    Dealer: This manual MUST be given to the enduser.User: BEFORE using this product, read this

    manual and save for future reference.

    Operators Manual

    For more information regardingInvacare products, parts, and services,

    please visit www.invacare.com

    Perfecto2

    SeriesOxygen Concentrators without SensO2,

    Oxygen Concentrators with SensO2

    HomeFill Compatible

    Model IRC5PO2AWModel IRC5PO2VAW

    Model IRC5PAW

    English . . . . . . . . . . . . . . . . . 1French . . . . . . . . . . . . . . . . 34Italian . . . . . . . . . . . . . . . . . 66

    Spanish. . . . . . . . . . . . . . . . 97Dutch. . . . . . . . . . . . . . . . 129Greek . . . . . . . . . . . . . . . 161Portuguese . . . . . . . . . . . 195

    0

    434

  • 7/24/2019 Perfecto2 Manual de Uso

    2/228

    WARNING

    Perfecto2Series 2 Part No 1163146

    WARNINGDONOTusethisproductoranyavailableoptionalequipment

    withoutfirstcompletelyreadingandunderstandingtheseinstructionsandanyadditionalinstructionalmaterialsuchasownersmanuals,servicemanualsorinstructionsheetssuppliedwiththisproductoroptionalequipment.Ifyouareunabletounderstandthewarnings,cautionsorinstructions,contactahealthcareprofessional,dealerortechnicalpersonnelbeforeattemptingtousethisequipment otherwise,injuryordamagemayoccur.

    ACCESSORIES WARNINGInvacareproductsarespecificallydesignedandmanufacturedforuseinconjunctionwithInvacareaccessories.Accessoriesdesignedbyothermanufacturers

    havenotbeentestedbyInvacareandarenotrecommendedforusewithInvacareproducts.

    ACCESSORIESTherearemanydifferenttypesofhumidifiers,oxygentubing,cannulasandmasksthatcanbeusedwiththisdevice.You

    shouldcontactyourlocalhomecareproviderforrecommendationsonwhichofthesedeviceswillbebestforyou.Theyshouldalsogiveyouadviceontheproperusage,maintenance,andcleaning.NOTE:Updatedversionsofthismanualcanbefoundatwww.invacare.com.

  • 7/24/2019 Perfecto2 Manual de Uso

    3/228

    TABLE OF CONTENTS

    Part No 1163146 3 Perfecto2Series

    ACCESSORIES ..................................................2

    SPECIAL NOTES ...............................................4

    LABEL LOCATION ............................................6

    SECTION 1GENERAL GUIDELINES ..................7

    Radio Frequency Interference.......................................................... 8

    SECTION 2FEATURES .................................. 10

    SECTION 3HANDLING ................................. 11

    Unpacking............................................................................................11

    Inspection............................................................................................11Storage.................................................................................................11

    SECTION 4TECHNICAL DESCRIPTION .......... 12

    SECTION 5TYPICAL PRODUCT

    PARAMETERS ............................................... 13

    SECTION 6OPERATING INSTRUCTIONS ....... 17

    Introduction........................................................................................17Select a Location ...............................................................................17

    Set Up ..................................................................................................18

    Flowrate...............................................................................................21

    SensO2Oxygen Purity Indicator - Model IRC5PO2AW &

    IRC5PO2VAW .........................................................................................22

    Initial Startup of the Concentrator ...............................................22

    Explanation of Indicator Lights - Model IRC5PAW ..................23

    Elapsed Time Meter (Hour Meter) ...............................................24

    SECTION 7MAINTENANCE .......................... 25

    Routine Maintenance........................................................................25

    SECTION 8TROUBLESHOOTING GUIDE ......... 28

    SECTION 9OPTIONAL ACCESSORIES ............ 30

    RECYCLING INFORMATION ........................... 31

    LIMITED WARRANTY .................................... 32

  • 7/24/2019 Perfecto2 Manual de Uso

    4/228

    SPECIAL NOTES

    Perfecto2Series 4 Part No 1163146

    SPECIAL NOTESSignalwordsareusedinthismanualandapplytohazardsorunsafepracticeswhichcouldresultinpersonalinjuryorpropertydamage.Refertothefollowingtablefordefinitionsofthesignalwords.

    NOTICETheinformationcontainedinthisdocumentissubjecttochangewithoutnotice.

    DANGERDO NOT SMOKE while using this device. Keep all

    matches, lit cigarettes or other sources of ignition out ofthe room in which this product is located and away fromwhere oxygen is being delivered.

    NO SMOKING signs should be prominently displayed.Textiles and other materials that normally would notburn are easily ignited and burn with great intensity inoxygen enriched air. Failure to observe this warning can

    result in severe fire, property damage and cause physicalinjury or death.

    SIGNAL WORD MEANING

    DANGER

    Danger indicates an imminently hazardous situationwhich, if not avoided, will result in death or seriousinjury.

    WARNING

    Warning indicates a potentially hazardous situationwhich, if not avoided, could result in death orserious injury.

    CAUTION

    Caution indicates a potentially hazardous situationwhich, if not avoided, may result in propertydamage or minor injury or both.

  • 7/24/2019 Perfecto2 Manual de Uso

    5/228

    SPECIAL NOTES

    Part No 1163146 5 Perfecto2Series

    CAUTION

    Caution:Statutorylawcanrestrictthisdevicetosalebyoron

    orderofaphysician,oranyotherpractitionerlicensedbythelawofthegovernmentalagencyinwhichhe/shepracticestouseorordertheuseofthisdevice.

    Invacarerecommendsanalternatesourceofsupplementaloxygenintheeventofapoweroutage,alarmconditionormechanicalfailure.Consultyourphysicianorequipmentproviderforthetypeofreservesystemrequired.

    Thisequipmentistobeusedasanoxygensupplementandisnotconsideredlifesupportingorlifesustaining.

  • 7/24/2019 Perfecto2 Manual de Uso

    6/228

    LABEL LOCATION

    Perfecto2Series 6 Part No 1163146

    LABEL LOCATION

    Serial Number Label islocated on the resonator

    intake assembly

    Specification Label is

    located on the backnear the bottom

    erfecto

    FOR OPERATING INSTRUCTIONS or ALARMS (See Operators Manual)

    DANGER - RISK OF FIRENO SMOKING - Keep ALL sources of ignition out of the room in which this product islocated and away from areas where oxygen is being delivered. Textiles, oil and othercombustiles are easily ignited and burn with great intensity in oxygen enriched air.

    DANGER - ELECTRIC SHOCKDO NOT remove cover. Refer servicing to qualified service personnel.

    Flowrates below 0.5 L/min will trigger Low Flow Alarm (rapid audible alarm).

    Set Flowmeter at 2.5 L/min or less when filling HomeFill cylinders.

    HomeFillCompatible

  • 7/24/2019 Perfecto2 Manual de Uso

    7/228

    SECTION 1GENERAL GUIDELINES

    Part No 1163146 7 Perfecto2Series

    SECTION 1GENERAL GUIDELINESInordertoensurethesafeinstallation,assemblyandoperationofthePerfecto2concentratortheseinstructionsMUSTbefollowed.

    WARNINGThissectioncontainsimportantinformationforthesafeoperationanduseofthisproduct.

    DANGERRiskofelectricshock.DONOTdisassemble.Referservicingtoqualifiedservicepersonnel.Nouserserviceableparts.TOREDUCETHERISKOFBURNS,ELECTROCUTION,FIREORINJURYTOPERSONS.Avoidusingwhilebathing.Ifcontinuoususageisrequiredbythephysiciansprescription,theconcentratorMUSTbelocatedinanotherroomatleast2.5m(7ft)fromthebath.DONOTcomeincontactwiththeconcentratorwhilewet.DONOTplaceorstoreproductwhereitcandropintowaterorotherliquid.DONOTreachforproductthathasfallenintowater.UnplugIMMEDIATELY.Iftheconcentratorhasadamagedcordorplug,ifitisnotworkingproperly,ifithasbeendroppedordamaged,ordroppedintowater,callqualifiedtechnicianforexaminationandrepair.Aspontaneousandviolentignitionmayoccurifoil,greaseorgreasysubstancescomeincontactwithoxygenunderpressure.ThesesubstancesMUSTbekeptawayfromtheoxygenconcentrator,tubingandconnections,andallotheroxygen

    equipment.DONOTuseanylubricantsunlessrecommendedbyInvacare.Avoidcreationofanysparknearmedicaloxygenequipment.Thisincludessparksfromstaticelectricitycreatedbyanytypeoffriction.

  • 7/24/2019 Perfecto2 Manual de Uso

    8/228

    SECTION 1GENERAL GUIDELINES

    Perfecto2Series 8 Part No 1163146

    Operating InformationForoptimumperformance,Invacarerecommendsthateachconcentratorbeonandrunningforaminimumof30minutes

    atatime.Shorterperiodsofoperationmayreducemaximumproductlife.

    Keeptheoxygentubing,cord,andunitoutfromundersuchitemsasblankets,bedcoverings,chaircushions,clothingandawayfromheatedorhotsurfaces,includingspaceheaters,stovesandsimilarelectricalappliances.

    DONOTmoveorrelocateconcentratorbypullingonthepowercord.

    NEVERdroporinsertanyobjectorliquidintoanyopening.

    InvacarerecommendsthatCrushProofoxygentubingbeusedwiththisproductandnotexceed15.2m(50ft)inlength.

    Therearenouserserviceableparts.Thisdoesnotinclude

    normalmaintenanceitems.Seemaintenancesectionforusermaintenanceitems.

    AproductshouldNEVERbeleftunattendedwhenpluggedin.MakesurethePerfecto2isOffwhennotinuse.

    Closesupervisionisnecessarywhenthisproductisusednearchildrenorphysicallychallengedindividuals.

    Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealarmsorcommunicatediscomfort.

    DONOTconnecttheconcentratorinparallelorserieswithotheroxygenconcentratorsoroxygentherapydevices.

    Radio Frequency InterferenceThisequipmenthasbeentestedandfoundtocomplywithEMClimitsspecifiedbyIEC/EN6060112.Theselimitsaredesignedtoprovideareasonableprotectionagainstelectromagneticinterferenceinatypicalmedicalinstallation.

  • 7/24/2019 Perfecto2 Manual de Uso

    9/228

    SECTION 1GENERAL GUIDELINES

    Part No 1163146 9 Perfecto2Series

    Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabovestandards.TodetermineiftheemissionsfromthePerfecto2is

    causingtheinterference,turnthePerfecto2Off.Iftheinterferencewiththeotherdevice(s)stops,thenthePerfecto2iscausingtheinterference.Insuchrarecases,interferencemaybereducedorcorrectedbyoneofthefollowingmeasures:

    Reposition,relocate,orincreasetheseparation

    betweentheequipment. Connecttheequipmentintoanoutletonacircuit

    differentfromthattowhichtheotherdevice(s)isconnected.

  • 7/24/2019 Perfecto2 Manual de Uso

    10/228

    SECTION 2FEATURES

    Perfecto2Series 10 Part No 1163146

    SECTION 2FEATURES

    Oxygen Outlet

    Flowmeter

    Oxygen Purity IndicatorLights / Fault and Power

    Indicator Lights

    Power Switch

    Circuit Breaker

    Elapsed TimeMeter

    CabinetFilter

    REAR VIEW

    *HF OutletFitting

    PowerCord

    *NOTE:ThisoutletfittingistobeusedonlyforfillingoxygencylinderswiththeHomeFillhomeoxygencompressor.Theoutlet

    fittingdoesnotaffectconcentratorperformance.RefertotheHomeFillownersmanual,partnumber1145805,forconnectionandoperatinginstructions.Whennotinuse,theplugprovidedwith

    theconcentratorshouldbeinsertedintotheoutletfitting.FormoreinformationabouttheHomeFill,contactyourInvacaredealer.

  • 7/24/2019 Perfecto2 Manual de Uso

    11/228

    SECTION 3HANDLING

    Part No 1163146 11 Perfecto2Series

    SECTION 3HANDLINGTheconcentratorshouldALWAYSbekeptintheuprightpositiontopreventcabinetdamagewhilebeingtransported.

    Iftheconcentratoristobereshippedbycommoncarrier,additionalcartonsareavailablefromInvacare.

    UnpackingNOTE:Forthisprocedure,refertoFIGURE3.1.

    1. Checkforanyobvious

    damagetothecartonoritscontents.Ifdamageisevident,notifythecarrier,oryourlocaldealer.

    2. Removeallloosepackingfromthecarton.

    3. Carefullyremoveallthecomponentsfromthecarton. FIGURE 3.1 Unpacking

    NOTE:UnlesstheoxygenconcentratoristobeusedIMMEDIATELY,leaveconcentratorinitspackagingforstorageuntiluseoftheconcentratorisrequired.

    Inspection1. Inspect/examineexterioroftheoxygenconcentratorfor

    nicks,dents,scratchesorotherdamages.Inspectallcomponents.

    Storage1. Storetherepackagedoxygenconcentratorinadryarea.

    2. DONOTplaceobjectsontopofrepackagedconcentrator.

  • 7/24/2019 Perfecto2 Manual de Uso

    12/228

    SECTION 4TECHNICAL DESCRIPTION

    Perfecto2Series 12 Part No 1163146

    SECTION 4TECHNICALDESCRIPTION

    TheInvacarePerfecto2concentratorisusedbypatientswithrespiratorydisorderswhorequiresupplementaloxygen.Thedeviceisnotintendedtosustainorsupportlife.

    Theoxygenconcentrationleveloftheoutputgasrangesfrom87%to95.6%.Theoxygenisdeliveredtothepatientthroughtheuseofanasalcannula.

    TheInvacarePerfecto2concentratorusesamolecularsieveandpressureswingadsorptionmethodologytoproducetheoxygengasoutput.Ambientairentersthedevice,isfilteredandthencompressed.Thiscompressedairisthendirectedtowardoneoftwonitrogenadsorbingsievebeds.Concentratedoxygenexitstheoppositeendoftheactivesievebedandisdirectedintoanoxygenreservoirwhereitis

    deliveredtothepatient.TheInvacarePerfecto2concentratoriscapableofoperationbythepatientinahomeenvironmentorinaninstitutionalenvironment.Deviceoperatesatanominal230VAC/50Hertzsupply.

    Serviceinformationwillbeavailableuponrequestto

    qualifiedtechnicalpersonnelonly.

  • 7/24/2019 Perfecto2 Manual de Uso

    13/228

    SECTION 5TYPICAL PRODUCT PARAMETERS

    Part No 1163146 13 Perfecto2Series

    SECTION 5TYPICAL PRODUCTPARAMETERS

    Alternating Current

    Type BF equipment

    Unit running

    Unit not running

    Attention - Consider Accompanying Documents

    DO NOT smoke

    Class II, Double Insulated

    Protected against dripping water

    ElectricalRequirements: 230 VAC 10% (253 VAC/207 VAC), 50 Hz

    Rated Current Input:1.4 A

    Sound Level: 39.5 dBA Average

    Altitude: All Perfecto2Models:Up to 8,000 ft (2438 meters) above sea

    level without degradation of concentration levels. From 8,000 ft to13,129 (4000 meters) below 90% efficiency.

    All Perfecto2Models while filling a HomeFill System:

    Up to 6,000 ft (1828 meters) above sea level without degradationof concentration levels. From 6,000 ft (1828 meters) to 13,129 ft

    (4000 meters) below 90% efficiency

    IPX1

  • 7/24/2019 Perfecto2 Manual de Uso

    14/228

    SECTION 5TYPICAL PRODUCT PARAMETERS

    Perfecto2Series 14 Part No 1163146

    Oxygen OutputConcentrationLevels:NOTE:Stated

    concentrationlevelsachievedafterinitialwarmupperiod(approximately30minutes)

    87% to 95.6% at 0.5 to 5 L/min93% 3% at 2 L/min

    Maximum OutletPressure: 34.5 kPa 3.45 kPa (5 psi 0.5 psi)

    Flow Range: 0.5 to 5 L/min (maximum)For flowrates less than 1 L/min, we recommend the use of the

    Invacare Pediatric Flowmeter Accessory (IRCPF16AW)Low Flow Alarm: 0 L/min to 0.5 L/min

    Rapid Audible Alarm Beeping (No Accessories Connected)

    Average PowerConsumption: Perfecto

    2AW - 300 W @ 5 L/min, 280 W @ 3 L/min

    Perfecto2VAW - 320 W@ 5 L/min

    Pressure Relief

    MechanismOperational at: 241 kPa 24.1 kPa (35 psi 3.5 psi)

    Change in maximumrecommended flowwhen back pressureof 7kPa is applied: 0.7 L/min

    Filters: Cabinet, Outlet HEPA and Compressor Inlet

    Safety System: Current overload or line surge shutdownHigh temperature compressor shutdownHigh Pressure Alarm w/compressor shutdownLow Pressure Alarm w/compressor shutdownBattery Free Power Loss Alarm. SensO

    2Oxygen System (SensO

    2

    Model) Low Flow Alarm.

    Width: 38.1 cm 1 cm (15 in 3/8 in)

    Height: 58.4 cm 1 cm (23 in 3/8 in)

    Depth: 30.5 cm 1 cm (12 in 3/8 in)

    Weight: Perfecto2AW Models - 20.5 kg 1 kg (45 lbs 2 lbs)

    Perfecto2VAW Models - 18.1 kg 1 kg (40 lbs 2 lbs)

    Shipping Weight: Perfecto2AW Models - 22.7 kg 1 kg (50 lbs 2 lbs)

    Perfecto2VAW Models - 20.4 kg 1 kg (45 lbs 2 lbs)

    >

  • 7/24/2019 Perfecto2 Manual de Uso

    15/228

    SECTION 5TYPICAL PRODUCT PARAMETERS

    Part No 1163146 15 Perfecto2Series

    Operating AmbientTemperature: 10C - 35C (50F - 95F) at 20-60% relative humidity

    Exhaust Cooling AirTemperature: Less than Ambient +19C (+ 35F)

    Oxygen OutputTemperature: Less than Ambient +3C(+ 6F)

    Cabinet: Impact Resistant flame-retardant plastic cabinet that conforms toUL 94-V0

    No AP/APG Not suitable for use in the presence of a flammable anaestheticmixture

    Standards andRegulatory Listing: IRC5PO2AW/IRC5PAW/IRC5PO2VAW:

    IEC/EN 60601-1, A1, A2IEC/EN 60601-1-2IEC/EN61000-3-2IEC/EN61000-3-3

    IRC5PO2AW/IRC5PO2VAW Models OnlyISO8359

    MDD 93/42/EEC, Annex I and IXCE marked model IRC5PO2AW & IRC5PO2VAW

    Electrical: No extension cords

    Placement: No closer than 7.5 cm (3 in) from any wall, furniture, draperies, orfurniture to assure sufficient air flowAvoid deep pile carpets and heaters, radiators or hot air registersFloor location onlyNo confined spaces (Example: no closets)

    Tubing: 2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-ProofTubing (DO NOT pinch)

    Time of Operation: Up to 24 hours per day

    RecommendedStorage and ShippingTemperature: -29C to 65C (-20F to 150F) at 15-95% relative humidity

    Environment: Smoke and soot-freeNo confined spaces (Example: No closets).

    Minimum OperatingTime: 30 Minutes

  • 7/24/2019 Perfecto2 Manual de Uso

    16/228

    SECTION 5TYPICAL PRODUCT PARAMETERS

    Perfecto2Series 16 Part No 1163146

    IRC5PO2AW & IRC5PO2VAW INDICATORS

    IRC5PAW INDICATORS

    LABELSYMBOL

    O2PURITY INDICATOR LIGHTS

    (LED)

    SYSTEM OKAYO

    2over 85% ( 2%)

    GREEN Indicator Light

    O2Between 73% ( 3%)

    to 85% (2%)

    YELLOW Indicator lightA. YELLOW SolidB. YELLOW Flashing Sensor FailureCall a qualified technician.

    SYSTEM FAILUREO2Below 73% (3%) RED Indicator LightContinuous Audible Alarm

    Sieve-GARD Compressor Shutdown

    LABELSYMBOL

    STATUS INDICATOR LIGHTS(LED)

    SYSTEM OKAY GREEN Indicator Light

    SYSTEM FAILURE RED Indicator lightContinuous Audible AlarmSieve-GARDCompressorShutdownCall a qualified technician.

    O2

    I/O

  • 7/24/2019 Perfecto2 Manual de Uso

    17/228

    SECTION 6OPERATING INSTRUCTIONS

    Part No 1163146 17 Perfecto2Series

    SECTION 6OPERATINGINSTRUCTIONS

    IntroductionYouroxygenconcentratorisintendedforindividualuse.Itisanelectronicallyoperateddevicethatseparatesoxygenfromroomair.Itprovideshighconcentrationofoxygendirectlytoyouthroughanasalcannula.Clinicalstudieshavedocumentedthatoxygenconcentratorsaretherapeutically

    equivalenttoothertypesofoxygendeliverysystems.Yourproviderwillshowyouhowtouseyouroxygenconcentrator.He/Sheshouldbecontactedwithanyquestionsorproblemsregardingyouroxygenconcentrator.Thisownersmanualwilltellyouaboutyourconcentratorandwillserveasareferenceasyouuseyourconcentrator.

    Select a Location

    WARNING

    NEVERblocktheairopeningsoftheproductorplaceitonasoftsurface,suchasabedorcouch,wheretheairopeningmay

    beblocked.Keeptheopeningsfreefromlint,hairandthelike.Keepunitatleast7.5cm(3in)awayfromwalls,draperies,furniture,andthelike.

    Youmayselectaroominyourhousewhereusingyouroxygenconcentratorwouldbemostconvenient.Yourconcentratorcan

    beeasilyrolledfromroomtoroomonitscasters.

  • 7/24/2019 Perfecto2 Manual de Uso

    18/228

    SECTION 6OPERATING INSTRUCTIONS

    Perfecto2Series 18 Part No 1163146

    YouroxygenconcentratorwillperformbestwhenoperatedundertheconditionsoutlinedintheTypicalProductParametersonpage 13.Usageinenvironmentsotherthan

    thosedescribedmayresultintheneedforincreasedequipmentmaintenance.Theairintakeoftheunitshouldbelocatedinawellventilatedareatoavoidairbornepollutantsand/orfumes.DONOTuseinacloset.

    Set Up

    1. Pluginpowercordtoanelectricaloutlet.2. ConnectHumidifier(IfSoPrescribed)

    WARNING

    DONOToverfillhumidifier.

    DONOTreversetheoxygeninputandoutputconnections.

    Waterfromthehumidifierbottlewilltravelthroughthecannulabacktothepatient.

    NOTE:Forthisprocedure,refertoFIGURE6.1,FIGURE6.2onpage 19,FIGURE6.3onpage 20.

    1. Removecapfrombottle.

    2. Fillhumidifierwithdistilledwatertothelevelindicatedbythemanufacturer.Replacethehumidifiercapandsecurelytighten.

    FIGURE 6.1 Filling the Humidifier

    HumidifierBottle w/o CapHumidifier

    Bottle with Cap

  • 7/24/2019 Perfecto2 Manual de Uso

    19/228

    SECTION 6OPERATING INSTRUCTIONS

    Part No 1163146 19 Perfecto2Series

    3. Insertaflatheadscrewdriverintheplategrooveonthetopedgeofthefilteraccessdoorandgentlyprythefilteraccessdooroff(FIGURE6.2).

    4. Pullupandremovethehumidifierbottleadapter(FIGURE6.2).5. Replacethefilteraccessdoorandinstalltheairscoopontothefilteraccessdoorbyinsertingthesixtabsontheairscoopintothesixslotsonthesidesofthefilteraccessdoor.

    FIGURE 6.2 Humidifer Bottle Adapter6. Attachthehumidifierbottleadaptertothehumidifier

    bottlebyturningthewingnutonthehumidifierbottlecounterclockwiseuntilitissecurelyfastened.SeeDetailAinFIGURE6.3.

    7. Placethehumidifierbottle/adapterassemblyinthehumidifiercompartmentontheconcentrator.SeeDetailBinFIGURE6.3.

    8. Attachoxygentubingfromthehumidifierbottle/adapter

    assemblytooxygenoutletconnectorontheconcentrator.SeeDetailBinFIGURE6.3.9. Attachthecannula/patientsupplytubingtothe

    humidifierbottleoutlet.SeeDetailBinFIGURE6.3.10. Afterassembly,ensurethatoxygenisflowingthroughthe

    cannula.

    Insert FlatHead

    ScrewdriverHere Filter AccessDoor

    Humidifier BottleAdapter

    Cabinet Filter

    Air Scoop

    Tabs

    Slots

  • 7/24/2019 Perfecto2 Manual de Uso

    20/228

    SECTION 6OPERATING INSTRUCTIONS

    Perfecto2Series 20 Part No 1163146

    FIGURE 6.3 Humidifier Compartment

    Power SwitchNOTE:Forthisprocedure,refertoFIGURE6.4.

    1. PresspowerswitchtoOnposition.Allthepanellightsandtheaudiblealarmwillcomeonforonesecond,indicatingthattheunitis

    functioningproperly.

    FIGURE 6.4 Power Switch

    HumidifierBottle

    Humidifier

    Compartment

    HumidifierBottle

    AdapterHumidifier

    Bottle

    DETAIL A DETAIL B

    OxygenTubing

    OxygenOutlet

    Connector

    HumidifierBottleOutlet

    On/Off

    (I/O)PowerSwitch

  • 7/24/2019 Perfecto2 Manual de Uso

    21/228

    SECTION 6OPERATING INSTRUCTIONS

    Part No 1163146 21 Perfecto2Series

    FlowrateNOTE:Forthisprocedure,refertoFIGURE6.5.NOTE:TakecarenottosettheflowaboveREDring.Anoxygen

    flowgreaterthan5L/minwilldecreasetheoxygenconcentration.1. Turntheflowrateknobtothesettingprescribedbyyour

    physicianortherapist.

    WARNINGDONOTchangetheL/minsettingontheflowmeterunlessachangehasbeenprescribedbyyourphysicianortherapist.

    NOTE:Toproperlyreadtheflowmeter,locatetheprescribedflowratelineontheflowmeter.Next,turntheflowknobuntiltheballrisestotheline.Now,centertheballontheL/minlineprescribed.2. Iftheflowrateontheflowmetereverfallsbelow0.5L/minfor

    morethanaboutoneminute,theLOWFLOWalarmwillbetriggered.Thisisarapidbeepingoftheaudiblealarm.Check

    yourtubingoraccessoriesforblockedorkinkedtubingoradefectivehumidifierbottle.Afterratedflowisrestoredtomorethan0.5L/min,theLOWFLOWaudiblealarmwillgooff.

    NOTE:TheuseofsomeaccessoriessuchasthepediatricflowstandandtheHomeFillcompressorwilldeactivatetheLowFlowAlarm.

  • 7/24/2019 Perfecto2 Manual de Uso

    22/228

    SECTION 6OPERATING INSTRUCTIONS

    Perfecto2Series 22 Part No 1163146

    FIGURE 6.5 Flowrate

    SensO2Oxygen Purity Indicator - Model

    IRC5PO2AW & IRC5PO2VAWThisfeaturemonitorsthepurityofoxygengeneratedbytheoxygenconcentrator.Ifpurityfallsbelowfactorypresetstandards,indicatorlightsonthecontrolpanelwillilluminate.

    Initial Startup of the ConcentratorNOTE:Concentratormaybeusedduringtheinitialstartwarmuptime(approximately30min.)whilewaitingfortheO2puritytoreachmaximum.

    Whentheunitisturnedon,theGREENlightwillcomeon

    (SYSTEMOK/O2greaterthan85%).Afterfiveminutes,theoxygensensorwillbeoperatingnormallyandwillcontroltheindicatorlightsdependingonoxygenconcentrationvalues.

    Flow Knob

    Ball

    Flowmeter

    Front Panel

    0.5 L/min

  • 7/24/2019 Perfecto2 Manual de Uso

    23/228

    SECTION 6OPERATING INSTRUCTIONS

    Part No 1163146 23 Perfecto2Series

    Explanation of Oxygen Purity Indicator Lights -Model IRC5PO2AW & IRC5PO2VAWNOTE:Forthisprocedure,refertoFIGURE6.6onpage 23.

    GREENlight(O2) NormalOperation.

    YELLOWlight( ) Immediatelycallsupplier.Youmaycontinuetousetheconcentratorunlessinstructedotherwisebyyoursupplier.Becertainthatbackupoxygenisnearby.

    REDlight( ) Totalunitshutdown.Immediatelyswitchto

    backupoxygensupplyandcallsupplier.GREENlightwithYELLOWlightflashingImmediatelycallsupplier.Oxygensensormalfunctioning;youmaycontinuetousetheconcentrator.

    FIGURE 6.6 Explanation of Oxygen Purity Indicator Lights -Model IRC5PO2AW & IRC5PO2VAW

    Explanation of Indicator Lights - ModelIRC5PAWNOTE:Forthisprocedure,refertoFIGURE6.7.

    IRC5PAW Model Indicator Light Explanation

    REDlight( ) TotalUnitShutDown.Immediatelyswitchtoabackupoxygensupplyandcallsupplier.

    GREENlight(I/O) On/Off.Systemokay.

    O2

    Auto Shut Down

    Use Backup

    BelowNormal

    Normal

    RED

    Call Supplier

    YELLOW

    GREEN

    OXYGEN PURITY

  • 7/24/2019 Perfecto2 Manual de Uso

    24/228

    SECTION 6OPERATING INSTRUCTIONS

    Perfecto2Series 24 Part No 1163146

    FIGURE 6.7 Explanation of Indicator Lights - Model IRC5PAW

    Elapsed Time Meter (Hour Meter)

    Thehourmeterdisplaysthecumulativenumberofhourstheunithasoperated.

    Auto Shut DownUse Backup

    Normal

    RED

    GREEN

    Call Supplier

  • 7/24/2019 Perfecto2 Manual de Uso

    25/228

    SECTION 7MAINTENANCE

    Part No 1163146 25 Perfecto2Series

    SECTION 7MAINTENANCE

    WARNING

    TheInvacareconcentratorsarespecificallydesignedtominimizeroutinepreventivemaintenance.Onlyqualifiedpersonnelshouldperformpreventivemaintenanceontheconcentrator.

    Unplugtheconcentratorwhencleaning.Toavoidelectricalshock,DONOTremovecabinet.

    NOTE:Ataminimum,preventivemaintenanceMUSTbeperformedaccordingtothemaintenancerecordguidelines.Inplaceswithhighdustorsootlevels,maintenancemayneedtobeperformedmoreoften.RefertoPreventiveMaintenanceRecordonpage 27.

    Routine MaintenanceThefollowingroutinemaintenanceshouldbeperformedmore

    frequently.SeerecommendedintervalsineachsectionandPreventiveMaintenanceRecordonpage 27.

    Cleaning the Cabinet Filter

    CAUTIONDONOToperatetheconcentratorwithoutthefilterinstalled.

    NOTE:Forthisprocedure,refertoFIGURE7.1onpage 26.

    NOTE:Thereisonecabinetfilterlocatedonthebackofthecabinet.

    1. Inspectthefiltereverymonth.Removethefilterandcleanasneeded.

    NOTE:Enviromentalconditionsthatmayrequiremorefrequentinspectionandcleaningofthefilterinclude,butarenotlimitedto:highdust,airpollutants,etc.

  • 7/24/2019 Perfecto2 Manual de Uso

    26/228

    SECTION 7MAINTENANCE

    Perfecto2Series 26 Part No 1163146

    2. Cleanthecabinetfilterwithavacuumcleanerorwashinwarmsoapywaterandrinsethoroughly.

    3. Drythefilterthoroughlybeforereinstallation.

    FIGURE 7.1 Cleaning the Cabinet Filter

    Cleaning the Cabinet

    1. Cleanthecabinetwithamildhouseholdcleanerandnonabrasiveclothorsponge.

    Cleaning the HumidifierNOTE:Tocleantheoxygenhumidifier,followtheinstructionsprovidedbythemanufacturer.Ifnoneareprovided,followtheseSTEPS:

    1. Cleanthehumidifiereveryday.

    2. Washitinsoapywaterandrinseitwithasolutionoften

    partswaterandonepartvinegar.3. Rinsethoroughlywithhotwaterandrefillwithdistilled

    watertothelevelshownonthehumidifier.

  • 7/24/2019 Perfecto2 Manual de Uso

    27/228

    SECTION 7MAINTENANCE

    Part No 1163146 27 Perfecto2Series

    FIGURE 7.2 Preventive Maintenance Record

  • 7/24/2019 Perfecto2 Manual de Uso

    28/228

    SECTION 8TROUBLESHOOTING GUIDE

    Perfecto2Series 28 Part No 1163146

    SECTION 8TROUBLESHOOTINGGUIDE

    SYMPTOM: PROBABLE CAUSE: SOLUTION:Alarm:

    Short beeps, longpause

    Concentrator notoperating, power

    switch On.

    Beep......Beep......

    Main Power Loss:

    1. Power cord notplugged in.

    1. Insert plug into outlet.

    2. No power at outlet. 2. Inspect house circuitbreakers or fuses. If problemrecurs, use a different outlet.

    3. Tripped circuitbreaker.

    3. Push/reset circuit breaker.If problem recurs, call ServiceProvider.

    Alarm:Continuous

    Concentrator not

    operating, PowerSwitch On.

    Beeeeeeeep....

    System Failure:

    1. Unit overheating dueto blocked air intake.

    1a. Remove and clean cabinetfilters.

    1b. Move oxygenconcentrator at least threeinches away from walls,draperies or furniture.

    2. Insufficient power atoutlet.

    2. DO NOT use extensioncords. Move to anotherelectrical outlet or circuit.

    3. Internal repairsrequired.

    3. Call Service Provider.

  • 7/24/2019 Perfecto2 Manual de Uso

    29/228

    SECTION 8TROUBLESHOOTING GUIDE

    Part No 1163146 29 Perfecto2Series

    YELLOW or REDlight Illuminated.

    Alarm:ContinuousOn RED lightonly.

    * Only applies toIRC5PO2AW &IRC5PO2VAWmodels

    1. Low oxygen purity.* 1. Clean or Replace filters.

    2. Kinked or blockedtubing, cannula orhumidifier.*

    2. Inspect for kinks orblockages. Correct, clean orreplace item. Oncecorrected, turn power Off for60 seconds and then turnpower back On.

    3. Flowmeter set at 0.5

    L/min *.

    3. Check flowmeter is set to

    1.0 L/min or more. Refer toTypical Product Parameterson page 13.

    4. Unit overheating dueto blocked air intake.

    4a. Remove and clean cabinetfilters.4b. Move concentrator atleast three inches from walls,draperies, and furniture.

    5. Insufficient power atoutlet.

    5. DO NOT use extensioncords. Move to anotherelectrical outlet or circuit.

    6. Internal repairsrequired.

    6. Call Service Provider.

    GREEN light withYELLOW Light

    Flashing.

    1. Internal repairsrequired.

    1. Call Service Provider.

    Alarm:

    Rapid

    Beep....Beep...Beep...Beep

    LOW FLOW ALARM

    1. Kinked or blockedtubing, cannula orhumidifier.

    1. Inspect for kinks orblockages. Correct, clean orreplace item. Oncecorrected, turn power Off for60 seconds and then turnpower back ON.

    2. Flowmeter set at 0.5L/min.

    2. Flowrates less than 1 L/minare not recommend.

    SYMPTOM: PROBABLE CAUSE: SOLUTION:

  • 7/24/2019 Perfecto2 Manual de Uso

    30/228

    SECTION 9OPTIONAL ACCESSORIES

    Perfecto2Series 30 Part No 1163146

    SECTION 9OPTIONALACCESSORIES

    Thefollowingoptionalaccessoriesarealsoavailable: StandardAdultNasalCannula,2.1m(7feet) M3120

    StandardHumidifier M5120

    PreciseRXPediatricHumidifier/FlowmeterAccessory IRCPF16AW

    HomeFillhomeoxygencompressor IOH200AW

  • 7/24/2019 Perfecto2 Manual de Uso

    31/228

    RECYCLING INFORMATION

    Part No 1163146 31 Perfecto2Series

    RECYCLING INFORMATIONThisproducthasbeensuppliedfromanenvironmentallyawaremanufacturerthatcomplieswiththeWasteElectricalandElectronicEquipment(WEEE)Directive2002/96/CE.Thisproductmaycontainsubstancesthatcouldbeharmfultotheenvironmentifdisposedofinplaces(landfills)thatarenotappropriateaccordingtolegislation.

    Thecrossedoutwheeliebin symbolisplacedonthisproducttoencourageyoutorecyclewhereverpossible.

    Pleasebeenvironmentallyresponsibleandrecyclethisproductthroughyourrecyclingfacilityatitsendoflife.

  • 7/24/2019 Perfecto2 Manual de Uso

    32/228

    Perfecto2Series 32 Part No 1163146

    LIMITED WARRANTYBelgium&Luxemburg:Invacarenv,Autobaan22,B8210LoppemTel:(32)(0)50831010

    Fax:(32)(0)[email protected]:InvacareA/S,Sdr.Ringvej37,DK2605BrndbyTel:(45)(0)36900000Fax:(45)(0)[email protected]:InvacareAquatecGmbH,Alemannenstrae10,D88316Isny

    Tel:(49)(0)75627000Fax:(49)(0)[email protected]:Invacare,Kleiststrae49,D32457PortaWestfalicaTel:(49)(0)5731754540Fax:(49)(0)[email protected]:InvacareSA,c/Arenys/n,Polgon

    IndustrialdeCelr,E17460Celr(Girona)Tel:(34)(0)972493200Fax:(34)(0)[email protected]:InvacarePoirierSAS,Route deStRoch,F37230 FondettesTel:(33)(0)247626466Fax:(33)(0)[email protected]:

    InvacareAG,Benkenstrasse260,CH4108 WitterswilTel:(41)(0)614877080Fax:(41)(0)[email protected]:InvacareIrelandLtd,Unit5SeatownBusinessCampus,SeatownRoad,Swords,CountyDublinIrelandTel:(353)18107084

    Fax:(353)[email protected]:InvacareMeccSans.r.l.,Via deiPini62,I36016Thiene(VI)Tel:(39)0445380059Fax:(39)[email protected]

    Nederland:InvacareBV,Celsiusstraat46,NL6716BZEde

    Tel:(31)(0)318695757Fax:(31)(0)[email protected]@invacare.comNorge:InvacareAS,Grensesvingen 9,Postboks6230,Etterstad,N0603OsloTel:(47)(0)22579500Fax:(47)(0)[email protected]@invacare.com

    sterreich:InvacareAustriaGmbH,HerzogOdilostrasse 101,A5310 MondseeTel:(43)623255350Fax:(43)[email protected]&Suomi:InvacareAB,Fagerstagatan 9,S163 91 SpngaTel:(46)(0)87617090Fax:(46)(0)87618108

    [email protected]@invacare.comUnitedKingdom:InvacareLimited,PencoedTechnologyPark,Pencoed,BridgendCF355HZTel:(44)(0)1656776200Fax:(44)(0)1656776201CustomerServices:Tel:(44)(0)1656776222Fax:(44)(0)[email protected]

    Portugal:InvacareLda,RuaEstradaVelha,949,P4465784LeadoBalioTel:(351)(0)225105946/47Fax:(351)(0)[email protected]

    mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]
  • 7/24/2019 Perfecto2 Manual de Uso

    33/228

    Part No 1163146 33 Perfecto2Series

    NOTES

  • 7/24/2019 Perfecto2 Manual de Uso

    34/228

    Distributeur : Ce manuel DOIT tre remis lutilisateur final.Utilisateur

    : Lire ce manuel AVANT d'utiliser cet

    appareil, et le conserver en cas de besoin.

    Manuel dutilisation

    Pour de plus amples informations sur lesproduits Invacare, les pices dtaches et le

    service client, veuillez vous rendre surwww.invacare.fr

    Perfecto2

    SrieConcentrateurs oxygne sans SensO2,

    Concentrateurs oxygne avec SensO2

    Compatible HomeFill

    Modle IRC5PO2AW

    Modle IRC5PO2VAW

    Modle IRC5PAW

    0

    434

  • 7/24/2019 Perfecto2 Manual de Uso

    35/228

    MISE EN GARDE

    Part No 1163146 35 Perfecto2 Srie

    MISE EN GARDENEPASutilisercetappareiloutoutautrequipement

    optionnelsansavoirpralablementluetcompriscesinstructionsettoutdocumentdinstructionadditionneltelquelemanueldelutilisateur,manueldentretienoufeuilletsfournisavecceproduitoutoutmatrieloptionnel.Silutilisateurnecomprendpaslesmisesengarde,messagesdattentionouinstructions,ildoitcontacterunprofessionneldelasant,ledistributeuroulepersonneltechniqueavantdetenterdutilisercetappareil Sinon,ilyaunrisquedeblessurecorporelleoudedommagematriel.

    MISE EN GARDE POUR LESACCESSOIRES

    LesproduitsInvacareonttspcifiquementconusetfabriquspourtreutilissconjointementaveclesaccessoiresInvacare.LesaccessoiresconuspardautresfabricantsnontpasttestsparInvacareetilnestpasrecommanddelesutiliseraveclesproduitsInvacare.

    ACCESSOIRES

    Ilyaunelargegammedhumidificateurs,detubesoxygne,decanulesetdemasquespouvanttreutilissaveccetappareil.Vousdevriezcontactervotreprestatairedesoinsdomicileafinquilvousindiquequelssontlesdispositifsadaptsvotrecas.Ilsdevraientgalementvousrenseignerconcernantlamthodedutilisation,lamaintenanceetlenettoyageappropris.REMARQUE:Lesversionsmisesjourdecemanuelsontdisponiblessurwww.invacare.fr.

  • 7/24/2019 Perfecto2 Manual de Uso

    36/228

    SOMMAIRE

    Perfecto2 Srie 36 Part No 1163146

    ACCESSOIRES ................................................ 35REMARQUES SPCIALES ................................37

    EMPLACEMENT DU LABEL ............................. 39SECTION 1REGLES GENERALES .................... 40

    Informations de fonctionnement ...................................................41

    Interfrence radiolectrique...........................................................42

    SECTION 2CARACTRISTIQUES ................... 43SECTION 3MANIPULATION ......................... 44

    Dballage.............................................................................................44Contrle..............................................................................................44

    Rangement ..........................................................................................44

    SECTION 4DESCRIPTION TECHNIQUE .......... 45SECTION 5PARAMTRES TYPES DUPRODUIT ...................................................... 46SECTION 6CONSIGNES D'UTILISATION ........ 50

    Introduction........................................................................................50

    Choisir Un Endroit............................................................................50

    Montage...............................................................................................51

    Interrupteur........................................................................................53

    Dbit.....................................................................................................54

    Tmoin de la puret de l'oxygne SensO2

    - Modle

    IRC5PO2AW & IRC5PO2VAW....................................................55

    Mise en marche initiale du concentrateur...................................55

    Explication des tmoins lumineux - Modle IRC5PAW ..........57

    Compteur horaire.............................................................................57

    SECTION 7ENTRETIEN ................................ 58Entretien de routine .........................................................................58

    SECTION 8GUIDE DE DPANNAGE .............. 61SECTION 9ACCESSOIRES EN OPTION ...........63INFORMATIONS DE RECYCLAGE .................... 64LIMITES DE GARANTIE .................................. 65

  • 7/24/2019 Perfecto2 Manual de Uso

    37/228

    REMARQUES SPCIALES

    Part No 1163146 37 Perfecto2 Srie

    REMARQUES SPCIALESDesmotsdalertesontutilissdanscemanueletsappliquentauxpratiquesrisqueoudangereusespouvantentrainerdesblessurescorporellesoudesdgtsmatriels.Sereporterautableausuivantpourlesdfinitionsdesmotsindicateurs.

    AVIS

    Lesinformationscontenuesdanscedocumentsontsusceptiblesdtremodifiessanspravis.

    DANGERNE FUMEZ PAS pendant l'utilisation de cet appareil.Conservez allumettes, cigarettes allumes ou autressources d'ignition en dehors de la pice dans laquelle setrouve cet appareil et loignes de l'endroit ou del'oxygne est distribu.

    MOT-INDICATEUR DEFINITION

    DANGER

    Danger indique une situation dangereuseimminente qui, si elle n'est pas carte, entrainerala mort ou des blessures graves.

    MISE EN GARDE

    Mise en garde indique une situationpotentiellement dangereuse qui, si elle n'est pascarte, peut entrainer la mort ou des blessuresgraves.

    ATTENTION

    Attention indique une situation de risque potentielqui, s'il n'est pas cart, peut entrainer des dgtspersonnels ou des lsions mineures voire les

    deux.

  • 7/24/2019 Perfecto2 Manual de Uso

    38/228

    REMARQUES SPCIALES

    Perfecto2 Srie 38 Part No 1163146

    DANGERDes panneaux INTERDIT DE FUMER doivent tre

    exposs bien en vue. Les textiles et autres matriaux quine brleraient pas dans des conditions normalessenflamment facilement et brlent avec une grandeintensit dans un endroit ou l'air est enrichi en oxygne.Le non-respect de cette mise en garde peut entraner unincendie et des dgts matriels graves, ou provoquer desblessures corporelles voire la mort.

    ATTENTION

    Attention:Laloipeutrestreindrelaventedecetappareildesortequilnepuissetrevenduqueparunmdecinousursonordonnance,oupartoutautrepraticienagrsurlelieudexercice.

    Invacarerecommandeunesourcealternativedapportenoxygnedanslecasdunecoupuredecourant,dunesituationdurgenceoudunepannemcanique.Consultezvotremdecinouvotrefournisseurconcernantlesystmederservequevoussouhaitez.

    Cematrieldoittreutilispourassureruncomplmentdoxygneetnonpourservirdappareilderanimationetdemaintiendesfonctionsvitales.

  • 7/24/2019 Perfecto2 Manual de Uso

    39/228

    EMPLACEMENT DU LABEL

    Part No 1163146 39 Perfecto2 Srie

    EMPLACEMENT DU LABEL

    L'tiquette avec le numrode srie est situe sur

    L'tiquette descaractristiques est

    situe sur l'arrire versle bas

    erfecto

    POUR LES INSTRUCTIONS DE FONCTIONNEMENT ou ALARMES (voir le manuel de l'oprateur)

    DANGER - RISQUE D'INCENDIENE PAS FUMER - Ecarter TOUTE source d'ignition de la pice o se trouve cet appareil et distance des zones de distribution d'oxygne. Les textiles, huiles et autres combustibles reprsententdes sources inflammables pouvant brler avec grande intensit dans un environnement o l'air estenrichi en oxygne.DANGER - ELECTROCUTIONNE PAS enlever le couvercle. Confier toute rparation un personnel de maintenance qualifi.

    Un taus de dbit infrieur 0.5 L/min dclenchera une alarme sonore de signalisation de bas dbit(alarme rapide audible).Positionnez le dbitmtre sur 2.5 L/min ou moins lorsque vous remplissez les bouteilles HomeFill.

    CompatibleHomeFill

  • 7/24/2019 Perfecto2 Manual de Uso

    40/228

    SECTION 1REGLES GENERALES

    Perfecto2 Srie 40 Part No 1163146

    SECTION 1REGLES GENERALESPourgarantirlinstallation,lassemblageetlefonctionnementsrsduconcentrateurPerfecto2,ilconvientderespectercesinstructionsIMPRATIVEMENT.

    MISE EN GARDECechaptrecontientdesinformationsimportantesconcernantlefonctionnementetlutilisationsansdangerdeceproduit.

    DANGER

    Risquedlectrocution.NEPASdmonter.Confiertouterparationunpersonneldemaintenancequalifi.Aucunepicerparableparlutilisateur.

    POURREDUIRELERISQUEDEBRLURES,

    DELECTROCUTION,DINCENDIE

    OU

    DE

    DOMMAGES

    CORPORELS.

    Nejamaisutiliserlappareilenprenantunbain.Silaprescriptionmdicalerequiertuneutilisationcontinue,leconcentrateurDOITtreplacdansuneautrepiceloignedaumoins2.5m(7ft)delasalledebains.

    NE

    PAS

    toucher

    le

    compresseur

    en

    tant

    mouill.NEPASplacerourangerlappareilproximitdeleauoudunautreliquide.

    NEJAMAIStenterdereprendrelappareilsiltombedansleau.DbrancherIMMEDIATEMENT.

    Silecordonoulapriseduconcentrateurestendommag,

    sil

    ne

    fonctionne

    pas

    correctement,

    sil

    heurte

    le

    sol,

    sil

    est

    endommagouplongdansleau,lappareildoittrerenvoyuntechnicienqualifipourtreinspectetrpar.

  • 7/24/2019 Perfecto2 Manual de Uso

    41/228

    SECTION 1REGLES GENERALES

    Part No 1163146 41 Perfecto2 Srie

    DANGER

    Uneinflammationspontaneetviolentepeutseproduiresi

    de

    lhuile,

    de

    la

    graisse

    ou

    des

    substances

    grasses

    entrent

    en

    contactavecloxygnesouspression.CessubstancesDOIVENTtretenueslcartduconcentrateurdoxygne,destubesetconnexions,etdetoutmatrieldoxygnationengnral.NEPASutiliserdelubrifiantssaufencasderecommandationsparInvacare.

    viterdeprovoquerdestincellesproximitdelappareil

    mdicaldoxygnation.Celainclutlestincellesdlectricitstatiqueprovoquespardesfrictions.

    Informations de fonctionnementPouruneperformanceoptimale,Invacarerecommandeque

    chaqueconcentrateurfonctionnependantunminimumde30minuteschaquefois.Despriodesdefonctionnementpluscourtespeuventrduireladuredeviedecetappareil.

    Eloignezlestubesoxygne,lecordonetlappareildesouslescouvertures,lesdessusdelit,lescoussins,lesvtementsetcartsdessurfaceschauffesouchaudes,comprenantlesradiateurs,lesfoursetappareilslectriquessimilaires.

    NEPASdplacerouchangerleconcentrateurdeplaceentirantsurlecordonlectrique.

    NEJAMAISlaissertomberouinsrerunobjetouduliquidedanslesouvertures.

    Invacarerecommandequuntubeoxygneanticrasementsoitutilisavecceproduitetquilnedpassepas15.2m

    (50ft)delongueur.Ilnyapasdepicesrparablesparlutilisateur.Cecinincluepasleslmentsdentretiennormal.Voirlasectionentretienpourlentretiendeslmentsparlutilisateur.

  • 7/24/2019 Perfecto2 Manual de Uso

    42/228

    SECTION 1REGLES GENERALES

    Perfecto2 Srie 42 Part No 1163146

    NEJAMAISlaisserlappareilsanssurveillancelorsquilestbranch.AssurezvousquelePerfecto2estdbranchlorsquilnestpasutilis.

    Unesupervisionrapprocheestncessairelorsqueceproduitestutilisparouproximitdesenfantsoudesindividussouffrantdehandicapsphysiques.

    Uncontrleouuneattentionparticuliredoittreapporteauxpatientsquiutilisentcetappareilouquisontincapablesdentendreoudevoirdesalarmesoudecommuniquerleurinconfort.

    NEPASbrancherleconcentrateurenparallleouensrieavecdautresconcentrateursdoxygneouappareilsdoxygnothrapie.

    Interfrence radiolectrique

    CetquipementattestettrouvconformeauxlimitesEMCspcifiesparIEC/EN6060112.Ceslimitessontdestinesprotgerraisonnablementcontrelesinterfrenceslectromagntiquesdansuneinstallationmdicaletypique.

    Dautresappareilspeuventrecevoirdesinterfrencesmmedesplusbasniveauxdesmissionslectromagntiquesautorisesparlesstandardscidessus.Pourdterminersiles

    missionsduPerfecto2causentlinterfrence,arrtezlePerfecto2.Silinterfrenceaveclautreappareilsarrte,alorslePerfecto2causelinterfrence.Danscescasrares,linterfrencepeuttrerduiteoucorrigedelunedesfaonssuivantes:

    Repositionnezlamachine,changezladeplace,ouloignezlespacemententrelesmachines.

    Branchezlappareildansuneprisediffrentedecellesurlaquellelautreappareilestbranch.

  • 7/24/2019 Perfecto2 Manual de Uso

    43/228

    SECTION 2CARACTRISTIQUES

    Part No 1163146 43 Perfecto2 Srie

    SECTION 2CARACTRISTIQUES

    Sortie oxygne

    Dbitmtre

    Voyants lumineux de lapuret de l'oxygne /Voyants lumineux defonctionnement et de

    panne

    Interrupteur

    Disjoncteur

    Compteurhoraire

    Filtres

    VUE ARRIRE

    * Raccord desortie HomeFillCordon

    d'alimentation

    *REMARQUE:Ceraccorddesortienedoittreutilisque

    pourremplirlesbouteillesdoxygneaveclecompresseurHomeFill.Leraccorddesortienaffectepaslaperformancedu

    concentrateur.SerfreraumanueldutilisationHomeFill,rfrence1145805,pourlesinstructionsderaccordementetdutilisation.Lobturateur

    fourniavecleconcentrateurdoittreintroduitdansleraccorddesortielorsquilnest

    pasutilis.PourplusdinformationsconcernantleHomeFill,contactezvotrevendeurInvacare.

  • 7/24/2019 Perfecto2 Manual de Uso

    44/228

    SECTION 3MANIPULATION

    Perfecto2 Srie 44 Part No 1163146

    SECTION 3MANIPULATIONLeconcentrateurdoitTOUJOURStremaintenuenpositionverticaleafindviterdendommagerlecaissonlorsdutransport.

    Sileconcentrateurdoittrerexpdipartransporteur,descartonssupplmentairessontdisponibleschezInvacare.

    DballageREMARQUE:Pourcetteprocdure,sereporterFIGURE3.1.

    1. Contrlertoutdommage

    videntducartonoudesoncontenu.Sidesdgtssontvisibles,informezenlelivreur,ouvotrevendeurlocal.

    2. Retirertouteslesgarnituresdescartons.

    3. Retirersoigneusementtousleslmentsducarton.

    FIGURE 3.1 DballageREMARQUE:AmoinsqueleconcentrateuroxygnedoivetreutilisIMMEDIATEMENT,laisserleconcentrateurdanssonemballagepourlerangerjusqucequelutilisationdu

    concentrateursoitncessaire.Contrle1. Examiner/contrlerlextrieurduconcentrateurdoxygne

    pourreprerdventuellesrayures,bosses,rafluresouautresdommages.Contrlertousleslments.

    Rangement1. Rangerleconcentrateurdoxygneremballdansunendroitsec.

    2. NEPASplacerdautresobjetsaudessusduconcentrateurdoxygneremisdanssonemballage.

    http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    45/228

    SECTION 4DESCRIPTION TECHNIQUE

    Part No 1163146 45 Perfecto2 Srie

    SECTION 4DESCRIPTIONTECHNIQUE

    LeconcentrateurInvacarePerfecto2estutilisparlespatientsatteintsdedsordresrespiratoiresquincessitentunapportdoxygne.Lappareilnestpasprvupourmaintenirouprolongerlavie.

    Letauxconcentrationdoxygnevariede87%95,5%.Lapportenoxygneaupatientsefaitlaidedunecannulenasale.

    LeconcentrateurInvacarePerfecto2utiliseuntamismollculaireetunemthodedabsorptiondelapressionpourproduirelasortiedoxygne.Lairambiantentredanslappareil;estfiltrpuiscompress.Laircompressestensuitedirigverslundesdeuxtamisabsorbantslazote.Loxygneconcentrschappedutamisopposetestdirig

    dansunrservoiroxygneoilesttransmisaupatient.LeconcentrateurInvacarePerfecto2estcapabledtreutilisparlepatientdansunenvironnementdhabitationouunenvironnementinstitutionnel.Lappareilfonctionneunepuissancenominalede230VAC/50Hertz.

    Lesinformationsdentretienserontfourniessurdemandeau

    personneltechniquequalifiuniquement.

  • 7/24/2019 Perfecto2 Manual de Uso

    46/228

    SECTION 5PARAMTRES TYPES DU PRODUIT

    Perfecto2 Srie 46 Part No 1163146

    SECTION 5PARAMTRES TYPES DUPRODUIT

    Courant alternatif

    quipement de Type BF

    Appareil en marche

    Appareil l'arrt

    Attention - Consulter la documentation jointe

    NE PAS fumer

    Classe II, double isolation

    Protger contre les projections d'eau

    Caractristiqueslectriques : 230 V c.a. 10 % (253 V c.a./207 V c.a.), 50 Hz

    Consommation decourant nominal : 1.4 A

    Niveau sonore : 39 Dba de moyenne

    Altitude : Tout le Perfecto2 Modle : Jusqu' 8.000 pi (2438 mtres)au-dessus du niveau de la mer sans la dgradation de niveaux deconcentration. De 8.000 pi 13.129 pi (4000 mtres) au dessousde 90% efficacit.

    Tout le Perfecto2 Modle en remplissant un Systme de HomeFill :Jusqu' 6.000 pi (1828 mtres) au-dessus du niveau de la mer sans la

    dgradation de niveaux de concentration. De 6.000 pi (1828mtres) 13.129 pi (4000 mtres) au dessous de 90% efficacit.

    IPX1

  • 7/24/2019 Perfecto2 Manual de Uso

    47/228

    SECTION 5PARAMTRES TYPES DU PRODUIT

    Part No 1163146 47 Perfecto2 Srie

    Taux deconcentrationd'oxygne :

    REMARQUE:Lestauxdeconcentrationindiqusonttobtenusaprsune

    priodedechauffeinitialedenviron30minutes.

    87 % 95.6 % de 0.5 5 L/min93% 3% 2 L/min

    Pression de sortiemaximale : 34.5 kPa 3.45 kPa (5 psi 0.5 psi)

    Plage de dbit : De 0.5 to 5 L/min (maximum)Pour les taux de dbit infrieurs 1L/min, nous recommandonsl'utilisation du dbitmtre pdiatrique Invacare (IRCPF16AW)

    Alarme de baisse dedbit :

    De 0 L/min to 0.5 L/minBips d'alarme rapides et audibles(Aucuns accessoires branchs)

    Puissance moyenneConsommation: Perfecto

    2AW - 300 W @ 5 L/min, 280 W @ 3 L/min

    Perfecto2VAW - 320 W @ 5 L/min

    Mcanisme dedcompressionfonctionnant : 241 kPa 24,1 kPa (35 psi 3,5 psi)

    Modification du dbitmaximalrecommand

    lorsquunecontre-pression de 7kPa est applique : 0,7 l/mn.

    Filtres : Caisson, sortie HEPA et admission

    Systme de scurit : Arrt en cas de surcharge de courant ou de surtensionArrt du compresseur en cas de temprature leveAlarme de haute pression avec arrt du compresseurAlarme de basse pression avec arrt du compresseur

    Alerte de perte dalimentation auxiliaireAlarme de dbit faible du systme doxygne SensO

    2(modle SensO

    2)

    Largeur : 38.1 cm 1 cm (15 in 3/8 in)

    Hauteur : 58.4 cm 1 cm (23 in 3/8 in)

    Profondeur : 30.5 cm 1 cm (12 in 3/8 in)

    >

  • 7/24/2019 Perfecto2 Manual de Uso

    48/228

    SECTION 5PARAMTRES TYPES DU PRODUIT

    Perfecto2 Srie 48 Part No 1163146

    Poids : Perfecto2AW - 20.5 kg 1 kg (45 lbs 2 lbs)

    Perfecto2VAW - 18.1 kg 1 kg (40 lbs 2 lbs)

    Poids dexpdition: Perfecto2AW - 22.7 kg 1 kg (50 lbs 2 lbs)

    Perfecto2VAW - 20.4 kg 1 kg (45 lbs 2 lbs)Tempratureambiante : 10 C - 35 C (50 F - 95 F) une humidit relative de 20 60 %

    Temprature del'chappement del'air derefroidissement : Infrieur la temprature ambiante +19 C

    Temprature desortie de l'oxygne : Infrieure la temprature ambiante +3 C (+ 6 F)

    Caisson : Coffrage en plastique anti-chocs et retardateur de flammesconforme la norme UL 94-V0

    Pas de No AP/APG Non adapt pour une utilisation en prsence d'un mlangeanesthsique inflammable

    Liste standard etrglementaire:

    IRC5PO2AW/IRC5PAW/IRC5PO2VAWIEC/EN 60601-1, A1, A2

    IEC/EN 60601-1-2IEC/EN61000-3-2IEC/EN61000-3-3

    Uniquement les modles IRC5PO2AW/IRC5PO2VAWISO8359MDD 93/42/EEC, Annexe I et IX

    Modles CE : IRC5PO2AW & IRC5PO2VAW

    lectricit : Pas de rallongePositionnement : loigner le concentrateur doxygne au moins 7,5 cm des murs,

    rideaux ou meubles pour assurer un dbit d'air suffisantviter les moquettes paisses et les appareils de chauffage,radiateurs ou registres dair chaudUniquement sur le solNe pas utiliser l'appareil dans des espaces confins (Exemple:Placards ou Toilettes)

    Lunette oxygne : Canule de 2 m (7 ft) indformable de 15 m (50 ft) max. de tubage(NE PAS pincer).

    Dure defonctionnement : Jusqu 24 heures par jour

  • 7/24/2019 Perfecto2 Manual de Uso

    49/228

    SECTION 5PARAMTRES TYPES DU PRODUIT

    Part No 1163146 49 Perfecto2 Srie

    TMOINS IRC5PO2AW & IRC5PO2VAW

    VOYANT LUMINEUX IRC5PAW

    Tempraturerecommande pourle stockage etlexpdition : -29 C 65 C, une humidit relative de 15 95 %.

    Environnement : Sans fume ni suieNe pas utiliser l'appareil dans des espaces confins (Exemple:Placards ou Toilettes)

    Dure defonctionnementminimum : 30 minutes

    SYMBOLEFIGURANT

    SURLTIQUETTE

    PURET O2 VOYANTS LUMINEUX(DEL)

    SYSTEM OKAY(SYSTME OKAY)O

    2au dessus de 85 % ( 2 %)

    Voyant lumineux VERT

    O2entre 73 % ( 3 %) et

    85 % ( 2 %)

    Tmoin lumineux JAUNEA. JAUNE fixeB. Voyant clignottant JAUNEPanneAppelez un technicien qualifi.

    SYSTME DEFAILLANTO

    2au dessous de 73 % (3 %)

    Voyant lumineux ROUGEAlarme sonore continueSieve-GARD arrt du compresseur

    SYMBOLEFIGURANT

    SURLTIQUETTE

    TAT VOYANTS LUMINEUX(DEL)

    SYSTEM OKAY(SYSTME OKAY)

    Voyant lumineux VERT

    SYSTME DEFAILLANT Voyant lumineux ROUGEAlarme sonore continueSieve-GARDArrt du compresseurAppelez un technicien qualifi

    O2

    I/O

  • 7/24/2019 Perfecto2 Manual de Uso

    50/228

    SECTION 6CONSIGNES D'UTILISATION

    Perfecto2 Srie 50 Part No 1163146

    SECTION 6CONSIGNESD'UTILISATION

    IntroductionVotreconcentrateuroxygneestprvupourunusageindividuel.Cestunappareillectroniquespareloxygnedelairambiant.Ildlivreuneforteconcentrationdoxygneparunecanulenasale.Destudescliniquesontmontrquelesconcentrateursdoxygnesontquivalentsenmatirede

    thrapiedautrestypesdesystmededlivrancedoxygne.Leprestatairedmontreracommentutiliserleconcentrateurdoxygne.Ildoittreconsultpourtoutesquestionsouproblmespossparleconcentrateurdoxygne.Lemanueldelutilisateurvousinstruirasurvotreconcentrateuretvousserviraderfrencepourlutilisationdevotreconcentrateur.

    Choisir Un Endroit

    MISE EN GARDE

    NEJAMAISbloquerlesouverturesdairdelappareilnileplacersurunesurfacemolle,unlitouuncanapparexemple,

    pournepasbloquerlessortiesdair.loignerlespeluches,cheveuxettexturessimilairesdessortiesdair.

    Tenirlappareilloigndesmurs,rideaux,mobilieretautresdaumoins7,5cm.

    Choisirunepicedanslamaisonolutilisationdu

    concentrateurdoxygneseralapluspratique.Leconcentrateurdoitpouvoirtrefacilementdplacdepiceenpicesursesroulettes.

  • 7/24/2019 Perfecto2 Manual de Uso

    51/228

    SECTION 6CONSIGNES D'UTILISATION

    Part No 1163146 51 Perfecto2 Srie

    VotreconcentrateuroxygneseraplusperformantsivouslefaitesfonctionnerdanslesconditionssuivantesParamtrestypesduproduitenpage46.Uneutilisationdansunenvironnement

    autrequeceluidcritpourraitncessiterunemaintenancesupplmentaire.Ladmissiondairdelappareildoittresituedansunendroitbienventilpourviterlespolluantsatmosphriqueset/oulesfumes.NEPASlutiliserdansuncagibi.

    Montage

    1. Brancherlecordondansunepriselectrique.

    2. Raccorderlhumidificateur(Sicelaatrecommand).

    MISE EN GARDE

    NEPAStropremplirlhumidificateur.

    NEPASinverserlesconnexionsentreetsortiedoxygne.Leauprovenantdelhumidificateurrepasseraparlacanulejusqulutilisateur.

    REMARQUE:Pourcetteprocdure,sereporterFIGURE6.1,FIGURE6.2enpage52,FIGURE6.3enpage53.

    1. Enlevezlebouchondelabouteille.2. Remplirlhumidificateurdeaudistillejusquauniveau

    indiquparlefabricant.Remettrelebouchondelhumidificateuretbienlefermer.

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    52/228

    SECTION 6CONSIGNES D'UTILISATION

    Perfecto2 Srie 52 Part No 1163146

    FIGURE 6.1 Remplissage de lhumidificateur3. Insreruntournevistteplatedanslarainureplaquesur

    lebordsuprieurdelatrappedaccsetsouleverdlicatementlaportedaccsdufiltrepourouvrirlatrappe(FIGURE6.2).

    4. Retirerleraccorddelhumidificateur(FIGURE6.2).5. Replacezlatrappedaccsaufiltreetremettezloue

    darationsurlatrappedaccsdufiltreeninsrantlessixtaquetssetrouvantsurlouedarationdanslessixencoches

    situessurlesctsdelatrappedaccsaufiltre.

    FIGURE 6.2 Adaptateur de lhumidificateur

    Humidificateursans bouchonHumidificateur

    avec bouchon

    Insrez letournevis tte plate ici Trappe daccs

    au filtre

    Raccord delhumidificateur

    Filtres

    Oued'aration

    Taquets

    Encoches

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    53/228

    SECTION 6CONSIGNES D'UTILISATION

    Part No 1163146 53 Perfecto2 Srie

    6. Fixezleraccorddelhumidificateursurlhumidificateurentournantlavisoreillesdelhumidificateurdanslesensdesaiguillesdunemontrejusqucequilsoitbienfix.VoirDtailA dansFIGURE6.3.

    7. Placerlhumidificateursurlecompartimenthumidificateurduconcentrateur.VoirDtailB dansFIGURE6.3.

    8. Fixerletubeoxygnelhumidificateurousurlemboutdesortiedeloxygnesurleconcentrateur.VoirDtailBdansFIGURE6.3.

    9. Fixerlacanule/letubedalimentationdupatientlasortie

    delhumidificateur.VoirDtailB dansFIGURE6.3.10. Aprsmontage,sassurerqueloxygnepassebienparlacanule.

    FIGURE 6.3 Compartiment de lhumidificateur

    InterrupteurREMARQUE:Pourcetteproc dure,sereporterFIGURE6.4.

    Humidifi-cateur

    Compartiment delhumidificateur

    Raccord delhumidifi-

    cateur

    Humidifi-cateur

    DETAIL "A" DETAIL "B"

    Tube oxygne

    Connecteur desortie

    doxygne

    Ouverture

    de l'humidifi-cateur

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    54/228

    SECTION 6CONSIGNES D'UTILISATION

    Perfecto2 Srie 54 Part No 1163146

    1. MettrelinterrupteursurlapositionMarche.Touslesvoyantslumineuxet

    lalarmesonoresallumentpendantunesecondepourindiquerquelappareilfonctionnenormalement.

    FIGURE 6.4 Interrupteur

    DbitREMARQUE:Pourcetteprocdure,sereporterFIGURE6.5.REMARQUE:VeillernepasrglerledbitaudessusdelanneauROUGE.Untauxdedbitdoxygnesuprieur5L/mindiminueralaconcentrationdoxygne.1. Tournerleboutondedbitsurlerglagerecommandpar

    lemdecin.

    MISE EN GARDENEPASmodifierlerglageL/mndudbitmtretantquunchangementnapastdemandparlemdecin.

    REMARQUE:Pourbienlireledbitmtre,reprerletraitdedbitrecommande.Tournerensuiteleboutondedbitdefaonvoirlabillemonterjusquautrait.Maintenant,centrerlabillesurletraitL/mnrecommand.

    Interrupteurmarche/arrt

    (I/O)

    http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    55/228

    SECTION 6CONSIGNES D'UTILISATION

    Part No 1163146 55 Perfecto2 Srie

    2. Siledbitindiquparledbitmtrechuteaudessousde0,5L/mnpendantplusduneminuteenviron,lalarmedeDBITMINIMUMsedclenche.Lalarmesonoremetdessignauxrapides.Contrlerletubeoulesaccessoirespourreprertoutblocageoupincementdutubeouundfautdelhumidificateur.Unefoisledbitnormalrtabli,savoirplusde0,5L/mn,lalarmesonoredeDEBITMINIMUMsarrte.

    REMARQUE:LutilisationdaccessoirestelsqueledbitmtrepdiatriqueetlecompresseurHomeFilldsactivelalarmedebaissededbit.

    FIGURE 6.5 Dbit

    Tmoin de la puret de l'oxygne SensO2-

    Modle IRC5PO2AW & IRC5PO2VAWCedispositifcontrlelapuretdeloxygnegnrparleconcentrateurdoxygne.Silapurettombeaudessousdesnormesprrgleslusine,lesvoyantsdutableaudecontrlesallument.

    Mise en marche initiale du concentrateurREMARQUE:Leconcentrateurpeuttreutilispendantladuredechauffedudmarrageinitial(environ30mn)enattendantquelevoyantdepuretO2vertsallume.

    Bouton dedbit

    Bille

    Dbitmtre

    Panneau avant

    0.5 L/min

  • 7/24/2019 Perfecto2 Manual de Uso

    56/228

    SECTION 6CONSIGNES D'UTILISATION

    Perfecto2 Srie 56 Part No 1163146

    Lorsquelappareilestmisenmarche,levoyantVERTsallume(SYSTMECORRECT/O2suprieur85%).Aprs5minutes,lecapteurdoxygnefonctionnenormalementetcommandeles

    voyantslumineuxenfonctiondelaconcentrationdoxygnedfinie.

    Explication des tmoins lumineux de la puret del'ocygne - Modle IRC5PO2AW & IRC5PO2VAWREMARQUE:Pourcetteprocdure,sereporterFIGURE6.6en

    page56.

    VoyantVERT(O2) Fonctionnementnormal.

    TmoinlumineuxJAUNE( ) Appelezlefournisseurimmdiatement.Lepatientpeutcontinuerutiliserleconcentrateursaufinstructionscontrairesdelapartduprestataire.Sassurerquelarservedoxygneestproximit.

    VoyantROUGE( ) Arrttotaldelappareil.Passezimmdiatementsurunapportdoxygnedesecoursetappelezlefournisseur.TmoinlumineuxVERT avecunclignottantJAUNE appelezlefournisseurimmdiatement.Mauvaisfonctionnementducapteurdoxygne;lepatientpeutcontinuerdutiliserleconcentrateur.

    FIGURE 6.6 Explication des tmoins lumineux de la puret del'ocygne - Modle IRC5PO2AW & IRC5PO2VAW

    O2

    Arrt automatiqueUtilisation de la

    rserve

    Infrieure lanormale

    Normale

    ROUGE

    Appeler leprestataire

    JAUNE

    VERTE

    PURET DE LOXYGNE

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    57/228

    SECTION 6CONSIGNES D'UTILISATION

    Part No 1163146 57 Perfecto2 Srie

    Explication des tmoins lumineux - ModleIRC5PAWREMARQUE:Pourcetteprocdure,sereporterFIGURE6.7.

    Modle IRC5PAW - Explication des tmoins lumineux

    VoyantROUGE( ) Arrttotaldelappareil.Passezimmdiatementsurunapportdoxygnedesecoursetappelezlefournisseur.

    VoyantVERT(I/O) Marche/arrt.Systmecorrect.

    FIGURE 6.7 Explication des tmoins lumineux - ModleIRC5PAW

    Compteur horaire

    Lecompteurhoraireaffichelecumuldesheuresdefonctionnementdelamachine.

    Arrt automatiqueUtilisation de la

    rserve

    Normale

    ROUGE

    VERTE

    Appeler leprestataire

    http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    58/228

    SECTION 7ENTRETIEN

    Perfecto2 Srie 58 Part No 1163146

    SECTION 7ENTRETIEN

    MISE EN GARDE

    LesconcentrateursInvacaresontspcialementconuspourminimiserlesoprationsdemaintenanceprventivederoutine.Seuleslespersonnesqualifiessontautoriseseffectuerlentretienderoutineduconcentrateur.

    Dbrancherleconcentrateurlorsdunettoyage.Pourviterlesrisquesdlectrocution,NEPASretirerlecapotsuprieur.

    REMARQUE:LamaintenanceprventiveDOITaumoinstreeffectueconformmentauxinstructionsducarnetdemaintenance.Unemaintenanceplusfrquentepeutsavrerncessairesurlessitestrspoussireuxouimpliquantdesniveauxdesuieslevs.RefertoRapportdemaintenanceprventiveonpage 60.

    Entretien de routineLentretienderoutinecidessousdoittreeffectuplus

    frquemment.RegardezlesintervallesrecommandsdanschaquesectionetRapportdemaintenanceprventiveenpage60.

    Nettoyage des filtres

    ATTENTION

    NEPASfairefonctionnerleconcentrateursilesfiltresnesontpasinstalls.

    REMARQUE:Pourcetteprocdure,sereporterFIGURE7.1en

    page59.REMARQUE:Ilyaunfiltresitularrireducaisson.

    1. Vrifiezlefiltrechaquemois.Enlevezlefiltreetnettoyezlesincessaire.

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 7/24/2019 Perfecto2 Manual de Uso

    59/228

    SECTION 7ENTRETIEN

    Part No 1163146 59 Perfecto2 Srie

    REMARQUE:Lesconditionsenvironnementalesncessitantunevrificationplusfrquenteetlenettoyagedufiltreincluent,maisneselimitentpas:poussiredense,polluantsatmosphriques,etc.

    2. Nettoyerlefiltrelaspirateurouleauchaudesavonneuseetlesrincersoigneusement.

    3. Schezlefiltresoigneusementavantdeleremettre.

    FIGURE 7.1 Nettoyage des filtres

    Nettoyer le caisson suprieur

    1. Nettoyerlecaissonsuprieuravecunproduitnettoyantdouxetunchiffonouunepongedouce.

    Nettoyer l'humidificateurREMARQUE:Pournettoyerlhumidificateurdoxygne,suivrelesinstructionsfourniesparlefabricant.Sinon,suivreces

    TAPES:1. Nettoyerlhumidificateurtouslesjours.

    2. Lelaverleausavonneuseetlerinceravecunesolutiondedix(10)volumesdeaupourun(1)devinaigre.

    3. Lerincersoigneusementleauchaudeetleremplirdeaudistillejusquauniveauindiqusurlhumidificateur.

  • 7/24/2019 Perfecto2 Manual de Uso

    60/228

    SECTION 7ENTRETIEN

    Perfecto2 Srie 60 Part No 1163146

    FIGURE 7.2 Rapport de maintenance prventive

  • 7/24/2019 Perfecto2 Manual de Uso

    61/228

    SECTION 8GUIDE DE DPANNAGE

    Part No 1163146 61 Perfecto2 Srie

    SECTION 8GUIDE DE DPANNAGESYMPTME : CAUSE PROBABLE : SOLUTION :

    Alarme :

    Bips sonores,courts, longue

    Le concentrateurne fonctionne pas,Linterrupteur estsur marche

    Bip.Bip.

    Perte dalimentation

    principale :

    1. Le cordon nest pasbranch.

    1. Insrer la fiche dans laprise.

    2. Pas de courant lasortie.

    2. Contrler le disjoncteur oules fusibles. Si le problme seproduit de nouveau, utiliserune autre prise.

    3. Disjoncteur dclench. 3. Pousser/rinitialiser lecoupe circuit. Si le problmepersiste, appelez votreservice client.

    Alarme :

    Continue

    Le concentrateurne fonctionne pas.Linterrupteur estsur marche

    Biiiiiiiip

    Dfaillance dusystme :

    1. Surchauffe de lappareildu fait du blocage deladmission dair.

    1a. Retirer le filtre caisson etle nettoyer.

    1b. loigner le concentrateurdoxygne au moins 7,5 cmdes murs, rideaux oumeubles.

    2. Puissance insuffisante

    la sortie.

    2. NE PAS utiliser de rallonge.

    Passer une autre circuit ouprise lectrique.

    3. Rparations internesrequises.

    3. Appeler le serviceaprs-vente.

  • 7/24/2019 Perfecto2 Manual de Uso

    62/228

    SECTION 8GUIDE DE DPANNAGE

    Perfecto2 Srie 62 Part No 1163146

    Voyant JAUNE ouROUGE allum.

    Alarme :ContinueVoyant ROUGEuniquement.

    * S'appliqueuniquement surles modlesIRC5PO2AW &IRC5PO2VAW

    1. Puret doxygnefaible.*

    1. Nettoyer ou remplacer lefiltre.

    2. Tube, canule ouhumidificateur coud oubloqu.*

    2. Contrler labsence decoudes ou blocages. Rectifier,nettoyer ou remplacer larticle.Une fois la correctionexcute, mettre hors tensionpendant 60 secondes, puis denouveau sous tension.

    3. Dbitmtre

    0.5 L/min *.

    3. Vrifier que le dbitmtre

    est rgl sur 1.0 L/mn ou plus.Se reporter Paramtrestypes du produit en page46.

    4. Surchauffe de lappareildu fait du blocage deladmission dair.

    4a. Retirer le filtre caisson etle nettoyer.4b. loigner le concentrateur au moins 7,5 cm des murs,rideaux ou meubles.

    5. Puissance insuffisante la sortie.

    5. NE PAS utiliser de rallonge.Passer une autre circuit ouprise lectrique.

    6. Rparations internesrequises.

    6. Appeler le serviceaprs-vente.

    Voyant VERTwavec

    clignotementvoyant JAUNE

    1. Rparations internesrequises.

    1. Appeler le serviceaprs-vente.

    Alarme :

    Rapide

    Bip BipBip Bip

    ALARME DE BAISSEDE DEBIT

    1. Tube, canule ouhumidificateur coud oubloqu.

    1. Contrler labsence decoudes ou blocages.Rectifier, nettoyer ou remplacer

    larticle. Une fois la correctionexcute, mettre hors tensionpendant 60 secondes, puis denouveau sous tension.

    2. Dbitmtre 0.5 L/min 2. Les dbits infrieurs 1 l/mnne sont pas recommands

    SYMPTME : CAUSE PROBABLE : SOLUTION :

  • 7/24/2019 Perfecto2 Manual de Uso

    63/228

    SECTION 9ACCESSOIRES EN OPTION

    Part No 1163146 63 Perfecto2 Srie

    SECTION 9ACCESSOIRES ENOPTION

    Lesaccessoiresenoptioncidessoussontgalementdisponibles:

    Cannulenasaleadultestandard,2.1m(7feet) M3120

    Humidificateurstandard M5120

    DbitmtrepdiatriquePreciseRX/DbitmtreIRCPF16AW

    CompresseuroxygneparticulierHomeFillIOH200AW

  • 7/24/2019 Perfecto2 Manual de Uso

    64/228

    INFORMATIONS DE RECYCLAGE

    Perfecto2 Srie 64 Part No 1163146

    INFORMATIONS DE RECYCLAGECeproduitatlivrparunfabricantsoucieuxdelenvironnementetconformeladirective2002/96/CEsurlamiseaurebutdesquipementslectroniquesetlectriquesWEEE.Ceproduitestsusceptibledecontenirdessubstancespotentiellementnocivespourlenvironmentsilestmisaurebutdansdesendroits(dcharges)nonconformeslalgislationenvigueur.

    Lesymbolepoubellebarreestappossurceproduitpourvousencouragerlerecyclerdanslesstructuresdecollecteappropries.

    Veuillezvouscomporterdemanireresponsablevisvisdelenvironnementetrecyclerceproduitarrivenfindevieparlebiaisduncentrederecyclage.

  • 7/24/2019 Perfecto2 Manual de Uso

    65/228

    Part No 1163146 65 Perfecto2 Srie

    LIMITES DE GARANTIEBelgium&Luxemburg:Invacarenv,Autobaan22,B8210LoppemTel:(32)(0)50831010

    Fax:(32)(0)[email protected]:InvacareA/S,Sdr.Ringvej37,DK2605BrndbyTel:(45)(0)36900000Fax:(45)(0)[email protected]:InvacareAquatecGmbH,Alemannenstrae10,D88316Isny

    Tel:(49)(0)75627000Fax:(49)(0)[email protected]:Invacare,Kleiststrae49,D32457PortaWestfalicaTel:(49)(0)5731754540Fax:(49)(0)[email protected]:InvacareSA,c/Arenys/n,Polgon

    IndustrialdeCelr,E17460Celr(Girona)Tel:(34)(0)972493200Fax:(34)(0)[email protected]:InvacarePoirierSAS,Route deStRoch,F37230 FondettesTel:(33)(0)247626466Fax:(33)(0)[email protected]:

    InvacareAG,Benkenstrasse260,CH4108 WitterswilTel:(41)(0)614877080Fax:(41)(0)[email protected]:InvacareIrelandLtd,Unit5SeatownBusinessCampus,SeatownRoad,Swords,CountyDublinIrelandTel:(353)18107084

    Fax:(353)[email protected]:InvacareMeccSans.r.l.,Via deiPini62,I36016Thiene(VI)Tel:(39)0445380059Fax:(39)[email protected]

    Nederland:InvacareBV,Celsiusstraat46,NL6716BZEde

    Tel:(31)(0)318695757Fax:(31)(0)[email protected]@invacare.comNorge:InvacareAS,Grensesvingen 9,Postboks6230,Etterstad,N0603OsloTel:(47)(0)22579500Fax:(47)(0)[email protected]@invacare.com

    sterreich:InvacareAustriaGmbH,HerzogOdilostrasse 101,A5310 MondseeTel:(43)623255350Fax:(43)[email protected]&Suomi:InvacareAB,Fagerstagatan 9,S163 91 SpngaTel:(46)(0)87617090Fax:(46)(0)87618108

    [email protected]@invacare.comUnitedKingdom:InvacareLimited,PencoedTechnologyPark,Pencoed,BridgendCF355HZTel:(44)(0)1656776200Fax:(44)(0)1656776201CustomerServices:Tel:(44)(0)1656776222Fax:(44)(0)[email protected]:InvacareLda,RuaEstradaVelha,949,P4465784LeadoBalioTel:(351)(0)225105946/47Fax:(351)(0)[email protected]

    mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:conta