peter regier seminar - ic
TRANSCRIPT
-
7/30/2019 Peter Regier Seminar - IC
1/37
Regier 1
Development of an Environmental Sample Analysis
Method Using Ion-Exchange Chromatography
Peter Regier
Bethel College
Advisor: Dr. Gary Histand
-
7/30/2019 Peter Regier Seminar - IC
2/37
Regier 2
Table of Contents
I. AbstractII. Introduction
III. Ion-Exchange Chromatography TheoryIV. Methods
a. Development of Operation Procedureb. Instrumentationc. Reagentsd. Calibratione. Testing an Unknownf. Sample Preparation
V. ResultsVI. Discussion
VII. ConclusionVIII. Acknowledgements
IX. ReferencesX. Appendices A-K
-
7/30/2019 Peter Regier Seminar - IC
3/37
Regier 3
I. Abstract
Ion Chromatography is a rapidly expanding analytical technique, used for simultaneous
determination of ions. Bethel College recently purchased a Metrohm Ion Chromatograph (IC) and the
successful development of proper operating conditions and techniques was met through completion of
this seminar. All aspects for creating a precise, accurate and repeatable analysis environment weretaken into account. As IC analysis detects ionic compounds on the sub-ppm (parts per million) level,
great care was taken in all steps, from cleaning of glassware to general maintenance and creation of
necessary solutions. The method was verified through use of standard anion solution*. Soil samples
were taken from a tallgrass prairie restoration plot at Bethel College to provide environmental samples.
As nitrogen is one of the most important nutrients for a healthy, functioning prairie[1] and is often the
limiting nutrient in terrestrial environments[2], nitrate and nitrite were analyzed. Substantial research
into ionic species in aqueous environments already exists[3][4][5]
but less for environmental investigation
of ionic concentrations in soil, especially cationic compounds[6]
. As there is concern of the health of the
restoration plot, nitrate and nitrite levels are key to understanding what is wrong and possible remedies
to restore the health of the plot.
*See Sec. IVc: Reagents for solution details
-
7/30/2019 Peter Regier Seminar - IC
4/37
Regier 4
II. Introduction
Chromatography is a common and diverse method of separation where a mobile phase flows
over a stationary phase. Ion Chromatography is a branch of Liquid Chromatography that has become
increasingly popular in recent years. Ion Chromatography separates ionic compounds in solution based
on ionic bond strength ability[6]
. As all ions have different bond strengths, each individual compound is
retained inside a column packed with a stationary phase coated in an ion-exchange resin for a certain
amount of time (stronger ionic bonds result in longer retention times). Samples to be separated and
analyzed are injected into a continuous stream of eluent (also referred to as the mobile phase) in a
sharp plug via sample loop on an injection valve. One of the major advantages of this process is the
ability to simultaneously determine multiple ionic species at sub-ppm (parts per million) concentrations.
In addition, IC has become common in environmental analysis, being used to determine ionic
concentrations in a wide variety of samples, from sea water to polar ice to food products[7]
. Indeed, the
ability of IC to handle samples from air, soil and water accounts for the rapid growth of the IC as an
accurate tool for environmental monitoring[6]
. Ion chromatographic methods can be applied to a variety
of ionic solutes, and can separate anions (negatively charged ions, eg. Br
-
) or cations (positively charged
ions, eg. Ca+). This paper focuses on quantitative determination of anionic compounds and qualitative
discussion of environmental samples. Of particular interest is a series of prairie restoration plots which
have received experimental treatments. One plot is currently severely damaged with nitrogen
deficiency being a possible culprit.
Nitrogen is the most important nutrient to a healthily functioning prairie[8]
. Nitrate and nitrate
are the two most prevalent anionic compounds found in soil. Nitrogen is deposited into soil via
Eluent: solvent used to carry extracted ions through chromatograph
[10][11]
-
7/30/2019 Peter Regier Seminar - IC
5/37
Regier 5
atmospheric deposition or anthropological sources and doesnt adsorb with soil particles, meaning it
stays in soil until used by plants or transported by groundwater[9]
. During the growing season, nitrate is
absorbed into plant material. Unless this growth is removed from the area, such as harvesting of crops,
the nitrate concentration will mostly return to the soil through decomposing plant material[9]. During
the winter, nitrate concentrations tend to remain fairly constant due to limited use by plants.
Substantial research into analyzing ionic species in water already exists[3][4][5]
but less for
environmental investigation of ionic concentrations in soil[6]
. The purpose of this research is to develop
and test a method for analyzing nitrate and nitrite concentrations in tallgrass prairie soil samples using
Ion Chromatography. In order to accomplish this, a good operating environment must be established,
then proper calibration must be demonstrated. When these steps are complete, known and unknown
samples will be analyzed, including environmental samples from the prairie.
III. Ion-Exchange Chromatography Theory
Note: IC methods can be used for separation of cations (C+) or anions (A
-). However, for simplicity, the
following theory details anions separated on a positively charged column, as this was the primary area ofresearch. For cationic separations, all charges would be switched (C
+ A
-, A
- C
+)
The theory of Ion Exchange Chromatography is relatively simple. There is a stationary phase and
a mobile phase. The stationary phase is a packed column which is coated with positively charged ions.
A sample is injected into the column, separated out in the column and recorded by the detector. The
sample flow path through the instrument is shown in Fig. 1
For a more detailed flow path, seeAppendix K: Ion Chromatograph Flow Diagram (detailed)
EluentPum In ector
Sam le
Column Su ressor Detector
Figure 1: Flow Path
-
7/30/2019 Peter Regier Seminar - IC
6/37
Regier 6
The sample is injected from the sample loop into a constant flow of charged mobile phase
(eluent). As the anions enter the column, they are attracted to the positively charged ions on the
column and must compete for binding locations. All ions have different binding strength based on the
equilibrium constant for the exchange reaction (Kex.)[12][13]
. Ions are sent into the column in a sharp plug
and separate along the column based on binding strength. When the samples anions are flushed from
the column, they come out separated into species and pass through the detector.
Figure 2: Four steps to ion exchange within the column. The column is coated with positive charges, represented
by plus signs. The smaller, thinner minus signs represent the anions in the eluent, while the thicker, bigger minus
signs are the anions in the sample being analyzed.
+++++++++++++++++++++
+++++++++++++++++++++
Flow Direction
A B
+++++++++++++++++++++
+++++++++++++++++++++
Flow Direction
+++++++++++++++++++++
+++++++++++++++++++++
Flow Direction
D
+++++++++++++++++++++
+++++++++++++++++++++
Flow Direction
C
Fig. 2
-
7/30/2019 Peter Regier Seminar - IC
7/37
Regier 7
In Fig. 2, A is the column before injection of the sample. B is at the moment of sample injection
where the strong anion from the sample is attracted to the wall of the column. C shows the sample
anion bonded to the positively charged wall of the column after kicking the weaker eluent anion off the
bonding site. D shows the sample anion moving down the column after its bond is broken with the
cation by an eluent anion. The exchange equilibrium for the column used in this experiment isxNR3+-
OCO2H-+ Ax
- [NR3
+]x-A
x- + OCO2H-where NR3
+is the stationary phase, OCO2H
-is the mobile phase and
Ax- is the anionic compounds in the injection sample to be analyzed[14].
IV. Methods
a. Development of Operation Procedure:The initial phase of research included the development of a proper operating environment
for the Ion Chromatograph as well as clear and precise instructions for running the machine and
maintenance. During this process, a cleaned stock of glassware was assembled, procedures for
mixing of solutions were developed, several trouble-shooting sessions resolved issues with
software, hardware and interfacing and general maintenance was conducted. To document this
information for future operators, all important procedures were recorded to facilitate continued
operation of the IC. These procedures are attached asAppendices (A-H). Appendix J:
Troubleshooting provides a table which details problems that were encountered in the
development of the instrument method and the indications and steps taken to fix the issue.
b. Instrumentation:All samples were analyzed using a Metrohm (Herisau, Switzerland) 820 IC Separation Center
equipped with a Metrohm 819 IC detector for conductivity detection. Samples were placed in
-
7/30/2019 Peter Regier Seminar - IC
8/37
Regier 8
Metrohm 11 mL sample vials and automatically drawn into the column using a Metrohm 838
Advanced Sample Processor. A Metrosep A Supp 5 - 50/4.0mm anion separation column was
used for separation with a mobile-phase eluent of 3.2mM Na2CO3 / 1.0 mM NaHCO2. All
samples were run under standard operating conditions of flow rate: 0.70 mL/min, pressure: 7.0
MPa, column temperature: . The sample was injected by a 20 L injection loop mounted on
a 6-port rotary injector valve (flow path rotation detailed in Fig. 3). In the fill position (3A),
sample ran through the sample loop until the injector valve rotated to the inject position (3B).
In inject position, contents of the sample loop were injected as a sharp plug into the mobile
phase and enters the column.
c. Reagents:All water used for washing glassware, dilutions and making standards was filtered with a
Barnstead (Dubuque, Iowa) EASYpure line-fed ultra-pure water (UPW) system equipped with a
0.2 m irradiated filter producing de-ionized UPW. All water used for this experiment had a
Fig. 3A: Valve in Fill Position Fig. 3B: Valve in Inject Position
Fig. 3
-
7/30/2019 Peter Regier Seminar - IC
9/37
Regier 9
resistivity between 17.8 and 18.2 M-cm. Anion standards were mixed from Metrohm MCal
Custom Anion Mix consisting of the ions at the concentrations presented in Table 1:
Table 1
Concentration
(mg/L)
Anion
Name
Anion
Formula
10 Bromide Br-
10 Nitrate NO3-
10 Phosphate PO4-3
10 Sulphate SO4-2
5 Chloride Cl-
5 Nitrite NO2-
2 Fluoride F-
Four concentrations of standards were then diluted from the MCal Anion Mix using a 100-
10000.5% L micro-pipette and 50.000.05 mL volumetric flasks filled with UPW. All
concentrations were based on the relative concentration of nitrate in each standard in mg/L and
are presented in Table 2:
Table 2
Level Dilution Concentration
NO3-(mg/L)
1 1:50 0.2
2 1:20 0.5
3 1:10 1
4 1:5 2
d. Calibration:The four standards were run at the beginning of every sample queue and then batch-
reprocessed to achieve a best-fit calibration curve. An example of a calibration curve is
presented in Fig. 4.
As solutions are mostly water, which has a density of 1 kg/L, 1 mg/L 1 part per million
-
7/30/2019 Peter Regier Seminar - IC
10/37
Regier 10
e. Unknown TestingOnce a satisfactory multi-point calibration curve had been produced, a sample of the 1:10
dilution standard was added to a sample queue but with the level set to 0, meaning that it was
treated as a sample, not a standard (seeAppendix B: Setting up a Sample Queue for detailed
information). The resulting 1:10 dilution sample chromatogram was compared with the 1:10
dilution standard to check that the machine could accurately analyze an unknown sample.
f. Environmental Sample Preparation:All soil samples were taken using a T-sampler at the experimental tallgrass prairie restoration
plots of Kauffman Museum on the grounds of Bethel College (North Newton, KS) on October 24,
2011. Three plots were examined, labeled Plot 0, Plot 1 and Plot 2 for this experiment. Plot 0
and Plot 2 were older sections of prairie that were relatively undisturbed, while Plot 1 had been
recently seeded and was less healthy than the other two plots. To obtain each core, the T-
sampler was pushed vertically eight inches into the soil and then pulled vertically back up. For
each plot, 6 core samples were taken, each for a randomly selected location within the plot.
The cores were placed in a clean bucket and broken by hand into small aggregates, forming a
homogenous mixture[7]. Approximately five grams of this mixture were placed in a clean paper
bag and allowed to air dry for 72 hours[6]. The following is the extraction process for one
sample. This process was repeated for all analyzed samples:
Ion extraction: The sample was ground by mortar and pestle into a fine matrix and
0.2500 grams were measured out on a Ohaus (Pine Brook, NG) AV264 Analytical Balance
(capacity of 260.0000g, readability of 0.0001g) and transferred gravimetrically into a
sterile Corning (Corning, NY) 15 mL polypropylene centrifuge tube. The weigh boat was
-
7/30/2019 Peter Regier Seminar - IC
11/37
Regier 11
weighed both before and after sample transfer to give a net weight of soil transferred,
presented in Table 3:
Table 3
Sample Full Weight (g) Empty Weight (g) Soil Weight (g)
Plot 0: South 1 0.2518 0.0038 0.2480
Plot 0: South 2 0.2493 0.0125 0.2368
Plot 0: North 1 0.2488 0.0089 0.2399
Plot 0: North 2 0.2503 0.0003 0.2500
Plot 1: East 0.2504 0.0007 0.2497
Plot 1: West 0.2493 0.0025 0.2468
Plot 2: East 0.2493 0.0015 0.2478
Plot 2: West 10.2507 0.0025 0.2482
Plot 2: West 2 0.2518 0.0034 0.2484
A 10.000.04 mL volumetric flask was used to measure 10 mL of UPW into the sample
tube which was capped and agitated on a rotary shaker for 1 hour[8]
. UPW was used
instead of eluent (as suggested[8]
), because of the possibility of solute dispersion which
reduces efficiency of the method[9]
. At the end of this time period, the sample was
centrifuged for 5 minutes. Approximately 2 mL of the resulting solution were decanted
and filtered through a Whatman Uniprep 0.2 m PTFE membrane syringeless filter
device and 1 mL of this filtered solution was transferred via micropipette to a clean 10
mL volumetric flask which was filled to the line with UPW, capped and shaken. The
solution was then poured into a clean Metrohm 11 mL sample vial.
After the samples were run, they were reprocessed with the standards using the best-fit
calibration curve obtained in the Calibration section. During the reprocessing, retention times
for samples were automatically adjusted to adhere to the retention times of the calibration
-
7/30/2019 Peter Regier Seminar - IC
12/37
Regier 12
standards. Concentration values in mg/L were assigned to each sample based upon the values
obtained from the calibration curve.
-
7/30/2019 Peter Regier Seminar - IC
13/37
Regier 13
V. Results
The four anion standards were run as shown below:
1:20 Dilution
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 mi
0.0
0.5
1.0
1.5
2.0
2.5
3.0
3.5
4.0
mV
ch1
fluoride
0.1
chloride
0.
nitrite
0.2
43
brom
ide
0.5
00
nitrate
0.5
01
sulp
hate
0.5
04
phosphate
0.5
01
1:50 Dilution
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19
-0.2
0.0
0.2
0.4
0.6
0.8
1.0
1.2
mV
ch1
fluoride
0.
chloride
0
nitrite
0.1
20
brom
ide
0.1
nitrate
0.1
sulphate
0.1
81
phosphate
0.1
-
7/30/2019 Peter Regier Seminar - IC
14/37
Regier 14
1:5 Dilution
0 1 2 3 4 5 6 7 8 9 10 1 1 1 2 13 1 4 15 1 6 17 1 8 1 9 mi
0
2
4
6
8
10
12
14
16
18
mV
ch1
1
fluoride
0.4
03
chloride
1.
nitrite1.0
01
brom
ide
2.0
09
nitra
te
2.0
51
sulphate
2.0
16
phosphat
e
2.0
09
1:10 Dilution
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 mi
0
1
2
3
4
5
6
7
8
9
mV
ch1
fluoride
0.2
02
2
chloride
0.
nitrite
0.4
94
brom
ide
1.0
03
ni
trate
1.0
06
sulphate
1.0
07
phosp
hate
1.0
01
-
7/30/2019 Peter Regier Seminar - IC
15/37
Regier 15
The standard solutions were then batch-reprocessed**
and multi-point calibration curves were
created for each anion based upon the four concentration values obtained for each standard dilution
level. Fig. 4 shows an example calibration curve where each points number cooresponds to its
calibration level:
The peak area for each standard level for a specific anion is plotted against the concentration
given by the method. This data was fitted to a best-fit linear formula with a y-intercept of 0 and the
process was repeated for all anions. The correlation in this case is 0.99998 with a residual standard
deviation (RSD) of 2.494%. The results for each anion tested are presented below inTable 4:
Table 4
Anion RSD (%) Correlation
Retention
(min) Concentration (mg/L)
F- 2.494 0.99998 4.184 0.403
Cl- 6.675 0.99996 6.149 1.017
NO2- 2.655 0.99969 7.189 1.001
Br- 1.216 0.99997 8.997 2.009
NO3- 1.548 0.99996 9.977 2.051
SO4-2
2.443 0.99996 13.754 2.016PO4
-3 1.177 0.99997 17.357 2.009
MeanSD 2.6011.901 0.999930.00011
**SeeAppendix E: Batch Reprocessing for procedure
Fig. 4
-
7/30/2019 Peter Regier Seminar - IC
16/37
Regier 16
Once a standard curve was successfully developed, an unknown sample was run along with the
standard solutions. The unknown solution was a repeat of one of the standard solutions but the level
was not entered in the sample queue, making its anion concentrations unknown to the instrument. As
ambient conditions and variation of parameters in the instrument environment can significantly affect
chromatographic output, the 4-point standard curve was generated for every sample queue. The
unknown sample run was 1:10 dilution, and this is compared with the 1:10 standard run in Table 5. As
the concentrations are sub-ppm, the data is presented with units of parts per billion (ppb):
Table 5
Anion (ppb)
Sample F-
Cl-
NO2-
Br-
NO3-
SO4-2
PO4-3
1:10 Standard 190 211 544 1002 984 994 988
1:10 Sample 189 203 543 1003 988 975 996
Difference 1 8 1 1 4 19 8
Once the method was developed and an unknown sample was tested and experimentally
proved the reproducibility and accuracy of the method, environmental samples were taken. However,
the soil samples were run using water from the UPW machine that wasnt properly cleaned due to worn
out filters. The samples run with this water saturated the column with ions which were subsequently
flushed by making new solution and running many blank samples of clean UPW. At completion of the
project, the column was not completely cleaned and so repeat extraction and analysis of the soil
samples was not possible. As the eluent was more conductive due to excess ions, the two most diluted
standards run for the environmental sample queue were removed due to excessive peak interference.
The resulting sample chromatograms are presented inAppendix I: Sample Chromatograms and the
values for the calibration curve are given in Table 6:
-
7/30/2019 Peter Regier Seminar - IC
17/37
Regier 17
Table 6
Anion RSD (%) Correlation Retention (min) Concentration (mg/L)
F- 3.628 1.00000 4.184 0.405
Cl- 46.473 1.00000 5.811 1.058
NO2-
6.677 1.00000 6.805 0.975
Br- 0.149 1.00000 8.470 1.999
NO3- 1.163 1.00000 9.632 2.008
SO4-2
0.482 1.00000 12.336 2.003
PO4-3
0.912 1.00000 15.089 2.006
Using the data from Table 4 and Table 6, the differences between the standard curve concentrations
were determined and are presented in Table 7:
Table 7
Anion Actual conc. (ppb) |Actual - Table 4| (ppb) |Actual - Table 6| (ppb)
F- 400 3 5
Cl- 1000 17 58
NO2- 1000 1 25
Br- 2000 9 1
NO3-
2000 51 8SO4
-2 2000 16 3
PO4-3
2000 9 6
Finally, the nitrate and nitrite concentrations of all environmental samples are shown in Table 8:
Table 8
Sample NO2-(ppm) NO3
-(ppm)
Plot 0: South 1 0.014 0.204
Plot 0: South 2 0.016 0.107
Plot 0: North 1 0.009 0.049
Plot 0: North 2 0.008 0.008
Plot 1: East 0.012 0.209
Plot 1: West 0.013 0.098
-
7/30/2019 Peter Regier Seminar - IC
18/37
Regier 18
Plot 2: East 0.007 0.090
Plot 2: West 1 NA 0.015
Plot 2: West 2 NA 0.091
Mean 0.011 0.097
VI. Discussion
Initial results for running a standard curve are very precise. The chromatograms all have strong,
sharp peaks with limited broadening and no strong tails. Metrohm states that a value for RSD of less
than 5% and a Correlation Value of 0.999 or better is expected for a strong standard curve[15]
. The
samples provide good RSD values for each anion except chloride, with a RSD of 6.675%. Though this is
close to the target value of 5%, it must be noted that this is a possible point of concern. When observing
the chromatograms a negative peak is observed (most strongly in the weak dilutions) on the end of the
nitrite peak. A negative peak generally means that something less conductive than the eluent is running
through the detector. It is possible that some sort of contaminant in small quantities with a similar
retention time to nitrite caused this peak.
The chromatograms for the samples run are much more concerning than the chromatograms
run simply for the standard. This is due in part to the nature of environmental samples. It is relatively
easy to get strong, even peaks when the sample is a standard anion solution of certified concentration
diluted with UPW. However, environmental samples will inevitably have competing anionic and cationic
compounds and a more complex matrix filled with other compounds that can interfere with accurate
detection. In addition, the sample runs coincided with the failure of the filters on the UPW machine,
leading to even the tested standards giving poor results. The large negative peaks present on each
chromatogram are most likely the result of an eluent with higher ionic activity than desired. This
presence of extra ions increases the general conductivity of the eluent. As the conductivity is increased,
-
7/30/2019 Peter Regier Seminar - IC
19/37
Regier 19
the baseline increases as well and the bottoms of peaks are either less clear or partially cut off. Despite
these difficulties, the two strongest standard dilutions built a curve detailed in Table 6 that is acceptable
except chloride. The correlation values are all at the maximum possible value as there are only two
points on which the curve was based around, and therefore, correlation data are ignored. The resulting
RSD values show that all anions except chloride and nitrite have under 5% residual standard deviation.
The first standard multi-point calibration curve (data presented in Table 4) was assumed to be precise
and accurate based on using all four points for calibration and virtually all requirements for a decent
curve having been successfully met. Based on strong evidence against the precision of the
environmental standard method, the difference between the curves detailed inTable 4 and Table 6 and
the actual concentration value were calculated in Table 7. As the concentrations of anions are known,
the difference between the actual values and the values provided inTable 4 and Table 6 were
calculated. This provided a comparison of the two sample curve concentrations using a shared
reference point. Based upon the evidence against the strength of the two-point curve, it was used to
batch-reprocess all of the environmental samples to label peaks and assign concentrations, but the data
are considered to be inaccurate. Therefore, the data are analyzed qualitatively based on relative shape
and peak strength rather than quantitatively.
Most noticeable of all the anion peaks in the environmental samples is the chloride peak. While
it is possible that there is contamination from the problems with UPW, the chloride peaks do not seem
to be spiked in the standards. Therefore, it may be inferred that the chloride levels in the prairie plots
appear to be high. It must be noted that the field duplicate for Plot 0: North shows a dramatic
difference between samples in terms of chloride level. As it is a duplicate, it is assumed that both
samples will have the same concentration, while the chloride differs by 212 ppb. Chloride is generally
anthropologically deposited and in this case, it is quite plausible that irrigation of grass in close proximity
-
7/30/2019 Peter Regier Seminar - IC
20/37
Regier 20
to the prairie plots has led to a marked increase in chloride concentration within the plots over time.
The same possibility applies to the fluoride peaks, though they are much less pronounced than the
chloride values.
The discrepancies in Plot 0: North samples regarding chloride raise serious questions about the
reproducibility of the soil analysis method. However, in general, the duplicates appear to be reasonably
similar in terms of present peaks and their relative concentrations. Plot 0: South, Plot 0: North and Plot
2: West samples all have relatively similar concentration patterns. It is also noted that Plot 0 has no
sulfate peaks recognized while the other two plots samples all have sulfate peaks. As Plot 0 and Plot 2
are supposed to be similar, this is surprising, though not alarming as the concentration of sulfate is
uniformly low across all samples.
However, of all the anions, nitrogenous compounds are of the most interest for this study.
Nitrate is present in all samples with an average concentration of 97 ppb. Nitrite is present in all
samples except Plot 2: West and its field duplicate and had an average of 11 ppb. As nitrite is located at
the beginning of a negative peak on all chromatograms, its likely that the contamination causing the
negative spike also interfered with the resolution of the nitrite peak. However, nitrate peaks are
consistently taller and clearer than nitrite, even at the beginning of the peak where nitrite would be
unaffected by the negative peak on its tail. This leads to the tentative conclusion that nitrate is either
more common in higher concentrations in the prairie soil, or that the extraction method or instrumental
procedure retain or detect nitrate better than nitrite. As the original concern was that nitrogen species
may be low in the experimental Plot 1, the data are compared between plots. It is observed that, when
the nitrate and nitrite values are summed for each plot and then divided by the number of plots, the
experimental plot has the highest average nitrogenous species concentration. This is opposite of the
initial prediction that nitrogen deficiency is causing ecological problems in the plot. However, as the
-
7/30/2019 Peter Regier Seminar - IC
21/37
Regier 21
concentrations of both compounds are so small, so varied, and the operating state of the machine is in
question, these observations do not hold statistical significance.
VII. Conclusion
The development of the method was successful as far as it was tested. A standard curve with
strong correlation was created, demonstrating an ability to accurately get reproducible results.
Furthermore, the ability to identify an unlabeled standard via a batch reprocess proves that the curve
can be used to accurately calculate ionic concentrations in an unknown. Soil samples were run with
limited success due to interference from poorly filtered UPW, but the developed extraction method is
ready to be tested. Using the given operating protocols and research, accurate analysis of
environmental samples using Ion Chromatography should be quite easily reached with limited further
study, possibly leading to data that can be used to diagnose problems in the Kauffman prairie plots.
VIII. Acknowledgments
Dr. Gary Histand
Dr. Richard Zerger
Derrick Law
Carrie Shultz
Martin Olson
Matthew Carda
Charles Mayer
-
7/30/2019 Peter Regier Seminar - IC
22/37
Regier 22
IX. References
1. P.G. Risser, W.J. Parton, Ecosystem Analysis of the Tallgrass Prairie: Nitrogen Cycle, Ecology, 63,1342-1351 (1982).
2. P.M. Vitousek, R.W. Howarth, Nitrogen limitation on land and in the sea: how can it occur?Biogeochem. 13, 87115 (1991).
3. A. Eaton, M. Carter, A. Fitchett, J. Oppenheimer, M. Bollinger, J. Sepikas, Comparability of IonChromatography and Conventional Methods for Drinking Water Analysis in Proc. AWWA
Water Qual. Tech. Conf., 175-188, (1984).
4. USEPA Method 300.1, The Determination of Inorganic Anions in Water by Ion Chromatography,USEPA, Cincinnati, OH, (1997).
5. R. Michalski, I. Kurzyca, 'Determination of Nitrogen Species (Nitrate, Nitrite and Ammonia Ions)in Environmental Samples by Ion Chromatography', Polish Journal of Environmental Studies,
15, 5-18 (2006)
6. P.E. Jackson, Ion Chromatography in Environmental Analysis in Encyclopedia of AnalyticalChemistry, ed. R.A. Meyers, John Wiley & Sons Ltd, Chichester, (2000).
7. B. Lpez-Ruiz, Advances in the determination of inorganic anions by ion chromatography,Journal of chromatography, A, 881-1, 607 (2000)
8. P.G. Risser, W.J. Parton, Ecosystem Analysis of the Tallgrass Prairie: Nitrogen Cycle, EcologicalSociety of America, 63-5, 1342-1351 (1982).
9. A. Wild, Soils and the Environment, Cambridge University Press, New York (1993).10.H P.E. Jackson, Ion Chromatography in Environmental Analaysis in Encyclopedia of Analytical
Chemistry, ed. R.A. Meyers, John Wiley & Sons Ltd, Chichester, (2000).
11.Chemistry Dictionary, Definition of eluent, ChemiCool, viewed November 7, 2011, (2011).
12. J.S. Fritz 1999, Analytical Solid-Phase Extraction, Wiley-VCH, New York.13.Metrohm, IC Theory, Metrohm IC Tutorials, Herisau, Switzerland, CD-ROM, (2011).14.D. Skoog, F.J. Holler, S.R. Crouch, Principles of Instrumental Analysis, 6th edn., Brooks Cole
(2006).
15.Metrohm, Multipoint Calibration, Metrohm IC Tutorials, Herisau, Switzerland, CD-ROM,(2011).
16.Metrohm, IC Flow Path, Metrohm IC Tutorials, Herisau, Switzerland, CD-ROM, (2011).
http://www.chemicool.com/definition/eluent.htmlhttp://www.chemicool.com/definition/eluent.html -
7/30/2019 Peter Regier Seminar - IC
23/37
Regier 23
X. Appendices:
Appendix A: Startup and Run the IC
Appendix B: Setting up a Sample Queue
Appendix C: Setting up Standard Concentrations
Appendix D: Peak Labeling
Appendix E: Batch Reprocessing
Appendix F: Preparation of Anion Standard
Appendix G: Preparation of Sulfuric acid standard
Appendix H: Replenishing Ultrapure H2O
Appendix I: Sample Chromatograms
Appendix J: Troubleshooting
Appendix KIon Chromatograph Flow Diagram (detailed)
-
7/30/2019 Peter Regier Seminar - IC
24/37
Regier 24
Appendix A: Startup and Run the IC
1. Double-click IC Net 2.3.exe on desktop2. Enter user-name and password (located at the bottom of the monitor) into the dialog box. Click
Log In or hit Return.
3. Right-Click on the graphic and select Z to make it disappear4. Open System: File > Open > System Select the Bethel folder, then either anions.smt for
the anion column, or cations.smt for the cation column. It is important to select the right one,
as it controls which parts of the machine operate. IMPORTANT. A box with cute little drawings
of the instruments parts will show up. This is your control panel.
Note: if the control panel window is closed, it is easy to bring it back by simply clicking
right button on the toolbar. It will be the farthest right of three buttons showing three
green boxes.
5. Start up the instrumenta. Turn the power switch on behind the instrument. This will turn on all components
EXCEPT the 830 IC interface. Check to make sure all component lights come on.
b. Turn on the power switch located at the back of the 830 IC Interface and check the lighton the front. If it is blinking rapidly, turn 830 off and then turn it back on. It should blink
twice slowly then stay on if the interface is talking to the computer.
6. Interface the software and instrument: in the control panel, click Control > Connect ToWorkplace. A dialog box in the lower left corner of the screen will show On-line [830 ICInterface] in gray. If it does not, check that the light on the 830 component isnt blinking.
7. Start hardware: in the control panel, click Control > Connect Hardware (Measure Baseline). Adialog box in the lower left corner of the screen will show the status of each component. Red
means a component is not connected, yellow means it is connecting, and gray means its
connected. Wait until everything is ready (gray) before proceeding to the next step.
8. Prepare peristaltic pumps: before starting the pumps, the clamps for each need to be attached.The two lines that need to be tight for measuring the baseline are the acid and water lines.
Thee pump and lines are located on the front right of the 833 IC Liquid Handling Unit. Make
sure that each line (yellow and black) have their colored tabs in the cutouts on the left-hand side
of the plastic clamp and are lined up. Then, make sure that the lines on the other side of the
pump are in between the two arms for each clamp. These steps are important to make sure
that the lines arent damaged during the tightening process. Carefully slip the arms on the right-
hand side down past the metal bar until the indention lines up and snaps into place. Next,
-
7/30/2019 Peter Regier Seminar - IC
25/37
Regier 25
tighten the clamps down using the circular gears on top of each one. Make them hand-tight but
DO NOT OVERTIGHTEN. It would be a good idea to have an instructor demonstrate proper
tightening technique before you do this. The end result should be that the lines each run in the
middle of their respective clamps (not pinched anywhere) and are tightened against the rollers
of the peristaltic pump.
Note: To make sure that the clamps are tightened and that pumps are drawing correctly,
watch for any bubbles moving through the lines. If there is a bubble that is sitting in the
line and not moving, no liquid is moving through. Additionally, check the waste lines to
make sure that they are dripping constantly.
9. Measure baseline: in the control panel, click Control > Connect Hardware (Measure Baseline)again. This time, a chromatogram window will pop up and the 833 pump will engage.
Additionally, you will hear a strange sucking sound from inside the machine. Thats okay. Its
supposed to happen. A status window will pop up displaying the conductivity, pump pressure
and column temperature. The column temperature will be red until it is close to 35 degrees. No
samples should be run until the column reaches prime operating temperature. Let the baseline
run for about half an hour remembering to step the injector every 10 minutes or so (see next
step)
10.Check acid and water lines with litmus paper. Acid should produce red on blue and water lineshould produce nothing on either.
11.Step injector (as needed): in the control panel, double-click on the 820 IC Separation Centericon. This will bring up a dialog box. Under the Manual tab, click STEP. You will here the
injector valve rotate inside the machine and this is supposed to be a fairly strong peak
associated with a STEP event. This step should only be done during the measuring of the
baseline. DO NOT do this any time you are running a sample as you will RUIN IT FOREVER.
12.While your baseline is running, make sure to take off all lids for any samples that need to be runand prepare, label and place your samples in the autosampler. You may need to move the
autosampler manually to get access to the rinse bottles or slots you need. To do this:
a. In the control panel, double-click the 838 Advanced Sample Processorb. In the dialog box the pops up, select the Autosampler tabc. In the drop-down box in the Move column, select the position that you wish to move tod. Click Start
13.Once your baseline has leveled out, around 16-20 mV, check your Method (see Setting up aMethod), load your Sample Queue (see Setting up a Sample Queue), tighten the peristaltic
pump on the auto-sampler (see Step 8 for instructions) and run your samples.
Note: make sure that this one is tight or it wont properly. Check the sample waste line
to make sure sample is running
-
7/30/2019 Peter Regier Seminar - IC
26/37
Regier 26
14.To shut down the machine:a. In control panel, click Control > Shut down Hardware. The pump should stop runningb. Click Control > Disconnect from Workplacec. Close all open windows in ICNet and the program, saving if promptedd. Turn off IC Detector unite. Turn off power stripf. Release all peristaltic clamps. This is IMPORTANT to keep the tubing in good shape
without indentions
Appendix B: Setting up a Sample Queue
1. Refer to Startup and Run the ICfor proper instrument setup.2. Once these steps are completed, go to File > Open> Sample Queue3. Navigate to the current folder for Sample Queues (currently C:\Program Files\Metrohm\IC Net
2.3\IC Net\Systems\Bethel)
4. Enter the name of your sample queue, leaving the .que5. In the dialog box that pops up, click edit to open the Queue Editor6. Under System for all samples, type the name of the system being used (currently
anions.smt)
7. Under Ident type a short name for each sample8. Under Vial enter the location of the sample vial on the autosampler rack9. All other categories EXCEPTLevel can be copied down for all samples10.For level, all samples will be 0 except for standards being used to build a standard curve. For 4
standards, the lowest concentration is 1, the highest is 4. If you are not sure if your standardsare correct in the method, check Setting up Standard Concentrations and follow the
instructions for adjusting the concentrations of each level
Appendix C: Setting up Standard Concentrations
1. Refer to Startup and Run the ICfor proper instrument setup.2. Once these steps are completed, go to File > Open> Chromatogram3. Choose a chromatogram that was created using the method whose standard concentrations you
want to change
4. With the chromatogram window active, go to Method > Calibration > Concentrations which willopen a Concentrations dialog box
5. For any levels that you wish to change, double-click each cell and change the concentration6. If all levels are to be changed, simply click a cell within a level and click the Delete button.
Then, click add and add a blank level column
7. If prompted to save, make sure to save every change. Add appropriate comments as prompted.
-
7/30/2019 Peter Regier Seminar - IC
27/37
Regier 27
Appendix D: Peak Labeling
1. Refer to Startup and Run the ICfor proper instrument setup.2. Once these steps are completed, go to File > Open> Chromatogram3. Choose the chromatogram that you wish to edit and double-click8. With the chromatogram window active, go to Method > Calibration > Components which will
open a table below the chromatogram.
4. All anions will be listed but may be assigned to the wrong peak, or no peak at all. Leave peakassignments that are correct.
5. Under the Peak column, double-click the anion row that you wish to change. Enter the correctpeak number and press Return.
6. Click the OK button and exit the chromatogram7. If prompted to save, make sure to save every change. Add appropriate comments as prompted.
Appendix E: Batch Reprocessing
1. Refer to Startup and Run the ICfor proper instrument setup.2. Refer to Setting up Standard Concentrations and Peak Labeling to make sure that all standard
chromatograms to be processed have been properly labeled and are using the same standard
concentration levels for calibration (sample chromatograms will be automatically labeled during
the batch reprocessing)
3. Once these steps are completed, go to File > Open> Chromatogram4. Hold and click on each chromatogram to be included in the batch reprocess, including all
standards and samples and click To Batch on the right-hand side of the dialog box
5. Give the batch a name and press OK6. Click the Edit sample table button to view the chromatograms selected. Make sure that all
samples to be reprocessed are included and that the levels are correct (1-4 for standards, 0 for
all samples)
7. Select Reintegrate (Recalibrate is automatically selected as well)8. Press the Reprocess button9. If prompted to save, make sure to save every change. Add appropriate comments as prompted.10.To double-check, select a chromatogram and check the calibration table and that components
are properly labeled
-
7/30/2019 Peter Regier Seminar - IC
28/37
Regier 28
Appendix F: Preparation of Anion Standard
Empty eluent bottle of any remaining eluent (but do not rinse)
Obtain 1 L clean volumetric flask labeled IC Eluent only (located near Ultrapure machine)
Remove parafilm, dump out contents
Rinse 3 times with UltraPure (UP), shaking vigorously when dumpingFill the volumetric half full of UP.
Make sure that the eluent snip (located on the south shelf in the instrument room) has all liquid
at bottom. If not, try inverting or flicking the tube, as complete volumetric transfer is key.
Cut off the top of the tube and dump contents into the volumetric flask
Rinse the tube three times and dump this into the flask as well
Fill Volumetric to line and cover with parafilm
Invert 10 times
Dump into open eluent bottle, and cap the bottle
Rinse the flask out 3 times, swirling it, with UP
Fill to top with UP and cap with parafilm
Caution:
Do NOT touch any part of the rim or inside of the flask with anything other than solution and
clean parafilm. Contamination from your hands, gloves or parafilm can effect the eluent!
This is a very dilute, very specific solution. Take great care.
Dont rinse the eluent bottle. The extra water will mess up the concentration
Dont let the tube inside touch anything, make sure it is resting on Kimwipes, not the machine
Make sure that the UP machine reads between 17.8-18.2 M-cm. If it goes below this
conductivity, give it some time to get back into this zone.ALWAYS use Ultrapure. NEVER use distilled or tap water, as they will contaminate the lines.
Appendix G: Preparation of Sulfuric acid standard
Obtain 1L clean volumetric flask labeled IC Sulfuric Acid (located near Ultrapure machine)
Remove parafilm, dump out contents
Rinse 3 times with UltraPure (UP), shaking vigorously when dumping
Measure out 5.55 mL (5.5 is good enough) of 18M Sulfuric Acid into graduated cylinder
Add sulfuric acid to empty flask
SLOWLY add a small amount of water and allow the flask to sit until it cools down
Repeat the above step with increasing amounts of water, making sure the flask stays cool
Fill volumetric to line and cover with parafilm
Invert 10 times
Dump into open acid bottle and cap bottle
-
7/30/2019 Peter Regier Seminar - IC
29/37
Regier 29
Rinse the flask out 3 times, swirling it, with UP
Fill to top with UP and cap with parafilm
Caution:
Do NOT touch any part of the rim or inside of the flask with anything other than solution and
clean parafilm.
Dont let the tube inside touch anything. Make sure it is resting on Kimwipes, not the machine
Make sure that the UP machine reads between 17.8-18.2 M-cm. If it goes below this
conductivity, give it some time to get back into this zone.
ALWAYS use Ultrapure. NEVER use distilled or tap water, as they will contaminate the lines.
Use caution, Sulfuric acid is nasty stuff
Appendix H: Replenishing Ultrapure H2O
Fill bottle labeled Ultrapure H2O at Millipore machine
Caution:
Do NOT touch any part of the rim or inside of the flask with anything other than solution and
clean parafilm.
Dont let the tube inside touch anything. Make sure it is resting on Kimwipes, not the machine
Make sure that the UP machine reads between 17.8-18.2 M-cm. If it goes below this
conductivity, give it some time to get back into this zone.
ALWAYS use Ultrapure. NEVER use distilled or tap water, as they will contaminate the lines.
-
7/30/2019 Peter Regier Seminar - IC
30/37
Regier 30
Appendix I: Sample Chromatograms
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 mi
0
2
4
6
8
10
12
14
16
mV
ch1
1
fluoride
0.
chloride
1.
nitrite
0.9
75
5
brom
ide
1.
nitrate
2.0
sulphate
2.0
03
phosphat
e
2
1:5 Standard Dilution
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 16 17 18 19 min
-1
0
1
2
3
4
5
6
7
8
mV
ch1
1
fluoride
0.1
chloride
0.
211
4
nitrite
0.5
44
6
bromide
1.
nitrat
e
0.9
sulphate
0.9
94
phosph
ate
1:10 Standard Dilution
-
7/30/2019 Peter Regier Seminar - IC
31/37
Regier 31
Plot 0: South 2
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 min
-1
0
1
2
3
4
5
mV
ch1
1 2flu
oride
0.0
23
chlor
ide
0.
5 ni
trite
0.0
16
78 br
om
ide
0.0
23
nitrate
0.1
07 p
hosphate
0.4
32
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 min
0
5
10
15
20
25
mV
ch1
1
fluoride
0.0
84
3
chl
oride
1.
nitrite
0.0
14
6 7 bro
m
ide
0.0
09
nitrate
0.2
04
10p
hosphate
0.2
84
Plot 0: South 1
-
7/30/2019 Peter Regier Seminar - IC
32/37
Regier 32
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 14 1 5 1 6 17 18 19 min
-1
0
1
2
3
4
mV
ch1
1 2fluoride
0.0
12
chloride
0.0
87
5 nitrite
0.0
08
7
8 brom
ide
0.0
17
10 nitrate
0.0
08
phosphate
0
13
Plot 0: North 2
Plot 0: North 1
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 14 15 16 17 1 8 19 min
-1
0
1
2
3
4
5
mV
ch1
1 2fl
uoride
0.0
15
4
chlo
ride
0.
6 ni
trite
0.0
09
89 br
om
ide
0.0
19
n
itrate
0.
049
pho
sphate
13
-
7/30/2019 Peter Regier Seminar - IC
33/37
Regier 33
Plot 1: West
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 min
0
2
4
6
8
10
12
14
16
mV
ch1
fluoride
0.0
89
chlor
ide
1.
nitrit
e
0.0
13
4 brom
ide
0.0
26
nitrate
0.0
98
sulph
ate
0.0
92
ph
osphate
0.2
72
Plot 1: East
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 14 1 5 1 6 17 1 8 19 min
-1
0
1
2
3
4
5
6
7
8
mV
ch1
1 2 fluo
ride
0.0
20
chloride
0.
nitrite
0.0
12
6 brom
ide
0.0
25
nitrate
0.2
09
sulphate
0.0
88
phosphate
0.5
55
11
-
7/30/2019 Peter Regier Seminar - IC
34/37
Regier 34
1:10 Sample
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 min
-1
0
1
2
3
4
5
6
7
8
mV
ch1
1
fluoride
0.1
3
chloride
0.2
03
5
nitrite
0.5
43
7
brom
id
e
1.
nitrate
0.9
sulphate
0.9
75
phospha
te
0
Plot 2: East
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 14 1 5 1 6 17 1 8 19 min
0
2
4
6
8
10
12
mV
ch1
fluoride
0.0
85
chloride
0.
34
brom
ide
0.0
07
nit
rate
0.0
90
7 sulp
hate
0.0
98
p
hosphate
0.2
00
-
7/30/2019 Peter Regier Seminar - IC
35/37
Regier 35
Plot 2: West 2
0 1 2 3 4 5 6 7 8 9 10 1 1 12 1 3 1 4 15 1 6 17 1 8 19 min
0
5
10
15
20
mV
ch1
f
luoride
0.0
84
c
hloride
1.
3 4 brom
ide
0.0
13
6nitrate
0.0
91
sulp
hate
0.0
44
phosphate
0.4
15
10
Plot 2: West 1
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 min
0
5
10
15
20
mV
ch1
1
fl
uoride
0.0
80
chlorid
e
1.
4 5 brom
ide
0.0
12
nitra
te
0.0
15
sulph
ate
0.0
99
ph
osphate
0.3
40
10
-
7/30/2019 Peter Regier Seminar - IC
36/37
Regier 36
Appendix J - Troubleshooting
Problem Diagnosis Action Result
Peaks slipping off the
end of the
Chromatogram
Bad Eluent Mixed New Eluent Shorter Retention
Rates
Peaks unclear/blend in
with baseline
Need more
sensitivity Full Scale
for 819 Detector
Change Control Panel >
819 IC Detector > Method
Parameter > Full Scale
Stronger peaks
Baseline decreases at
beginning of
chromatogram
Acid and Water lines
are switched
Test with litmus. Blue
litmus should turn red for
acid
No acid interfering
with the beginning of a
chromatogram
Weak peaks on lower
dilutions
Need stronger
standards
Mixed new standards with
higher concentration
Stronger, clearer peaks
but more
contamination from
residual ions left in
column
-
7/30/2019 Peter Regier Seminar - IC
37/37
Appendix K Ion Chromatograph Flow Diagram (detailedi)
iImage from IC Flow Path
[16]