Dealer: This manual MUST be given to the enduser.User: BEFORE using this product, read this
manual and save for future reference.
Operators Manual
For more information regardingInvacare products, parts, and services,
please visit www.invacare.com
Perfecto2
SeriesOxygen Concentrators without SensO2,
Oxygen Concentrators with SensO2
HomeFill Compatible
Model IRC5PO2AWModel IRC5PO2VAW
Model IRC5PAW
English . . . . . . . . . . . . . . . . . 1French . . . . . . . . . . . . . . . . 34Italian . . . . . . . . . . . . . . . . . 66
Spanish. . . . . . . . . . . . . . . . 97Dutch. . . . . . . . . . . . . . . . 129Greek . . . . . . . . . . . . . . . 161Portuguese . . . . . . . . . . . 195
0
434
-
7/24/2019 Perfecto2 Manual de Uso
2/228
WARNING
Perfecto2Series 2 Part No 1163146
WARNINGDONOTusethisproductoranyavailableoptionalequipment
withoutfirstcompletelyreadingandunderstandingtheseinstructionsandanyadditionalinstructionalmaterialsuchasownersmanuals,servicemanualsorinstructionsheetssuppliedwiththisproductoroptionalequipment.Ifyouareunabletounderstandthewarnings,cautionsorinstructions,contactahealthcareprofessional,dealerortechnicalpersonnelbeforeattemptingtousethisequipment otherwise,injuryordamagemayoccur.
ACCESSORIES WARNINGInvacareproductsarespecificallydesignedandmanufacturedforuseinconjunctionwithInvacareaccessories.Accessoriesdesignedbyothermanufacturers
havenotbeentestedbyInvacareandarenotrecommendedforusewithInvacareproducts.
ACCESSORIESTherearemanydifferenttypesofhumidifiers,oxygentubing,cannulasandmasksthatcanbeusedwiththisdevice.You
shouldcontactyourlocalhomecareproviderforrecommendationsonwhichofthesedeviceswillbebestforyou.Theyshouldalsogiveyouadviceontheproperusage,maintenance,andcleaning.NOTE:Updatedversionsofthismanualcanbefoundatwww.invacare.com.
-
7/24/2019 Perfecto2 Manual de Uso
3/228
TABLE OF CONTENTS
Part No 1163146 3 Perfecto2Series
ACCESSORIES ..................................................2
SPECIAL NOTES ...............................................4
LABEL LOCATION ............................................6
SECTION 1GENERAL GUIDELINES ..................7
Radio Frequency Interference.......................................................... 8
SECTION 2FEATURES .................................. 10
SECTION 3HANDLING ................................. 11
Unpacking............................................................................................11
Inspection............................................................................................11Storage.................................................................................................11
SECTION 4TECHNICAL DESCRIPTION .......... 12
SECTION 5TYPICAL PRODUCT
PARAMETERS ............................................... 13
SECTION 6OPERATING INSTRUCTIONS ....... 17
Introduction........................................................................................17Select a Location ...............................................................................17
Set Up ..................................................................................................18
Flowrate...............................................................................................21
SensO2Oxygen Purity Indicator - Model IRC5PO2AW &
IRC5PO2VAW .........................................................................................22
Initial Startup of the Concentrator ...............................................22
Explanation of Indicator Lights - Model IRC5PAW ..................23
Elapsed Time Meter (Hour Meter) ...............................................24
SECTION 7MAINTENANCE .......................... 25
Routine Maintenance........................................................................25
SECTION 8TROUBLESHOOTING GUIDE ......... 28
SECTION 9OPTIONAL ACCESSORIES ............ 30
RECYCLING INFORMATION ........................... 31
LIMITED WARRANTY .................................... 32
-
7/24/2019 Perfecto2 Manual de Uso
4/228
SPECIAL NOTES
Perfecto2Series 4 Part No 1163146
SPECIAL NOTESSignalwordsareusedinthismanualandapplytohazardsorunsafepracticeswhichcouldresultinpersonalinjuryorpropertydamage.Refertothefollowingtablefordefinitionsofthesignalwords.
NOTICETheinformationcontainedinthisdocumentissubjecttochangewithoutnotice.
DANGERDO NOT SMOKE while using this device. Keep all
matches, lit cigarettes or other sources of ignition out ofthe room in which this product is located and away fromwhere oxygen is being delivered.
NO SMOKING signs should be prominently displayed.Textiles and other materials that normally would notburn are easily ignited and burn with great intensity inoxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physicalinjury or death.
SIGNAL WORD MEANING
DANGER
Danger indicates an imminently hazardous situationwhich, if not avoided, will result in death or seriousinjury.
WARNING
Warning indicates a potentially hazardous situationwhich, if not avoided, could result in death orserious injury.
CAUTION
Caution indicates a potentially hazardous situationwhich, if not avoided, may result in propertydamage or minor injury or both.
-
7/24/2019 Perfecto2 Manual de Uso
5/228
SPECIAL NOTES
Part No 1163146 5 Perfecto2Series
CAUTION
Caution:Statutorylawcanrestrictthisdevicetosalebyoron
orderofaphysician,oranyotherpractitionerlicensedbythelawofthegovernmentalagencyinwhichhe/shepracticestouseorordertheuseofthisdevice.
Invacarerecommendsanalternatesourceofsupplementaloxygenintheeventofapoweroutage,alarmconditionormechanicalfailure.Consultyourphysicianorequipmentproviderforthetypeofreservesystemrequired.
Thisequipmentistobeusedasanoxygensupplementandisnotconsideredlifesupportingorlifesustaining.
-
7/24/2019 Perfecto2 Manual de Uso
6/228
LABEL LOCATION
Perfecto2Series 6 Part No 1163146
LABEL LOCATION
Serial Number Label islocated on the resonator
intake assembly
Specification Label is
located on the backnear the bottom
erfecto
FOR OPERATING INSTRUCTIONS or ALARMS (See Operators Manual)
DANGER - RISK OF FIRENO SMOKING - Keep ALL sources of ignition out of the room in which this product islocated and away from areas where oxygen is being delivered. Textiles, oil and othercombustiles are easily ignited and burn with great intensity in oxygen enriched air.
DANGER - ELECTRIC SHOCKDO NOT remove cover. Refer servicing to qualified service personnel.
Flowrates below 0.5 L/min will trigger Low Flow Alarm (rapid audible alarm).
Set Flowmeter at 2.5 L/min or less when filling HomeFill cylinders.
HomeFillCompatible
-
7/24/2019 Perfecto2 Manual de Uso
7/228
SECTION 1GENERAL GUIDELINES
Part No 1163146 7 Perfecto2Series
SECTION 1GENERAL GUIDELINESInordertoensurethesafeinstallation,assemblyandoperationofthePerfecto2concentratortheseinstructionsMUSTbefollowed.
WARNINGThissectioncontainsimportantinformationforthesafeoperationanduseofthisproduct.
DANGERRiskofelectricshock.DONOTdisassemble.Referservicingtoqualifiedservicepersonnel.Nouserserviceableparts.TOREDUCETHERISKOFBURNS,ELECTROCUTION,FIREORINJURYTOPERSONS.Avoidusingwhilebathing.Ifcontinuoususageisrequiredbythephysiciansprescription,theconcentratorMUSTbelocatedinanotherroomatleast2.5m(7ft)fromthebath.DONOTcomeincontactwiththeconcentratorwhilewet.DONOTplaceorstoreproductwhereitcandropintowaterorotherliquid.DONOTreachforproductthathasfallenintowater.UnplugIMMEDIATELY.Iftheconcentratorhasadamagedcordorplug,ifitisnotworkingproperly,ifithasbeendroppedordamaged,ordroppedintowater,callqualifiedtechnicianforexaminationandrepair.Aspontaneousandviolentignitionmayoccurifoil,greaseorgreasysubstancescomeincontactwithoxygenunderpressure.ThesesubstancesMUSTbekeptawayfromtheoxygenconcentrator,tubingandconnections,andallotheroxygen
equipment.DONOTuseanylubricantsunlessrecommendedbyInvacare.Avoidcreationofanysparknearmedicaloxygenequipment.Thisincludessparksfromstaticelectricitycreatedbyanytypeoffriction.
-
7/24/2019 Perfecto2 Manual de Uso
8/228
SECTION 1GENERAL GUIDELINES
Perfecto2Series 8 Part No 1163146
Operating InformationForoptimumperformance,Invacarerecommendsthateachconcentratorbeonandrunningforaminimumof30minutes
atatime.Shorterperiodsofoperationmayreducemaximumproductlife.
Keeptheoxygentubing,cord,andunitoutfromundersuchitemsasblankets,bedcoverings,chaircushions,clothingandawayfromheatedorhotsurfaces,includingspaceheaters,stovesandsimilarelectricalappliances.
DONOTmoveorrelocateconcentratorbypullingonthepowercord.
NEVERdroporinsertanyobjectorliquidintoanyopening.
InvacarerecommendsthatCrushProofoxygentubingbeusedwiththisproductandnotexceed15.2m(50ft)inlength.
Therearenouserserviceableparts.Thisdoesnotinclude
normalmaintenanceitems.Seemaintenancesectionforusermaintenanceitems.
AproductshouldNEVERbeleftunattendedwhenpluggedin.MakesurethePerfecto2isOffwhennotinuse.
Closesupervisionisnecessarywhenthisproductisusednearchildrenorphysicallychallengedindividuals.
Additionalmonitoringorattentionmayberequiredforpatientsusingthisdevicewhoareunabletohearorseealarmsorcommunicatediscomfort.
DONOTconnecttheconcentratorinparallelorserieswithotheroxygenconcentratorsoroxygentherapydevices.
Radio Frequency InterferenceThisequipmenthasbeentestedandfoundtocomplywithEMClimitsspecifiedbyIEC/EN6060112.Theselimitsaredesignedtoprovideareasonableprotectionagainstelectromagneticinterferenceinatypicalmedicalinstallation.
-
7/24/2019 Perfecto2 Manual de Uso
9/228
SECTION 1GENERAL GUIDELINES
Part No 1163146 9 Perfecto2Series
Otherdevicesmayexperienceinterferencefromeventhelowlevelsofelectromagneticemissionspermittedbytheabovestandards.TodetermineiftheemissionsfromthePerfecto2is
causingtheinterference,turnthePerfecto2Off.Iftheinterferencewiththeotherdevice(s)stops,thenthePerfecto2iscausingtheinterference.Insuchrarecases,interferencemaybereducedorcorrectedbyoneofthefollowingmeasures:
Reposition,relocate,orincreasetheseparation
betweentheequipment. Connecttheequipmentintoanoutletonacircuit
differentfromthattowhichtheotherdevice(s)isconnected.
-
7/24/2019 Perfecto2 Manual de Uso
10/228
SECTION 2FEATURES
Perfecto2Series 10 Part No 1163146
SECTION 2FEATURES
Oxygen Outlet
Flowmeter
Oxygen Purity IndicatorLights / Fault and Power
Indicator Lights
Power Switch
Circuit Breaker
Elapsed TimeMeter
CabinetFilter
REAR VIEW
*HF OutletFitting
PowerCord
*NOTE:ThisoutletfittingistobeusedonlyforfillingoxygencylinderswiththeHomeFillhomeoxygencompressor.Theoutlet
fittingdoesnotaffectconcentratorperformance.RefertotheHomeFillownersmanual,partnumber1145805,forconnectionandoperatinginstructions.Whennotinuse,theplugprovidedwith
theconcentratorshouldbeinsertedintotheoutletfitting.FormoreinformationabouttheHomeFill,contactyourInvacaredealer.
-
7/24/2019 Perfecto2 Manual de Uso
11/228
SECTION 3HANDLING
Part No 1163146 11 Perfecto2Series
SECTION 3HANDLINGTheconcentratorshouldALWAYSbekeptintheuprightpositiontopreventcabinetdamagewhilebeingtransported.
Iftheconcentratoristobereshippedbycommoncarrier,additionalcartonsareavailablefromInvacare.
UnpackingNOTE:Forthisprocedure,refertoFIGURE3.1.
1. Checkforanyobvious
damagetothecartonoritscontents.Ifdamageisevident,notifythecarrier,oryourlocaldealer.
2. Removeallloosepackingfromthecarton.
3. Carefullyremoveallthecomponentsfromthecarton. FIGURE 3.1 Unpacking
NOTE:UnlesstheoxygenconcentratoristobeusedIMMEDIATELY,leaveconcentratorinitspackagingforstorageuntiluseoftheconcentratorisrequired.
Inspection1. Inspect/examineexterioroftheoxygenconcentratorfor
nicks,dents,scratchesorotherdamages.Inspectallcomponents.
Storage1. Storetherepackagedoxygenconcentratorinadryarea.
2. DONOTplaceobjectsontopofrepackagedconcentrator.
-
7/24/2019 Perfecto2 Manual de Uso
12/228
SECTION 4TECHNICAL DESCRIPTION
Perfecto2Series 12 Part No 1163146
SECTION 4TECHNICALDESCRIPTION
TheInvacarePerfecto2concentratorisusedbypatientswithrespiratorydisorderswhorequiresupplementaloxygen.Thedeviceisnotintendedtosustainorsupportlife.
Theoxygenconcentrationleveloftheoutputgasrangesfrom87%to95.6%.Theoxygenisdeliveredtothepatientthroughtheuseofanasalcannula.
TheInvacarePerfecto2concentratorusesamolecularsieveandpressureswingadsorptionmethodologytoproducetheoxygengasoutput.Ambientairentersthedevice,isfilteredandthencompressed.Thiscompressedairisthendirectedtowardoneoftwonitrogenadsorbingsievebeds.Concentratedoxygenexitstheoppositeendoftheactivesievebedandisdirectedintoanoxygenreservoirwhereitis
deliveredtothepatient.TheInvacarePerfecto2concentratoriscapableofoperationbythepatientinahomeenvironmentorinaninstitutionalenvironment.Deviceoperatesatanominal230VAC/50Hertzsupply.
Serviceinformationwillbeavailableuponrequestto
qualifiedtechnicalpersonnelonly.
-
7/24/2019 Perfecto2 Manual de Uso
13/228
SECTION 5TYPICAL PRODUCT PARAMETERS
Part No 1163146 13 Perfecto2Series
SECTION 5TYPICAL PRODUCTPARAMETERS
Alternating Current
Type BF equipment
Unit running
Unit not running
Attention - Consider Accompanying Documents
DO NOT smoke
Class II, Double Insulated
Protected against dripping water
ElectricalRequirements: 230 VAC 10% (253 VAC/207 VAC), 50 Hz
Rated Current Input:1.4 A
Sound Level: 39.5 dBA Average
Altitude: All Perfecto2Models:Up to 8,000 ft (2438 meters) above sea
level without degradation of concentration levels. From 8,000 ft to13,129 (4000 meters) below 90% efficiency.
All Perfecto2Models while filling a HomeFill System:
Up to 6,000 ft (1828 meters) above sea level without degradationof concentration levels. From 6,000 ft (1828 meters) to 13,129 ft
(4000 meters) below 90% efficiency
IPX1
-
7/24/2019 Perfecto2 Manual de Uso
14/228
SECTION 5TYPICAL PRODUCT PARAMETERS
Perfecto2Series 14 Part No 1163146
Oxygen OutputConcentrationLevels:NOTE:Stated
concentrationlevelsachievedafterinitialwarmupperiod(approximately30minutes)
87% to 95.6% at 0.5 to 5 L/min93% 3% at 2 L/min
Maximum OutletPressure: 34.5 kPa 3.45 kPa (5 psi 0.5 psi)
Flow Range: 0.5 to 5 L/min (maximum)For flowrates less than 1 L/min, we recommend the use of the
Invacare Pediatric Flowmeter Accessory (IRCPF16AW)Low Flow Alarm: 0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No Accessories Connected)
Average PowerConsumption: Perfecto
2AW - 300 W @ 5 L/min, 280 W @ 3 L/min
Perfecto2VAW - 320 W@ 5 L/min
Pressure Relief
MechanismOperational at: 241 kPa 24.1 kPa (35 psi 3.5 psi)
Change in maximumrecommended flowwhen back pressureof 7kPa is applied: 0.7 L/min
Filters: Cabinet, Outlet HEPA and Compressor Inlet
Safety System: Current overload or line surge shutdownHigh temperature compressor shutdownHigh Pressure Alarm w/compressor shutdownLow Pressure Alarm w/compressor shutdownBattery Free Power Loss Alarm. SensO
2Oxygen System (SensO
2
Model) Low Flow Alarm.
Width: 38.1 cm 1 cm (15 in 3/8 in)
Height: 58.4 cm 1 cm (23 in 3/8 in)
Depth: 30.5 cm 1 cm (12 in 3/8 in)
Weight: Perfecto2AW Models - 20.5 kg 1 kg (45 lbs 2 lbs)
Perfecto2VAW Models - 18.1 kg 1 kg (40 lbs 2 lbs)
Shipping Weight: Perfecto2AW Models - 22.7 kg 1 kg (50 lbs 2 lbs)
Perfecto2VAW Models - 20.4 kg 1 kg (45 lbs 2 lbs)
>
-
7/24/2019 Perfecto2 Manual de Uso
15/228
SECTION 5TYPICAL PRODUCT PARAMETERS
Part No 1163146 15 Perfecto2Series
Operating AmbientTemperature: 10C - 35C (50F - 95F) at 20-60% relative humidity
Exhaust Cooling AirTemperature: Less than Ambient +19C (+ 35F)
Oxygen OutputTemperature: Less than Ambient +3C(+ 6F)
Cabinet: Impact Resistant flame-retardant plastic cabinet that conforms toUL 94-V0
No AP/APG Not suitable for use in the presence of a flammable anaestheticmixture
Standards andRegulatory Listing: IRC5PO2AW/IRC5PAW/IRC5PO2VAW:
IEC/EN 60601-1, A1, A2IEC/EN 60601-1-2IEC/EN61000-3-2IEC/EN61000-3-3
IRC5PO2AW/IRC5PO2VAW Models OnlyISO8359
MDD 93/42/EEC, Annex I and IXCE marked model IRC5PO2AW & IRC5PO2VAW
Electrical: No extension cords
Placement: No closer than 7.5 cm (3 in) from any wall, furniture, draperies, orfurniture to assure sufficient air flowAvoid deep pile carpets and heaters, radiators or hot air registersFloor location onlyNo confined spaces (Example: no closets)
Tubing: 2 m (7 ft) cannula with a maximum 15 m (50 ft) of Crush-ProofTubing (DO NOT pinch)
Time of Operation: Up to 24 hours per day
RecommendedStorage and ShippingTemperature: -29C to 65C (-20F to 150F) at 15-95% relative humidity
Environment: Smoke and soot-freeNo confined spaces (Example: No closets).
Minimum OperatingTime: 30 Minutes
-
7/24/2019 Perfecto2 Manual de Uso
16/228
SECTION 5TYPICAL PRODUCT PARAMETERS
Perfecto2Series 16 Part No 1163146
IRC5PO2AW & IRC5PO2VAW INDICATORS
IRC5PAW INDICATORS
LABELSYMBOL
O2PURITY INDICATOR LIGHTS
(LED)
SYSTEM OKAYO
2over 85% ( 2%)
GREEN Indicator Light
O2Between 73% ( 3%)
to 85% (2%)
YELLOW Indicator lightA. YELLOW SolidB. YELLOW Flashing Sensor FailureCall a qualified technician.
SYSTEM FAILUREO2Below 73% (3%) RED Indicator LightContinuous Audible Alarm
Sieve-GARD Compressor Shutdown
LABELSYMBOL
STATUS INDICATOR LIGHTS(LED)
SYSTEM OKAY GREEN Indicator Light
SYSTEM FAILURE RED Indicator lightContinuous Audible AlarmSieve-GARDCompressorShutdownCall a qualified technician.
O2
I/O
-
7/24/2019 Perfecto2 Manual de Uso
17/228
SECTION 6OPERATING INSTRUCTIONS
Part No 1163146 17 Perfecto2Series
SECTION 6OPERATINGINSTRUCTIONS
IntroductionYouroxygenconcentratorisintendedforindividualuse.Itisanelectronicallyoperateddevicethatseparatesoxygenfromroomair.Itprovideshighconcentrationofoxygendirectlytoyouthroughanasalcannula.Clinicalstudieshavedocumentedthatoxygenconcentratorsaretherapeutically
equivalenttoothertypesofoxygendeliverysystems.Yourproviderwillshowyouhowtouseyouroxygenconcentrator.He/Sheshouldbecontactedwithanyquestionsorproblemsregardingyouroxygenconcentrator.Thisownersmanualwilltellyouaboutyourconcentratorandwillserveasareferenceasyouuseyourconcentrator.
Select a Location
WARNING
NEVERblocktheairopeningsoftheproductorplaceitonasoftsurface,suchasabedorcouch,wheretheairopeningmay
beblocked.Keeptheopeningsfreefromlint,hairandthelike.Keepunitatleast7.5cm(3in)awayfromwalls,draperies,furniture,andthelike.
Youmayselectaroominyourhousewhereusingyouroxygenconcentratorwouldbemostconvenient.Yourconcentratorcan
beeasilyrolledfromroomtoroomonitscasters.
-
7/24/2019 Perfecto2 Manual de Uso
18/228
SECTION 6OPERATING INSTRUCTIONS
Perfecto2Series 18 Part No 1163146
YouroxygenconcentratorwillperformbestwhenoperatedundertheconditionsoutlinedintheTypicalProductParametersonpage 13.Usageinenvironmentsotherthan
thosedescribedmayresultintheneedforincreasedequipmentmaintenance.Theairintakeoftheunitshouldbelocatedinawellventilatedareatoavoidairbornepollutantsand/orfumes.DONOTuseinacloset.
Set Up
1. Pluginpowercordtoanelectricaloutlet.2. ConnectHumidifier(IfSoPrescribed)
WARNING
DONOToverfillhumidifier.
DONOTreversetheoxygeninputandoutputconnections.
Waterfromthehumidifierbottlewilltravelthroughthecannulabacktothepatient.
NOTE:Forthisprocedure,refertoFIGURE6.1,FIGURE6.2onpage 19,FIGURE6.3onpage 20.
1. Removecapfrombottle.
2. Fillhumidifierwithdistilledwatertothelevelindicatedbythemanufacturer.Replacethehumidifiercapandsecurelytighten.
FIGURE 6.1 Filling the Humidifier
HumidifierBottle w/o CapHumidifier
Bottle with Cap
-
7/24/2019 Perfecto2 Manual de Uso
19/228
SECTION 6OPERATING INSTRUCTIONS
Part No 1163146 19 Perfecto2Series
3. Insertaflatheadscrewdriverintheplategrooveonthetopedgeofthefilteraccessdoorandgentlyprythefilteraccessdooroff(FIGURE6.2).
4. Pullupandremovethehumidifierbottleadapter(FIGURE6.2).5. Replacethefilteraccessdoorandinstalltheairscoopontothefilteraccessdoorbyinsertingthesixtabsontheairscoopintothesixslotsonthesidesofthefilteraccessdoor.
FIGURE 6.2 Humidifer Bottle Adapter6. Attachthehumidifierbottleadaptertothehumidifier
bottlebyturningthewingnutonthehumidifierbottlecounterclockwiseuntilitissecurelyfastened.SeeDetailAinFIGURE6.3.
7. Placethehumidifierbottle/adapterassemblyinthehumidifiercompartmentontheconcentrator.SeeDetailBinFIGURE6.3.
8. Attachoxygentubingfromthehumidifierbottle/adapter
assemblytooxygenoutletconnectorontheconcentrator.SeeDetailBinFIGURE6.3.9. Attachthecannula/patientsupplytubingtothe
humidifierbottleoutlet.SeeDetailBinFIGURE6.3.10. Afterassembly,ensurethatoxygenisflowingthroughthe
cannula.
Insert FlatHead
ScrewdriverHere Filter AccessDoor
Humidifier BottleAdapter
Cabinet Filter
Air Scoop
Tabs
Slots
-
7/24/2019 Perfecto2 Manual de Uso
20/228
SECTION 6OPERATING INSTRUCTIONS
Perfecto2Series 20 Part No 1163146
FIGURE 6.3 Humidifier Compartment
Power SwitchNOTE:Forthisprocedure,refertoFIGURE6.4.
1. PresspowerswitchtoOnposition.Allthepanellightsandtheaudiblealarmwillcomeonforonesecond,indicatingthattheunitis
functioningproperly.
FIGURE 6.4 Power Switch
HumidifierBottle
Humidifier
Compartment
HumidifierBottle
AdapterHumidifier
Bottle
DETAIL A DETAIL B
OxygenTubing
OxygenOutlet
Connector
HumidifierBottleOutlet
On/Off
(I/O)PowerSwitch
-
7/24/2019 Perfecto2 Manual de Uso
21/228
SECTION 6OPERATING INSTRUCTIONS
Part No 1163146 21 Perfecto2Series
FlowrateNOTE:Forthisprocedure,refertoFIGURE6.5.NOTE:TakecarenottosettheflowaboveREDring.Anoxygen
flowgreaterthan5L/minwilldecreasetheoxygenconcentration.1. Turntheflowrateknobtothesettingprescribedbyyour
physicianortherapist.
WARNINGDONOTchangetheL/minsettingontheflowmeterunlessachangehasbeenprescribedbyyourphysicianortherapist.
NOTE:Toproperlyreadtheflowmeter,locatetheprescribedflowratelineontheflowmeter.Next,turntheflowknobuntiltheballrisestotheline.Now,centertheballontheL/minlineprescribed.2. Iftheflowrateontheflowmetereverfallsbelow0.5L/minfor
morethanaboutoneminute,theLOWFLOWalarmwillbetriggered.Thisisarapidbeepingoftheaudiblealarm.Check
yourtubingoraccessoriesforblockedorkinkedtubingoradefectivehumidifierbottle.Afterratedflowisrestoredtomorethan0.5L/min,theLOWFLOWaudiblealarmwillgooff.
NOTE:TheuseofsomeaccessoriessuchasthepediatricflowstandandtheHomeFillcompressorwilldeactivatetheLowFlowAlarm.
-
7/24/2019 Perfecto2 Manual de Uso
22/228
SECTION 6OPERATING INSTRUCTIONS
Perfecto2Series 22 Part No 1163146
FIGURE 6.5 Flowrate
SensO2Oxygen Purity Indicator - Model
IRC5PO2AW & IRC5PO2VAWThisfeaturemonitorsthepurityofoxygengeneratedbytheoxygenconcentrator.Ifpurityfallsbelowfactorypresetstandards,indicatorlightsonthecontrolpanelwillilluminate.
Initial Startup of the ConcentratorNOTE:Concentratormaybeusedduringtheinitialstartwarmuptime(approximately30min.)whilewaitingfortheO2puritytoreachmaximum.
Whentheunitisturnedon,theGREENlightwillcomeon
(SYSTEMOK/O2greaterthan85%).Afterfiveminutes,theoxygensensorwillbeoperatingnormallyandwillcontroltheindicatorlightsdependingonoxygenconcentrationvalues.
Flow Knob
Ball
Flowmeter
Front Panel
0.5 L/min
-
7/24/2019 Perfecto2 Manual de Uso
23/228
SECTION 6OPERATING INSTRUCTIONS
Part No 1163146 23 Perfecto2Series
Explanation of Oxygen Purity Indicator Lights -Model IRC5PO2AW & IRC5PO2VAWNOTE:Forthisprocedure,refertoFIGURE6.6onpage 23.
GREENlight(O2) NormalOperation.
YELLOWlight( ) Immediatelycallsupplier.Youmaycontinuetousetheconcentratorunlessinstructedotherwisebyyoursupplier.Becertainthatbackupoxygenisnearby.
REDlight( ) Totalunitshutdown.Immediatelyswitchto
backupoxygensupplyandcallsupplier.GREENlightwithYELLOWlightflashingImmediatelycallsupplier.Oxygensensormalfunctioning;youmaycontinuetousetheconcentrator.
FIGURE 6.6 Explanation of Oxygen Purity Indicator Lights -Model IRC5PO2AW & IRC5PO2VAW
Explanation of Indicator Lights - ModelIRC5PAWNOTE:Forthisprocedure,refertoFIGURE6.7.
IRC5PAW Model Indicator Light Explanation
REDlight( ) TotalUnitShutDown.Immediatelyswitchtoabackupoxygensupplyandcallsupplier.
GREENlight(I/O) On/Off.Systemokay.
O2
Auto Shut Down
Use Backup
BelowNormal
Normal
RED
Call Supplier
YELLOW
GREEN
OXYGEN PURITY
-
7/24/2019 Perfecto2 Manual de Uso
24/228
SECTION 6OPERATING INSTRUCTIONS
Perfecto2Series 24 Part No 1163146
FIGURE 6.7 Explanation of Indicator Lights - Model IRC5PAW
Elapsed Time Meter (Hour Meter)
Thehourmeterdisplaysthecumulativenumberofhourstheunithasoperated.
Auto Shut DownUse Backup
Normal
RED
GREEN
Call Supplier
-
7/24/2019 Perfecto2 Manual de Uso
25/228
SECTION 7MAINTENANCE
Part No 1163146 25 Perfecto2Series
SECTION 7MAINTENANCE
WARNING
TheInvacareconcentratorsarespecificallydesignedtominimizeroutinepreventivemaintenance.Onlyqualifiedpersonnelshouldperformpreventivemaintenanceontheconcentrator.
Unplugtheconcentratorwhencleaning.Toavoidelectricalshock,DONOTremovecabinet.
NOTE:Ataminimum,preventivemaintenanceMUSTbeperformedaccordingtothemaintenancerecordguidelines.Inplaceswithhighdustorsootlevels,maintenancemayneedtobeperformedmoreoften.RefertoPreventiveMaintenanceRecordonpage 27.
Routine MaintenanceThefollowingroutinemaintenanceshouldbeperformedmore
frequently.SeerecommendedintervalsineachsectionandPreventiveMaintenanceRecordonpage 27.
Cleaning the Cabinet Filter
CAUTIONDONOToperatetheconcentratorwithoutthefilterinstalled.
NOTE:Forthisprocedure,refertoFIGURE7.1onpage 26.
NOTE:Thereisonecabinetfilterlocatedonthebackofthecabinet.
1. Inspectthefiltereverymonth.Removethefilterandcleanasneeded.
NOTE:Enviromentalconditionsthatmayrequiremorefrequentinspectionandcleaningofthefilterinclude,butarenotlimitedto:highdust,airpollutants,etc.
-
7/24/2019 Perfecto2 Manual de Uso
26/228
SECTION 7MAINTENANCE
Perfecto2Series 26 Part No 1163146
2. Cleanthecabinetfilterwithavacuumcleanerorwashinwarmsoapywaterandrinsethoroughly.
3. Drythefilterthoroughlybeforereinstallation.
FIGURE 7.1 Cleaning the Cabinet Filter
Cleaning the Cabinet
1. Cleanthecabinetwithamildhouseholdcleanerandnonabrasiveclothorsponge.
Cleaning the HumidifierNOTE:Tocleantheoxygenhumidifier,followtheinstructionsprovidedbythemanufacturer.Ifnoneareprovided,followtheseSTEPS:
1. Cleanthehumidifiereveryday.
2. Washitinsoapywaterandrinseitwithasolutionoften
partswaterandonepartvinegar.3. Rinsethoroughlywithhotwaterandrefillwithdistilled
watertothelevelshownonthehumidifier.
-
7/24/2019 Perfecto2 Manual de Uso
27/228
SECTION 7MAINTENANCE
Part No 1163146 27 Perfecto2Series
FIGURE 7.2 Preventive Maintenance Record
-
7/24/2019 Perfecto2 Manual de Uso
28/228
SECTION 8TROUBLESHOOTING GUIDE
Perfecto2Series 28 Part No 1163146
SECTION 8TROUBLESHOOTINGGUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:Alarm:
Short beeps, longpause
Concentrator notoperating, power
switch On.
Beep......Beep......
Main Power Loss:
1. Power cord notplugged in.
1. Insert plug into outlet.
2. No power at outlet. 2. Inspect house circuitbreakers or fuses. If problemrecurs, use a different outlet.
3. Tripped circuitbreaker.
3. Push/reset circuit breaker.If problem recurs, call ServiceProvider.
Alarm:Continuous
Concentrator not
operating, PowerSwitch On.
Beeeeeeeep....
System Failure:
1. Unit overheating dueto blocked air intake.
1a. Remove and clean cabinetfilters.
1b. Move oxygenconcentrator at least threeinches away from walls,draperies or furniture.
2. Insufficient power atoutlet.
2. DO NOT use extensioncords. Move to anotherelectrical outlet or circuit.
3. Internal repairsrequired.
3. Call Service Provider.
-
7/24/2019 Perfecto2 Manual de Uso
29/228
SECTION 8TROUBLESHOOTING GUIDE
Part No 1163146 29 Perfecto2Series
YELLOW or REDlight Illuminated.
Alarm:ContinuousOn RED lightonly.
* Only applies toIRC5PO2AW &IRC5PO2VAWmodels
1. Low oxygen purity.* 1. Clean or Replace filters.
2. Kinked or blockedtubing, cannula orhumidifier.*
2. Inspect for kinks orblockages. Correct, clean orreplace item. Oncecorrected, turn power Off for60 seconds and then turnpower back On.
3. Flowmeter set at 0.5
L/min *.
3. Check flowmeter is set to
1.0 L/min or more. Refer toTypical Product Parameterson page 13.
4. Unit overheating dueto blocked air intake.
4a. Remove and clean cabinetfilters.4b. Move concentrator atleast three inches from walls,draperies, and furniture.
5. Insufficient power atoutlet.
5. DO NOT use extensioncords. Move to anotherelectrical outlet or circuit.
6. Internal repairsrequired.
6. Call Service Provider.
GREEN light withYELLOW Light
Flashing.
1. Internal repairsrequired.
1. Call Service Provider.
Alarm:
Rapid
Beep....Beep...Beep...Beep
LOW FLOW ALARM
1. Kinked or blockedtubing, cannula orhumidifier.
1. Inspect for kinks orblockages. Correct, clean orreplace item. Oncecorrected, turn power Off for60 seconds and then turnpower back ON.
2. Flowmeter set at 0.5L/min.
2. Flowrates less than 1 L/minare not recommend.
SYMPTOM: PROBABLE CAUSE: SOLUTION:
-
7/24/2019 Perfecto2 Manual de Uso
30/228
SECTION 9OPTIONAL ACCESSORIES
Perfecto2Series 30 Part No 1163146
SECTION 9OPTIONALACCESSORIES
Thefollowingoptionalaccessoriesarealsoavailable: StandardAdultNasalCannula,2.1m(7feet) M3120
StandardHumidifier M5120
PreciseRXPediatricHumidifier/FlowmeterAccessory IRCPF16AW
HomeFillhomeoxygencompressor IOH200AW
-
7/24/2019 Perfecto2 Manual de Uso
31/228
RECYCLING INFORMATION
Part No 1163146 31 Perfecto2Series
RECYCLING INFORMATIONThisproducthasbeensuppliedfromanenvironmentallyawaremanufacturerthatcomplieswiththeWasteElectricalandElectronicEquipment(WEEE)Directive2002/96/CE.Thisproductmaycontainsubstancesthatcouldbeharmfultotheenvironmentifdisposedofinplaces(landfills)thatarenotappropriateaccordingtolegislation.
Thecrossedoutwheeliebin symbolisplacedonthisproducttoencourageyoutorecyclewhereverpossible.
Pleasebeenvironmentallyresponsibleandrecyclethisproductthroughyourrecyclingfacilityatitsendoflife.
-
7/24/2019 Perfecto2 Manual de Uso
32/228
Perfecto2Series 32 Part No 1163146
LIMITED WARRANTYBelgium&Luxemburg:Invacarenv,Autobaan22,B8210LoppemTel:(32)(0)50831010
Fax:(32)(0)[email protected]:InvacareA/S,Sdr.Ringvej37,DK2605BrndbyTel:(45)(0)36900000Fax:(45)(0)[email protected]:InvacareAquatecGmbH,Alemannenstrae10,D88316Isny
Tel:(49)(0)75627000Fax:(49)(0)[email protected]:Invacare,Kleiststrae49,D32457PortaWestfalicaTel:(49)(0)5731754540Fax:(49)(0)[email protected]:InvacareSA,c/Arenys/n,Polgon
IndustrialdeCelr,E17460Celr(Girona)Tel:(34)(0)972493200Fax:(34)(0)[email protected]:InvacarePoirierSAS,Route deStRoch,F37230 FondettesTel:(33)(0)247626466Fax:(33)(0)[email protected]:
InvacareAG,Benkenstrasse260,CH4108 WitterswilTel:(41)(0)614877080Fax:(41)(0)[email protected]:InvacareIrelandLtd,Unit5SeatownBusinessCampus,SeatownRoad,Swords,CountyDublinIrelandTel:(353)18107084
Fax:(353)[email protected]:InvacareMeccSans.r.l.,Via deiPini62,I36016Thiene(VI)Tel:(39)0445380059Fax:(39)[email protected]
Nederland:InvacareBV,Celsiusstraat46,NL6716BZEde
Tel:(31)(0)318695757Fax:(31)(0)[email protected]@invacare.comNorge:InvacareAS,Grensesvingen 9,Postboks6230,Etterstad,N0603OsloTel:(47)(0)22579500Fax:(47)(0)[email protected]@invacare.com
sterreich:InvacareAustriaGmbH,HerzogOdilostrasse 101,A5310 MondseeTel:(43)623255350Fax:(43)[email protected]&Suomi:InvacareAB,Fagerstagatan 9,S163 91 SpngaTel:(46)(0)87617090Fax:(46)(0)87618108
[email protected]@invacare.comUnitedKingdom:InvacareLimited,PencoedTechnologyPark,Pencoed,BridgendCF355HZTel:(44)(0)1656776200Fax:(44)(0)1656776201CustomerServices:Tel:(44)(0)1656776222Fax:(44)(0)[email protected]
Portugal:InvacareLda,RuaEstradaVelha,949,P4465784LeadoBalioTel:(351)(0)225105946/47Fax:(351)(0)[email protected]
mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected] -
7/24/2019 Perfecto2 Manual de Uso
33/228
Part No 1163146 33 Perfecto2Series
NOTES
-
7/24/2019 Perfecto2 Manual de Uso
34/228
Distributeur : Ce manuel DOIT tre remis lutilisateur final.Utilisateur
: Lire ce manuel AVANT d'utiliser cet
appareil, et le conserver en cas de besoin.
Manuel dutilisation
Pour de plus amples informations sur lesproduits Invacare, les pices dtaches et le
service client, veuillez vous rendre surwww.invacare.fr
Perfecto2
SrieConcentrateurs oxygne sans SensO2,
Concentrateurs oxygne avec SensO2
Compatible HomeFill
Modle IRC5PO2AW
Modle IRC5PO2VAW
Modle IRC5PAW
0
434
-
7/24/2019 Perfecto2 Manual de Uso
35/228
MISE EN GARDE
Part No 1163146 35 Perfecto2 Srie
MISE EN GARDENEPASutilisercetappareiloutoutautrequipement
optionnelsansavoirpralablementluetcompriscesinstructionsettoutdocumentdinstructionadditionneltelquelemanueldelutilisateur,manueldentretienoufeuilletsfournisavecceproduitoutoutmatrieloptionnel.Silutilisateurnecomprendpaslesmisesengarde,messagesdattentionouinstructions,ildoitcontacterunprofessionneldelasant,ledistributeuroulepersonneltechniqueavantdetenterdutilisercetappareil Sinon,ilyaunrisquedeblessurecorporelleoudedommagematriel.
MISE EN GARDE POUR LESACCESSOIRES
LesproduitsInvacareonttspcifiquementconusetfabriquspourtreutilissconjointementaveclesaccessoiresInvacare.LesaccessoiresconuspardautresfabricantsnontpasttestsparInvacareetilnestpasrecommanddelesutiliseraveclesproduitsInvacare.
ACCESSOIRES
Ilyaunelargegammedhumidificateurs,detubesoxygne,decanulesetdemasquespouvanttreutilissaveccetappareil.Vousdevriezcontactervotreprestatairedesoinsdomicileafinquilvousindiquequelssontlesdispositifsadaptsvotrecas.Ilsdevraientgalementvousrenseignerconcernantlamthodedutilisation,lamaintenanceetlenettoyageappropris.REMARQUE:Lesversionsmisesjourdecemanuelsontdisponiblessurwww.invacare.fr.
-
7/24/2019 Perfecto2 Manual de Uso
36/228
SOMMAIRE
Perfecto2 Srie 36 Part No 1163146
ACCESSOIRES ................................................ 35REMARQUES SPCIALES ................................37
EMPLACEMENT DU LABEL ............................. 39SECTION 1REGLES GENERALES .................... 40
Informations de fonctionnement ...................................................41
Interfrence radiolectrique...........................................................42
SECTION 2CARACTRISTIQUES ................... 43SECTION 3MANIPULATION ......................... 44
Dballage.............................................................................................44Contrle..............................................................................................44
Rangement ..........................................................................................44
SECTION 4DESCRIPTION TECHNIQUE .......... 45SECTION 5PARAMTRES TYPES DUPRODUIT ...................................................... 46SECTION 6CONSIGNES D'UTILISATION ........ 50
Introduction........................................................................................50
Choisir Un Endroit............................................................................50
Montage...............................................................................................51
Interrupteur........................................................................................53
Dbit.....................................................................................................54
Tmoin de la puret de l'oxygne SensO2
- Modle
IRC5PO2AW & IRC5PO2VAW....................................................55
Mise en marche initiale du concentrateur...................................55
Explication des tmoins lumineux - Modle IRC5PAW ..........57
Compteur horaire.............................................................................57
SECTION 7ENTRETIEN ................................ 58Entretien de routine .........................................................................58
SECTION 8GUIDE DE DPANNAGE .............. 61SECTION 9ACCESSOIRES EN OPTION ...........63INFORMATIONS DE RECYCLAGE .................... 64LIMITES DE GARANTIE .................................. 65
-
7/24/2019 Perfecto2 Manual de Uso
37/228
REMARQUES SPCIALES
Part No 1163146 37 Perfecto2 Srie
REMARQUES SPCIALESDesmotsdalertesontutilissdanscemanueletsappliquentauxpratiquesrisqueoudangereusespouvantentrainerdesblessurescorporellesoudesdgtsmatriels.Sereporterautableausuivantpourlesdfinitionsdesmotsindicateurs.
AVIS
Lesinformationscontenuesdanscedocumentsontsusceptiblesdtremodifiessanspravis.
DANGERNE FUMEZ PAS pendant l'utilisation de cet appareil.Conservez allumettes, cigarettes allumes ou autressources d'ignition en dehors de la pice dans laquelle setrouve cet appareil et loignes de l'endroit ou del'oxygne est distribu.
MOT-INDICATEUR DEFINITION
DANGER
Danger indique une situation dangereuseimminente qui, si elle n'est pas carte, entrainerala mort ou des blessures graves.
MISE EN GARDE
Mise en garde indique une situationpotentiellement dangereuse qui, si elle n'est pascarte, peut entrainer la mort ou des blessuresgraves.
ATTENTION
Attention indique une situation de risque potentielqui, s'il n'est pas cart, peut entrainer des dgtspersonnels ou des lsions mineures voire les
deux.
-
7/24/2019 Perfecto2 Manual de Uso
38/228
REMARQUES SPCIALES
Perfecto2 Srie 38 Part No 1163146
DANGERDes panneaux INTERDIT DE FUMER doivent tre
exposs bien en vue. Les textiles et autres matriaux quine brleraient pas dans des conditions normalessenflamment facilement et brlent avec une grandeintensit dans un endroit ou l'air est enrichi en oxygne.Le non-respect de cette mise en garde peut entraner unincendie et des dgts matriels graves, ou provoquer desblessures corporelles voire la mort.
ATTENTION
Attention:Laloipeutrestreindrelaventedecetappareildesortequilnepuissetrevenduqueparunmdecinousursonordonnance,oupartoutautrepraticienagrsurlelieudexercice.
Invacarerecommandeunesourcealternativedapportenoxygnedanslecasdunecoupuredecourant,dunesituationdurgenceoudunepannemcanique.Consultezvotremdecinouvotrefournisseurconcernantlesystmederservequevoussouhaitez.
Cematrieldoittreutilispourassureruncomplmentdoxygneetnonpourservirdappareilderanimationetdemaintiendesfonctionsvitales.
-
7/24/2019 Perfecto2 Manual de Uso
39/228
EMPLACEMENT DU LABEL
Part No 1163146 39 Perfecto2 Srie
EMPLACEMENT DU LABEL
L'tiquette avec le numrode srie est situe sur
L'tiquette descaractristiques est
situe sur l'arrire versle bas
erfecto
POUR LES INSTRUCTIONS DE FONCTIONNEMENT ou ALARMES (voir le manuel de l'oprateur)
DANGER - RISQUE D'INCENDIENE PAS FUMER - Ecarter TOUTE source d'ignition de la pice o se trouve cet appareil et distance des zones de distribution d'oxygne. Les textiles, huiles et autres combustibles reprsententdes sources inflammables pouvant brler avec grande intensit dans un environnement o l'air estenrichi en oxygne.DANGER - ELECTROCUTIONNE PAS enlever le couvercle. Confier toute rparation un personnel de maintenance qualifi.
Un taus de dbit infrieur 0.5 L/min dclenchera une alarme sonore de signalisation de bas dbit(alarme rapide audible).Positionnez le dbitmtre sur 2.5 L/min ou moins lorsque vous remplissez les bouteilles HomeFill.
CompatibleHomeFill
-
7/24/2019 Perfecto2 Manual de Uso
40/228
SECTION 1REGLES GENERALES
Perfecto2 Srie 40 Part No 1163146
SECTION 1REGLES GENERALESPourgarantirlinstallation,lassemblageetlefonctionnementsrsduconcentrateurPerfecto2,ilconvientderespectercesinstructionsIMPRATIVEMENT.
MISE EN GARDECechaptrecontientdesinformationsimportantesconcernantlefonctionnementetlutilisationsansdangerdeceproduit.
DANGER
Risquedlectrocution.NEPASdmonter.Confiertouterparationunpersonneldemaintenancequalifi.Aucunepicerparableparlutilisateur.
POURREDUIRELERISQUEDEBRLURES,
DELECTROCUTION,DINCENDIE
OU
DE
DOMMAGES
CORPORELS.
Nejamaisutiliserlappareilenprenantunbain.Silaprescriptionmdicalerequiertuneutilisationcontinue,leconcentrateurDOITtreplacdansuneautrepiceloignedaumoins2.5m(7ft)delasalledebains.
NE
PAS
toucher
le
compresseur
en
tant
mouill.NEPASplacerourangerlappareilproximitdeleauoudunautreliquide.
NEJAMAIStenterdereprendrelappareilsiltombedansleau.DbrancherIMMEDIATEMENT.
Silecordonoulapriseduconcentrateurestendommag,
sil
ne
fonctionne
pas
correctement,
sil
heurte
le
sol,
sil
est
endommagouplongdansleau,lappareildoittrerenvoyuntechnicienqualifipourtreinspectetrpar.
-
7/24/2019 Perfecto2 Manual de Uso
41/228
SECTION 1REGLES GENERALES
Part No 1163146 41 Perfecto2 Srie
DANGER
Uneinflammationspontaneetviolentepeutseproduiresi
de
lhuile,
de
la
graisse
ou
des
substances
grasses
entrent
en
contactavecloxygnesouspression.CessubstancesDOIVENTtretenueslcartduconcentrateurdoxygne,destubesetconnexions,etdetoutmatrieldoxygnationengnral.NEPASutiliserdelubrifiantssaufencasderecommandationsparInvacare.
viterdeprovoquerdestincellesproximitdelappareil
mdicaldoxygnation.Celainclutlestincellesdlectricitstatiqueprovoquespardesfrictions.
Informations de fonctionnementPouruneperformanceoptimale,Invacarerecommandeque
chaqueconcentrateurfonctionnependantunminimumde30minuteschaquefois.Despriodesdefonctionnementpluscourtespeuventrduireladuredeviedecetappareil.
Eloignezlestubesoxygne,lecordonetlappareildesouslescouvertures,lesdessusdelit,lescoussins,lesvtementsetcartsdessurfaceschauffesouchaudes,comprenantlesradiateurs,lesfoursetappareilslectriquessimilaires.
NEPASdplacerouchangerleconcentrateurdeplaceentirantsurlecordonlectrique.
NEJAMAISlaissertomberouinsrerunobjetouduliquidedanslesouvertures.
Invacarerecommandequuntubeoxygneanticrasementsoitutilisavecceproduitetquilnedpassepas15.2m
(50ft)delongueur.Ilnyapasdepicesrparablesparlutilisateur.Cecinincluepasleslmentsdentretiennormal.Voirlasectionentretienpourlentretiendeslmentsparlutilisateur.
-
7/24/2019 Perfecto2 Manual de Uso
42/228
SECTION 1REGLES GENERALES
Perfecto2 Srie 42 Part No 1163146
NEJAMAISlaisserlappareilsanssurveillancelorsquilestbranch.AssurezvousquelePerfecto2estdbranchlorsquilnestpasutilis.
Unesupervisionrapprocheestncessairelorsqueceproduitestutilisparouproximitdesenfantsoudesindividussouffrantdehandicapsphysiques.
Uncontrleouuneattentionparticuliredoittreapporteauxpatientsquiutilisentcetappareilouquisontincapablesdentendreoudevoirdesalarmesoudecommuniquerleurinconfort.
NEPASbrancherleconcentrateurenparallleouensrieavecdautresconcentrateursdoxygneouappareilsdoxygnothrapie.
Interfrence radiolectrique
CetquipementattestettrouvconformeauxlimitesEMCspcifiesparIEC/EN6060112.Ceslimitessontdestinesprotgerraisonnablementcontrelesinterfrenceslectromagntiquesdansuneinstallationmdicaletypique.
Dautresappareilspeuventrecevoirdesinterfrencesmmedesplusbasniveauxdesmissionslectromagntiquesautorisesparlesstandardscidessus.Pourdterminersiles
missionsduPerfecto2causentlinterfrence,arrtezlePerfecto2.Silinterfrenceaveclautreappareilsarrte,alorslePerfecto2causelinterfrence.Danscescasrares,linterfrencepeuttrerduiteoucorrigedelunedesfaonssuivantes:
Repositionnezlamachine,changezladeplace,ouloignezlespacemententrelesmachines.
Branchezlappareildansuneprisediffrentedecellesurlaquellelautreappareilestbranch.
-
7/24/2019 Perfecto2 Manual de Uso
43/228
SECTION 2CARACTRISTIQUES
Part No 1163146 43 Perfecto2 Srie
SECTION 2CARACTRISTIQUES
Sortie oxygne
Dbitmtre
Voyants lumineux de lapuret de l'oxygne /Voyants lumineux defonctionnement et de
panne
Interrupteur
Disjoncteur
Compteurhoraire
Filtres
VUE ARRIRE
* Raccord desortie HomeFillCordon
d'alimentation
*REMARQUE:Ceraccorddesortienedoittreutilisque
pourremplirlesbouteillesdoxygneaveclecompresseurHomeFill.Leraccorddesortienaffectepaslaperformancedu
concentrateur.SerfreraumanueldutilisationHomeFill,rfrence1145805,pourlesinstructionsderaccordementetdutilisation.Lobturateur
fourniavecleconcentrateurdoittreintroduitdansleraccorddesortielorsquilnest
pasutilis.PourplusdinformationsconcernantleHomeFill,contactezvotrevendeurInvacare.
-
7/24/2019 Perfecto2 Manual de Uso
44/228
SECTION 3MANIPULATION
Perfecto2 Srie 44 Part No 1163146
SECTION 3MANIPULATIONLeconcentrateurdoitTOUJOURStremaintenuenpositionverticaleafindviterdendommagerlecaissonlorsdutransport.
Sileconcentrateurdoittrerexpdipartransporteur,descartonssupplmentairessontdisponibleschezInvacare.
DballageREMARQUE:Pourcetteprocdure,sereporterFIGURE3.1.
1. Contrlertoutdommage
videntducartonoudesoncontenu.Sidesdgtssontvisibles,informezenlelivreur,ouvotrevendeurlocal.
2. Retirertouteslesgarnituresdescartons.
3. Retirersoigneusementtousleslmentsducarton.
FIGURE 3.1 DballageREMARQUE:AmoinsqueleconcentrateuroxygnedoivetreutilisIMMEDIATEMENT,laisserleconcentrateurdanssonemballagepourlerangerjusqucequelutilisationdu
concentrateursoitncessaire.Contrle1. Examiner/contrlerlextrieurduconcentrateurdoxygne
pourreprerdventuellesrayures,bosses,rafluresouautresdommages.Contrlertousleslments.
Rangement1. Rangerleconcentrateurdoxygneremballdansunendroitsec.
2. NEPASplacerdautresobjetsaudessusduconcentrateurdoxygneremisdanssonemballage.
http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
45/228
SECTION 4DESCRIPTION TECHNIQUE
Part No 1163146 45 Perfecto2 Srie
SECTION 4DESCRIPTIONTECHNIQUE
LeconcentrateurInvacarePerfecto2estutilisparlespatientsatteintsdedsordresrespiratoiresquincessitentunapportdoxygne.Lappareilnestpasprvupourmaintenirouprolongerlavie.
Letauxconcentrationdoxygnevariede87%95,5%.Lapportenoxygneaupatientsefaitlaidedunecannulenasale.
LeconcentrateurInvacarePerfecto2utiliseuntamismollculaireetunemthodedabsorptiondelapressionpourproduirelasortiedoxygne.Lairambiantentredanslappareil;estfiltrpuiscompress.Laircompressestensuitedirigverslundesdeuxtamisabsorbantslazote.Loxygneconcentrschappedutamisopposetestdirig
dansunrservoiroxygneoilesttransmisaupatient.LeconcentrateurInvacarePerfecto2estcapabledtreutilisparlepatientdansunenvironnementdhabitationouunenvironnementinstitutionnel.Lappareilfonctionneunepuissancenominalede230VAC/50Hertz.
Lesinformationsdentretienserontfourniessurdemandeau
personneltechniquequalifiuniquement.
-
7/24/2019 Perfecto2 Manual de Uso
46/228
SECTION 5PARAMTRES TYPES DU PRODUIT
Perfecto2 Srie 46 Part No 1163146
SECTION 5PARAMTRES TYPES DUPRODUIT
Courant alternatif
quipement de Type BF
Appareil en marche
Appareil l'arrt
Attention - Consulter la documentation jointe
NE PAS fumer
Classe II, double isolation
Protger contre les projections d'eau
Caractristiqueslectriques : 230 V c.a. 10 % (253 V c.a./207 V c.a.), 50 Hz
Consommation decourant nominal : 1.4 A
Niveau sonore : 39 Dba de moyenne
Altitude : Tout le Perfecto2 Modle : Jusqu' 8.000 pi (2438 mtres)au-dessus du niveau de la mer sans la dgradation de niveaux deconcentration. De 8.000 pi 13.129 pi (4000 mtres) au dessousde 90% efficacit.
Tout le Perfecto2 Modle en remplissant un Systme de HomeFill :Jusqu' 6.000 pi (1828 mtres) au-dessus du niveau de la mer sans la
dgradation de niveaux de concentration. De 6.000 pi (1828mtres) 13.129 pi (4000 mtres) au dessous de 90% efficacit.
IPX1
-
7/24/2019 Perfecto2 Manual de Uso
47/228
SECTION 5PARAMTRES TYPES DU PRODUIT
Part No 1163146 47 Perfecto2 Srie
Taux deconcentrationd'oxygne :
REMARQUE:Lestauxdeconcentrationindiqusonttobtenusaprsune
priodedechauffeinitialedenviron30minutes.
87 % 95.6 % de 0.5 5 L/min93% 3% 2 L/min
Pression de sortiemaximale : 34.5 kPa 3.45 kPa (5 psi 0.5 psi)
Plage de dbit : De 0.5 to 5 L/min (maximum)Pour les taux de dbit infrieurs 1L/min, nous recommandonsl'utilisation du dbitmtre pdiatrique Invacare (IRCPF16AW)
Alarme de baisse dedbit :
De 0 L/min to 0.5 L/minBips d'alarme rapides et audibles(Aucuns accessoires branchs)
Puissance moyenneConsommation: Perfecto
2AW - 300 W @ 5 L/min, 280 W @ 3 L/min
Perfecto2VAW - 320 W @ 5 L/min
Mcanisme dedcompressionfonctionnant : 241 kPa 24,1 kPa (35 psi 3,5 psi)
Modification du dbitmaximalrecommand
lorsquunecontre-pression de 7kPa est applique : 0,7 l/mn.
Filtres : Caisson, sortie HEPA et admission
Systme de scurit : Arrt en cas de surcharge de courant ou de surtensionArrt du compresseur en cas de temprature leveAlarme de haute pression avec arrt du compresseurAlarme de basse pression avec arrt du compresseur
Alerte de perte dalimentation auxiliaireAlarme de dbit faible du systme doxygne SensO
2(modle SensO
2)
Largeur : 38.1 cm 1 cm (15 in 3/8 in)
Hauteur : 58.4 cm 1 cm (23 in 3/8 in)
Profondeur : 30.5 cm 1 cm (12 in 3/8 in)
>
-
7/24/2019 Perfecto2 Manual de Uso
48/228
SECTION 5PARAMTRES TYPES DU PRODUIT
Perfecto2 Srie 48 Part No 1163146
Poids : Perfecto2AW - 20.5 kg 1 kg (45 lbs 2 lbs)
Perfecto2VAW - 18.1 kg 1 kg (40 lbs 2 lbs)
Poids dexpdition: Perfecto2AW - 22.7 kg 1 kg (50 lbs 2 lbs)
Perfecto2VAW - 20.4 kg 1 kg (45 lbs 2 lbs)Tempratureambiante : 10 C - 35 C (50 F - 95 F) une humidit relative de 20 60 %
Temprature del'chappement del'air derefroidissement : Infrieur la temprature ambiante +19 C
Temprature desortie de l'oxygne : Infrieure la temprature ambiante +3 C (+ 6 F)
Caisson : Coffrage en plastique anti-chocs et retardateur de flammesconforme la norme UL 94-V0
Pas de No AP/APG Non adapt pour une utilisation en prsence d'un mlangeanesthsique inflammable
Liste standard etrglementaire:
IRC5PO2AW/IRC5PAW/IRC5PO2VAWIEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2IEC/EN61000-3-2IEC/EN61000-3-3
Uniquement les modles IRC5PO2AW/IRC5PO2VAWISO8359MDD 93/42/EEC, Annexe I et IX
Modles CE : IRC5PO2AW & IRC5PO2VAW
lectricit : Pas de rallongePositionnement : loigner le concentrateur doxygne au moins 7,5 cm des murs,
rideaux ou meubles pour assurer un dbit d'air suffisantviter les moquettes paisses et les appareils de chauffage,radiateurs ou registres dair chaudUniquement sur le solNe pas utiliser l'appareil dans des espaces confins (Exemple:Placards ou Toilettes)
Lunette oxygne : Canule de 2 m (7 ft) indformable de 15 m (50 ft) max. de tubage(NE PAS pincer).
Dure defonctionnement : Jusqu 24 heures par jour
-
7/24/2019 Perfecto2 Manual de Uso
49/228
SECTION 5PARAMTRES TYPES DU PRODUIT
Part No 1163146 49 Perfecto2 Srie
TMOINS IRC5PO2AW & IRC5PO2VAW
VOYANT LUMINEUX IRC5PAW
Tempraturerecommande pourle stockage etlexpdition : -29 C 65 C, une humidit relative de 15 95 %.
Environnement : Sans fume ni suieNe pas utiliser l'appareil dans des espaces confins (Exemple:Placards ou Toilettes)
Dure defonctionnementminimum : 30 minutes
SYMBOLEFIGURANT
SURLTIQUETTE
PURET O2 VOYANTS LUMINEUX(DEL)
SYSTEM OKAY(SYSTME OKAY)O
2au dessus de 85 % ( 2 %)
Voyant lumineux VERT
O2entre 73 % ( 3 %) et
85 % ( 2 %)
Tmoin lumineux JAUNEA. JAUNE fixeB. Voyant clignottant JAUNEPanneAppelez un technicien qualifi.
SYSTME DEFAILLANTO
2au dessous de 73 % (3 %)
Voyant lumineux ROUGEAlarme sonore continueSieve-GARD arrt du compresseur
SYMBOLEFIGURANT
SURLTIQUETTE
TAT VOYANTS LUMINEUX(DEL)
SYSTEM OKAY(SYSTME OKAY)
Voyant lumineux VERT
SYSTME DEFAILLANT Voyant lumineux ROUGEAlarme sonore continueSieve-GARDArrt du compresseurAppelez un technicien qualifi
O2
I/O
-
7/24/2019 Perfecto2 Manual de Uso
50/228
SECTION 6CONSIGNES D'UTILISATION
Perfecto2 Srie 50 Part No 1163146
SECTION 6CONSIGNESD'UTILISATION
IntroductionVotreconcentrateuroxygneestprvupourunusageindividuel.Cestunappareillectroniquespareloxygnedelairambiant.Ildlivreuneforteconcentrationdoxygneparunecanulenasale.Destudescliniquesontmontrquelesconcentrateursdoxygnesontquivalentsenmatirede
thrapiedautrestypesdesystmededlivrancedoxygne.Leprestatairedmontreracommentutiliserleconcentrateurdoxygne.Ildoittreconsultpourtoutesquestionsouproblmespossparleconcentrateurdoxygne.Lemanueldelutilisateurvousinstruirasurvotreconcentrateuretvousserviraderfrencepourlutilisationdevotreconcentrateur.
Choisir Un Endroit
MISE EN GARDE
NEJAMAISbloquerlesouverturesdairdelappareilnileplacersurunesurfacemolle,unlitouuncanapparexemple,
pournepasbloquerlessortiesdair.loignerlespeluches,cheveuxettexturessimilairesdessortiesdair.
Tenirlappareilloigndesmurs,rideaux,mobilieretautresdaumoins7,5cm.
Choisirunepicedanslamaisonolutilisationdu
concentrateurdoxygneseralapluspratique.Leconcentrateurdoitpouvoirtrefacilementdplacdepiceenpicesursesroulettes.
-
7/24/2019 Perfecto2 Manual de Uso
51/228
SECTION 6CONSIGNES D'UTILISATION
Part No 1163146 51 Perfecto2 Srie
VotreconcentrateuroxygneseraplusperformantsivouslefaitesfonctionnerdanslesconditionssuivantesParamtrestypesduproduitenpage46.Uneutilisationdansunenvironnement
autrequeceluidcritpourraitncessiterunemaintenancesupplmentaire.Ladmissiondairdelappareildoittresituedansunendroitbienventilpourviterlespolluantsatmosphriqueset/oulesfumes.NEPASlutiliserdansuncagibi.
Montage
1. Brancherlecordondansunepriselectrique.
2. Raccorderlhumidificateur(Sicelaatrecommand).
MISE EN GARDE
NEPAStropremplirlhumidificateur.
NEPASinverserlesconnexionsentreetsortiedoxygne.Leauprovenantdelhumidificateurrepasseraparlacanulejusqulutilisateur.
REMARQUE:Pourcetteprocdure,sereporterFIGURE6.1,FIGURE6.2enpage52,FIGURE6.3enpage53.
1. Enlevezlebouchondelabouteille.2. Remplirlhumidificateurdeaudistillejusquauniveau
indiquparlefabricant.Remettrelebouchondelhumidificateuretbienlefermer.
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
52/228
SECTION 6CONSIGNES D'UTILISATION
Perfecto2 Srie 52 Part No 1163146
FIGURE 6.1 Remplissage de lhumidificateur3. Insreruntournevistteplatedanslarainureplaquesur
lebordsuprieurdelatrappedaccsetsouleverdlicatementlaportedaccsdufiltrepourouvrirlatrappe(FIGURE6.2).
4. Retirerleraccorddelhumidificateur(FIGURE6.2).5. Replacezlatrappedaccsaufiltreetremettezloue
darationsurlatrappedaccsdufiltreeninsrantlessixtaquetssetrouvantsurlouedarationdanslessixencoches
situessurlesctsdelatrappedaccsaufiltre.
FIGURE 6.2 Adaptateur de lhumidificateur
Humidificateursans bouchonHumidificateur
avec bouchon
Insrez letournevis tte plate ici Trappe daccs
au filtre
Raccord delhumidificateur
Filtres
Oued'aration
Taquets
Encoches
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
53/228
SECTION 6CONSIGNES D'UTILISATION
Part No 1163146 53 Perfecto2 Srie
6. Fixezleraccorddelhumidificateursurlhumidificateurentournantlavisoreillesdelhumidificateurdanslesensdesaiguillesdunemontrejusqucequilsoitbienfix.VoirDtailA dansFIGURE6.3.
7. Placerlhumidificateursurlecompartimenthumidificateurduconcentrateur.VoirDtailB dansFIGURE6.3.
8. Fixerletubeoxygnelhumidificateurousurlemboutdesortiedeloxygnesurleconcentrateur.VoirDtailBdansFIGURE6.3.
9. Fixerlacanule/letubedalimentationdupatientlasortie
delhumidificateur.VoirDtailB dansFIGURE6.3.10. Aprsmontage,sassurerqueloxygnepassebienparlacanule.
FIGURE 6.3 Compartiment de lhumidificateur
InterrupteurREMARQUE:Pourcetteproc dure,sereporterFIGURE6.4.
Humidifi-cateur
Compartiment delhumidificateur
Raccord delhumidifi-
cateur
Humidifi-cateur
DETAIL "A" DETAIL "B"
Tube oxygne
Connecteur desortie
doxygne
Ouverture
de l'humidifi-cateur
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
54/228
SECTION 6CONSIGNES D'UTILISATION
Perfecto2 Srie 54 Part No 1163146
1. MettrelinterrupteursurlapositionMarche.Touslesvoyantslumineuxet
lalarmesonoresallumentpendantunesecondepourindiquerquelappareilfonctionnenormalement.
FIGURE 6.4 Interrupteur
DbitREMARQUE:Pourcetteprocdure,sereporterFIGURE6.5.REMARQUE:VeillernepasrglerledbitaudessusdelanneauROUGE.Untauxdedbitdoxygnesuprieur5L/mindiminueralaconcentrationdoxygne.1. Tournerleboutondedbitsurlerglagerecommandpar
lemdecin.
MISE EN GARDENEPASmodifierlerglageL/mndudbitmtretantquunchangementnapastdemandparlemdecin.
REMARQUE:Pourbienlireledbitmtre,reprerletraitdedbitrecommande.Tournerensuiteleboutondedbitdefaonvoirlabillemonterjusquautrait.Maintenant,centrerlabillesurletraitL/mnrecommand.
Interrupteurmarche/arrt
(I/O)
http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
55/228
SECTION 6CONSIGNES D'UTILISATION
Part No 1163146 55 Perfecto2 Srie
2. Siledbitindiquparledbitmtrechuteaudessousde0,5L/mnpendantplusduneminuteenviron,lalarmedeDBITMINIMUMsedclenche.Lalarmesonoremetdessignauxrapides.Contrlerletubeoulesaccessoirespourreprertoutblocageoupincementdutubeouundfautdelhumidificateur.Unefoisledbitnormalrtabli,savoirplusde0,5L/mn,lalarmesonoredeDEBITMINIMUMsarrte.
REMARQUE:LutilisationdaccessoirestelsqueledbitmtrepdiatriqueetlecompresseurHomeFilldsactivelalarmedebaissededbit.
FIGURE 6.5 Dbit
Tmoin de la puret de l'oxygne SensO2-
Modle IRC5PO2AW & IRC5PO2VAWCedispositifcontrlelapuretdeloxygnegnrparleconcentrateurdoxygne.Silapurettombeaudessousdesnormesprrgleslusine,lesvoyantsdutableaudecontrlesallument.
Mise en marche initiale du concentrateurREMARQUE:Leconcentrateurpeuttreutilispendantladuredechauffedudmarrageinitial(environ30mn)enattendantquelevoyantdepuretO2vertsallume.
Bouton dedbit
Bille
Dbitmtre
Panneau avant
0.5 L/min
-
7/24/2019 Perfecto2 Manual de Uso
56/228
SECTION 6CONSIGNES D'UTILISATION
Perfecto2 Srie 56 Part No 1163146
Lorsquelappareilestmisenmarche,levoyantVERTsallume(SYSTMECORRECT/O2suprieur85%).Aprs5minutes,lecapteurdoxygnefonctionnenormalementetcommandeles
voyantslumineuxenfonctiondelaconcentrationdoxygnedfinie.
Explication des tmoins lumineux de la puret del'ocygne - Modle IRC5PO2AW & IRC5PO2VAWREMARQUE:Pourcetteprocdure,sereporterFIGURE6.6en
page56.
VoyantVERT(O2) Fonctionnementnormal.
TmoinlumineuxJAUNE( ) Appelezlefournisseurimmdiatement.Lepatientpeutcontinuerutiliserleconcentrateursaufinstructionscontrairesdelapartduprestataire.Sassurerquelarservedoxygneestproximit.
VoyantROUGE( ) Arrttotaldelappareil.Passezimmdiatementsurunapportdoxygnedesecoursetappelezlefournisseur.TmoinlumineuxVERT avecunclignottantJAUNE appelezlefournisseurimmdiatement.Mauvaisfonctionnementducapteurdoxygne;lepatientpeutcontinuerdutiliserleconcentrateur.
FIGURE 6.6 Explication des tmoins lumineux de la puret del'ocygne - Modle IRC5PO2AW & IRC5PO2VAW
O2
Arrt automatiqueUtilisation de la
rserve
Infrieure lanormale
Normale
ROUGE
Appeler leprestataire
JAUNE
VERTE
PURET DE LOXYGNE
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
57/228
SECTION 6CONSIGNES D'UTILISATION
Part No 1163146 57 Perfecto2 Srie
Explication des tmoins lumineux - ModleIRC5PAWREMARQUE:Pourcetteprocdure,sereporterFIGURE6.7.
Modle IRC5PAW - Explication des tmoins lumineux
VoyantROUGE( ) Arrttotaldelappareil.Passezimmdiatementsurunapportdoxygnedesecoursetappelezlefournisseur.
VoyantVERT(I/O) Marche/arrt.Systmecorrect.
FIGURE 6.7 Explication des tmoins lumineux - ModleIRC5PAW
Compteur horaire
Lecompteurhoraireaffichelecumuldesheuresdefonctionnementdelamachine.
Arrt automatiqueUtilisation de la
rserve
Normale
ROUGE
VERTE
Appeler leprestataire
http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
58/228
SECTION 7ENTRETIEN
Perfecto2 Srie 58 Part No 1163146
SECTION 7ENTRETIEN
MISE EN GARDE
LesconcentrateursInvacaresontspcialementconuspourminimiserlesoprationsdemaintenanceprventivederoutine.Seuleslespersonnesqualifiessontautoriseseffectuerlentretienderoutineduconcentrateur.
Dbrancherleconcentrateurlorsdunettoyage.Pourviterlesrisquesdlectrocution,NEPASretirerlecapotsuprieur.
REMARQUE:LamaintenanceprventiveDOITaumoinstreeffectueconformmentauxinstructionsducarnetdemaintenance.Unemaintenanceplusfrquentepeutsavrerncessairesurlessitestrspoussireuxouimpliquantdesniveauxdesuieslevs.RefertoRapportdemaintenanceprventiveonpage 60.
Entretien de routineLentretienderoutinecidessousdoittreeffectuplus
frquemment.RegardezlesintervallesrecommandsdanschaquesectionetRapportdemaintenanceprventiveenpage60.
Nettoyage des filtres
ATTENTION
NEPASfairefonctionnerleconcentrateursilesfiltresnesontpasinstalls.
REMARQUE:Pourcetteprocdure,sereporterFIGURE7.1en
page59.REMARQUE:Ilyaunfiltresitularrireducaisson.
1. Vrifiezlefiltrechaquemois.Enlevezlefiltreetnettoyezlesincessaire.
http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?- -
7/24/2019 Perfecto2 Manual de Uso
59/228
SECTION 7ENTRETIEN
Part No 1163146 59 Perfecto2 Srie
REMARQUE:Lesconditionsenvironnementalesncessitantunevrificationplusfrquenteetlenettoyagedufiltreincluent,maisneselimitentpas:poussiredense,polluantsatmosphriques,etc.
2. Nettoyerlefiltrelaspirateurouleauchaudesavonneuseetlesrincersoigneusement.
3. Schezlefiltresoigneusementavantdeleremettre.
FIGURE 7.1 Nettoyage des filtres
Nettoyer le caisson suprieur
1. Nettoyerlecaissonsuprieuravecunproduitnettoyantdouxetunchiffonouunepongedouce.
Nettoyer l'humidificateurREMARQUE:Pournettoyerlhumidificateurdoxygne,suivrelesinstructionsfourniesparlefabricant.Sinon,suivreces
TAPES:1. Nettoyerlhumidificateurtouslesjours.
2. Lelaverleausavonneuseetlerinceravecunesolutiondedix(10)volumesdeaupourun(1)devinaigre.
3. Lerincersoigneusementleauchaudeetleremplirdeaudistillejusquauniveauindiqusurlhumidificateur.
-
7/24/2019 Perfecto2 Manual de Uso
60/228
SECTION 7ENTRETIEN
Perfecto2 Srie 60 Part No 1163146
FIGURE 7.2 Rapport de maintenance prventive
-
7/24/2019 Perfecto2 Manual de Uso
61/228
SECTION 8GUIDE DE DPANNAGE
Part No 1163146 61 Perfecto2 Srie
SECTION 8GUIDE DE DPANNAGESYMPTME : CAUSE PROBABLE : SOLUTION :
Alarme :
Bips sonores,courts, longue
Le concentrateurne fonctionne pas,Linterrupteur estsur marche
Bip.Bip.
Perte dalimentation
principale :
1. Le cordon nest pasbranch.
1. Insrer la fiche dans laprise.
2. Pas de courant lasortie.
2. Contrler le disjoncteur oules fusibles. Si le problme seproduit de nouveau, utiliserune autre prise.
3. Disjoncteur dclench. 3. Pousser/rinitialiser lecoupe circuit. Si le problmepersiste, appelez votreservice client.
Alarme :
Continue
Le concentrateurne fonctionne pas.Linterrupteur estsur marche
Biiiiiiiip
Dfaillance dusystme :
1. Surchauffe de lappareildu fait du blocage deladmission dair.
1a. Retirer le filtre caisson etle nettoyer.
1b. loigner le concentrateurdoxygne au moins 7,5 cmdes murs, rideaux oumeubles.
2. Puissance insuffisante
la sortie.
2. NE PAS utiliser de rallonge.
Passer une autre circuit ouprise lectrique.
3. Rparations internesrequises.
3. Appeler le serviceaprs-vente.
-
7/24/2019 Perfecto2 Manual de Uso
62/228
SECTION 8GUIDE DE DPANNAGE
Perfecto2 Srie 62 Part No 1163146
Voyant JAUNE ouROUGE allum.
Alarme :ContinueVoyant ROUGEuniquement.
* S'appliqueuniquement surles modlesIRC5PO2AW &IRC5PO2VAW
1. Puret doxygnefaible.*
1. Nettoyer ou remplacer lefiltre.
2. Tube, canule ouhumidificateur coud oubloqu.*
2. Contrler labsence decoudes ou blocages. Rectifier,nettoyer ou remplacer larticle.Une fois la correctionexcute, mettre hors tensionpendant 60 secondes, puis denouveau sous tension.
3. Dbitmtre
0.5 L/min *.
3. Vrifier que le dbitmtre
est rgl sur 1.0 L/mn ou plus.Se reporter Paramtrestypes du produit en page46.
4. Surchauffe de lappareildu fait du blocage deladmission dair.
4a. Retirer le filtre caisson etle nettoyer.4b. loigner le concentrateur au moins 7,5 cm des murs,rideaux ou meubles.
5. Puissance insuffisante la sortie.
5. NE PAS utiliser de rallonge.Passer une autre circuit ouprise lectrique.
6. Rparations internesrequises.
6. Appeler le serviceaprs-vente.
Voyant VERTwavec
clignotementvoyant JAUNE
1. Rparations internesrequises.
1. Appeler le serviceaprs-vente.
Alarme :
Rapide
Bip BipBip Bip
ALARME DE BAISSEDE DEBIT
1. Tube, canule ouhumidificateur coud oubloqu.
1. Contrler labsence decoudes ou blocages.Rectifier, nettoyer ou remplacer
larticle. Une fois la correctionexcute, mettre hors tensionpendant 60 secondes, puis denouveau sous tension.
2. Dbitmtre 0.5 L/min 2. Les dbits infrieurs 1 l/mnne sont pas recommands
SYMPTME : CAUSE PROBABLE : SOLUTION :
-
7/24/2019 Perfecto2 Manual de Uso
63/228
SECTION 9ACCESSOIRES EN OPTION
Part No 1163146 63 Perfecto2 Srie
SECTION 9ACCESSOIRES ENOPTION
Lesaccessoiresenoptioncidessoussontgalementdisponibles:
Cannulenasaleadultestandard,2.1m(7feet) M3120
Humidificateurstandard M5120
DbitmtrepdiatriquePreciseRX/DbitmtreIRCPF16AW
CompresseuroxygneparticulierHomeFillIOH200AW
-
7/24/2019 Perfecto2 Manual de Uso
64/228
INFORMATIONS DE RECYCLAGE
Perfecto2 Srie 64 Part No 1163146
INFORMATIONS DE RECYCLAGECeproduitatlivrparunfabricantsoucieuxdelenvironnementetconformeladirective2002/96/CEsurlamiseaurebutdesquipementslectroniquesetlectriquesWEEE.Ceproduitestsusceptibledecontenirdessubstancespotentiellementnocivespourlenvironmentsilestmisaurebutdansdesendroits(dcharges)nonconformeslalgislationenvigueur.
Lesymbolepoubellebarreestappossurceproduitpourvousencouragerlerecyclerdanslesstructuresdecollecteappropries.
Veuillezvouscomporterdemanireresponsablevisvisdelenvironnementetrecyclerceproduitarrivenfindevieparlebiaisduncentrederecyclage.
-
7/24/2019 Perfecto2 Manual de Uso
65/228
Part No 1163146 65 Perfecto2 Srie
LIMITES DE GARANTIEBelgium&Luxemburg:Invacarenv,Autobaan22,B8210LoppemTel:(32)(0)50831010
Fax:(32)(0)[email protected]:InvacareA/S,Sdr.Ringvej37,DK2605BrndbyTel:(45)(0)36900000Fax:(45)(0)[email protected]:InvacareAquatecGmbH,Alemannenstrae10,D88316Isny
Tel:(49)(0)75627000Fax:(49)(0)[email protected]:Invacare,Kleiststrae49,D32457PortaWestfalicaTel:(49)(0)5731754540Fax:(49)(0)[email protected]:InvacareSA,c/Arenys/n,Polgon
IndustrialdeCelr,E17460Celr(Girona)Tel:(34)(0)972493200Fax:(34)(0)[email protected]:InvacarePoirierSAS,Route deStRoch,F37230 FondettesTel:(33)(0)247626466Fax:(33)(0)[email protected]:
InvacareAG,Benkenstrasse260,CH4108 WitterswilTel:(41)(0)614877080Fax:(41)(0)[email protected]:InvacareIrelandLtd,Unit5SeatownBusinessCampus,SeatownRoad,Swords,CountyDublinIrelandTel:(353)18107084
Fax:(353)[email protected]:InvacareMeccSans.r.l.,Via deiPini62,I36016Thiene(VI)Tel:(39)0445380059Fax:(39)[email protected]
Nederland:InvacareBV,Celsiusstraat46,NL6716BZEde
Tel:(31)(0)318695757Fax:(31)(0)[email protected]@invacare.comNorge:InvacareAS,Grensesvingen 9,Postboks6230,Etterstad,N0603OsloTel:(47)(0)22579500Fax:(47)(0)[email protected]@invacare.com
sterreich:InvacareAustriaGmbH,HerzogOdilostrasse 101,A5310 MondseeTel:(43)623255350Fax:(43)[email protected]&Suomi:InvacareAB,Fagerstagatan 9,S163 91 SpngaTel:(46)(0)87617090Fax:(46)(0)87618108
[email protected]@invacare.comUnitedKingdom:InvacareLimited,PencoedTechnologyPark,Pencoed,BridgendCF355HZTel:(44)(0)1656776200Fax:(44)(0)1656776201CustomerServices:Tel:(44)(0)1656776222Fax:(44)(0)[email protected]:InvacareLda,RuaEstradaVelha,949,P4465784LeadoBalioTel:(351)(0)225105946/47Fax:(351)(0)[email protected]
mailto:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:[email protected]:conta