R = /C = 1140 M

    Amplitude ratio at 5 Hz = 0.605

    Q. 19: - A capacitance transducer of two parallel plates of

    overlapping area of

    5104

    m2

    is immersed in water. Thecapacitance C has been found to be 9.5 pF. Calculate the

    separation d between the plates & the sensitivity, S. Given1r for water = 81;

    10=8.854p/m %

    (CSVTU Nov-Dec 2011)

    Ans:

    C = 0rA/d

    On substituting the values we get the value of displacement as

    d=37.75 mm

    Sensitivity = dC/dd = - 0rA/d2= 0.025 * 10-8F/m

    Q. 20: Discuss piezoelectric transducer and derive the

    expression for output voltage? (CSVTU Nov-Dec 2009)

    Ans:A piezoelectric material is one in which an electric potential

    appears across certain surfaces of a crystal if the dimensions of

    the crystal are changed by the application of a mechanical force.

    This potential is produced by the displacement of charges. The

    effect is reversible i.e. conversely, if a varying potential is applied to

    the proper axis of the crystal, it will change the dimension of the

    crystal thereby deforming it. This effect is known as piezoelectriceffect. Element exhibiting piezoelectric qualities are called as

    electro resistive elements.

    Jainam Publication TMPage 72

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    73/247

    Jainam Solved Paper Electronic Instrumentation

    The piezoelectric effect can be made to respond tomechanical deformations of the material in many different modes.

    The modes can be thickness expansion, thickness shear and face

    shear.

    The piezoelectric effect is direction sensitive. The tensile

    force produces a voltage of one polarity while a compressive force

    produces a voltage of opposite polarity. The magnitude and

    polarity of the induced surface charges are proportional to the

    magnitude and direction of the applied force F.

    A piezoelectric element generates a charge and this charge

    appears as a voltage across the electrodes. i.e.

    E=Q/C

    We know that charge Q= d*F Coulamb (i)

    Where, d= charge sensitivity of the crystal (C/N)

    And F=force applied in N.

    F= AE t/t

    Where A= cross sectional area and E= Youngs modulus

    Also, E=stress/strain = (F/A)/( t/t)

    So, from equation (i) Q = dAE t/t

    Jainam Publication TMPage 73

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    74/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    75/247

    Jainam Solved Paper Electronic Instrumentation

    The voltage output of the transducer under no load conditions, is

    thereforeE0.Under conditions of load: -

    Impedance of load ZL=RL

    1+(23 0LRL )

    Total impedance of circuit Zt=1

    230p +RL

    1+23 RL 0L

    Voltage across the load ELwill be

    EL=4L

    4t )

    0

    EL=23 0pRL

    1+23 (0L+0p)RL )

    0

    The Magnitude of voltage across the load is:

    EL=

    p+0L

    02RL

    2

    1+32

    3 0pRL

    We know that E0=

    d

    0p

    Then EL=

    p+0L0

    2RL2

    1+32

    3 0pRL

    Jainam Publication TMPage 75

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    76/247

    Jainam Solved Paper Electronic Instrumentation

    EL=

    p+0L0

    2RL2

    1+32

    3 RL

    At medium and High Frequencyp+0L

    0

    32

    EL=0p

    (0L+0p) )

    0

    Q. 22: - A Barium titanate pickup has the dimension of 5 mm

    5 mm 1.25 mm. The force acting on it is 5 N. The

    charge sensitivity of Barium titanate is 150 pc/N & its

    permittivity is 12.5 10

    9

    F/m. If the modulus of elasticityof Barium titanate is 120 10

    6N/m

    2. Calculate strain,

    charge & capacitance.

    (CSVTU April-May 2012, Nov-Dec 2011, April-May 2011, April-

    May 2010)

    Ans:

    Given:

    1.Dimension of barium titanate = 5mm 5mm 1.25mm

    2.Force F =5N

    3.Charge Sensitivity d =150pC/N

    4.Permittivity = 12.5109

    F/m

    5.Modulus of elasticity= 12 x 106N/m2

    To Calculate:

    Jainam Publication TMPage 76

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    77/247

    Jainam Solved Paper Electronic Instrumentation

    1.Strain

    2.Charge

    3.Capacitance

    Solution:

    Area of Plate (A) = 5mm 5mm = 25106

    m2

    Pressure (P) = F/A=5/ 25106

    =0.2 MN/m2

    Voltage sensitivity g = d/ = (1501012

    )/12.510

    9

    )

    = 12103

    Vm/N

    Voltage Generated Eo= g t P =12 x 10-3x 1.25 x 10-3x 0.2 x 106

    =3 V

    Strain= Stress/Youngs Modulus = (0.2103

    )/1210

    6

    )

    = 0.0167

    Charge Q= d F = 1501012 )* 5 ) = 750 pC

    Capacitance C= Q/Eo= (750 pC)/3 =250 pF

    Q. 23: - A piezo electric transducer has a capacitance of

    1000 pF and a change sensitivity of 40 10-3C/m. The

    connecting cable has a capacitance of 300 pF while the

    oscilloscope used for readout has a read-out input

    resistance of 1 m with a parallel capacitance of 50 pF.

    (i) What is the high frequency sensitivity (V/m) of the

    entire measuring system?

    (ii) What is the sensitivity (V/m) of the transducer alone?

    (iii)What is the lowest frequency that can be measured

    with 5% amplitude error by the entire system?

    Jainam Publication TMPage 77

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    78/247

    Jainam Solved Paper Electronic Instrumentation

    (iv) What is the value of an external shunt capacitance

    that can be connected in order to extend the range of

    5% error down to 10 Hz? (CSVTU Nov-Dec 2008)

    Ans:

    Given:

    1.A piezo electric transducer has a capacitance of 1000 pF

    2.a change sensitivity of 4010-3C/m.

    3.capacitance of 300 pF

    4.input resistance of 1 m

    5.a parallel capacitance of 50 pF.

    Calculate:

    (v) What is the high frequency sensitivity (V/m) of the entire

    measuring system?

    (vi) What is the sensitivity (V/m) of the transducer alone?

    (vii)What is the lowest frequency that can be measured with

    5% amplitude error by the entire system?(viii)What is the value of an external shunt capacitance that

    can be connected in order to extend the range of 5% error

    down to 10 Hz

    Solution:

    1.Charge sensitivity of transducer Kq = 4010

    3 C/m

    Capacitance of transducer Cp = 100010

    12 F

    Hence Sensitivity Of Transducer K = Kq/Cp = (4010

    3 )/(1000

    1012

    )

    Jainam Publication TMPage 78

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    79/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    80/247

    Jainam Solved Paper Electronic Instrumentation

    5.High Frequency sensitivity with external capacitance = 40

    103

    /48.380 = 827kV/m

    Q. 24: - Explain photovoltaic cells?

    Ans:

    The photo-voltaic or solar cell, produces an electrical

    current when connected to a load. Both silicon (Si) and selenium

    (Se) types are known for these purposes.

    Multiple unit silicon photo-voltaic devices may be used for

    sensing light in applications such as reading punched cards in the

    data processing industry.

    Gold-doped germanium cells with controlled spectral

    response characteristics act as photo-voltaic devices in the infra-

    red region of the spectrum and may be used as infra-red

    detectors.

    The silicon solar cell converts the radiant energy of the

    sun into electrical power. The solar cell consists of a thin slice of

    single crystal P-type silicon, up to 2 cm2into which a very thin

    (0.5 micron) layer of N-type material is diffused. The conversion

    efficiency depends on the spectral content and intensity of

    illumination

    Jainam Publication TMPage 80

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    81/247

    Jainam Solved Paper Electronic Instrumentation

    Q. 25: - What do you mean by Hall Effect? ExplainHall Effect

    transducers?

    (CSVTU April-May 2011, Nov-Dec 2008)

    Ans:

    When a current carrying conductor is placed in a

    magnetic field, a transverse effect is noted. This effect is called

    Hall Effect (discovered by Hall in 1879). Hall found that: "When a

    magnetic field is applied at right angles to the direction of electric

    current an electric field is set up which is perpendicular to both

    the direction of electric current and the applied magnetic field".

    In other words:

    "When any specimen carrying a current I is placed in the

    transverse magnetic field B, then an electric field E is induced in

    the specimen in the direction perpendicular to both I and B. The

    phenomenon is known as Hall effect".

    The principle of working of a Hall Effect transducer is

    that if a strip of conducting material carries a current in the

    presence of a transverse magnetic field, a difference of potential is

    produced between the opposite edges of the conductor. The

    magnitude of the voltage depends upon the current, the strength

    of magnetic field and the property of the conductor called Hall

    Effect. The Hall Effect is present in the metals and

    Jainam Publication TMPage 81

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    82/247

    Jainam Solved Paper Electronic Instrumentation

    semiconductors in varying amounts, depending upon the

    densities and motilities of carriers.

    When the transverse magnetic field passes through the

    strip, an output voltage across the output leads. This voltage is

    proportional to the current and the field strength.

    The output voltage ) H

    $6 7 B

    #

    Where, KH= Hall Effect coefficient

    T = thickness of strip

    I = current

    B = flux density

    Thus, the voltage produced may be used for measurement

    of either the current or the magnetic field strength B.

    Hall effect transducers are the transducers in which Hall

    effect is utilized to measure various electrical or non-electrical

    quantities.

    Commercial Hall effect transducers are made fromgermanium or other semiconductor materials.

    Q. 26: - Write some applications of Hall Effect transducers?

    Ans:

    The following are the applications of Hall Effect transducers:

    1. Displacement measurement:

    Hall Effect transducer may be used to measure a linear

    displacement or to locate a structural element is cases where it ispossible to change the magnetic field strength by variation in the

    geometry of a magnetic structure.

    2. Current measurement:

    Hall Effect transducer can be used to measure current in a

    conductor without interrupting the circuit and without making

    electrical connection between the conductor circuit and the meter.

    When a D.C. or A.C. current flows through the conductor,

    it sets up a magnetic field around. This magnetic field is

    Jainam Publication TMPage 82

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    83/247

    Jainam Solved Paper Electronic Instrumentation

    proportional to the current. A Hall effect transducer is insert it in

    a slotted ferromagnetic tube which acts as a magnetic

    concentrator. The voltage produce* at the output terminals is

    proportional to the magnetic field strength and hence is

    proportion to the current, flowing through the conductor.

    3. Magnetic flux measurement:

    The magnetic flux can be measured by using Hall Effect

    transducer. Here, i semiconductor plate is inserted into the

    magnetic field which is to be measured. Tin magnetic lines of force

    are perpendicular to the semiconductor. The transducer gives an

    output voltage which is proportional to the magnetic field intensity

    (B).4. Fluid level measurement:

    Hall Effect sensors can be used as position, displacement

    and proximity sensors if object is being sensed with a small

    permanent magnet. Such a sensor can be used to determine the

    level of fuel in an automobile fuel tank.

    Q. 27: - Describe semiconductor photo-diode?(CSVTU Nov-Dec 2010, Nov-Dec 2007)

    Ans:

    The photodiode incorporatesa P and an N type - layer. The system

    has the electrical characteristics of a rectifier. Radiation directed

    in the vicinity of the PN junction and causes a flow of current. Fig.

    shows the circuit of a photodiode.

    Jainam Publication TMPage 83

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    84/247

    Jainam Solved Paper Electronic Instrumentation

    The photodiode is reverse biased. The reverse biased

    saturation current is dependent upon the intensity of the incident

    light.

    Fig. shows the typical characteristics of a photodiode.

    The photodiode is very useful for applications where the

    space is restricted. The effective area of a photodiode is about 0.2

    mm2and it has a pinhead (serving as electrode) of a diameter of

    0.5 mm.

    The photocurrent versus light relationship is linear over

    a wide range. In order to maintain the linearity the bias voltage

    should be kept constant. From Fig. it is clear that the output

    resistanceR = AV/AI,is very high and is of the order of tens of

    M. The d.c. resistance,V/I,is the diode leakage resistance and

    that too is very high. This d.c. resistance depends upon the lightintensity.

    The frequency response of a photodiode is largely

    dependent upon the intrinsic capacity which is typically 2 pF for a

    reverse bias of - 10 V.

    The cut off frequency is given byfc= 1/2RLCwhereC

    is capacitance of photodiode and RLis the load resistance. The cut

    off frequency is of the order of MHz. Even in dark there is alwaysleakage current of the photodiode and this current is known as

    dark current:The dark current doubles about every 10C increase

    Jainam Publication TMPage 84

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    85/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    86/247

    Jainam Solved Paper Electronic Instrumentation

    the photodiode is varied. The highest mode that can be

    reproduced is thus dependent upon the response of the

    photodiode as well as the film speed, the width of slit and the

    definition of photographic emulsion.

    4.Photodiodes can be used as detectors of modulated light in

    optical communication systems.

    5.The photodiodes can be used in switching circuits as they have

    a fast response time. A switching circuit using a photodiode is

    the transistor is normally 'ON* due to the bias resistor. When

    the photodiode is illuminated, the base current is reduced

    taming the transistor OFF.

    Q. 28: Explain phototransistor? (CSVTU Nov-Dec 2007)

    Ans:

    A phototransistor is a normal transistor in which the

    envelopeenclosing the junction is transparent to allow light to fall

    on the base emitter junction . At any PN junction hole-electron

    pairs are generated when light fall on the junction, so that any

    light falling on the base-emitter junction, produces a current

    which is amplified by transistor action, making the device way

    sensitive.

    Jainam Publication TMPage 86

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    87/247

    Jainam Solved Paper Electronic Instrumentation

    Illumination of the central region causes the release of

    electron hole pairs. This lowers the barrier potential across both

    junctions, causing an increase in the flow of electrons from the

    left region into the centre region and on to the right region.The

    sensitivity of a photo diode can be increased by as much as 100

    times bv adding a junction, resulting in an NPN device.

    The advantages of the phototransistor are: low power

    consumption, small size, immediate operation on switching on,

    low voltage operation and long life.

    A phototransistor gives a high gain. These transistor digital

    applications because of the small rise and fall times. The rise

    time, which represents the response to dark-to-light irradiance

    is about 1 microseconds and the fall time which represents

    light-to-dark light irradiance is about 10 microseconds.

    Q. 29: -What are frequency generating transducers?

    Ans:

    PRESSURE INDUCTIVE TRANSDUCER

    A simple, arrangement, where in a change in the inductance of a

    sensing element is produced by a pressure change, is given in Fig.

    below

    Jainam Publication TMPage 87

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    88/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    89/247

    Jainam Solved Paper Electronic Instrumentation

    angle measurement. All these devices work essentially on the

    same principle that is of a rotating transformer. A Synchro

    appears like an AC motor consisting of a rotor and a stator.

    Synchro's are normally used in control system, but have

    properties that can be used in instrumentation also.

    A Synchro can be an angular position transducer working

    on inductive principle, wherein a variable coupling between

    primary and secondary winding is obtained by changing the

    relative orientation of the windings.

    Internally, most synchro's are similar in construction.

    They have a rotor with one or three windings capable of revolvinginside a fixed stator. There are two common types of rotors, the

    salient poleand thewound rotor.

    The primary winding is a single phase winding wound on

    a rotor made of laminations. The connection to the rotor windings

    are made through precision slip rings.

    The stator has a 3-phase winding with the windings of the

    3-phase displaced by 120o. The Synchro may be viewed as a

    variable coupling transformer. A Synchro is also called asSelsyn.

    Jainam Publication TMPage 89

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    90/247

    Jainam Solved Paper Electronic Instrumentation

    Synchro systems consists of two or more interconnected

    Synchro. They are grouped or connected together according to the

    purpose to be used.

    A Synchro system formed by interconnection of the

    devices called the Synchro transmitter and Synchro controltransmitter is perhaps the most widely med error detector in

    feedback control system. It measures and compares two angular

    displacements and its output voltage is approximately linear with

    angular displacement.

    The conventional Synchro transmitter (TX) uses a salient

    pole rotor with sleeved slot When an ac excitation voltage is

    applied to the rotor, the resultant current produces a magnetic

    field and by transformer action induces voltage in the stator coils.

    The effective voltage induced in any stator coil depends upon die

    angular position of the coil axis with respect to the rotor axis

    (when the coil voltage is known, the induced voltage at any

    angular displacement can be determined).

    Q. 31: - Discuss the method for measurement of angularvelocity?

    (CSVTU Nov-Dec 2009, April-May 2009)

    Ans:

    The measurement of angular velocity is more prominent than that

    of linear velocity. In many cases the only way for measuring linear

    velocity is by its conversion into an angular velocity.

    The main problem with linear velocity measurement is in use of a

    fixed reference and in detection in case of moving body travelling

    over a long distance. The various devices used in measurement of

    angular velocity are described below.

    1.DC Generator Tachometer.

    This generator consisting of a small armature which is coupled to

    the shaft of the machine whose speed is to be measured.

    It is an ordinary miniature de generator consisting of a smallarmature rotating in a constant magnetic field.

    Jainam Publication TMPage 90

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    91/247

  • 8/10/2019 PAPER EI(2) by Pankaj Sir

    92/247

    Jainam Solved Paper Electronic Instrumentation

    and an aluminum disc facing the poles of the permanent magnet,

    the disc being mounted on the shaft carrying the instrument

    pointer.

    3. Reluctance pulse pick-ups:This transducer is very suitable for the measurement of shaft

    speed and liquid flow. It is based on the principle that if the field

    of any magnet is varied momentarily by the motion of an external

    magnetic body near it, a voltage pulse is generated at the coil of

    the magnet because of the change in flux surrounding the coil.

    The transducer consists of a permanent magnet on which a coil is

    wound. The output voltage depends upon the rate of change of

    magnetic flux and the number of turns. Flux depends upon theclearance between the pick up and the actuating medium, the rate

    of movement and size of the actuating medium. The output voltage

    is inversely proportional to the distance between the head of the

    pick up and the actuating medium.

    The pick up is actuated by the teeth of a gear or blades of

    turbines. In rpm measurement, the pick up is placed near the

    teeth of a gear. Th