personal surveyor 200 (ps200) series

47
Personal Surveyor 200 (PS200) Series User Training

Upload: nasya

Post on 01-Feb-2016

66 views

Category:

Documents


0 download

DESCRIPTION

Personal Surveyor 200 (PS200) Series. User Training. PS200 User Training - Objectives. Convey a clear understanding of the operational features of the instrument Teach users how to get the most out of their instrument Train trainees to become trainers. Overview. Single button user operation - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Personal Surveyor 200 (PS200) Series

Personal Surveyor 200 (PS200) Series

User Training

Page 2: Personal Surveyor 200 (PS200) Series

PS200 User Training - Objectives

•Convey a clear understanding of the operational features of the instrument•Teach users how to get the most out of their instrument•Train trainees to become trainers

Page 3: Personal Surveyor 200 (PS200) Series

Overview

• Single button user operation• Robust rubber overmould• LCD with bi-colour backlight• Up to 4-gases• Audible/visual/vibrating alarms• Integrated Bump-Test feature• Datalogging included• Optional integral pump

Page 4: Personal Surveyor 200 (PS200) Series

-20°C Ta 50 ° C• ATEX: Sira 11ATEX2041X

II2GEx ia d IIC T4 Gb

• IECEx: SIR 11.0019X

• UL*: 913 (7th Edition)• MED*: => A.1 / 3.30• CSA*: C22.2 No 152

Certification

*pending

Page 5: Personal Surveyor 200 (PS200) Series

Ranges of operation

• 0-100 %LEL hydrocarbons(CH4, C3H8, C4H10, C5H12)

• 0-25 % Oxygen (O2)• 0-100 PPM Hydrogen Sulphide (H2S)• 0-1000 PPM Carbon monoxide (CO)

1-gas 2-gas 3-gas 4-gas

Page 6: Personal Surveyor 200 (PS200) Series

Operating procedures

• Check the following before use:• instrument is clean and in good condition• hydrophobic and inlet filter are clean and in good

condition• sample line and probe, if used, are in good condition

• Switch ON in fresh air• battery is fully charged• instrument is within the calibration period• all gas ranges are operational and the instrument is

zeroed• Perform leak check by blocking inlet of sample intake

Page 7: Personal Surveyor 200 (PS200) Series

Switching ON

• Press and Hold Right Hand Button for 1 second• Instrument begins warm-up routine with

the GMI and PS200 logo.• During 30 second warm-up a

countdown timer appears on the screen• The instrument display identifies the

serial number, software version,datalogging, battery status andflammable cal gas during warm-up.

Page 8: Personal Surveyor 200 (PS200) Series

Warm-up information

• The time and datefrom the instrument’sbuilt-in clock is displayed

• The calibration due dateis shown on the display

• If the calibration due date has expired, the audible and visual alarm activates

• Press RH Button once to acknowledge overdue calibration due date, cancel alarms and continue with warm-up

Page 9: Personal Surveyor 200 (PS200) Series

Warm-up information (2)

• If the calibration is overdue, and the instrument is configured to be switched off in that situation, the display below appears (audible / visual alarms are also activated). It must be switched off.

Page 10: Personal Surveyor 200 (PS200) Series

Sensor check

• A diamond symbol appears next to each sensor to confirm that the sensor

– is working correctly– is being zeroed

• When sensors are zeroed correctly a tick appears next to each sensor

Page 11: Personal Surveyor 200 (PS200) Series

Zero fault

• If sensor(s) fail the zero checkat the end of warm-up,the alarms activate.The display will show a spanner.

• To acknowledge the alarm, press the RH button once.The instrument continues to show a flashing spanner symbol alternating with the faulty sensor zero reading.

Page 12: Personal Surveyor 200 (PS200) Series

Normal operating display

• When warm-up is completed successfully, back-light switches off and normal operating screen is shown as follows

• Each gas range is displayed• Display back-light can be manually switched on with

a single press of the RH button. It remains on for 20 seconds and then switches off

Page 13: Personal Surveyor 200 (PS200) Series

Normal operating display (2)

• During normal operation where the calibration date has expired, a warning flag will flash on the display every 30 seconds to warn the user that the expiry date has been exceeded.

Page 14: Personal Surveyor 200 (PS200) Series

Viewing Max & Min values

• Start from normal operating screen.Press the RH button to switch screenlight on. Press RH button againwhile light is on to view maximumvalues for each gas.

• Press RH button again to view theminimum gas values for each range.

• Stored readings can be removed from memory by pressing holding the RH button for 2 seconds

Page 15: Personal Surveyor 200 (PS200) Series

Remote sampling / pre-entry check

• Remote sampling is carried out with the in-built electric pump or by the hand aspirator– Press and hold the RH button and the pump (if fitted) will run at normal

speed for remote sampling.– When the pump is running at normal speed, a pump symbol rotates

in the display.

Page 16: Personal Surveyor 200 (PS200) Series

• Remote sampling– Tygon tubing with ID of 4mm is used in combination

with the bayonet connector– Push quick connector

on to the inlet nozzle– Turn clockwise until

locked in place

Attach tubing

Page 17: Personal Surveyor 200 (PS200) Series

Alarms

• All gas ranges have alarm limits that trigger an alarm when the measured value exceeds the set level

• The audible alarm sounds, LEDs flash red, the instrument vibrates and the gas range in alarm flashes on the display– Flammable LEL Alarm Limits - Up to 2 alarm levels are programmable– Over-Range Flammable Gas Alarm - If the flammable gas readings

exceed 120% LEL, a safety alarm is activated– Oxygen Alarm Limits - Up to 1 upper and 2 lower alarm limits are

programmable– Toxic Alarm Limits - Up to 2 instantaneous and 2 Time Weighted

Average (TWA) alarms are programmable for each toxic range

Page 18: Personal Surveyor 200 (PS200) Series

Alarms (2)

• The instrument display will signal the range in alarm. – Example of Lo Lo Oxygen Alarm Display

– Example of HiHi LEL Alarm Display

Page 19: Personal Surveyor 200 (PS200) Series

Alarms (3)

• Alarms can be latching or non-latching– Latching - alarm will stay on until the user resets by a press and hold

of the RH button when the gas level return to within the pre-set alarm limits

– Non-Latching - the audible and visual alarm will reset automatically when the reading returns to within the pre-set alarm limits

Page 20: Personal Surveyor 200 (PS200) Series

Alarms (4)

• Acknowledging Gas Alarms– Once in a safe gas free area, or the gas reading has returned to within

pre-set limits, press and hold the RH button to acknowledge/clear the alarm sounder and extinguish the red LED’s.

– Alarms can have the mute function enabled or disabled

Page 21: Personal Surveyor 200 (PS200) Series

Alarms (5)

• Muting Gas Alarms– Alarms can have the mute function enabled or disabled– This means the audible alarm can be silenced for 60s.– If the alarm is non-latching, after 60s it will operate according to the

gas concentration at that moment. I.e. if safe, it will remain silent, if above alarm level it will start sounding again.

– If the alarm is latching, after 60s the audible alarm sounds again, regardless of the concentration. The user can then mute it again (if the concentration is above the alarm level, or acknowledge it if it is below).

Page 22: Personal Surveyor 200 (PS200) Series

Alarms (6)

• High Flammable Gas Over-Range Alarms– Protects the user from danger in event of the LEL sensor being over

exposed to a high concentration of flammable gas– When over 120% LEL has been detected,

the displayed value will change tofour rising arrows, tone of audible alarm will changeand a flashing message will be shown

– The instrument must be returned to a gas free areaand switched off by a press of both buttons together. A timer from 10 to zero will appear with the messages below alternating

Page 23: Personal Surveyor 200 (PS200) Series

Alarms (7)

• Default Alarm Settings (Table 1)

Page 24: Personal Surveyor 200 (PS200) Series

Alarms (8)

• Man Down Alarm (Motion sensor) - October onwards– Protects the user from danger in event of becoming unconscious or

unable to move.– When the instrument has not detected a movement for a pre-set

period of time, the displayed back-light will flash red, the audible alarm will sound every two seconds and the red LED’s will flash

– The Man Down Alarm is latching and can only be acknowledged after the instrument has moved.

Page 25: Personal Surveyor 200 (PS200) Series

Confidence signal

• During normal operation, the instrument sounds a confidence beep and illuminates the green LEDs briefly every 15 seconds.

• This function makes the useraware that the instrument isworking correctly and alarmsare functioning.

Page 26: Personal Surveyor 200 (PS200) Series

• Self-Testing the PS200– Press & hold the LH button– The following screens will be shown successively

– The alarms will be activated for a brief period– Followed by the flammable cal gas being shown

Self-Test

Page 27: Personal Surveyor 200 (PS200) Series

Fault warnings

• Low Battery– ‘LOW BATTERY’ flashes approximately 30 minutes

operating time remains. The display flashes red,the audible alarm sounds once every 2 seconds andthe red LED’s flash.

– ‘BAT FAULT’ flashes when approximately 3 minutes operating time remains. The display will be red, the audible alarm sounds continuously and the red LED’s remain on. After approximately 3 minutes the instrument automatically switches off.

Page 28: Personal Surveyor 200 (PS200) Series

Battery life

• Battery life varies depending on instrument configuration– Electrochemical sensors only– LEL sensor included– Pump included

Page 29: Personal Surveyor 200 (PS200) Series

Fault warnings (2)

• ZERO FAULT– Appears after warm-up if the instrument is switched on in gas or the

instrument has been unable to zero all sensors. The display will be red, the audible alarm sounds once every 2 seconds, and the red LED’s flash (start-up only).

– The instrument can still be used to detect and alarm on the other sensor(s) fitted.

– Press the RH button, as per the screen prompt, to continue. A flashing spanner symbol will be displayed against the faulty sensor.

– To clear, the instrument should be taken into a gas free area, switched off and then back on.

Page 30: Personal Surveyor 200 (PS200) Series

Fault warnings (3)

• There are 3 types of sensor fault– If a flashing symbol is alternating with a gas value,

leave instrument on for 30 minutes.Switch instrument Off and On again in fresh air.If the fault remains, return instrument to aGMI approved Service/Repair facility.

– If a symbol is displayed constantly,then the sensor may require replacement.Return instrument to aGMI approved Service/Repair facility.

Page 31: Personal Surveyor 200 (PS200) Series

Fault warnings (4)

• FLOW FAULT (pumped instruments only)– The pump symbol on the display will change and FLOW FAULT

flashes to indicate a sample fault– The display will be red, the audible alarm sounds continuously and the

red LED’s remain on.

– Check sample line / filters / and probe for blockage.– Clear the blockage and restart the pump by a press and hold of the

RH Button

Page 32: Personal Surveyor 200 (PS200) Series

Switching off

• Press and hold the LH and RH button together to switch the instrument off.

• The instrument will start a countdown from three (3) to OFF. Both buttons must be pressedtogether until the displaygoes blank.

• While both buttons arepressed, the audible alarm sounds to alert the user that the instrument is switching OFF.

Page 33: Personal Surveyor 200 (PS200) Series

• Recharge (Li-ion) Battery– Connect standard charger lead to battery pack– Switch on power supply to the charger– Leave up to 4 hours to recharge the discharged

battery pack from depleted

Charging

Page 34: Personal Surveyor 200 (PS200) Series

PS200

MANUAL BUMP-TEST MODE

(using pre-set limits)

Page 35: Personal Surveyor 200 (PS200) Series

MANUAL BUMP-TEST MODE

Objectives

• Convey a clear understanding of the bump-test features of the instrument

• Teach users how to test their instrument and know it is safe

• Train trainees to become trainers

Page 36: Personal Surveyor 200 (PS200) Series

Bump-Test mode:Switching On

• Hold the LH button to switch the instrument on in Bump-Test mode

• The display will flash Bump Testalong the top of the screen

• Wait for the warm-up to complete• Accept any messages

during warm-up

Page 37: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Apply gas

• After completion of warm-up, the screen will prompt the user to start applying gas.

Page 38: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Gas application (LEL)

• If the LEL being applied peaks within the pre-set limits, within the given time period, a check mark will appear in front of the range, intermittent with the actual value.

• If the pre-set higher limit is overshot, the bump-test will fail.

Page 39: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Gas application (O2)

• If the O2 test-gas being applied causes the oxygen reading to dip within the pre-set limits, within the given time period, a check mark will appear in front of the range, intermittent with the actual value.

• If the pre-set lower limit is not reached, the bump-test will fail.

• The back-light turns red and a cross is shown in front of the range.

Page 40: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Gas application (toxic)

• It is possible that this was done in combination with the LEL, in form of a combi-gas (or mixed gas).

• If the toxic gases being applied peak within the pre-set limits, within the given time period, a check mark will appear in front of the range, intermittent with the actual value.

• If the pre-set higher limit is overshot,the bump-test will fail.

Page 41: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Timed out

• If the gas application is not completed within a pre-set time, the instrument will time out.

• The screen shows all ranges that were not bumped with a spanner and the back-light turns red.

Page 42: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Confirm Alarms

• After the gas application is completed, gas is disconnected and switched off, the user is asked to confirm that both audible and visual alarms were activated and detectable.

• Press the LH button for NO• Press the RH button for YES

Page 43: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Confirm Alarms

• After the alarms are confirmed (or not), the overall result for the Bump-Test will be shown.

• After, the instrument proceeds tonormal operation:

Page 44: Personal Surveyor 200 (PS200) Series

PS200

MANUAL BUMP-TEST MODE

(quick trigger alarm)

Page 45: Personal Surveyor 200 (PS200) Series

Bump-Test mode: Quick

• Operates the same as the full bump-test, but does not have upper limits and uses alarm level as lower limit.

Page 46: Personal Surveyor 200 (PS200) Series

PS200

DATALOGGING

Page 47: Personal Surveyor 200 (PS200) Series

Datalogging

• The instrument comes by default with datalogging.• Automatic logging is set for every minute; this can be

changed.• Manual logging is available by pressing the LH

button.• Briefly, the word ‘LOGGING’ is shown.

• A manual log is clearly identified in the datalogging software.